Ventilator System SERVO-i V3.0: User's Manual
Ventilator System SERVO-i V3.0: User's Manual
Ventilator System SERVO-i V3.0: User's Manual
VENTILATOR SYSTEM
SERVO-i V3.0
Contents
1 • Introduction . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Device Description . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.1.1 Device Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.2 Device Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.3 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.4 Intended User . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.5 Intended Use Environment . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.6 Servicing Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.7 Disclaimers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
1.2.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.3 Fire Hazard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.4 Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.5 Auxiliary Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Version & Configurations . . . . . . . . . . . . . . . . . . . . . . . .4
1.3.1 Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2 • System Overview . . . . . . . . . . . . . . . . . . . . 7
2.1 Ventilator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.2 User Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.2.1 User Interface Components . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2.2 User Interface Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.3 User Interface Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3 Navigating the User Interface . . . . . . . . . . . . . . . . . . . . 10
2.3.1 Touch Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3.2 Main Rotary Dial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3.3 Fixed Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3.4 Direct Access Knobs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3.5 Menu Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.3.6 Status Touchpad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.4 Patient Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.4.1 Patient Unit Components . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.4.2 Patient Unit Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.4.3 Patient Unit Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.5 Transport and Storage . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.5.1 Before Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.5.2 During Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.5.3 Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
i
3 • Power Supply. . . . . . . . . . . . . . . . . . . . . . 19
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.1.1 Power Supply Specifications . . . . . . . . . . . . . . . . . . . . . . . . 19
3.1.2 Battery Data Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Viewing Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.3 Alarms and Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3.1 AC Power Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3.2 Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3.3 Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4 • Operation Overview . . . . . . . . . . . . . . . . . 23
4.1 Work Flow Summary . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Pre-Use Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.1 Performing a Pre-Use Check . . . . . . . . . . . . . . . . . . . . . . . . 24
4.3 Patient Circuit Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4.4 Start-up Configuration . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.4.1 Accessing the Start-Up Configuration . . . . . . . . . . . . . . . . . . 31
4.4.2 Editing the Start-up Configuration . . . . . . . . . . . . . . . . . . . . 31
4.5 Entering Patient Data. . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.6 Setting Ventilation Mode . . . . . . . . . . . . . . . . . . . . . . . 35
4.7 Setting Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.8 Starting Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.8.1 Starting Invasive Ventilation . . . . . . . . . . . . . . . . . . . . . . . 37
4.8.2 Starting Noninvasive Ventilation (NIV) . . . . . . . . . . . . . . . . . 37
4.9 Additional Settings Window . . . . . . . . . . . . . . . . . . . . . 38
4.10 Using Suction Support . . . . . . . . . . . . . . . . . . . . . . . . 39
4.10.1 Preparation Phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.10.2 Disconnect Phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.10.3 Post-Oxygen Phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.11 Recalibrating the Oxygen Cell . . . . . . . . . . . . . . . . . . . 40
4.12 Disconnecting the Patient . . . . . . . . . . . . . . . . . . . . . . 41
ii
5.2.3 Adjusting Scale/Sweep Speed . . . . . . . . . . . . . . . . . . . . . . 46
5.3 Showing the Event Log . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.4 Showing Loops . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.5 Showing Trends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.6 The Open Lung Tool. . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.6.1 Using the Open Lung Tool. . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.6.2 Adjusting Scales in the Open Lung Tool Display . . . . . . . . . . . 48
5.7 Saving Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.7.1 Recording Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.7.2 Using Recorded Waveforms. . . . . . . . . . . . . . . . . . . . . . . . . 50
5.7.3 Ventilation Record Card. . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6 • Ventilation Modes. . . . . . . . . . . . . . . . . . . 53
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.1.1 Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.1.2 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6.1.3 Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.2 Ventilator Operation . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.2.1 Setting Ventilation Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.2.2 Recalling a Previous Ventilation Mode. . . . . . . . . . . . . . . . . . 56
6.2.3 Special Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.2.4 Backup Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.3 Breathing Parameters . . . . . . . . . . . . . . . . . . . . . . . . . 57
7 • Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . 61
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
7.1.1 Alarm Output Connection Option . . . . . . . . . . . . . . . . . . . . . 61
7.1.2 Visual Alarm Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
7.2 Handling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.2.1 Viewing the Current Alarms Window. . . . . . . . . . . . . . . . . . . 62
7.2.2 Resetting Latched Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.2.3 Responding to Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.2.4 Presilencing Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
7.2.5 Permanently Silencing Alarms . . . . . . . . . . . . . . . . . . . . . . . 63
7.2.6 Turning Off the Apnea Alarm . . . . . . . . . . . . . . . . . . . . . . . 64
7.2.7 Responding to Technical Alarms . . . . . . . . . . . . . . . . . . . . . 64
7.3 Alarm Settings for Breathing Parameters . . . . . . . . . . . . 64
7.3.1 Viewing Alarm Limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
7.3.2 Setting Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
7.3.3 List of Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
7.3.4 Conditions Leading to Default Alarm Settings. . . . . . . . . . . . . 65
7.3.5 Alarm Ranges and Defaults . . . . . . . . . . . . . . . . . . . . . . . . . 66
iii
8 • Accessories . . . . . . . . . . . . . . . . . . . . . . . 67
8.1 Servo Ultra Nebulizer. . . . . . . . . . . . . . . . . . . . . . . . . . 67
8.1.1 Nebulizer Use Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 67
8.1.2 Nebulizer Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.1.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.2 CO2 Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.2.1 CO2 Analyzer Use Guidelines. . . . . . . . . . . . . . . . . . . . . . . . 69
8.2.2 CO2 Analyzer Components . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.2.3 CO2 Analyzer Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.3 Y Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
8.3.1 Y Sensor Use Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
8.3.2 Y Sensor Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
9 • System Messages . . . . . . . . . . . . . . . . . . . 73
9.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.2 High Priority Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.3 Medium Priority Alarms . . . . . . . . . . . . . . . . . . . . . . . . 78
9.4 Low Priority Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
9.5 Pre-Use Check Messages: General . . . . . . . . . . . . . . . . . 82
9.6 Pre-Use Check Messages: Specific . . . . . . . . . . . . . . . . . 83
9.7 CO2 Analyzer: Calibration Error Messages . . . . . . . . . . . . 85
9.8 Technical Error Messages . . . . . . . . . . . . . . . . . . . . . . . 86
10 • Specifications. . . . . . . . . . . . . . . . . . . . . 87
10.1 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.2 Operating Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.3 Nonoperating conditions. . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.4 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.2 Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
10.2.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
10.2.2 Gas Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
10.2.3 Patient System Connectors . . . . . . . . . . . . . . . . . . . . . . . . 88
10.2.4 User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
10.3 Standard Condition Specification . . . . . . . . . . . . . . . . . 88
10.4 Inspiratory Channel . . . . . . . . . . . . . . . . . . . . . . . . . . 88
10.5 Expiratory channel . . . . . . . . . . . . . . . . . . . . . . . . . . 89
10.6 Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
10.7 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
iv
10.7.1 Allowed Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 89
10.7.2 Autoset Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
10.7.3 Alarms Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
10.8 Ventilation modes . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
10.8.1 Controlled Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
10.8.2 Supported ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
10.8.3 Combined ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
10.9 Trend Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
10.10 Open Lung Tool Trend . . . . . . . . . . . . . . . . . . . . . . . 91
10.11 Log function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
10.12 Immediate functions . . . . . . . . . . . . . . . . . . . . . . . . 92
10.13 Communication/Interface . . . . . . . . . . . . . . . . . . . . . 92
10.14 Servo Ultra Nebulizer . . . . . . . . . . . . . . . . . . . . . . . . 92
10.15 CO2 Analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.16 Y Sensor Measuring . . . . . . . . . . . . . . . . . . . . . . . . . 94
10.17 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
10.18 Breathing Parameters: Default Values & Allowed Settings
(Standard Configuration) . . . . . . . . . . . . . . . . . . . . . . . . . . 95
10.19 Alarm Limits: Default Settings & Allowed Ranges . . . . . 97
11 • Definitions. . . . . . . . . . . . . . . . . . . . . . . 99
v
vi
1 • Introduction
1.1 Device Description • are a professional health care provider,
and
This section provides general information
about the Servo-i Ventilator System along • have received training in the use of this
with guidelines for appropriate use. system, and
• have experience with ventilation
1.1.1 Device Diagram treatment.
1.2.4 Gases
• The gases used in the system must be free
from particles.
• The gases used in the system must
conform to the following standards for
concentrations of water vapor and oil:
• Air: H2O < 7 g/m3; Oil < 0.5 mg/m3
• Oxygen: H2O < 20 mg/m3
• The system is not intended to be used
with any anesthetic agent.
Function or Accessory
Basic Extended
Alarm output connection option t t t t
Automode, pressure t t t %
Automode, PRVC t t t %
Automode, volume t t t %
Bi-Vent t t t t
CO2 Analyzer t t t t
NIV (Non Invasive Ventilation) t t t t
Nasal CPAP t - t t
Open Lung Tool t t t %
Pressure Control % t % %
Pressure Support % % % %
PRVC (Pressure Reg. Volume Control) t t t %
SIMV (PC) + Pressure Support % t % %
SIMV (PRVC) + Pressure Support t t % %
SIMV (VC) + Pressure Support t % % %
Suction Support % % % %
Upgrade to universal (all patient categories) t t
Volume Control t % % %
Volume Support t t t %
Y Sensor measuring t t t t
% : standard
t : optional
Infant configuration
Adult configuration
Universal configuration
ON/OFF switch
To adjust ventilator settings:
1. Activate the desired menu by touching
Trigger indication—appears in the one of the pads at the top of the screen.
message/alarm field when the 2. Activate the desired parameter by
patient triggers a breath pressing its touch pad.
The pad is now highlighted in white with a
NIV symbol—appears in the Mode blue frame and it is possible to enter a new
pad field during Non-Invasive value.
Ventilation.
3. Turn the Main Rotary Dial to the desired
value or line.
4. Confirm your setting by pressing the
parameter touch pad or by pressing the
Main Rotary dial.
The parameter touch pad turns blue again
indicating the new setting has been entered.
5. Touch Accept to activate your settings.
6. Press Cancel to start over.
SVX-6021_XX
SVX-5089_EN
Nebulizer Connector
SVX-9032
Check battery status There is a problem with the Open the Battery Status
(Informative Text Message) battery modules. One or more Window for information.
battery modules must be Replace and discard defective
replaced. battery module(s).
Battery operation (Medium AC power is off line due to a Check the connection to AC
Priority Alarm) power failure or power.
disconnection.
Limited battery capacity (High Less than 10 minutes left of Insert new battery modules or
Priority Alarm) battery operation. connect to AC power.
No battery capacity (High Less than 3 minutes left of Insert new battery modules or
Priority Alarm) battery operation. connect to AC power.
Low battery voltage (High Battery voltage too low. Insert new battery modules or
Priority Alarm) Cannot guarantee continued connect to AC power.
ventilator operation. This alarm may indicate worn
out or damaged batteries
(requiring replacement with
new ones) if it appears in
isolation; if this alarm appears
after a Limited battery
capacity or No battery
capacity alarm, then the
batteries only need to be
recharged.
SVX-155
7. Connect a complete breathing system 10. Unblock the Y Sensor and follow the on-
including (if available) a humidifier and screen instructions. The circuit
a Servo Ultra Nebulizer. compliance is automatically measured.
Important: When blocking the Y piece or Y Go step 11.
Sensor, make sure there is no leakage.
Leakage will affect the circuit compliance
compensation calculation.
If no Y Sensor is connected then:
8. Block the Y piece and follow the online
instructions. The circuit compliance is
automatically measured. Go to step 11.
If a Y Sensor is connected then:
9. Block the Y Sensor and follow the on-
screen instructions.
SVX-6091_XX
SVX-9042
WARNINGS!
SVX-9046
SVX-9028_EN
SVX-9058
5.1.1 Description
20
15 10
8.5
6.2
8.5
8.5 6.5
6.5
40 SVX.5092_EN
30
11
5.2.1 Description
The default waveform display has the
following characteristics:
• The value of a measured parameter vs.
time is displayed.
• The displayed parameter and the scale
are indicated on the y-axis.
• The pressure vs. time display is dark
yellow.
• The flow vs. time display is green.
• The volume vs. time display is light blue.
• The CO2 concentration vs. time display is
light yellow.
To show or hide the volume waveform or the
CO2 waveform display:
1. Press the fixed key Quick access.
2. Press the Waveform configuration
touchpad.
3. Press the touchpad corresponding to the
waveform you wish to show or hide.
Note: The pressure waveform and the flow
waveform are always displayed. The volume
and CO2 waveforms may be hidden. Thus,
2, 3, or 4 waveforms may be displayed.
Note: When you hide a waveform, the
remaining waveforms are expanded to use
all available screen space.
To set the sweep speed and amplitude for To view the Event Log:
displayed waveforms:
1. Press the fixed key Menu
1. Press the fixed key Quick access.
2. Press the Review touchpad
2. Press the Scales touchpad.
3. Press the Event log touchpad to view all
3. Press the touchpad corresponding to the
logged events.
waveform whose scale you wish to
change or select a sweep speed (6, 10 or 4. Use the arrows to scroll.
20 mm/s).
4. To adjust the scale of a waveform, turn
the Main Rotary Dial to the desired
value or use auto scale (press Auto).
Important: MAQUET does not recommend
using auto scale in Bi-Vent mode, when
patient breathing is spontaneous on both
levels.
To show trends:
1. Press the fixed key Trends.
2. Use the Main Rotary Dial to adjust
display parameters.
3. View trended measured values.
4. Use the up and down arrows to scroll.
5. To quit the Trends Window press Close.
6. To adjust the time resolution press the
Hours touchpad and turn the Main
Rotary Dial.
To activate the Loops function: 7. Activate the Cursor. Move it back and
1. Press the fixed key Quick access. forth on the time axis using the Main
2. Press the Loops touchpad Rotary Dial or touch screen.
8. Move the cursor to display the time,
3. Press [reference loop] to store a event type, and ventilation mode. For
reference loop. event stamps, an explanation appears.
4. Press [overlay loops] to see the two 9. View logged event stamps.
previous loops simultaneously. 10. If a recording is saved at a time
5. To close the window, press Close. corresponding to the cursor position, a
recording button is shown. To view the
recording, press the button.
Pressure Regulated Volume PRVC TV/MV; RR; PEEP; O2; IE/IT; IRT; TF/TP
Control
Volume Control VC TV/MV; RR; PEEP; O2; IE/IT; PT; IRT; TF/TP
Continuous Positive Airway CPAP PS+; PEEP; O2; IRT; TF/TP; ICO; PC+
Pressure
Synchronized Intermittent SIMV (PRVC) + TV/MV; SIMVR; O2; IE/IT; IRT; BCT; TF/TP; ICO;
Mandatory Ventilation PS PS+
(PRVC) + PS
SIMV (VC) + PS SIMV (VC) + PS TV/MV; SIMVR; O2; IE/IT; IRT; BCT; TF/TP; ICO;
PS+
SIMV (PC) + PS SIMV (PC) + PS TV/MV; SIMVR; O2; IE/IT; IRT; BCT; TF/TP; ICO;
PS+
Automode Automode
a. TV/MV = Tidal Volume or Minute Volume; RR = Respiratory Rate; PEEP = Positive End Expiratory Pressure;
IE/IT = I:E Ratio or Inspiration Time; IRT = Inspiratory Rise Time; TF/TP = Trigg. Flow or Trigg. Pressure;
PT = Pause Time; PC+ = Pressure Control Level Above PEEP; PS+ = Pressure Support Level Above PEEP;
ICO = Inspiratory Cycle Off; SIMVR = SIMV Rate; BCT = Breath Cycle Time; P_HIGH = Maximum Pressure for
Higher Pressure Level; PEEP_LOW = PEEP for the Lower Pressure Level; T_HIGH = Time at the Higher Pressure
Level; T_PEEP = Time at the Lower Pressure Level; P_HIGH+ = Pressure Support Level Above Maximum
Pressure
Pressure Control PC IMPORTANT to set alarm limits for MVol; cm; eb; df
Continuous Positive CPAP IMPORTANT to set alarm limits for MVol, Apnea; sb
Airway Pressure
Bi-Vent BV IMPORTANT to set alarm limits for MVol, Auto Scale not
recommended, Apnea alarm and back-up ventilation
not available, two pressure levels and durations are
set, alarms are handled for each Bi-Vent cycle
Non-Invasive NIV-PC for ALL NIV modes: avoid high inspiratory pressure, use
Ventilation - PC of nebulizer not recommended, excessive leakage will
cause a high-priority ALARM, trigger sensitivity and
cycle off cannot be set; NIV - PC is a controlled
breathing mode (cm)
a. MVol = Minute Volume, UPL = Upper Pressure Limit, TVol = Tidal Volume, cm = controlled ventilation mode, pib
= patient-initiated breathing mode, sb = spontaneous breathing mode, eb = patient can trigger extra breaths,
df = decelerating inspiratory flow, cf = constant inspiratory flow
7.2.1 Viewing the Current Alarms Note: The NIV alarm Leakage out of range is
Window. not latched.
SVX-5098_EN
SVX-5099_EN
High continuous pressure (HP) NA; (Set PEEP level+15 cmH2O for > 15 sec) No
6
1
2 3
SVX-136_XX
Notes:
• The etCO2 concentration low alarm can
be permanently silenced (Audio off) when
the message Silence alarm permanently?
is shown.
• During calibration no CO2 waveforms or
measured CO2 values will be displayed.
• During zero calibration the adapter must
contain room air only.
SVX-5078_XX
Note: A Pre-Use Check or a Patient Circuit The two versions of the disposable, single-
Test is required to prior to using the Y use Y Sensor —Adult (1) and Infant (2)—are
Sensor. shown in the figure above. The Infant sensor
includes an adaptor for use with the neonate
CO2 adapter.
8.3.1 Y Sensor Use Guidelines
These guidelines are reminders for health
care professionals who have already been
trained to use the Y Sensor.
• The Y Sensor is intended for single-patient
use only.
• Do not insert two Y Sensor modules at the
same time. The Servo-i Ventilator System
can handle only one Y Sensor module at a
time.
• Do not apply tension to the Y Sensor
tubing.
• Make sure there are no kinks in the Y
Sensor tubing.
• If the Y Sensor is not connected to the
module, do not connect to the patient
circuit as this may cause leakage.
• Frequently check for condensed water or
other fluids in the Y Sensor. Fluids in the Y
Excessive leakage.
Expiratory cassette The expiratory cassette is disconnected Connect the expiratory cassette.
disconnected or not connected properly. Replace the expiratory cassette.
Perform a Pre-Use Check if a new
expiratory cassette is inserted.
Expiratory Minute Preset or default alarm limit exceeded. Check patient and breathing
Volume: High Increased patient activity. system.
1. at RR < 60 b/min
2. at RR < 100 b/min
3. at constant leakage fraction <30%
4. at RR < 100 b/min 9. The apnea alarm can be turned off in Nasal
5. at constant leakage fraction <30% CPAP.
6. at constant leakage fraction <30% 10. Note: Setting the alarm to 0 (zero) is equal to
7. at Expiration time < 4 s and RR <100 b/min alarm off.
8. at Expiration time < 1 s and RR <100 b/min 11. In NIV low limit can be set to 0 (zero).
Servo… User´s manual
English edition
Order No: 65 14 892 E313E Infant Adult Universal Options 89
10.7.2 Autoset Alarm Limits 10.8 Ventilation modes
High Airway Pressure:
Mean peak pressure +10 cmH2O or at least 35 cmH2O. 10.8.1 Controlled Ventilation
Expiratory Minute Volume (upper) Pressure Control (PC)
+ 50% Pressure controlled ventilation.
Expiratory Minute Volume (lower) Volume Control (VC)
- 50% Volume controlled ventilation.
Respiratory Frequency (upper) Pressure Regulated Volume Control (PRVC)
+ 40% Pressure regulated volume controlled ventilation.
Respiratory Frequency (lower) NIV Pressure Control
- 40% Noninvasive pressure controlled ventilation.
Mean End Expiratory Pressure (high)
+5 cmH2O 10.8.2 Supported ventilation
Mean End Expiratory Pressure (low) Volume Support (VS)
-3 cmH2O. Volume supported ventilation.
etCO2 Concentration (upper) Pressure Support (PS)/CPAP
+ 25% Pressure supported ventilation / Continuous positive
airway pressure ventilation.
etCO2 Concentration (lower)
- 25% NIV Pressure Support
Noninvasive pressure supported ventilation.
10.7.3 Alarms Miscellaneous Nasal CPAP
Nasal continuous positive airway pressure ventilation.
Audio Pause (Alarm silence/reset)
Two-minute silence and reset of latched alarms.
10.8.3 Combined ventilation
SIMV (VC) + PS
Synchronized intermittent mandatory ventilation based
on volume controlled ventilation with pressure support.
SIMV (PC) + PS
Synchronized intermittent mandatory ventilation based
on pressure controlled ventilation with pressure
support.
SIMV (PRVC) + PS
Synchronized intermittent mandatory ventilation based
on pressure regulated volume controlled ventilation
with pressure support.
Bi-Vent
Pressure controlled ventilation that allows the patient
the opportunity of unrestricted spontaneous breathing.
Automode
Control mode Support mode
VC <----------- ----------> VS
PC <----------- ----------> PS
PRVC <----------- ----------> VS
In Servo-i flow measurements all preset and indicated
volumes are referenced to ambient pressure at +21˚C
(AP21).
Elastance E
Inspiratory Resistance Ri
Expiratory Resistance Re
P0.1 P0.1
Drawer Kit
Weight (kg)
4.5
Dimensions (mm)
300 W x 210 L x 240 H
Automode ON/OFF OFF OFF OFF OFF ON/OFF ON/OFF ON/OFF ON/OFF
Backup Ti (s) 0.5 1.0 0.5 1.0 0.3-1 0.5-2 0.3-1 0.5-2
Compensate for OFF OFF OFF OFF ON/OFF ON/OFF ON/OFF ON/OFF
compliance
Flow trig sensitivity level 50% 50% 50% 50% 0-100% 0-100% 0-100% 0-100%
(fraction of bias flow)
I:E ratio 1:2 1:2 1:2 1:2 1:10-4:1 1:10-4:1 1:10-4:1 1:10-4:1
Inspiratory cycle-off (% 30 30 30 30 1 - 70 1 - 70 1 - 70 1 - 70
of peak flow)
Inspiratory rise time (s) 0.15 0.15 0.15 0.15 0 - 0.2 0 - 0.4 0 - 0.2 0 - 0.4
Inspiratory rise time (s) 0.15 0.2 0.15 0.2 0 - 0.2 0 - 0.4 0 - 0.2 0 - 0.4
in NIV
Mode (Invasive PC VC PC VC - - - -
ventilation)
PEEP (cmH2O) 5 5 5 5 0 - 50 0 - 50 0 - 50 0 - 50
Tpause (%) - 10 10 10 - 0 - 30 0 - 30 0 - 30
T—Time
Tc—Time constant
Ti—Inspiration time
Ti/Ttot—Duty cycle or ratio of inspiration
time to total breathing cycle time (only
during spontaneous breathing)
Tidal Volume—Volume per breath or target
volume (ml)
Thigh—Time at Phigh level in Bi-Vent (s)
TPEEP—Time at PEEP level in Bi-Vent (s)
—Flow
CO2—CO2 Minute elimination
ee—End expiratory flow
leak—Leakage flow (l/min)
Ventilation record card—Documentation
card
VTCO2—CO2 tidal elimination
Va—Alveolar ventilation
VC—Volume Control
VDaw—Airway dead space
VS—Volume Support
VTA—Alveolar Tidal Volume
VTe —Expiratory Tidal Volume
VTi—Inspiratory Tidal Volume
SVX-580_EN
SVX-583_EN Expiratory hold is activated by manually pressing the fixed key.The maximum time
is 30 seconds.The inspiratory and expiratory valves close after expiration.
Exp.
This function can provide an exact measurement of the end expiratory lung pressure.
hold
It can be used for static compliance measurement and to determine the total PEEP.
Note. Expiratory hold is not available in Nasal CPAP.
Inspiratory hold is activated by manually pressing the fixed key.The maximum time
is 30 seconds.The inspiratory and expiratory valves close after inspiration.
Insp.
This function can provide an exact measurement of the end inspiratory lung pressure.It
hold
can be used during x-ray or to determine Plateau pressure, or static compliance
calculation.
Note. Inspiratory hold is not available in Nasal CPAP.
Loops
Note. A reference loop can be saved by
the user.
Scales
Open Lung
Tool
Open Lung
Tool scales Note. Open Lung Tool is not available in NIV
Main
screen Main
screen
The Main screen fixed key will return you to the Main screen,
windows.
SVX-582_EN
Trends
Recorded
waveforms Review trends, recorded waveforms, event log or configuration.
Event log
View
configuration
SVX-632_EN
Technical alarms
Alarm
sound level
NIV adult
alarm
NIV infant
alarm
Start-up
configuration
SVX-579_EN
O2 cell/sensor
Batteries
Status of O2cell/sensor
CO2 Module Status ofY Sensor module
(if available)
Status of batteries
Y Sensor
Measuring
NIV Backup This pad is only visible during NIV Pressure Support mode.