INSTITUTE OF PHYSICS PUBLISHING
PHYSIOLOGICAL MEASUREMENT
Physiol. Meas. 27 (2006) R1–R35
doi:10.1088/0967-3334/27/1/R01
TOPICAL REVIEW
Pulse oximetry in the oesophagus
P A Kyriacou
School of Engineering and Mathematical Sciences, City University, London EC1V 0HB, UK
E-mail: P.Kyriacou@city.ac.uk
Received 27 July 2005, accepted for publication 31 August 2005
Published 29 November 2005
Online at stacks.iop.org/PM/27/R1
Abstract
Pulse oximetry has been one of the most significant technological advances
in clinical monitoring in the last two decades. Pulse oximetry is a noninvasive photometric technique that provides information about the arterial
blood oxygen saturation (SpO2) and heart rate, and has widespread clinical
applications. When peripheral perfusion is poor, as in states of hypovolaemia,
hypothermia and vasoconstriction, oxygenation readings become unreliable or
cease. The problem arises because conventional pulse oximetry sensors must
be attached to the most peripheral parts of the body, such as finger, ear or toe,
where pulsatile flow is most easily compromised. Since central blood flow
may be preferentially preserved, this review explores a new alternative site, the
oesophagus, for monitoring blood oxygen saturation by pulse oximetry. This
review article presents the basic physics, technology and applications of pulse
oximetry including photoplethysmography. The limitations of this technique
are also discussed leading to the proposed development of the oesophageal
pulse oximeter. In the majority, the report will be focused on the description of
a new oesophageal photoplethysmographic/SpO2 probe, which was developed
to investigate the suitability of the oesophagus as an alternative monitoring site
for the continuous measurement of SpO2 in cases of poor peripheral circulation.
The article concludes with a review of reported clinical investigations of the
oesophageal pulse oximeter.
Keywords: pulse oximetry, photoplethysmography, perfusion, oesophagus
1. Introduction
Pulse oximetry has been one of the most significant technological advances in clinical
monitoring in the last two decades (Alexander et al 1989, Bowes et al 1989, Anonymous
2003, Tremper and Barker 1989, Welch 2005). Pulse oximetry is a non-invasive photometric
technique that provides information about the arterial blood oxygen saturation (SpO2) and
0967-3334/06/010001+35$30.00 © 2006 IOP Publishing Ltd Printed in the UK
R1
R2
Topical Review
heart rate, and has widespread clinical applications. The use of pulse oximeters has been
described in many settings: hospital, outpatient, domiciliary use and in veterinary clinics.
In the early 1990s pulse oximetry became a mandated international standard for monitoring
during anaesthesia following the publication in 1986 of the Harvard minimum standards for
monitoring. Kelleher (1989) reviewed 220 references in an article published in 1989. In a
follow-up review in 1992, Severinghaus and Kelleher (1992) found more than 500 new reports
between 1989 and October 1991. Nearly 5000 further reports on pulse oximetry have been
published since October 1991.
Although generally reliable, pulse oximeters do fail, in particular, in patients undergoing
prolonged procedures such as cardiac, vascular, reconstructive or neuro-surgery, at just the time
when the measurement of blood oxygen saturation would be clinically of most value (Ralston
et al 1991, Reich et al 1996). Many of their limitations, both physiological and technical,
will be discussed in this review. The hypothesis underlying this review is that a more central
site, such as the oesophagus, will remain adequately perfused in the above-mentioned clinical
situations, giving the possibility of monitoring SpO2 at the oesophagus when conventional
peripheral oximetry fails.
This review will first outline the basic physics, technology and applications of pulse
oximetry including photoplethysmography. The limitations of this technique are also discussed
leading to the proposed development of the oesophageal pulse oximeter. In the majority, the
report will be focused on the description of a new oesophageal photoplethysmographic/SpO2
probe, which was developed to investigate the suitability of the oesophagus as an alternative
monitoring site for the continuous measurement of SpO2 in cases of poor peripheral circulation.
The technological developments of such a pulse oximeter and the results from clinical
investigations will be presented.
2. Photoplethysmography and pulse oximetry
Photoplethysmography is a non-invasive optical technique widely used in the study and
monitoring of the pulsations associated with changes in blood volume in a peripheral vascular
bed (Roberts 1982, Dorlas and Nijboer 1985, Higgins and Fronek 1986, Lindberg and
Oberg 1991). Whether the term ‘plethysmography’ is a misnomer is a matter of debate,
yet the title has received general consent. Challoner (1979) made an excellent review of
photoplethysmography. As discussed in Challoner’s review, Hertzman in 1937 first coined
the term plethysmograph. It was pointed out above that there is not total agreement that this
is a strictly accurate name. An etymological definition would suggest that a plethysmograph
records volume; thus, volumetric changes are recorded in the blood vessels of an organ.
However, whether photoplethysmography measures only blood volume changes is open to
question. The origin of the photoplethysmographic (PPG) signal has been the subject of
continuing debate (Challoner 1979, Roberts 1982).
In photoplethysmography the emitted light, which is made to transverse the skin, is
reflected, absorbed and scattered in the tissue and blood. The modulated light level which
emerges, is measured using a suitable photodetector. It is possible for the hand or finger
to be directly transilluminated where the light source, usually in the region of 800 nm to
960 nm, is on one side of the skin and the detector on the other side. This method, also
called transmission mode, is limited to areas such as the finger, the ear lobe or the toe (Nijboer
et al 1981, Mendelson and Ochs 1988). However, when light is directed down into the skin a
proportion of this is backscattered so that it emerges from the skin adjacent to the light source.
The light source and the photodetector can be positioned side by side. This method, also
called the reflection mode, allows measurements on virtually any skin area (Nijboer et al 1981,
Topical Review
R3
Figure 1. Photoplethysmographic (PPG) waveform as measured by transmission through tissue.
Mendelson and Ochs 1988). The intensity of the transmitted or reflected light which reaches
the photodetector is measured and the variations in the photodetector current are assumed
to be related to blood volume changes underneath the probe (Nijboer et al 1981, Roberts
1982). These variations are electronically amplified and recorded as a voltage signal called
the photoplethysmograph. The photoplethysmographic signal (figure 1) is divided into two
components:
(i) A dc PPG component, a relatively constant voltage offset of which the magnitude is
determined by the nature of the material through which the tissue passes (skin, cartilage,
venous blood, etc).
(ii) A pulsatile or ac PPG component synchronous with the heart rate is often assumed to
be related to the arterial blood volume pulse. The ac PPG pulse shapes are indicative of
vessel compliance and cardiac performance.
Pulse oximeters, as will be discussed in more detail in the following sections, estimate
arterial blood oxygen saturation by shining light at two different wavelengths, red and infrared,
through vascular tissue. In this method, the ac pulsatile PPG signal associated with cardiac
contraction is assumed attributable solely to the arterial blood component. The amplitudes
of the red and infrared ac PPG signals are sensitive to changes in arterial oxygen saturation
because of differences in the light absorption of oxygenated and deoxygenated haemoglobin
at these two wavelengths. From the ratios of these amplitudes, and the corresponding dc
photoplethysmographic components, arterial blood oxygen saturation (SpO2) is estimated.
Hence, the technique of pulse oximetry relies on the presence of adequate peripheral arterial
pulsations, which are detected as photoplethysmographic signals (Mendelson and Ochs 1988,
Webster 1997).
3. Physics and technology of pulse oximetry
Theoretical descriptions of pulse oximetry often begin with a discussion of the Beer–Lambert
law of light absorption (Webster 1997). It is beyond the scope of this review to describe in
detail the Beer and Lambert approach, as it has been discussed and explained in detail in
the literature (Webster 1997). It will be useful though to note that by itself this approach is
incomplete, as it does not adequately account for the effects of a physical phenomenon called
light scattering present within the tissue region under investigation. Biological tissue is a
highly light-scattering medium and little information about its optical properties is available
R4
Topical Review
Figure 2. Extinction coefficients (Lmmol−1 cm−1) of the four most common haemoglobin
species; oxyhaemoglobin (HbO2), deoxyhaemoglobin (Hb), carboxyhaemoglobin (COHb) and
methemoglobin (MetHb) at the wavelengths of interest in pulse oximetry.
(Shimada et al 1984, Delpy 1988, Wray et al 1988, Schmitt 1991, Shvartsman and Fine 2003,
Finlay and Foster 2004). In conventional practice, the effects of light scattering are accounted
for by empirically calibrating the sensor and oximeter, and this appears to work well, but only
up to a certain point (Fine and Weinreb 1995, Webster 1997). Assumptions inherently made
during an empirical calibration are valid only for a limited range of saturations, and become
invalid under extreme conditions. Nevertheless, the Beer–Lambert law approach helps to
develop an understanding of the absorbance of light as it passes through living tissue and why
and how pulse oximetry works.
3.1. Wavelengths for pulse oximetry
Pulse oximeters determine arterial blood oxygen saturation by measuring the light absorbance
of tissue at two different wavelengths and using the arterial blood pulsation to differentiate
between absorbance of arterial blood and other absorbers (skin, bone, venous blood). A
good choice of wavelength is where there are large differences in the extinction coefficients
of deoxyhaemoglobin and oxyhaemoglobin (Mannheimer et al 1997a) (figure 2). Another
criterion for the wavelength selection is the relative flatness of the absorption spectra around the
chosen wavelength (Moyle 1994, Mannheimer et al 1997b). The two conventional wavelengths
used in pulse oximetry are 660 nm (red) and 940 nm (near infrared).
3.2. Calibration of commercial pulse oximeters
It could be considered that accurate values of SpO2 can be obtained by application of
simultaneous equations describing the Beer–Lambert law, but this is not the case, and most
modern pulse oximeters apply the ‘red: infrared ratio’ (R) to a ‘look-up table’ (Moyle 1994).
R(ratio) =
ac660 /dc660
.
ac940 /dc940
(1)
Topical Review
R5
Figure 3. Empirical relationship between arterial saturation and ‘red:infrared’ ratio (R).
The manufacturers calibrate pulse oximeters empirically by correlating the measured ratio
(R) of ac/dc signals from the red and infrared photoplethysmographs obtained from a large
group of healthy volunteers with arterial oxygen saturation (SaO2) values generally greater
than 70%. Blood SaO2 values are obtained directly from a standard in vitro CO-oximeter. This
calibration procedure involves desaturating the subjects by asking them to breathe hypoxic
gas mixtures and collecting optical measurements of blood samples at different steady-state
oxygenation levels. Consequently, different brands of pulse oximeters may display slightly
different values, depending on the internal calibration of the oximeter. A typical relationship
between the ratio (R) and SpO2 is shown in figure 3.
One of the limitations of this traditional calibration method is the limited range of oxygen
saturation that can be acquired. Ethical issues prevent intentional desaturation of healthy
subjects below a certain point due to the risk of hypoxic brain damage.
3.3. Technical developments of commercial pulse oximeters
A typical commercial pulse oximeter consists of an opto-electronic probe that operates either
in reflectance or in transmittance mode and a microprocessor-controlled electronic system
(figure 4). The technology of pulse oximetry and the advancements in signal processing for
eliminating motion artefact or dealing with low amplitude PPGs is well reviewed (Pologe
1987, Wukitsch et al 1988, Severinghaus 1993, Webster 1997, Goldman et al 2000). This
section describes briefly the main blocks comprising the pulse oximeter system, such as the
analogue and digital processing, the microprocessor and the front display (figure 4).
The pulse oximeter probe has a single photodetector receiving signals from the infrared
(IR) and red (R) emitters. The currents through the emitters are controlled by the emitter driver,
which comprises a pair of current sources (R and IR) and a multiplexer. Emitter drive currents
may vary between 40 mA and 120 mA. The multiplexer turns the red and infrared emitters on
and off at a rate of approximately 325 Hz in sequence (but varies with different manufacturers)
(Webster 1997). The multiplexed PPG signals from the photodetector are received by a
transimpedance amplifier that converts the current output into a signal voltage. The mixed PPG
signals are separated into red and infrared with the use of a demultiplexer. The signals are then
R6
Topical Review
Red
Infrared
(940nm) (660nm)
Current
source
R
Current
source
R
Emitter Driver
Timing signal
Multiplexer
Photodetector
Demultiplexer
ADC
transimpedance
amplifier
Red
MICROPROCESSOR
Infrared
Digital display
SpO2
Heart rate
Figure 4. Basic block diagram of a commercial transmission mode pulse oximeter.
converted by an analogue-to-digital converter (ADC) into a form suitable for manipulation by
the microprocessor. The digitized information is processed by the microprocessor to compute
the blood oxygen saturation from the ratio (R) derived from the signal at the red wavelength
compared with the signal at the infrared wavelength. Pulse oximeters usually display blood
oxygen saturation (SpO2) in per cent together with heart rate, and photoplethysmograph.
4. Pulse oximeter probes and their applications
4.1. Pulse oximetry probes
The probe of the original pulse oximeter as described by Yoshiya et al (1980) was based on
a bulky fibre optic cable. The fibre optic cable in this implementation was used only as a
guide to conduct light from a quartz halogen lamp to the remote measurement side and to
conduct the light transmitted through the tissue back to the photodetector. The light source
and the photodetector were both housed inside the oximeter. Narrow-bandpass interference
optical filters were used in combination with a mechanical chopper to select properly the red
and infrared wavelengths (Mendelson 1992). An improved design of a non-invasive pulse
oximeter probe was introduced in the United States in the early 1980s (figure 5).
This much simpler design, which dominates most of the commercial pulse oximeters
nowadays, consists of a pair of small red and infrared emitters and a single highly sensitive
silicon photodetector mounted inside either a reusable spring loaded clip (figure 5), or a
disposable adhesive wrap (Mendelson 1992). A flexible cable connecting the probe and the
pulse oximeter unit carries electric power to the emitters and the signal from the photodetector.
A synopsis of different types of pulse oximetry probes used today in the clinical setting is
presented below.
4.1.1. MRI probes. When a pulse oximeter is used during magnetic resonance imaging
(MRI), the very high magnetic field strengths involved with this imaging modality may
give erroneous readings or no readings at all (Wahr et al 1995). To overcome the
problem manufacturers have produced pulse oximeters where all of the electronic and optical
components are in the housing of the main unit. The light energy is directed to and from the
patient by optical fibres. The complete pulse oximeter is kept beyond the influence of the
magnetic field (approximately 3 m) of the MRI scanner (Webster 1997).
Topical Review
R7
Figure 5. Typical reusable spring loaded clip pulse oximetry probe.
4.1.2. Foetal probes. This is an area of continued research since no probe yet is used in
routine foetal monitoring (Dildy et al 1994, Dildy 2004). Several designs of foetal pulse
oximetry probes have been developed over the last few years. The probes are guided into the
cervix and must be placed beyond the presenting part and the transcervical region (just beyond
the cervix). The probe is placed on the temple of the foetus and therefore has less interference
from hair (Webster 1997).
4.1.3. Reusable probes. Generally, all probes with nonadhesive or disposable adhesive
sensors are reusable probes. The main advantage of reusable pulse oximetry probes is the low
use cost involved. However, reusable probes require cleaning between patients to minimize
the risk of cross contamination (Kelleher 1989).
4.1.4. Disposable probes. In the past few years, many pulse oximeter manufacturers have
produced disposable probes. One of the advantages of disposable probes is the elimination of
any form of cross contamination between patients since disposable probes are used on a single
patient.
4.2. Applications of pulse oximetry
Pulse oximeters, as has been discussed above, are non-invasive, easy to use and readily
available. Due to these characteristics, they have an abundance of clinical applications. This
may be seen as an overstatement, but as more and more pulse oximeters are coming in use,
more and more hypoxaemic events are being seen as precursors of pathological events. Some
of the main areas in which they are used are anaesthesia, patient transport, childbirth, neonatal
and paediatric care, dentistry and oral surgery, sleep studies and many other applications.
Some of them are described briefly.
4.2.1. Anaesthesia. The pulse oximeter is most often and most importantly used when
anaesthesia is given. Several studies have shown that desaturation is often a problem
R8
Topical Review
during induction of anaesthesia (Drummond and Park 1984). This is the most critical
time of general anaesthetic because of the side effects of the induction agents, onset of
neuromuscular blockade, loss of protective reflexes, mechanical problems with the airway,
respiratory depression and problems with tracheal intubation.
4.2.2. Emergency medicine. Anaesthetists were quick to see the enormous benefit to the
safety of the patient of using pulse oximeters, and doctors, nurses and medical technicians
involved in emergency care soon recognized the advantage of pulse oximetry (Lambert and
Crinnion 1989). A retrospective study by Anderson et al (1988) showed that failure to
recognize hypoxaemia has been identified as one of the major avoidable causes of death
in trauma patients. Pulse oximeters are now regularly used in accident and emergency
departments and in the pre-hospital care of the sick and injured.
4.2.3. Postoperative recovery. The period between the end of surgery and when the patient
is fully conscious is often when hypoxaemia is most likely to go unnoticed. The greatest risk
during this period is respiratory failure. For these reasons the pulse oximeter must be used
regularly in the postoperative recovery phase (Tyler et al 1985, Pullerits et al 1987, Kelleher
1989, Moyle 1994).
4.2.4. Childbirth. The process of labour and delivery is a stressful time for both the mother
and the foetus. If the foetus does not have sufficient metabolic reserve to withstand this
ordeal, it is at risk of hypoxia and of sustaining subsequent brain damage (Minnich et al 1990).
Many difficulties have been encountered when attempting to monitor foetal pulse oximetry.
The obvious problem is that the foetus is not accessible. The potential application of pulse
oximetry to foetal monitoring during labour has been demonstrated by several groups (Gardosi
et al 1991, Klauser et al 2005).
4.2.5. Neonatal and paediatric care. Low and high arterial oxygen levels can both be
damaging to newborn infants. Infants who are hypoxic may develop organ damage, and
hypoxia may also cause pulmonary hypertension (Moyle 1996). The great concern of neonatal
paediatricians is to prevent retinopathy of prematurity, which is caused by high levels of retinal
oxygenation and may lead to blindness (Moyle 1994). It is, therefore, essential to monitor
accurately the oxygen levels of sick and premature newborn infants.
4.2.6. Dentistry and oral surgery. Many papers have recommended the use of pulse oximetry
during general anaesthesia for dental procedures, pointing out that pulse oximeters should be
applied to all patients even for short procedures (Hovagim et al 1989).
4.2.7. Sleep studies and exercise. Many people become desaturated during sleep or heavy
exercise. The cause of desaturation during sleep is due to a disorder known as sleep apnoea
(Webster 1997). Desaturation can occur during heavy exercise due to poor ventilation or
chronic obstructive pulmonary disease (COPD). The use of pulse oximetry during sleep and
exercise aids in the diagnosis of these respiratory problems (Trang et al 2004).
4.3. Future applications of pulse oximetry
Although pulse oximetry seems to be at the peak of its development, there is always room
for further improvement and optimization. Many of these improvements relate to specific
Topical Review
R9
applications. Improvements, which will increase the performance of pulse oximetry during
transport, are to lengthen the battery life in portable units. Reducing the occurrence of false
alarms would be beneficial in all applications, but especially during long-term monitoring when
staff cannot always be in the room. The application of foetal pulse oximetry is an exciting
on-going research by many groups (Dildy 2001, Reuss 2004). In this unique application,
sensor attachment to the foetal scalp must be improved. Other potential applications of pulse
oximetry have been suggested for monitoring oxygen saturation via the retinal fundus in the
eye (De Kock et al 1993, Nelson et al 2005). The direct application of pulse oximetry to
an organ, such as the liver, the kidney, the gut or the colon, will be a very useful application
in determining organ SpO2, regardless if the patients SpO2 as measured from an extremity
(finger) is normal (Denobile et al 1990, Yilmaz et al 1999, Crerar-Gilbert et al 2002).
5. Limitations of pulse oximeters
5.1. Calibration assumptions
Initially, the conversion from absorbancy ratios to arterial oxygen saturation as described in
previous sections is based on experimentally derived calibration curves (figure 3). These
curves are based on measurements in healthy young volunteers after induction of hypoxaemia
(Moyle 1994, Sinex 1999). An unavoidable limitation is that pulse oximeters can only be
as accurate as their empirical calibration curves. Understandably, researchers were limited
in the degree of haemoxaemia inducible in these volunteers, to a SaO2 of approximately
75% to 80%. Therefore the shape of the curve (figure 3) below these levels (75% to 85%)
must be extrapolated, with obvious implications for the accuracy of pulse oximetry at low
saturation levels. Studies showed such great inaccuracy and bias at low oxygen saturation that
manufacturers revised early calibration curves and software (Severinghaus and Naifeh 1987).
More studies, however, continued to show significant bias, increasing as oxygen saturation
decreases, although it has been justifiably pointed out that few, if any, clinical treatment
decisions will hinge on whether the oxygen saturation is actually 50% or 60% (Kelleher
1989).
5.2. Dyshaemoglobinaemias
Many substances in the blood can interfere optically with pulse oximetry. This interference
generally takes the form of false absorbers, or components besides deoxyhaemoglobin or
oxyhaemoglobin that will absorb light within the red and near-infrared wavelengths used
in pulse oximetry. The most significant potential false absorbers in the circulation are
carboxyhaemoglobin (COHb) and methaemoglobin (MetHb) (Kelleher 1989, Richardson
2005). Being two-wavelength devices, pulse oximeters can only deal with two haemoglobin
species (deoxyhaemoglobin and oxyhaemoglobin). Therefore, both COHb and MetHb will
cause errors in the pulse oximeter readings (Barker and Tremper 1987, Barker et al 1989).
When the presence of these Hb species is suspected, pulse oximetry should be supplemented
by in vitro multiwavelength CO-oximetry.
5.3. Intravenous dyes
Intravenous dyes are known to have potentially profound effects on pulse oximetry readings,
resulting in falsely low measured oxygen saturations. Methylene blue, with very high
absorbance at 660 nm, causes spurious decrease in SpO2 (Kessler et al 1986, Scheller et al
1986, Sidi et al 1987).
R10
Topical Review
5.4. Anaemia
Anaemia appears to adversely affect the accuracy of pulse oximetry, although the mechanism
is unclear, and it may do so only in the presence of hypoxia. In theory, anaemia should
not affect pulse oximetry at all, as the ratio of relative absorbances should be preserved and
unchanged by changes in total haemoglobin concentration within the sample (Severinghaus
and Koh 1990). A study of non-hypoxic human patients with acute anaemia showed good
accuracy with pulse oximetry (Jay et al 1994).
5.5. Skin pigmentation and nail polish
Skin pigmentation has shown variable effects on pulse oximetry (Volgyesi and Spahr-Schopfer
1991, Bickler et al 2005). Primarily, dark pigmentation appears to make adequate light
penetration more difficult, with significantly more signal detection failures. More concerning
are some studies of pigmentation effects which have shown overestimation of oxygen
saturation, both in models and in human subjects (Ries et al 1989). Similarly, there is
evidence for the effects of nail polish on pulse oximetry, but little consensus on occurrence or
degree (Kataria and Lampkins 1986, Cote et al 1988, Chan et al 2003).
5.6. Electromagnetic interference
Electromagnetic interference (EMI) can affect the accuracy of pulse oximeters and other
medical devices. It may be generated by many sources, mostly man made but it may also
result from atmospheric events.
5.7. Interference due to electrocautery
Electrocautery can interfere with pulse oximetry by artifactually decreasing SpO2 readings,
or by setting off false alarms (Wahr et al 1995). The cause is the wide spectrum of radio
frequency emissions picked up directly by the photodetector in the pulse oximeter probe
(Block and Detko 1986). New signal extraction technology (MASIMO) claims to be more
accurate in the presence of electrocautery (Wahr et al 1995).
5.8. Signal artefact
Most commonly, problems in pulse oximetry arise from signal artefact. Signal artefact
results from false sources of signal or from a low signal-to-noise ratio. False signal can
arise from detection of non-transmitted light (ambient sources or optical shunt) or from nonarterial sources of alternating signal. A low signal-to-noise ratio results from inadequate
signal complicated by an excess of physiological or technical noise. The oximetry system
as outlined in the previous sections assumes that the sum of the light absorbed and the light
transmitted is equal to the incident light, with no other light loss or gain affecting the detector.
Ambient light, however, is potentially a major source of interference (Fluck et al 2003).
Recognizing this, designers of pulse oximeters divided the emitter and detector activities into
three sensing periods, cycling at hundreds of times per second. Two of these periods use light
emitted by the emitters at each of the two incident wavelengths. In the third period neither
emitter is activated and the photodiode detector measures only ambient light, the influence
of which is subsequently eliminated from the emitter-illuminated sensing periods. However,
cases of ambient light interference still occur (Hanowell et al 1987). Implicated sources
include fluorescent lighting, surgical lamps, fiber optic instruments and sunlight. Covering
Topical Review
R11
the probe with an opaque shield offers a simple solution. Optical shunting occurs when light
from the emitters reaches the photodetector without passing through an arterial bed. This
occurs most commonly with inappropriate probe selection, as when a digit probe is placed
on the ear lobe, or with probe misplacement (Barker et al 1993, Webster 1997). In both
ambient light interference and optical shunt, the result is the addition of false signal to both
wavelengths emitted (Kelleher and Ruff 1989). Non-arterial sources of alternating signal
most commonly result from motion artefact. Motion artefacts, such as during shivering,
seizure activity or exercise, are usually recognized by false or erratic heart rate displays or
by distorted photoplethysmographic waveforms (Severinghaus and Kelleher 1992, Hanning
and Alexander-Williams 1995). When interference does occur from motion artefact it again
tends to be additive into both the red and infrared wavelength channels (Pologe 1987), with
a resultant absorbance ratio (R) becoming nearer to 1 and an SpO2 approaching 85% (see
figure 3).
Although different oximeter models employ different data processing, the saturation
values acquired per second are typically averaged over a period of 2–16 s before a reading
is given, serving in part to limit the impact of motion artefact (Kidd and Vickers 1989).
Newly developed signal processing algorithms by MASIMO technologies have demonstrated
successfully that can minimize or eliminate erroneous estimation of SpO2 due to movement
artefact (Goldman et al 2000).
5.9. Inadequate pulsatile signals
Apart from the physiological and technical limitations of pulse oximeters described in this
review, they have also been reported to fail in patients with compromised peripheral perfusion
(Severinghaus and Spellman 1990, Freund et al 1991, Moller et al 1993, Reich et al 1996).
Pulse oximetry is a pulse-dependent technique, and any significant reduction in the amplitude
of the pulsatile component of the photoplethysmographic signal can lead to dubious values for
blood oxygen saturation (SpO2) or complete failure. Hence, pulse oximeters require adequate
peripheral perfusion to operate accurately. When peripheral perfusion is poor, as in states
of hypovolaemia, hypothermia and vasoconstriction, oxygenation readings become unreliable
or cease (Palve and Vuori 1989, Palve 1992a, 1992b). Such clinical situations occur, for
example, after prolonged operations, especially hypothermic cardiopulmonary bypass surgery.
The problem arises because conventional pulse oximetry sensors must be attached to the most
peripheral parts of the body, such as finger, ear or toe, where pulsatile flow is most easily
compromised. Measurements at sites other than the finger or ear, such as the forehead and
nose, give no improvement in poorly perfused patients (Rosenberg and Pedersen 1990, Clayton
et al 1991). Thus, SpO2 readings are often unobtainable at just the time when they would be
most valuable.
Therefore, the question becomes what to do in such cases, particularly those in which
the oximeter is unable to find an adequate pulse at any of the available peripheral sites.
There is, therefore, a need to find a means of solving this frustrating and serious clinical
problem.
6. Oesophageal pulse oximetry
In an attempt to overcome the difficulties associated with conventional measurements of
arterial blood oxygen saturation during conditions of poor peripheral perfusion and pulsation,
the oesophagus has been proposed as a potential measurement site on the hypothesis that
R12
Topical Review
Figure 6. The reflectance transoesophageal probe (oesophageal probe is an 18-gauge stethoscope
with a temperature sensor at the distal tip and an oximetry sensor approximately 7 cm proximally).
perfusion may be better preserved at this central site. The oesophagus is perfused directly by
main arteries. The cervical oesophagus is supplied mainly by branches of the inferior thyroid
artery. The chief blood supply of the thoracic oesophagus is either by branches of the bronchial
arteries, or direct from the aorta, and there are branches of the left gastric and inferior phrenic
arteries running along the surface of the lower oesophagus for 2–3 cm above the diaphragm
(Romanes 1972, Marieb 1992).
Atlee and Bratanow (1995) proposed to use the upper oesophagus as a measurement
site. They presented results of blood oxygen saturation measurements obtained at the
cricopharyngeus muscle in the upper oesophagus (14 ± 1 cm from incisors) using a reflectance
‘transoesophageal’ probe where the optical components of the oximetry sensor were
incorporated into a traditional anaesthesia oesophageal stethoscope (figure 6). Technical
details of the probe or processing system are not available in the literature.
Clinical studies using the ‘transoesophageal’ probe compared SpO2 measurements with
simultaneous SpO2 measurements from conventional pulse oximetry probes (Nellcor N-200:
N-200F) and arterial oxygen saturation (SaO2) measurements using an in vitro CO-oximeter
in 16 anaesthetized adult patients. The results showed that the ‘transoesophageal’ probe
underestimated or overestimated SpO2 values depending on the geometry of the sensor
(Prielipp et al 1996, Borum 1997). Another limitation of this design was the difficulty
in placing the probe accurately at the cricopharyngeus muscle, as the procedure required
considerable expertise. It was also found that electrocautery interference resulted in more
frequent signal dropout and delayed signal reacquisition than for a peripheral pulse oximetry
probe. The successful use of the ‘transoesophageal’ pulse oximeter has been demonstrated
in a single patient study in which peripheral oximetry was unobtainable (Borum 1997). A
more recent study (Prielipp et al 2000) used the upper oesophageal pulse oximeter in a
group of ten patients undergoing coronary bypass surgery and compared their results with
two commercial finger pulse oximeters (Criticare 504-US and Nellcor N-200). The study,
however, could not confirm that the upper oesophageal pulse oximeter was superior to
conventional peripheral pulse oximeters. Vicenzi et al (2000) used the same probe on a study
of 40 severely traumatized or diseased intensive care patients. They compared SpO2 values
from the upper oesophagus with those from a finger pulse oximeter (Hewlett-Packard). The
Topical Review
R13
Figure 7. Reflectance oesophageal pulse oximetry probe.
authors concluded that assuming correct positioning of the ‘transoesophageal’ probe, readings
from the upper oesophagus were more consistent with SpO2 values from the finger pulse
oximeter.
All the studies using the ‘transoesophageal’ probe did not present any investigations into
the morphology or the quality of PPG signals at the cricopharyngeus muscle or at any other
depths within the oesophagus.
To overcome the drawbacks of the upper oesophageal pulse oximeter, and the difficulties
which are associated with attempts to measure arterial blood oxygen saturation in the poorly
perfused peripheral circulation, Kyriacou (2001) describes a new oesophageal pulse oximetry
system. This new system comprised a miniaturized oesophageal pulse oximeter probe and a
custom made processing unit which allows the detailed investigation of photoplethysmographic
signals and SpO2 values within the whole depth of the oesophagus in healthy and critically ill
patients, and this will be the subject of the following sections.
7. Technical developments of the oesophageal pulse oximeter
7.1. Oesophageal pulse oximeter probe
A new reflectance oesophageal pulse oximeter probe was constructed (Kyriacou et al 1999,
2002c). This new probe comprised two infrared and two red surface mount emitters and
a surface mount photodetector (figure 7). The peak emission wavelengths of the infrared
and red emitters used were 880 nm and 655 nm, respectively. The oesophageal probe with
attached cable was designed to fit into a plastic transparent disposable stomach/oesophageal
tube (Pennine Healthcare, Derby, UK). The oesophageal tube used was a size 20 French gauge
(external diameter: 6.66 mm, internal diameter: 4.66 mm, length: 780 mm, without x-ray
detectable line) mainly used for gastric lavage (washout) or other gastric surgical procedures.
A finger reflectance probe, optically and electronically identical to the oesophageal probe has
also been developed to facilitate comparisons between the two sites (oesophagus and finger)
(Kyriacou et al 2002c).
R14
Topical Review
Figure 8. Main block diagram of the oesophageal processing system including the identical finger
channel.
7.2. Oesophageal processing system: hardware
A processing system was constructed to pre-process, record and display oesophageal and
finger PPG signals and estimate SpO2 values on a laptop personal computer (figure 8). Both
channels were identical and only the oesophageal will be described.
7.2.1. Pre-isolation circuitry of the processing system. The master clock and timing generator
circuit (figure 9) of the processing system was used to generate the timing signals for switching
(ON/OFF) the red (R) and infrared (IR) emitters. These timing control signals were also used
for synchronizing the demultiplexer that separated the mixed (red and infrared) PPG signals
at the output of the current-to-voltage (I –V ) amplifier circuit into red and infrared PPG
signals. The emitters were driven by a pair of identical constant current sources one for each
wavelength. Analogue switches were used to time multiplex the red and infrared emitters at
75 Hz (Kyriacou 2001, Kyriacou et al 2002c).
The mixed PPG signal from the output of the I–V amplifier was demultiplexed to separate
the PPG signals into two independent channels (red and infrared). Low-pass filters were used
to eliminate the high frequency switching noise from the demultiplexer. The red and infrared
PPG signals were then passed through the isolation barrier of the analogue isolation amplifier
to the post-isolation circuitry of the processing system.
Topical Review
R15
Figure 9. Oesophageal pulse oximeter processing system.
7.2.2. Post-isolation circuitry of the processing system. Two analogue isolation amplifiers
were used (figure 9). The infrared and red PPG signals were split into independent channels
(infrared ac and dc and red ac and dc) using active filters (Kyriacou 2001, Kyriacou et al
2002c).
7.3. Oesophageal processing system: software
All PPG output signals were digitized (100 samples per second) by a 16-bit data acquisition
card on a laptop personal computer. The digitized PPG signals were analysed by a Virtual
Instrument (VI) implemented in LabVIEW (National Instruments Corporation, Austin, TX).
This VI read the oesophageal and finger PPG data, converted them into a spreadsheet format
and saved them into a file specified by the user and displayed the signals in real time on the
screen of the laptop computer (Kyriacou 2001, Kyriacou et al 2002c). Algorithms were also
developed in the VI for the online estimation of oesophageal SpO2. The algorithm used to
estimate oesophageal SpO2 calculated the ratio (R) (see equation (1)) of the quotients of the
ac and dc PPG amplitudes at the red (655 nm) and infrared (880 nm) wavelengths. The ratio
(R) was then used to compute the arterial oxygen saturation (SpO2) using the equation
SpO2 = 110 − 25R.
(2)
This equation is a linear approximation to an empirical calibration curve established by
measurements on a large group of healthy volunteers with arterial blood oxygen saturation
(SaO2) values generally greater than 70% (Webster 1997). Figure 10 depicts a typical VI
screen of the developed oesophageal pulse oximeter processing system incorporating a finger
and an ECG channel.
R16
Topical Review
Figure 10. Virtual instrument (VI) screen of the oesophageal pulse oximetry system.
The following sections will describe the various clinical investigations performed using
the above-described oesophageal pulse oximeter to investigate for the first time PPG signals
and SpO2 values from the whole length of the oesophagus from healthy and critically ill
patients.
8. Clinical investigation 1: comparison of oesophageal and finger PPGs in healthy
anaesthetized patients
8.1. Clinical method
Twenty consented healthy (ASA 1 or 2) adult patients who were to undergo tracheal intubation
as a routine part of general anaesthesia were studied (Kyriacou et al 1999, Kyriacou 2001).
Following induction of general anaesthesia the oesophageal probe was advanced under direct
vision into the oesophagus until the end of the probe was between 25 cm and 30 cm from the
upper incisors (figure 11). The identical reflectance finger probe was also placed on the index
finger of the patient.
8.2. Results
Figure 12 shows typical red and infrared ac PPG traces obtained from the middle third of
the oesophagus and the finger of an anaesthetized patient with the mechanical ventilator
temporarily switched off (Kyriacou et al 1999). When the ventilator was switched on, the
Topical Review
R17
Figure 11. The oesophageal PPG probe contained within the stomach tube is seen placed in the
oesophagus via the mouth.
Figure 12. Typical PPG traces for the red and infrared wavelengths from the middle third of the
oesophagus and the finger of an anaesthetized patient with the mechanical ventilator temporarily
switched off.
oesophageal PPG traces were modulated by an artefact synchronous with the approximately
5 s period of the ventilator, as shown in figure 13.
Figure 14 shows the means, standard deviations (SD) of the peak-to-peak amplitudes of
the red and infrared ac PPGs from the oesophagus and the finger of the 20 patients studied
(Kyriacou et al 1999).
Statistically significant differences were found between the PPG amplitudes in the midoesophagus and the PPG amplitudes at the finger at the infrared wavelength. There were
R18
Topical Review
Figure 13. PPG traces for the red and infrared wavelengths from the middle third of the oesophagus
and the finger of an anaesthetized patient with the mechanical ventilator switched on.
Figure 14. Mean peak-to-peak PPG signals (±SD) at red and infrared wavelengths from the
mid-third of the oesophagus and the finger.
no significant differences between the PPG amplitudes in the mid-oesophagus and the PPG
amplitudes at the finger at the red wavelength. The amplitudes of the oesophageal PPGs were
on average approximately three times larger than those obtained simultaneously from a finger
for both wavelengths.
Topical Review
R19
9. Clinical investigation 2: ‘PPG mapping of the oesophagus’
9.1. Clinical method
A new clinical study was performed by Kyriacou et al (2001) to investigate in more detail the
quality of PPG signals within the whole length of the oesophagus. The preliminary objective
of this study was to characterize PPGs in healthy (ASA 1) anaesthetized patients undergoing
low risk surgery and to determine whether there would be sufficient PPG amplitudes at red
and infrared wavelengths throughout the oesophagus to make pulse oximetry feasible.
Thirteen adult elective surgery patients were recruited for this study. In this group of
patients, the oesophageal probe was advanced into the oesophagus until the end of the probe
was 35 cm from the upper incisors. PPG traces from the oesophagus at both wavelengths were
recorded for approximately 5 min at this depth. Measurements were repeated at 30, 25, 20
and 15 cm from the upper incisors (Kyriacou et al 2001).
9.2. Results
Measurable ac PPG traces at both wavelengths were obtained in the oesophagus at all five
depths in all patients. Figure 15 depicts typical traces from one patient for the five depths.
Figure 16 gives the mean ±SE of the ac PPG amplitudes at both wavelengths at the
five oesophageal depths for the 13 patients. The ac PPGs in the mid to lower oesophagus
(depths of 20 cm or greater) have significantly larger mean amplitudes at both wavelengths
than those in the upper oesophagus (15 cm). The maximum mean oesophageal amplitude for
each wavelength occurs at the depth of 25 cm. The mean value of the ac PPG amplitude at
25 cm is a factor 4.8 higher than that at 15 cm at the infrared wavelength and a factor of 6.7
higher at the red wavelength.
Statistically significant differences between the PPG amplitudes in the upper oesophagus
(15 cm) and the amplitudes at all other depths at the infrared wavelength were found. This
was also true for the red wavelength except that there is no significant difference between the
amplitudes at the depths of 15 cm and 35 cm (Kyriacou et al 2001).
10. Clinical investigation 3: oesophageal PPG signals and blood oxygen saturation
measurements in cardiothoracic surgery patients
10.1. Clinical method
This study (Kyriacou et al 2002b, 2002c, 2003) investigated and compared oesophageal and
finger PPGs and SpO2s in patients undergoing high-risk operations, such as hypothermic
cardiothoracic bypass surgery, in whom conventional pulse oximetry might fail due to
poor peripheral circulation. Forty nine adult patients were recruited for this study
(Kyriacou et al 2003). Having previously found (Kyriacou et al 1999, 2001) that PPG
signals in the mid oesophagus (20–25 cm from the upper lip) are of large amplitude, the
oesophageal pulse oximeter probe was advanced into the oesophagus at 30 cm from the
lips. Photoplethysmographic signals were observed at various depths in the oesophagus
until the site that provided the highest amplitude PPG signals and small ventilator artefact
(synchronous modulation of the oesophageal PPG traces at the frequency of the ventilator)
was determined. During the oesophageal measurements, values of blood oxygen saturation
from a commercial transmission type finger pulse oximeter (Marquette, Tram 200A; Marquette
Electronics, Milwaukee, WI) were also recorded.
R20
Topical Review
Figure 15. PPG signals at red and infrared wavelengths from five oesophageal depths. The infrared
PPG trace is at the top and has the larger amplitude in each case. The amplitudes of both red and
infrared signals increase as the depth increases from 15 cm reaching a maximum at 25 cm.
Monitoring with the oesophageal pulse oximeter was performed intermittently (Kyriacou
et al 2003) during the various periods of the operation (during induction of anaesthesia, prior
to commencing cardiopulmonary bypass, after bypass and postoperative in the intensive care
unit). During the above recording periods, samples of arterial blood were taken and analysed
by an Instrumentation Laboratories IL 482 CO-oximeter or an Instrumentation Laboratories IL
BG-1400 Blood Gas Analyser (BGA) (Instrumentation Laboratories, Lexington, MA, USA).
Topical Review
R21
800
Mean Infrared & Red AC PPG Amplidutes (mV)
Infrared (880 nm)
Red (655 nm)
700
600
500
400
300
200
100
0
15.0 cm
20.0 cm
25.0 cm
30.0 cm
35.0 cm
Oesophageal Depths
Figure 16. Mean (std error) PPG peak-to-peak amplitudes at two wavelengths and five oesophageal
depths.
10.2. Results
10.2.1. Results from the investigation of PPG signals in cardiac patients. Measurable
PPG traces at red and infrared wavelengths were obtained in the oesophagus in all 49
patients (Kyriacou et al 2002b). Figure 17 depicts typical traces from one patient undergoing
cardiopulmonary bypass surgery during the various monitoring periods as described above
(probe depth 17 cm from the lips). Figure 17(a) shows oesophageal and finger ac PPGs,
obtained at both wavelengths, and ECG signals recorded prior to skin incision. The signals in
figure 17(b) were recorded just before sternotomy. In figure 17(c) the signals were recorded
after the chest was open. Figure 17(d) shows the transition from before bypass to being on
cardiopulmonary bypass. When the heart–lung machine was switched on (indicated in the
figure as ‘on bypass’) the pulsatile PPG signals disappeared within the next 20–25 s. The
ECG trace on bypass shows a variable high frequency activity as expected. Figures 17(e) and
(f) show PPG and ECG signals after bypass, during closing of the chest and postoperatively
in the intensive care unit, respectively.
The chosen oesophageal monitoring depths ranged from 14 cm to 28 cm, measured from
the upper lip (mean ± SD: 17.8 ± 3.3 cm). The optimal oesophageal monitoring depth for each
patient was considered the depth at which oesophageal PPGs with a good signal-to-noise ratio
and acceptable ventilator artefact (synchronous modulation in the form of a sinusoidal baseline
shift in time with the approximately 5 s period of the ventilatory cycle) could be obtained
(Kyriacou et al 1999, 2001). The mean ± SD of the ac PPG amplitudes at both wavelengths
at the different oesophageal monitoring depths for all cardiothoracic patients were within the
same ranges as those obtained in an earlier PPG amplitude study at five oesophageal depths
in healthy anaesthetized patients (Kyriacou et al, 2001, 2002b).
R22
Topical Review
18.0
18.0
17.0
17.0
16.0
16.0
15.0
15.0
14.0
14.0
13.0
13.0
12.0
12.0
11.0
11.0
10.0
V
10.0
V
9.0
9.0
8.0
8.0
7.0
7.0
6.0
6.0
5.0
5.0
4.0
4.0
3.0
3.0
2.0
2.0
1.0
0.0
15
1.0
16
17
18
19
20
21
22
23
44
0.0
25
17
18
19
20
Time/s
21
(a)
18.0
18.0
17.0
17.0
16.0
16.0
15.0
15.0
14.0
14.0
13.0
13.0
12.0
12.0
25
26
27
10.0
V
9.0
9.0
8.0
8.0
7.0
7.0
6.0
6.0
5.0
5.0
4.0
4.0
3.0
3.0
2.0
2.0
1.0
1.0
27
28
29
30
31
32
33
34
35
0.0
00
36
05
10
15
Time/s
20
25
30
35
40
Time/s
(d)
(c)
18.0
18.0
17.0
17.0
16.0
16.0
15.0
15.0
14.0
14.0
13.0
13.0
12.0
12.0
11.0
11.0
10.0
10.0
V
24
11.0
10.0
0.0
26
23
(b)
11.0
V
22
Time/s
V
9.0
9.0
8.0
8.0
7.0
7.0
6.0
6.0
5.0
5.0
4.0
4.0
3.0
3.0
2.0
2.0
1.0
1.0
0.0
22
0.0
23
24
25
26
Time/s
(e)
27
28
29
30
31
32
16
17
18
19
20
21
22
23
24
25
26
Time/s
(f)
Figure 17. Oesophageal, finger and ECG traces obtained from an anaesthetized patient undergoing
cardiopulmonary bypass surgery (probe depth 17 cm from lips): (a) prior to skin incision,
(b) in operating theatre before sternotomy, (c) after sternotomy, (d) during bypass transition (before
bypass, on bypass), (e) closing the chest, (f) in intensive care unit. O/IR/AC—oesophageal infrared
ac PPG, O/R/AC—oesophageal red ac PPG, F/IR/AC—finger infrared ac PPG, F/R/AC—finger
red ac PPG.
Topical Review
R23
Oesophageal SpO2%
100
98
96
94
92
90
90
92
94
96
98
100
Blood Gas Analysis SaO2%
Figure 18. Plot of SpO2 measurements obtained from the oesophageal pulse oximeter (OES SpO2)
against SaO2 from blood gas analysis (BGA SaO2) for 49 patients. The solid line represents the
best fit linear regression line: (OES SpO2) = 12.3 + 0.88 (BGA SaO2), r 2 = 0.74, SEE = 0.86,
n = 155, p < 0.001. The dashed line represents identity.
10.2.2. Comparisons of blood oxygen saturation measurements from the oesophageal and
commercial finger pulse oximeters with those from blood gas analysis. A total of 155 sets of
data points from the 49 patients were used for the regression analysis (Kyriacou et al 200c). A
plot of SpO2 readings obtained from the reflectance oesophageal pulse oximeter (OES SpO2)
against the SaO2 values from the blood gas analyser (BGA SaO2) is shown in figure 18. The
equation of the best fit linear regression line is: (OES SpO2) = 12.3 + 0.88 (BGA SaO2)
(the solid line in figure 18); r 2 = 0.74; standard error of estimate (SEE) = 0.86; p < 0.001.
The dashed line represents the line of identity. The median (interquartile range [range]) of
the differences between the blood oxygen saturation values obtained with the oesophageal
pulse oximeter and those obtained from blood gas analysis (OES SpO2 − BGA SaO2) is 0.00
(−0.30, 0.3 [7.07]).
Figure 19 shows a plot of SpO2 readings obtained from the commercial finger pulse
oximeter (CF SpO2) against the SaO2 values from the blood gas analyser. The equation of the
best fit linear regression line was: (CF SpO2) = 26.8 + 0.73 (BGA SaO2); r 2 = 0.39; SEE =
1.48; p < 0.001. The dashed line represents the equal value line. The median (interquartile
range [range]) of the differences between the blood gas analyser and the commercial finger
pulse oximeter results (BGA SaO2 − CF SpO2) is 0.10 (−0.40, 1.07 [10.60]).
10.2.3. Comparisons of blood oxygen saturation measurements from the oesophageal and
commercial finger pulse oximeter with those from CO-oximetry. In a subset of 17 patients,
arterial blood was also analysed by CO-oximetry, providing 36 sets of data for regression
analysis.
A plot of oesophageal SpO2 readings against the SaO2 values from the CO-oximeter is
shown in figure 20. The equation of the best fit linear regression line is: (OES SpO2) = 10.1 +
1.1 (CO-ox SaO2) (the solid line in figure 20); r 2 = 0.83; SEE = 0.71; p < 0.001. The
dashed line represents the equal value line. The mean (± SD) of the differences between the
oesophageal pulse oximeter SpO2 values and the corresponding CO-oximeter readings (OES
SpO2 − CO-ox SaO2) is 0.73 ± 0.72%.
R24
Topical Review
Commercial Finger SpO2 %
100
98
96
94
92
90
90
92
94
96
98
100
Blood Gas Analysis SaO2%
Figure 19. Plot of SpO2 measurements obtained from the commercial finger pulse oximeter
(CF SpO2) against SaO2 from blood gas analysis (BGA SaO2) for 49 patients. The solid line
represents the best fit linear regression line: (CF SpO2) = 26.8 + 0.73 (BGA SaO2), r 2 = 0.39,
SEE = 1.48, n = 155, p < 0.001. The dashed line represents identity.
Oesophageal SpO2%
100
98
96
94
92
90
90
92
94
96
98
100
CO-oximetry SaO2%
Figure 20. Plot of SpO2 measurements obtained from the oesophageal pulse oximeter (OES SpO2)
against SaO2 from the CO-oximetry (CO-ox SaO2) for 17 patients. The solid line represents the
best fit linear regression line: (OES SpO2) = 10.1 + 1.1 (CO-ox SaO2), r 2 = 0.83, SEE = 0.71,
n = 36, p < 0.001. The dashed line represents identity.
Figure 21 shows a plot of commercial finger SpO2 readings against the SaO2 values from
the CO-oximeter. The equation of the best fit linear regression line is: (CF SpO2) = 4.9 + 0.9
(CO-ox SaO2); r 2 = 0.55; SEE = 1.26; p < 0.001. The dashed line represents the equal value
line. The mean and standard deviation of the differences between the commercial finger pulse
oximeter and the CO-oximeter readings (CF SpO2 − CO-ox SaO2) are 0.61 ± 1.23%.
Topical Review
R25
Commercial Finger SpO2 %
100
98
96
94
92
90
90
92
94
96
98
100
CO-oximetry SaO2%
Figure 21. Plot of SpO2 measurements obtained from the commercial finger pulse oximeter
(CF SpO2) against SaO2 from the CO-oximetry (CO-ox SaO2) for 17 patients. The solid line
represents the best fit linear regression line: (CF SpO2) = 4.9 + 0.9 (CO-ox SaO2), r 2 = 0.55,
SEE = 1.26, n = 36, p < 0.001. The dashed line represents identity.
6
(CF SpO2 - OES SpO2) %
4
Mean + 2SD
2
0
Mean
-2
Mean - 2SD
-4
-6
-8
90
92
94
96
98
100
102
(CF SpO2 + OES SpO2)/2 %
Figure 22. The difference between blood oxygen saturation values from the commercial finger
pulse oximeter (CF SpO2) and SpO2 readings obtained from the reflectance oesophageal pulse
oximeter (OES SpO2) plotted against their mean for 49 patients.
10.2.4. Comparisons of blood oxygen saturation measurements from the oesophageal and
commercial finger pulse oximeters. The 155 sets of blood oxygen saturation data points
from 49 patients were used to compare the oesophageal and the commercial finger pulse
oximeters. Since neither can be regarded as a ‘gold’ standard, the between-method differences
analysis as suggested by Bland and Altman (1986) was used to compare these two pulse
oximeters. Figure 22 is a plot of the difference between the commercial finger (CF) and
oesophageal (OES) SpO2 values against their mean. As no obvious relation between the
difference and the mean is revealed in figure 22, a calculation of the bias, estimated by the
mean difference (d) and the standard deviation of the differences (s) was performed to assess
the degree of agreement between the two methods. The bias (d) (commercial pulse oximeter
minus oesophageal pulse oximeter) was −0.3% and the standard deviation (s) was 1.5%. The
R26
Topical Review
Table 1. Blood oxygen saturation measurements from the oesophageal pulse oximeter and blood
gas analysis in patients in whom peripheral pulse oximetry failed.
Patients
BGA SaO2%
Oesophageal SpO2%
Finger SpO2%
1
2
3
97.0
97.7
98.7
99.0
97.9
98.9
98.0
97.1
97.0
98.5
99.0
97.9
98.8
98.9
Failed
Failed
Failed
Failed
Failed
Failed
Failed
4
5
limits of agreement for the SpO2 data (commercial finger and oesophageal) were
d − 2s = −0.3 − (2 × 1.5) = − 3.3%
d + 2s = −0.3 + (2 × 1.5) = 2.7%.
10.2.5. Failure of commercial pulse oximeter. Of the 49 patients included in the study, it was
found that five patients (10.2%) had one or more periods of at least ten consecutive minutes,
during which the commercial finger pulse oximeter failed to record pulsatile PPG signals and
display SpO2 values, despite being correctly positioned on the finger. The oesophageal pulse
oximeter operated successfully throughout these periods of finger monitoring failure. In four
of these patients, the finger pulse oximeter failed postoperatively in the intensive care unit
(within the first half hour after completion of the surgery), and in the fifth patient, the failure
occurred in the operating theatre before bypass. Results from arterial blood gas analysis
performed during these periods of failed finger pulse oximetry are shown in table 1, and
demonstrate good agreement (mean difference = 0.0%) between the oxygen saturation values
obtained from the oesophageal pulse oximeter and the blood gas analyser.
11. Clinical investigation 4: oesophageal blood oxygen saturation measurements in
burned patients
The oesophageal pulse oximeter developed by Kyriacou (2001) has also been used in a small
clinical study on burned patients (Pal et al 2005). In this group of patients, standard sites
for monitoring SpO2 such as fingers or toes may be affected by the burn or they might be
unsuitable due to the use of tourniquets during surgery. Seven patients with major burns were
recruited for the study. The total body surface area burnt ranged between 28% and 90%.
The oesophageal PPG signals recorded from this group of patients were of good quality
and large amplitude. A plot of SpO2 readings obtained from the reflectance oesophageal
pulse oximeter (OES SpO2) against the SaO2 values from the CO-oximeter is shown in figure
23. The equation of the best fit linear regression line is: (OES SpO2) = 33.278 + 0.666
(CO-ox SaO2) (the solid line in figure 23), r 2 = 0.49, standard error of estimate (SEE) = 0.64,
p < 0.001. The dashed line represents the line of identity.
12. Clinical investigation 5: oesophageal pulse oximetry in neonatal and paediatric
patients
The oesophageal pulse oximeter probe developed by Kyriacou (2001) has been modified
(Kyriacou et al 2002a) to fit into a conventional disposable transparent stomach tube, 12
French gauge. Such a size stomach tube is used routinely in anaesthetized babies and children.
Topical Review
R27
Oesophageal SpO2 %
100
98
96
94
92
90
90
92
94
96
98
100
CO-ox SaO2 %
Figure 23. Plot of SpO2 measurements obtained from the oesophageal pulse oximeter against SaO2
from the CO-oximetry for seven patients. The solid line represents the best fit linear regression
line: (OES SpO2) = 33.278 + 0.666 (CO-ox SaO2), r 2 = 0.49, standard error of estimate
(SEE) = 0.64, p < 0.001. The dashed line represents the line of identity.
Figure 24. Typical PPG traces obtained from a neonatal human oesophagus at two wavelengths,
infrared (top trace) and red (bottom trace).
In this pilot study, five patients were studied in the intensive care unit (Kyriacou et al 2002a).
The oesophageal SpO2 probe was advanced through the mouth to a maximum depth of 15 cm
from the lips. During the oesophageal measurements, values of blood oxygen saturation from
a commercial foot pulse oximeter were also recorded.
Measurable PPG traces of good quality were obtained in the oesophagus in all patients.
Figure 24 depicts typical PPG signals from the oesophagus of a 3.2 kg, 5 day old neonate. An
Altman and Bland (1983) plot of the difference between blood oxygen saturation values from
the commercial pulse oximeter and those from the oesophageal pulse oximeter against their
mean showed that the bias and the limits of agreement between the oesophageal and toe pulse
oximeters were −0.3% and −1.7% to 1.0%.
13. Other applications of the oesophageal pulse oximeter
13.1. Introduction
The oesophageal pulse oximeter (Kyriacou 2001) has also been used to investigate
photoplethysmographic signals in human visceral organs (Crerar-Gilbert et al 2002). The
R28
Topical Review
Figure 25. Reflectance oesophageal pulse oximeter probe placed on the surface of the human
bowel.
hypothesis underlying that investigation is that blood oxygen saturation from an extremity
such as the finger may not accurately reflect splanchnic oxygen saturation values. In many
critically ill patients, poor tissue oxygenation is due to disordered regional distribution of
blood flow, despite high global blood flow and oxygen delivery. Splanchnic ischaemia may
ultimately lead to cellular hypoxia and necrosis and may well contribute to the development
of multiple organ failure and increased mortality (Lemaire et al 1996). Rapid detection of a
significant change in tissue oxygenation could enable earlier and more successful intervention
and restoration of splanchnic blood flow and should improve survival in critically ill patients
(Lemaire et al 1996).
Techniques used to measure tissue oxygenation such as polarographic oxygen electrodes,
luminescent oxygen probes, magnetic resonance spectroscopy and positron emission
tomography remain research tools (Lemaire et al 1996). Manual fluid tonometry for estimating
intestinal hypoxia is expensive, intermittent, operator dependent and time consuming; the
recently introduced automatic device is more convenient but is even more expensive (Lemaire
et al 1996). Methods such as laser Doppler, Doppler ultrasound and intravenous fluorescein
have been previously explored to assess intestinal ischaemia in animals (Pearce et al 1987,
Ferrara et al 1988, Denobile et al 1990, Macdonald et al 1993). Many of these techniques
are complex and expensive and none of them directly measures oxygenation. Therefore, there
is a need for a simple, reliable and continuous method for estimating visceral organ SpO2.
Animal studies have also shown that pulse oximetry could be used to monitor intestinal oxygen
saturation (Macdonald et al 1993). The feasibility of estimating blood oxygen saturation in
humans has been demonstrated by a study using a commercial transmission pulse oximeter
on the colon (Ouriel et al 1988). However, there are difficulties in applying commercial pulse
oximeters to measurements in abdominal human organs because the probes are unsuitable and
are not easily sterilizable. Moreover, none of the currently available probes could be left in
the abdomen for prolonged postoperative monitoring.
As a preliminary to constructing a suitable pulse oximeter for monitoring abdominal organ
SpO2, the oesophageal pulse oximeter has been used for the measurement of PPG signals from
the surface of the bowel, liver and kidney. The aim was to develop techniques to facilitate
Topical Review
R29
Figure 26. PPG traces from simultaneous measurements at various abdominal organs (bowel,
kidney and liver) and the finger.
measurements on patients with compromised splanchnic circulation, which will be useful both
intraoperatively and in intensive care.
Topical Review
Mean Infrared & Red AC PPG Amplitudes (V)
R30
2.5
Infrared (880 nm)
Red (655 nm)
2.0
1.5
1.0
0.5
0.0
Bowel
Liver
(n = 12)
(n = 8)
Kidney
(n = 6)
Finger
(n = 11)
Organs
Figure 27. AC peak-to-peak amplitudes, mean (± SD), at two wavelengths from the three
abdominal organs and the finger.
13.2. Clinical method
Twelve adult patients undergoing elective laparotomy under general anaesthesia were studied
(Crerar-Gilbert et al 2002). The oesophageal probe was inserted into a sealed and sterilized
disposable size 20 French gauge gastric tube. The gastric tube containing the probe was then
applied gently to the surface of each abdominal organ so that the emitted light was reflected
from its surfaces (figure 25). The identical reflectance finger probe was placed on the finger of
the patient. Simultaneous PPG traces from each abdominal organ and the finger were recorded
for approximately 2 min.
13.3. Results
Crerar-Gilbert et al (2002) reported that measurable PPG signals were always obtained from
the surface of the bowel in all 12 patients, depending on intra-operative accessibility from the
liver (eight patients) and the kidney (six patients). PPG signals with similar amplitudes and
reasonably high signal-to-noise ratios were obtained from all investigated abdominal organs
(figure 26). The PPG amplitudes from both hollow and solid abdominal organs were, on
average, approximately the same as those obtained simultaneously from a finger for both
wavelengths, although there is considerable variability.
Figure 27 shows the mean peak-to-peak PPG amplitudes and standard deviations from
all investigated abdominal organs and the finger. Paired t-tests showed that there were no
statistically significant differences between the PPG amplitudes recorded from the abdominal
organs and those from the finger (Crerar-Gilbert et al 2002).
14. Conclusions
Pulse oximetry is a non-invasive photometric technique that provides information about
the arterial blood oxygen saturation (SpO2) and heart rate, and has widespread clinical
Topical Review
R31
applications. Pulse oximeters estimate arterial oxygen saturation by shining light at two
different wavelengths, red and infrared, through vascular tissue. In this method, the ac
pulsatile photoplethysmographic (PPG) signal associated with cardiac contraction is assumed
attributable solely to the arterial blood component. The amplitudes of the red and infrared ac
PPG signals are sensitive to changes in arterial oxygen saturation because of differences in the
light absorption of oxygenated and deoxygenated haemoglobin at these two wavelengths. From
the ratios of these amplitudes, and the corresponding dc photoplethysmographic components,
arterial blood oxygen saturation (SpO2) is estimated.
Although generally reliable, pulse oximeters do fail in patients with compromised
peripheral perfusion. Pulse oximetry is a pulse-dependent technique, and any significant
reduction in the amplitude of the pulsatile component of the photoplethysmographic signal
can lead to dubious values for blood oxygen saturation (SpO2) or complete failure. Hence,
pulse oximeters require adequate peripheral perfusion to operate accurately. When peripheral
perfusion is poor, as in states of hypovolaemia, hypothermia and vasoconstriction, oxygenation
readings become unreliable or cease. Such clinical situations occur, for example, after
prolonged operations such as cardiac, vascular, reconstructive or neuro-surgery. The problem
arises because conventional pulse oximetry sensors must be attached to the most peripheral
parts of the body, such as finger, ear or toe, where pulsatile flow is most easily compromised.
The introduction of the oesophageal pulse oximeter was found to be reliable and accurate in
cases of poor peripheral perfusion where peripheral pulse oximeters failed to estimate oxygen
saturation. These results show that in general the arterial blood circulation to the oesophagus
is less subject to peripheral vasoconstriction and decreased PPG amplitudes than are the
peripheral sites used for pulse oximetry such as the finger. Therefore, the human oesophagus
not only can be used as an alternative SpO2 monitoring site but also can be used as a possible
SpO2 monitoring site in cases of poor peripheral circulation where peripheral pulse oximeters
fail. This novel monitoring site, the oesophagus, can also find applications in patients who
have burns or other serious injuries where the oesophagus may be the only available site for
pulse oximetry monitoring. In addition, the application of the oesophageal pulse oximeter in
hollow and solid abdominal organs supported the hypothesis that pulse oximetry may be used
as a blood oxygen saturation monitoring technique for abdominal organs for intraoperative
and prolonged postoperative monitoring.
In summary, the use of this novel pulse oximeter has proven for the first time that the
whole of the oesophagus is a reliable and accurate monitoring site for blood oxygen saturation
in healthy patients and in sick patients in whom conventional pulse oximetry might fail due to
poor peripheral circulation.
References
Alexander C M, Teller L E and Gross J B 1989 Principles of pulse oximetry: theoretical and practical considerations
Anesth. Analg. 68 368–76
Altman D G and Bland J M 1983 Measurement in medicine: the analysis of method comparison studies Statistician
32 307–17
Anderson I D, Woodford M, De Dombal F T and Irving M 1988 Retrospective study of 1000 deaths from injury in
England and Wales Br. Med. J. 296 1305–8
Anonymous 2003 Next generation pulse oximetry Health Devices 32 49–103
Atlee J L and Bratanow N 1995 Comparison of surface and esophageal oximetry in man [abstract] Anesthesiology 83
A455
Barker S J, Hyatt J, Shah N K and Kao Y J 1993 The effect of sensor malpositioning on pulse oximeter accuracy
during hypoxemia Anesthesiology 79 248–54
Barker S J and Tremper K K 1987 The effect of carbon monoxide inhalation on pulse oximetry and transcuteneous
PO2 Anesthesiology 66 677–9
R32
Topical Review
Barker S J, Tremper K K and Hyatt J 1989 Effects of methemoglobinemia on pulse oximetry and mixed venous
oximetry Anesthesiology 70 112–7
Bickler P E, Feiner J R and Severinghaus J W 2005 Effects of skin pigmentation on pulse oximeter accuracy at low
saturation Anesthesiology 102 715–9
Bland J M and Altman D G 1986 Statistical methods for assessing agreement between two methods of clinical
measurement Lancet 1 307–10
Block F E and Detko G J 1986 Minimizing interference and false alarms from electrocautery in the Nellcor N-100
pulse oximeter J. Clin. Monit. 2 203–5
Borum S E 1997 The successful use of transesophageal pulse oximetry in a patient in whom peripheral pulse oximetry
was unobtainable [case report] Anesth. Analg. 85 514–6
Bowes W A, Corke B C and Hulka J 1989 Pulse oximetry: a review of the theory, accuracy and clinical applications
Obstet. Gynaecol. 74 541–6
Challoner A V J 1979 Photoelectric Plethysmography for Estimating Cutaneous Blood Flow (New York: Academic)
chapter 6 pp 125–51
Chan M M, Chan M M and Chan E D 2003 What is the effect of fingernail polish on pulse oximetry? Chest 123
2163–4
Clayton D G, Webb R K, Ralston A C, Duthie D and Runciman W B 1991 A comparison of the performance of 20
pulse oximeters under conditions of poor perfusion Anaesthesia 46 3–10
Cote C J, Goldstein E A, Fuchsman W H and Hoaglin D C 1988 The effect of nail polish on pulse oximetry Anesth.
Analg. 67 683–6
Crerar-Gilbert A, Kyriacou P A, Jones D P and Langford R M 2002 Assessment of photoplethysmographic signals
for the determination of splanchnic oxygen saturation in humans Anaesthesia 57 442–5
Delpy D T 1988 Developments in oxygen monitoring J. Biomed. Eng. 10 533–40
Denobile J, Guzzetta P and Patterson K 1990 Pulse oximetry as a means of assessing bowel viability J. Surg. Res. 48
21–3
De Kock J P, Tarassenko L, Glynn C J and Hill A R 1993 Reflectance pulse oximetry measurements from the retinal
fundus IEEE Trans. Biomed. Eng. 40 817–23
Dildy G A 2001 Fetal pulse oximetry: current issues J. Perinat. Med. 29 5–13
Dildy G A 2004 Fetal pulse oximetry: a critical appraisal Best Pract. Res. Clin. Obstet. Gynaecol. 8 477–84
Dildy G A, Clark S L and Loucks C A 1994 Intrapartum fetal pulse oximetry: the effects of maternal hyperoxia on
fetal arterial oxygen saturation Am. J. Obstet. Gynecol. 171 1120–4
Dorlas J C and Nijboer J A 1985 Photo-electric plethysmography as a monitoring device in anaesthesia: application
and interpretation Br. J. Anaesth. 57 524–30
Drummond G B and Park G R 1984 Arterial oxygen saturation before intubation of the trachea Br. J. Anaesth. 56
987–93
Ferrara J J, Dyess D L, Lasecki M, Kinsey S, Donnell C and Jurkovich G J 1988 Surface oximetry: a new method to
evaluate intestinal perfusion Am. Surg. 54 10–4
Fine I and Weinreb A 1995 Multiple scattering effect in transmission pulse oximetry Med. Biol. Eng. Comput. 33
709–12
Finlay J C and Foster T H 2004 Hemoglobin oxygen saturations in phantoms and in vivo from measurements of
steady-state diffuse reflectance at a single, short source–detector separation Med. Phys. 31 1949–59
Fluck R R Jr, Schroeder C, Frani G, Kropf B and Engbretson B 2003 Does ambient light affect the accuracy of pulse
oximetry? Respir. Care 48 677–80
Freund P R, Overand P T, Cooper J, Jacobson L, Bosse S, Walker B, Posner K L and Cheney F W 1991 A prospective
study of intraoperative pulse oximetry failure J. Clin. Monit. 7 253–8
Gardosi J O, Schram C M and Symonds E M 1991 Adaptation of pulse oximetry for fetal monitoring during labour
Lancet 337 1265–7
Goldman J M, Petterson M T, Kopotic R J and Barker S J 2000 Masimo signal extraction pulse oximetry J. Clin.
Monit. Comput. 16 475–83
Hanning C D and Alexander-Williams J M 1995 Pulse oximetry: a practical review Br. Med. J. 311 367–70
Hanowell L, Eisele J H and Downs D 1987 Ambient light affects pulse oximeters Anesthesiology 67 864–5 (letter)
Higgins J L and Fronek A 1986 Photoplethysmographic evaluation of the relationship between skin reflectance and
blood volume J. Biomed. Eng. 8 130–6
Hovagim A R, Vitkun S A, Manecke G R and Reiner R 1989 Arterial oxygen saturation in adult dental patients
receiving conscious sedation J. Oral Maxillofac. Surg. 47 936–9
Jay G D, Hughes L and Renzi F P 1994 Pulse oximetry is accurate in acute anemia from hemorrhage Ann. Emerg.
Med. 24 32–5
Kataria B K and Lampkins R 1986 Nail polish does not affect pulse oximetry saturation Anesth. Analg. 65 824 (letter)
Topical Review
R33
Kelleher J F 1989 Pulse oximetry J. Clin. Monit. 5 37–62
Kelleher J F and Ruff R H 1989 The penumbra effect: vasomotion-dependent pulse oximeter artifact due to probe
malposition Anesthesiology 71 787–91
Kessler M R, Eide T, Humayan B and Poppers P J 1986 Spurious pulse oximeter desaturation with methylene blue
injection Anesthesiology 65 435–6
Kidd J F and Vickers M D 1989 Pulse oximeters: essential monitors with limitations Br. J. Anaesth. 62 355–7
Klauser C K, Christensen E E, Chauhan S P, Bufkin L and Magann E F 2005 Use of fetal pulse oximetry among
high-risk women in labor: a randomized clinical trial Am. J. Obstet. Gynecol. 192 1810–7
Kyriacou P A 2001 Investigation of new electro-optical techniques for monitoring patients with compromised
peripheral perfusion in anaesthesia PhD Thesis University of London
Kyriacou P A, Moye A R, Choi D M A, Langford R M and Jones D P 2001 Investigation of the human oesophagus
as a new monitoring site for blood oxygen saturation Physiol. Meas. 22 223–32
Kyriacou P A, Moye A R, Gregg A, Choi D M A, Langford R M and Jones D P 1999 A system for investigating
oesophageal photoplethysmographic signals in anaesthetised patients Med. Biol. Eng. Comput. 37 639–43
Kyriacou P A, Powell S L, Jones D P and Langford R M 2003 Evaluation of oesophageal pulse oximetry in
cardiothoracic surgery patients Anaesthesia 58 422–7
Kyriacou P A, Powell S, Langford R M and Jones D P 2002b Investigation of oesophageal photoplethysmographic
signals and blood oxygen saturation measurements in cardiothoracic surgery patients Physiol. Meas. 23
533–45
Kyriacou P A, Powell S, Langford R M and Jones D P 2002c Esophageal pulse oximetry utilizing reflectance
photoplethysmography IEEE Trans. Biomed. Eng. 49 1360–8
Kyriacou P A, Wardhaugh A, Jones D P, Langford R M and Petros A J 2002a Investigation of photoplethysmographic
signals in neonatal and paediatric patients Intensive Care Med. 28 (Suppl. 1) S111
Lambert M A and Crinnion J 1989 The role of pulse oximetry in the accident and emergency department Arch. Emerg.
Med. 6 211–5
Lemaire F et al 1996 Tissue hypoxia. How to detect, how to correct, how to prevent? Intensive Care Med. 22
250–1257
Lindberg L G and Oberg P A 1991 Photoplethysmography: part 2. Influence of light source wavelength Med. Biol.
Eng. Comput. 29 48–54
Macdonald P H, Dinda P K, Beck I T and Mercer C D 1993 The use of oximetry in determining intestinal blood flow
Surg. Gynecol. Obstet. 176 451–8
Mannheimer P D, Casciani J R, Fein M E and Nierlich S L 1997a Wavelength selection for low-saturation pulse
oximetry IEEE Trans. Biomed. Eng. 44 148–58
Mannheimer P D, Fein M E and Casciani J R 1997b Physio-optical considerations in the design of fetal pulse oximetry
sensors Eur. J. Obstet. Gynecol. Reprod. Biol. 72 9–19
Marieb E N 1992 Human Anatomy and Physiology 2nd edn (Redwood City, CA: Benjamin/Cumming)
Mendelson Y 1992 Pulse oximetry: theory and applications for noninvasive monitoring Clin. Chem. 38 1601–7
Mendelson Y and Ochs B D 1988 Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography IEEE
Trans. Biomed. Eng. 35 798–805
Minnich M E, Clark R B, Miller F C and Thompson D S 1990 Oxygen desaturation in women in labor J. Reprod.
Med. 35 693–6
Moller J T, Johannessen N W, Espersen R, Ravlo O, Pedersen B D, Jensen P F, Rasmussen N H, Rasmussen L S,
Pedersen T and Cooper J B 1993 Randomized evaluation of pulse oximetry in 20802 patients: II. Perioperative
events and postoperative complications Anesthesiology 78 445–53
Moyle J T B 1994 Principles and Practice Series: Pulse Oximetry (London: BMJ Publishing Group)
Moyle J T B 1996 Uses and abuses of pulse oximetry Arch. Dis. Child. 74 77–80
Nelson D A, Krupsky S, Pollack A, Aloni E, Belkin M, Vanzetta I, Rosner M and Grinvald 2005 Special report:
noninvasive multi-parameter functional optical imaging of the eye Ophthal. Surg. Lasers Imaging 36 57–66
Nijboer J A, Dorlas J C and Mahieu H F 1981 Photoelectric plethysmography—some fundamental aspects of the
reflection and transmission method Clin. Phys. Physiol. Meas. 2 205–15
Ouriel K, Fiore W M and Geary J E 1988 Detection of occult colonic ischaemia during aortic procedures: use of an
intraoperative photoplethysmographic technique J. Vas. Surg. 7 5–9
Pal S K, Kyriacou P A, Kumaran S, Fadheel S, Emamdee R, Langford R M and Jones D P 2005 Evaluation of
oesophageal reflectance pulse oximetry in major burns patients Burns 31 337–41
Palve H 1992a Reflection and transmission pulse oximetry during compromised peripheral perfusion J. Clin. Monit.
8 12–5
Palve H 1992b Comparison of reflection and transmission pulse oximetry after open-heart surgery Crit. Care Med.
29 48–51
R34
Topical Review
Palve H and Vuori A 1989 Pulse oximetry during low cardiac output and hypothermia states immediately after open
heart surgery Crit. Care Med. 17 66–9
Pearce W H, Jones D N, Warren G H, Bartle E J, Whitehill T A and Rutherford R B 1987 The use of infrared
photoplethysmography in identifying early intestinal ischaemia Arch. Surg. 122 308–10
Pologe J A 1987 Pulse oximetry: technical aspects of machine design Int. Anesthesiol. Clin. 35 137–53
Prielipp R C, Scuderi P E, Butterworth J F, Royster R L and Atlee J L 1996 Comparison of transesophageal pulse
oximetry (TEPO) with peripheral surface pulse oximetry in CABG patients [abstract] Anesthesiology 85 3A
Prielipp R C, Scuderi P E, Hines M H, Atlee J L and Butterworth J F 2000 Comparison of a prototype esophageal
oximetry probe with two conventional digital pulse oximetry monitors in aortocoronary bypass patients J. Clin.
Monit. Comput. 16 201–9
Pullerits J, Burrows F A and Roy W L 1987 Arterial desaturation in healthy children during transfer to the recovery
room Can. J. Anaesth. 34 470–3
Ralston A C, Webb R K and Runciman W B 1991 Potential errors in pulse oximetry: I. Pulse oximeter evaluation
Anaesthesia 46 202–6
Reich D L, Imcenko A, Bodian C A, Kraidin J, Hofman J B, Deperio M, Konstadt S N, Kurki T and Eisenkraft J B
1996 Predictors of pulse oximetry data failure Anesthesiology 84 859–64
Reuss J L 2004 Factors influencing fetal pulse oximetry performance J. Clin. Monit. Comput. 8 13–24
Richardson T 2005 Methemoglobinemia and pulse oximetry Resp. Care 50 386
Ries A L, Prewitt L M and Johnson J J 1989 Skin color and ear oximetry Chest 96 287–90
Roberts V C 1982 Photoplethysmography—fundamental aspects of the optical properties of blood in motion Trans.
Inst. M C 4 101–6
Romanes G J 1972 Cunninghamn’s Textbook of Anatomy (London: Oxford University Press)
Rosenberg J and Pedersen M H 1990 Nasal pulse oximetry overestimates oxygen saturation Anaesthesia 45 1070–1
Scheller M S, Unger R J and Kelner M J 1986 Effects of intravenously administered dyes on pulse oximeter readings
Anesthesiology 65 550–2
Schmitt J M 1991 Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetry IEEE Trans.
Biomed. Eng. 38 1194–203
Severinghaus J W 1993 History and recent developments in pulse oximetry Scand. J. Clin. Lab. Invest. Suppl. 214
105–11
Severinghaus J W and Kelleher J F 1992 Recent developments in pulse oximetry Anesthesiology 76 1018–38
Severinghaus J W and Koh S O 1990 Effect of anemia on pulse oximeter accuracy at low saturations J. Clin. Monit.
6 85–8
Severinghaus J W and Naifeh K H 1987 Accuracy of response of six pulse oximeters to profound hypoxia
Anesthesiology 67 551–8
Severinghaus J W and Spellman M J 1990 Pulse oximeter failure thresholds in hypotension and vasoconstriction
Anesthesiology 73 532–7
Shimada Y, Yoshiya I, Oka N and Hamaguri K 1984 Effects of multiple scattering and peripheral circulation on
arterial oxygen saturation measured with pulse-type oximeter Med. Biol. Eng. Comput. 22 475–8
Shvartsman L D and Fine I 2003 Optical transmission of blood: effect of erythrocyte aggregation IEEE Trans. Biomed.
Eng. 50 1026–33
Sidi A, Paulus D A, Rush W, Gravenstein N N and Davis R F 1987 Methylene blue and indocyanine green artifactually
lower oximetry readings of oxygen saturation J. Clin. Monit. 3 249–56
Sinex J E 1999 Pulse oximetry: principles and limitations Am. J. Emerg. Med. 17 59–66
Trang H, Boureghda S and Leske V 2004 Sleep desaturation: comparison of two oximeters Pediatr. Pulmonol. 37
76–80
Tremper K K and Barker S J 1989 Pulse oximetry Anaesthesiology 70 98–108
Tyler I L, Tantisira B, Winter P M and Motoyama E K 1985 Continuous monitoring of arterial oxygen saturation with
pulse oximetry during transfer to the recovery room Anesth. Analg. 64 1108–12
Vicenzi M N, Gombotz H, Krenn H, Dorn C and Rehak P 2000 Transesophageal versus surface pulse oximetry in
intensive care unit patients Crit. Care Med. 28 2268–70
Volgyesi G A and Spahr-Schopfer I 1991 Does skin pigmentation affect the accuracy of pulse oximetry? An in vitro
study Anesthesiology 75 A406 (abstract)
Wahr J A, Tremper K K and Diab M 1995 Pulse oximetry Respir. Care Clin. N Am. 1 77–105
Webster J G 1997 Design of Pulse Oximeters (Bristol: Institute of Physics Publishing)
Welch J 2005 Pulse oximeters Biomed. Instrum. Technol. 39 125–30
Wray S, Cope M, Delpy D T, Wyatt J S and Reynolds E O R 1988 Characterization of the near infrared absorption
spectra of cytochrome aa3 and hemoglobin for the non-invasive monitoring of cerebral oxygenation Biochim.
Biophys. Acta 933 184–92
Topical Review
R35
Wukitsch M W, Petterson M T, Tobler D R and Pologe J A 1988 Pulse oximetry: analysis of theory, technology and
practice J. Clin. Monit. 4 290–301
Yilmaz E N, Vahl A C, van Rij G, Nauta S H, Brom H L and Rauwerda J A 1999 Endoluminal pulse oximetry of the
sigmoid colon and the monitoring of the colonic circulation Cardiovasc. Surg. 7 704–9
Yoshiya I, Shimada Y and Tanaka K 1980 Spectrophotometric monitoring of arterial oxygen saturation in the fingertip
Med. Biol. Eng. Comput. 18 27–32