Comparative Bench Study Evaluation of a Modi ed Snorkeling Mask Used
During COVID-19 Pandemia and Standard Interfaces for Non-invasive
Ventilation.
Giuliano Ferrone ( giulianoferrone@gmail.com )
Fondazione Policlinico Gemelli, IRCCS, Università cattolica Del sacro Cuore https://orcid.org/0000-0001-8371-7127
Giorgia Spinazzola
Fondazione Policlinico Universitario A. Gemelli
Roberta Costa
Università Cattolica del Sacro Cuore: Universita Cattolica del Sacro Cuore
Edoardo Piervincezi
Fondazione policlinico Universitario A. Gemelli
Antonio Gullì
Fondazione Policlinico Universitario A. Gemelli, Roma
Andrea Scapigliati
Fondazione policlinico Universitario A. Gemelli, Roma
Massimo Antonelli
Università Cattolica del Sacro Cuore: Universita Cattolica del Sacro Cuore
Giorgio Conti
Università Cattolica del Sacro Cuore: Universita Cattolica del Sacro Cuore
Research
Keywords: Covid-19, non-invasive ventilation, mechanical ventilation, ARDS
DOI: https://doi.org/10.21203/rs.3.rs-102370/v1
License: This work is licensed under a Creative Commons Attribution 4.0 International License. Read Full License
Page 1/11
Abstract
Background
COVID-19 infection has put enormous pressure on the healthcare systems worldwide and especially on Intensive Care Units (ICUs). In this particular situation,
a modi ed snorkeling mask into a mask for non-invasive continuous positive airway pressure (nCPAP) and non-invasive ventilation (NIV), with the help of 3D
printers, has been proposed for clinical use. We designed a bench study to compare a Helmet (H), a Full face mask (RFF) and a modi ed full face snorkeling
mask (MFF) for delivering nCPAP and NIV in pressure support mode (PSV).
Methods
A mannequin was connected to an active lung simulator. The inspiratory and expiratory variations in airways pressure observed with a high simulated effort
(Pmus), were determined relative to the preset CPAP level. NIV was applied in PSV at two simulated respiratory rates and two cycling-off flow thresholds.
During the bench study we measured the variables de ning patient-ventilator interaction and performance.
Results
During nCPAP, the MFF presented signi cantly lower values of ΔPawi and ∆Pawe compared to the other interfaces tested (H and RFF). During NIV, the MFF
demonstrated a better patient-ventilator interaction compared to RFF, as shown by signi cantly shorter Timepress and Delaytrexp (p<0.01), but no signi cant
differences were found in terms of Delaytrinsp and Timesync between the interfaces tested. At RR 20sim, the MFF presented the shorter ΔPtrigger (p<0.01),
moreover during all the conditions tested the MFF showed the longer PTP 200, 300 and 500 compared to RFF (p<0.01). A major limitation of MFF is that
during NIV with this interface it is possible to reach at maximum 18 cmH2O of peak inspiratory pressure. Over this pressure value, the presence of air leaks
determined important asynchrony phenomena. For this reason, our test was limited to one pressure support and PEEP set (PS 10 cmH2O and PEEP 8 cmH2O).
Conclusions
The modi ed full face snorkeling mask can be used as an acceptable alternative to other interfaces for both nCPAP and NIV in emergency situations.
Introduction
On March 2020, the World Health Organization (WHO) declared the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) outbreak a pandemic, due
to the increasing number of cases reported worldwide1, with high rates of hospitalization and ICU admission 2. The high number of SARS-CoV-2 cases initially
in Lombardy, and subsequently throughout the whole country, made Italy one of the most affected country in Europe 3.
As happened in March 2020, where in Lombardy in few days a total of 1593 patients, affected by severe respiratory failure due to COVID-19, were admitted to
the ICU4, out of a total of 1202 ICU beds available2, the National health care system is now again under pressure from a growing second wave of patients
hospitalized or admitted to ICU for acute respiratory failure .
COVID-19 is characterized by a viral interstitial pneumonia 5 with fever, dry cough, dyspnoea, and bilateral ground-glass opacities6, with about 67% of patients
evolving to a severe pneumonia 7,8.
Preliminary reports described that COVID-19 patients, compared to conventional Acute Respiratory Distress Syndrome (ARDS), are characterized by moderate
to severe hypoxemia despite a relatively high pulmonary compliance 9,10. Due to the enormous number of COVID-19 patients with acute respiratory failure and
to the shortage of ICU beds and ventilators, in many Italian hospital the management of patients with respiratory failure was entrusted to Non-Invasive
Ventilation (NIV) or Non-invasive continuous positive airway pressure (nCPAP).
In particular, in a scenario of discrepancy between facilities and a large number of casualties, as with COVID-19 pandemic, the application of NIV or nCPAP
has been useful as respiratory supportive strategy, especially in patients with mild to moderate ARDS and a PaO2 to Inspired oxygen fraction ratio (PaO2/FiO2)
>150. In our country, the Helmet 4, 10 has been the most widely used device for non-invasive respiratory support during COVID-19 outbreak both in general
wards and in ICU 10.
Unfortunately, a major problem of ventilator and device for NIV shortages rapidly emerged because of the further spread of the virus in other regions of Italy. In
order to relieve the pressure on our national Health System, a device converting a full face snorkeling masks into a mask for CPAP or NIV has been designed
and proposed for clinical use, with the help of 3D printers.
Given the large diffusion of this modi ed full face snorkeling mask in COVID-19 patients, we designed this bench study to evaluate and compare a helmet, a
full face mask and a modi ed full face snorkeling mask in delivering nCPAP, and a modi ed full face snorkeling masks and a full face mask in delivering NIV
in Pressure Support Ventilation mode (PSV).
Materials And Methods
The study was performed at the Respiratory Mechanics Lab (Ventilab) of the Fondazione Policlinico Universitario A. Gemelli IRCCS, Università Cattolica del
Sacro Cuore in Rome, Italy.
Bench Study
Page 2/11
Non-invasive CPAP and non-invasive positive pressure ventilation delivered in PSV mode were applied to a mannequin (LaerdalMedical AS, Stavanger,
Norway) connected to an active test lung system (ASL 5000; Ingmar Medical, Pittsburgh, PA) set using a single-compartment model, an active inspiration
simulated by a semi-sinusoidal pressure waveform (Rise Time 15 %, Pause 0 % and Release Hold 25 %) and the following mechanical properties of the
respiratory system : resistance 5 cmH2O/l/s and compliance 40 ml/cmH2O.
nCPAP was applied via Helmet (H) (CPAP-Castar Starmed, Mirandola, Italy), PerforMax Full face mask (Philips Respironics, Murrysville, PA, USA) (RFF) and a
modi ed full face snorkeling mask (MFF)(SEA VU DRY, Mares Spa, Rapallo, Italy), while non-invasive PSV was delivered through RFF and MFF. The Helmet
used for this bench study is a transparent latex-free polyvinylchloride hood, joined by a rigid plastic ring to a soft collar and secured by two padded armpit
braces at four hooks (two in the front and two in the back of the plastic ring). The helmet used was the size Small to attain a good seal and avoid air-leaks.
The modifed full face snorkeling mask differs from Performax full face mask for shape and design characteristic; it presents a complete separation between
inspiratory and expiratory circuits with the following main features: hypoallergenic silicone mouth-and-nose pocket connected to a polycarbonate transparent
main body; quick-release buckles for easy do ng and a polycarbonate Charlotte valve with an inspiratory and an expiratory channel. See gure 1 for details.
MFF presents two parallel connections with a complete separation between inspiratory and expiratory limbs, while Performax Full face mask is characterized
by a single limb connected to the Y piece. The measure of masks used were medium size for RFF and large size for MFF to attain a good seal and avoid airleaks.
nCPAP (10 cmH2O) was applied at a simulated respiratory rate (RRsim) of 20 breaths per min (b/min) and a simulated level of inspiratory effort (Pmus) of 12
cmH2O, using a standard CPAP device delivering a flow rate of 60 l/min with reservoir (Drager CF 800 Continous Flow CPAP System; Dragerwerk AG & Co,
Lubeck, Germany).
NIV in PSV mode was delivered at 2 RRsim (20 and 30 b/min) and a Pmus of 12 cmH2O with the mechanical ventilator (Puritan Bennet 840; Covidien HealthCare, Mansfield, MA) set in inspiratory pressure support (iPS) of 10 cmH2O, Positive End-Expiratory Pressure (PEEP) of 8 cmH2O, the fastest rate of
pressurization, and a cycling-off flow threshold of 25 and 50 % of the peak inspiratory flow. We set the inspiratory ow trigger at the lowest value not
determining auto-cycling: this threshold was 5L/min during all conditions tested. This setting was chosen for comparing the performance of these interface
under condition of highest pressurization rate and fast or slow cycling-off criteria 11.
Measurements
Air ow (V′) was measured with a pneumotachograph (Fleish No.1, Metabo, Epalinges, Switzerland), while airway pressure (Paw) was measured by a pressure
transducer with a differential pressure of ±100 cmH2O (Digima Clic-1, ICULab system; KleisTek Engineering, Bari, Italy), placed distally from the
pneumotachograph. Air ow (V’) and airway pressure (Paw) at the helmet inlet during the inspiratory phase were measured using a pneumotachograph
(Fleisch n.2; Metabo, Epalinges, Switzerland) and a pressure transducer with differential pressure of ±100 cmH2O (Digima Clic-1; KleisTEK, ICU-Lab System,
Italy) sited at the distal end of the inspiratory limb of the circuit. When the mannequin was ventilated through the RFF, the pneumotachograph and the
pressure transducer were positioned at the Y-connection of the ventilator circuit, instead when we tested the MFF the pneumotachograph and the pressure
transducer were positioned on the inspiratory channel. All these signals were acquired, ampli ed, ltered, digitized at 100 Hz, recorded on a dedicated personal
computer, and analyzed with a speci c software (ICU lab 2.3; KleisTEK Advanced Electronic System, Italy and Analysis Plus).
Each trial lasted 5 minutes; the breaths of the last minute (20 or 30 depending on the trial) were recorded and averaged for analysis.
The measured variables assessed during nCPAP were the maximum inspiratory de ection (∆Pawi, inspiratory drop) and the expiratory peak (∆Pawe),
calculated as differences from the preset CPAP level.
During the NIV test, we evaluated the following variables: Ventilator inspiratory and expiratory time (mechanical TI and mechanical TE, respectively), and
ventilator rate of cycling were all determined on the ow tracing. The inspiratory duty cycle (mechanical TI/Ttot) was calculated as the ratio between
mechanical TI and the total mechanical breath duration (Ttot). Air ow (V′) and tidal volume (VT) delivered to the simulator, airway opening pressure (Paw),
and inspiratory muscles effort were displayed online on the computer screen. The signals obtained with the ASL were transmitted to a PC host via 10/100MBit
Ethernet, sampled and processed in real time by means of speci c software (Lab View, Ingmar Medical). The signals obtained with the ASL were integrated
with the signals from the ICULab system by using a speci c application of the ICULab (ICULab 2.7, KleisTek). The numerical integration of ow over time
determined the mechanical tidal volume (mechanical VT). The amount of tidal volume delivered to the simulator during its active inspiration (ie, the neural
tidal volume, VTneu) was calculated as the volume generated from the onset of inspiratory muscle effort negative de ection to its nadir.
Interfaces performance was evaluated using the following parameters11-13:
1)Trigger pressure drop (ΔPtrigger), de ned as the pressure swing generated by the simulator inspiratory effort in the airway during the triggering phase; 2)
Inspiratory pressure–time product (PTPtrigger), de ned as the area under the Paw curve relative to the time between the onset of inspiratory effort and the
start of mechanical assistance; 3) pressure–time product at 200 ms from the onset of the ventilator pressurization (PTP200), as the index of pure
pressurization performance14; 4)Pressure-time product at 300 ms (PTP300) de ned as the integration of Paw over time during the rst 300 msec and
representing the speediness of the ventilator in reaching the preset level of pressure support; 5) Pressure-time product at 500 ms (PTP500), de ned as the
integral Paw area over insu ation time from the simulated effort onset, representing the ventilator capability of maintaining the pressurization; 6) PTP500
ideal index, expressed as a percentage of the ideal PTP, which is unattainable because it would imply a trigger pressure drop and an instantaneous
pressurization of the ventilator.
Page 3/11
Patient–ventilator interaction was evaluated by determining:
1) Pressurization time (Timepress), de ned as the time necessary to achieve the pre-set level of pressure support from the baseline value; 2) Inspiratory trigger
delay (Delaytrinsp), calculated as the time lag between the onset of inspiratory muscle effort negative swing and the start of the ventilator support (i.e., Paw
positive de ection); 3) Expiratory trigger delay (Delaytrexp), assessed as the delay between the end of the inspiratory effort and the end of the mechanical
insu ations (i.e., ow de ection); 4) Time of synchrony (Timesync), de ned as the time during which inspiratory muscle effort and Paw are in phase (ideally
100%); 5) SimulatorVT/mechanicalVT, intended as the percentage of VT delivered during inspiratory muscle effort negative de ection; 6) The time during
which simulator respiratory effort and ventilator assistance were synchronous, indexed to simulated inspiratory time (Timesync/Tineu) was also computed 1517
.7) Wasted efforts, de ned as ineffective inspiratory efforts, not assisted by the ventilator; 8) Auto-triggering, namely a mechanical insu ation in absence of
inspiratory effort.
Statistical analysis
Continuous data were expressed as mean ± standard deviation (SD). Categorical data were presented as number and percentage in brackets. All variables
were compared with each interface used. Comparison were made by Student’s 𝑡-test and Chi test, as appropriate. The analysis of variance (ANOVA) for
repeated measures was used to detect signi cant differences between the different experimental conditions. When signi cant differences were detected, posthoc analysis was performed using the Bonferroni test; p values < 0.05 were considered statistically signi cant. Statistical analysis was performed using
MEDcalc version 18.6.
Results
As shown in Fig.2, during nCPAP, the MFF demonstrated signi cantly lower ∆Pawi and ∆Pawe values, compared to H and RFF (p< 0.01). Of note, the
differences between H and RFF, in terms of ∆Pawi and ∆Pawe, although statistically significant, were overall small and of limited clinical relevance.
Patient-ventilator interaction described by the measurement of Delaytrinsp, Timepress and Delaytrexp during NIV delivered through MFF and RFF are shown in
Fig 3 and 4. At both RRsim tested and both ventilator settings, the MFF showed a signi cantly shorter Timepress and Delaytrexp compared to RFF (p<0.01).
During all conditions, no signi cant differences were found in terms of Delaytrinsp, Timesync (Fig. 4) and Timesynch/Tineu between all the interfaces tested.
During all study conditions, the volume delivered with MFF was higher than that delivered with the RFF (p<0.01) (Table 1).
The trigger and pressurization performances of the two masks during NIV are shown in Tables 2 and 3. At RRsim 20, the MFF showed a signi cantly shorter
ΔPtrigger than RFF (p<0.01), while no difference was found at RRsim 30. No signi cant differences were found in terms of PTPtrigger between the interfaces
during all tested conditions.
Signi cant differences in PTP 200, PTP 300 and PTP 500 were found between MFF and RFF (p<0.01) in all conditions tested. The MFF demonstrated a
signi cantly higher capacity to maintain the pressurization at 200, 300 and 500 ms after opening the inspiratory valve.
In terms of PTP 500 ideal index, the MFF showed a higher value compared to RFF during all settings, although not statistically signi cant. During all settings,
no asynchrony phenomena were detected with both interfaces.
During NIV test, with the MFF was possible to reach at maximum 18 cmH2O of peak inspiratory applied pressure. Over this pressure, the presence of air leaks
determined important asynchrony phenomena (as autocycling and double triggering), not modi able with change in ow or pressure trigger.
Discussion
In this bench study, during nCPAP, the modi ed full face snorkeling mask presented a best performance compared to the other interfaces tested. Instead,
during NIV, the MFF outperformed the RFF for most of the variables considered and in most of the simulated settings.
During the recent SARS-CoV-2 pandemic, an early non-invasive respiratory support allowed to treat a large number of patients with respiratory distress in order
to prevent ICU admission. For several reasons the CPAP application by Helmet represented the rst choice of treatment in patients with mild to moderate ARDS
COVID-19 related. The Helmet demonstrated to be an effective alternative to face mask in recruiting alveolar units and improve hypoxemia. It can also limit airleaks and room contamination, increasing patient’s comfort, and is more tolerated than the face mask requiring fewer discontinuations 10,18,19.
Unfortunately, the enormous request for helmet CPAP resulted in a rapid lack of supply, so many laboratories, in order to try to ll this gap, started to readjust
snorkeling masks into devices for respiratory support.
We therefore decided to compare the modi ed snorkeling mask with the Helmet and one of the most popular full face mask (PerforMax full face mask), during
CPAP. We applied CPAP at 10 cmH2O because this is a value most commonly applied in this clinical setting 20,21.
In our nCPAP bench study evaluation, the MFF presented signi cantly lower values of ΔPawi and ΔPawe compared to the other interfaces tested, showing a
minor depressurization during the inspiratory and expiratory phases. These results suggest an increased stability in maintaining the PEEP level applied during
nCPAP. Probably, this phenomenon may be due to the low internal volume of MFF, due to the presence of the so-called mouth-and-nose pocket, that reduces
the mask internal volume. In fact, the MFF total internal volume is 1350 ml, but the mouth-and-nose pocket internal volume is only 80 ml, while the RFF
internal volume is 500 ml and the H internal volume is 15000 ml (a real pressurized gas reservoir during inspiration).
Page 4/11
The MFF has been initially proposed as an alternative to the helmet for CPAP, but during the emergency it has been often used also for NIV. Therefore, we
decided to compare this interface with one of the most popular interfaces used in ICU for NIV (RFF). We limited the comparison only to the RFF because, as
previously demonstrated by Chiumello et al.22, in pressure support mode, the mask was more e cient than the helmet. In fact, with the helmet the initial part
of the inspiratory pressure applied is dissipated to pressurize its soft wall. Accordingly, Navalesi et al. 23 demonstrated that the helmet significantly worsens
patient– ventilator synchrony, when compared to the facemask, as indicated by longer delays between inspiratory muscle effort and support delivery, both at
the onset and at the end of inspiration, and by the occurrence of wasted efforts.
For NIV settings, the values of the simulated effort, RRsim, resistance, and compliance, were those already utilized in previous investigations24, 25, 26. The MFF
demonstrated the better simulated patient-ventilator interaction compared to RFF, as shown by shorter Timepress and Delaytrexp. These results are further
validated by interface performance data. Probably the better performance underlined by shorter ΔPtrigger, and longer PTP 200, 300 and 500 can be explained
by the reduced MFF inner volumes and its materials (Figure 1). In fact, a prerogative of this mask is the presence of two separate limbs. The fresh air enters
through the snorkel’s central channel, passing the lens, and keeping it fog-free. It then enters the oral-nasal pocket through non-return valves and ows towards
the nose and mouth. The oral-nasal pocket’s structure guides exhaled CO2 through 4 no-return valves:
2 no-return valves, one each side of the nose, to prevent backward ow,
2 no-return valves at the entrance of the exhalation tubes.
Exhaled CO2 is channeled into the soft silicone side tubes through the second pair of no-return valves. This con guration could explain our results in terms of
simulated patient-ventilator interaction and performance device during NIV settings.
A major limitation of MFF must be underlined: during NIV with this interface it is possible to reach at maximum 18 cmH2O of peak inspiratory applied
pressure. Over this pressure, the presence of air leaks determines important asynchrony phenomena (as autocycling and double triggering), not modi able
with change in ow or pressure trigger. For this reason, our test was limited to a single pressure support and PEEP level (PS 10 cmH2O and PEEP 8 cmH2O).
Conclusion
The results of this comparative bench study suggest that this modi ed snorkeling mask can be used as an alternative to other interfaces for both CPAP and
NIV in emergency situations. The MFF can be proposed in the event of new pandemic surge or for countries where the number of COVID-19 patients is such
that all usable interfaces for non-invasive respiratory support must be used. Particular attention should be paid to the use of this modi ed snorkeling mask
during NIV, related to the maximum pressure limit of 18 cmH20.
Abbreviations
ICUs: Intensive Care Units; nCPAP: Non-invasive continuous positive airway pressure, NIV: Non-invasive ventilation; Pmus: Simulated effort; PSV: Pressure
support ventilation; WHO: World Health Organization; ARDS: Acute respiratory distress syndrome; PaO2/FiO2: Inspired oxygen fraction ratio; RRsim: Simulated
respiratory rate; iPS: Inspiratory pressure support; ∆Pawi: Maximum inspiratory de ection; ∆Pawe: Expiratory peak; TI: Mechanical inspiratory time; TE:
Mechanical expiratory time, VT: Tidal volume; VTneu: Neural tidal volume; Ttot: Total mechanical breath duration; Paw: Airway opening pressure; ΔPtrigger:
Trigger pressure drop; PTPt: Inspiratory pressure–time product; PTP: Pressure time product; Timepress: Pressurization time; Delaytrinsp: Inspiratory trigger delay;
Delaytrexp: Expiratory trigger delay; Timesync: Time of synchrony.
Declarations
Acknowledgements
The authors are grateful to all ICU doctors, residents, and nurses who ght this COVID 19 pandemia with their efforts and passion.
Authors’ contributions
GF, AG, AS, GS and GC conceived the study. GF, GS, AS, AG and EP conducted the bench study. GF and GS analyzed the data. GF, GS and GC interpreted the
data and wrote the rst draft of the manuscript. GC and MA critically revised the manuscript. All the authors reviewed the nal draft of the manuscript and
agreed on submitting it to the Critical Care.
Funding
This study was funded solely by institutional/departmental resources.
Availability of data and materials
The datasets used and/or analyzed during the current bench study are available from the corresponding author on reasonable request.
Ethics approval and consent to participate
Not applicable
Consent for publication
Page 5/11
Not applicable
Competing interests
The authors declare that they have no competing interests
References
1. WHO Director-General' s opening remarks at the media brie ng on COVID-19:11 March 2020. Published March 11, 2020. Accessed March 30, 2020.
https://www.who.int/dg/speeches/detail/who-director-general-s-opening-remarks-at-themedia-brie ng-on-covid-19---11-march-2020
2. Grasselli G, Pesenti A, Cecconi M. Critical care utilization for the COVID-19 outbreak in Lombardy, Italy: early experience and forecast during an emergency
response. JAMA. 2020; 323(16):1545-1546.
3. Coronavirus COVID-19 global cases by the Center for Systems Science and Engineering (CSSE) at Johns Hopkins. Up dated March 30, 2020. Accessed
March 30, 2020.
4. Grasselli G, Zangrillo A, Zanella A, Antonelli M, Cabrini L et al. Baseline Characteristics and Outcomes of 1591 Patients Infected With SARS-CoV-2
Admitted to ICUs of the Lombardy Region, Italy. 2020; 323(16):1574-81.
5. Shi Y, Wang Y, Li X, Huang J, Gan J, Huang X et al. Covid 19 infection: the perspectives on immune response. Cell Death Differ. 2020; 27(5):1451-1454.
6. Huang C, Wang Y, Li X, Lili Ren, Jianping Zhao, Yi Hu et al. Clinical features of patients affected with 2019 novel coronavirus in Wuhan, China. Lancet.
2020; 395: 497-506.
7. Yang X, Yu Y, Shu H, Shu H, Xia J, Liu H et al. Clinical course and outcomes of critically ill patients with Sars-CoV2 pneumonia in Whuan, China: a single
centered, retrospective, observational study. Lancet Respir Med. 2020; 8(5):475-481.
8. Bhrataju PK, Ghassemieh BJ, Nichols M, Kim R, Jerome KR, Nalla AH et al. Covid-19 in critically ill patients in the seattle region-case series. N Eng J Med.
2020; 382(21):2012-2022.
9. Gattinoni L, Coppola S, Cressoni M, Busana M, Rossi S and Chiumello D. Covid-19 does not lead to “typical” acute respiratory distress syndrome. Am J
Respir Crit Care Med. 2020; 201(10):1299-1300.
10. Lucchini A, Giani M, Isgro S, Rona R and Foti G. The “helmet bundle” in COVID-19 patients undergoing non invasive ventilation. Intensive Crit Care Nurs.
2020; 58:102859
11. Olivieri C, Costa R, Spinazzola G, Ferrone G, Longhini F, Cammarota G et al. Bench comparative evaluation of a new generation and standard helmet for
delivering non-invasive ventilation. Intensive Care Med. 2013; 39:734–8.
12. Vignaux L, Vargas F, Roeseler J, Tassaux D, Thille AW, Kossowsky MP et al. Patient–ventilator asynchrony during non-invasive ventilation for acute
respiratory failure: a multicenter study. Intensive Care Med. 2009; 35:840. 21.
13. Vignaux L, Piquilloud L, Tourneux P, Jolliet P and Rimensberger PC. Neonatal and adult ICU ventilators to provide ventilation in neonates, infants, and
children: a bench model study. Respir Care. 2014; 59:1463–75. 22.
14. Olivieri C, Costa R, Conti G, Navalesi P. Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives. Intensive
Care Med. 2012; 38:160–167.
15. Cammarota G, Longhini F, Perucca R, Ronco C, Colombo D, Messina A et al. New setting of neurally adjusted ventilatory assist during noninvasive
ventilation through a helmet. Anesthesiology 2016; 125 (6):1181-9.
16. Olivieri C, Longhini F, Cena T, Cammarota G, Vaschetto R, Messina A et al. New versus conventional helmet for delivering noninvasive ventilation: a
physiologic, crossover randomized study in critically ill patients. Anesthesiology. 2016; 124 (1):101–8.
17. Longhini F, Pan C, Xie J, Cammarota G, Bruni A, Garofalo E et al. New setting of neurally adjusted ventilatory assist for noninvasive ventilation by facial
mask: a physiologic study. Crit Care. 2017; 21(1):170.
18. Antonelli M, Conti G, Pelosi P, Gregoretti C, Pennisi MA, Costa Ret al. New treatment of acute hypoxemic respiratory failure: noninvasive pressure support
ventilation delivered by helmet: a pilot controlled trial. Crit Care Med. 2002; 30(3): 602-608.
19. Antonelli M, Pennisi MA, Pelosi P, Gregoretti C, Squadrone V, Rocco M et al. Noninvasive positive pressure ventilation using a helmet in patients with acute
exacerbation of chronic obstructive pulmonary disease: a feasibility study. Anesthesiology. 2004;100 (1): 16-24.
20. Ferrari G, Oliveri F, De Filippi G, Apri F, Boccuzzi A, Converso M et al. Non-invasive positive airway pressure and risk of myocardial infarction in acute
cardiogenic pulmonary edema: continuous positive airway pressure vs. non-invasive positive pressure ventilation. Chest. 2007; 132:1804–1809 19.
21. Cammarota G, Vaschetto R, Turucz E, Dellapiazza F, Colombo D, Blando C et al. Influence of lung collapse distribution on the physiologic response to
recruitment maneuvers during noninvasive continuous positive airway pressure. Intensive Care Med. 2011; 37:1095–1102.
22. Chiumello D, Pelosi P, Carlesso E, Severgnini P, Aspesi M, Gamberoni C et al. Noninvasive positive pressure ventilation delivered by helmet vs standard
face mask. Intensive Care Med. 2003; 29:1671–1679.
23. Navalesi P, Costa R, Ceriana P, Carlucci A, Prinianakis G, Antonelli M et al. Noninvasive ventilation in chronic obstructive pulmonary disease patients:
helmet versus facial mask. Intensive Care Med. 2006; 33:74–81.
24. Costa R, Navalesi P, Spinazzola G, Rossi M, Cavaliere F, Antonelli M et al. Comparative evaluation of different helmets on patient–ventilator interaction
during non-invasive ventilation. Intensive Care Med. 2008; 34:1102–1108 14.
25. Costa R, Navalesi P, Spinazzola G, Ferrone G, Pellegrini A, Cavaliere F et al. Influence of ventilator settings on patient–ventilator synchrony during pressure
support ventilation with different interfaces. Intensive Care Med. 2010; 36:1363–1370.
Page 6/11
26. Grasselli G, Tonetti T, Protti A, Langer T, Girardis M, Bellani G et al. Pathophysiology of COVID-19-associated acute respiratory distress syndrome: a
multicentre prospective observational study. Lancet Respir Med. 2020 Aug 27: S2213-2600(20)30370-2.
Tables
Table 1: Tidal volume delivered with the different interfaces tested
MFF
RFF
RR 20
RR20
Timepress100%/Trexp50%
P
7.54.33±5.77
676±0.00
0.001
606.33± 5.77
459±5.20
80%
68%
MFF
RFF
RR 30
RR30
Timepress100%/Trexp50%
P
724.33±5.77
586±5.2
417.33±5.77
58%
Timepress100%/Trexp50%
VTmech
MFF
RFF
RR 20
RR20
Timepress100%/Trexp25%
P
8.14.33±5.77
694±0.00
0.001
0.001
603±0.00
459±5.2
0.001
0.001
74%
66%
0.001
MFF
RFF
RR 30
RR30
Timepress100%/Trexp25%
Timepress100%/Trexp25%
0.001
739.67±6.35
566.33±5.77
0.001
304±9.00
0.001
378±0.00
276±0.00
0.001
52%
0.001
51%
49%
0.02
Timepress100%/Trexp25%
(ml)
VTneu
(ml)
VTneu/VTmech
(%)
Timepress100%/Trexp50%
VTmech
P
(ml)
VTneu
(ml)
VTneu/VTmech
(%)
VTmech: mechanical tidal volume; VTneu: neural tidal volume; VTneu/VTmech: the percentage of tidal volume delivered during inspiratory simulated muscle
effort negative de ection.
MFF: Modi ed Mares Full Face Mask, RFF: Respironics PerforMax Full face mask.
RR Respiratory rates.
Table 2: Performance of the interfaces during NIV at 20 RRsim
MFF
RFF
MFF
RFF
RR 20
RR20
Timepress100%/Trexp50%
P
RR 20
RR20
Timepress100%/Trexp25%
P
1.05±0.01
0.72±0.05
0.004
1.03±0.05
0.78±0.04
0.014
0.04±0.01
0.03±0.01
1
0.05±0.00
0.03±0.00
0.001
0.88±0.05
0.57±0.09
0.013
0.87±0.12
0.57±0.05
0.04
2.04±0.08
1.22±0.09
0.004
1.97±0.03
1.21±0.05
0.001
4.49±0.09
2.66±0.11
0.001
4.35±0.03
2.67±0.03
0.001
72
56
0.45
71
51
0.5
Timepress100%/Trexp50%
ΔPtrigger
Timepress100%/Trexp25%
(cmH2O)
PTPt
(cmH2O/s)
PTP200
(cmH2O/s)
PTP300
(cmH2O/s)
PTP500
(cmH2O/s)
PTP500 ideal
index (%)
Page 7/11
ΔPtrigger: trigger pressure drop, PTPt: Pressure Time Product during the triggering phase, Paw: airway pressure, PTP200, PTP300 and PTP500: Pressure Time
Product during the initial 200, 300 and 500 ms from the onset of the ventilator pressurization expressed as the absolute value, PTP500 ideal index: Pressure
Time Product during the initial 500 ms from the onset of the simulated effort, expressed as the percentage of the area of ideal pressurization, with different
ventilator settings (see text).
MFF: Modi ed Mares Full Face Mask, RFF: Respironics PerforMax Full face mask.
RR Respiratory rates.
Table 3:Performance of the interfaces during NIV at 30 RRsim
MFF
RFF
MFF
RFF
RR 30
RR30
Timepress100%/Trexp50%
P
RR 30
RR30
Timepress100%/Trexp25%
P
1.28±0.04
1.75±0.47
0.11
1.94±0.23
2.05±0.11
0.31
0.06±0.01
0.10±0.04
0.12
0.11±0.02
0.13±0.01
0.16
0.98±0.07
0.56±0.10
0.02
0.63±0.15
0.32±0.03
0.04
2.26±0.07
1.29±0.10
0.002
1.91±0.08
1.07±0.10
0.006
5.09±0.03
3.05±0.12
0.007
4.69±0.16
2.77±0.11
0.003
71
56
0.22
68
48
0.3
Timepress100%/Trexp50%
ΔPtrigger
Timepress100%/Trexp25%
(cmH2O)
PTPt
(cmH2O/s)
PTP200
(cmH2O/s)
PTP300
(cmH2O/s)
PTP500
(cmH2O/s)
PTP500 ideal
index (%)
ΔPtrigger: trigger pressure drop, PTPt: Pressure Time Product during the triggering phase, Paw: airway pressure, PTP200, PTP300 and PTP500: Pressure Time
Product during the initial 200, 300 and 500 ms from the onset of the ventilator pressurization expressed as the absolute value, PTP500 ideal index: Pressure
Time Product during the initial 500 ms from the onset of the simulated effort, expressed as the percentage of the area of ideal pressurization, with different
ventilator settings (see text).
MFF: Modi ed Mares Full Face Mask, RFF: Respironics PerforMax Full face mask.
RR Respiratory rates
Figures
Page 8/11
Figure 1
Mares snorkeling mask (MFF) features: presence of two separate limbs; the inspiratory one is highlighted by blue arrows, the expiratory one by red arrows. The
adapter created with 3 d printers features two channels: the central upward-pointing channel is the inspiratory channel, the side channel is the expiratory
channel.
Figure 2
Page 9/11
Maximum inspiratory de ection (∆Pawi, inspiratory drop) and the expiratory peak (∆Pawe), during CPAP with Helmet (light grey column), PerforMax Full face
mask (dark grey column) and Mares Mask (black column) at 20 breaths/min.
Figure 3
Inspiratory trigger delay (Delaytrinsp), Expiratory Trigger delay (Dealytrexp), and Pressurization Time (Timepress) with the Mares Mask (MFF) (black column)
and the PerforMax Full face mask (RFF) (grey column) at 20 breaths/min, with 2 ventilator settings.
Page 10/11
Figure 4
Inspiratory trigger delay (Delaytrinsp), Expiratory Trigger delay (Dealytrexp), and Pressurization Time (Timepress) with the Mares Mask (MFF) (black column)
and the PerforMax Full face mask (RFF) (grey column) at 30 breaths/min, with 2 ventilator settings.
Figure 5
Time of synchrony with the Mares Mask (MFF) (black column) and the PerforMax Full face mask (RFF) (grey column) at two respiratory rates (RR 20 and 30
breaths/min), with 2 ventilator settings.
Page 11/11