346105223-Swelab-Alfa-Manual-1504154-Apr-2006.en.fr
346105223-Swelab-Alfa-Manual-1504154-Apr-2006.en.fr
346105223-Swelab-Alfa-Manual-1504154-Apr-2006.en.fr
PRÉFACE ....................................................................................................................................................... 3
introduction ............................................................................................................................................. 3
1
SECTION 8: CPENCHÉ, MNTRETIEN & TRANSPORT ................................................................................ 53
Aperçu de la section ................................................................................................................................ 53
8.1 Nettoyage quotidien ........................................................................................................................... 53
8.2 Nettoyage mensuel............................................................................................................................. 54
8.3 Six (6) Nettoyage du mois ................................................................................................................... 55
8.4 Entretien de l'instrument ..................................................................................................................... 56
8,5 Re-localisation d'instrument (au laboratoire) ........................................................................................... 56
8.6 Transport à court terme (<12H) ............................................................................................................ 57
8.7 Re-emballage et le transport à long terme (> 12H) ................................................................................... 57
8,8 arrêt et stockage permanent ................................................................................................................. 58
8.9 Informations sur l'élimination .............................................................................................................. 59
jeNDEX ........................................................................................................................................................ 84
2
Préface
introduction
Numéro de série
Figure 1.1
Une version de logiciel La version du logiciel est affichée lors du démarrage de l'instrument.
Figure 1.2
3
Documentation Une documentation supplémentaire est disponible auprès de votre distributeur
supplémentaire agréé.
Documentation supplémentaire actuelle est ci-dessous:
Service manuel
Hématologie de base
fiches produit
Distributeur
Les distributeurs sont inscrites à la cote http://www.boule.se
détails
garantie Le service doit être effectuée par Medical AB Boulé (ci-après dénommé
limites Boule) ou par le personnel de service autorisé par Boulé.
Utilisez uniquement des pièces de rechange d'origine et réactifs, autorisé
Boulé contrôles sanguins,
calibrateurs et produits de nettoyage. (Si ces produits sont substitués, il peut
annuler
garantie)
Les opérateurs et les superviseurs de laboratoire sont responsables que les
produits sont Boule
utilisé et entretenu conformément aux modes opératoires décrits dans les
manuels,
inserts de contrôle et des bulletins techniques.
Limitations Chaque système est testé à l'aide Boulé réactifs recommandés, les
de garantie contrôles sanguins, calibreurs et produits de nettoyage. Toutes les
en demandes de performance sont générés dans le cadre de ce système
profondeur
complet.
produits Boule ne font pas des diagnostics sur les patients. Boule entend
ses produits de diagnostic (systèmes, logiciels et matériels) à utiliser pour
recueillir hématologique du patient. Ces données, conjointement avec d'autres
des données informations de diagnostic et l'évaluation de l'état du patient, peut être
reflétant utilisé par un clinicien formé pour établir le diagnostic d'un patient et de
l'état définir le traitement clinique.
5
1.2 Consignes de sécurité
1.3 Biohazards
Manipulation Toujours porter des gants et des lunettes. Suivez les réglementations locales.
de matières Manipuler des échantillons avec grand soin. Rapport d'incident
infectieuses
conformément aux réglementations locales.
Ne touchez pas le liquide des déchets lors de l'élimination des déchets.
Si le sang entre en contact avec les yeux ou coupe ouverte, laver
abondamment avec de l'eau.
action obligatoire
Si le liquide des déchets est par inadvertance touché, laver la zone
affectée avec une solution désinfectante d'abord et suivre avec du
savon.
En cas de S'il y a des signes évidents de dysfonctionnement tels que des fuites de
urgence fumée ou de liquide hors de l'appareil, procédez comme suit:
Étape action
1 Débranchez l'alimentation principale immédiatement en tirant le
cordon de l'alimentation principale.
2 Contactez votre distributeur agréé.
Panneaux de signalisation Les signes d'avertissement suivants dans le manuel sont utilisés
pour identifier les dangers possibles et de faire appel à l'attention de
l'opérateur à cette condition.
Une
Signe fonction
La description Les signes placés sur l'instrument définissent des zones qui nécessitent une
attention particulière ou des zones qui contiennent danger. Voir IVD Table
des symboles à la page 9.
8
Figure 1.7 IVD Symbole Tableau
9
Section 2: Installation
Aperçu de la section
introduction Cette section décrit comment déballer et installer l'instrument Swelab Alfa.
Assujettir Sujet
Déballage dix
Contraintes d'installation 11
Connections electriques 12
Source de courant 12
Connexion de l'imprimante 13
Connexion, Changement et réactifs Amorçage 14
2.1 Déballer
La description L'appareil est emballé dans une boîte de protection spécialement conçu.
Inclus Instrument
Matériel Manuel d'utilisation
Tube de déchets
Capteur de niveau de réactif et de bouchons de réactifs pour diluant
isotonique (Diluant)
Capteur de niveau réactif et bouchons de réactifs pour réactif hémolyse (Lyse)
Cordon d'alimentation
Formulaire d'installation
Lecteur externe code-barres
Optionnel Imprimante
Matériel Papier d'imprimante
kit MCI
Clavier externe
Réactifs, contrôles sanguins, calibreurs et produits de nettoyage
dix
2.2 Contraintes d'installation
Installation
/ Placement L'instrument doit être placé dans un environnement de laboratoire selon les
d'exploitatio directives ci-dessous:
n Placez l'instrument sur une surface horizontale propre.
Évitez l'exposition au soleil.
Assurez-vous que l'appareil dispose d'un accès à une bonne ventilation.
L'instrument doit avoir au moins 5 cm (2 pouces) de l'air au-dessus.
Placer l'arrière de l'instrument de sorte qu'il a au moins 10 cm (4
pouces) d'espace libre derrière elle.
5 cm
10 cm
Figure 2.1
1
2
3
4
7
6
Figure 2.2
2.4 Alimentation
environneme L'alimentation principale est située en interne et conçu pour être utilisé
nt principal à l'intérieur. L'alimentation est sans danger pour la tension transitoire
d'alimentati telle que définie dans la norme CEI 801-4.
on
Des lignes Les directives sont données dans le manuel d'entretien «dispositifs auxiliaires
directrices d'installation »
section. Contactez votre distributeur agréé dans un tel cas.
Dans le cas d'une perte de puissance soudaine il n'y aura pas de dommages
Puissance causés à la
instrument. constantes d'étalonnage et d'autres paramètres nécessaires à
interruptions l'exploitation
sont protégés contre la perte principale
d'approvisionnement.
13
2,5 Connexion de l'imprimante (A continué)
Prise en charge DPU 411/2 et 414 DPU (fourni par un accessoire Boule en option).
Suivez les instructions dans le manuel de l'utilisateur de l'imprimante à
imprimantes installer.
La Les réactifs pour l'instrument sont livrés dans des boîtes de cube
description formés avec des bouchons en plastique.
réactif de liaison Cette section décrit comment connecter les conteneurs de réactifs pour
conteneurs
une utilisation.
Étap
e Relier
1 Le capteur de niveau de réactif Lyse (jaune) et le capteur électronique à
l'analyseur.
2 Le capteur de niveau de diluant (rouge) et le capteur électronique à
l'analyseur.
Figure 2.3
Suite à la page suivante
14
2.6 Connexion, réactifs et changement Amorçage (A continué)
Étape Insérer
Les capteurs de niveau de réactif dans les récipients de réactif
3 correspondant.
Figure 2.4
système de
remplissage
Pour le remplissage initial de l'analyseur, branchez l'analyseur et activer /
désactiver l'interrupteur sur ON.
Appuyez sur la touche [EXIT] lors de l'affichage de l'invite de
remplissage, et suivez les instructions ci-dessous pour remplir l'analyseur.
Étap
e action
Sélectionnez l'onglet
1 MENU.
2 Appuyez sur [REACTIFS SETUP] puis appuyez sur [ENTRER LE
REACTIFS].
Numériser des codes-barres sur les conteneurs de réactifs, lorsque tous
3 les codes-barres sont entrées
un écran affiche les codes-barres que réactifs ont été acceptés.
Pour exécuter un cycle premier onglet, sélectionnez MENU puis appuyez sur
[SYSTEM PRIME].
16
Section 3: Présentation générale
Aperçu de la section
introduction Cette section contient des informations générales sur l'instrument et en option
accessoires.
7
6
4
2 3 8
Figure 3.1
Partie Une fonction
LCD à écran tactile, monochrome ou couleur, avec incorporé
1. Afficher clavier et pavé numérique.
2. Le total aiguille artérielle Aspirats sang total.
aiguille / distributeur de Aspirats échantillons pré-dilués en rendant superflue
3. pré-dilué diluant.
Micro capillaire d'entrée permet à l'utilisateur d'analyser 20 ul de
4. MCI (facultatif) sang.
Imprime les exemples de résultats. (Non représenté, le modèle
5. Imprimante (en option) dépend utilisateur)
lecteur de code à barres permet à l'utilisateur d'entrer rapidement
le patient, le contrôle et
6. code à barre identifications de paquets de réactifs, et utilisent le
programme QC.
7. Mixer (facultatif) mélanges d'échantillons de façon uniforme.
8. Échantillonneur (facultatif) Permet échantillons consécutifs à analyser
automatiquement.
Analyse des échantillons avec diminution du risque de
9. Cap Piercer (en option) contact avec le sang.
17
Structure 3.2 Menu
Q/C Menu
View Con/Cal >
View Xb Stats >
Enter Con/Cal > Reagent Barcode Input
View Assays >
Exit >
Exit <
Inactivate Reagent
Yes >
No >
18
3.2 Menu System (Continued)
Calibration
Whole Blood >
Predilute >
Capillary Device >
Closed Tube Device >
Calibration Log >
Exit <
Maintenance
Prime System
Clean Orifice
Fill System ¤
Clean Cycle >
Empty System
Clot Prevention ¤
Exit <
Service Menu 2
Service Menu Serial no
Serial no Firmware
Firmware Blood Detector >
Clot Removal > Level Detectors >
Noise test > Beaker Detector >
PTest > Reagent Detector >
Advanced Pump & Valve > HGB >
Calibration > Instrument Log > Shear Valve >
Maintenance > Service Setup 2 > Needle (if applicable) >
Service > Exit < Exit <
Setup >
Exit <
19
3.3 System Flow
Description This section contains the system flow concerning standby and cleaning cycles.
After 15 minutes*
Screen Saver
Mode**
Press anywhere on
screen to view screen Press [EXIT]
20
3.4 Sample Volume, Throughput, and Parameters
Leukocyte parameters 20 16 10
WBC Total White Blood Cell Count Yes Yes Yes
LYM% Lymphocytes percentage Yes Yes No
LYM# Lymphocytes (absolute) Yes Yes No
MID% Mid Cell Population percentage Yes Yes No
MID# Mid Cell Population (absolute) Yes Yes No
GRAN% Granulocytes percentage Yes Yes No
GRAN# Granulocytes (absolute) Yes Yes No
Erythrocyte parameters 20 16 10
RBC Total Red Blood Cell Count Yes Yes Yes
HGB Hemoglobin Concentration Yes Yes Yes
HCT Hematocrit Yes Yes Yes
MCV Mean Cell Volume of RBCs Yes Yes Yes
MCH Mean Cell Hemoglobin Yes Yes Yes
Mean Cell Hemoglobin
MCHC Concentration Yes Yes Yes
Red Blood Cells distribution
RDW% width percentage Yes Yes Yes
Red Blood Cells distribution
RDWa width (absolute) Yes No No
Thrombocyte parameters 20 16 10
PLT Total Platelet Count Yes Yes Yes
MPV Mean Platelet Volume Yes Yes Yes
PDW Platelet Distribution Width Yes No No
PCT Platelet Crit Yes No No
Large Platelet Concentration
LPCR Ratio Yes No No
21
Section 4: Instrument Setup
Section Overview
Introduction This section covers the initial configuration needed to customize the
instrument settings.
22
4.2 Initial Setup
Setting up
Change of display language is performed by following the instructions below:
language
Step Action
1 Start by pressing [ADVANCED] from the MENU tab.
2 Press [SETUP].
3 Press [SETUP MENU 2].
4 Press [REGIONAL SETUP], a list of local settings will be displayed.
5 Press [MORE] until language button is displayed.
6 Press [LANGUAGE] to enter language screen.
7 Choose the number that corresponds with the language desired
and press OK to save.
Menu
Activate Mixer
To activate mixer follow the instruction below:
(optional)
Step Action
1 Start by pressing [ADVANCED] from the MENU tab.
2 Press [SETUP] and then [SETUP MENU 2].
3 Press [SETUP MENU 3].
4 Press [MIXER].
If the mixer is not activated the button will have empty
5 brackets ( [ ] ). To activate press button and select [X].
Upon sample aspiration mixer will discontinue rotation
Note until sample analysis is complete.
It is recommended that whole blood samples are mixed for
10 – 15 minutes and then analyzed. Mixing for more than
4 hours may cause erroneous results.
Important
23
4.2 Initial Setup (Continued)
Setting up The date/time function is shown on all samples and printouts and should
date/time always be setup correctly. To set date/time follow the instruction below:
Step Action
1 Start by pressing [ADVANCED] from the MENU tab.
2 Press [SETUP], then press [SETUP MENU 2].
3 Press [DATE/TIME SETUP] to enter the set date/time menu.
4 Press [DATE FORMAT] to select date specific setting.
1 = DD/MM/YY; 2 = YY/MM/DD, 3 = YY/DD/MM, 4 = MM/DD/YY
5 Press on the item that you want to change and enter the changes on the
numerical pad. See menus below.
Menus
Description This section describes how to install and configure external components
such as barcode readers, printers, data communication, etc.
Step Action
1 Start by pressing [ADVANCED] from the MENU tab.
2 Press [SETUP] and then [PRINT SETUP] to enter the Print Setup
menu.
3 Press [MORE] to view Printer type. Printer types are as follows:
1 = Seiko DPU 411/12 and 414
2 = IBM proprinter / Epson compatible
3 = HP PCL 3 and 5 protocol compatible
4 To change printer type press [PRINTER TYPE], enter the
correct number and press [OK] to save.
24
4.3 Advanced Setup (Continued)
Step Action
1 Start by pressing [ADVANCED] from the MENU tab.
2 Press [SETUP].
3 Press [PRINT SETUP] to enter the printer setup menu.
4 To set Manual Print Mode function select from the following:
0 = None, 1 = Without Histograms, or 2 = With Histograms.
5 To select Auto Print Mode function select from the following:
0 = None, 1 = Without Histograms, or 2 = With Histograms.
Note Extended printer format settings and user definable print layouts
are also available. Please refer to Document 02012.
(www.Swelab.se/support/html/whatsnew.htm), for detailed
information on how to set up a user definable format.
Requirements: English language only. Requires basic
understanding of printer protocol, formats and font definitions.
Barcode Setup To setup the barcode reader follow the instructions below. (Note that
the default barcode setting is 9600N81).
Step Action
1 Start by pressing [ADVANCED] from the MENU tab.
2 Press [SETUP].
3 Press [SETUP MENU 2].
4 Press [BARCODE SETUP] to enter the barcode setup menu.
5 Choose the format that is appropriate for the barcode reader being
installed. (The generic driver is most common and is Compatible
with most readers).
0 No barcode reader
1 Generic barcode reader (9600N81)
2 Panasonic ZE-84RMSM (9600O72)
Keyboard To setup the keyboard follow manufacturer instruction for setup and plug
Setup (optional) into analyzer keyboard port. See Section 2.3 for details.
25
4.3 Advanced Setup (Continued)
To select options for sending results and data follow instruction below:
1 Start by pressing [ADVANCED] from the MENU tab.
2 Press [SETUP].
3 Press [SERIAL SETUP] to enter the serial setup menu.
4 To set Manual Send Mode function select from the following:
0 = None, 1 = Without Histograms, or 2 = With Histograms.
5 To select Auto Send Mode function select from the following:
0 = None, 1 = Without Histograms, or 2 = With Histograms.
26
4.4 Reagent Setup
Description This section describes the functions of the reagent setup menu and how
to access reagent statistics.
Reagent Input The Swelab Alfa System is interlocked with specified Boule reagents for
(Enter New optimal performance. The reagent containers must be identified by the
Reagents) instrument before analysis of samples can begin. To identify reagents scan in
or manually enter the barcodes on the reagent containers. See section 2.6.
Step Action
1 Start by pressing [REAGENT SETUP] from the MENU tab.
2 On the lower left-hand side of the Reagent Setup Menu, both the
remaining cycles for Diluent and Lyse are displayed. (It is important to
remember that cycles include analyses, wash cycles, background
counts, primes, exit standbys, etc.)
3
27
4.4 Reagent Setup (Continued)
Inactivate It is possible for the operator to inactivate the current reagent box by pressing
Reagent the [INACTIVATE REAGENT] button and then [YES]. Once deactivated the
operator must scan in or manually enter another reagent container before
analysis of samples can begin.
Reagent The interlocked reagent system displays indicator and warning messages to
Indicators alert the operator when reagents are running low and need to be changed. See
Section 12.2 and 12.3.
Description This section describes the functions of available menus in the instrument that
have not been described in any other section of this manual.
Analysis Profile It shall be possible for operators to customize analysis profiles. See following
menu options:
Step Action
1 Start by pressing [ADVANCED] from the MENU tab.
2 Press [SETUP], then [ANALYSIS PROFILE] to enter the Analysis
Profile Setup menu.
3
28
4.5 User Interface (Continued)
Sample Memory The following procedures explain how to search for previous sample analyses
and statistics, and print, send, and delete samples.
Step Action
1 To view previous analyses at a quick glance press [PREV] or [NEXT]
buttons to scroll through samples in either Sample or List menus.
2 To view a specific sample or a group of samples press [CHANGE] in
List Menu. In this menu samples can be selected by Sample ID,
SEQ, Date, and Sample profile. Press corresponding button to select.
29
4.5 User Interface (Continued)
All Settings From Menu tab press [ADVANCED] and then [SETUP] to enter Setup Menu.
To print all instrument settings, verify instrument is connected to a
printer and press [PRINT ALL SETTINGS].
To send all instrument settings, verify instrument is connected to a
computer and press [SEND ALL SETTINGS].
Change From Menu tab press [ADVANCED] and then [SETUP] to enter Setup Menu.
Sequence To change sequence number press [SEQ NUMBER SETUP], press [NEXT
Number SEQ NUMBER], enter in new sequence number and press OK to save.
Platelet Contact local distributor for more information on Platelet Concentrate Mode
Concentrate activation.
Mode
30
Section 5: Sample Analysis
Section Overview
Introduction This section covers the sample analysis routine, including how to analyze a
sample in the five different modes offered in the Swelab Alfa.
Limitations Samples drawn in an open tube or vacuum tube should be analyzed between
15 minutes and 6 hours for most accurate results.
Handling of The blood should be allowed to adapt to the EDTA for 10-15 minutes after
samples sampling.
The sample should be thoroughly and gently mixed before analysis. It is
recommended to use a mixer.
The sample should be mixed for 10-15 minutes. A sample not correctly
handled may give erroneous results.
31
5.1 Preparations before Analysis (Continued)
Background The following sequence is performed to check that the background count
Check is low enough to run a sample.
Step Action
1 From the main screen press [NEW SAMPLE].
2 Press [NEXT PROFILE] or [PREV PROFILE] to scroll
to Background and press OK to save.
3 Press the whole blood start plate, which is located behind
whole blood aspiration needle. (See Figure 5.1 below)
Figure 5.1
The aspiration time is approximately 10 seconds. After ~ 10
seconds the instrument will time out due to no detection of
blood, and continue its cycle.
Accepted
The background count should not be higher than the figures shown
Background
values below, assuming that at least 2 “blanks” are run after a sample.
32
5.3 Sample Identification
Step Action
1 From the main screen press [NEW SAMPLE] or begin sample
aspiration, which automatically opens NEW SAMPLE menu.
2 Press numerical keys to enter sample ID or scan in the ID barcode
from the sample tube.
3 Press [NEXT PROFILE] or [PREV PROFILE] to scroll to desired
profile.
4 Press OK to save profile and sample ID or begin sample aspiration.
Menu
33
5.4 Analyzing the Sample (Open Tube)
Description This section describes how to aspirate and analyze a sample with the “Open
Tube” procedure.
Starting Refer to Section 5.1 for blood sample preparation and then follow the
procedure procedure below:
Step Action
Choose List, Sample, or Main menu to begin sample analysis.
1 Analyzer must be in one of these operation modes to aspirate.
Aspirate the sample through the aspiration needle by gently inserting
2 aspiration needle into the sample tube, press the whole blood start
plate behind the left aspiration needle. (See Figure 5.4)
Follow the instruction on the menu when to remove the sample tube.
3 A beep should be heard indicating sample removal.
Make sure that the blood sample tube is not touching the upper part of the
aspiration needle.
Not removing the sample tube could result in incorrect washing sequence
of the aspiration needle.
Important Do not remove sample prior to instruction, incomplete aspiration could
occur, causing erroneous results.
Sample Aspiration
4
Figure 5.4
As there are no assurances of the absence of HIV, Hepatitis B or C viruses
or other infectious agents in blood samples, controls, and calibrators these
products should be handled as potentially biohazardous.
Warning
Always wear protective gloves and goggles. Follow local regulations.
34
5.4 Analyzing the Sample (Open Tube) (Continued)
35
5.5 Analyzing the Sample (Pre-dilution procedure)
Description This section describes how to analyze a pre-diluted sample through the “pre-
dilute” aspiration needle and how to use the dispense function. There are two
ways of pre-diluting a sample. The recommended pre-dilute method is using
the dispense function, which uses the factory calibrated dilution ratio of 1:225
(20 µl in 4.5 ml diluent). The second method is performing an external pre-
dilution using in-house dilution procedures, dilution ratios between 1:200 –
1:300, and re-calibrating system using selected dilution ratio.
Time limitations Pre-dilute procedures are generally less precise than open and closed tube
procedures and results may vary depending on local laboratory procedures and
conditions. Blood cells may shrink and/or swell during the time between mixing
in the beaker and the actual analysis, resulting in compromised values of MCV,
MPV and the distribution between lymphocytes/mid-cells/ granulocytes (with
indirect effect on calculated parameters, e.g. HCT) . Thus, the time between
mixing and analysis should be minimized and under no circumstances exceed
60 minutes, since RBC, PLT, HGB and WBC may also be affected.
Externally Pre- Pre-dilute volumes 4.5ml – 5.0ml. The dilution ratio must always be the
diluted volumes same as the dilution it is calibrated to in order to avoid erroneous result;
and preparation any dilution variation in an externally diluted sample will affect the
parameter test results.
Prepare pre-dilute sample according to internal documentation and
time limitations section above.
Dispense This feature is to be used as a precision dispenser for dilution of
Function blood samples.
Dispense amount: 4.5 ml.
Dilution: 20 µl in 4.5 ml diluent (1:225)
Follow the instruction below:
Step Action
1 Press the [DISPENSE] button from the MENU tab.
Before pressing the pre-dilute start plate make sure that a waste
2 beaker is placed under the pre-dilute aspiration needle.
Press the pre-dilute start plate (right start lever) to enable dispense
3 mode. (The instrument will fill the waste beaker with a small
amount of diluent, this is to be discarded)
Fill the pre-dilute beaker by pressing the start plate again. If more
4 than one beaker is to be filled repeat this step.
36
5.5 Analyzing the Sample (Pre-dilute procedure) (Continued)
Menus
Pre-dilute
Start by selecting pre-diluted sample beaker and follow the procedure below:
procedure
Step Action
1 Choose List, Sample, or Main menu to begin sample analysis.
Analyzer must be in one of these operation modes to aspirate.
2 Aspirate the pre-diluted sample through the pre-dilute aspiration
needle by pressing and holding the pre-dilute start plate behind the
right-side aspiration needle for five seconds. (See Figure 5.11)
Figure 5.11
3 Follow the instruction on the menu when to remove the sample
tube. A beep should be heard indicating sample removal.
4 Refer to Section 5.4 Steps 5 - 10 for remainder of
analysis sequence.
37
5.6 Analyzing the Sample (Micro Capillary Inlet, MCI)
Description This section describes how to analyze capillary whole blood samples with
the use of the Micro Capillary Inlet (MCI).
Micropipettes ONLY Boule supplied, plastic, high precision EDTA micropipettes should
be used when running MCI. Glass micropipettes can cause damage to
instrument if inserted incorrectly.
Starting
Follow the procedure below to operate MCI:
procedure
Step Action
1 Choose List, Sample, or Main menu to begin sample analysis.
Analyzer must be in one of these operation modes to aspirate.
2 Pull out the MCI adapter. (The instrument will give an instruction
to put back the MCI adapter to start).
3 Remove the previous sample micropipette. (If applicable)
4 Place the adapter on the table.
Figure 5.12
Figure 5.13
38
5.6 Analyzing the Sample (Micro Capillary Inlet, MCI) (Continued)
Fill the micropipette completely with fresh whole blood and wipe
off excessive blood on the outside surface.
Be careful not to wick blood from open ends of the micropipette.
Important
Ignoring these instructions might cause incorrect and non-
reproducible results.
Insert the micropipette into the MCI device as shown below:
Figure 5.14
Insert the MCI into its holder and the instrument will
automatically start the analyzing sequence.
Figure 5.15
Do not remove MCI during sample aspiration or analysis. Removal prior
to completion of analysis may cause erroneous results.
Important
5 Refer to Section 5.4 Steps 6 - 10 for remainder of analysis
sequence.
Further To find more information, visit www.boule.se The Boule website contains
information and detailed video shots on how to use the MCI and cleaning procedures.
demos
39
5.7 Analyzing the Sample (Cap Piercing Device)
Description This section describes how to analyze whole blood samples using the
Cap Piercing Device.
Sample tube Most standard 3.0 ml to 4.5 ml tubes, with a maximum length of 77 mm,
description can be used in the cap piercing device. The minimum volume in the closed
tube should be approximately 1 ml.
Starting
Follow the procedure below to operate the Cap Piercing Device.
procedure
Step Action
1 Choose List, Sample, or Main menu to begin sample analysis.
Analyzer must be in one of these operation modes to aspirate.
2 Open door to cap piercer and insert vacuum tube upside down,
pressing the tube in place, aligning with lower support.
3
40
5.8 Analyzing the Sample (Auto Sampler)
Description This section describes how to analyze whole blood samples using the Auto
Sampler.
Sample tube Only standard 4.5 ml tubes can be used in the Auto Sampler. Do not use
description Sarstedt tubes. The minimum volume in the closed tube should
be approximately 1 ml.
Starting
Follow the procedure below to operate the Auto Sampler:
procedure
Step Action
1 Unlock the center piece by turning it counterclockwise.
2 Load the vacuum tube samples by placing the capped end towards
outer edge of sample wheel and fitting it into designated slot.
3 Lock in samples by turning center piece clockwise.
4 Press [SAMPLER] button from the NEW SAMPLE Menu.
5 Press [START] to immediately begin analysis or press
[EXTRA MIX] if extra mixing of samples is needed.
Do not touch sample wheels or samples during operation.
Handling and operation by unauthorized personnel may result in injury.
Warning 6 Auto sampler begins analysis with the sample tube placed in
the lowest position number.
41
5.9 Results
Description This section describes the information that can be obtained from the sample
analysis results.
After sample After a sample has been analyzed the result information can be viewed in the
analyze following three screen displays:
Sample View 1
HGB parameters
RBC histogram
Figure 5.20
Sample View 2
Analysis mode
Sample ID and analysis
profile
Primary Diagnostic
Parameters
Press on to view
Displays date and time of different views of
sample analysis, and WBC same sample.
and RBC counting times.
42
5.9 Results (Continued)
Sample View 3
Figure 5.22
43
Section 6: Quality Control (QC) and Blood Control
Memory
Section Overview
Introduction The Swelab Alfa is equipped with a QC memory capable of displaying and
printing X-B and Levey Jennings plots.
Introduction This section describes the procedures to be performed for running control
samples.
QC Menu and Follow the instruction below to access the QC menu and to input
CBD input Control/Calibrator Blood Definitions (CBD) from the Assay sheet.
Step Action
1 Enter the QC menu by pressing [QC] from the menu tab.
2 Press [ENTER CON/CAL].
3 Refer to the Assay sheet for instructions on how to input control
definitions. (These pages are delivered with authorized Boule controls).
44
6.1 Quality Control (QC) (Continued)
Control Analysis It is advisable that the performance of the Swelab Alfa system is checked
daily with a certified blood control authorized by Boule. Comparing the
analyzer results to the known values on the Boule control assay sheet is a
good assurance that the system is functioning properly.
Step Action
1 Follow directions on Assay Sheet to scan in values of CBDs.
2 Choose either List, Sample, or Main Menu to begin
control analysis.
3 Using installed barcode reader, scan the Control ID from the
blood control vial label.
4 Aspirate the blood control and wait for the results. The Swelab
Alfa will identify this ID and match the results with the
previously defined control blood definitions.
Search Function Each blood control type can be found by control definition (Lot number
and level), date or sequence number.
Step Action
1 Enter the QC menu and press [VIEW CON/CAL].
2 Input the search criteria to be used.
3 Pressing on the SEQ bar will display Figure 6.4, in which one
particular lot or level can be selected.
45
6.1 Quality Control (QC) (Continued)
L-J Plots Levey-Jennings (L-J) plots are used to monitor the long term stability of the
instrument using Boule blood controls.
Blood controls To be able to use L-J plots, the Control/Calibrator Blood Definition for the
blood controls, must be scanned with the installed barcode reader or manual
entered in. Follow direction on Assay Sheet to scan in values of CBDs.
Step Action
1 Enter the QC menu and press [VIEW CON/CAL].
2 Scan the barcode label on the blood control tube, with the
barcode reader, select control from Select Con/Cal Sample
Menu, or manually enter in value.
3 Press [L-J VIEW] to display the Levey - Jennings plots.
46
6.2 Levey-Jennings Plots (Continued)
L-J plot The image below is constructed from several samples and will not
Diagrams be shown as below until a sufficient amount of samples have been
analyzed.
Figure 6.5
4 Scroll through parameters by choosing [MORE].
5 Print diagrams by choosing [PRINT].
Parameters The L-J plots are displayed for all parameters defined in the CBD page except
displayed on the WBC differential parameter “MID”.
L-J Plots
Note In case a control shows an error or warning flag SE, DE, FD, OF, LO, HI, NG,
TU, TL or TB; the parameter values of such control will not be included in the
L-J plots.
Description The X-B function in the Swelab Alfa follows strictly the Bull algorithm for
the parameters MCV, MCH and MCHC. These parameters should not drift
as a function of time within a large patient population. The recommended
range setting is ± 3% from the expected mean value of these parameters.
47
6.3 Initialization and Use of X-B Function (Continued)
Step Action
1 Enter the QC menu and press [VIEW Xb STATS].
2 Select X-b points by Date or by default all sample data is selected.
3 Press [LJ VIEW] to display Xb L – J diagrams.
Xb L-J The image below is constructed from several samples and will not
Diagrams be shown as below until a sufficient amount of samples have been
analyzed.
Reference Bull BS, Hay KL. The blood count, its quality control and related methods: X-bar calibration
and control of the multichannel hematology analysers. In: Clangoring I. editor. Laboratory
Hematology: An account of Laboratory Techniques. Edinburgh.
48
Section 7: Calibration
Section Overview
Introduction This section describes the step-by-step procedure for calibration of the Swelab
Alfa. The instrument has been calibrated by Boule prior to shipment. Good
laboratory practice, however, requires regular checks and calibration of the
measured parameters
49
7.2 Calibration
Input calibrator Follow the instruction in Section 6.1 Quality Control to access the QC
definitions menu and to input Control/Calibrator Blood Definitions (CBD) from the
Assay sheet.
Whole Blood The following instructions calibrate Open Tube, Cap Piercer, and
Calibration Auto Sampler modes. Follow the instructions below to calibrate:
Step Action
1 Follow directions on Assay Sheet to scan in calibrator definitions.
2 Choose either List, Sample, or Main menu to begin calibrator analysis.
3 Using installed barcode reader, scan the Calibrator ID from the
calibrator vial label.
4 To perform calibration, it is recommended that five calibration
analyses be performed in consecutive order through the open tube
mode.
Important
5 When analyses are complete press [ADVANCED] from the MENU
tab.
6 Press [CALIBRATION] and then choose [WHOLE BLOOD].
50
7.2 Calibration (Continued)
Figure 7.3
51
7.2 Calibration (Continued)
14 Perform steps 9-12 for RBC, MCV, PLT, MPV, HGB, and WBC
parameters. To move to the next parameter press [MORE].
15 It is recommended to not change preset calibration factors for
RDW%, RDWa, and PDW. If necessary, please contact local
distributor or Boule service technician for procedure.
16 Once parameters are calibrated, a screen will be displayed asking
operator if a calibration report is wanted, [SEND], [PRINT], or
[EXIT] can be selected.
Figure 7.4
17 It is recommended to run controls after calibration to verify that all
parameters have been calibrated correctly. See section 6.1 to
perform QC.
Capillary Device To calibrate MCI follow Steps 1-16 above except select [CALIBRATION]
Calibration and then choose [CAPILLARY DEVICE] instead of Whole Blood
calibration in Step 6 and use MCI mode for analysis.
52
Section 8: Cleaning, Maintenance & Transport
Section Overview
Introduction This section contains information that is crucial for maintaining, transporting
and storing the Swelab Alfa.
Description The majority of the instruments cleaning procedures are automated to keep
the user maintenance to an absolute minimum.
Cleaning
The Daily Cleaning takes only a few minutes, the instructions are as follows:
Procedure
Step Action
1 Clean the aspiration needles using a paper tissue with a
70% alcohol solution.
2 Remove possible traces of salt crystals or blood using a paper
tissue with a disinfecting solution.
53
8.2 Monthly Cleaning
Description This section describes the cleaning procedure to be used to secure the correct
function of the instrument on a monthly basis.
Step Action
1 Clean the aspiration needles using a paper tissue with a
70% alcohol solution.
2 Fill a cup with 10 ml 2% hypochlorite (bleach), certified by Boule.
3 Aspirate the hypochlorite as a whole blood sample.
4 Repeat step 2 but aspirate the hypochlorite as a pre-diluted sample.
5 Run 2 blank samples by aspirating 10 ml diluent as a whole
blood sample.
6 Repeat step 5, but aspirate diluent as a pre-diluted sample.
Clot Prevention This process will decrease the risk of system to build up debris material in
the instrument system. This should be preformed at least once a month or
every 1000 samples. This procedure will take 15 minutes to complete.
Once this procedure is started the operator will be unable to abort the
cycle until it is completed.
Prematurely aborted the cycle could cause erroneous patient results if system
Importantis not cleaned properly.
Step Action
1 Fill a small container with 5 ml of Enzymatic Cleaner.
(Enzymatic Cleaner from the cleaning kit can be used.)
2 From Main Menu press [ADVANCED] and then
press [MAINTENANCE].
3 Hold the container under the OT needle, submerged in cleaner,
press [CLOT PREVENTION] and then [OK] to confirm. Do not
remove container with cleaner for at least 5 seconds after
aspiration has stopped.
4 The system will then perform the cleaning process, and upon
completion instrument is ready for next sample analysis.
54
8.3 Six (6) Month Cleaning
Description To increase the life of internal tubing in the instrument, the following cleaning
procedure is strongly recommended.
Cleaning Press [ADVANCED] from Main menu, then press [MAINTENANCE], and
Procedure then press [CLEAN CYCLE] to enter the Cleaning Menu.
Follow the instruction for the Boule Cleaning kit to clean the instrument.
(Instructions for use are supplied with the Boule Cleaning kit solutions).
The Six Month Cleaning procedure takes approximately one hour and 15
minutes to complete.
Boule Cleaning The Boule Cleaning Kit contains the following items:
Kit
Hypochlorite (2%)
Enzymatic cleaner
Detergent cleaner
55
8.4 Instrument Maintenance
Description This section describes the maintenance that is required to maintain and
increase the life of the instrument. Refer to local distributor for
warranty requirements.
Description This section describes the procedure performed to move the instrument
over very short distances. (From table to table).
Before the If the instrument is in “standby” mode do not unplug instrument. Make
re-location sure that the instrument is in Sample or List menu before turning off.
1. Do not detach the reagent level sensors or waste tube, place the
sensors on top of the instrument when moving. (Avoid reagent level
sensor contact.)
2. Remove the waste tube from waste container or drain.
3. Disconnect all electrical connections.
Re-location Make sure that the instrument is lifted from beneath to avoid
unnecessary stress on the front cover.
56
8.6 Short Term Transport (<12h)
Description This section describes the procedure performed before transporting the
instrument over short distances. This procedure only describes the
preparations performed before transporting the instrument for less than
12 hours.
Empty System 1. Remove the reagent level sensors from the reagent containers.
2. Press [ADVANCED] button on MENU tab.
3. Press [MAINTENANCE] and then [EMPTY SYSTEM].
4. When empty procedure is complete, the following statement will appear
on screen: ‘System is empty and ready for fill or power off.’
5. Switch off power and then unplug analyzer.
Before the After instrument is powered off, detach reagent level sensors, waste tubing, and
re-location all electrical connections. Package all components carefully for transport.
Description This section describes the procedure when transporting or shutting down
the instrument for a longer period of time (>12 hours).
57
8.7 Re-packaging and Long Term Transport (>12h) (Continued)
Step Action
1 Select [EMPTY SYSTEM] from MAINTENANCE Menu. See Section
7.5 “Short term transport” for emptying instructions.
2 Remove the reagent sensors from the reagent containers and follow the
instructions for the Boule cleaning kit. (Instruction is supplied with the
Boule cleaning kit solutions).
3 After completing the cleaning of the instrument, insert the reagent
sensors into distilled water. Select [FILL SYSTEM] from
MAINTENANCE Menu.
4 When the instrument has been filled with distilled water select
[EMPTY SYSTEM] from MAINTENANCE Menu.
5 When system is emptied, disconnect the main supply cable and other
connections such as reagent sensors and waste tubing.
6 Pack the instrument using the original shipping container.
7 Mark the container with DELICATE INSTRUMENT, FRAGILE and
THIS SIDE UP.
8 Follow Guidelines for transport below.
Guidelines for The instrument in its export package should fulfill the following
transport transport/storage conditions:
Does not exceed - 40°C for ≥ 24 hours.
Does not exceed a Dry heat of + 70°C for ≥ 24 hours.
Dramatic change of temperature between - 40°C and + 30°C.
Does not exceed a Damp heat steady state of 90% RH and +
40°C during 48 hours.
Does not exceed a Damp heat cyclic of 90-100% RH and +
25°/+40°C 12+12 hours.
Permanent Shut-
Down and Storing See Section 8.7 Long Term Transportation.
58
8.9 Disposal Information
Manufacturer Place the instrument close to a waste container or drain suitable for disposal
Guidelines of used reagents.
Check that the drainage is suitable for disposal of chemical and biological
waste.
Check that the waste tubing is securely fastened in the drain.
Always use protective gloves when working with the waste container and
the waste tubing.
Mandatory Action
Disposal
Used reagents
Materials Reagents mixed with potentially biohazardous material
Instrument and instrument components
Controls and calibration material
59
Section 9: Parameter Flags
Section Overview
Introduction The Swelab Alfa has several error and warning flags related to the measured
parameters. These flags alert the operator of possible pathologic samples and
parameter value errors.
Description The instrument has several parameter error flags related to the measured
parameters. The flags are shown on the display and printouts.
Resolving Error The sample should be re-analyzed. If the problem persists see list below to
Flags identify the issue and/or refer to Section 12 Troubleshooting.
Note Note that a parameter that is outside the “Normal Range”, refer to Section 9.4
for User Interface setup, is either marked with “H” or “L” on the printout and
display to indicate if the value is higher or lower than the pre-set “Normal
Range” values.
60
9.1 Short Description of Flags and Troubleshooting (Continued)
Parameter Flags
61
9.2 Detailed Description of Flags
Description This section covers the full description of the parameter warning and error flags
that might occur. The list below is in order of the severity of the flag.
62
9.2 Detailed Description of Flags (Continued)
63
9.2 Detailed Description of Flags (Continued)
64
9.3 Flagging Capabilities
Description This section describes the limitations of the flagging capabilities of the
Swelab Alfa series.
Abnormalities All samples with anomalies and /or abnormal distributions signaled by the
instrument should be analyzed manually by a blood smear. Pathological cells
may vary in their stability towards lysing of their cytoplasmic membranes
compared to normal cells, which may cause aberrations in the automated
analysis. This also applies to the presence of normal non-pathological cells
that have been subjected to chemotherapy or other treatments.
65
Section 10: Technology
Section Overview
Introduction This section describes the different methods and principles of measurement
and calculations.
Description This section describes the measuring principles of the Swelab Alfa.
General The measuring principles of the Swelab Alfa are based on impedance and
Measuring spectrophotometry principles.
Principles
Whole Blood The number of cells for determining RBC and WBC values are counted from a
Dilution suspension of 1:40,000 for the RBC and 1:400 for the WBC dilution ratio of
whole blood.
Theoretical If a sample contains 5 million red blood cells per µl, a dilution of 1:40 000
Principles will give a final concentration of 5 million divided by 40,000 = 125 cells per
(RBC Example) µl. Each µl containing 125 cells, drawn through the aperture, will generate 125
pulses.
66
10.1 Measuring Principles (Continued)
Measured The measured volume drawn through the aperture is 270 µl (Manufacturer
Volumes calibrated). Based on the assumption made above, the system will count
(Example) 270*125 = 33,750 pulses. The analyzer uses a fixed division factor of 67.5
calculated as 33,750 / 67.5 = 500 which is the correct value. (Based upon
this calculation the instrument would show RBC = 5.0x106 cells/µl).
Stop sensor
Measured Flow
Volume
Start sensor
Figure 10.1
Theoretical The calculation principle for white blood cells is the same but with a
Principles (WBC difference in dilution ratio and cell quantity. An example of this could be
Example) as follows: 5,000 cells/ µl diluted 1:400 =12.5.
Description The counting time is defined as being the time needed for the sample to fill the
metering unit from the start to the stop detector.
Counting Time The normal counting time limits for the RBC and WBC metering units are
Limits between 13 – 18 seconds and 10 – 13 seconds respectively. If the counting
time is below or exceeds the above mentioned limits, the flag “LO” or “HI”
will be displayed.
Note The ´counting time´ is not related to the actual result. Atmospheric pressure
variations, protein built up within the orifice (aperture) and other secondary
effects that might cause pressure changes will NOT affect the counted
parameters RBC, PLT and WBC.
67
10.3 WBC Differentials
Fixed The differentiation of the WBC cells into lymphocytes, mid-cells and granulo-
Discriminators cytes is presented in the number of cells per liter or cubic millimeter and in the
percentage of the total number of WBC cells. The MID discriminator of WBC
is set to 140 and 180 fl. The WBC histogram is automatically adjusted
depending on the number of cells, i.e. expanded for low values and
compressed for high values.
68
Section 11: Specifications
Section Overview
Introduction This section describes the specifications for the Swelab Alfa and its parameters.
11.1 General
Description This section describes the Swelab Alfa and its parts in general.
User The operator works with a menu from which the desired program is chosen,
Environment e.g. discriminator settings.
3-Part WBC The instrument performs a 3-part WBC differential by means of a cyanide free
hemolyzing reagent.
Protected A sample memory is available and protected against main power failures. The
Sample Memory sample memory also contains a search function with selective printing and QC
Options.
69
11.2 Short List of Specifications
Specifications (Short)
70
11.3 Parameter Ranges
Measuring Range
The correlation is performed using a Bayer/Advia 120 as reference.
and Correlation
Reproducibility (typical)
Measured as an average of 10 measurements each on 3 different vein K2-
EDTA collected normal samples, on three instruments. Values shown
have been corrected to show 95% confidence limits.
Parameter X-mean (CGS units) CV (%)
WBC 8.4 < 3.5
RBC 4.34 < 1.8
MCV 94.4 < 1.5
PLT 313 < 4.8
HGB 13.7 < 1.5
Description This section describes the reagent consumption for the Swelab
Alfa depending on a sample per day calculation.
Supported Use only Boule authorized reagents. Erroneous results and damage
Reagents may occur if other reagents are used.
Diluent
Approximately 22 ml per analysis cycle
Consumption
71
11.4 Reagents and Reagent Consumption (Continued)
Lyse
Approximately 4.5 ml per analysis cycle.
Consumption
Diluent
Consumption Diluent Consumption
30
25
20
ml/sample
15
10
0
25 50 75 100 125 150 200
Samples/day
Figure 11.1
Lyse
Consumption Lyse Consumption
6
4
ml/sample
0
25 50 75 100 125 150 200
Samples/day
Figure 11.2
72
Section 12: Troubleshooting
Section Overview
Introduction This section contains information needed to troubleshoot the Swelab Alfa
instrument.
Printer Issues See Section 4.3 Printer Modes for further detail.
73
12.1 Communication Issues (Continued)
Serial Data Issues See Section 4.3 Data Communication for further detail.
Barcode Issues See Section 4.3 Barcode Setup for further detail.
General General information displays are informative screen displays that appear after
Information a function has been completed. Instruction is then displayed for the operator
Displays on next step or function to be performed.
74
12.2 General Information Displays (Continued)
The system is empty from all liquid The system is filled with liquid and is The system has not been used during
and prepared to be filled with other prepared for power off. Press [PWR the preset display saver time. Press
liquid or be stored away. Press UP] if you want to return the system [RESUME] to activate the
[FILL] if you want to refill system to active status or [EXIT] if you want instrument. Once activated, the
or [EXIT] if you want to return to to return to instrument menu. It is instrument is ready to perform an
instrument menu. No analyze can be recommended to use [ENTER analysis.
performed before the instrument is STANDBY] and that power is left on,
refilled with reagents. instead of using this feature.
Instrument will enter Standby mode The instrument is in the process of The system is in Standby. Press
in 2 minutes. Press [CANCEL] to going into Standby. Please wait. [EXIT STANDBY] to activate the
return to instrument menu. instrument. Once activated, the
instrument is ready to perform an
analysis.
75
12.2 General Information Displays (Continued)
The system is preparing the The instrument is in process of The instrument is in process
instrument for analysis mode. If the powering down. Please wait. of powering up. Please wait.
background check is activated,
background result will be displayed.
Once activated, the instrument is
ready to perform an analysis.
Diluent Dispense Informational Displays
The instrument is preparing to The instrument is now dispensing The instrument is exiting
dispense diluent. Dispose of 4.5 ml of diluent. Please wait. dispense function. Please wait.
first dispense for best results.
76
12.2 General Information Displays (Continued)
The instrument is priming the The instrument is filling the system. The instrument is cleaning the Open
system. Please wait. Please wait. Tube needle. Please wait.
Every twelve hour the instrument The system has finished the count of The printer is in the process of
performs a wash of the system. cells and displays the results. The printing. Please wait.
During wash cycle the instrument analysis cycle in not yet completed, as
can not be used for performing an the system still needs to perform wash
analysis. cycle for an accurate next sample
result. Please wait until the [NEW
SAMPLE] button is activated. If
needle was submerged in next sample
by mistake, perform a background
count before continuing with the next
analysis.
77
12.2 General Information Displays (Continued)
Instrument displays this notice to Instrument displays this notice to Instrument displays this notice to
inform operator that ComboPack inform operator that Diluent reagent inform operator that Lyse reagent
reagents will soon need to be will soon need to be changed. will soon need to be changed.
changed.
Instrument displays this notice The reagent barcodes were scanned in The CBDs were scanned in correctly
when reagent container or correctly using the barcode reader and using the barcode reader and the
containers need to be changed. Not the instrument has accepted the instrument has accepted the values.
changing reagents at this time could values.
cause erroneous results or possible
damage the instrument.
78
12.3 Warning Displays
Warning Warning displays appear after a function has been performed incorrectly or
Displays to inform the operator that further action is needed to complete the desired
task. The warning display describes the situation and instructs the operator
on next step or function to resolve issue.
The system has been switched off The system was switched off The system was manually switched
for a long time period. The incorrectly. Perform a prime to off with system emptied of reagents.
instrument has been powered down prepare the system for analysis. Check Fill the instrument with reagents to
with all valves open and filled with method for correct instrument power prepare for analysis or exit if only a
liquid. Empty and refill the system down procedure. search of instrument menus is
with reagents, and perform a needed.
background count.
The instrument has been switched The system was powered down with The regular 12 hour wash has failed.
off with power down function liquid in system and has been unused Make sure that reagent containers
before power was switched off. for long period of time. Perform the are filled and the detectors are
Perform a power up to prepare the cleaning procedure according to inserted correctly.
reagent system for analysis. cleaning kit instruction. Perform a
background check.
79
12.3 Warning Displays (Continued)
The regular 12 hour wash has not The reagent container or containers are This message is displayed if reagent
been preformed. Check if reagent empty. Check if the containers are container or containers are empty
containers are empty and if the empty and if level sensors and reagent when coming out of Standby. Check
reagent detectors are in contact contact plugs are inserted correctly. if the containers are empty and if
with reagent. level sensors and reagent contact
plugs are inserted correctly.
ComboPack container needs to be Diluent container need to be changed. Lyse container needs to be changed.
changed. Not changing reagents Not changing reagents at this time Not changing reagents at this time
at this time could cause erroneous could cause erroneous results or could cause erroneous results or
results or possible damage the possible damage the instrument. possible damage the instrument.
instrument. Connect new reagent Connect new reagent container and Connect new reagent container and
container and scan in barcode on scan in barcode on container. scan in barcode on container.
container.
80
12.3 Warning Displays (Continued)
No more space is available to scan CBD barcode scanning failed. The Reagent barcode scanning failed.
in new CBD. Follow the CBD or order of scanning in the Barcode printing or order of
recommendation or manually delete barcodes may have been incorrect. scanning in the barcodes may have
all the controls with same ID, to Verify that setups on the instrument been incorrect. Verify that setups on
free space for scanning the new match the required setup for the the instrument match the required
CBD. barcode reader. setup for the barcode reader.
Open Tube Warning Displays
The instrument was unable to wash The instrument was unable to wash The instrument is unable to wash the
the Open Tube aspiration needle. the Open Tube aspiration needle. Open Tube aspiration needle. Verify
Verify that tube is removed and Verify that tube is removed and wash that tube is removed and wash
wash device is in correct position, device is in correct position. It is device is in correct position. It is
then perform OT Wash. recommended that background count recommended that background count
is performed before next sample is performed before next sample
analysis. analysis.
81
12.3 Warning Displays (Continued)
The MCI was opened during an The MCI was opened during a cycle The MCI holder was opened
inappropriate time. It is or analysis. Re-insert holder, and during an inappropriate menu. The
recommended to perform a follow suggested recommendation. MCI holder should only be opened
prime cycle before next analysis. in List, Sample or Main menu.
Cap Piercer and Auto Sampler Warning Displays
The Cap Piercer door was opened The aspiration wheel has been Three aspirations have been
before the CAP door lock had been interfered with during mixing. Press attempted. All have failed. Verify
activated. Close the Cap Piercer [OK] to return to the sample menu. that sample tubes contain at least
door to continue with the analysis. To proceed with the analyses press 1 ml of blood.
[CONTINUE] in Auto Sampler List
Menu.
82
12.3 Warning Displays (Continued)
83
Index
A mixer ................................................................................ 23, 31
21, 36, 52, 70
Aspiration Issues ................................................... 73, 83 MPV ......................................................................
aspiration needle ......................... 32, 34, 36, 37, 40, 45, 49, 83 N
Assay sheet .......................................................... 44, 50
Auto Sampler ..................................... 31, 41, 50, 56, 82, 83 NEW SAMPLE ................................................. 32, 33, 35, 41, 77
B O
background count................................. 32, 63, 77, 79, 81, 87 Open Tube................................................... 31, 34, 35, 50, 77, 81
barcode.................. 24, 25, 33, 41, 45, 46, 50, 73, 74, 78, 80, 81 Operator ID ............................................................................. 51
barcode reader .................................... 25, 41, 45, 46, 50, 81 P
blood controls ............................................ 5, 6, 10, 32, 46
Parameter Ranges .............................................................. 69, 71
C PCT .................................................................................. 21, 70
calibration ......................................... 48, 49, 50, 51, 52, 59 PDW ............................................................................ 21, 52, 70
PLT...................... 13, 21, 29, 32, 36, 50, 52, 61, 62, 64, 67, 70, 71
Calibration Code........................................................ 51
75, 76, 79
calibrators ....................................... 5, 6, 10, 32, 34, 45, 49 Power Down.................................................................
7, 12, 13
Cap Piercer ............................................ 17, 40, 50, 70, 82 power supply..................................................................
75, 76
CBD ........................................... 44, 47, 50, 51, 74, 78, 81 Power Up ..........................................................................
Clot Removal ........................................................... 85 printer .............................................. 13, 14, 24, 25, 30, 73, 74, 77
CV ............................................................... 51, 70, 71 Q
D 17, 44, 45, 46, 48, 50, 52, 69, 70
QC. ............................................
date/time function ...................................................... 24 R
DF. ............................................................... ….61, 64
Diluent Dispense ....................................................... 76 RBC................ 13, 21, 29, 32, 36, 50, 52, 61, 62, 64, 66, 67, 70, 71
Dilution Rates ........................................................... 36 RDW ........................................................................... 21, 52, 70
dispenser ................................................................. 36 RDWa..................................................................................... 21
Disposal......................................................... 15, 53, 59 reagent barcodes ...................................................................... 16
distributor.............................................4, 7, 13, 51, 52, 56 Reagent Consumption ........................................................ 69, 71
DP. ................................................................... 61, 64 reagent level sensor 14, 56
................................... ........................ .
reagents.................... 5, 6, 14, 16, 49, 59, 61, 64, 71, 75, 78, 79, 80
E
S
EDTA ........................................................... 31, 38, 71
Electrical Connections ............................................. 10, 12 safety features........................................................................ 5, 6
Emergency Procedure.................................................... 7 sample analysis........... 23, 31, 34, 35, 37, 38, 40, 42, 50, 54, 74, 81
erroneous results ................... 6, 7, 23, 31, 32, 34, 39, 57, 78, 80 Sample collection .................................................................... 31
Sample Memory ................................................................ 29, 69
F sample statistics....................................................................... 29
fill.......................................... 15, 16, 36, 57, 61, 64, 67, 87 Sample View ................................................................ ..... 42, 43
26, 74
fixed discriminator ..................................................... 68 Send Mode ........................................................................
30, 45
flags .........................................................51, 60, 62, 70 sequence number ...............................................................
serial output............................................................................. 26
G Service ................................................................. 4, 5, 13, 83, 87
service technician ................................................... 51, 52, 56, 83
General Information Displays .................. 73, 74, 75, 76, 77, 78 Setup................................................ 22, 23, 24, 25, 26, 28, 30, 74
GRAN.......................................................21, 63, 64, 70 Specifications .................................................................... 69, 70
H Standby ............................................................................. 75, 76
Storage.............................................................................. 53, 58
HCT ............................................................. 21, 36, 70 system flow....................................................................... 14, 20
HGB .............................. 21, 32, 36, 50, 52, 61, 63, 64, 70, 71
T
I
target values ............................................................................ 51
Installation...................................................... 10, 11, 13 TL................................................................................ 47, 61, 62
instrument settings ................................................. 22, 30 Transport...................................................................... 53, 57, 58
Troubleshooting............................................................ 60, 73, 83
L
TU. ............................................................................ ..47, 61, 62
language ............................................................. 23, 25
W
Levey-Jennings Plots ......................................... 44, 46, 47
LPCR................................................................. 21, 70 Warning Displays ........................................ 73, 79, 80, 81, 82, 83
LYM ........................................................21, 63, 64, 68 warning signs ............................................................................ 7
warranty .............................................................................. 5, 56
M 77
wash cycle...............................................................................
maintenance ................................................53, 56, 73, 87 waste.............................................. 6, 7, 15, 32, 36, 56, 57, 58, 59
MCH ............................................................ 21, 47, 70 waste tube ......................................................................... 15, 56
MCHC .......................................................... 21, 47, 70 WBC... 13, 21, 29, 32, 36, 47, 50, 52, 61, 62, 63, 64, 66, 67, 68, 69,
MCI ....................................... 10, 17, 31, 38, 39, 52, 70, 82 70, 71
MCV .......................................... 21, 36, 47, 50, 52, 70, 71 X
measuring principles ................................................... 66
Menu Structure ..................................................... 18, 19 X-B function ........................................................................... 47
micropipette ......................................................... 38, 39
MID ................................................ 21, 47, 63, 64, 68, 70
84
Appendix A
Clot Removal This process will help operator to remove a clot from the system. This should
only be used when the OT aspiration needle is blocked and Clot Prevention
procedure can not be performed. THIS SHOULD ONLY BE PERFORMED
BY A SERVICE TECHNICIAN OR AUTHORIZED PERSONNEL.
Step Action
1 Remove outer cover:
Press release lever on underside of cover.
Figure 13.1
While pressing release lever, place one hand on top of analyzer to stabilize
and then gently pull bottom of cover forward (only enough to slide pass
release lever)
Place both hands on upper sides of cover and carefully pull towards you.
Figure 13.4
Place cover aside.
85
CLOT REMOVAL PROCEDURE (Continued)
Step Action
Be very careful when removing cover to not damage analyzer.
Follow directions and do not force.
Be aware of aspiration and pre-dilute needles.
Important
2 Prepare a syringe by attaching a piece of maintenance tubing to syringe tip and fill
syringe with 2% Hypochlorite solution. (Hypochlorite from the cleaning kit can be
used.)
3 Locate the Valve 27, the lower valve directly to the left of shear valve.
4 Locate the L (elbow) connector on the right-hand side of this valve and disconnect
the L connector from ONLY the tubing that is threaded through valve.
5 From Main Menu press [ADVANCED] and then press [SERVICE].
6 Attach prepared syringe tubing to L connector, press [CLOT REMOVAL], press
[OK], and gently apply pressure back and forth to syringe until clot is loosened.
Figure 13.7
8 Disconnect syringe and reattach L connector to valve tubing.
9 Replace analyzer cover:
Carefully align top edge of analyzer and display with cover.
Gently, partial push on upper part of cover to fit over display.
Using both hands on sides of covers, slowly press on, fitting over
aspiration plates.
If aligned properly release lever will automatically click into place, there will be
no spacing between cover and display, and aspiration plates will move freely.
10 Once cover it replaced Exit out of Service menu. Press [MAINTENANCE]
and then perform two Clot Prevention cycles following instructions above.
11 Run a background count and check that it is within limits (See Section 5.2), and
if necessary a control to verify that clot removal was successful.
86
DF or DP ERRORS
NO
NO Re-analyze
DF/DP flag? DF/DP flag?
sample
YES NO
NO
Is there
YES Contact Service
any liquid in air
Representative
bin?
NO
NO
DONE
87
Discordant Results
Questions to ask:
1. Was sample collected and handled properly?
2. Was the same sample used for both in-house and outside lab analysis?
3. Different blood draw and/or different tube?
4. Could the sample have been switched with another patient?
5. Could discordance be due to age changes during shipping or time periods
between blood draw and analysis (RBC swelling, platelet clumping,
deterioration of WBC for differential)?
Go to control out of
Are controls NO range Troubleshooting Repeat using correct
Check that sample is
within range?
YES Is MCHC protocol procedure
mixed correctly and no out-of-range?
hemolysis or lipemia is
present. YES NO
NO
NO Are all parameters YES Was calibration YES Was calibration
Is HGB recently protocol followed
Perform HGB YES discordant?
out-of-range? performed? correctly?
blank.
NO
NO YES
1. Verify by looking at
Is PLT
blood film. NO
2. Check sample for clots.
YES out-of-range? Change out Are reagent NO Were all
3. Review PLT histogram. reagent and parameters
levels ok?
re-run sample. low?
(verify bell-shaped curve)
NO
YES Perform
YES Maintenance
Is HCT low
Verify that sample is
mixed correctly and tube YES compared to Review differences that can Is monthly or 6- YES
spun PCV? be expected when comparing
has correct ratio of blood month Maintenance
results from same system,
to anticoagulant. due?
NO from different systems, etc.
NO
Are
Examine blood film Perform Clot
RBC and PLT both
high or both Prevention procedure
YES low? and rerun sample.
Are
RBC and HGB
Contact Service
both low?
Representative
Contact Service
NO Representative
88
Display Issues
Usual Cause:
1. Keypad flex-cable loose
2. Static electricity
3. Power Outage/Lightening
Verify/record with
clinic if power
surge was noted
Is green power
NO Contact Service
light on front of
Representative
analyzer lit?
YES
YES
NO
Turn off analyzer with
On/Off switch.
YES NO
Contact Service
Representative
89
HIGH BACKGROUND COUNTS
Initial Procedure:
1. Check Diluent Lot Number and expiration date.
2. Check age of Diluent (i.e. when was it opened?)
3. Check that level sensors are pushed all the way to the bottom of the reagent containers and firmly
tightened on back of analyzer.
4. Check that level sensors are in correct reagent containers (red=diluent, yellow = lyse)
5. Check reagent level.
6. Check environmental condition (i.e. extreme temperature fluctuations?)
Run total of 3
backgrounds to See Clot Removal
verify procedure
YES
YES
Did YES
background DONE
pass?
Repeat procedure NO Was procedure YES
using correct protocol followed correctly?
NO
YES Did
DONE background Is maintenance YES Perform scheduled
pass? procedure scheduled to maintenance and re-
be performed? analyze background.
NO
Run 3 NO
backgrounds. Did
Try another box of NO
background
diluent.
pass?
Did NO YES
DONE
YES background Did YES
pass? background DONE
pass?
NO
Perform heated
detergent procedure,
using detergent from
6-month cleaning kit
90
Noise Issues
Usual Cause:
1. Bad electrical outlet in clinic
2. Power Outage/Lightening
NO
YES
Are
Are the YES any cables loose
values for the Noise
on back of
as follows?
analyzer?
NO
Try plugging
analyzer into
different outlet
Is Noise test
NO
positive?
YES
YES
Is Noise test Recommend trying Did line
conditioner
positive? line conditioner work?
NO
NO
Contact Service
Representative
91
TU or TL ERRORS
Perform
NO
background count
YES
Remove right-side Check tubing to
YES
Perform MPA TU/TL flag? cover see instruction measuring chambers,
blockage removal in manual. re-attach if necessary.
procedure
NO
Is
DONE NO liquid level above
orifice in measuring
chambers?
YES
YES
Contact Service
Representative
92
Art no 1504154 Apr 2006