Thermo 2230XP Silica Analyzer
Thermo 2230XP Silica Analyzer
Thermo 2230XP Silica Analyzer
Now
!
Available
• Boiler water
2 Engineering Specifications
1. The silica analyzer measures silica concentrations using a 10. Analyzer has a heated sample cell with sample
Heteropoly-molybdenum blue colorimetric method. presence detection.
2. Measurement is semi continuous and is <15 minutes. 11. Display is a digital graphics LCD with backlight.
3. The range of measurement is between 0 and 5000 ppb 12. Sample delivery requirement is between 50 and 100 mL/min.
or µg/L.
13. Analyzer is capable of grab sample analysis.
4. The detection limit is ±.5 ppb or 5 % of reading, which ever is
14. Analyzer can be wall or panel mounted.
greater, in the range of 0 – 300 ppb and 10 % of range from
300 – 5000 ppb. 15. Analyzer fluidics cabinet is constructed to NEMA 4X/IP65.
Enclosure is constructed of fiberglass with clear poly windows.
5. Repeatability is <±2 % or ±.5 ppb, whichever is greater,
in the range of 0 – 500 ppb and <±5 % or range from 16. Power requirement is 100 – 240 V AC, 100 W, 50/60 Hz,
500 – 5000 ppb. auto-detection.
6. Analyzer is capable of auto ranging and auto calibration. 17. Analyzer has cTUVus approvals to meet UL61010-1 and CSA
C22.2 No. 61010-1 certifications and FCC Class A requirement.
7. Analyzer takes blank samples before every measurement
and compares and corrects the measurement against the 18. Analyzer shall meet EC Directives 2006/95/EC and
blank sample. 2004/108/EC.
8. Analyzer provides 2 isolated 4/20 mA outputs and 19. Analyzer is aware of sample availably.
4 programmable alarm relays. Relays are rated for
5 A @ 240 V AC.20 INCHES
ALLOW
(510 mm) CLEARANCE
9. Analyzer has an hourly and daily maintenance mode that will
FOR DOOR SWING
move solutions and purge lines to keep the analyzer ready while
offline for up to 30 days.
406mm (16")
304.8mm (12.0")
162mm (6.4")
70mm(2.75 ")
POWER IN
I/O SIGNALS
674.6mm (26.5")
698mm (27.5")
AIR PURGE
SAM PL
E
STAND ARD
330.3mm (13.0 ")
CAUT
I ON A ir p u
r g
e o f e
n c lo
s u re is r e
c m
o me n d
e d .
Repeatability Error: Less than ±2 % of reading or ±0.5 ppb, whichever is greater from 0 to 300 ppb.
Less than ±5 % of reading from 300 ppb to 5000 ppb.
Limit of Detection: 0.5 ppb
Data and Control Current Loops: Two 0/4-20 mA – Direct or Reverse Acting (Isolated). Maximum 900 ohm load.
FCC: Class A
Product Ordering Information
223020 200 ppb Calibration Standard, 0.5 L 2230XPTK Main Tubing Kit – for 2230XP
223030 1000 ppb Calibration Standard, 0.5 L 2230BC Empty Auto Cal Bottle (Validation Sample Bottle)
2230SF Fittings, Sample In, Sample Out and Drain
2230RC Reagent Bottle Cap – includes fittings
2230SC Cal Bottle Cap – includes fittings
2230SAC Sample Chamber