1084
IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009
Single-Photon Avalanche Diode CMOS Sensor for
Time-Resolved Fluorescence Measurements
David Stoppa, Member, IEEE, Daniel Mosconi, Lucio Pancheri, and Lorenzo Gonzo, Member, IEEE
Abstract—A single-photon avalanche diode–based pixel array
for the analysis of fluorescence phenomena is presented. Each
150
m2 pixel integrates a single photon detector
180
combined with an active quenching circuit and a 17-bit digital
events counter. On-chip master logic provides the digital control
phases required by the pixel array with a full programmability of
the main timing synchronisms. The pixel exhibits an average dark
count rate of 3 kcps and a dynamic range of over 120-dB in time
uncorrelated operation. A complete characterization of the single
photon avalanche diode characteristics is reported. Time-resolved
fluorescence measurements have been demonstrated by detecting
the fluorescence decay of quantum-dot samples without the aid of
any optical filters for excitation laser light cutoff.
Index Terms—CMOS, fluorescence lifetime, single-photon
avalanche diode (SPAD), time-gating.
I. INTRODUCTION
N recent years, there has been a growing interest in fast
and low-cost biological testing. However, existing analysisequipment are expensive, bulky and time-consuming, making
their use exploitable for research applications only. Among the
many methods used for biological testing, optical detection is
the most commonly used. In particular, fluorescence lifetime
imaging is an investigation tool of paramount importance in
imaging of molecular processes in physics and life sciences research [1], allowing the mapping of many cell parameters (such
as pH, ion concentrations, etc.) and the detection of pathologies
or DNA sequencing [2].
A typical fluorescence lifetime experiment uses a pulsed
or modulated laser pulse to excite the fluorophores and the
emitted light is revealed by means of intensified CCD cameras
or photomultipler tubes [3], in order to achieve the required
time-resolution and light sensitivity specifications. The performance of these laboratory instruments are excellent in terms
of time accuracy and spatial resolution but they are expensive
and bulky, while there is an increasing demand for portable
and inexpensive biosensors for environmental and biomedical
diagnostics. In the last few years, there have been many efforts
in realizing integrated circuits aimed at biosensing applications
[4], [5], [11]–[15]. The use of a CMOS integrated optical
I
Manuscript received October 02, 2008; revised December 17, 2008; accepted
February 03, 2009. Current version published August 05, 2009. The associate
editor coordinating the review of this paper and approving it for publication was
Dr. Krikor Ozanyan.
The authors are with Fondazione Bruno Kessler—Centro per la Ricerca
Scientifica e Tecnologica (FBK-irst), I-38100, Povo (TN), Italy (e-mail:
stoppa@fbk.eu; mosconi@fbk.eu; pancheri@fbk.eu; lgonzo@fbk.eu).
Digital Object Identifier 10.1109/JSEN.2009.2025581
sensor based on simple photodiodes has been proposed for
fluorescence decay detection [5], but the time resolution was
limited by the RC time constant of the photodiode. High accuracy time resolution can be achieved exploiting single-photon
avalanche diodes (SPADs) [6], but most of the efforts in the
implementation of such systems reported in the literature regard
the integration of the SPAD with quenching circuits, and of
a simple voltage comparator for the digital conversion of the
received Geiger pulse [7]–[9]. Only a few examples of SPAD
arrays with integrated readout channels have been reported so
far. In [10], the authors report a linear SPAD array with an
in-pixel Time-to-Amplitude Converter, used to detect the photons arrival time. In [12], a 128 128 pixel array and a column
Time-to-Digital Converter have been demonstrated. A 16 4
SPAD pixel array with two pixel-level gated counters have been
recently proposed for fluorescence lifetime measurement [13],
and also integrated with micropixellated light-emitting diodes
in a very compact analysis system [14]. We have recently
presented a chip capable of performing time-gated fluorescence
detection [15], which integrates, at the pixel level, an actively
quenched SPAD and a 17-bit digital counter. The chip, which
includes a 7 2-pixel array, was fabricated in a high voltage
CMOS technology.
0.35In this paper, a complete characterization of the chip proposed in [15] is reported, using an FPGA-based control board
and a USB interface for data acquisition. Using this system,
lifetime measurements of quantum dot fluorophores without the
use of optical filter for laser light cutoff have been successfully
performed.
The time-gated fluorescence lifetime measurement technique
is briefly reviewed in Section II, while an overview of the pixel
architecture is carried out in Section III. The chip architecture
is presented in Section IV. Finally, a complete characterization
of the system is reported in Section V.
II. MEASURING TECHNIQUE
The adopted measuring technique is sketched in Fig. 1. A
sub-ns laser pulse illuminates the biological sample containing
the fluorophores, and the low signal associated with the fluorescence emission is detected by the sensor. The adopted measuring technique is based on a time-gated detection method,
where the light signal is detected by using two or more observation windows (OWs). Each window has an externally programmable time width and can be delayed with respect to the
trigger laser pulse by a user-defined time value. The time offset
between the laser pulse and the beginning of the observation
1530-437X/$26.00 © 2009 IEEE
STOPPA et al.: SINGLE-PHOTON AVALANCHE DIODE CMOS SENSOR FOR TIME-RESOLVED FLUORESCENCE MEASUREMENTS
1085
Fig. 2. Cross section of the implemented SPAD.
Fig. 1. Adopted time-gating measuring technique.
window offers the possibility of suppressing unwanted background signals like scattering and auto-fluorescence. This improves the signal-to-background ratio when the light intensity
is measured.
The measurement starts setting the first OW synchronized
with the laser trigger. If an avalanche event is generated within
this window it will be detected by the in-pixel event counter.
The measurement is then repeated for a programmable number
so that a significant statistical population can be obof times
tained and, finally, the pixel-array is read out. After that, other
measurement cycles are performed by using time delayed
observation windows. At the end of the full measurement,
it is possible to sketch a histogram reporting the number of detected events within each OWs. With a proper choice of the OW
time width, it is possible to estimate not only the light intensity (given by the integral of the histogram), but also the time
constant of the fluorescence process [16]. If only one window
is used, the sensor provides an average intensity measurement
which is often satisfactory for many applications (e.g., DNA
micro-arrays).
III. PIXEL ARCHITECTURE
The proposed sensor was designed and fabricated in a high
CMOS technology. The structure of the fabvoltage, 0.35ricated SPADs is similar to that presented in other works in
0.8high voltage CMOS technologies [7]–[10]. A schematic
cross section of the device is shown in Fig. 2. The active region
of the device is made by a
over deep
junction. A
implantation is formed using the
guard-ring surrounding the
layer which is available inside a deep
in this kind
of fabrication processes. The geometry of the device is square
to optimize the area occupation, but the corners are smoothed
so as to avoid electric field peaks at the junction corners. The
active area is defined by means of an optical window opened in
the top-metal light shield only in correspondence with the region where avalanche multiplication occurs.
Fig. 3. Schematic diagram of the proposed pixel.
The schematic diagram of the proposed pixel is sketched in
Fig. 3. It basically consists of a SPAD, a quenching diode-connected transistor Mp2 to limit the avalanche current, a voltage
comparator (INV1) for the avalanche event detection, and a digital counter able to count and store the number of avalanche
events.
To avoid the use of high-voltage transistors, expensive in
terms of area occupation, the cathode of the SPAD is connected
, and the biasing
through the quenching transistor to
of the SPAD is assured
above the breakdown voltage
biased at a negative
by means of the external line
. The main
voltage whose typical value is
drawback of this approach is that the maximum excess voltage
available is below
. The
feedback loop, consisting of INV1 and Mn4, realizes an active
quenching mechanism, able to force the input node to ground
as soon as an avalanche event is detected.
The two control signals Precharge and SPADoff are used to
implement the active reset and quenching functions, respectively. When a photon is absorbed by the SPAD, the avalanche
is triggered and the node OUT is pulled up. The feedback loop
forces this state until the arrival of the next SPADoff pulse, at the
end of each repetition period (quenching-function). This prevents the input node from unwanted recharging. At every rising
edge of the signal CKstop, the event-counter is incremented only
if OUT is high. By so doing, an avalanche event is counted only
within the observation window defined by the rising edges of
Precharge and CKstop.
To reduce the area occupation, the in-pixel 17-bit digital
events counter consists of a pseudorandom number generator,
based on a LFSR (Linear Feedback Shift Register). With a
suitable choice of the bits used for the XOR mask [17], the
different states, so
designed LFSR allows generating
1086
IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009
Fig. 4. Die microphotograph and pixel layout.
m SPADs with V =
Fig. 6. Dark count rate distribution measured on the 20
.
4V
Fig. 5. Block diagram of the chip testing board.
that a maximum of 131 k events can be detected in the proposed pixel. The 17-bit shift registers are based on a dynamic
latch [18] where the initial logic state is set by means
of an internal reset switch. A more detailed description of pixel
architecture and working principle are reported in [15].
IV. CHIP ARCHITECTURE
The chip core is based on two different arrays of SPAD sensors, each one composed by seven pixels. SPADs of the first row
have an active area of about
, while in the second
row
devices have been used. An internal logic
generator provides all the waveforms necessary to the pixel operation, while a configuration register has been implemented for
full programmability of the on-chip digital part. Digital control
signals can optionally be generated externally and provided to
the chip. The microphotograph of the fabricated sensor and the
pixel layout are shown in Fig. 4.
V. EXPERIMENTAL CHARACTERIZATION
The proposed sensor has been mounted on an field programmable gate array (FPGA)-based board, which is schematically shown in Fig. 5. The FPGA was used to generate timing
and configuration signals as well as to manage the communication with a PC via a USB channel. The use of an FPGA
allows the operation of the SPAD in both asynchronous and
synchronous mode using the same hardware, implementing
only firmware modifications. In asynchronous operation, each
SPAD is recharged after a fixed dead time from the occurrence
of an avalanche event. In synchronous operation, the SPAD is
activated only during the observation window and is held in an
off state for the rest of the cycle. The characterization of the
SPAD performance has been conducted in asynchronous operation, while fluorescence measurements have been performed
in synchronous operation mode.
The breakdown voltage of the SPADs, which has been measured on several devices, varies from 27.5 to 28.4 V, and no dependence on the device area has been found.
SPADs is in the order
The dark count rate of
cps at room temperature with an applied excess bias
of
and 1 dead time. This is indeed a very large value,
which renders this detector not suitable for asynchronous operation, even considering cooling of the device. However, these
large SPADs can be used synchronously, activating the devices
for the small amount of time necessary to observe the desired
event and shutting them off immediately after the event has finished. In a 10-ns OW, for example, the probability of observing
a dark event will be about 1% and could be further decreased
with device cooling.
On the contrary, smaller SPADs can be used also in asynchronous operation, thanks to the lower dark count rate. HereSPADs is reported.
after, the characterization of
Dark count rate has been measured on several samples, to take
into account its variability across different SPADs (Fig. 6) the
complete DCR characterization of 17 devices tested on different
.
chips with an excess bias voltage
Almost 60% of the devices have a dark count rate lower than
4 kHz, which is a fairly good value for a SPAD of this size
fabricated in a CMOS technology.
Fig. 7 reports the temperature dependence of dark count rate.
At high temperatures, the slope of the curve is typical of SRH
generation, while at low temperatures the dominant generation
mechanism is tunneling [20]. The corner temperature, at which
the two contributions have the same weight, is around 10 .
Afterpulsing probability density was measured using a digital
oscilloscope to histogram the time delay distribution between
two subsequent pulses, and is shown in Fig. 8. As can be seen,
the detrapping time constants are quite long, in the order of 1 .
STOPPA et al.: SINGLE-PHOTON AVALANCHE DIODE CMOS SENSOR FOR TIME-RESOLVED FLUORESCENCE MEASUREMENTS
m SPADs with
Fig. 7. Dependence of dark count rate on temperature for a 20
.
V = 4V
1087
Fig. 9. Photon detection probability measured at different excess bias voltages.
Fig. 8. Afterpulsing probability density.
Fig. 10. Measured SPAD response to an optical power sweep under a wide
spectrum illumination. The SPAD is biased at
with 1- dead time
and 25-ms time window.
V =4V
Total afterpulsing probability can be reduced by increasing the
dead time, as shown in [8]. In our device, this figure is 4.5% at
setting a 500-ns dead time, but it can be reduced
down to 0.5% by increasing the dead time to 2 .
The spectral response of the SPAD has been measured using
an electro-optical bench composed by a white-light source
(QTH lamp, 250 W), a monochromator filter and a calibrated
reference detector. With this setup, spectral response can be
measured with a maximum relative error of 5% within the
wavelength range 350–1000 nm. The measured spectral dependence of photon detection probability at different applied
excess bias voltages is shown in Fig. 9. The maximum is
reached at about 450 nm, because of the shallow depth of the
/ntub active junction.
The dynamic range of the SPAD in asynchronous operation
has been measured using a wide spectrum lamp. An illumination
power density spanning more than seven orders of magnitude
was obtained on a dedicated electro-optical bench using a set
of neutral filters and varying the distance of the sensor from the
light source. The measured counts in a 25-ms time window as a
function of optical power density incident on the SPAD is shown
s
in Fig. 10, together with the measured noise, obtained as the
standard deviation over a large number of acquisitions.
This curve can be corrected at low count rates by subtracting
the dark counts, In this way, at small illumination power the
dynamic range is limited by the Poisson noise related to dark
counts. At large count rates, when the sum of all the dead times
becomes a relevant fraction of the measuring time window ,
the curve can be corrected by using the relation
(1)
is the corrected counts, and
where is the measured counts,
is the dead time. The noise relative to the corrected counts
can be obtained by applying error propagation to the measured
noise. Both the corrected signal and noise are shown in Fig. 10.
It is worth noting that, with this linearity correction, the sensor
has a dynamic range of six orders of magnitude with a linearity
error lower than 5%.
1088
IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009
Fig. 11. Measured time resolution using a 470-nm pulsed laser with 80-ps pulse
width.
m
Fig. 13. Fluorescence decay measurement performed with a 20- SPAD
coupled to a TCSPC instrument and with our time gated readout channel.
Fig. 12. Fluorescence measurement setup.
The time resolution of the SPAD has been measured using
a pulsed semiconductor laser at 470 nm as a light source. The
laser pulse width was 80 ps FWHM and the histogram was accumulated using a TCSPC instrument [21]. The measurement
is shown in Fig. 11.
at
A fluorescence decay measurement was performed using the
setup shown in Fig. 12. A glass capillary (internal diameter of
) containing a fluorophore solution (CdSe/ZnS quantum
550
dots in toluene [22]) was mounted in close proximity to the
chip surface and a pulsed laser light (470-nm, 80-ps FWHM,
1-MHz repetition frequency, 13-pJ pulse energy) was used to
excite the fluorescence. To demonstrate the sensor capability of
discriminating the fluorescence light from the laser excitation,
and to characterize the sensor performance independently from
the collection optics, a very simple setup has been used as shown
in Fig. 12. As can be seen, neither focusing optics nor optical filters are present. To validate this setup, the photon arrival time
histogram has been measured with a commercial TCSPC instrument [21] (see Fig. 13). Although the sharp laser line, due to
scattered laser photons, is visible in the measurement because
of the absence of optical filters, fluorescence decay could still
be measured accurately.
Fig. 14. Fluorescence decay measurement at different fluorophore concentra, performed with a 20- SPAD and time gated readout channel.
tions
C
m
A second measurement was then performed with the proposed system using the FPGA for data acquisition and transmission to a PC. In this measurement, a 10-ns observation window
has been swept at 60-ps time steps across the time range of interest. The 60-ps time delays could be generated with very good
accuracy by the FPGA. In this way, the convolution of the light
signal with the 10-ns observation window has been measured
and the result is shown in Fig. 13. Because of that, while the
TCSPC instrument produces a narrow line in correspondence
of the laser pulse, the proposed system generates a 10-ns knee
in the left-part of the graph. The average extracted lifetimes are
in good agreement, being 19.54 ns 50 ps and 19.97 ns 40 ps
for the TCSPC and time gated systems, respectively.
Fluorescence decay measurements performed on fluorophore
solutions with different concentrations are shown in Fig. 14. As
STOPPA et al.: SINGLE-PHOTON AVALANCHE DIODE CMOS SENSOR FOR TIME-RESOLVED FLUORESCENCE MEASUREMENTS
expected, fluorescence intensity changes with fluorophore concentration, while the decay curve remains the same. Quantum
dot fluorophores exhibit a multi-exponential decay [23], but the
main signal contribute is within the first tens of nanoseconds,
where the decay curve approximates a mono-exponential decay.
It is worth noting that both intensity and lifetime information
can be extracted from the measurement.
The lifetime measurement setup presented so far offers the
possibility of array integration, at a cost that is much lower than
currently used TCSPC instruments. Moreover, time gated operation allows laser light cutoff without the use of optical filters, with a relevant simplification of measurement setup. The
proposed readout channel can be further improved to include
multiple gating needed to maximize the overall collection efficiency, thus minimizing acquisition times and reducing fluorophore photobleaching.
VI. CONCLUSION
A pixel architecture implementing a SPAD detector and
a dedicated read out circuitry for fluorescence lifetime measurements has been presented. Each pixel is capable of single
photon detection and measures the number of events generated
within a user-defined observation window. A preliminary
test-chip, consisting of a 14-pixel array, has been fabricated
CMOS technology. The characterin a high-voltage 0.35ization of the proposed sensor with an FPGA-based control
board has been shown. The SPADs have shown a six orders
of magnitude dynamic range under asynchronous operation.
Time-gated lifetime measurements of quantum dot fluorophores
without the use of optical filter for laser light cutoff have been
demonstrated.
ACKNOWLEDGMENT
The authors would like to thank Austria Micro Systems for
its support in designing layout structures for the SPADs implementation and Dr. G. Pedretti for his valuable assistance in the
sensor testing.
REFERENCES
[1] J. R. Lakowicz, Principles of Fluorescence Spectroscopy, 2nd ed.
New York: Springer-Verlag, 1998.
[2] R. Cubeddu, D. Comelli, C. D’Andrea, P. Taroni, and G. Valentini,
“Time-resolved fluorescence imaging in biology and medicine,” J.
Phys. D: Appl. Phys., no. 35, pp. R61–R76, 2002.
[3] M. Straub and S. W. Hell, “Fluorescence lifetime three-dimensional
microscopy with picosecond precision using a multifocal multiphoton
microscope,” Appl. Phys. Lett., vol. 73, no. 13, pp. 1769–1771, 1998.
[4] H. Eltoukhy, K. Salama, A. El Gamal, M. Ronaghi, and R. Davis, “A
0.18 um CMOS 1E-6 lux bioluminescence detection system-on-chip,”
in Proc. ISSCC, Feb. 2004, pp. 222–223.
[5] G. Patounakis, K. L. Shepard, and R. Levicky, “Active CMOS array
sensor for time-resolved fluorescence detection,” IEEE J. Solid-State
Circuits, vol. 41, no. 11, pp. 2521–2530, Nov. 2006.
[6] K. K. Ng, Complete Guide to Semiconductor Devices. New York:
Wiley, 2002.
[7] A. Rochas, M. Gani, B. Furrer, P. A. Besse, R. S. Popovic, G. Ribordy,
and N. Gisin, “Single photon detector fabricated in a complementary
metal-oxide-semiconductor high-voltage technology,” Rev. Scientific
Instruments, vol. 74, no. 7, pp. 3263–3270, 2003.
[8] F. Zappa, S. Tisa, A. Gulinatti, A. Gallivanoni, and S. Cova, “Monolithic CMOS detector module for photon counting and picosecond
timing,” in IEEE Proc. ESSDERC, 2004, pp. 341–344.
1089
[9] C. L. Niclass, A. Rochas, P. A. Besse, and E. Charbon, “A CMOS
single photon avalanche diode array for 3D imaging,” in Proc. ISSCC
Digest Tech. Papers, Feb. 2004, pp. 120–121.
[10] D. Stoppa, L. Pancheri, M. Scandiuzzo, L. Gonzo, G.-F. Dalla Betta,
and A. Simoni, “A CMOS 3-D imager based on single photon
avalanche diode,” IEEE Trans. Circuits Syst. I, vol. 54, no. 1, pp. 4–12,
Jan. 2007.
[11] D. E. Schwartz, E. Charbon, and K. L. Shepard, “A single-photon
avalanche diode imager for fluorescence lifetime applications,” in
Proc. Digest Symp. VLSI Circuits, 2007, pp. 144–145.
[12] C. Niclass, C. Favi, T. Kluter, M. Gersbach, and E. Charbon, “A 128
128 single-photon imager with on-chip column-level 10b time-todigital converter array capable of 97 ps resolution,” in Proc. ISSCC,
2008, pp. 44–45.
[13] B. R. Rae, C. Griffin, K. R. Muir, J. M. Girkin, E. Gu, D. R. Renshaw,
E. Charbon, M. D. Dawson, and R. K. Henderson, “A microsystem
for time-resolved fluorescence analysis using CMOS single-photon
avalanche diodes and micro-LEDs,” in Proc. ISSCC, 2008, pp.
166–167.
[14] B. R. Rae, C. Griffin, J. McKendry, J. M. Girkin, H. X. Zhang, E. Gu,
D. R. Renshaw, E. Charbon, M. D. Dawson, and R. K. Henderson,
“CMOS driven micro-pixel LEDs integrated with single photon
avalanche diodes for time resolved fluorescence measurements,” J.
Phys. D: Appl. Phys., no. 41, pp. 1–6, 2008.
[15] D. Mosconi, D. Stoppa, L. Pancheri, L. Gonzo, and A. Simoni, “CMOS
single-photon avalanche diode array for time-resolved fluorescence detection,” in Proc. ESSCIRC, 2006, pp. 564–567.
[16] H. C. Gerritsen, M. A. H. Asselbergs, A. V. Agronskaia, and W. G.
J. H. M. Van Sark, “Fluorescence lifetime imaging in scanning microscopes: Acquisition speed, photon economy and lifetime resolution,”
J. Microscopy, vol. 206, pp. 218–224, 2002.
[17] P. Kitsos, N. Sklavos, N. Zervas, and O. Koufopavlou, “A reconfigurable linear feedback shift register (LFSR) for the bluetooth system,”
in Proc. IEEE Conf. Electron. Circuits Syst., Sep. 2001, vol. 2, pp.
991–994.
[18] J. M. Rabaey, Digital Integrated Circuits. Englewood Cliffs, NJ:
Prentice-Hall, 1996, pp. 351–353.
[19] D. Stoppa et al., “A single-photon avalanche-diode 3D imager,” in
Proc. IEEE ESSCIRC, 2005, pp. 487–490.
[20] A GulinattiI. Rech, P. Maccagnani, M. Ghioni, and S. Cova,
“Large-area avalanche diodes for picosecond time-correlated photon
counting,” in Proc. IEEE ESSDERC, 2005, pp. 355–358.
[21] PicoQuant PicoHarp 300, [Online]. Available: www.picoquant.com
[22] [Online]. Available: www.evidenttech.com/products/evidots.html
[23] U. Resch-Genger, M. Grabolle, S. Cavaliere-Jaricot, R. Nitschke, and
T. Nann, “Quantum dots versus organic dyes as fluorescent labels,”
Nature Methods, vol. 5, pp. 763–775, 2008.
2
David Stoppa (SM’97–M’02) received the Laurea degree in electronics engineering from Politecnico of Milan, Milan, Italy, in 1998, and the Ph.D. degree
in microelectronics from the University of Trento, Trento, Italy, in 2002.
In 2002, he joined the Integrated Optical Sensors Group, Fondazione Bruno
Kessler (FBK-IRST, formerly ITC-IRST), as a Research Scientist, working
on the development of CMOS image sensors for advanced applications. Since
2000, he has been teaching at the Telecommunications Engineering faculty of
the University of Trento courses on analogue electronics and microelectronics.
His research interests are in the design of high-performance CMOS imagers and
biosensors. He has authored or coauthored more than 40 papers in international
journals and presentations at international conferences, and holds two patents
in the field of image sensors.
Dr. Stoppa was a recipient of the Best Paper Award at the 2006 European
Solid-State Circuits Conference.
Daniel Mosconi received the Laurea degree in electronics engineering from the
University of Padua, Padua, Italy, in 2002.
In 2003, he was with the Altran Group, where he worked as Consultant in
Selex Communications, as a DSP Firmware Engineer. In 2005, he joined the
Integrated Optical Sensors Group, Fondazione Bruno Kessler (FBK-IRST), as
a Researcher working on integrated optical sensors and biosensors in CMOS
technology. Since July 2007, he has been with Zobele Holding Spa as an Electronic Engineer. He is currently the Supervisor of the Electronic Center of the
Zobele Group.
1090
Lucio Pancheri received the Laurea degree in materials engineering from the
University of Trento, Italy, in 2002, and the Ph.D. degree in information and
communication technologies from the University of Trento in 2006. During his
Ph.D., he has worked on the design and characterization of novel silicon photodetectors for 3-D imaging applications.
In 2006, he joined the Integrated Optical Sensors Group, Fondazione Bruno
Kessler (FBK-IRST), as a Research Scientist. He is currently a Postdoctoral
Fellow at FBK-IRST, where is involved in the development of optical sensors
for advanced imaging and biomedical applications.
Lorenzo Gonzo (M’97) received the degree in physics from the University of
Trento, Trento,Italy, in 1983 and the Ph.D. degree in physics from the Technical
University of Munich, Munich, Germany, in 1992.
IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009
He is a Senior Scientist at FBK-IRST and Head of the Research Unit Smart
Optical Sensors and Interfaces—SOI. In 1998, he joined the Department
of Surface Sciences, FBK-IRST, where he has been working on tunneling
microscopy and surfaces investigation techniques applied to microelectronics.
Since 1993, he has been working with the Integrated Optical Sensor Group,
FBK-IRST, where he was involved in the development of integrated optical
sensors in CMOS technology. His main research activity has been focused on
sensors for imaging both in 2–D and 3–D and with on chip signal processing.
Since 2006, as Head of the Research Unit SOI, he is coordinating a research
activity on high sensitivity optical sensors, single-photon avalanche diodes,
time gated and multispectral imaging. He has been working also in the field
of metrology with particular attention on 3–D measurements for realistic
multiscale 3–D modeling of objects, buildings and land. He has authored or
coauthored more than 70 papers in international journals and presentations at
international conferences, and holds two patents in the field of image sensors.