Jessica K. Camacho-Zavala
◽
Ana L. Perez-Medina
◽
Jorge A. Mercado-Gutierrez
◽
Mario I. Gutierrez
◽
Josefina Gutierrez-Martinez
◽
...
BACKGROUND: Complex personalized Functional Electrical Stimulation (FES) protocols for calibrating parameters and electrode positioning have been proposed, most being time-consuming or technically cumbersome for clinical settings. Therefore, there is a need for new personalized FES protocols that generate comfortable, functional hand movements, while being feasible for clinical translation. OBJECTIVE: To develop a personalized FES protocol, comprising electrode placement and parameter selection, to generate hand opening (HO), power grasp (PW) and precision grip (PG) movements, and compare in a pilot feasibility study its performance to a non-personalized protocol based on standard FES guidelines. METHODS: Two FES protocols, one personalized (P1) and one non-personalized (P2), were used to produce hand movements in twenty-three healthy participants. FES-induced movements were assessed with a new scoring scale which comprises items for selectivity, functionality, and comfort. RESULTS: Higher FES-HSS scores were obtained with P1 for all movements: HO (p= 0.00013), PW (p= 0.00007), PG (p= 0.00460). Electrode placement time was significantly shorter for P2 (p= 0.00003). Comfort scores were similar for both protocols. CONCLUSIONS: The personalized protocol for electrode placement and parameter selection enabled functional FES-induced hand movements and presented advantages over a non-personalized protocol. This protocol warrants further investigation to confirm its suitability for developing upper-limb rehabilitation interventions with clinical translational potential.