Volt/mA Calibrator
Instruction Sheet
Introduction
The Fluke 715 Volt/mA Calibrator is a source and measurement tool for 0 to 24 mA current loop testing and dc voltage from 0 to 20/25 V. The calibrator does not source and measure simultaneously. Your calibrator is supplied with a holster, a set of test leads, an installed 9 V alkaline battery, and this instruction sheet.
Summary of Calibrator Capabilities Function dc mV input 0 to 200 mV dc mV output dc V input dc V output dc mA input 0 to 24 mA dc mA output Loop power output 24 V dc output N/A 0.001 mA 0 to 25 V 0.001 V 0 to 20 V 0.01 mV Range Resolution
PN 650314 July 1997 Rev. 3, 8/05 1997-2005 Fluke Corporation. All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.
If the calibrator is damaged or something is missing, contact the place of purchase immediately. Contact your Fluke distributor for information about accessories. To order replacement parts or spares, see Replacement Parts. To contact Fluke, call one of the following telephone numbers: USA and Canada: 1-888-99-FLUKE (1-888-993-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-*-276-6196 Anywhere in the world: +1-425-356-5500 Address correspondence to: Fluke Corporation P.O. Box 9090, Everett, WA 98206-9090 U.S.A. Fluke Europe B.V. P.O. Box 1186, 5602 BD Eindhoven The Netherlands
J I M W T
$
P
Safety Information
Use the calibrator only as specified in this sheet, otherwise the protection provided by the calibrator may be impaired. A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the calibrator or the equipment under test.
W Warning
To avoid possible electric shock or personal injury: Never apply more than 30 V between any two jacks (terminals), or between any jack and earth ground. Make sure the battery door is closed and latched before you operate the calibrator. Remove test leads from the calibrator before you open the battery door. Do not operate calibrator if it is damaged. Do not operate the calibrator around explosive gas, vapor, or dust. When servicing the calibrator, use only specified replacement parts.
Caution
Use the proper jacks, function, and range for your measurement or output application.
1.
With the Calibrator OFF, press the O key. P.S.xx is displayed, where xx is the turn-off time in minutes. OFF means the power saver is disabled. Press ] and/or } to increase or decrease the turnoff time in minutes. To disable, press } until the display shows OFF.
2. 3.
Now using the ] } controls set the output to the desired output value for 100% Again, press the ] } scroll keys simultaneously to store the 100% value.
Measuring dc Volts
INPUT mV or V depending on range
V mV
mA %
V mV
it01i.eps
Sourcing dc Volts
OUTPUT mV or V depending on range
V mV
mA %
V mV
it02i.eps
Measuring dc mA
INPUT
mA %
V mV mA %
it03i.eps
mA %
V mV mA %
+24V output
Two-wire transmitter
HARTTM Communicator
it06i.eps
Sourcing mA
Use source mode whenever you need to supply current into a passive circuit such as a current loop with no loop supply. Insert the test leads into the OUTPUT + and mA jacks as shown below. Note A path must exist for current to flow between the OUTPUT + and mA jacks, otherwise the display will indicate an overload (OL) when you set an output value.
OUTPUT
715
VOLT/mA CALIBRATOR
mA %
V mV
mA %
+ LOOP
30V MAX
mA
30V MAX
HARTTM Communicator
it04i.eps
Simulating a Transmitter
Use simulate mode when an external 24 to 30 V loop power supply is available. Insert the test leads into the mA SIMULATE and + jacks as shown below.
OUTPUT
mA %
V mV
mA %
COM
30V dc max.
Power supply
it05i.eps
Maintenance
For maintenance procedures not described in this sheet, contact a Fluke Service Center.
In Case of Difficulty
Check the battery and test leads. Replace as necessary. Review this sheet to make sure you are using the correct jacks and pushbuttons.
If the calibrator needs repair, contact a Fluke Service Center. If the calibrator is under warranty, see the warranty statement below for terms. If the warranty has lapsed, the calibrator will be repaired and returned for a fixed fee. Contact a Fluke Service Center for information and price.
Cleaning
Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.
Calibration
Calibrate your calibrator once a year to ensure that it performs according to its specifications. A calibration manual is available (PN 686540). Call 1-800-526-4731 from the USA and Canada. In other countries, contact a Fluke Service Center.
W Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator (B) appears.
it07i.eps
BT1 H80M MP85 MP86 H2, 3, 4 MP89, 90 MP8 MP92 H5, 6 S1 TL75 AC72 TL20 -
9V battery, ANSI/NEDA 1604A or IEC 6LR61 Holster, Yellow, hanging Case top Case bottom Case screw Non-skid foot O-ring for input/output receptacle Battery door Battery door fasteners Keypad Test lead set Instruction sheet Alligator clips Industrial test lead set 71X Calibration Manual
1 1 1 1 3 2 1 1 2 1 1 1 1 Option Option
MP85
S1
V V m A m %
Holster
it10c.eps
Specifications
Specifications are based on a one year calibration cycle and apply from +18C to +28C unless stated otherwise. Counts means number of increments or decrements of the least significant digit.
200 mV 20 V output
0.01 mV 0.001 V
25 V input Input impedance: 1 M (nominal), < 100 pF Overvoltage protection: Fuseless Voltage drive capability: 1 mA
DC mA Input
Range Resolution Accuracy, (% of Reading + Counts) 0.01% + 2
24 mA
0.001 mA
DC mA Output
Range: 0 mA to 24 mA Percent display: 0% = 4 mA, 100% = 20 mA Accuracy: (0.01% of reading + 2 counts) Source mode: Compliance: 1000 at 20 mA for battery voltage 6.8 V (700 at 20 mA for battery voltage 5.8 to 6.8 V) Simulate mode: External loop voltage requirement: 24 V nominal, 30 V maximum, 12 V minimum
Loop Power
24 V 10%
General Specifications
Maximum voltage applied between any jack and earth ground or between any two jacks: 30 V Storage temperature: -40C to 60C Operating temperature: -10C to 55C Operating altitude: 3000 meters maximum Temperature coefficient: 0.005% of range per C for the temperature ranges -10 to 18C and 28 to 55C Relative humidity: 95% up to 30C, 75% up to 40C, 45% up to 50C, and 35% up to 55C Vibration: Random 2 g, 5 to 500 Hz Shock: 1 meter drop test Safety: Certified as compliant to CAN/CSA C22.2 No. 1010.1:1992. Complies with ANSI/ISA S82.01-1994. Power requirements: Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61) Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x 7.35 in L); With holster: 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L) Weight: 349 g (12.3 oz); With holster: 601 g (21.2 oz)
LIMITED WARRANTY & LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Flukes behalf. To obtain service during the warranty period, send your defective calibrator to the nearest Fluke Authorized Service Center with a description of the problem. THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.