212
B. Hiebl, S. Bog, R. Mikut, C. Bauer, O. Gemeinhardt, F. Jung, T. Krüger
Applied Cardiopulmonary Pathophysiology 14: 212-219, 2010
In vivo assessment of tissue compatibility and
functionality of a polyimide cuff electrode for recording
afferent peripheral nerve signals
B. Hiebl1, S. Bog2, R. Mikut3, C. Bauer3, O. Gemeinhardt4, F. Jung1,
T. Krüger5
1
Center for Biomaterial Development and Berlin-Brandenburg Center for Regenerative
Therapies, Helmholtz-Zentrum Geesthacht, Centre for Materials and Costal Research, Teltow, Germany; 2University Heidelberg, Interdisciplinary Biomedical Research Facility, Heidelberg, Germany; 3Karlsruhe Institute of Technology (KIT), Institute for Applied Computer
Science, Karlsruhe, Germany; 4Institute of Veterinary Anatomy, Faculty of Veterinary Medicine, Freie Universität Berlin, Germany; 5inomed Medizintechnik GmbH, Teningen, Germany
Abstract
Minimally-invasively implantable spiral cuff electrodes coated with polyimide (PI) for insulation
in combination with suitably designed amplifiers were recently reported to allow signalling of
electroneurograms (ENGs). These cuff electrodes can be used to record the respiratory drive
after implantation on the phrenic nerve, but PI can cause epineurial fibrosis, fiber loss, and limited reproducibility of recordings.
This study aimed to explore the tissue reaction in response to a flexible tripolar cuff electrode
embedded in a thin (10 µm) PI insulating carrier after implantation around the sciatic nerves
of rats (n=4) at the branching into the N. peroneus communis, the N. tibialis and N. cutaneus
surae caudalis. 28 days after implantation the electrode functionality was proven and ENG signals recorded. In addition, transverse sections of the implantation site were analysed after
hematoxylin-eosin (HE) and Weigert´s fibrin staining for changes in the tissue morphology and
the number of myelinated nerve fibres.
28 days after implantation the electrode contact to the nerve still was sufficient for signal
recording and no changes in the nerve morphology and the number of myelinated nerve fibres could be noted. However, cuff functionality for nerve signal transfer was limited by a fibrous capsule, which covered the whole electrode and which was composed mainly of tightly packed fibroblasts and fibrin. The formation of such a fibrous capsule is known to be caused
by the foreign body response and shows a limited tissue compatibility which might be mainly
related to the PI insulating material. Further studies will address the investigation of alternative
elastic matrix-materials to achieve a strong integration of the electrode in the nerve tissue and
in this way a long term functionality of the cuff electrode.
Key words: polyimide, cuff sensor, peripheral nerve, neural damage, functional electric stimulation
In vivo assessment of tissue compatibility and functionality of a polyimide cuff electrode ...
Introduction
The respiratory drive of anesthetized animals
can be measured by using electroneurograms
(ENGs) which are obtained from the phrenic
nerve using minimally-invasively implantable
spiral cuff electrodes [1]. They are advantageous compared to invasive nerve electrodes
(e.g. hook electrode), because they can provide signals without cutting or desheathing of
the nerve [2-10]. The spiral cuff electrode is
designed to wrap around the nerve and, due
to its self-sizing property, it adapts its diameter to the size of the nerve [4]. However, ENG
recordings from cuff electrodes have a low
signal-to-noise ratio and are often affected by
artefact signals, mostly generated by the muscles nearby [11]. Artefact reduction might be
achieved by using proper amplifiers [11-12].
The long-term efficacy of the cuff electrode
for ENG measurements depends on the tissue-material interaction especially the chronic inflammatory response induced by the
electrode material. Polyimide (PI) is broadly
explored as membrane or particle forming
biomaterial [13-15] as well as substrate and
insulation material for neural implants due to
its chemical and mechanical durability [1619], but epineurial fibrosis, fiber loss, limited
reproducibility of recordings and variability of
stimulation conditions have been documented after extraneural cuff electrode implantation [20-21].
In this study we explored the foreign body
response/tissue compatibility of adult female
rats in response to a new flexible tripolar cuff
electrode design composed of a thin (10 µm)
and flexible PI insulating carrier and three circumneural platinum electrodes after 28 days
of implantation at the branching of the sciatic nerve into the N. peroneus communis, the
N. tibialis and N. cutaneus surae caudalis.
213
Materials and methods
Polyimide cuff electrode
Nerve signals were recorded from the sciatic
nerve with a tripolar cuff electrode which
consisted of a cuff electrode and an electrode extension cable (length 150 mm, fig. 3).
The electrode to electrode distance in the
cuff was 1.5 mm. The fabrication of the polyimide-based microstructure with a single metallization layer and a PI coating (Pyralin PI
2611, HD Microsystems, Bad Homburg, Germany) has been described in detail previously [12, 18].
Animals
The study was approved by the regional
council of Karlsruhe (Germany) and performed in accordance with the EN DIN ISO
10993-6. A total of four female rats of the inbred strain BDX were used in the present
study. The animals were provided from a
commercial breeder (Charles River, Germany). They were 7 ± 1 months old and had
a body weight of 220 ± 10 g before surgery.
Housing and animal care
The animals were group-housed in MacrolonTM-based type III cages with at least two
animals per cage and ad libitum supply of water and food to guarantee homeostasis as
well as exclude acute and chronic medical
conditions. According to the European
Guidelines for care and use of laboratory animals, the animals were kept at a temperature
of 20 ± 1 °C and a humidity of 55%. Artificial
lighting was used to maintain a rhythm of 12
h day and 12 h night.
Surgical procedure
For electrode implantation, animals were
anaesthetized with isoflurane (1.5 v/v%). After shaving and disinfecting the dorsal neck
214
B. Hiebl, S. Bog, R. Mikut, C. Bauer, O. Gemeinhardt, F. Jung, T. Krüger
region and the right lateral Regio femoris with
poly(1-vinyl-2-pyrrolidon)iodine complex (Betaisodona®, Mundipharma, Germany), firstly,
the sciatic nerve was freed from surrounding
tissue via an incision of about 20±5 mm
through the skin and the Musculus biceps
femoris. The initially tube shaped cuff was
bent with forceps and positioned behind the
nerve. Thereafter the forceps were removed
allowing the cuff to wrap around the branching of the sciatic nerve into the N. peroneus
communis, the N. tibialis and N. cutaneus
surae caudalis avoiding compression and
stretching. After release, the spiral cuff was
closed covering the whole nerve perimeter,
to which it adhered by surface tension. The
electrode interconnect line was passed
through the incision of the Musculus biceps
femoris avoiding tension. The connector A
bundle (see fig. 2) was placed and fixed by a
single stitch subcutaneously on the upper lateral side of the hind limb. The incision of the
muscle was closed by a continuous suture using blue dyed poly(lactide-co-glycolide)
(VicrylPlus, 3-0, Ethicon, Germany) and the
sensor bundle with the connector B (see fig.
2) was subcutaneously placed on the back
and exit out of the body through an incision
at the prepared neck region which was subcutaneously stabilised by a polypropylene
mesh (15×15mm, Parietene®, Sofradim, Germany). To close the skin at the surgical site of
the limb and the neck, suture clips (Michel
Suture Clip, 7.5 mm, Harvard Apparatus,
USA) were used. Additionally a rat jacket
(Harvard Apparatus, USA) was used to protect the surgical site at the neck and also to
protect the animal in the post surgical or
treatment situation.
Electrode functionality testing
The functionality of the electrode to record
afferent ENG signal of the rat sciatic nerve
was tested after mechanical stimulation of the
rat’s paw using a 6.10 Semmes-Weinstein filament representing a force of linear pressure
corresponding to a mass of 75 g. [22]. The
method to investigate the functionality of the
cuff electrode especially recording of nerve
signals has been described in detail previously [12].
Histology
Four weeks (28 days) after the implantation,
the rats were sacrificed by the use of CO2.
The skin of the implantation site was opened
in the direction of the former incision, blood
was swabbed, and the electrode cuff together with the surrounding tissue (see fig. 3)
were immediately placed in a small jar containing the fixing agent, a buffered neutral 4
v/v% formalin solution (pH 7) for further histological examinations. Histological semi-sections were made for light microscopic analysis (phase contrast mode) after acrylic resin
embedding (Technovit 7200 VLC, Kulzer,
Germany). Sections were prepared in 5 µm
thickness and subjected to staining with
hematoxylin-eosin (HE) and Weigert´s fibrin
according to the protocol of Romeis [23].
Five slides per animal were prepared and
each slide was evaluated at five different
fields of view with a transmitted light microscope in the phase contrast mode (Zeiss,
Germany) using the image analysis software
AxioVisio (Zeiss, Germany).
Statistics
Results are expressed as mean value ± standard deviation for continuous variables of at
least three independent experiments. Differences were assessed using two-tailed Student´s t-tests for unpaired samples. In case of
multi-sample comparison, variance analysis
was performed. Significance was assumed if
p value was less than 0.05.
In vivo assessment of tissue compatibility and functionality of a polyimide cuff electrode ...
Results
Electrode functionality testing
The implantation of the cuff electrode around
the sciatic nerve took about 1-2 min. Immediately after the implantation, the contact interface between the electrode and the nerve
was sufficient to record afferent ENG signals
(sampling frequency 50 kHz) after stimulation with defined mechanical stimuli on the
naked skin of the paw. With a 6.10 SemmesWeinstein filament, a pressure was applied
approximately once per second inducing afferent signals in response to the mechanical
stimulus (see figure 1). Since it was difficult to
215
detect the raw signal, a filtered activity signal
was generated by computing absolute values
of the raw signal, followed by an infinite-input
response filter of first order (filter constant a
= 0.999). The induced afferent nerve signals
were seen in the filtered activity signal. At approximately 10.5 s, an external disturbance
occurred.
In addition, a spectrogram (windows size
1024 sample points) of the raw signal
showed the signal power for the different frequencies (fig. 2). Afferent nerve signals
ranged up to 7 kHz. All analyses were performed with the open-source Matlab toolbox
Gait-CAD [24-25].
Figure 1: Afferent nerve ENG signal of the rat sciatic nerve after
mechanical stimulation of the
paw using a 6.10 Semmes-Weinstein filament representing a force of a mass of 75 g, (A): raw signal amplitude, (B): filtered afferent activity signal; at 10.5 sec a
disturbance occurred.
Figure 2: Spectrogram of an ENG raw signal of the rat sciatic nerve after mechanical stimulation of
the paw using a Semmes-Weinstein filament in the size of 6.10 representing a force of a mass of
75 g.
216
B. Hiebl, S. Bog, R. Mikut, C. Bauer, O. Gemeinhardt, F. Jung, T. Krüger
At the end of the implantation period, the
electrode was still contacted to the nerve and
continued to measure current impedance.
However, signal recording was limited. After
28 days of implantation, the signal intensity
was decreased and the impedance was increased.
Macroscopical findings
In all animals, 28 days after the implantation,
the cuff electrode stayed at the position targeted by implantation and surrounded the
sciatic nerve perimeter. No pathological finding was observed in the periimplantary subcutaneous and muscular tissues. As shown in
figure 3, the cuff electrode and the electrode
expansion cable including the connectors appeared covered by thin fibrous tissue, well
vascularised and in continuity with the
epineurium (cuff) and perimysium (connectors and expansion cable).
Histological analysis
Semi-thin transverse sections of the cuffed
nerves (light microscopy-phase contrast
mode), showed that the general morphology
of the cuffed nerves was similar to that of the
intact contralateral nerves (fig. 4 A and D).
The transverse-sectional area of the cuffed
and non-cuffed contralateral sciatic nerve at
the branching into the N. peroneus communis, the N. tibialis and N. cutaneus surae caudalis clearly showed the distinct fascicles (tibial, peroneal, sural branches), and the endoneurium compartments encircled by the
perineurium and the outer loose epineurium.
This newly formed tissue layer covering
the cuff electrode was mainly composed of
fibrin (fig. 4 C), flattened fibroblasts and the
newly formed layer was significantly thicker
(26.7±6.8 µm, n=4) than the epineurium (fig.
5, 19.2+5.5 µm). The newly formed layer covered the whole electrode cuff surface including the apical tip of the electrode (fig. 3).
The density of myelinated fibres in the peripheral and central nerve area within the cuff
electrode ranged between 259 and 431 per
Figure 3: Implantation site 28 days after placing of a cuff electrode on the sciatic nerve of rats (n=4).
A: Spiral nerve cuff electrode wrapped around the sciatic nerve and covered by an artificial tissue layer. B: Electrode extension cable and connector (connector A) to the cuff electrode, both covered by
a newly formed tissue layer.
In vivo assessment of tissue compatibility and functionality of a polyimide cuff electrode ...
217
Figure 4: Transverse sectional area of the rat sciatic nerve 28 days after placing of a spiral cuff electrode on it (A) and without a cuff electrode (D, control), haematoxylin-eosin staining; (1a-c) distinct
sciatic nerve fascicles: (1a) tibial, (1b) common peroneal, (1c) sural; (2) Musculus biceps femoris, (3)
newly formed tissue layer covering the cuff electrode, (4) electrode cuff, (5) perimysium; (B, C) electrode-tissue interface after heamatoxylin-eosin staining (B) and fibrin staining (C); samples embedded
acrylic resin, phase contrast microscopy in transmission, primary magnification 10× (A, D) and 40×
(B, C).
Figure 5: Sciatic nerves of rats 28 days after implantation of spiral nerve cuff electrodes (n=4); (A)
Density of myelinated fibres; (B) Thickness of the epineurium and the artificial layer covering the cuff
electrode, gray: nerve with cuff electrode, shaded: nerve without cuff electrode; 5 fields of view analysed per section (n=4), mean ± standard deviation.
10×103 µm2 (peripheral 331 ± 72 per 10×103
µm2, central 332 ± 72 per 10×103 µm2, n=4)
and was similar to those found in the intact
contralateral nerves (peripheral 350 ± 51 per
10×103 µm2, central 373 ± 58 per 10×103
µm2) (fig. 5).
Discussion
The study showed that a spiral cuff electrode
which was placed on the sciatic nerve was
able to record signals for 28 days despite the
fact that the cuff electrode was covered by a
thin, tightly packed layer of fibroblasts and
fibrin as previously reported [33-34]. This tissue layer formation was the result of a foreign
body response which is known as a widely in-
vestigated chronic inflammatory response to
an implant [26-28] and is important in the
prevention of implant failure because the
capsule minimizes adverse effects of the implant on the periimplant tissue [29]. The thickness and cellular morphology of the encapsulation tissue dependent on the shape of the
implant [26-27], the surface texture of an implant [28], and the materials from which it is
fabricated. Changes in the functionality of implanted electrodes to sense signals have been
attributed to the growth of the fibrous tissue
capsule. Formation of the encapsulation tissue changed the signal-to-noise ratio of
recorded nerve compound action potentials
[36], and decreased the selectivity and increased the length dependence of epimysial
electrodes [37].
218
B. Hiebl, S. Bog, R. Mikut, C. Bauer, O. Gemeinhardt, F. Jung, T. Krüger
Further studies will focus on alternative
materials for PI to reduce the limitations in
electrode functionality. Cuff materials, which
allow strong tissue integration [38] might be
helpful to reduce tissue layer formation and
the use of multifunctional polymers with
shape-memory capabilities for cuff design
[39-40] might allow shorter implantation
times.
Conclusion
The minimally-invasively implantable spiral
cuff electrode coated with polyimide (PI) for
insulation could be shown in vivo to allow signal recording despite of a fibrous capsule formation between the electrodes and the
nerve. For this reason the cuff electrode can
be used to record electroneurograms (ENGs)
for instance on the phrenic nerve to record
the respiratory drive.
Acknowledgments
The authors thank Prof. Dr. Thomas Stieglitz,
Laboratory for Biomedical Microtechnology,
Department of Microsystems EngineeringIMTEK, University of Freiburg, Freiburg, Germany, for supporting the study.
References
1. Sahin M et al. Spiral nerve cuff electrode for
recordings of respiratory output. J Appl
Physiol 1997; 83 (1): 317-22
2. Haugland KH, Hoffer JA. Slip information
provided by nerve cuff signals: Application
in closed-loop control of functional electrical stimulation. IEEE Trans Rehabil Eng 1994;
2: 29-36
3. Haugland HK, Sinkjaer T. Cutaneous whole
nerve recordings used for correction of
footdrop in hemiplegic man. IEEE Trans Rehabil Eng 1995; 3: 307-17
4. Naples GG et al. A spiral nerve cuff electrode for peripheral nerve stimulation. IEEE
Trans Biomed Eng 1988; 35 (11): 905-16
5. Nikolic ZM et al. Instrumentation for ENG
and EMG recordings in FES systems. IEEE
Trans Biomed Eng 1994; 41 (7): 703-6
6. Popovic DB et al. Sensory nerve recording
for closed-loop control to restore motor
functions. IEEE Trans Biomed Eng 1993; 40
(10): 1024-31
7. Sahin M, Durand DM, Haxhiu MA. Whole
nerve recordings with the spiral nerve cuff
electrode. in Proceedings of the Sixteenth
Annual International Conference of the IEEE
Engineering in Medicine and Biology Society 1994. Baltimore IEEE
8. Stein RB et al. Principles underlying new
methods for chronic neural recording. Can J
Neurol Sci 1975; 2 (3): 235-44
9. Stein RB et al. Stable long-term recordings
from cat peripheral nerves. Brain Res 1977;
128 (1): 21-38
10. Stein RB, Pearson KG. Predicted amplitude
and form of action potentials recorded from
unmyelinated nerve fibres. J Theor Biol
1971; 32 (3): 539-58
11. Demosthenous A, Triantis IF. An adaptive
ENG amplifier for tripolar cuff electrodes.
IEEE Journal of Solid-state Circuits 2005; 40
(2): 412-21
12. Stieglitz T, Klausmann D, Krueger TB. Development of a simple low noise amplifier for
recording of sensory mass signals from peripheral nerves. Biomed Tech (Berl) 2009;
54 (1): 1-7
13. Albrecht W et al. Preparation of highly
asymmetric hollow fiber membranes from
poly(ether imide) by a modified dry-wet
phase inversion technique using a triple
spinneret. Journal of Membrane Science
2005; 262 (1-2): 69-80
14. Altankov G et al. On the tissue compatibility of poly(ether imide) membranes: an in
vitro study on their interaction with human
dermal fibroblasts and keratinocytes. Journal of Biomaterials Science-Polymer Edition
2005; 16 (1): 23-42
15. Trimpert C et al. Poly(ether imide) membranes modified with poly(ethylene imine)
as potential carriers for epidermal substitutes. Macromolecular Bioscience 2006; 6
(4): 274-284
16. Shamma-Donoghue SA et al. Thin-film multielectrode arrays for a cochlear prosthesis.
IEEE T Electron Dev 1982; 29 (1): 136-44
17. Boppart SA, Wheeler BC, Wallace CS. A
flexible perforated microelectrode array for
In vivo assessment of tissue compatibility and functionality of a polyimide cuff electrode ...
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
extended neural recordings. IEEE Trans Biomed Eng 1992; 39 (1): 37-42
Stieglitz T et al. Micromachined, polyimidebased devices for flexible neural interfaces.
Biomed Microdevices 2000; 2 (4): 283-94
Rubehn B et al. A MEMS-based flexible multichannel ECoG-electrode array. J Neural
Eng 2009; 6 (3): 036003
Grill WM, Veraart C, Mortimer JT. Selective
activation of peripheral nerve fascicles: use
of field steering currents. in Annual International Conference of the IEEE Engineering in
Medicine and Biology Society, 1991
Grill WM, Mortimer JT. Electrical impedance of electrode encapsulation tissue. in
Annual International Conference of the IEEE
Engineering in Medicine and Biology Society, 1992
Kumar S et al. Semmes-Weinstein monofilaments: a simple, effective and inexpensive
screening device for identifying diabetic patients at risk of foot ulceration. Diabetes Res
Clin Pract 1991; 13 (1-2): 63-7
Mulisch M, Welsch U (Eds.). Mikroskopische Technik. 18 ed. 2010, Spectrum: Baltimore
Mikut R et al. The open source Matlab toolbox Gait-CAD and its application to bioelectric signal processing. in DGBMT-Workshop
Biosignal Analysis. 2008. Potsdam
http://sourceforge.net/projects/gait-cad/
Wood NK, Kaminski EJ, Oglesby Rj. The significance of implant shape in experimental
testing of biological materials: disc vs. rod. J
Biomed Mater Res 1970; 4 (1): 1-12
Matlaga BF, Yasenchak LP, Salthouse TN.
Tissue response to implanted polymers: the
significance of sample shape. J Biomed
Mater Res 1976; 10 (3): 391-7
Taylor SR, Gibbons DF. Effect of surface texture of the soft tissue response to polymer
implants. J Biomed Mater Res 1983; 17:
205-27
Lago N et al. Assessment of biocompatibility of chronically implanted polyimide and
platinum intrafascicular electrodes. IEEE
Trans Biomed Eng 2007; 54 (2): 281-90
Klinge PM et al. Immunohistochemical characterization of axonal sprouting and reactive tissue changes after long-term implantation of a polyimide sieve electrode to the
transected adult rat sciatic nerve. Biomaterials 2001; 22 (17): 2333-43
View publication stats
219
31. Richardson RR Jr., Miller JA, Reichert WM.
Polyimides as biomaterials: preliminary biocompatibility testing. Biomaterials 1993; 14
(8): 627-35
32. Sun Y et al. Assessment of the biocompatibility of photosensitive polyimide for implantable medical device use. J Biomed
Mater Res-A 2008; 90 (3): 648-55
33. McNamara A, Williams DF. The response to
the intramuscular implantation of pure metals. Biomaterials 1981; 2 (1): 33-40
34. Stensaas SS, Stensaas LJ. Histopathological
evaluation of materials implanted in the
cerebral cortex. Acta Neuropathol 1978; 41
(2): 145-55
35. Anderson J. Mechanisms of inflammation
and infection wth implanted devices. Cardiovasc Pathol 1993; 2 (3): 33S-41S
36. Lefurge T et al. Chronically implanted intrafascicular recording electrodes. Ann Biomed Eng 1991; 19 (2): 197-207
37. Grandjean PA, Mortimer JT. Recruitment
properties of monopolar and bipolar
epimysial electrodes. Ann Biomed Eng
1986; 14 (1): 53-66
38. Hiebl B et al. Tissue Integration of Two Different Shape-Memory Polymers with Poly(εcaprolactone) Switching Segment in Rats.
Materials Research Society Symposium Proceedings 2009; 1140: 143-149
39. Lendlein A et al. Shape-memory polymers
as a technology platform for biomedical applications. Expert Rev Med Devices 2010; 7
(3): 357-79
40. Behl M, Razzaq MY, Lendlein A. Multifunctional Shape-memory Polymers. Advanced
Materials 2010; 22 (31): 3388-3410
Correspondence address:
Bernhard Hiebl, M.D.
Center for Biomaterial
Development and
Berlin Brandenburg Center for Regenerative
Therapies
Helmholtz-Zentrum Geesthacht
Zentrum für Material- und Küstenforschung
GmbH
Kantstr. 55
14513 Teltow
Germany
bernhard.hiebl@hzg.de