Chemical Geology 262 (2009) 344–354
Contents lists available at ScienceDirect
Chemical Geology
j o u r n a l h o m e p a g e : w w w. e l s ev i e r. c o m / l o c a t e / c h e m g e o
In situ Pb-isotope analysis of pyrite by laser ablation
(multi-collector and quadrupole) ICPMS
Jon Woodhead a,⁎, Janet Hergt a,b, Sebastien Meffre b, Ross R. Large b, Leonid Danyushevsky b, Sarah Gilbert b
a
b
School of Earth Sciences, The University of Melbourne, VIC 3010, Australia
ARC Centre of Excellence in Ore Deposits, The University of Tasmania, TAS 7001, Australia
a r t i c l e
i n f o
Article history:
Received 18 September 2008
Received in revised form 2 February 2009
Accepted 5 February 2009
Editor: R.L. Rudnick
Keywords:
Sulfide
Pyrite
Pb-isotope
In-situ analysis
Laser ablation
ICPMS
a b s t r a c t
Pb-isotope ratios, measured in the mineral pyrite, provide a valuable petrogenetic tool with widespread
applicability. In order to interpret complex structural and mineralogical textures, however, a method of insitu analysis is essential. While laser ablation ICPMS is ideally suited to this task, the low melting point of
sulfide, the highly variable and often high Pb contents, and the potential presence of relatively radiogenic
inclusions introduce a number of analytical problems unique to pyrite Pb-isotope analysis. Here we address
these issues using results obtained on two very different analytical systems based around multi-collector and
quadrupole ICPMS instruments respectively. We suggest that controlled ablation of pyrite is only achieved
at low laser fluence and that, under these conditions, standardisation using silicate reference materials is
inappropriate and natural pyrite standards are to be preferred. The inherent variability in Pb (and
sometimes U) concentrations in pyrite requires careful selection of detector systems for optimal analysis and
in this regard both quadrupole and multi-collector ICPMS instruments can play important and
complimentary roles. Multi-collector instruments provide higher precision analyses but detector configurations can prohibit simultaneous measurement of U, Th and Pb. Furthermore, the micrometer-scale variability
in Pb concentrations can cause problems for both Faraday cup and ion counting detection systems. In
contrast, quadrupole ICPMS systems allow simultaneous measurement of U, Th and Pb, and have more
flexible detection systems with many orders of magnitude dynamic range but are unable to produce high
precision data. Results are presented for two different analytical systems and demonstrate a very strong
dependence of data quality upon signal size. In addition they allow some estimation of the limiting precision
obtainable by these methods. Finally, a geological example is provided from the giant Sukhoi Log
sedimentary Au deposit of Russia.
© 2009 Elsevier B.V. All rights reserved.
1. Introduction
The application of radiogenic isotope systems as tracers of both
source and process in the field of economic geology is limited by a lack
of suitable sulfide minerals (i.e. those with low parent/daughter
element ratios enabling preservation of near-initial ratios over time),
coupled with the often uncertain relationship between silicate host
minerals, which may be more amenable to analysis, and sulfide ores.
Studies of Pb-isotope compositions in galena have been and continue
to be highly successful (e.g., Carignan et al., 1993; Carr et al., 1995,
Ayuso et al., 2004) but the mineral pyrite also holds considerable
promise as a phase which is commonly associated with many types of
mineralisation and may contain moderate to high quantities of Pb and
low U, providing the opportunity to determine near-initial Pb-isotope
ratios. Previous Pb-isotope studies of pyrite have largely reported the
use of bulk assemblages (e.g., Ho et al., 1994; Olivo et al., 2004) but the
⁎ Corresponding author.
E-mail address: jdwood@unimelb.edu.au (J. Woodhead).
0009-2541/$ – see front matter © 2009 Elsevier B.V. All rights reserved.
doi:10.1016/j.chemgeo.2009.02.003
typically complex textural relationships found in ore systems
ultimately demand a technique with high spatial resolution while
still retaining the potential for high levels of precision such that subtle
variations can be identified. Two recent studies have demonstrated
the feasibility of using laser ablation ICPMS technologies in this regard
(Mathez and Waight, 2003; Mathez and Kent, 2007) but these were
largely concerned with the application of the results to the petrogenesis of the Bushveld Complex rather optimising an analytical protocol
for sulfide analysis.
The laser ablation analysis of pyrite for Pb-isotope ratio determination is not without technical difficulties. In particular Pb contents
are often highly variable due to micro-inclusions of Pb-rich phases,
providing extreme conditions for most signal detection systems, and
U-rich domains sometimes occur, potentially requiring correction for
radiogenic Pb ingrowth (e.g., Fig. 1). In addition most laser ablation
systems couple very efficiently with sulfides and large-scale melting is
a common, but generally undesirable consequence leading to the
production of large particulates and the resultant possibility for
decreased sensitivity, and enhanced matrix effects (e.g. Jackson and
345
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
Table 1
Analytical conditions for the systems used in this study.
Laser
Type
Wavelength
Pulse width
Repetition rate
Fluence
Ablation
chamber
He gas flow
Ar gas flow
ICPMS
Type
RF power
Shield torch
Dwell time
University of Melbourne
University of Tasmania
Lambda Physik Compex 110
193 nm
25 ns
5 Hz
~ 1 J cm− 2
Helex, low volume (2.5 cm2)
New Wave solid state
213 nm
2 ns
10 Hz
1–2 J cm− 2
Custom made, low volume (3.4 cm2)
0.2 l/min
0.95 l/min
0.7 l/min
1.23 l/min
Nu Plasma MC-ICPMS
1325 W
No
Simultaneous detection
208, 207, 206, 204, 202, 200
Agillent 4500 quad ICPMS
1380 W
No
202, 204, 206, 208: 40 ms
56, 232, 238: 5 ms
System sensitivitya
Spot size
55 µm
Repetition rate
5 Hz
238
U
50 mV (~3,125,000 cps)
139
La
7
Li
248
ThO/232Th
110 µm
10 Hz
300,000 cps
200,000 cps
150,000 cps
0.15%
a
Typical system sensitivities were determined at slightly higher fluence (~ 3 J cm− 2)
than that used for pyrite analyses to enable efficient ablation of the NIST 612 reference
material.
Gunther, 2003). Finally, standardisation to correct for mass bias is an
important issue requiring careful consideration. In the following text
we discuss all of these issues, provide optimum schema for analysis of
pyrite and document our experiences using two different analytical
systems to achieve these goals.
2. Analytical instrumentation and methods
Table 1 provides a summary of key instrumental parameters for the
analytical systems used in this study. All analyses conducted at the
University of Melbourne were performed on a Nu Plasma MC-ICPMS
instrument. For solution analyses, Pb was separated on Eichrom™
SrResin and introduced to the mass spectrometer using an Aridus
desolvation unit, equipped with a Glass Expansion™ OpalMist Teflon
nebuliser operating at an uptake rate of ~ 30 µl min− 1 under free
aspiration. Analyses were corrected for mass bias effects using the
modified thallium doping technique described in Woodhead (2002),
and the SRM 981 values noted therein as reference. In-situ analyses at
the University of Melbourne utilised a HelEx ablation system,
constructed around a Lambda Physik™ Compex 110 excimer laser
operated with ArF providing an output wavelength of 193 nm. The
system has been described in some detail in a number of previous
publications (e.g. Eggins et al., 1998; Woodhead et al., 2004; Eggins
et al., 2005; Woodhead et al., 2005). The laser output energy and
power density were adjusted as described below and the laser was
typically operated at a repetition rate of 5 Hz, with spot sizes ranging
from 90 to 200 µm, depending upon pyrite Pb concentration. Data
were collected in time-resolved mode but all data deconvolution was
Fig. 1. Selected elemental maps for a pyrite grain from the Sukhoi Log sediment-hosted
Au deposit (Russia), with scales in counts per second. Image size is approximately
1.3 × 1.7 mm. The maps were produced at the University of Tasmania using a laser
ablation ICPMS method similar to that employed by Woodhead et al. (2007). The
shape of the pyrite grain is clearly delineated in the Fe concentration plot (top). Note
that the Pb concentration is quite heterogeneous and reaches some very high values
(~1e+ 8 cps). It is also clear that this particular grain contains some areas with
significant U content. Both these features must be accommodated in any successful
analytical protocol.
346
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
undertaken offline in the Iolite software package (Hellstrom et al.,
2008) which allows for detailed visualisation of ion beam intensities
versus time and integration of specific portions of the ablated signal.
In this way it was possible to avoid any complications due to surface
Pb contamination from sample polishing (surficial contamination at
some level is almost unavoidable during polishing of sulfides). Hg
interferences on the 204Pb isotope were corrected using an initial 60 s
‘on peak’ baseline followed by subtraction of any sample-derived Hg
signal by monitoring the 202Hg isotope and peak-stripping.
All analyses conducted at the University of Tasmania were
performed on an Agilent 4500 ICPMS coupled to a New Wave
213 nm solid state laser in a custom made, low volume (3.4 cm3),
barrel-shaped chamber in a He atmosphere. The laser was operated at
5 or 10 Hz using spot sizes between 15 and 110 µm. Data were
collected in time-resolved mode with 30 s gas blank measurement
followed by 60 s analysis time typically drilling at around ~1 µm s− 1.
Data deconvolution was performed using custom made excel-based
spreadsheets. Surface contamination was eliminated by pre-ablating
all spots at the same spot size as used for analysis but a lower
repetition rate (typically 5 s at 1 Hz). Hg interferences were stripped
using the same techniques as described above for the multicollector
instrument. Where necessary, corrections for radiogenic Pb ingrowth
resulting from in situ decay of U and Th within the pyrite or included
minerals were undertaken using measured U/Pb, Th/Pb and Pbisotope ratios and the known or estimated age of the deposit in
question.
3. Some observations regarding power density
When compared to many silicates, sulfides have relatively low
melting points (e.g. that of pyrite is ~1180 °C; Hurlbut et al., 1985). As a
consequence it is frequently observed in the laser ablation of sulfides
that extensive melting occurs with significant melt and condensate
accumulation around the ablation site: this is most readily observed as
a dark halo around the ablation site in reflected light but is also
apparent on SEM images (Fig. 2a and b). These observations clearly
demonstrate that, under such conditions, there is considerable
production of melt droplets often of a relatively large (up to
micrometer) size. Experience with silicate and metal laser ablation
strongly suggests that this is an undesirable trait since sample
accumulation around the ablation pit will not only result in decreased
sensitivity (as less sample is delivered to the plasma) but also any
large melt droplets that do enter the plasma are likely to induce
matrix effects and promote isotopic fractionation (e.g. Jackson and
Gunther, 2003). Although such effects may be less noticeable when
conducting relatively low-precision quadrupole ICPMS measurements, in order to obtain the highest precision and accuracy isotope
ratios by MC-ICPMS in particular it is clearly preferable to minimise
any such phenomena. To this end we performed a variety of ablation
experiments by varying the laser power density on the sample and
observing the effects using optical and SEM imaging. Similar
experiments were performed on both the excimer Ar–F gas
(193 nm) laser at Melbourne and the solid state Nd:YAG (213 nm)
laser at the University of Tasmania. On the excimer laser by using a low
laser output energy, in combination with a 25% transmission beam
splitter we were able to reduce melting effects almost entirely while
still maintaining controlled ablation. For typical silicate analysis the
Melbourne excimer laser system is generally operated at a fluence of
~ 5 J cm− 2 or slightly less whereas we estimate the ideal situation for
sulfide ablation to be in the region of ~ 1 J cm− 2. Using these
conditions, melt/condensate droplets external to the ablation site are
greatly reduced, and thus sample transport to the ICPMS and system
sensitivity are optimised.
Intriguingly, during these experiments, we observed considerable
structure in the ablation pits themselves; in particular, a variety of conelike structures typically appear (Fig. 2c to f). Although not widely
reported, almost identical structures have been observed previously
during the ablation of both polyamides and organic crystals (e.g. Dyer
et al., 1986) and have been termed ‘trulli’ after a form of traditional
dwelling with a cone-shaped roof from Apulia, Italy (Kampmeier et al.,
1997). Little is known of their origin but it is thought that small
imperfections in the sample surface may initially act to shield the
underlying sample from the incoming laser pulses and in this way
surface irregularities begin to develop. In materials that are highly
reflective further laser pulses tend to exacerbate such surface imperfections since glancing impacts on the sides of the cone are insufficient to
ablate the material. Certainly the enhanced ablation we observe around
the base of such structures, producing ‘moats’ lends some weight to this
hypothesis. We have noticed, on occasion, however, that trulli may
extend beyond the sample surface (e.g. Fig. 2f) which suggests that they
may, under some circumstances, be partly constructional in origin,
presumably by adhesion of condensate from the ablation plume.
Although the investigation of the exact origin of this unusual
phenomenon is beyond the scope of this study, it is universally agreed
in the existing literature that such features are only present at values of
laser fluence just above the ablation threshold (Dyer et al., 1986;
Kampmeier et al., 1997). As a result, the appearance of trulli, and lack of
significant ‘blackening’ around the ablation pit (the latter when viewed
in reflected light) can both be used to assess optimal levels of laser
power density for sulfide ablation. In this way efficient transport to the
mass spectrometer can be readily achieved. These findings are entirely
consistent with the recent work of Wohlgemuth-Ueberwasser et al.
(2007) who note that fluence above 3 J cm− 2 produces more melting
and promotes elemental fractionation in sulfides and Jackson and
Gunther (2003) who suggest that higher fluence promotes incomplete
vaporisation and ionisation of large particles in the plasma.
Trulli-like structures were also observed using the University of
Tasmania system. In this case we estimate that optimum fluence for
sulfide ablation will be in the range 1–2 J cm− 2. At low fluence the
count rates increase rapidly with small changes in fluence so that
analyses are strongly affected by small fluctuations in the laser power
and at high fluence, deposition and blackening occurs around the
ablation site, suggesting that melting, and deposition may be
occurring. Optimum ablation conditions were therefore chosen at
the lowest point where large increases in fluence make relatively
small differences to the count rates.
4. Standardisation and data quality
There has been much debate in the literature as to the viability of the
commonly used NIST glasses for calibration of samples with variable
matrices during LA-ICPMS. In the case of sulfides, whereas some
workers have used NIST glasses for external calibration of sulfide
ablation data (e.g. Halter et al., 2004; Sylvester, 2008) there is also a
broad view that, in order to obtain the most accurate elemental data,
some form of matrix-matched material is to be preferred and, indeed a
number of attempts have been made to produce sulfide reference
materials (e.g., Wilson et al., 2002; Wohlgemuth-Ueberwasser et al.,
2007; Danyushevsky et al., in press). While the corresponding case for
matrix effects during isotopic analysis by laser ablation is less welldeveloped (e.g., see Norman et al., 2006 for an example), we have
argued above that optimal ablation of sulfide materials is best performed
under conditions of low laser fluence and, under these circumstances,
calibration with a reference material of radically different matrix is not a
viable option. For sulfide ablation we reduce the laser power density by a
factor of around x5 compared to our normal ablation conditions. Using
these laser output energies it is simply not possible to ablate NIST glass
efficiently since it is close to or below its ablation threshold.
Rather than attempt to produce a synthetic reference material, we
chose instead to characterise a number of natural sulfides for use as
potential standards, an approach which proved relatively straightforward.
As a result of this work two pyrites were identified as likely candidates for
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
laser ablation studies: 700380, is a nodular pyrite from drill-hole 208
(depth 602.7 m) from the Jaguar deposit of western Australia and 110028
is a fine-grained massive sulfide from the CSA Mine in Cobar, New South
Wales. Jaguar is an Archean volcanic hosted massive sulfide deposit
located 4 km south of the Teutonic Mine in central Western Australia at
28.40 S, 121.15E (Barley, 1992). The pyrite nodule (2 cm in diameter) was
initially chosen because it has a high Pb concentration but this also proved
to be a perfect primary calibration standard with both homogeneous Pbisotope composition and relatively homogeneous Pb content (1000–
4000 ppm: Table 2), producing large and constant ion beams under most
ablation conditions. The CSA mine at Cobar is a Devonian Cu–Pb–Zn
deposit in central New South Wales, located at 31.41S, 148.80E (Giles and
347
Marshall, 2004). This pyrite sample was chosen because its Pb isotopic
composition is both homogeneous and contrasts with the Jaguar Deposit
pyrite but, in terms of Pb concentration, it is much more variable (0.2–
2000 ppm: Table 2). As a result it produces analyses more typical of ‘realworld’ pyrites and therefore can be used as a robust secondary standard.
Tables 2 and 3 contain representative trace element analyses from the
University of Tasmania quadrupole ICPMS and the solution based Pbisotope analyses of both materials using the Melbourne MC-ICPMS system,
respectively.
Fig. 3a shows the raw laser ablation MC-ICPMS Pb-isotope data
(corrected for Hg interference on 204Pb but not mass bias) obtained on
the 700380 pyrite—our current primary calibration material—for eight
Fig. 2. Optimisation of the Melbourne laser ablation system. SEM images of the ablation process in pyrites using the 193 nm excimer laser. All ablation pits are ~ 100 μm in diamater.
(a) and (b) demonstrate the results of sulfide ablation using laser fluence values appropriate for silicate analysis. This results in substantial melting and production of a large dark area
around the ablation pit, often visible in reflected light. Closer inspection reveals this area to be littered with melt droplets often approaching 1 μm in size. (c) and (d) show the effects
of lowering the laser fluence to just above the sulfide ablation threshold. Here melt production is greatly reduced, with far less droplet accumulation surrounding the pit, resulting in
higher sensitivity (note that none of these surfaces were cleaned prior to SEM study). Ablation under these conditions is also characterised by the appearance of ‘trulli’ structures
within the ablation pit itself. (e) and (f) demonstrate that, in some cases, trulli appear to extend beyond the rim of the ablation pit, leading to the conclusion that, at least in part, they
may be constructional in nature. The fact that significant ‘moats’ appear around the base of individual trulli is, however, certainly consistent with the conventional interpretation of
their formation. See text for discussion.
348
Table 2
Laser ablation-ICPMS trace element data (ppm), normalised to Fe.
Source
Cr
Mn
Co
Ni
Cu
700380
Jaguar deposit,
western Australia
Ti
14
25
17
5.2
4.8
6.2
4.5
4.4
6.4
4.9
5.9
7.1
5.9
0.8
1.0
2.7
b 0.61
b 2.22
b 2.80
b 2.15
b 2.06
b 2.35
b 2.59
b 2.55
b 2.50
b 3.13
5
7
12
3
3
3
3
3
3
4
4
3
3
261
178
167
109
44
111
423
381
371
375
355
314
339
179
177
164
220
151
158
165
169
195
185
166
157
181
19,482
13,546
18,048
8851
445
2428
2143
416
38,201
3410
4387
2845
11,656
110028
CSA Mine, Cobar,
New South Wales
7.9
5.9
6.2
5.3
5.5
8.6
4.7
14
298
7.6
b 2.43
b 2.34
b 2.19
b 2.49
b 2.30
b 2.50
b 2.20
b 1.8
b 1.1
b 1.9
12
3
2
15
b0.8
4
3
3
33
4
33
1.2
75
4.4
0.5
0.9
22
6.3
34
0.9
Analytical methods after Danyushevsky et al. (in press).
1.0
3.2
3.9
3.2
b 0.9
1.9
1.4
1.4
1.3
1.4
Zn
5.8
3.2
3.2
57
1.2
5.3
3.5
83
61
3.9
As
Se
Zr
Mo
Ag
Cd
285
2488
296
85
20
275
61
11
58
88
213
35
145
1624
1416
1449
1548
1230
1511
1833
1690
1722
1668
1693
1603
1726
17
18
20
22
8.8
25
18
13
22
27
16
20
20
0.3
0.8
2.8
0.01
0.1
0.2
b 0.05
0.1
b 0.04
b 0.04
b 0.04
0.1
b 0.05
0.1
0.2
0.2
b0.1
b0.4
b0.4
b0.3
b0.3
b0.5
b0.4
b0.5
b0.4
b0.5
109
96
85
115
40
92
128
76
114
101
92
79
141
1.3
9.1
1.5
0.8
b 1.0
b 1.6
b 1.2
b 1.5
b 1.6
b 1.7
b 1.1
b 1.3
1.3
4
b 0.7
b1
14,419
4
b 1.2
724
301
4229
15
175
211
112
1711
33
237
141
302
1052
94
42
b4
17
b5
6.9
5.8
7.7
6.2
14
9.0
1.5
0.2
0.9
b 0.06
b 0.05
0.8
3.2
1.1
1.8
2.2
b0.6
b0.4
b0.5
b0.5
b0.4
b0.4
b0.4
b0.2
b0.1
b0.1
b 0.1
b 0.1
b 0.1
7
b 0.1
0.1
b 0.1
0.3
6.1
0.1
b 1.3
b 1.5
b 1.2
20
b 1.2
b 1.3
b 1.3
0.5
7.3
b 0.1
Sn
Sb
48
212
113
48
15
76
45
14
26
47
87
39
33
1065
1022
987
1126
674
953
1183
1005
1134
1120
1119
1002
1115
0.2
2.2
0.1
2.2
b 0.1
0.2
b 0.1
2.6
1.7
0.3
8.2
16
3.9
66
b 0.08
9.7
3.9
6.6
37
4.0
Te
3.2
3.1
2.7
4.0
1.9
3.9
6.5
2.2
3.1
3.7
2.7
5.5
3.3
b1.5
b1.1
b0.8
b1.2
b1.0
b1.2
b1.1
b0.1
0.1
0.1
Ba
La
W
Au
3.8
6.2
4.6
2.8
1.3
3.3
2.5
1.9
1.7
2.8
4.4
2.2
2.2
0.06
0.14
0.21
0.03
0.07
0.06
0.04
b 0.01
b 0.01
0.02
0.09
0.04
0.02
0.2
0.1
0.1
0.2
0.1
0.2
0.2
0.1
0.2
0.2
0.2
0.2
0.1
5.1
4.9
4.0
5.7
3.4
4.6
6.0
4.4
5.2
5.0
4.1
4.6
4.9
b0.16
b0.12
b0.13
b0.09
b0.11
b0.15
b0.16
b0.04
b0.1
b0.1
b 0.01
b 0.01
b 0.02
b 0.02
b 0.01
b 0.01
b 0.01
0.01
0.27
0.18
b0.13
b0.08
b0.08
b0.08
b0.04
b0.06
b0.05
b0.02
1.2
b0.02
b 0.1
0.1
b 0.07
1.1
b 0.1
0.1
b 0.05
0.2
1.0
0.01
Tl
Pb
Bi
33
25
35
24
31
30
23
32
29
28
29
33
28
2570
2902
2720
2537
1141
3389
2528
1432
3091
3850
1500
2940
2835
189
206
175
229
130
195
219
179
210
209
181
192
202
0.1
0.2
b0.03
0.1
b0.04
0.7
0.0
0.1
10.5
0.1
143
215
83
2070
0.2
137
67
117
785
82
0.04
b0.02
b0.02
0.21
b0.02
b0.03
0.04
1.4
9.3
0.03
Th
U
0.01
0.07
0.06
0.01
b 0.01
b 0.01
b 0.01
b 0.01
b 0.01
0.02
b 0.01
b 0.02
b 0.01
0.01
0.02
0.05
0.01
b0.01
b0.02
b0.01
b0.01
b0.01
b0.01
b0.01
b0.01
b0.02
0.11
b 0.01
0.04
b 0.02
b 0.01
0.08
0.19
0.05
0.11
0.25
0.11
0.01
0.03
b0.01
b0.01
0.10
0.09
0.03
0.04
0.09
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
Sample
349
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
Table 3
Solution MC-ICPMS Pb-isotope ratios for our two reference pyrites, corrected for mass bias using Tl-normalisation methods as documented in Woodhead (2002).
Pb/206Pb
Sample
700380
110028
Aliquot 1
Aliquot 2
Aliquot 3
Mean and standard deviation
Aliquot 1
Aliquot 2
Aliquot 3
Mean and standard deviation
207
2 s.e.
208
Pb/206Pb
1.087757
1.087830
1.087857
1.087815
0.862775
0.862894
0.862853
0.862841
0.000014
0.000014
0.000014
0.000052
0.000009
0.000010
0.000009
0.000060
2.483143
2.483347
2.483453
2.483315
2.112382
2.112713
2.112636
2.112577
analytical sessions conducted over a period of 2 years. Although mean
values for each session can differ considerably as a result of variable
mass spectrometer parameters (e.g. cleanliness of cones, torch
position, gas flows) the data within any one analytical session show
a remarkable degree of consistency with levels of reproducibility
comparable to typical thermal ionisation mass spectrometry, which is
often considered the benchmark analytical method for Pb-isotope
2 s.e.
206
Pb/204Pb
2 s.e.
207
Pb/204Pb
2 s.e.
208
Pb/204Pb
0.000025
0.000037
0.000035
0.000157
0.000023
0.000026
0.000022
0.000173
13.3589
13.3606
13.3598
13.3598
18.1165
18.1160
18.1150
18.1158
0.0006
0.0007
0.0008
0.0009
0.0007
0.0006
0.0009
0.0007
14.5310
14.5340
14.5336
14.5329
15.6305
15.6324
15.6307
15.6312
0.0007
0.0007
0.0009
0.0017
0.0006
0.0005
0.0008
0.0011
33.1722
33.1793
33.1792
33.1769
38.2689
38.2747
38.2707
38.2715
2 s.e.
0.0015
0.0016
0.0019
0.0041
0.0015
0.0012
0.0018
0.0030
analysis (illustrated in the figure). We interpret this high level of
consistency as a direct result of our attempts to achieve stable and
reproducible laser ablation conditions (see above) although we also
readily admit that we have not conducted any extensive experiments
under conditions of higher fluence with which to compare these data
simply because we view this as being potentially detrimental to our
ICPMS detection systems. Certainly, however, there is evidence in the
Fig. 3. (a) Raw (corrected for Hg interference but not mass bias) Pb-isotope ratios obtained for the 700380 pyrite during 8 laser ablation MC-ICPMS analytical sessions conducted over
a period of 2 years. Separate sessions are distinguished using alternate open and closed symbols, and separated by vertical hatched lines. Numbers represent percent standard
deviation for any particular session. Although there is a significant range in values depending upon specific ICPMS conditions on any given occasion, variation within any given
analytical session is small and comparable to what might be observed during conventional TIMS Pb-isotope analysis (conventional TIMS and double spike-TIMS/solution MC-ICPMS
uncertainties provided for comparison). (b) Raw (corrected for Hg interference but not mass bias or drift) Pb-isotope ratios obtained for the 700380 pyrite during 16 quadrupoleICPMS analytical sessions, covering a similar time period. Note different scale. The precision of individual measurements is ~ 10 times larger than on the multi-collector instrument
but there nevertheless remains a high degree of internal consistency within individual analytical sessions.
350
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
literature for variable isotopic fractionation during ablation of copper
with different laser pulse energies (Jackson and Gunther, 2003) and
thus we prefer to avoid conditions of extensive sample melting. Fig. 3b
illustrates 700380 data for the quadrupole instrument analysed over a
similar period. These display within-run errors ~ 10 times greater than
the MC-ICPMS data as expected for a single-collector instrument but,
nevertheless, within any one analytical session, data exhibit a high
degree of consistency.
In our current methodology, analyses of this reference pyrite
throughout an analytical session are then used to determine mass bias
and hence make a so called ‘external’ correction for this effect during
processing of unknowns, using either an average value for the entire
analytical session (usually the case) or a calibration line versus time if
fractionation factors have drifted significantly during the session (which
we rarely observe). Using this procedure, corrected isotope ratios are
obtained and, as a result, these within-session variations in the analysis of
the calibration material represent a first-order control on the quality of
unknown analyses. As an illustration of this methodology using the MCICPMS instrument, Fig. 4 shows data for our secondary standard, which
closely approaches many pyrites in terms of Pb content and variability.
These data, now corrected for both Hg interference and mass bias, cover
the same analytical period as the calibration standards shown in Fig. 3a.
Levels of accuracy,1 that is the degree to which the data approach the
assumed true values (in this case the Tl-normalised bulk-pyrite solution
MC-ICPMS data of Table 3) are typically around ~1500 ppm for ratios
involving 204Pb, dropping to ~300 ppm for 208Pb/206Pb and ~900 ppm for
207
Pb/206Pb ratios. Throughout the period over which the work described
here was conducted, these fractionation and drift-corrected Pb-isotope
analyses show a relative standard deviation of 0.13 and 0.15% for 206Pb/
204
Pb and 207Pb/204Pb ratios and 0.09% and 0.03% for 207Pb/206Pb and
208
Pb/206Pb ratios respectively. Such reproducibility is only a factor of ~2–
3 larger that of conventional (i.e. not double or triple spiked) Pb isotope
analysis of bulk samples by thermal ionisation mass spectrometry, which
is remarkable considering the minimal sample preparation involved and
small amount of sample material consumed.
For most pyrite analyses, within-run analytical precision is highly
dependent upon Pb content. Fig. 5 shows typical 2 s.e. within-run
error estimates for analyses obtained using both quadrupole and MCICPMS instruments across a wide range of signal sizes. No attempt is
made here to compare in detail the sensitivities of the two systems
since both were coupled to quite different lasers and thus there are too
many variables involved to make any meaningful comparison
(although an approximate indication of relative sensitivities for
silicate analysis can be obtained from Table 1). These diagrams
illustrate the very strong relationship between signal size and withinrun precision which is common to both analytical systems. In addition
they provide an estimate of the limiting precision of each technique,
given maximum signal size. For MC-ICPMS, this is around 0.005 (2 s.
e.) for the 206Pb/204Pb ratio and 0.001 for the 207Pb/206Pb ratios
whereas, for quadrupole ICPMS, these figures are ~0.2 and ~0.005
respectively. For determination of the 207Pb/206Pb ratio the MC-ICPMS
limiting precision is thus ~ 10 times better than the quadrupole data
while for the 206Pb/204Pb ratio this is a factor of ~ 40, reflecting the
difficulties inherent in measurement of the 204Pb signal and accurate
correction for 204Hg interference using a less sensitive single collector
instrument with sequential rather than simultaneous signal detection.
We do note, however, that when using quadrupole (and indeed MC-)
ICPMS it is possible to improve these figures by pooling the results of
multiple analyses. There is an obvious loss of spatial resolution using
1
Accuracy is defined here as the difference between measured and fractionation
corrected ratios for this pyrite and the assumed true ratio as defined by solution
analysis. For a group of analyses, accuracy (usually expressed in ppm) can be calculated
as A = [ΣA2i /(N − 1)]1/2 where Ai = 1,000,000 × (Ri′ − Rtr)/Rtr, R′ is the measured and
corrected ratio for a standard, and Rtr is the true ratio. After Platzner et al. (1997), page
187.
such a method; however, if different pyrite generations are clearly
visible optically, this technique can improve precision considerably.
The pooling of the analyses also allows possible outliers to be
identified and can be performed quickly due to the rapidity of the
technique (1–2 min per analysis) but, of course, must be undertaken
with a degree of caution in terms of the interpretation of the resulting
‘average’ analyses.
5. Optimal analytical strategies and an example from Sukhoi Log
Based upon our observations in analysing many thousands of
grains, it is clear that there is no one ideal analytical strategy for in-situ
Pb-isotope determination of pyrites. Published (e.g. Large et al., 2007)
and unpublished laser ablation data obtained at the University of
Tasmania from a variety of ore deposit types suggest that pyrites can
be highly variable in Pb content. Although most ore deposits contain
pyrite with around 100 ppm Pb, VHMS, SEDEX, lode gold, IOCG and
epithermal Au deposits may be more enriched (50–10,000 ppm) and
porphyry Cu–Au and skarn systems less so (0.1–50 ppm). A rather
more limited dataset of analyses for pyrite not associated with ore
deposits suggests that somewhat lower Pb concentrations (0–
10 ppm) prevail in these types of sample, although diagenetic pyrite
in black shales can contain Pb up to 500 ppm. Studies of
metamorphosed ore deposits generally show that the paragenetically
early pyrite, with colloform, nodular or spongy textures, contains
more Pb than later recrystallised euhedral pyrite. In addition it is
frequently observed that individual pyrite crystals can show Pb
concentrations which vary by several orders of magnitude on the scale
of micrometers, largely due to the presence of microscopic galena
inclusions, as noted above (see Fig. 1, and Table 2). Together, the high
and inherent variability in Pb concentrations impose severe constraints upon any analytical method.
In our studies we have considered two different analytical approaches
using multi-collector and quadrupole ICPMS instruments respectively;
each has its advantages and disadvantages. Ultimately MC-ICPMS
instruments are capable of producing more precise data than quadrupole
mass spectrometers, due to flat-topped peaks, simultaneous detection of
all relevant ion beams eliminating noise associated with so-called ‘plasma
flicker,’ and the generally higher sensitivity of sector-based instruments.
Despite this, the extremely variable and often unstable signals observed
during pyrite ablation make the use of multi ion counting an unattractive
proposition with the real possibility of damage to the detectors. In
contrast, Faraday cup detection systems can potentially cope with
relatively large signals (10–50 V per channel on the commonly available
mass spectrometers) but have a much reduced dynamic range compared
to the electron multiplier units employed on quadrupole-ICP instruments
and, in addition, can have problems with rapidly changing beam sizes if socalled tau corrections are not correctly implemented. Tau effects result
from the relatively slow response of the Faraday cup preamplifiers to
changes in signal intensity (which can be a few seconds) compared to ion
counters. Although algorithms are usually incorporated into the instrument software to calibrate this response, problems sometimes occur with
the most typical symptom being an apparent change in isotope ratio
correlated with rapid change in signal intensity (e.g. Hirata et al., 2003;
Paul et al., 2005). The veracity of the tau correction is easily monitored by
measuring an isotope pair simultaneously in a Faraday cup-ion counter
configuration and opening/closing the laser shutter, thereby simulating
rapidly changing signal intensity. Measured isotope ratios should remain
constant throughout this process.
In contrast to multi-collector instruments, quadrupole ICP systems
are incapable of such high precision and the accuracy of measurements can be compromised when detectors switch mode (e.g. from
pulse to analog detection), or if they become saturated with high
count-rates. The likely occurrence of such scenarios is exacerbated by
the rapidly changing signals often observed during sulfide ablation.
Nevertheless, the use of quadrupole ICPMS instruments does confer a
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
351
Fig. 4. Pb isotope ratios for the 110028 pyrite, obtained using the multi-collector instrument, and corrected for Hg interference on 204Pb and mass bias. Levels of accuracy and
precision are noted on each panel, while the dashed line represents the ratio determined by high-precision solution MC-ICPMS. Within-run errors are highly variable and strongly
controlled by signal size (see later section).
352
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
Fig. 5. Typical Pb-isotope analyses by MC- and quadrupole-ICPMS (most analyses of 30–60 s duration) comparing within-run precision (quoted as 2 s.e.) relative to total ion beam
intensity (note different scales on both axes). All plots show the strong dependence of within-run precision on signal size, although those of the multi-collector instrument attain
lower absolute values for 2 s.e. (see arrows for estimate) at high signal size.
number of advantages. In particular, their detection systems are
optimised for use over a very large dynamic range (typically 8–9
orders of magnitude) and thus have the ability to analyse minerals
with a wide variety of Pb contents. Furthermore, as noted above,
pyrites can contain significant contents of U, which can be attributed
to micro-inclusions of other minerals such as zircon, monazite and
uraninite (Fig. 1). In such cases it may be necessary to apply an age
correction to the Pb-isotope data obtained and this is not possible
unless U, Th, and Pb contents are measured concurrently with Pbisotope ratios. Although some MC-ICPMS instruments do have
specialized collector blocks which are designed for static multicollection of U–Th–Pb isotope ratios (e.g., Simonetti et al., 2005) these
are generally inappropriate for pyrite ablation studies since they
incorporate ion counter detection for some or all of these ion beams.
Clearly in these cases the advantages of simultaneous U, Th, and Pb
determination by quadrupole ICPMS outweigh the disadvantages of
lower precision on isotope ratio determination.
Thus, one might consider the following strategy for optimal
analysis. Under circumstances in which large (N200 μm) grains are
available, with moderate Pb contents (hundreds to thousands of
ppm) MC-ICPMS will usually produce high-quality data comparable
to TIMS or solution-based MC-ICPMS analyses. In some circum-
stances, however, it will be more advantageous to utilise quadrupoleICPMS analysis, for example in samples where very large ranges in Pb
content are to be expected, or where Pb contents are very low (tens
of ppm) and so the Faraday collectors employed on MC-instruments
are inappropriate detection systems. In addition some pyrites will
have significant uranium concentrations and thus require an age
correction to the Pb-isotope data, a feature which can only be
assessed with reconnaissance trace element analyses or elemental
mapping techniques (e.g., Fig. 1). In this case the availability of
simultaneously determined U, Th, and Pb concentration data will
prove invaluable, even at the expense of relatively poorer precision
on the isotope ratio determination. In rare cases pyrite will prove
intractable to both MC- or quadrupole-ICPMS analysis, most often as
a result of extremely low Pb concentrations, possibly in conjunction
with significant levels of sample Hg. In such cases we would advise
the use of a micro-drilling technique coupled with low-blank Pb
chemistry and analysis using high-precision TIMS or solution MCICPMS methods.
As an example of the utility of pyrite Pb-isotope analysis we may
consider the analyses of pyrites from the giant Sukhoi Log sedimentary Au deposit in Russia. Petrogenetic interpretations of this dataset
have been published elsewhere (Meffre et al., 2008) but here we
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
simply consider the data themselves. Sukhoi Log is one of Russia's
largest gold deposits and contains textural, structural and chemical
evidence for multiple geological processes including diagenesis,
metamorphism and post-metamorphic influences. Mineralisation is
intimately associated with multiple pyrite generations and Large et al.
(2007) have documented the paragenesis and trace element geochemistry of these different pyrite phases. These authors have
identified an early stratiform arsenian pyrite containing structurallybound (invisible) gold from 0.5 to 12 ppm Au, and later coarsergrained pyrites relatively depleted in invisible gold, formed through
late diagenesis and metamorphism.
Fig. 6 shows the results of our combined quadrupole ICPMS and
MC-ICPMS analyses of the Sukhoi Log pyrites and illustrates many of
the pertinent features discussed above. Firstly we note that analyses
from both multi-collector and quadrupole instruments are highly
consistent, giving confidence in the analytical protocols at both the
University of Melbourne and CODES laboratory at the University of
Tasmania. Levels of precision approaching conventional TIMS analyses
can be obtained from single spots on the MC-ICPMS. Similar levels of
precision on quadrupole ICPMS analyses can be achieved by pooling
multiple analyses, as is the case in this study. Clear distinctions in
isotopic composition are noted between different pyrite morphologies
and generations. These variations are unrelated to radiogenic ingrowth
since the pyrites analysed in this case contain very little U and hence
are believed to represent primary compositions. The Pb isotope data,
when viewed in the context of mineralogical texture, provide
compelling evidence for a protracted history of pyrite formation, as
proposed by Large et al. (2007), involving multiple sources of Pb
throughout the history of the Sukhoi Log deposit. Combining these
data with U–Th–Pb geochronology of zircon and monazite from the
deposit, Meffre et al. (2008) were able to define a pre-deformation
phase of Au-bearing pyrite growth, formation of Pb-rich pyrite during
sedimentation or early diagenesis, and Pb bearing pyrite with free gold
inclusions, throughout late Neoproterozoic deformation and subsequent metamorphic and hydrothermal events in the Palaeozoic. Early
stratiform pyrites contain Pb with a rather distinctive signature
compared to the rest of the deposit which Meffre et al. (2008) suggest
353
hints at an old organic rich shale source. Elucidation of the details of
these processes would not have been possible without the ability to
perform in-situ analyses and thus investigate complex textural and
structural relationships between successive multiple pyrite generations. For further petrogenetic and geochronological interpretation
readers are referred to Large et al. (2007) and Meffre et al. (2008).
6. Conclusions
In-situ Pb isotope analysis of pyrite by laser ablation ICPMS can be a
valuable tool in the investigation of the often protracted and complex
paragenetic processes occurring in mineralised systems. Such analyses present some unique analytical challenges, however. Laser
ablation protocols should ideally be optimised to produce the least
possible sample melting during ablation, typically by lowering laser
power densities considerably compared to those used for silicate
ablation. Under such circumstances intriguing textures are observed
in the ablation pits, which have been termed trulli by some authors.
Although their exact origin is unclear, trulli provide a useful indication
of laser fluence just above the ablation threshold and which we
believe is ideal for pyrite analysis.
Using natural pyrite reference materials it is possible to obtain insitu Pb-isotope data of a quality approaching conventional TIMS data,
based upon bulk analyses. Within-run precision is strongly dependent
upon signal size, with multi-collector instruments providing better
precision than quadrupole instruments. Natural pyrites are, however,
characterised by highly variable and often high Pb contents and, in
some cases, significant U contents: optimal Pb isotope analysis
requires careful attention to these characteristics. Under ideal
circumstances, where pyrites contain hundreds to thousands of ppm
Pb, MC-ICPMS analyses provide highly precise and accurate data.
Quadrupole ICPMS data can be utilised where Pb concentrations are
inappropriate for Faraday cup detection or where age corrections for
radiogenic ingrowth may be required. In such cases, pooling of data
from multiple spot analyses can help to improve analytical precision
markedly but this ‘averaging’ of data points must always be undertaken with care.
Fig. 6. A 208Pb/204Pb vs. 206Pb/204Pb plot for pyrites from the Sukhoi Log stratiform Au deposit of Russian (data replotted after Meffre et al., 2008 to distinguish quadrupole, open
symbols, from MC-ICPMS, closed symbol, data). Errors shown are 95% confidence. Quadrupole analyses represent 4 to 6 pooled analyses. Four different pyrite types, distinguished on
textural and temporal characteristics show distinct isotopic signatures. Each pyrite type is indicated by a different symbol and shaded field; there is also a strong suggestion in the
data (not explored by Meffre et al., 2008), that the large euhedral pyrites may contain two distinct populations (separated in this figure by dashed line). Note that these data are
plotted uncorrected for any radiogenic ingrowth since the magnitude of this correction at Sukhoi Log is mostly small (b 0.2 difference in the 206Pb/204Pb ratio) compared to the
isotopic differences observed between the various pyrite types (see Meffre et al., 2008 for further analytical details and the text for discussion.
354
J. Woodhead et al. / Chemical Geology 262 (2009) 344–354
Acknowledgements
Mike Shelley is thanked for his insights into the origins of trulli
structures and Chad Paton for programming assistance with the Iolite
code. We thank Adam Kent and one anonymous reviewer for the
detailed comments which have helped to focus our thoughts and
improve the manuscript significantly. This work was funded by the
ARC CODES Centre of Excellence in Ore Deposits.
References
Ayuso, R.A., Kelley, K.D., Leach, D.L., Young, L.E., Slack, J.F., Wandless, J.F., Lyon, A.M.,
Dillingham, J.L., 2004. Origin of the Red Dog Zn–Pb–Ag Deposits, Brooks Range,
Alaska: evidence from regional Pb and Sr sources. Economic Geology 99,
1533–1553.
Barley, M.E., 1992. A review of Archean volcanic-hosted massive sulfide and sulfate
mineralization in Western Australia. Economic Geology 87 (3), 855–872.
Carignan, J., Machado, N., Gariepy, C., 1993. Pb isotope composition of Ni–Cu–Pb ore
deposits in an Archean greenstone bet; evidence for Proterozoic remobilisation in
the Pontiac Subprovince of Canada. Economic Geology 88, 709–715.
Carr, G.R., Dean, J.A., Suppel, D.W., Heithersay, P.S., 1995. Precise lead isotope
fingerprinting of hydrothermal activity associated with Ordovician to Carboniferous metallogenic events in the Lachlan Fold Belt of New South Wales. Economic
Geology 90, 1467–1505.
Danyushevsky, L., Robinson, P., Gilbert, S., Norman, M., Large, R., McGoldrick, P., Shelley, M., in
press. A technique for routine quantitative multi-element analysis of sulphide minerals by
laser ablation ICP-MS geochemistry, exploration, environment, analysis.
Dyer, P.E., Jenkins, S.D., Sidhu, J., 1986. Development and origin of conical structures on
XeCl laser ablated polyimide. Applied Physics Letters 49, 453–455.
Eggins, S.M., Kinsley, L.P.J., Shelley, J.M.G., 1998. Deposition and element fractionation
processes during atmospheric pressure laser sampling for analysis by ICP-MS.
Applied Surface Science 127–129, 278–286.
Eggins, S.M., Grun, R., MczCulloch, M.T., Pike, A.W.G., Chappell, J., Kinsley, L., Mortimer,
G., Shelley, M., Murray-Wallace, C.V., Spotl, C., Taylor, L., 2005. In situ U-series dating
by laser ablation multi-collector ICPMS: new prospects for Quaternary geochronology. Quaternary Science Reviews 24, 2523–2538.
Giles, A.D., Marshall, B., 2004. Genetic significance of fluid inclusions in the CSA Cu–Pb–
Zn deposit, Cobar, Australia. Ore Geology Reviews 24 (3–4), 241–266.
Halter, W.E., Pettke, T., Heinrich, C.A., 2004. Laser-ablation ICP-MS analysis of silicate
and sulfide melt inclusions in an andesitic complex I: analytical approach and data
evaluation. Contributions to Mineralogy and Petrology 147, 385–396.
Hellstrom, J., Paton, C., Woodhead, J., Hergt, J., 2008. Iolite: software for spatially
resolved LA-(quadrupole and MC) ICPMS analysis. Mineralogical Association of
Canada short course series 40, 343–348.
Hirata, T., Hayano, Y., Ohno, T., 2003. Improvements in precision of isotopic ratio
measurements using laser ablation-multiple collector-ICP-mass spectrometry:
reduction of changes in measured isotopic ratios. Journal of Analytical Atomic
Spectrometry 18, 1283–1288.
Ho, S.E., McNaughton, N.J., Groves, D.I., 1994. Criteria for determining initial lead
isotopic compositions of pyrite in Archaean lode–gold deposits: a case study at
Victory, Kambalda, Western Australia. Chemical Geology 111, 57–84.
Hurlbut, C.S., Klein, Cornelis, 1985. Manual of Mineralogy, 20th Ed. John Wiley and Sons,
New York, pp. 285–286.
Jackson, S.E., Gunther, D., 2003. The nature and sources of laser induced isotopic
fractionation in laser ablation multicollector inductively coupled mass spectrometry. Journal of Analytical Atomic Spectrometry 18, 205–212.
Kampmeier, J., Dreisewerd, K., Schurenberg, M., Strupat, K., 1997. Investigations of 2,5DHB and succinic acid as matrices for IR and UV MALDI. Part 1. UV and IR laser
ablation in the MALDI process. International Journal of Mass Spectrometry and Ion
Processes 169/170, 31–41.
Large, R.R., Maslennikov, V.V., Robert, F., Danyushevsky, L.V., Scott, R.J., Chang, Z., 2007.
Multi-stage sedimentary and metamorphic origin of pyrite and gold in the giant
Sukhoi Log deposit, Lena Goldfield, Russia. Economic Geology 102, 1233–1267.
Mathez, E.A., Waight, T.E., 2003. Lead isotopic disequilibrium between sulfide and
plagioclase in the Bushveld Complex and the chemical evolution of large layered
intrusions. Geochimica et Cosmochimica Acta 67, 1875–1888.
Mathez, E.A., Kent, A.J.R., 2007. Variable initial Pb isotopic compositions of rocks
associated with the UG2 chromitite, eastern Bushveld Complex. Geochimica et
Cosmochimica Acta 71, 5514–5527.
Meffre, S., Large, R.R., Scott, R., Woodhead, J., Chang, Z., Gilbert, S.E., Danyushevsky, L.V.,
Maslennikov, V., Hergt, J.M., 2008. Age and pyrite Pb-isotopic composition of the
giant Sukhoi Log sediment-hosted gold deposit, Russia. Geochimica et Cosmochimica Acta 72, 2377–2391.
Norman, M.D., McCulloch, M.T., O, ' Neill, H.St.C., Yaxley, G.M., 2006. Magnesium isotopic
analysis of olivine by laser ablation multi-collector ICP-MS: composition dependent
matrix effects and a comparison of the Earth and Moon. Journal of Analytical
Atomic Spectrometry 21, 50–54.
Olivo, G.R., Isnard, H., Williams-Jones, A.E., Gariépy, C., 2004. Pb isotope compositions of pyrite
from the C quartz–tourmaline vein of the Siscoe gold deposit, Val D'Or, Quebec:
constraints on the origin and age of gold mineralization. Economic Geology 102,137–146.
Paul, B., Woodhead, J.D., Hergt, J.M., 2005. Improved in situ isotope analysis of low-Pb
materials using LA-MC-ICPMS with parallel ion-counter and faraday detection.
Journal of Analytical Atomic Spectrometry 20, 1350–1357.
Platzner, I.T., Habfast, K., Walder, A.J., Goetz, A., 1997. Modern Isotope Ratio Mass
Spectrometry. John Wiley and Sons, U.K. 514pp.
Simonetti, A., Heaman, L., Hartlaub, R.P., Creaser, R.A., MacHattie, T.G., Böhm, C., 2005.
U–Pb zircon dating by laser ablation MC-ICP-MS using a new multiple ion counting
Faraday collector array. Journal of Analytical Atomic Spectrometry 20, 677–686.
Sylvester, P.J., 2008. Matrix effects in laser ablation ICPMS. Mineralogical Association of
Canada short course series 40, 67–78.
Wilson, S.A., Ridley, W.I., Koenig, A.E., 2002. Development of sulfide calibration
standards for the laser ablation inductively coupled plasma mass spectrometry
technique. Journal of Analytical Atomic Spectrometry 17, 4006–4009.
Wohlgemuth-Ueberwasser, C.C., Ballhaus, C., Berndt, J., Stotter nee Paliulionyte, V.,
Meisel, T., 2007. Synthesis of PGE sulfide standards for laser ablation inductively
coupled plasma mass spectrometry (LA-ICP-MS). Contributions to Mineralogy and
Petrology 154, 607–617.
Woodhead, J.D., 2002. A simple method for obtaining highly accurate Pb-isotope data by
MC-ICPMS. Journal of Analytical Atomic Spectrometry 17, 1381–1385.
Woodhead, J., Hergt, J., Shelley, M., Eggins, S., Kemp, R., 2004. Zircon Hf-isotope analysis
with an excimer laser, depth profiling, ablation of complex geometries, and
concomitant age estimation. Chemical Geology 209, 121–135.
Woodhead, J.D., Swearer, S., Hergt, J., Maas, R., 2005. In situ Sr-isotope analysis of
carbonates by LA-MC-ICP-MS: interference corrections, high spatial resolution and
an example from otolith studies. Journal of Analytical Atomic Spectrometry 20,
22–27.
Woodhead, J., Hellstrom, J., Hergt, J., Greig, A., Maas, R., 2007. Isotopic and elemental
imaging of geological materials by laser ablation Inductively Coupled Plasma mass
spectrometry. Journal of Geostandards and Geoanalytical Research 31, 331–343.