Velocity measurement errors were investigated for an array-based preclinical ultrasound scanner (... more Velocity measurement errors were investigated for an array-based preclinical ultrasound scanner (Vevo 2100, FUJIFILM VisualSonics, Toronto, ON, Canada). Using a small-size rotating phantom made from a tissue-mimicking material, errors in pulse-wave Doppler maximum velocity measurements were observed. The extent of these errors was dependent on the Doppler angle, gate length, gate depth, gate horizontal placement and phantom velocity. Errors were observed to be up to 172% at high beam-target angles. It was found that small gate lengths resulted in larger velocity errors than large gate lengths, a phenomenon that has not previously been reported (e.g., for a beam-target angle of 0°, the error was 27.8% with a 0.2-mm gate length and 5.4% with a 0.98-mm gate length). The error in the velocity measurement with sample volume depth changed depending on the operating frequency of the probe. Some edge effects were observed in the horizontal placement of the sample volume, indicating a change...
Wall-less flow phantoms are preferred for ultrasound (US) because tissue-mimicking material (TMM)... more Wall-less flow phantoms are preferred for ultrasound (US) because tissue-mimicking material (TMM) with good acoustical properties can be made and cast to form anatomical models. The construction and geometrical stability of wall-less TMM flow phantoms is described using a novel method of sealing to prevent leakage of the blood-mimicking fluid (BMF). Wall-less stenosis flow models were constructed using a robust
Arteriovenous prosthetic grafts are used in hemodialysis. Stenosis in the venous anastomosis is t... more Arteriovenous prosthetic grafts are used in hemodialysis. Stenosis in the venous anastomosis is the main cause of occlusion and the role of local hemodynamics in this is considered significant. A new spiral graft design has been proposed to stabilize the flow phenomena in the host vein. Cross-flow vortical structures in the outflow of this graft were compared with those from a control device. Both grafts were integrated in identical in-house ultrasound-compatible flow phantoms with realistic surgical configurations. Constant flow rates were applied. In-plane 2-D velocity and vorticity mapping was developed using a vector Doppler technique. One or two vortices were detected for the spiral graft and two to four for the control, along with reduced stagnation points for the former. The in-plane peak velocity and circulation were calculated and found to be greater for the spiral device, implying increased in-plane mixing, which is believed to inhibit thrombosis and neo-intimal hyperplasia.
Assessment of the severity of arterial disease is typically made using B-mode imaging for plaque ... more Assessment of the severity of arterial disease is typically made using B-mode imaging for plaque evaluation and Doppler ultrasound for estimation of maximum blood velocity. Measurement of other blood velocity related quantities which are thought to be important in plaque development, such as wall shear stress, has proven difficult using conventional 2D ultrasound. This paper describes a 3D ultrasound system
The aim of this work was to investigate the potential of a novel forward-viewing intravascular ul... more The aim of this work was to investigate the potential of a novel forward-viewing intravascular ultrasound (IVUS) system for flow quantification and colour flow imaging combined with B-mode imaging. A stiff 3.8-mm diameter catheter was used to scan a 72 degrees sector ahead of its tip. Operating at 30 MHz, the catheter was integrated with an IVUS scanner and a radiofrequency (RF) data-acquisition system. RF data were software processed for producing B-mode images and deriving velocity estimates. Steady flow in the range of 45 to 146 mL/min toward the catheter, was used in wall-less tissue-mimicking phantoms simulating healthy lumen (8-mm diameter), 30% diameter symmetrical stenosis and 37% diameter eccentric stenosis. The system provided colour flow images and good estimation of peak velocity and volumetric flows (within 1% to 9% and 16% to 48%, respectively, of calculated values) at 5 to 7 mm distal to the catheter. A sector forward-viewing IVUS imaging/Doppler system is suitable for combined anatomical and functional assessment of stenosed vessels.
Doppler test objects are used to characterise Doppler systems, both stand-alone systems and the D... more Doppler test objects are used to characterise Doppler systems, both stand-alone systems and the Doppler part of so-called duplex scanners. The aim of the project partially presented here is the development and validation of an example of a Doppler test object fulfilling the requirements of the IEC 1685. The project has been carried out by nine partners of five European countries and has been funded by the European Commission. The flow Doppler test object is composed of: tissue mimicking material (TMM), blood mimicking fluid (BMF), tube (embedded in the TMM and carrying the BMF), tank flow system, including a pump and a flow meter. In the normative part of the IEC 1685, requirements are given for the values of acoustical parameters of TMM and BMF such as sound velocity, attenuation and backscattering. For BMF, requirements are given also for values of density and viscosity. In an informative (but not compulsory) annex, a description is given of a flow test object meeting these requirements as an example. 'example test object' developed during the project is composed of TMM based on agar and including SiC- and Al2O3-powders, BMF based on nylon particles suspended in water and glycerine, and a tube of c-flex, a silicon copolymer. Two tube sizes are used: 4.0 mm ID and 8.0 mm ID. During the project, very precise recipes have been developed for the composition and preparation of both TMM and BMF. Based on these recipes and a description of the construction in a design five flow test objects have been constructed in the laboratories of five participants. The test objects have been compared by measurements of the physical parameters and by Doppler measurements of the five test objects with the Doppler system. The measurements have been carried out by five observers. Inter-test object and inter-observer variabilities are determined, yielding information about usefulness of the parameters.
A more accurate means of prediction of abdominal aortic aneurysm (AAA) rupture would improve the ... more A more accurate means of prediction of abdominal aortic aneurysm (AAA) rupture would improve the clinical and cost effectiveness of prophylactic repair. The purpose of this study was to determine whether AAA wall distensibility can be used to predict time to rupture independently of other recognized risk factors. A prospective, six-center study of 210 patients with AAA in whom blood pressure (BP), maximum AAA diameter (Dmax), and AAA distensibility (pressure strain elastic modulus [Ep] and stiffness [beta]) were measured at 6 months with an ultrasound scan-based echo-tracking technique. A stepwise, time-dependent, Cox proportional hazards model was used to determine the effect on time to rupture of age, gender, BP, Dmax, BP, Ep, beta, and change in Dmax, Ep, and beta adjusted for time between follow-up visits. Median (interquartile range) AAA diameter was 48 mm (41 to 54 mm), median age was 72 years (68 to 77 years), and median follow-up period was 19 months (9 to 30 months). In the Cox model, female gender (hazards ratio [HR], 2.78; 95% CI, 1.23 to 6.28; P =.014), larger Dmax (HR, 1.36 for 10% increase in Dmax; 95% CI, 1.12 to 1.66; P =.002), higher diastolic BP (HR, 1.13 for 10% increase in BP; 95% CI, 1.13 to 1.92; P =.004), and a decrease in Ep (increase in distensibility) over time (HR, 1.38 for 10% decrease in Ep over 6 months; 95% CI, 1.08 to 1.78; P =.010) significantly reduced the time to rupture (had a shorter time to rupture). Women have a shorter time to AAA rupture. The measurement of AAA distensibility, diastolic BP, and diameter may provide a more accurate assessment of rupture risk than diameter alone.
Velocity measurement errors were investigated for an array-based preclinical ultrasound scanner (... more Velocity measurement errors were investigated for an array-based preclinical ultrasound scanner (Vevo 2100, FUJIFILM VisualSonics, Toronto, ON, Canada). Using a small-size rotating phantom made from a tissue-mimicking material, errors in pulse-wave Doppler maximum velocity measurements were observed. The extent of these errors was dependent on the Doppler angle, gate length, gate depth, gate horizontal placement and phantom velocity. Errors were observed to be up to 172% at high beam-target angles. It was found that small gate lengths resulted in larger velocity errors than large gate lengths, a phenomenon that has not previously been reported (e.g., for a beam-target angle of 0°, the error was 27.8% with a 0.2-mm gate length and 5.4% with a 0.98-mm gate length). The error in the velocity measurement with sample volume depth changed depending on the operating frequency of the probe. Some edge effects were observed in the horizontal placement of the sample volume, indicating a change...
Wall-less flow phantoms are preferred for ultrasound (US) because tissue-mimicking material (TMM)... more Wall-less flow phantoms are preferred for ultrasound (US) because tissue-mimicking material (TMM) with good acoustical properties can be made and cast to form anatomical models. The construction and geometrical stability of wall-less TMM flow phantoms is described using a novel method of sealing to prevent leakage of the blood-mimicking fluid (BMF). Wall-less stenosis flow models were constructed using a robust
Arteriovenous prosthetic grafts are used in hemodialysis. Stenosis in the venous anastomosis is t... more Arteriovenous prosthetic grafts are used in hemodialysis. Stenosis in the venous anastomosis is the main cause of occlusion and the role of local hemodynamics in this is considered significant. A new spiral graft design has been proposed to stabilize the flow phenomena in the host vein. Cross-flow vortical structures in the outflow of this graft were compared with those from a control device. Both grafts were integrated in identical in-house ultrasound-compatible flow phantoms with realistic surgical configurations. Constant flow rates were applied. In-plane 2-D velocity and vorticity mapping was developed using a vector Doppler technique. One or two vortices were detected for the spiral graft and two to four for the control, along with reduced stagnation points for the former. The in-plane peak velocity and circulation were calculated and found to be greater for the spiral device, implying increased in-plane mixing, which is believed to inhibit thrombosis and neo-intimal hyperplasia.
Assessment of the severity of arterial disease is typically made using B-mode imaging for plaque ... more Assessment of the severity of arterial disease is typically made using B-mode imaging for plaque evaluation and Doppler ultrasound for estimation of maximum blood velocity. Measurement of other blood velocity related quantities which are thought to be important in plaque development, such as wall shear stress, has proven difficult using conventional 2D ultrasound. This paper describes a 3D ultrasound system
The aim of this work was to investigate the potential of a novel forward-viewing intravascular ul... more The aim of this work was to investigate the potential of a novel forward-viewing intravascular ultrasound (IVUS) system for flow quantification and colour flow imaging combined with B-mode imaging. A stiff 3.8-mm diameter catheter was used to scan a 72 degrees sector ahead of its tip. Operating at 30 MHz, the catheter was integrated with an IVUS scanner and a radiofrequency (RF) data-acquisition system. RF data were software processed for producing B-mode images and deriving velocity estimates. Steady flow in the range of 45 to 146 mL/min toward the catheter, was used in wall-less tissue-mimicking phantoms simulating healthy lumen (8-mm diameter), 30% diameter symmetrical stenosis and 37% diameter eccentric stenosis. The system provided colour flow images and good estimation of peak velocity and volumetric flows (within 1% to 9% and 16% to 48%, respectively, of calculated values) at 5 to 7 mm distal to the catheter. A sector forward-viewing IVUS imaging/Doppler system is suitable for combined anatomical and functional assessment of stenosed vessels.
Doppler test objects are used to characterise Doppler systems, both stand-alone systems and the D... more Doppler test objects are used to characterise Doppler systems, both stand-alone systems and the Doppler part of so-called duplex scanners. The aim of the project partially presented here is the development and validation of an example of a Doppler test object fulfilling the requirements of the IEC 1685. The project has been carried out by nine partners of five European countries and has been funded by the European Commission. The flow Doppler test object is composed of: tissue mimicking material (TMM), blood mimicking fluid (BMF), tube (embedded in the TMM and carrying the BMF), tank flow system, including a pump and a flow meter. In the normative part of the IEC 1685, requirements are given for the values of acoustical parameters of TMM and BMF such as sound velocity, attenuation and backscattering. For BMF, requirements are given also for values of density and viscosity. In an informative (but not compulsory) annex, a description is given of a flow test object meeting these requirements as an example. 'example test object' developed during the project is composed of TMM based on agar and including SiC- and Al2O3-powders, BMF based on nylon particles suspended in water and glycerine, and a tube of c-flex, a silicon copolymer. Two tube sizes are used: 4.0 mm ID and 8.0 mm ID. During the project, very precise recipes have been developed for the composition and preparation of both TMM and BMF. Based on these recipes and a description of the construction in a design five flow test objects have been constructed in the laboratories of five participants. The test objects have been compared by measurements of the physical parameters and by Doppler measurements of the five test objects with the Doppler system. The measurements have been carried out by five observers. Inter-test object and inter-observer variabilities are determined, yielding information about usefulness of the parameters.
A more accurate means of prediction of abdominal aortic aneurysm (AAA) rupture would improve the ... more A more accurate means of prediction of abdominal aortic aneurysm (AAA) rupture would improve the clinical and cost effectiveness of prophylactic repair. The purpose of this study was to determine whether AAA wall distensibility can be used to predict time to rupture independently of other recognized risk factors. A prospective, six-center study of 210 patients with AAA in whom blood pressure (BP), maximum AAA diameter (Dmax), and AAA distensibility (pressure strain elastic modulus [Ep] and stiffness [beta]) were measured at 6 months with an ultrasound scan-based echo-tracking technique. A stepwise, time-dependent, Cox proportional hazards model was used to determine the effect on time to rupture of age, gender, BP, Dmax, BP, Ep, beta, and change in Dmax, Ep, and beta adjusted for time between follow-up visits. Median (interquartile range) AAA diameter was 48 mm (41 to 54 mm), median age was 72 years (68 to 77 years), and median follow-up period was 19 months (9 to 30 months). In the Cox model, female gender (hazards ratio [HR], 2.78; 95% CI, 1.23 to 6.28; P =.014), larger Dmax (HR, 1.36 for 10% increase in Dmax; 95% CI, 1.12 to 1.66; P =.002), higher diastolic BP (HR, 1.13 for 10% increase in BP; 95% CI, 1.13 to 1.92; P =.004), and a decrease in Ep (increase in distensibility) over time (HR, 1.38 for 10% decrease in Ep over 6 months; 95% CI, 1.08 to 1.78; P =.010) significantly reduced the time to rupture (had a shorter time to rupture). Women have a shorter time to AAA rupture. The measurement of AAA distensibility, diastolic BP, and diameter may provide a more accurate assessment of rupture risk than diameter alone.