euspen’s 16th International Conference &
Exhibition, Nottingham, UK, May 2016
www.euspen.eu
Mechanical testing of orthodontic archwires
Višnja Katić, 1 Saša Zelenika, 2 Ervin Kamenar, 2 Kristina Marković, 2 Marko Perčić 2 and Stjepan Špalj 1
1 University
2 University
of Rijeka, School of Medicine, Braće Branchetta 20, 51000 Rijeka, CROATIA
of Rijeka, Faculty of Engineering & Centre for Micro and Nano Sciences and Technologies, Vukovarska 58, 51000 Rijeka, CROATIA
sasa.zelenika@riteh.hr
Abstract
Mechanical properties of orthodontic archwires are studied in this work in the micrometric domain. Measurements are performed
on nickel-titanium archwires without and with different coatings as well as subjected for four weeks to an artificial saliva solution
with different pH values (4.8 and 6.6). Since surface properties can have a meaningful influence on friction, corrosion or plaque
accumulation, surface texture of the wires is measured by employing atomic force microscopy. On the other hand, with the aim of
determining Young’s modulus and the hardness, nanoindentation tests are performed for different peak load values. It is hence
established that there is no statistically relevant deterioration of the surfaces after their exposure to saliva. Rh coating leads, in turn,
to an increase of surface roughness. Young’s moduli and hardness values tend to increase with increasing indentation depths, while
they are not meaningfully affected by the coating or the corrosion in saliva.
Nickel-titanium archwires, coating and corrosion, AFM and nanoindentation testing, surface roughness, hardness, Young’s modulus
1. Introduction
Orthodontic archwires with submillimetre cross-sections, used
to correct anomalies in teeth position, must be aesthetically
pleasing and biostable but also characterised by low friction,
formability as well as high compliance. Mechanical properties of
typically used archwire materials, such as their Young’s modulus,
hardness and especially surface texture, must thus be studied as
they influence archwires’ efficient use [1-2]. Three types of
0.02 inch x 0.02 inch archwires are hence studied:
- a superelastic nickel-titanium alloy with an untreated surface
(NiTi, i.e. Nitinol known as a shape memory alloy): has good
mechanical and clinical properties, sold as Sentalloy;
- NiTi archwire with an ion-implanted rhodium (Rh) coating: Rh
NiTi sold as High Aesthetic;
- and the nitrified NiTi archwire: N NiTi sold as IonGuard.
To measure the relevant mechanical properties of archwire
specimens, the equipment of the Precision Engineering Laboratory of the University of Rijeka, Croatia, is used. The surface texture of the samples is hence measured by employing the atomic
force microscopy (AFM) mode of Bruker’s Dimension Icon SPM,
whereas Young’s moduli and the hardness are determined by
using a Keysight G200 Nanoindenter. Conclusions on the influence of the used coatings and of the corrosion on the measured
mechanical properties of archwires are thus drawn.
2. Materials and methods
Specimens of each wire type are cut from arch forms. They are
then subjected to an artificial saliva solution (1.5 g/L KCl, 1.5 g/L
NaHCO3, 0.5 g/L NaH2PO4xH2O, 0.5 g/L KSCN, 0.9 g/L lactic acid)
for four weeks at 37 C. To simulate intraoral variations, two
solutions with different pH values of artificial saliva (4.8 and 6.6)
are adjusted with lactic acid and NaOH. Depending on the
coating and the pH value, 6 different wire types are hence used
in tests made at 23 C.
AFM is considered as an appropriate tool for measuring surface topography of orthodontic wires, since it allows a high-resolution qualitative characterisation. In fact, surface properties,
and in particular roughness, can have a meaningful influence not
only on friction, but also on corrosion, adhesion, plaque accumulation and aesthetics of the archwires [1, 3]. Bruker’s SPM apparatus is hence used in this work to perform two contact-mode
AFM measurements on four samples of each considered wire
type (affixed onto a sample holder – see Fig. 1 left). In total, 48
measurements are hence performed on 30 m x 30 m surfaces. The measurements are controlled via the instrument’s
NanoScope software that is also used to flatten measurement
data to filter the inclination of the probe with respect to the surface of the sample. The used probe is Bruker's SNL-10 highresolution probe with a 2 nm Si tip radius mounted on a 0.6 m
thick triangular Si3N4 cantilever with a bending stiffness of
0.12 N/m. Surface topographies, and in particular the values of
the arithmetic average surface roughness Ra, the RMS roughness
Rq and the maximum height RZ, are hence obtained.
Figure 1. Archwires mounted on the AFM sample holder (left) and
specimens mounted into the Keysight G200 Nanoindenter (right).
Nanoindentation is an accurate way to probe the mechanical
characteristics of the surface layers of orthodontic archwires [3].
In fact, the used Keysight device can generate forces with a resolution of 50 nN and measure displacements with a resolution
of 0.01 nm. The indentation hardness HIT and indentation
Young’s modulus EIT measurements are hence performed in accordance to ISO 14577 by using a 20 nm Berkovich tip;
nanoindenter frame compliance and tip sharpness calibration
are performed on a standardised fused-Si reference sample. A
10 s time to load, a peak hold time of 1 s and allowable drift rates
Obtained representative 3D AFM images of the surfaces of the
analysed archwires are shown in Fig. 2 whereas the Ra, Rq and RZ
values with the respective standard deviations are reported in
table 1 for each of the 6 wire type – pH value combinations. It
can thus be inferred that, contrary to what was reported in [2],
there is no significant correlation between corrosion in saliva
with variable pH values and surface roughness. It is clear,
however that, in accordance also with [1-2], the Rh coating tends
to increase the roughness. The values of the characteristic
parameters for the Rh NiTi archwires (Ra 155 nm, Rq 205 nm,
RZ 1770 nm) are thus higher than those of the other two
considered wire types, which are, in turn, approximately
equivalent
(Ra 110 nm,
Rq 140 nm,
RZ 1150 nm).
Roughness of the plain NiTi archwire tends, however, to be
higher than that reported in previous literature [1-2], while, in
general, the dispersion of data, related to optically visible
material inhomogeneity that could be related to the differences
in the production processes, is rather large.
Figure 2. Topographies of NiTi (~ N NiTi) and Rh NiTi wires with pH 6.6.
Table 1 Surface roughness parameters of the analysed archwires.
NiTi
Rh
NiTi
N
NiTi
Ra/nm
98.5 ± 14.8
120.3 ± 41.7
162.4 ± 40.7
152.8 ± 22.8
117.6 ± 21.7
109.0 ± 12.4
100
20 mN
100 mN
50
0
0
500
1000
1500
2000
Displacement into surface/nm
Figure 3. Typical nanoindentation curves for the considered archwires.
3. Results and discussion
pH
4.8
6.6
4.8
6.6
4.8
6.6
to the difference in the products evaluated in the two cases. The
values of HIT and EIT obtained in this work do not correlate with
the variation of surface coating either. Moreover, as in the case
of surface roughness, there is a large dispersion of measurement
values, confirming material inhomogeneity.
Load/mN
of 0.2 nm/s are set via instrument’s software that is used also to
acquire measurement data. Four wires of the same type are
hence bonded to a sample holder with a cyanoacrylate adhesive,
while three holders are positioned and bonded onto a standard
Al sample mount, as this solution has proven to be the most stable and insensitive to perturbations. Two sample mounts, each
with 12 wires, are hence carefully aligned in their seats to guarantee a flat mounting of all the samples (Fig. 1 right). Since in
literature it is evidenced that EIT and HIT vary depending on loading conditions [3], each of the 24 archwires is hence indented in
a 4 x 4 pattern of points with a peak load of 20 mN (corresponding to an average indentation depth of ~ 1 m) and then, in a
different area, with a 100 mN peak force (indentation depth of
~ 2 m). More than 750 indents are thus performed.
Rq/nm
125.1 ± 18.3
151.8 ± 47.4
211.4 ± 50.8
199.9 ± 30.2
150.8 ± 27.8
134.8 ± 15.6
RZ/nm
1175 ± 389
1147 ± 400
1757 ± 521
1784 ± 771
1614 ± 1023
1023 ± 237
A typical diagram of the performed nanoindentations is shown
in Fig. 3, while the obtained EIT and HIT values, with the respective standard deviations, are given in table 2. It can hence be
noted that there is seemingly no deterioration of the mechanical
properties of the archwires with their exposure to saliva. EIT and
HIT values are clearly lower (roughly 30 % in the case of EIT and
up to 3-6 times in the case of HIT) than those reported in [3]. In
that case, however, not only the influence of the pH value was
not considered, but the surfaces of the samples were also
ground prior to nanoindentation, which clearly influences surface properties by levelling the present irregularities and
inhomogeneity. Also, the herein obtained EIT and HIT values tend
to increase by rising the peak load (indentation depth), which is
in contrast with the results obtained in [3] and could be related
Table 2 Archwires’ indentation Young’s modulus EIT and hardness HIT.
pH
NiTi
Rh
NiTi
N
NiTi
4.8
6.6
4.8
6.6
4.8
6.6
EIN/GPa
20 mN
100 mN
45.5 ± 12.2 52.1 ± 9.8
44.7 ± 10.3 50.1 ± 26.4
31.4 ± 18.3 48.1 ± 14.5
42.7 ± 13.4 43.9 ± 21.1
33.8 ± 12.1 40.3 ± 14.1
52.1 ± 15.7 41.7 ± 9.6
HIN/GPa
20 mN
100 mN
0.8 ± 0.4 1.3 ± 0.5
0.9 ±0.4 2.0 ± 1.4
0.6 ± 0.8 1.1 ± 0.7
0.7 ± 0.4 1.5 ± 0.5
0.5 ± 0.4 1.0 ± 0.3
1.9 ± 1.1 1.3 ± 0.6
4. Conclusions and outlook
A thorough analysis of the surface properties of orthodontic
archwires with different coatings and pH values is performed in
this work on the microscale by using AFM and nanoindentation.
It is established that there is no deterioration of the markedly
inhomogeneous surfaces with their exposure to saliva, regardless of its level of acidity. The surface quality is, in turn, related
to non-standardised production and coating processes of the
wires. Confirming prior art, it is established that Rh coating leads
to an increase of surface roughness that could negatively affect
friction, corrosion and plaque formation. On the other hand, the
EIT and HIT values of archwires’ surfaces are not affected by
varying coating and corrosion caused by the acidulated (pH 4.8)
and neutral (pH 6.6) saliva, whereas the obtained values are
different from those reported in literature. This is related to the
fact that in prior art surfaces of the samples were ground before
measurements, which levels surface irregularities. EIT and HIT
tend to increase with rising loads/depths of indentation.
In future work, the herein presented data will be correlated to
tests on friction forces and other mechanical properties performed on the same specimens. Recently introduced diamondlike carbon coated archwires will also be considered, since they
should imply a decrease of friction while significantly increasing
hardness. An attempt will also be made to correlate the
measured mechanical properties to information about the
production process of the archwires, including information on
the deposition processes of the coating materials.
Acknowledgements
The work was enabled via the ERDF project RC.2.2.06-0001 (RISK) and
the Croatian Science Foundation project IP-2014-09-7500.
Essential bibliography
[1] D’Antò V et al. 2012 Angle Orthodontists 82/5 922-8
[2] Rongo R et al. 2014 Angle Orthodontists 84/4 665-72
[3] Iijima M, Muguruma T, Brantley W A and Mizoguchi I 2011 Am. J.
Orthodontics & Dentofacial Orthopedics 140/1 65-71