James - Spectrograph Design Fundamentals
James - Spectrograph Design Fundamentals
3
P
1
resonance line) are harmful and must be shielded, together with any
surfaces which may reect that radiation into the surrounding laboratory. Such sources
may also stimulate the production of ozone (O
3
) which has a characteristic electric smell
and has well-known corrosive properties on some materials such as rubber.
r
0.20.12 m, 20001200
A, 610 eV. The vacuum ultra-violet, where atmospheric gases
absorb the radiation (O
2
below 2424
A and N
2
below 1270
A), where most materials
have lost their transparency and where the reecting powers of metals are beginning to
decline. Optical instruments for this region must be in a vacuum chamber or in a non-
absorbing atmosphere such as helium, which is transparent down to 911
A, and other rare
gases similarly down to their respective ionisation potentials. It is an interesting and not
yet thoroughly explored region where autoionisation series spectra can be detected with
their concomitant superallowed transitions and broad line absorption spectra. Filling
the spectrograph with helium does not ease the vacuum problem so far as leak prevention
is concerned but it may ease the stress on thin-lm windows.
r
0.120.01 m, 1200100
A, 10100 eV. The far- or extreme-vacuum ultra-violet, some-
times called the XUV, where radiation is ionising, where there are no transparent ma-
terials except in the form of thin (12 m) lms, where reection coefcients are down
to 0.10.2 and apart from synchrotron radiation there are no reliable, steady sources of
radiation for absorption spectroscopy.
1
Ultra-Violet-Optical-Infra-Red.
2
Most optical glasses are opaque below 3500
A.
8 The relevant regions of the electromagnetic spectrum
Gas molecules have absorption cross-sections measured in megabarns (1 Mb =
10
18
cm
2
) and this, in an optical path of 1 m, implies a need for a vacuum better
than 10
4
mm Hg, which in turn requires a diffusion pump or, if especially pure samples
are to be examined, a turbo-molecular pump. This is probably the most difcult region of
the spectrum for which to design spectrographs and in which to do research. The vacuum
requirements are demanding and not to be undertaken lightly.
At wavelengths below 100
A materials start to become transparent again
3
and this
is the beginning of the soft X-ray region where different techniques supervene.
In each of these regions of the spectrum the optical design requirements have
much in common. There are exceptions.
r
In the infra-red, alternative techniques such as Fourier-transform spectroscopy, usually
with a Michelson interferometer,
4
are possible, are advantageous and are capable of enor-
mous resolving power. Awarning, however: the method lacks spectrometric simultaneity.
A ickering source can have catastrophic effects on the spectrum unless rapid sawtooth
scanning is used at a frequency well above the highest icker frequency or below the
lowest. In commercially available Fourier spectrometers this rapid scanning is unusual.
Fourier spectroscopy has no advantage elsewhere than in the infra-red.
r
In the UVOIR region the plane diffraction grating is the radiation analyser of choice
although there are circumstances where a prism has advantages. The FabryPerot inter-
ference spectrograph, too, has its place where high resolution is needed or where very
low light levels are encountered.
r
In the vacuum ultra-violet between 2000
A and 1200
A prism and lens materials such
as calcite and lithium uoride are available with the required transmission but generally
the concave grating possesses advantages which increase rapidly as the wavelength of
interest decreases. At wavelengths below 2000
A the concave grating is the analyser of
choice. Below 1200
A there is no realistic alternative.
So far as refracting materials are concerned for these various regions, technology
is advancing all the time, especially in the infra-red and the Internet should be
consulted for details of their transmitting and refracting properties and for the
names of their manufacturers.
2.1 The limits of optical spectrography
An optical spectrograph or spectrometer of one sort or another can be made to give
a resolution up to R 10
7
without a struggle. It is difcult to imagine a practical
application for a conventional spectrograph with a greater resolving power than
3
Or the radiation more penetrating!
4
But see Section 7.6.
2.1 The limits of optical spectrography 9
this.
5
A FabryPerot etalon with a 3-metre spacer and a nesse of 50 will give
4.10
8
at optical wavelengths, enoughtoreveal the natural widths of allowedatomic
transitions, providedof course that youcanrst eliminate all the temperature effects.
Extraordinary precautions are needed to ensure stability (in the source as well as the
instrument) since the etendue is tiny and the exposure time long. It has the merit of
being absolute, since the various quantities involved are directly measurable with
a metre-stick. But at this sort of resolving power, more sophisticated spectroscopic
techniques are available and preferable.
5
The measurement of small Doppler shifts, for example, does not require high resolving power, only high spatial
resolution at the focal plane. A wavelength shift of a small fraction of a resolved element is measurable.
3
Geometrical optics
This is the branch of optics which deals with image-forming instruments, includ-
ing of course spectrographs and interferometers. Such instruments employ lenses,
mirrors and prisms and it is the art of combining these elements to make useful
devices which is the subject of the next two chapters.
3.1 Rays and wavefronts
Image-forming instruments are intended to project images of real objects on a
screen or focal surface usually though not always plane, and instrumental optics
is the study of ways of doing this.
The subject comprises two chief parts: instrument design and lens design.
The former is the design of instruments so that light is conveyed through their
optical components from one to another to arrive eventually at the focal surface.
The technique is essentially a graphical one, using a drawing board or a computer
drafting program to lay out the components at their proper places.
The latter involves the accurate tracing of rays through various optical sur-
faces of different types of glass, and reections from mirrors of various shapes
to achieve a correction of the optical aberrations and to ensure that light of all
wavelengths from a point on an object is focused to a corresponding point on its
image.
Traditionally this was done by ray tracing, the accurate computation of ray paths
using seven-gure logarithms and trigonometrical tables, but now is done chiey
by iterative ray tracing in a small computer, using a program specically designed
for that purpose.
A word of warning: an optical system such as a spectrograph cannot be designed
on a ray-tracing program. The proper positions for the various elements must be
found by means of the drawing board or the CADprogram. An iterative ray-tracing
programcan then be used to optimise the performance of the instrument by altering
10
3.2 Instrumental optics 11
the positions of the optical elements, the curvatures and tilts of refracting and
reecting surfaces and their excentricities.
3.2 Instrumental optics
High-school teachingis that light is emittedfroma luminous point, travels instraight
lines outwards called rays, and that these rays can be caught by a lens and bent
to form bundles which then converge to a point. If a screen is placed at the proper
place, a point can be seen, the image of the original luminous object.
Light rays are a convenient ction of course like the lines of force in an electric
or magnetic eld, or the streamlines of uid mechanics. Following Huygens
principle, we postulate that light is emitted as wavefronts (which are also known as
surfaces of constant phase) which spread outwards spherically from each point
of an existing wavefront so that their envelope forms a new wavefront. They are
slowed down when they pass through a refracting transparent mediumsuch as glass
or water. A spherically expanding wavefront from a point source can be converted
by a properly shaped agglomeration of lenses to a spherical converging wavefront,
which converges to form a point image of the point object. The point source is
another convenient ction.
Other necessary concepts are:
r
Refractive index This is the ratio of the speed of light in vacuo divided by the speed
of light in a refracting medium.
1
Refractive index is generally measured indirectly by
observing the deviation of a ray as it passes through a prism for example.
r
Geometrical path This is the actual physical distance between a source of light and
the place where the light falls on a screen and is observed. This is the distance which you
measure with a ruler on a ray diagram on a drawing board.
r
Optical path This is the geometrical path multiplied by the local refractive index.
Usually it is the sum of several sections where the light passes through different media,
and if the refractive index, n, varies continuously it is the integral
_
b
a
n(s) ds
along the path of the ray.
Alternatively and this is useful it is the number of wavelengths of monochromatic
light along the path, multiplied by the vacuum wavelength of the light.
The basic principle of image formation in an optical instrument is this: that the paths
of the rays from a point on an object to its conjugate point on the image should all have
the same optical path length.
1
I use the word speed here, since the speed, unlike the velocity, can be measured directly by ordinary physical
experiments.
12 Geometrical optics
r
u, 0 0, 0 R, 0 v, 0
h
i
Figure 3.1 The basic diagram for ray tracing. It shows the various quantities in-
volved in tracing a ray through a surface, using Snells law to relate the angles i
and r.
r
Reducedpath This is the geometrical path divided by the refractive index. For example,
if you look into a swimming pool it appears to be shallower than it is. What you see or
would measure with a camera rangender for example is the reduced path to the bottom
of the pool.
All these are useful ideas in the eld of optical design.
3.3 Centred systems
Optical image-forming instruments such as telescopes and microscopes generally
have optical trains of spherical surfaces with the centres all lying on a straight
line, the optic axis of the system. The use of spherical surfaces derives from the
physical fact that these are the only surfaces which can be ground and polished
with high accuracy with comparatively simple apparatus. One piece of glass ground
against another with a suitable abrasive automatically produces a common spherical
interface. To achieve other shapes, such as paraboloids, more elaborate methods
are required but with contemporary technology are achieved at small extra cost.
3.4 Gaussian optics
The elementary theory of image formation is due to Carl Friedrich Gauss (1777
1855) and makes the assumption that all the angles between rays and the optic axis
are small, so that the approximations sin and cos 1 are good.
By convention, optical diagrams are drawn with light entering the system from
the left. Suppose (Fig. 3.1) that there are two transparent media of refractive indices
n
1
and n
2
, and a spherical interface between them of radius R with its vertex at the
origin of a Cartesian coordinate system and the centre of the sphere at the point
C(R. 0). Suppose that a ray of light starts from a point on the optic axis A(u. 0)
and reaches the interface at D where it is refracted according to Snells law and
continues in a new direction to meet the optic axis again at B(:. 0). Draw the rays
from C to D to B. Angles are measured clockwise from the optic axis to the ray.
3.4 Gaussian optics 13
Then,
h,u = . h,: = . h,R = .
and by ordinary geometry,
r = . i = +.
and with Snells law,
n
1
_
1
R
h
u
_
= n
2
_
1
R
h
:
_
.
giving
n
1
u
n
2
:
=
n
1
n
2
R
.
or, more usefully,
n
1
_
1
R
1
u
_
= n
2
_
1
R
1
:
_
.
and this is a fundamental formula connecting the object distance u, the image
distance :, the radius of curvature R of the surface and the two refractive indices
n
1
and n
2
.
3.4.1 The thin lens
We next apply this result to two surfaces in succession, in other words, to a simple
lens. The image distance after refraction by the rst surface becomes the object
distance for refraction at the second. The distance between the two vertices is taken
as negligible. Then, with sufxes for the two radii and ticks for the distances from
the second surface,
n
1
_
1
R
1
1
u
_
= n
2
_
1
R
1
1
:
_
.
n
2
_
1
R
2
1
u
_
= n
1
_
1
R
2
1
:
_
.
and u
, the object distance for the second surface, equals :, the image distance for
the rst. Then, as before,
n
1
_
1
R
1
1
u
_
= n
2
_
1
R
1
1
:
_
.
14 Geometrical optics
and for the second surface,
n
2
_
1
R
2
1
:
_
= n
1
_
1
R
2
1
:
_
.
Eliminating : between these and adding gives
n
1
R
1
n
1
u
n
2
R
1
+
n
2
R
2
n
1
R
2
+
n
1
:
= 0.
and putting, as usual for lenses in air, n
2
= n and n
1
= 1, the result is
1
:
1
u
= (n 1)
_
1
R
1
1
R
2
_
.
An obvious and probably familiar corollary follows if we put the object distance
equal to , for then
(n 1)
_
1
R
1
1
R
2
_
=
1
f
.
where f is the focal length of the lens. The plane perpendicular to the optic axis
where the image is formed is called the focal plane of the lens. There are obviously
two focal planes, one either side of the lens itself, at distances f and f from the
vertex. Where they cross the optic axis are the two foci or focal points of the lens.
The other familiar equation follows:
1
:
1
u
=
1
f
.
This is the Gaussian lens equation.
NB Distances to the left (object side) of the lens are negative and those to
the right (image side) are positive. This is the so-called sign convention; one of
many which plague geometrical optics, but the one which appears to work best and
which is used throughout here. It is known generally as the Cartesian convention.
Another part of the convention is that surface curvatures are positive if the surface
is convex to the incoming light, that is, if the centre of the sphere lies to the right
of the vertex of the surface.
Notice that in general the optic axis can be dened as the line joining the two
centres of the spheres which formthe lens surfaces. In practice one tries to construct
the instrument so that all the sphere centres lie on one line. This calls for some
subtlety in manufacture and the accuracy with which a given lens element has its
optic axis at the mechanical centre of its (circular) rim is called the centration.
Design note It is much more convenient when doing calculations to use cur-
vatures and powers rather than radii and focal lengths. The Gaussian equation is
3.4 Gaussian optics 15
d
f
Figure 3.2 The geometry of a thick lens. The position of a virtual equivalent thin
lens is one of the two principal planes which denes the position of refraction of a
ray by any lens or system of lenses. The distance from this plane to the focal point
is the focal length of the lens.
then
P = (n 1)(c
1
c
2
) = (n 1)Lc.
where P = 1,f and is measured by opticians in reciprocal metres, or dioptres.
3.4.2 Thick lenses
We assumed above that the vertices of the two surfaces were coincident. In practice
there is a distance d between them, where d is small compared with the object and
image distances.
Repeating the argument, but with u
1
R
2
+
(n 1)
2
n
d
R
1
R
2
_
.
This is sometimes called the lensmakers equation.
Confusion often arises when concave and convex mirrors are included. Acertain
way round the difculty is to treat every optical component, lens or mirror, as if it
were a lens, and to lay out the system on the drawing board as if it were a totally
refracting one. Then a mirror concave to the left behaves like a lens with a positive
focal length f and f = R,2 where R is the radius of the mirror. To be pedantic,
the mirror can be treated as a lens of refractive index 1 and a at rst surface.
An important tip is to remember that the angle through which a ray is deviated
on passage through a lens depends only on the height of incidence and the focal
length.
3.4.3 Aplanatic surfaces and lenses
The aplanatic points U and U
= h/f
Figure 3.3 The refraction angle of a ray entering a lens depends in the Gaussian
approximation only on the height of incidence and the focal length.
R
R(1 + n)
R(1 + n)
n
Figure 3.4 The converging aplanatic lens. The object is always virtual. Both object
point and image point are at nite distances from the rst surface vertex and are
related to each other as in the diagram by the refractive index. The second surface
position is not critical but it must be concentric with the paraxial image point.
Perfect imaging holds for any incident ray directed initially towards the object
point. The imaging is virtually aplanatic for a nite eld surrounding the paraxial
image point.
distances from the vertex of the surface are easily shown
2
to be related by
u = R(1 +n) : = R(1 +n),n. (3.1)
An image so formed in the vicinity of U
1
d the hs and the s can be eliminated to give the focal length
of the whole compound lens as
1
F
=
1
f
1
+
1
f
2
d
f
1
f
2
.
or, more usefully,
F =
_
f
1
f
2
f
1
+ f
2
d
_
.
For practical purposes we need to know more than this. The usual question is
where should one put a single thin lens, equivalent to the combination, which will
produce an image at the same position and with the same focal length? In other
words, where should a single lens of focal length F be placed to secure an image
at V?
18 Geometrical optics
A
f
2
P
2
P
1
f
1
f f
A
Figure 3.6 The principal planes andfocal planes of a systemof lenses. Onlythe rst
and last surfaces are shown and there may be any number of surfaces in between.
Any such system of lenses possesses such a pair of principal planes which serve
to determine the paths of rays traced through the system in either direction. Actual
ray paths are drawn in full, construction paths are shown interrupted.
3.4.5 Principal planes
Suppose that the proper position is at a distance x
1
to the right of the rst element.
Then (Fig. 3.5)
d x
1
=
h
1
h
2
1
+
2
=
1
d
1
+
2
=
Fd
f
1
.
so that
x
1
= d(1 F,f
1
) =
d( f
1
d)
f
1
+ f
2
d
.
and the simple equivalent lens must be placed at a distance x to the right of the rst
element to obtain the same-sized image of an object at at the same place.
If the light were coming from the right:
r
the focal length of the compound lens would be the same;
r
the position of the simple equivalent lens would be
x
2
=
d( f
2
d)
f
1
+ f
2
d
to the left of the second element.
These two positions are calledthe principal planes of the compoundlens. Where
they cross the optic axis are the principal points of the lens.
It can be shown that any centred system of lenses and mirrors, however compli-
cated, has two principal planes and two focal planes. They can be dened as follows
(Fig. 3.6):
A ray, parallel to the optic axis, entering the lens system at P
2
emerges at P
1
at
the same height above the axis and passes through the focus F
1
.
Similarly a ray fromthe right entering the systemat P
1
emerges at P
2
at the same
height and passes through F
2
.
3.4 Gaussian optics 19
f
x
1
(a)
f
x
1 (b)
Figure 3.7 Two examples of lenses with principal planes outside the lens itself.
(a) Represents a telephoto lens and (b) is a combination of positive lenses with a
negative focal length, albeit giving a real erect image of a point at . The single
equivalent lens would be at a distance x
1
to the right of the rst vertex, and would
be a negative lens giving a virtual image a distance f to the left of x
1
.
3.4.6 Nodal planes
Two further points and planes can be dened, although for most purposes they are
coincident with the principal planes. They are the nodal planes. There are several
denitions, but a simple and practical one is as follows:
A ray from the left, crossing the optic axis at the nodal point n
2
at an angle
to the optic axis, emerges from the other nodal point n
1
at the same angle. The
nodal points coincide with the principal points unless the initial and nal refractive
indices are different. This circumstance might happen if a lens were designed to
work with one end in water or oil for example.
For a single-element thick lens the principal planes are inside the glass at a dis-
tance x
i
= f [(n 1),n](d,R
i
) from the surface i (where i = 1 or 2, obviously).
In practice x
i
d,n.
The principal planes are not necessarily between the elements nor even near the
lens itself, as the examples in Fig. 3.7 show. A telescope in proper adjustment, for
example, has no focal length and its principal planes are at .
The principal planes are also called the planes of unit magnication. If some
other optical system to the left produces an image at the P
2
plane, the image will
be transferred with the same size to the P
1
plane and will appear to be there to any
subsequent system.
20 Geometrical optics
3.5 Optical layout
Once the positions of the focal planes and principal planes have been established,
rays can be drawn from an object point, through the system, to the image point.
The rules are:
r
A ray from a point A on the object, parallel to the axis, entering at any height h meets
the principal plane P
2
and emerges from P
1
at this same height and at an angle to make
it pass through the focal point f
1
.
r
If a ray from the same object point, drawn to the nodal (or, in practice, principal) point
of P
2
makes an angle with the optic axis it emerges from the nodal (principal) point of
P
1
at the same angle and continues until it meets the other, previously traced ray, at a
point A
.
Where they meet is the image of the object point A. The two points A and A
1
Figure 3.10 The HelmholtzLagrange invariant. It is a matter of elementary geo-
metry that the product h
i
i
is conserved throughout the system and there is a
similar product in the plane through the optic axis perpendicular to this diagram.
In practice the two invariants combine to give the law of conservation of etendue.
optical elements is a constant through a given centred system. To be precise,
n
1
h
1
tan
1
= n
2
h
2
tan
2
= n
3
h
3
tan
3
= = L.
where n
1
is the refractive index of the medium in the appropriate section. Usually
n = 1.
If the eld is anamorphic as in a wide-screen cine-projector for example
there are two such invariants, perpendicular to each other, and each including the
optic axis. The concept can be extended as follows:
The area of the eld, multiplied by the solid angle subtended there by the
pupil following it and by the square of the intervening refractive index, is a
constant of the system.
It can alternatively be dened as the area of the pupil multiplied by the solid
angle subtended by the preceding or following eld-stop.
This constant is called the etendue of the system. It is a particularly valuable
quantity to know when designing a photometer, since it immediately allows one to
calculate the ux of light that will be accepted.
In practice it is usually sufcient to dene it as
E =
(area of pupil) (area of eld)
(distance from pupil to eld stop)
2
.
and it is generally quoted as cm
2
sterad.
Etendue is of particular importance when applied to spectrometers. It is also
known as luminosit e and generally denoted by the letter L. In a spectroscope for
example it is easy to see that, when looking at a line emission spectrum, if the slit
width is doubled the amount of light entering the eye is doubled. At the same time
the resolving power the ability to see two lines as separate is halved. We shall
see later that if resolving power is dened as ,d where d is proportional to
the width of the slit, then the product, LR, of etendue and resolving power is a
constant for a particular design and E = LR is dened as the efciency of the
instrument.
3.10 Black body radiation 25
3.9 Surface brightness
The total power radiated by a source depends on its area and its surface brightness.
This is dened in two ways. It is either:
(1) the power emitted per unit area per steradian normal to the surface; or
(2) the photon ux emitted per unit area per steradian normal to the surface.
It can be measured in watts per unit area per steradian, or photons per second per
unit area per steradian. If the source is a Lambert radiator (it usually is not) then
the total power radiated into 2 steradians is I S, where I is the power per unit
area per unit solid angle normal to the surface, and S is the area of the surface.
This follows from Lamberts law.
3.9.1 Lamberts law
This states that the power per unit solid angle emitted by a luminous surface is
proportional to the cosine of the angle which the radiation makes with the surface
normal.
Seen from a distance, a tilted surface of nite area S appears foreshortened to
an area equal to S cos , and by Lamberts law, the radiation is similarly curtailed
in intensity.
Thus the surface brightness of a perfect Lambert radiator is independent of the
direction from which it is viewed. To a distant observer, for example, a radiating
spherical surface will be indistinguishable froma uniformly illuminated plane disc.
The idea is convenient geometrically, but in nature and in the laboratory Lambert
radiators are rare. The Sun, for example, is not a uniformly bright disc: its apparent
surface brightness at the edge is less than half that at the centre, and moreover is
wavelength dependent.
A perfect optical system will convert rays from a point source into a parallel
beam. Since the power radiated by a source is proportional to its area, there will
thus be no power in a parallel beam. The power will depend on the divergence of
the beam, which in turn is measured by the angular diameter of the source at the
entry pupil of the system.
3.10 Black body radiation
The most important practical consequence of Lamberts law is that if a Lambert
surface emits a total of I photons per second per unit area, then the apparent surface
brightness is I , photons per second per unit area per steradian. A black body for
example, emits T
4
watts per unit area into 2 steradians. is Stefans constant
26 Geometrical optics
and T the absolute temperature. The surface brightness normal to the surface is
then T
4
, watts per unit area per sterad.
5
3.10.1 Spectral power density
The total power emitted will depend on the nature of the emitting source. A black
body radiates power as a function of wavelength according to the well-known
Planck formula:
I () =
2hc
2
5
_
1
e
hc,kT
1
_
. (3.2)
and this can be converted to a formula for the photon ux:
N() =
2c
4
_
1
e
hc,kT
1
_
. (3.3)
The units in each case are watts (or photons per second) per unit area per unit
wavelength range radiated into 2 steradians. N or I must be divided by to give
the spectral surface brightness.
Most sources do not radiate as black bodies. A polished reecting surface has
a low emissivity and neither emits nor absorbs radiation efciently. Its emissivity
may be a function of wavelength, e(), which must multiply the Planck formula
above.
The important property of surface brightness is that when multiplied by the
etendue of the photometric instrument observing it, the result is the power or photon
ux entering the detector.
Remember that the units of etendue are (area) (solid angle) and that, multiplied
by the units of surface brightness (ux per unit area per unit solid angle), yields
ux.
These are the twofactors involvedincomputingthe yieldof anoptical photometer
or spectrophotometer with a given source.
Worked example Consider a photometer with a detector area of 1.5 cm
2
at the
focus of a lens 3 cm diameter and focal length 10 cm. Its etendue is (3,2)
2
1.5,100 = 0.106 cm
2
sterad. If it observes a distant extended source of luminosity
10
10
photons cm
2
s
1
sterad
1
it will receive radiationat the rate of 10
10
0.106 =
1.06 10
9
photons s
1
.
Notice that the answer does not depend on the distance of the photometer from
the source. In other words, if you sit facing a blank, uniformly illuminated wall
you cannot tell how far you are from it: its surface brightness does not change as
5
Not T
4
,2, because of the cosine factor.
3.10 Black body radiation 27
you move away. Each cell of the retina is a separate detector with an etendue in the
region of 10
6
cm
2
sterad. At a greater distance it will receive less light from unit
area of the wall, but there will be a correspondingly greater area of the wall imaged
on to each cell. The total light received by each cell is then unchanged.
However, when light arrives at a telescope from an apparently point object such
as a star it is the aperture of the telescope which determines the signal at the
detector, provided all the light from the object arrives at the same detector. The
noise generated by the detector depends on the sensitive area
6
and it is usually
desirable to have as small a detector area as possible. The numerical aperture or
focal ratio of the telescope is then important, as the diameter of the diffraction-
limited image is 1.22F where F is the focal ratio. Even when the size of the
image is determined by seeing to an angular size of about 1
y
T
S
P
3y
2
Ftan
Figure 4.2 The meridional or tangential coma of a lens. In the absence of other
aberrations, the marginal rays of an oblique bundle in the plane of the optic axis
meet at the tangential image point T, which generally is not on the oblique axis
the line joining the Gaussian image point P to the lens vertex. Tangential coma
is then measured as the distance in the focal plane from point P to the tangential
focus point T. Sometimes it is measured as the angle subtended by this distance
at the lens vertex. The oblique marginal rays perpendicular to this diagram meet
at S, the sagittal focus, and in the Seidel approximation the distance PS is always
1,3 of the distance PT.
eld angle of the chief-ray of a bundle and we therefore consider oblique bundles
of rays where the chief-ray is inclined at an angle to the optic axis.
In an oblique ray bundle, the paraxial rays arrive as before at the paraxial focus.
If there is no spherical aberration, the marginal rays converge to the same focal
surface, but they arrive at different points on that surface. Two pairs of marginal
rays are particularly important and may be said to dene the amount of coma.
Tangential coma
When the two marginal rays in the plane of the optic axis meet at a point which is
not on the paraxial ray, the perpendicular distance of this point from the paraxial
ray is the tangential coma. The height of the coma patch is
L
coma
= 3y
2
Ftan . (4.3)
F, the coma coefcient, and B, the spherical aberration coefcient, are the two
most important design parameters in a spectrograph. Appendix 1 gives formulae
for calculating B and F for single lenses and mirrors.
Sagittal coma
In a simple case, the two marginal rays from the ends of the diameter perpendicular
to the plane of the chief-ray arrive at a point between the paraxial focus and the
point where the meridional marginal rays meet. Its distance from the paraxial focus
is one-third of the distance to the meridional marginal focus and, by symmetry, is
on the line joining them. Other pairs of marginal rays arrive at points on a circle,
of which the meridional and sagittal foci are at the ends of a diameter. Ray pairs
from each annulus of the lens form circles of different diameter and the resulting
32 Optical aberrations
M
S
G
Figure 4.3 The sagittal coma of a lens. This is dened by the point S, where
marginal rays in the plane normal to the plane of Fig. 4.2 meet. Pairs of marginal
rays from other mutually antipodal points on the rim of the lens meet on a small
circle which has the tangential and sagittal foci at the ends of a diameter. Ray pairs
from other, smaller annuli form similar, smaller circles all lying cradled between
two lines of 60
. The illumination is uneven, being fainter nearer the outer edge, and it is from
this comet-like form that the aberration gets its name.
4.3.3 Astigmatism
This aberration is closely linked to eld curvature and the two are usually considered
together. Like the B and F coefcients of spherical aberration and coma, the other
two are C which vanishes for zero astigmatism and D which vanishes if the eld is
at.
Suppose that a distant wheel-like object in the form of concentric black circles
and radial lines on a white eld, centred on the optic axis and perpendicular to it, is
imaged by the system. If astigmatism is present, there are two focal surfaces, both
spherical. One of them will show the rings in focus and the other the radial lines.
The two surfaces have a common tangent at the Gaussian focus.
In neither case does a point on the object correspond to a point in the image.
An annular parallel bundle of rays inclined to the optic axis will rst converge
to form a short line image; a line which, if projected, will cut the optic axis: and
the bundle will subsequently converge to a short line image perpendicular to the
former. This is positive astigmatism. The position of the rst short line is the sagittal
focal surface the spokes of the wheel and the other the meridional or tangential
surface the rimof the wheel. The two do not necessarily happen in the order given
here. Astigmatism, like coma, may be negative.
In classical prism and grating spectroscopes it is the tangential surface which
chiey concerns us. This is where the spectrum lines are focused.
4.3 First-order aberrations 33
M
G
S
Figure 4.4 Astigmatism. In the absence of coma the tangential and sagittal ray
fans meet at different points along the oblique axis forming line images of a point
object. Where the rays of the tangential bundle meet, the tangential line image M
is perpendicular to the optic axis, and at the sagittal focus the sagittal line image S
if produced, would cut the optic axis. There is a circle of least confusion G, lying
between the two foci. In spectroscopy it is the tangential focus which is almost
always required. Consequently a spectrum line is not a precise image of the entry
slit but the aberration, being along the slit direction, does not interfere with the
spectral resolution.
Astigmatism, like coma, is an aberration of the eld and vanishes on the optic
axis. In order for it to exist at all there must simultaneously be eld curvature.
3
One
of the two eld curvatures may be zero. In a grating spectrograph, for example,
the tangential eld may be at and the entry slit will then be imaged, in various
colours, on to a plane. The image of a point on the slit is a short line lying along
the direction of the spectrum line image. Thus we do not notice it unless the slit is
very long or the spectral resolution very high.
4.3.4 Astigmatism and eld curvature
Field curvature on its own means that the image of a scene at innity will not lie
on a plane but on a sphere which has its centre on the optic axis and which cuts the
3
Although the latter may exist even in the absence of astigmatism.
34 Optical aberrations
Focal surfaces
T = tangential or meridional surface
S = sagittal surface
P = (fictitious) Petzval surface.
T S P
Figure 4.5 Field curvature. This is inseparable from astigmatism as the tangential
and sagittal images lie on two different spherical surfaces. If there is no astigmatism
the two surfaces are the same, and if eld curvature is also eliminated the eld is
at. There is a third, ctitious surface called the Petzval surface, the radius of which
is dened in the text, and the order T-S-P of the curvatures is always preserved.
optic axis at the Gaussian focus. The eld curvature is dened by the radius of this
sphere or by its reciprocal.
This sphere may be convex or concave towards the lens, i.e. its radius of curva-
ture may be positive or negative. For many purposes, such as photography, CCD
imaging, etc., it is highly desirable that the radius be innite so that the focal surface
is a plane. There is one notable exception: the Schmidt camera, where all aberra-
tions except eld curvature are perfectly corrected and a plane glass photographic
plate is bent to form a spherical surface (of fairly large radius, although glass is
more elastic than most people realise). The symbol c is used in some books for eld
curvature as well as surface curvature. To avoid confusion lower-case c is reserved
for surface curvatures of lenses and mirrors, and capital C
(sufx)
is used for eld
curvatures. C
P
, C
s
and C
t
refer to Petzval, sagittal and tangential eld curvature
respectively. C refers to the astigmatism coefcient and, like the other aberration
coefcients, is always in bold font.
4.3.5 Distortion
This is characterised by a fth coefcient E, and occurs when the position of the
principal plane depends on the eld angle, so that the focal length is a function of
the eld angle. In the absence of all other aberrations the focal length may either
increase or decrease with eld angle. If it increases with eld angle the result is
4.4 Theorems 35
object
object
sagittal focus tangential focus
barrel
distortion
pincushion
distortion
Figure 4.6 Images of a wheel-object, showing the practical effects of astigmatism
and distortion in the images which they produce. Distortion, usually barrel distor-
tion, is inevitable in lenses corrected to give very wide elds but is small or absent
in so-called symmetrical lenses such as the double Gauss camera objective.
pincushion distortion and if it decreases it produces barrel distortion. In other
words the image of a square object is shaped as the name implies. Barrel distortion
is difcult to avoid with wide-angle lenses.
4
4.4 Theorems
4.4.1 Fraunhofers theorem
This is an important theorem which shows how it is possible to place a stop to
control the coma of a system. If the entry pupil of a simple lens is at the lens itself,
the pupil distance is zero and the coma coefcient is written as F
0
.
Fraunhofers theoremis that if the entry pupil is at a distance Z fromthe principal
point, the coma coefcient becomes
F = (F
0
+ ZB). (4.4)
It follows then that if a stop is placed a distance Z from the vertex, given by
Z = F
0
,B, the lens will be free from coma. The distance is negative and so, by
the Cartesian sign convention, the stop must be to the left of the vertex.
Consider for example a spherical interface between glass and air, with a stop
at the centre of curvature of the glass sphere.
5
Parallel bundles of rays, no matter
4
Take especial care when choosing a lens for a FabryPerot spectrograph, where the dispersion is radial and only
a camera lens with freedom from distortion gives a true square-law wavelength scale.
5
This example is repeated in Appendix 1 with attendant mathematical details.
36 Optical aberrations
Figure 4.7 Controlling coma with a stop. In the upper gure a stop is placed at the
centre of curvature of a glass sphere so that collimated ray bundles coming fromthe
left are always symmetrical and are imaged with spherical aberration but no coma
on a spherical focal surface. In the lower gure the sphere has been truncated to
give a plano-convex lens with a plane rst surface, and the stop position is moved
so that the ray bundles reaching the second surface are unchanged. A stop at this
position in front of a plano-convex lens will thus eliminate coma.
in which direction they are travelling, are symmetrical about their chief-ray since,
because of the spherical symmetry, there is no unique optic axis. There must then
be no coma when the bundle comes to a focus after refraction, although there will
be spherical aberration.
Now place a second, plane, surface to the left of the spherical surface to make
a plano-convex lens. The stop, now in air to the left of the lens, can be moved a
distance R[(n 1),n] towards the lens so that parallel bundles of rays from it will
arrive at the second spherical surface just as if they had come from its centre of
curvature. To put it another way, the stop should be a distance f [(n 1),n] from
the plane surface of the lens. Again there will be no coma and the stop is at the
position required by Fraunhofers theorem.
This is an elementary example showing the underlying physics: the theorem is
of more general use than this.
A word of warning: coma can only be corrected with a stop if there is some
spherical aberration. It happens occasionally that it is worth leaving some spher-
ical aberration in the system deliberately in order that coma may be removed.
For example, the Ross corrector, which was used in large astronomical tele-
scopes with paraboloidal (and therefore spherical-aberration-free) mirrors, did just
this. The pupil distance is so large (from the telescope primary to the vicinity of
the focal plane) that the introduction of a totally insignicant amount of spher-
ical aberration is sufcient to allow coma correction over a large area on the
image.
4.5 Aberration coefcients for mirrors 37
On its own a singlet lens has no coma provided that its surface curvatures are
connected by
c
2
c
1
=
(n
2
n 1)
n
2
.
andtypically, for n = 1.516728, c
2
= 0.94c
1
will give coma belowthe diffraction
limit.
If the refractive index is 1.618 034 the golden section of the pyramid-builders
and Greek architects the coma of a plano-convex lens is zero when the rst surface
is convex. This is no gift of the gods however: all the other aberrations are there in
full measure!
4.4.2 Petzvals theorem
When astigmatism is present the tangential focal surface is a sphere of curvature
C
t
= 4C +2D, and the sagittal surface curvature is C
s
= 2D.
Petzvals theorem relates the two coefcients C and D by the equation
C D =
i
1
n
i
f
i
. (4.5)
where n
i
is the refractive index of the i th component. This called the Petzval sum
and denoted by the symbol
P
z
.
A surface can be dened, the Petzval surface, of which the curvature, C
P
, is
2(C D) and the two astigmatic surfaces have curvatures given by
C
t
= C
P
+
3
2
C.
C
s
= C
P
+
1
2
C.
C D = 0 is the condition for freedom from astigmatism, and C = D = 0 are the
conditions for freedom from both astigmatism and eld curvature.
4.5 Aberration coefcients for mirrors
These are collected here as they can be given in closed form and are most likely to
be used in spectrograph design. Formulae from which the equivalent coefcients
for lenses may be computed are collected in Appendix 1. When using them bear in
mind that these are only rst-order improvements on Gaussian optics and are not
exact. They give reasonable results at apertures slower than F,12 and eld angles
up to 5
1
1)C
1
+(n
2
1)C
2
.
6
These wavelengths correspond to the similarly lettered Fraunhofer lines in the solar spectrum.
40 Optical aberrations
where n
1
and n
2
are the refractive indices at the second wavelength. Then
Ln
1
C
1
= Ln
2
C
2
.
P = (n
1
1)C
1
(Ln
1
,Ln
2
)(n
2
1)C
1
= Ln
1
C
1
_
n
1
1
Ln
1
n
2
1
Ln
2
_
C
1
Ln
1
(:
1
:
2
)
and the sign indicates that an intermediate wavelength has been used for the
factors (n
i
1).
After a few more lines of algebra you will nd
P
1
=
_
:
1
:
1
:
2
_
P
and
P
2
=
_
:
2
:
1
:
2
_
P.
From these two equations the curvatures of the two elements can be computed.
However, nothing has been said about the individual curvatures of the four sur-
faces. These can be adjusted to make the spherical aberrations equal and opposite.
7
A minor caution, however: the chromatic aberration is not perfectly corrected. Al-
though the formulae above provide the same focal length for the two colours, the
principal planes and hence the focal planes are not exactly coincident and small
longitudinal colour remains. At F,10, though, this is comparable with the diffrac-
tion disc and is not signicant. Yet the ray tracer will correct for this also and will
remove one or other of these two aberrations. Finding tool radii to match is another
problem entirely and best left to the lensmakers.
The search for a respectable at eld of 10
20
f (t )e
2it
dt.
f (t ) =
_
()e
2it
d.
which we usually write symbolically as
f (t ) ().
Note that this particular notation frees the denitions fromthe normalising constants
which otherwise appear in front of the integral sign.
In geometrical optics we also have to deal with a spatial dimension and its
inverse, spatial frequency. On a grating for example, we may have a ruled width
x measured in centimetres and the neness of the grating measured in lines per
41
42 Fourier transforms: a brief revision
centimetre p. Similarly in photography, the resolution of a lens or a photographic
emulsion may be dened in line-pairs per centimetre, and by convention this uses
the symbol s.
5.2 Theorems
Various elementary theorems follow here and for brevity are given without proof.
Proofs are generally simple and to be found in any treatise on Fourier theory.
1
5.2.1 The addition theorem
f
1
(t ) + f
2
(t )
1
() +
2
(). (5.1)
5.2.2 The multiplication theorem
_
1
()
2
() d =
_
f
1
(t ) f
2
(t ) dt. (5.2)
where the asterisk denotes a complex conjugate. An important special case of this is.
5.2.3 Parsevals theorem
_
| () |
2
d =
_
| f
1
(t ) |
2
dt.
5.2.4 The shift theorem
f (t +t
0
) ()e
2it
0
. (5.3)
where t
0
is a constant. Its corollary is important:
f (t +t
0
) + f (t t
0
) 2() cos 2t
0
.
5.3 Convolutions
These are conveniently described by a spectroscopic example. No spectrometer is
perfect. It it were and if monochromatic light really existed then the spectrum
1
The exact and rigorous proof of the fundamental inversion theorem is anything but simple and to be found for
example in Titchmarsh, a bible to be consulted only by the most didactic reader.
5.3 Convolutions 43
I ()
(a)
(
0
)
I ()
(b)
C(
0
)
I ()
(c)
S(
1
)
1
S()
I ()
(d)
C(
2
1
) S(
1
)
1
1
2
Figure 5.1 Convolution of an instrumental prole with a continuous spectrum.
Each element of the spectrumis treated as a monochromatic emission line which is
converted by the shortcomings of the spectrograph to the shape of the instrumental
prole. The sum of all these conversions is added to give the observed spectrum.
of the light would be a Dirac delta function A(
0
) at the appropriate place in the
spectrum, where A is the intensity. In the real world the spectrometer produces a
curve of intensity against wavelength which we denote by AI (
0
) and which
either peaks at
0
or has its centre of gravity at
0
. The function I () is called the
instrumental prole of the spectrometer and
_
I () d = 1. The instrumental
prole in general does not hold its shape over the whole spectrum but the shape
changes slowly enough with wavelength that it serves to demonstrate the idea of a
convolution.
Consider now a real spectrum, S(), and in particular an innitesimal part, of
intensity S(
1
) d
1
at wavelength
1
. This approximates to a -function and will be
converted by the defects of the spectrometer to a function S(
1
) d
1
I (
1
) and
this in turn will contribute an innitesimal amount of power S(
1
)I (
2
1
) d
1
apparently at a wavelength
2
. The apparent spectrum, which is the total power
appearing at
2
from all parts of the true spectrum, is then the integral
P(
2
) =
_
0
S(
1
)I (
2
1
) d
1
.
and this is a practical example of the mathematical operation of convolution:
c(x
) =
_
f
1
(x) f
2
(x
x) dx. (5.4)
44 Fourier transforms: a brief revision
usually written symbolically as
c(x
) = f
1
(x) f
2
(x).
The operation can be visualised by drawing the function f
1
(x) on a sheet of paper,
drawing f
2
(x) on transparent paper, turning the drawing over and laying it on top
of the sheet with f
1
(x) and sliding it along, integrating the product of the two as
you go. Each integral gives one point on the convolution.
5.3.1 The convolution theorem
This theorem is a most important and powerful one, particularly in Fraunhofer
diffraction theory, and it greatly simplies the mathematics of diffraction.
It states that if c(x) is the convolution of f
1
(x) and f
2
(x) as above, then their
respective Fourier transforms ( p),
1
( p) and
2
( p) are related by
( p) =
1
( p)
2
( p). (5.5)
In words: the Fourier transform of a convolution is the product of the Fourier
transforms of the components.
5.3.2 Convolution algebra
There is an algebra of convolutions which can be used to manipulate them. It is not
difcult. The rules are as follows:
The commutative law
f
1
(x) f
2
(x) = f
2
(x) f
1
(x).
The associative law
f
1
(x) [ f
2
(x) f
3
(x)] = [ f
1
(x) f
2
(x)] f
3
(x).
The distributive law
f
1
(x) [ f
2
(x) + f
3
(x)] = f
1
(x) f
2
(x) + f
1
(x) f
3
(x).
but note that convolutions and multiplications must be kept separate. The distribu-
tive law does not hold in a mixture of them:
f
1
(x) [ f
2
(x) f
3
(x)] = [ f
1
(x) f
2
(x)] f
3
(x).
Nevertheless the algebra can be extended. For example,
[ f
1
(x) f
2
(x)] [ f
3
(x) f
4
(x)] [
1
( p)
2
( p)][
3
( p)
4
( p)]
where Greek and Roman letters denote Fourier pairs as before.
5.5 Useful functions 45
5.4 The WienerKhinchine theorem
We rst of all dene spectral power density (SPD), which is what a spectrometer
measures.
The autocorrelation function of a function f (t ) is similar to the convolution of a
function with itself, and is dened as
A() = lim
T
1
2T
_
T
T
f (t ) f (t +) dt.
where the factor before the integral implies a time average.
If the Fourier pair of f (t ) is (), the shift theorem gives the Fourier transform
of f (t +) as ()e
2i
and the multiplication theorem then gives
_
T
T
f (t ) f (t +) dt =
_
()()e
2i
d =
_
|()|
2
e
2i
d.
On the (reasonable) assumption that the power is zero before the beginning and
after the end of the signal receipt, we dene the SPD to be G() given by
G() =
|()|
2
2T
.
The WienerKhinchine theorem states, as in the equation above, that the SPD is
the Fourier transform of the autocorrelation function of the signal f (t ):
A() G().
or, more practically,
1
2T
_
T
T
f (t ) f (t +) dt
|()|
2
2T
. (5.6)
5.5 Useful functions
There are some mathematical expressions which recur frequently in physical sci-
ence in conjunction with their Fourier transforms or Fourier pairs. Some of them
are described here.
5.5.1 The top-hat function
This is a function which is zero from to a,2 and from a,2 to . Between
a,2 and a,2 it has the value 1. It is generally denoted by H
a
(x).
46 Fourier transforms: a brief revision
Its Fourier transform is
_
H(x)e
2i px
dx =
_
a,2
a,2
e
2i px
dx =
e
i pa
e
i pa
2i p
= a
sin pa
pa
= a sinc(pa).
and the so-called sinc function is much encountered in physics and engineering,
not least because data strings have a nite length and the length in turn limits the
frequency resolution. This is because the observed spectrum the Fourier transform
of the data is the convolution of the true spectrum with the sinc function.
5.5.2 The Gaussian function
The Fourier transform of the Gaussian function e
x
2
,a
2
is obtained directly by
completing the square of the exponent in the integrand:
_
e
x
2
,a
2
e
2i px
dx = e
2
p
2
a
2
_
e
(x,a+i pa)
2
dx
= a
2
p
2
a
2
.
The full width at half maximum(FWHM) of a Gaussian is 2a
ln 2, and numeri-
cally is 1.386a where a is the width parameter. The convolution of two Gaussians
of width parameters a and b is another Gaussian of width parameter
a
2
+b
2
, i.e.
the Pythagorean sum of the two width parameters.
5.5.3 The Lorentz function
When a damped electrical oscillator decays the amplitude it radiates varies with time
according to f (t ) = A
0
e
2i
0
t
e
t ,
, where
0
is its natural oscillation frequency and
the decay constant.
The spectrum of the oscillation is the Fourier transform of this:
A() = A
0
_
0
e
2i
0
t
e
t ,
e
2it
dt.
and the integral lower limit is 0 because the oscillation is deemed to begin at this
time. The integral is then
A() =
A
0
2i(
0
) 1,
.
5.5 Useful functions 47
This gives the spectral amplitude of the radiation. The spectral power density, the
square modulus of this, is A()
A(), which is
I () =
| A
0
|
2
4
2
(
0
)
2
+1,
2
.
and this is the prole of a spectrum line emitted in an atomic dipole transition.
The same function can be derived quantum mechanically. It is known as a Lorentz
prole.
5.5.4 The Voigt prole
This is a common shape for spectrum lines which have been broadened by temper-
ature in the emitting gas or plasma.
Temperature alone will broaden a monochromatic line to a Gaussian:
I () = I
0
e
(
0
)
2
,a
2
.
The width parameter, a, comes from the Maxwellian distribution of velocities, and
a
2
= 2
2
kT,mc
2
, where T is the absolute temperature, k is Boltzmanns constant
and m the mass of the emitting species.
Substituting values we nd that a monochromatic spectrum line becomes a
Gaussian with a FWHM L given by L, = 7.16 10
7
T,M where M is
the atomic or molecular weight of the emitter.
The observed spectrum line shape is the convolution of this with the Lorentz
prole resulting from the nite decay time of the oscillator, the so-called natural
width of the line.
5.5.5 Separation of the components of a Voigt prole
Given a spectrum line with a Voigt prole the width parameters of the two com-
ponents can be separated. Suppose that the width parameter of the Gaussian com-
ponent is a and that of the Lorentz component is b. Then the Voigt prole is the
convolution
V() = G() V() = e
2
,a
2
1,(
2
+b
2
).
Let n be the Fourier pair of . Then g(n) G() and l(n) L().
The Lorentz prole L() is, by the WienerKhinchine theorem, the Fourier trans-
form of the autocorrelation of the truncated exponential function which represents
the decaying amplitude of a damped oscillator. The truncation is because the decay
48 Fourier transforms: a brief revision
is deemed to begin at n = 0. Then,
l(n
) =
_
n
e
bn
e
(nn
),b
dn
= b,2e
n
,b
for positive values of n
2
n
2
a
2
e
n
,b
.
A graph of log :(n) is a parabola and the values of a and b can be extracted by
simple measurement.
5.5.6 The Dirac delta function
This has the property that (x) is zero everywhere except at the origin x = 0, and
at x = 0 is innite. The important property is
_
(x) dx = 1.
It may be regarded as the limiting case of a top-hat function as it becomes
innitesimally wide and innitely high with its area always unity, or as the limiting
case of a Gaussian:
lim
a0
1
a
e
x
2
,a
2
.
In either case it is an improper function which gives angst to mathematicians but
is useful to physicists.
2
5.5.7 The Dirac comb
This is an innite set of equally spaced -functions each separated from its neigh-
bours by a distance a. It is generally denoted by the Cyrillic letter III (Shah) and
written III
a
(x):
III
a
(x) =
n=
(x na).
Its Fourier transform is then
n=
_
(x na)e
2i px
dx =
n=
e
2inpa
. (5.7)
2
For example, it has no upper bound, which violates one of the Dirichlet conditions that dene a function to be
Fourier-transformable.
5.6 More theorems 49
f (x)
f
p
(x)
a
0
a
1
a
2
a
3
( p)
( p).III( p)
Figure 5.2 The sampling theorem. A continuous truncated function together
with its mirror image produces a real Fourier pair. The same function repeated
indenitely can be synthesised from the fundamental a
0
and its overtones
a
1
. a
2
. a
3
. a
4
. . . . . and these have the same values as the Fourier pair at the appro-
priate values of p. In practice it is the set of regular samples which provides the
experimental database and the function which produced them can be computed
via a FFT computer program.
and it can be shown
3
that this is another Dirac comb, (1,a)III
1,a
( p), so that we can
write
III
a
(x) (1,a)III
1,a
(x).
It is useful for example when describing the aperture of a diffraction grating as
the convolution of a top-hat function of width a with a Dirac comb of spacing b,
the convoluted product then being multiplied by a wide top-hat function of width
Nb to limit the width to N rulings.
5.6 More theorems
5.6.1 The sampling theorem
This is especially important in Fouriermultiplex spectrometry but it applies also
to the theory of diffraction grating spectrographs. It comes from Whittakers inter-
polary function theory
4
where it is called the cardinal theorem, and states, in effect,
that the amplitudes of all the frequencies that a signal contains can be measured
provided it is measured (sampled) at intervals of 1,2
f
, that is, to the reciprocal
of twice the highest frequency present.
5
It serves also to link Fourier series with
the Fourier transform.
3
R. N. Bracewell, The Fourier Transform and Its Applications (New York: McGraw-Hill, 1965).
4
J. M. Whittaker, Interpolary Function Theory (Cambridge University Press, 1935).
5
Sometimes called the Nyquist or folding frequency.
50 Fourier transforms: a brief revision
The signal which enters a spectrograph or spectrometer has an alternating electric
eld which can be described by F(t ) and this has a spectrum, S(), which may
extend from = 0 to
0
.
Consider a periodic function S
p
() which is identical with this spectrum from 0
to
0
, is the mirror image from
0
to 2
0
and is thereafter periodic with period 2
0
.
In other words, it is the convolution of S() and its mirror image in the origin with
a Dirac comb of interval 2
0
.
This periodic function can be written as the sum of an innite series:
S
p
() =
a
0
2
+
n=1
a
n
cos
2n
2
0
=
a
0
2
+
n=1
a
n
cos
n
0
. (5.8)
and then the amplitudes a
n
are given by
a
n
=
1
0
_
0
0
S() cos
n
0
d.
and the integral here need go only from 0 to
0
because the integrand is zero from
there on.
Now suppose we have a signal, F(t ), which contains frequencies up to
0
and
is empty from then on. Its spectrum, S(), is continuous from = 0 to =
0
.
Corresponding to this spectrum we may construct a symmetrical function S
s
()
such that S
s
() = 1,2(S() + S()), which is symmetric and stretches between
0
and
0
. The convolution of this function with III
2
0
is symmetric and periodic
with a period 2
0
and can therefore be synthesised from a discrete innite set of
cosines with coefcients which are a set of numbers a
n
. In Fourier language:
S
s
() =
1
2
[S() + S()] III
2
0
and the corresponding F
s
(t ) is
F
s
(t ) =
1
2
[F(t ) + F
(t )]III 1
2
0
(t ).
Since F(t ) is real, F(t ) = F
(t ) and F
p
(t ) = F(t )III 1
2
0
(t ), which is the set, F(nt
0
),
of samples of F(t ) at intervals t = 1,2
0
. From this set of samples we recover
S
s
() by the Fourier sum:
S
s
() =
F(nt
0
)e
2in,
0
=
F(nt
0
)e
int
0
.
5.7 Aliasing 51
Thus, in principle, a measurement of the incoming signal at intervals 1,2
0
is
sufcient to give the spectrum of the signal. The spectral resolution depends on the
number of samples taken of the signal, when the signs in the sum are replaced
by 0 and N, the (nite) number of samples taken.
The application of this theorem is particularly apparent in Fourier spectroscopy,
but the analogy in grating spectroscopy is apparent also when considering the sum
of amplitudes reected from a grating. The angle of diffraction and the implied
phase delay in the signal are analogous to the exponential in the Fourier sum and
different values of correspond to different angles.
The detector receives the sum of the complex amplitudes from the rulings, in
other words the sumof the samples, each delayed in time fromits neighbours by the
extra path difference it has travelled. This sum multiplied by its complex conjugate
is the power detected at each point of the spectrum.
5.7 Aliasing
The highest frequency,
0
, in the last section is also called the folding frequency
of the spectrum: the frequency beyond which there should be no power. If the sam-
pling interval is too large, the corresponding folding frequency may be too low for
the spectrum and the power beyond it at frequency appears in the spectrum at
frequency
0
. Power at frequency 2
0
appears at frequency 0. For example,
consider a cosine of amplitude a. It must be sampled at 0. . 2. 3 etc., and alter-
nate samples will have values a. a. a. a . . . From this the amplitude is correctly
recovered by the summing process. Had the samples been taken at 0. 2. 4. 6 . . .
they would all have had the same value, a, and the cosine would appear to be at
zero frequency.
Again, had the samples been taken at 0. 3. 6. . . the frequency would have
appeared in its proper place.
Aliasing can thus be put to good use. If the spectrum is conned to a small part
of the frequency domain it can be under-sampled deliberately. It may in fact occupy
the third, fth etc. period of the periodic function, S
s
(), but the Fourier-summing
process will put it apparently in the rst period. This is perfectly satisfactory pro-
vided we knowbeforehand what the sampling interval and the range of the spectrum
are, and it avoids the tediumof sampling for the empty part of the spectrumat lower
frequencies.
Again the analogy with grating spectra will be clear: we may look at the second
or third orders of diffraction, knowing beforehand that the rst-order spectrum at
longer wavelengths is absent, either by ltering or by detector impotence.
6
Physical optics and diffraction
6.1 Fraunhofer diffraction
Elementary textbooks dealing with a transparent aperture in physical optics reduce
the question of Fraunhofer or far-eld diffraction to a simple two-dimensional
form in which a plane wavefront passes through an opaque surface. There is no
variation perpendicular to the plane of the diagram and in this third dimension the
aperture is assumed to have unit length. A narrow strip of width dx along the x-
direction is a source of secondary wavelets and transmits, according to Huygens
principle, a wave with an amplitude proportional to the area A dx, where A is the
amplitude per unit area on the incident wavefront. In the far-eld approximation
the inverse-square law is ignored and the amplitude, including the phase,
1
at the
point P is dU, where
dU = A dx e
2i
(Rx sin )
.
where R is large compared with and x. If the amplitude A varies from point to
point along the aperture, the total amplitude arriving at point P is
U() = e
2i
R
_
A(x) e
2i
x sin
dx.
and the integral is over the whole of the transparent part of the aperture.
If A varies in the x-direction and if we replace sin , by a new variable p and
take the innite integral we nd
U( p) = C
_
A(x)e
2i px
dx (6.1)
so that
The amplitude U( p) diffracted in a direction is proportional to the Fourier
transform of the incident amplitude A(x) across the aperture
1
Remember that phase change =
2
path change.
52
6.1 Fraunhofer diffraction 53
x
s
in
x
dx
R
P
Figure 6.1 Fraunhofer diffraction by a slit aperture. Each element of the aperture is
treated as a separate source coherent with all the others and the several amplitudes
are added at the image point to produce a resultant which depends on the various
phases as well as the magnitudes of the elements.
and as usual in optics, A(x), the variation of incident amplitude with x, when
multiplied by its complex conjugate, is the transparency of the aperture.
2
In practice it is the intensity of the diffracted light which we measure and this
in turn is the (complex) diffracted amplitude multiplied by its complex conjugate,
and we write, remembering that C = e
2i
R
so that CC
= 1,
I ( p) = U( p)U
( p).
This can now be applied to various apertures and the usual diffraction patterns are
described.
6.1.1 Single-slit diffraction
A single slit of width n will diffract an amplitude a( p) where
a( p) = A
_
n,2
n,2
e
2i px
dx
= A
1
2i p
(e
i pn
e
i pn
)
= An
1
pn
sin(pn)
= Ansinc(pn).
where the sinc function (sin x,x) is as described in the previous chapter.
2
A(x) itself may be complex, since there may be variations in the optical thickness in what is otherwise a
transparent part of the aperture, and this will cause the phase to vary with x. This happens for instance in the
blazed transmission grating.
54 Physical optics and diffraction
The diffracted intensity then varies according to
I ( p) = I (0)sinc
2
(pn).
which is the familiar expression for diffraction by a single slit.
6.1.2 Two-slit diffraction
Two slits, as in the Youngs slits experiment, are represented by an aperture function
which is the convolution H
n
(x) [(x a,2) +(x +a,2)], where nis the width
of each slit and a their separation.
The Fourier transform of this is 2 sinc(pn) cos(2pa), and this, when multi-
plied by its complex conjugate, describes the familiar two-slit diffraction pattern.
Bear in mind that we can rejoin physical reality at any time by replacing
the abstract variable p by sin ,.
6.1.3 N-slit diffraction: the diffraction grating aperture
The original crude diffraction gratings were transparent slits ruled on an opaque
screen of smoked glass. Such an aperture can be represented by narrow top-hat
functions representing a single ruling, a Dirac comb representing the grid of aper-
tures and a broad top-hat function dening the limited width of the grating. The
result is
G(x) = H
Na
(x)[H
n
(x) III
a
(x)]. (6.2)
The width of a ruling is n, the spacing between adjacent rulings (the grating con-
stant) is a and there are a total of N rulings.
The Fourier transform of this is
I( p) = Na sinc(Npa)
_
nsinc(pn)
1
a
III
1,a
( p)
_
. (6.3)
and with p = , sin as before, this is the amplitude diffracted in the direction
. Notice in particular that because of the convolution with a Dirac comb, I( p)
will have a maximum whenever p = n,a, that is when sin = n,a. n is any
integer, positive or negative and ,a is the ratio of wavelength to slit spacing. The
application of this to diffraction grating theory will be elaborated in Chapter 8.
6.1.4 Diffraction with oblique incidence
If the wavefronts are incident on the diffracting aperture at an angle to the aperture
normal, we must take account of the phase at the aperture. The complex amplitude
6.2 Two-dimensional apertures and oblique incidence 55
x
s
i
n
x
s
in
R
P
R
Figure 6.2 Fraunhofer diffraction with oblique incidence. The different times of
arrival of the incoming wavefront at the different elements of the aperture imply
an extra phase term in the eventual sum of the elementary amplitudes at the image
point.
at z = 0 is then
A(x)e
2
x sin
.
and the diffracted amplitude is
U( p) =
_
A(x)e
2
x sin
e
2
x sin
dx.
For N slits, where A(x) = H
Na
(x)[H
n
(x) III
a
(x)], the diffracted amplitude has
a maximum whenever sin +sin = n,a.
6.2 Two-dimensional apertures and oblique incidence
Assume that the diffracting aperture in three-dimensional Cartesian coordinates
occupies the plane z = 0. A plane wavefront, incident obliquely, is conveniently
described by its direction cosines, l. m. and n. If the phase is zero at the origin,
the wavefront occupies the plane l x +my +nz = 0 and the phase at the point
P(x. y. 0) is (2,) times the perpendicular distance (l x +my) from the point to
the diffracting plane. The two-dimensional complex amplitude at the diffracting
aperture is then
A(x. y)e
2i
(l x+my)
.
and the diffracted amplitude is
U( p. q) =
_
A(x. y) e
2i
(l x+my)
e
2i
( px+qy)
dx dy.
where p and q are, as before, the direction cosines of the diffracted wavefront.
56 Physical optics and diffraction
A(x. y) for a diffraction grating is separable into A
1
(x) A
2
(y) and A
2
(y) is a top-
hat function with a width equal to the length of the grating rulings. The condition
nowfor a maximumof the diffracted amplitude is obtained as before, when l + p =
n,a and m +q = 0, but l and p are now direction cosines. l = sin i cos and
p = sinr cos , and the equation for a maximum, in Cartesian coordinates, becomes
(sin i +sinr) cos = n,a. (6.4)
where is the angle of inclination of the incoming chief-ray to the xz plane.
As with the diffraction grating, the amplitude at the diffracting aperture is often
separable into functions of x and y, and the integrals are separated and elementary.
However, if there is circular symmetry, the whole integral equation must be referred
to polar coordinates. The problem is not usually to be found in spectrograph design
and will not be pursued further here.
7
The prism spectrograph
7.1 Introduction
This is the traditional form of spectrograph or spectroscope, rst perfected by
Bunsen and Kirchhoff. In its transmission formthe prismis generally an equilateral
triangle in section. There is no absolute virtue in this as it is the prism base length
which determines the resolving power, and the equilateral is a compromise between
too great an apex angle, which results in more light reection at the surfaces, and too
small an apex angle, which results in an extra weight of glass at no extra resolution.
The angle of 60
=
cos
_
(A+D)
2
_
sin
_
A
2
_
1
2
D
.
1
It would be exact for a refractive index of
3 = 1.732 . . .
57
58 The prism spectrograph
A
h
D
Figure 7.1 The spectrograph derived fromthe classic prismspectroscope of Kirch-
hoff and Bunsen. The resolution depends on the width, h, of the emerging beam.
The ratio ,h determines the optical resolution via diffraction theory and the
dispersion requires an angle greater than this between two resolved wavelengths.
and on making the necessary substitutions:
b n, = h D,.
The Rayleigh criterion for resolution is D = ,h in this instance so that
d
= R = b
n
.
where b is the base length of the prism. Visual inspection with the telescope on a
goniometer table divided into minutes of arc will allow the refractive index to be
measured with a precision of 1 part in 10
4
, and interpolation between two known
lines on a photograph will achieve a wavelength measurement with an accuracy
of a few parts in 10
6
, bearing in mind that the centre of a diffraction-limited line
image can be estimated to 1%.
n, tends to vary rapidly with wavelength in highly dispersive glasses and
dispersioncurves for UV-qualityfusedquartz anda typical short int glass suitable
for prisms are shown in Fig. 7.2.
One of the chief reasons why prism spectrographs are still worth considering is
their freedom from grating ghosts. The line prole, provided that the prism is the
limiting aperture, is a sinc
2
function, and may have its secondary maxima further
reduced by a mask. This makes it sensitive to faint satellite lines on either side of a
parent line and this, together with the relative freedom from scattering when best-
quality glass is used, makes it suitable for Raman spectrography, where spectral
purityandnot highresolutionis the major criterion. Ina Littrowmounting, where the
base of the 30
60
90
Figure 7.2 The wavelength dispersion curves for two materials suitable for spec-
trograph prisms. The short int SF11 is satisfactory at wavelengths above
3200
A and fused quartz is unsurpassable between 3000
A and 2000
A. The
curves show resolving power of a prism with a 1 cm base length.
7.3 The focal curve theorem
This derives from the elementary consideration of chromatic aberration in a lens. If
most of the lens is removed, the images of a point object will appear in focus along
the optic axis, the red image most distant from the lens vertex. By extrapolation, a
prismwith lens of the same glass type to collimate and focus the light will showthe
spectrum of a point source lying along a line joining the red image to the source.
This is not exact of course, but it explains why the detector in such a spectrograph
lies at a steep angle to the emerging chief-rays. The focal surface is generally not
at but is a gentle curve which must be determined by calculation or ray tracing.
Achromatic or apochromatic lenses may be computed for collimator and camera
which will give a at focal surface, and camera lenses can be pressed into service
on both sides to give a low focal ratio, but it must be borne in mind that such lenses
are computed strictly for the visible region and lose resolution very rapidly outside
their intended spectral extent, especially at the UV end of the spectrum.
7.4 The Littrow mounting
This uses a 30
60
90
prism with half the weight of glass. With a high focal ratio and a singlet lens
60 The prism spectrograph
Figure 7.3 The traditional quartz Littrow mounting. With a resolution of about
0.1
A throughout the visible and near UV, this was state-of-the-art spectrography
in the mid twentieth century and much used for the analysis of molecular spectra.
acting as collimator and camera diffraction-limited resolution is possible, although
long exposure times are needed, and the photographic plate is at a steep angle to
the output chief-rays.
The Hilger 1
1
2
-metre quartz Littrow spectrograph (Fig. 7.3), which gave yeoman
service in the rst half of the twentieth century, was a typical example of a high-
performance Littrow prism spectrograph, working at a focal ratio of F,30 and
giving resolving power in the region of 40 000 in the near UV. The recording plates
were 250 mm long by 100 mm wide, were curved gently by pressing them against
a mandrel in the plateholder and gave full resolution along the whole length of
the spectrum. Ten or more spectra could be recorded on one plate by moving the
plateholder vertically with a rack-and-pinion motion.
7.5 The PellinBroca prism
This was an ingenious device (Fig. 7.4) for producing a monochromator with con-
stant deviation in this case 90
45
45
60
90
60
90
a
3
a
P 2
a
1
a
2
a
1
a
0
Figure 8.2 (a) The shape of the spectrumline when strictly monochromatic light is
diffracted. This is the rst factor in the grating equation and the width (FWHM) is
determined by the whole ruled width Na of the grating. (b) The second component
of the convolution in Eq. (8.1). The amplitudes of teeth of the Dirac comb are con-
trolled by the overall sinc
2
function derived from the width, n, of one ruling. The
abscissae here are the sines of the angles of diffraction divided by the wavelength
of the diffracted light. (c) When each tooth of the comb is replaced by the sinc
2
function illustrated in (a), the diffraction pattern of the grating is displayed.
66 The plane grating spectrograph
(a)
(b)
Figure 8.3 (a) Two sinc
2
functions representing two adjacent spectrum lines lie
with one principal maximum lying on the rst zero of the other. (b) When the
two are added the combined intensity prole has a 20% dip between the principal
maxima.
A more elaborate criterion may be taken as a measure of resolution, similar to
the optical transfer function used for photographic lenses, in which an image is
formed of an object that is a grid of alternate black and white lines. The visibility
2
of the fringes in the image, as a function of spatial frequency, is called the contrast
transfer function and is regarded as a measure of the spatial lter response of the
lens.
In the spectrograph a set of EdserButler fringes
3
may be observed, using a
Michelson interferometer rather than the usual FabryPerot etalon as the fringe
generator. With a small diameter source of white light, well collimated as it passes
through the interferometer, the intensity at the entry slit of the spectrograph will
vary sinusoidally with wavenumber or approximately with wavelength over a
short range of the spectrum. The visibility of the EdserButler fringes as a function
of path difference in the interferometer is then a measure of the spectral transfer
function, which in the absence of aberrations should fall to zero when the Michelson
path difference is equal to the ruled width, W, of the grating.
8.1.1 The grating equation
The elementary analysis assumed that light was incident normally on to the grating.
If the incidence is oblique, at an angle to the grating normal, there is an additional
phase term
4
covering the whole aperture and then a = ,(sin +sin ) or, more
2
Dened as : = (I
max
I
min
),(I
max
+ I
min
).
3
E. Edser & C. P. Butler, Phil. Mag., 46 (1898), 207.
4
As in Fig. 6.2.
8.2 Blazing of gratings 67
b
a
reflecting face
scarp face
Figure 8.4 A section through the reecting surface of a blazed grating. The height
of the step between rulings, the scarp face, is one half-wavelength of the blaze
or rst-order blaze wavelength. At this wavelength incident light at an angle b to
the grating normal is all reected at this same angle and in rst order of diffraction.
This is the angle and wavelength of maximum efciency.
familiarly, the grating equation:
sin +sin = n,a. (8.2)
Remember that and are on the same side of the grating normal.
8.2 Blazing of gratings
In early gratings where the width of each ruling was about 1,2 the ruling interval,
i.e. when n was a,2, only odd orders were present, the even orders having been
suppressed by the zeros of the broad sinc function lying on top of the diffraction
maxima.
In practice the grating rulings have the same width a as the spacing and are
inclined at the blaze angle b of the grating, so that a cross-section of the surface
has the appearance of Fig. 8.4. The effect of this blazing on a reection grating
can be described by changing in Eq. (6.2) the amplitude factor of one ruling from
H
a
(x) to H
a
(x)e
(2,)2x sin b
. The factor of 2 allows for the reection. Replacing
2 sin b, by q changes the corresponding Fourier transformfactor in Eq. (6.3) from
nsinc(pn) to nsinc(pa) ( p q) which, by the shift theorem (Eq. (5.3)) is
the same as nsinc(( p q)a). The maximumof the diffraction pattern then comes
when p = q, or =2b.
Thus for one wavelength, the blaze wavelength, all the incident light is diffracted
into rst order and the zero order is suppressed as are all other orders, by zeros of
the sinc function.
For neighbouring wavelengths the enveloping broad sinc function changes its
width at the same time as the diffraction peaks move in the opposite direction so
68 The plane grating spectrograph
that the maxima are not at the peak transmission. The efciency is less, and some
light is lost into other orders and scattered inside the spectrograph, but nevertheless,
as a rule-of-thumb, the efciency is greater than 50% over a range of wavelengths:
_
2
2n +1
_
b
- -
_
2
2n 1
_
b
.
where n is the order of diffraction and
b
the blaze wavelength.
5
8.3 Apodising
It is the rst factor in the diffraction equation which describes the shape of each
spectrum line. A purely monochromatic line is thus presented as a sinc
2
function
of FWHM ,Nn and with secondary maxima on either side of approximately
4% of the height of the parent. It is possible that these secondary maxima may
disguise real satellite lines or impersonate ctitious ones, and some effort has
been devoted to their suppression. The sinc
2
function occurs when the grating is
uniformly illuminated, giving a H
a
(x)-factor in the aperture function. This factor
can be modied by masking the grating with an aperture of a shape such that its
Fourier transform alters the shape of the line prole. A diamond-shaped mask for
instance will change the rst term of the aperture function from H
Na
(x) to A
Na
(x)
where the A-function, or triangle function, is the convolution of two H-functions:
A
a
(x) = H
a,2
(x) H
a,2
(x).
Here, two top-hat functions each of width Na,2 are combined to form A
Na
(x).
By the convolution theorem the Fourier transform of A
Na
(x) is
(N
2
a
2
,4)sinc
2
(pNa).
and the intensity distribution in a spectrum line is
(N
2
a
2
,4)
2
sinc
4
(pNa).
This has the merit of suppressing the secondary maxima from 4.5% to 0.2%
of the principal line, a possibly useful gain if the ne structure of a line is being
examined. The concomitant demerits are
(1) that by covering half the grating area, half the intensity is lost and
(2) that by restricting the effective width, the resolving power is reduced by 40%.
There are various shaped masks, known as windows, illustrated in Fig. 8.5,
which will redistribute the power in the secondary maxima in various ways and the
5
For ne details of the often complicated efciency curves of diffraction gratings see: C. Palmer, Diffraction
Grating Handbook, 6th edn. (Rochester, NY: Newport Corporation, 2005).
8.4 Order overlap and free spectral range 69
Figure 8.5 Apodising masks for reection gratings. The Hanning mask with
the sinusoidal variation of amplitude across the face of the grating has many
advantages, but the area is still open to experiment. The light which fails to reach
the principal maximum can be redistributed almost at will among the secondary
maxima.
problem of side-lobes is one that is common to all spectroscopy. The Hanning
and Hamming windows of communication theory
6
are typical but in the normal
spectrograph the combination of the rectangular grating aperture and the usually
circular pupils which are superimposed on it compromise all these windows.
The masking process is generally known as apodising
7
and there are legions
of apodising masks known and advocated, chiey by their inventors.
8.4 Order overlap and free spectral range
Despite blazingandother improvements it is not possible tosuppress bythese means
unwanted radiation from other orders of diffraction from appearing in a spectrum.
A wavelength diffracted in rst order is accompanied by ,2 in second order, by
,3 in third order and so on, to the confusion of the investigator. It may well be that
the detector is insensitive to this unwanted radiation but in general precautions must
be taken to exclude it. Usually simple colour ltering is adequate and in particular in
the visible region in rst-order working, a glass which absorbs UVlight is sufcient
6
See R. B. Blackman & J. E. Tukey, The Measurement of Power Spectra (New York: Dover, 1958).
7
From the Greek without feet. The same apodising process is used in Fourier-multiplex spectrometry where
it is much more effective and more necessary, since the raw instrumental prole there is a sinc function. In
that case apodising is part of the computation process and involves interferogram samples being appropriately
weighted to form the apodising mask.
70 The plane grating spectrograph
to allow an octave of the spectrum from 8000 to 4000 to be photographed
unambiguously. However, it may be necessary to use a higher order with an ap-
propriately higher blaze angle to obtain the necessary resolution, and in this case
narrow-band lters are required to suppress both higher and lower orders.
8.4.1 Free spectral range
If a wavelength is to be examined in nth order, then sin i +sinr = n,a. At
these same angles i and r, n will appear in rst order, n,2 in second order, and
generally, n,m in mth order. To avoid confusion all these other wavelengths must
be excluded. The extent of the spectrogram, the range of wavelengths that is, which
can be examined without confusion in nth order, stretches from to L at
which point appears again in (n +1)th order. L comes from
(n +1) = n( L) whence L = ,n.
8.5 Grating ghosts and periodic errors
8.5.1 Rowland ghosts
It sometimes happens that the rulings of a grating are not separated by precisely
equal spaces. Particularly in older ruling engines which relied on mechanical pre-
cision, a periodic error in the master-screw would cause a cyclic variation in the
position of successive rulings. When this happens a monochromatic spectrum line
may be accompanied on each side by faint satellite lines known as ghosts, some-
times by a whole series of themof steadily decreasing intensity. There is an analogy
with the side-bands of a frequency-modulated radio transmission, where the side-
bands contain the information signal.
The problemmay be described using Fourier theory. The Dirac comb which rep-
resented the positions of the rulings is replaced by a set of -functions in which the
nth is displaced from its proper place by a cyclically varying amount sin(2sna).
is the amplitude of the periodic error and 1,s, the pitch of the lead-screw of the
ruling engine, is its period. The aperture function for the grating then replaces the
Dirac comb III
a
by
A(x) =
n=
[na + sin(2sna)]. (8.3)
and its Fourier transform is
I( p) =
n=
e
2i p[na+ sin(2sna)]
. (8.4)
8.5 Grating ghosts and periodic errors 71
with p = sin , as usual. From Bessel function theory we have the Jacobi
expansion:
e
ix sin y
=
m=
J
m
(x)e
imy
.
where J
m
(x) is the mth-order Bessel function. Using this with x = 2pn and
y = 2sna, the sum becomes
I( p) =
n=
e
2i pna
m=
J
m
(2p)e
2imsna
. (8.5)
which, when rearranged is
I( p) =
n=
m=
J
m
(2p)e
2ina( p+ms)
. (8.6)
We suppose now that the amplitude, , of the periodic error is small compared
with the grating constant a. For small values of the argument the Bessel functions
approximate to
J
0
(x) = 1. J
1
(x) =
x
2
. J
2
(x) =
x
2
4
. etc.
and
I( p) =
n=
e
2inap
+
n=
p e
2ian( p+s)
+
n=
p e
2ian( ps)
+
n=
2
p
2
2
e
2ian( p+2s)
+
n=
2
p
2
2
e
2ian( p2s)
+ (8.7)
This represents an innite series of Dirac combs. The rst is the one which
appears in Eq. (8.1), representing the perfect grating. The second pair represent
Dirac combs with amplitudes attenuated by a factor p and with periods 1,( p +s)
and 1,( p s). The third pair are attenuated by
2
p
2
2
, with periods 1,( p +2s)
and 1,( p 2s). These describe the ghost lines, and in rst order the strongest
are separated from the parent line by L =
2
,(lead-screw pitch). The effect is
shown in Fig. 8.6.
8.5.2 Lyman ghosts
These are faint images of a line which appear apparently at sub-integer orders and
are chiey found in XUV spectroscopy when older, mechanically ruled concave
gratings are used in rst order. They may be ascribed to defects on the ruling tip
72 The plane grating spectrograph
x
2
x
2
4
(x = 2 p)
Figure 8.6 Rowland ghosts surrounding a spectrum line. These are separated by
amounts which depend on the pitch of the ruling engine lead-screw and by ampli-
tudes which depend on the amplitude of the pitch error. In a modern ruling engine,
where the position of the ruling tip is optically controlled, there should be no such
periodic error and hence no Rowland ghosts.
which make an irregular but constantly repeated ruling prole so that, perhaps as a
result of aliasing, the III(x)-function which describes the grating appears to contain
another such function in its interstices.
8
This imitates a grating with two, three or
more times the number of rulings and consequently two, three or more times the
dispersion.
Lyman ghosts are a possible but very unlikely source of confusion unless very
large dynamic range and sensitivity are factors in the investigation.
8.6 The complete grating equation
A complete description of the diffraction grating must take account of the length
of the rulings as well as the ruled width. The Fraunhofer diffraction pattern is then
two dimensional and the two-dimensional Fourier transform must be used with
allowance made for incidence which is oblique in both axes.
8
Crudely, if the H-function which naively describes one ruling were replaced by two narrower H-functions,
the grating would have twice the number of rulings, not necessarily uniformly spaced. If, for example, the
H-functions were separated by 1,3 of the grating constant, there would be a contribution to the spectrum at
order 1,3.
8.6 The complete grating equation 73
R
2
R
1
B
P(x, y, 0)
A
R
1
R
2
y
x
z
1
and R
2
are the same rays after reection.
Perpendiculars from the reection point of R
2
on to R
1
and R
1
lie in the incoming
and outgoing wavefronts and cut off lengths which add to give the path difference
between the two. When the reection point is one ruling away from the origin the
path difference must be one wavelength.
However, the alternative derivation of the full grating equation is both simple
and instructive. Suppose that the grating occupies the plane z = 0 and that the
rulings are parallel to the y-axis. Consider two parallel rays from the collimated
incident beam incident on the grating, ray R
1
at the origin and ray R
2
at the point
P(x. y. 0) (Fig. 8.7). Suppose that their direction cosines are l
1
. m
1
. n
1
and that the
directioncosines of the correspondingdiffractedrays R
1
and R
2
arel
2
. m
2
. n
2
. Byan
elementarytheoreminCartesiangeometry, perpendiculars fromthe point P onto R
1
and R
1
intercept lengths PA = l
1
x +m
1
y +0z and PB = l
2
x +m
2
y +0z. This
path difference between the two rays must be an integer n number of wavelengths.
Thus,
AO + BO = (l
1
+l
2
)x +(m
1
+m
2
)y = n.
where n is the order of diffraction.
This must hold for any value of y, so that (m
1
+m
2
) = 0 and (l
1
+l
2
)x = n.
If point P is on a ruling adjacent to the y-axis, then the grating equation appears:
(l
1
+l
2
)x = n,a. or (sin i +sinr) cos = n,a.
8.6.1 Distortion of line shapes
Some important consequences come from this equation. Consider rst of all the
chief-rays of wavelength incident on to the grating at its vertex from points on
74 The plane grating spectrograph
a
(a,i)
(a, r)
i
r
0
Figure 8.8 The diffraction of rays fromdifferent points on a straight entry slit. The
complete grating equation requires a conical fan of diffracted rays, which in turn
results in a curved image of a straight entry slit.
a straight entry slit parallel to the y-axis. All the x-direction cosines, l
1
, are zero.
The diffracted chief-rays will therefore all have the same value of l
2
, so that they
generate a circular cone with its axis on the x-axis of coordinates. If the focal plane
of the spectrograph is parallel to the grating surface, the images of the entry slit will
occupy a hyperboloid, where the focal plane intersects the chief-rays. Alternatively,
if the focal plane is perpendicular to the surface and parallel to the rulings, the
images will be arcs of circles. This curvature of line images is potentially a major
cause of loss of resolution. The image curvature varies with wavelength and in the
calibration of the spectrograph due account must be taken of it.
The proper descriptive tool here is spherical trigonometry, in which the grating
normal lies on the equator and the rulings are parallel to the polar axis. In Fig. 8.8,
angles like i and r are longitudes and chief-rays approach the centre of the unit
sphere from various latitudes and longitudes. A chief-ray coming from latitude
and longitude i is diffracted to latitude and longitude r, connected by
sinr = (n,a) sec sin i.
and for small values of the approximation sec = 1 +
2
,2 is adequate. If
the equatorial chief-ray is diffracted to r
0
, the inclined one arrives at longitude r
where
sinr sinr
0
= (n,2a)
2
.
r = (n,2a)
2
sec r
0
. (8.8)
For small the curve of against r is a parabola and describes the shape of
the Gaussian image of a straight input slit. In practice this may sometimes be
8.6 The complete grating equation 75
i
,
Figure 8.9 An entry slit which is the arc of a circle will produce an image at one
wavelength which is a continuation of the same circle. In a scanning monochro-
mator, images of one half-circle will appear on the other half-circle at whatever
wavelength is returned from the tilted grating. The wavelength scanning is then
done simply by rotating the grating about an axis parallel to the rulings.
compromised by the aberrations of the system but is nevertheless a useful guide to
the correction process. The constant of the parabola changes with wavelength and
at high resolution each wavelength requires separate treatment to straighten its line
image.
8.6.2 The circular slit theorem
Another theorem, chiey of interest to monochromator constructors, is worthy of
mention in this context.
In Fig. 8.9, consider a point at latitude 0, longitude 0 and suppose that this makes
an angle i with the optic axis of the spectrometer. Draw a small circle of angular
radius with its centre at this point. The coordinates (. ) of a point on this circle
are connected by another well-known theorem in spherical trigonometry:
cos cos = cos .
Two diametrically opposite points on the circumference of the small circle then
have coordinates (. ) and (. ), and if they are also points where incoming
and outgoing chief-rays cut the sphere, then
cos [sin(i +) +sin(i )] = 2 cos sin i cos = m,a.
or
2 cos sin i = m,a.
76 The plane grating spectrograph
Since i is constant, every diametrically opposite pair of points on the small circle
are conjugate points for the same wavelength. An input slit which is the arc of a
circle will result in an image at wavelength on another arc of the same circle.
Different wavelengths will provide images at the same circular output slit as the
grating is turned about its axis.
8.7 Differential dispersion
In normal practice the spectrograph slit widths are not determined by the ultimate
resolving power theoretically available. The entry slit width in such an instance
would be such that the rst zeros of its diffraction pattern the sinc
2
function
would fall at the edges of the grating. However, in a spectrograph the resolution
may be set by the width of the CCD pixels or the average size of the silver halide
crystals of a photographic emulsion. The slit width is then chosen to match this
width and there is no resolution to gain and much light to lose in making it narrower.
From the grating equation
(sin i +sinr) = n,a
we get
cos i di = n d,a.
whichrelates the resolutiondtothe angular slit widthdi andfor a xedwavelength,
cos i di = cos r dr. (8.9)
which relates the width of the input slit to the width of its image. This is the lateral
magnication, which is a function of wavelength, and must be taken into account
during the design.
8.8 Mounting congurations
Since the time of Bunsen and Kirchhoff and until the end of the nineteenth century
the optical arrangement for a spectroscope had comprised a slit, a collimator, a
disperser, and a telescope. For photography the eyepiece could be replaced by a
photographic emulsion. It was Ebert, following Rowlands production of research-
quality diffraction gratings, who proposed an achromatic spectrograph using a
single spherical mirror both as collimator and telescope and eliminating the un-
wanted longitudinal dispersion inherent in a simple lens system.
9
9
H. Ebert, Annalen der Physik, 38 (1889), 489.
8.8 Mounting congurations 77
entry slit
grating
Ebert mirror
photographic
plate
Figure 8.10 The EbertFastie mounting. The rst achromatic method of mounting
a high-resolution spectrograph, with almost miraculous correction of the gross
aberrations. However, it proved to be endishly difcult to achieve good optical
alignment and it consequently lost ground to the
CzernyTurner mounting with
similar achromatic properties, where the separate mirrors gave more degrees of
freedom and greater exibility in practical alignment.
Strangely it seems to have been ignored until it was rediscovered by Fastie,
10
and meanwhile
Czerny and Turner had replaced the two traditional lenses by
two spherical mirrors.
11
Littrow similarly had evolved an arrangement using
a single spherical or paraboloidal mirror which would yield diffraction-limited
resolution.
8.8.1 The EbertFastie mounting
One of the features of this mounting (Fig. 8.10), which it shares with the
Czerny
Turner mounting, is the partial correction of tangential coma, leaving spherical
aberration and astigmatism as the two main aberrations. A proper choice of pos-
ition for the grating ensures a at meridional eld. Astigmatism is tolerated be-
cause only the meridional focus is required and distortion is ignored since the
spectral dispersion is in any case non-linear. Although it is capable of yielding
a high resolution at a fairly high numerical aperture, the restriction imposed by
the single mirror makes the optical alignment of the Ebert mounting unnecessarily
difcult and it has largely been discarded in favour of the
CzernyTurner mount-
ing in most commercial designs. Only two portions of the mirror are used there
can be no overlap between incoming and outgoing beams and a mask must be
inserted in the centre of the aperture to prevent undispersed light falling on to
the detector. This reduces the device essentially to the same conguration as the
sin =
cos
3
r
cos
3
sin . (8.10)
where i and r are as usual the incident and diffraction angles, and and are the
eld angles of the incoming and outgoing beams at the M1 and M2 mirrors. The
condition can be derived from elementary Seidel theory as follows.
In Fig. 8.14, parallel rays starting from the grating and reected from the M1
mirror will converge to a comatic image at the entry slit, the marginal rays showing
an angular coma
coma
. Conversely, marginal rays starting precisely from the entry
slit will arrive at the grating edge at this same angle
coma
to the paraxial ray. After
diffraction this angle will be multiplied by the factor (cos i,cos r) of Eq. (8.7)
and this must be added to the coma produced by the output mirror M2. The total
tangential coma is then
C
m
= 3F
1
y
2
1
tan (cos i , cos r) +3F
2
y
2
2
tan . (8.11)
where subscripts refer to the respective mirrors.
The coma coefcient F (Eq. 4.8) is given by
F = f
2
,4 Z(1 e
2
),8 f
2
.
14
G. R. Rosendahl, J. Opt. Soc. Am., 52 (1962), 412.
82 The plane grating spectrograph
r
i
Figure 8.14 The angles involved in the calculation of the coma of the
Czerny
Turner mounting. The rays here are of course the chief-rays of each bundle.
For paraboloidal mirrors, which as we shall see later are preferable to spherical
mirrors, this reduces to f
2
,4.
The geometry of the
CzernyTurner arrangement provides that y
1
=
(W,2)(cos i,cos ) and y
2
= (W,2)(cos r,cos ) and if as usual (though not neces-
sarily) both mirrors are identical, then F
1
= F
2
= F and the total tangential coma is
C
m
= (3,4)F[(cos i,cos )
2
(cos i,cos r) tan (cos r,cos )
2
tan ]. (8.12)
and when C
m
= 0 this is equivalent to the Rosendahl condition.
The condition for vanishing of tangential coma derived here is more general
than the Rosendahl condition as it allows for M1 and M2 mirrors of different
focal lengths and it also allows exact computation of the coma when there is some
offence against the Rosendahl condition. The condition holds strictly only for one
wavelength and it is only the tangential coma which is corrected. Sagittal coma
remains, albeit somewhat reduced. There is no foreshortening of the aperture on
diffraction and the sagittal coma, one-third of the tangential coma, is F(L,2)
2
tan .
This is converted to F(L,2)
2
tan (cos i,cos r) on diffraction, and supplemented by
F(L,2)
2
tan from the M2 mirror. When the Rosendahl condition is met this gives
C
s
= F(L,2)
2
tan
_
1 (sin ,sin )
2,3
_
. (8.13)
which is substantially less than the coma of a paraboloid at the same eld angle
away from the zero-coma wavelength.
8.8.6 The generalised Rosendahl condition
To satisfy the Rosendahl condition we abandon the cherished concept of the sym-
metrical
CzernyTurner conguration and allow the grating to be nearer (usually)
8.8 Mounting congurations 83
to the input-side optic axis. If we also make allowance for the M1 and M2 mirrors
to be of different focal lengths, the condition can be rearranged in a form more
convenient for computation:
_
cos i
cos r
_
3
=
_
f
1
f
2
_
2
_
cos
cos
_
3
_
sin
sin
_
. (8.14)
Although it is not necessary for the input and output chief-rays to be parallel, it is
convenient and it gives us another constraint:
r i = 2( +) = Q. (8.15)
With r replaced by Q +i , the grating equation becomes
2 cos(i + Q),2 sin(i Q),2 = ,a. (8.16)
where is the desired zero tangential coma wavelength. FromEq. (8.14), i is found
by iteration. From Eq. (8.13), r is found. Then by substituting values in Eq. (8.12),
is obtained to any degree of accuracy by another iteration. Finally the grating tilt,
t , is given by t = ,2 2 i .
8.8.7 The at-eld condition
While photographic plates and lms can be bent gently to t the eld curvature of
an optical instrument, the crystalline nature of a CCD or similar detector requires
an accurately at focal surface. The expression for tangential eld curvature of a
convex mirror is (Eq. 4.12)
C
t
=
2
f
+
3Z
f
2
+
3
4
Z
2
(1 e
2
)
f
3
.
and for a spherical mirror C
t
is zero when Z = 2 f (1 1,
= ,a
. Both diffracted
rays intersect the Rowland circle at y. Then by geometry,
sinr =
H
2R
. sinr
=
H
r
g
.
R
g
2R
= cos .
1
H. A. Rowland, Phil. Mag., 16 (1883), 197.
89
90 The concave grating spectrograph
r
r
H
o
y
x
r
g
r
g
r
g
sec
r
g
sec
Figure 9.1 The basic geometry of the concave grating. The grating has been moved
slightly (about 1 mm) to ease the geometry. Elementary diffraction theory predicts
a focus at the point y provided the grating constant changes appropriately from
the vertex to the edge.
By grating theory,
sinr =
a
. sinr
=
a
.
and by elimination, a = a
sec .
The focusing property therefore follows from the requirement for the spacing a
to be constant along a chord. This argument, like the theorem, is not exact but the
device works well enough within the limits of practical grating construction.
In practice it is the tangential focal surface which is supposed to lie on the
Rowland circle, and a monochromatic point source is imaged as a line segment,
long or short according to the dimensions of the grating.
2
A theory of the focusing property of the Rowland grating is to be found, with
some approximations, in R. W. Woods Physical Optics.
3
However, the strict, three-
dimensional theory is derived by Beutler
4
and a restricted form of this is presented
here as an illustration of the different ways in which the grating may be used.
Beutler used the so-called eikonal method in which the length of a ray is traced
from an object point to an image point and measured in terms of the positions of
2
To be more precise the tangential focal surface is a cylinder coaxial with the Rowland circle and containing it,
and the sagittal focal surface is a similar cylinder turned through a right angle about the grating normal. These
are in accordance with Eqs. (4.11) and (4.12).
3
R. W. Wood, Physical Optics (New York: MacMillan, 1905).
4
H. G. J. Beutler, J. Opt. Soc. Am., 35 (1945), 311.
9.1 The Rowland grating 91
x, w, h
r
cos r, sin r
cos i, sin i
i
Figure 9.2 The geometry of the eikonal method used by Beutler for a strict descrip-
tion of the geometry of the concave grating. The method measures the path length
from an arbitrary point on the Rowland circle to another arbitrary point where
an image is formed, and the length is given as a function of wavelength, grating
constant and the coordinates of the point of reection. In the present analysis a
simplied eikonal, ignoring the third dimension, is used.
the object and image and the coordinates of the intersection with the grating.
5
Polar
and Cartesian coordinates are used as in Fig. 9.2.
6
The object and image positions
are (. i. 0) and (
n(sin i +sinr) +
mn
a
+
n
2
2
0
__
sin i
_
n
_
cos
2
i
cos i
R
_
+
_
sinr
_
n
_
cos
2
r
cos r
R
__
.
(9.1)
The vanishing of various powers of n gives ever more stringent conditions for
an image to be formed at the image point.
(1) The vanishingof the rst power of ngives the standardgratingequation(sin i +sinr)
m
a
= 0.
(2) The vanishing of the second power gives
_
cos
2
i
cos i
R
_
+
_
cos
2
r
cos r
R
_
= 0. (9.2)
5
See e.g. M. Born & E. Wolf, Principles of Optics (Cambridge University Press, 1999), Chapter 3. Not always
to be recommended because although it is precise, it involves miserably complicated and tedious elementary
algebra and in practice yields little that is not available through Seidel theory.
6
Beutler has the x-coordinate as the optic axis.
92 The concave grating spectrograph
(3) The vanishing of the third power gives
sin i
_
cos
2
i
cos i
R
_
+
sinr
_
cos
2
r
cos r
R
_
= 0. (9.3)
and conditions (2) and (3) can be satised simultaneously by
_
cos
2
i
cos i
R
_ _
sin i
sinr
_
= 0. (9.4)
The rst factor gives the Rowland condition. The second factor yields some
alternative congurations which do not use the Rowland circle. This factor can be
combined with Eq. (9.2) to yield
=
R[cos
2
i sinr +cos
2
r sin i ]
sinr[cos i +cos r]
(9.5)
and
=
R[cos
2
i sinr +cos
2
r sin i ]
sin i [cos i +cos r]
. (9.6)
Two results follow:
(1) If i = r, then =
30
and the offset of the rotation axis from the grating vertex as
l = R sin [1 tan (tan
0
tan
0
),2].
10
This is a device for wavelength calibration. It is a plate covering the input slit with three rectangular holes,
staggered, in it. It can be slid across the slit so that the lower third, the middle third or the upper third of the
slit can be photographed in succession, with a reference spectrum in the rst and last cases. Line curvature
sometimes makes it of dubious value.
11
J. F. James, J. Sci. Instrum., 36 (1959), 186.
12
M. Seya, Sci. Light, 2 (1951), 8. Seyas mounting follows fromthe solution of a transcendental equation derived
from the Beutler eikonal.
13
P. D. Johnson, Rev. Sci. Instrum., 28 (1957), 833.
14
R. Onaka & I. Fujita, Sci. Light, 9 (1960), 31.
9.4 The aberrations of the Rowland grating 97
grating rotates
about its vertex
70 30
0.8179 r
g
both slits are fixed
Figure 9.7 The SeyaNamioka mounting. This tooabandons the Rowlandfocusing
property, but, as discovered by Seya, two points are conjugate provided they are
on the Rowland circle and separated by an angle of 70
30
. Wavelength scanning
is then by rotation of the grating about its vertex. The astigmatism however is a
serious drawback with this type of mounting.
However, a great demerit appears when these designs are subjected to ray tracing.
Giventhe usual lengthof gratingrulings, there is a verylarge amount of astigmatism,
up to half the length of the grating rulings, with its accompanying defocus. It is
sufcient to dilute line images to an unacceptable degree and only a few milli-
metres of each ruling on either side of the plane of symmetry make any effective
contribution to the measurable line images.
9.4 The aberrations of the Rowland grating
Astigmatism is the chief source of imperfection in any mounting. The full Beutler
eikonal is required to discuss it, and if this is done an expression in closed form for
astigmatism emerges. A point source on the Rowland circle with the usual angles
of incidence and diffraction i and r, will produce an image at the tangential focus,
also on the Rowland circle, of length L where
L = l
_
sin
2
r +sin
2
i
cos r
cos i
_
= Al. (9.7)
where l is the length of the rulings. The coefcient A is required later.
9.4.1 Astigmatic curvature
The astigmatic image of a point source is not straight and neither is the image of a
long input slit, for the reason outlined in Subsection 8.6.1. The two curvatures are
98 The concave grating spectrograph
in opposite directions and the overlapping images from different parts of the input
slit yield a coma-like aberration which limits the resolution.
15
There are two cases
to consider.
(1) Grating rulings of nite length and a point source on the Rowland circle.
(2) A long straight slit and innitesimal grating rulings.
To simplify the analysis we dene by
1
2
= [i sin
2
i +sinr tan
2
r +sin i sin
2
r tan
2
r sin 2i sinr tanr].
Then dr, the width of the line image, is in each instance:
(1) Long rulings, short input slit:
dr =
l
2
2R
2
whence l =
2R dr.
and l is the length of the longest rulings that can be used.
(2) Short rulings, long input slit:
dr =
s
2
2Rl
2
2
whence s = A
2R dr.
where A is dened in Eq. (9.7) and s is the length of the longest slit that can be used.
It must be emphasised that these computations are based solely on geometry,
and the effective resolution, d, in a concave grating spectrograph is obtained in
each case from the usual formula d = (a,m) cos r dr.
9.5 Practical details of design
There are some peculiarities of vacuum grating spectrography which are worthy of
particular mention, beyond what is discussed in Chapter 14.
There are various ways of constructing vacuum spectrographs and some provide
more practical difculties than others. It is a sine qua non in vacuum technology
that you have as few vacuum seals as possible, particularly if they are to be broken
frequently.
So far as possible, remote operation of any interior moving parts by electric
motor or magnetic coupling is to be preferred to a rotating vacuum seal. Low-
voltage operation is essential, especially if low-pressure gases are to be tolerated
or even required perhaps, in absorption spectrography.
15
Coma itself depends on the coefcient of nl
2
in the Beutler eikonal, and as l is ignored, it does not appear here.
9.5 Practical details of design 99
source, detector etc.
to cold trap and diffusion pump
Figure 9.8 Apossible arrangement for an Eagle mounting, with a separate vacuum
chamber which can be removed for adjustments to be made to the optical system.
All controls are led to the optical table through the end plate via vacuum seals
which in general are not rotating and therefore leak-proof.
It is imperative that the instrument be designed so that it can be adjusted and
focused in visible light. The awful alternative is to reduce the pressure to a vacuum
each time a focus adjustment is made.
The author offers this advice from his own experience.
Build a monochromator rather than a spectrograph. Make it as small as possible,
compatible with the research end in view. Choose the Eagle mounting, with a
1-metre radius grating with 6000 rulings/cm. Design for 10 mm-long slits centred
15 mmabove andbelowthe plane of symmetry. The defocus andcoma are negligible
out to 5000
A. The resolving power is of the order of 5 10
4
. Build the instrument
as shown schematically in Fig. 9.8, on a horizontal table cantilevered out from a
vertical face-plate which itself is carried on a suitable structure. Allow for a grating
adjustment and traverse for a maximum wavelength of at least 5461
A to permit
visual adjustment of the optical alignment and focus. The table should have large
holes for faster pumping. The vacuum chamber should be a simple cylinder able
to run its whole length on horizontal rollers, wheels or linear bearings to make a
single large-diameter vacuum joint at the face-plate. The seal may then be like that
of a simple evaporator plant. This may be expensive, but saves much time in the
long run.
Pass all controls through the face-plate in seals which need not normally be
demountable. The face-plate also holds the slits, fore-optics and detectors. The
lower half of the face-plate below the optical table holds the vacuum pump
port, and the pump should have the largest pumping speed that the budget will bear.
100 The concave grating spectrograph
It is unlikely that a vacuum lower than 10
6
mm Hg will be required, so that a
turbo-molecular pump is probably unnecessary. Nevertheless for a 1-metre concave
grating spectrograph in a 500-litre stainless steel chamber, a diffusion pump with at
least 1520 cm throat diameter should be included in the design. A cold-trap (LN
2
)
must be included.
10
The interference spectrograph
Interferometers were originally conceived as devices for measuring the wavelength
of light or the precise measurement of small distances, and with spatial resolution
and localised fringes,
1
contour maps of reecting surfaces could be made using
Fizeau fringes or Tolansky fringes. They also served as very high resolution spec-
trographs with which the hyperne structure of spectral lines could be studied. It
was only later that their very high efciency was noted and that they could be used
with advantage at moderate resolution for the spectral examination of very feeble
sources of radiation such as chemical phosphorescence, the night airglow and the
zodiacal light.
10.1 The phase angle
There is an abstract angle, the phase angle, dened by
= (2,)2nd cos . (10.1)
This is the most important general concept in interferometry, and describes all the
varieties of interferometer. The phase angle comprises four variables: (1) the path
difference, 2d; (2) the angle of incidence, ; (3) the incident wavelength, ; and
(4) the refractive index in the gap, n. The product 2nd represents the optical path
difference between interfering wavefronts and (2,)2nd is the consequent phase
difference.
Holding two of these four quantities constant and using a third as independent
variable, the behaviour of the fourth describes one of a large number of possible
fringe patterns. These include Fizeau fringes, constant inclination fringes, Mach
Zehnder fringes, EdserButler fringes, FabryPerot fringes, Tolansky fringes and
1
There is a myth that in localised fringes the interference takes place in the gap between two reecting surfaces
rather than at innity as in fringes of constant inclination. The interference of course takes place everywhere
in space and in particular at the detector, in the retina of the eye or on the plate of the camera.
101
102 The interference spectrograph
so on. Six types of interferometer result. So far as spectroscopy is concerned, is
the dependent variable and can be made to depend on the variation of . d or n.
Interferometers for spectroscopy can be divided into two categories: spectrom-
eters and spectrographs. In the former category there is a further subdivision into
Michelson Fourier-transform spectrometers and etalon scanning monochromators,
and in the latter category the FabryPerot spectrograph is the sole survivor of a
century of experiment. Other devices such as Michelsons echelon and Lummers
plate, although they yielded very high resolution, were difcult to make, inconve-
nient to use and had far lower etendue than the etalon spectrograph. They are all
now museum pieces.
10.2 The FabryPerot etalon spectrograph
The FabryPerot etalon was rst described by Charles Fabry and Alfred Perot
in 1899.
2
It comprises two at transparent plates worked optically so that the
gap between them can be held constant over the whole surface to an accuracy
of better than ,120. The constant gap is the important aspect: the surfaces of
the individual plates need not be at to better than ,4.
3
When the device is
dismountable, the angular orientation of the two plates may be important and is
then indicated by a pair of arrows inscribed on the sides to ensure correct alignment.
Etalons are oftenmanufacturedandassembledas sealedunitarydevices withthe gap
permanently xed so that no adjustment is required or possible, and these, dedicated
to a particular investigation, are to be preferred when constructing a FabryPerot
spectrograph.
The obvious advantage of a dismountable etalon is that the gap can be changed
for different purposes and when this is needed it is desirable that there be a servo-
stabilising mechanismfor holding the gap parallel and constant during observations.
10.3 FabryPerot theory
The theory of the FabryPerot etalon is to be found in every textbook on physical
optics. The transmitted intensity is given by the Airy formula:
I () =
I (0)
1 +
4r
(1r)
2
sin
2
_
2
_. (10.2)
where r is the reection coefcient for intensity at the reecting surfaces and, as
before, = (2,)2nd cos . For values of close to zero, sin(,2) (,2) and
2
C. Fabry & A. Perot, Ann. Chim. Phys., 16 (1899), 115.
3
Although the modern tendency is to make the plates individually so that they can be assembled without regard
to orientation.
10.3 FabryPerot theory 103
0 2
Figure 10.1 The Airy function of Eq. (10.2), illustrating the free spectral range
L and the resolution, .
the Airy function can be approximated by
A() = I (0)
U
2
U
2
+
2
.
where U
2
= (1 r)
2
,r.
This is the equation of a Lorentz curve, and the Airy function (Fig. 10.1), which is
periodic, may be regarded as the convolution of a Lorentz curve with a Dirac comb
of period 2. As we shall see in a moment, this Lorentz curve is the instrumental
prole of the FabryPerot spectrometer and the number of such proles that can be
tted into one order of interference is the ratio of the FWHM to the order spacing.
The FWHM, expressed as an angle, is 2(1 r),
r
2(1 r)
=
r
1 r
3
1 r
.
This quantity F is called the nesse of the interferometer and is a measure of
the number of resolved spectral elements in one order, that is, in one free spectral
range (FSR). The nesse is affected by the quality of the reecting coatings, by
the wavelength to be examined and by the atness of the gap. In addition there is
a small effect caused by the polarisation of the light, as the effective gap width for
oblique incidence at the edge of the fringe pattern that is depends on the state of
polarisation, that is, whether the electric vector is radial or tangential. The effect,
which vanishes on the optic axis, in the end controls the outer ring diameters and
limits the available nesse.
For purely practical reasons, larger etalons of 50 mm diameter and above are
likely to have a lower nesse than smaller ones of less than 25 mmaperture. Typical
values in practice for a 50 mm etalon are F = 3545.
104 The interference spectrograph
The free spectral range L is the range of wavelengths that can be tted between
two orders of the same wavelength. The gap width, d = n,2, is also equal to
(n 1)( +L),2, so that L = ,n =
2
,2d.
The resolution, the smallest range of wavelengths, , that can be observed as
two distinct lines, is equal to the FWHM of the Airy function maximum, that is to
L,F, and can be written in terms of wavelength as
=
nF
=
2
2dF
=
2
(1 r)
6d
and r, the reection coefcient of the plates, is usually in the region of 0.93.
4
10.4 The FabryPerot monochromator
10.4.1 Scanning in wavelength
This is done by changing the optical path between the two plates.
If the etalon is followed by a focusing lens the FabryPerot fringes from
monochromatic light appear at the focus as ne concentric circles. An annular
hole might be made in the screen with its inner and outer edges at the half-power
points of a ring, and its diameters come from the Airy formula (Eq. 10.1) with the
usual approximation cos = 1
2
,2. At the maximum of intensity,
2
2
= 1
n
2d
.
and at the half-power points,
2
1
2
= 1
n( ,2)
2d
.
2
2
2
= 1
n( +,2)
2d
.
whence
2
1
2
2
2
=
n
2d
=
_
n
2d
__
_
=
cos
R
.
It is usual but not universal to let the inner diameter of the annulus go to zero
so that the wavelength-dening stop is a circle. Then
2
= 0 and cos
1
1 and the
angular diameter, , of the circular stop is given by
2
8
=
1
R
. (10.3)
4
With multilayer reecting coatings this high reection coefcient holds good only over a hundred angstroms or
so, and must be specied when ordering the plates.
10.4 The FabryPerot monochromator 105
FabryPerot talon
chief-ray collimator
Jacquinot stop
and detector
interference filter
as order-sorter
Figure 10.2 The classical mounting for a FabryPerot spectrometer. The incoming
light, which must be from an extended source, is focused on to the Jacquinot stop
but rst passes through a short section where the chief-rays are collimated before
passing through the interference lter which is the order-sorter in this type of
arrangement. Wavelength scanning is by one of various methods of changing the
optical path between the plates. Fore-optics, usually in the formof an astronomical
telescope, could be added if the source was of small extent or if ne detail in the
source was to be examined.
These relationships between the various quantities . d. r and can be used to
dene the important quantities associated with the FabryPerot spectrometer and
spectrograph.
(a) The free spectral range (FSR) = L. = ,n. = 2d,n
2
. =
2
,2d.
(b) The nesse, F = L,. =
2
p
2
n
2
,1.92
.
These two curves can not be overlaid and matched exactly but a reasonable approx-
imation is that they cross at the 1,e-point. Then,
n,1.386 = 1,b
and the FWHM of the FabryPerot instrumental prole should be that of the ob-
served line FWHM divided by 1.386.
This is still approximate, and in practice much the same efciency is obtained
provided that the etalon line width is about the same as the observed line or some-
what less (rather than more). It is impossible to be more precise as the line shapes
being observed are not always conveniently Gaussian and the spectrummay contain
lines of different half-widths.
10.10 The crossed FabryPerot spectrograph
A technique for improving the resolution of emission line spectra in a spectrograph
by an order of magnitude was developed in the early days by imaging the rings
of a FabryPerot etalon on to the entry slit of a Littrow spectrograph, the centre
of the ring system lying on the line of the slit. The slit was then opened wide,
subject to not allowing line images at different wavelengths to overlap and the
spectrogram revealed such things as pressure broadening, Zeeman splitting and
hyperne structure simultaneously at a large number of wavelengths for example
in a molecular band spectrum. The technique depends essentially on having small or
zero astigmatismin the grating or prismspectrograph and is therefore most suitable
for the Littrow conguration. The image of the ring system should contain both
parts of the innermost ring in order that the centre of the system be located.
12
The
same criteria apply as for a normal FabryPerot spectrograph: the number of rings
should be at least two, and with a CCD detector the instrumental prole should be
at least 5 pixels wide at FWHM if the line prole is required.
The technique is not yet moribund and, if applied, the same rules apply for
imaging the source with the proviso that if the source has structure of its own an
oxy-acetylene ame for example the magnication and defocus should be such
that the local variations do not affect the interferogram.
12
The alternative is a tedious amount of programming to linearise the square-law wavelength dispersion along
the slit direction.
11
The multiplex spectrometer
The theory and practice of Michelson Fourier-transform spectrometry is strictly
speaking beyond the purview of this book, but it is in such universal use that a
brief description is warranted in order that a reasoned choice can be made when
contemplating infra-red spectroscopy.
11.1 The principles of Fourier spectrometry
When a monochromatic beam of light passes through an ideal Michelson inter-
ferometer with no absorption and exactly 50% reection at the beam-splitter, the
intensity of the beam which emerges depends upon the path difference between the
two recombined components.
Suppose that the incident beam has wavenumber and complex amplitude A.
Suppose, for simplicity, that the phase is zero at the beam-splitter. After passing
the beam-splitter the two emerging amplitudes will be
A
2
e
i
1
and
A
2
e
i
2
.
where
1
and
2
are the phase changes on reection and transmission.
After passing through the two arms of the interferometer the two beams recom-
bine at the beam-splitter. Before recombination the two amplitudes are
1
A
2
e
i
1
e
2id
1
and
A
2
e
i
2
e
2id
2
.
and after recombination, when each beam has undergone one transmission and one
reection, the resultant amplitude transmitted is
A
2
e
i(
1
+
2
)
_
e
2id
1
+e
2id
2
_
.
1
Bearing in mind the precept that phase change = 2 path change.
114
11.1 The principles of Fourier spectrometry 115
This quantity, multiplied by its complex conjugate gives an expression for the
transmitted intensity:
I (L) =
AA
2
[1 +cos(2L)].
where L = 2(d
1
d
2
) is the path difference between the two beams.
The transmitted intensity can be seen to vary sinusoidally with L. The intensity
and the frequency of this wave depend on the rate at which the path difference
changes and on the intensity and wavenumber of the incident radiation.
In practice the input has a continuous spectrum S() and if the intensity in the
range + is S() d, it will contribute an innitesimal power dI (L) =
(S() d,2)[1 +cos(2L)] to the output beam.
The total transmitted intensity will then be
I (L) =
1
2
_
=0
S() d[1 +cos 2L]
=
1
2
_
=0
S() d +
1
2
_
=0
S() cos 2Ld
and the second term is the Fourier cosine transform of the spectral power density,
S(), as a function of L.
An alternative way of looking at Fourier spectrometry is to regard the signal
strength at the detector as the time average of the product of the signal and its
delayed counterpart:
I (t
) = [ f (t ) + f (t +t
)][ f
(t ) + f
(t +t
)]
= f (t ) f
(t ) + f (t +t
) f
(t +t
) +2| f (t ) f (t +t
)|.
where t
= L,c and L is the path difference. The third term is the autocorrelation
function of f (t ) and by the WienerKhinchine theorem its Fourier transform is the
SPD, or the spectrum of the signal.
The basis of Fourier-transform spectroscopy is that the output signal from the
interferometer is recorded as a function of L (the interferogram) and the spectrum
is obtained from this by a Fourier transform, usually in a digital computer.
The sampling theoremis used, and samples of the transmitted intensity are taken
at equal intervals of path difference. The number of samples may vary from 1024
to 10
6
, depending on the investigation being followed. This set of numbers takes
the place of the interferogram and is sent as a one-dimensional array or vector to
the computer.
The art of multiplex spectrometry lies in three directions.
Firstly, the optical alignment must be maintained to interferometric accuracy
during the whole displacement phase of several centimetres or more while the
interferogram is being recorded.
116 The multiplex spectrometer
(a)
beam-splitter
R1
R2
output port
Detector
(c)
beam-splitter
R1
R2
output port
input port
Detector
Detector
(b)
(d)
beam-splitter
R1
R1
R2
R2
output port
Detector
input port input port
input port
output port
beam-splitter
Figure 11.1 Variations on a theme. Four different types of Michelson Fourier
spectrometer which have found commercial employment in different wavelength
regions. (a) The classical Michelson interferometer; (b) the same, with cube-corner
reectors for automatic alignment maintenance; (c) the cats-eye autocollima-
tor reector which returns any incidence ray anti-parallel to its input direction;
(d) the tilting Michelson, in which the path difference is changed without
physically moving the reectors.
Secondly, the samples must be collected at exactly equal intervals of path dif-
ference. The process is analogous to ruling a new diffraction grating every time
a spectrum is observed, and the precision required is the same as that of a ruling
engine,
2
although the time devoted to each spectrum is considerably less.
Thirdly, an efcient way of doing the Fourier transformmechanically is required.
Much research was devoted to these three problems in the 1960s and 1970s and
various solutions were found. In far IR spectroscopy, simple mechanical displace-
ment of a reector is adequate. At shorter wavelengths, cats-eye reectors and
cube-corner reectors allow linear translation of the Michelson mirrors without
misalignment, and the tilting Michelson changes the path difference by rotating
a jig containing the beam-splitter and a rigorously parallel mirror about a hinge.
Examples of successful applications of these ideas in manufactured Fourier spec-
trometers are shown in Fig. 11.1.
If samples of the interferogram must be taken when the reectors are stationary,
piezo-electric transducers can be made to oppose a smooth steady change in path
2
The late Dr H. J. J. Braddick once remarked that designing a grating-ruling engine was the same problem as
designing a screw-cutting lathe to be made of rubber.
11.2 The multiplex advantage 117
difference with servo-stabilisation of both the alignment and path difference for
each sample. Wavelength-stabilised lasers offer a standard of step length either
by ensuring that the step length is a constant number of half-wavelength fringes
or, by an optical screw
3
or ratchet device, a constant fraction (often vulgar) of a
fringe.
The Fourier transformation of the interferogram is by digital computer, usually
a dedicated laptop, using the fast Fourier transform or FFT, one of a number of
routines derived from the CooleyTukey algorithm of 1964.
4
The reasonfor doingspectroscopyinthis wayis the enormous gaininsignal/noise
ratio over grating spectroscopy which is obtained when the dominant noise source
is the detector noise. This follows from P. J. Fellgetts discovery in the 1950s of
the multiplex advantage.
5
This merits some discussion as it is typical of the way in
which detection methods are assessed.
11.2 The multiplex advantage
Consider a grating spectrometer and a multiplex spectrometer of the same etendue
E and observing the same spectral source S(), in each case to be explored over a
bandwidth
0
. The spectrum is everywhere dense and the total power is S.
Suppose each instrument dwells at each sample point for a time T. Suppose there
are to be N resolved elements in the spectrum.
In the monochromator the ux to the detector is
F
g
=
ES
0
T
N
.
which is the total signal energy in each sample. The detector noise is proportional
to
T
=
ES
0
T
Nk
.
In the multiplex spectrometer the total signal in each sample is much greater:
F
m
=
ES
0
T
2
.
and again the noise energy at each sample is k
T.
3
Y. P. Elsworth & J. F. James, J. Phys. E: Sci. Instrum., 6 (1973), 1134.
4
J. W. Cooley and J. W. Tukey, Mathematics of Computation, 19 (1965), 297. See also: E. Oran Brigham, The
Fast Fourier Transform (Prentice-Hall, 1974).
5
P. Fellgett, J. de Phys., 19 (1958), 187, 237.
118 The multiplex spectrometer
In this case the samples are used to compute the energy in each element of the
spectrum by
S(m) =
N1
n=0
_
ES
0
T
2
_
cos
2mn
N
.
and since cosines are as often negative as positive the total energy in the computed
sample will be
F
m
=
ES
0
T
2
.
The noise samples, also negative as often as positive will average to 0 but the
RMS value will be
_
N1
n=0
k
2
T cos
2
2mn
N
=
_
Nk
2
T
2
since cos
2
=
1
2
.
and the signal/noise ratio is then
S,N =
ES
0
2k
_
T
N
.
which is greater than that of the monochromator by a factor
N,2.
Asimilar calculation for the same spectrometers with photoelectron shot-noise as
the principal source of noise shows that there is no such multiplex advantage and that
the S,N ratio is apparently the same for each type of instrument. The calculation
however does not show that there is in fact a multiplex disadvantage since the
shot-noise on any one strong emission line is spread throughout the interferogram
and hence throughout the spectrum, and it effectively drowns any faint emission
lines which may be present. Similarly, the shot-noise from a continuous emission
spectrum lls in or obscures any narrow shallow absorption lines which may be
present.
The above calculations are not exact since source strengths may vary from point
to point in the spectrum as well as from moment to moment during the observation,
and the multiplex advantage is usually present only in order of magnitude. When
the number of samples (or the resolving power) is high, 100 000 for example,
the multiplex advantage is 300, the time taken to obtain the spectrum is reduced
from 5 minutes to 1 second, and in the infra-red beyond 1 m, where sensitive area
detectors still await invention, the Michelson Fourier spectrometer is the instrument
of choice.
Fourier spectroscopy is most suited to absorption spectroscopy where (usually)
vapour samples are observed with a constant, well-regulated infra-red source
11.2 The multiplex advantage 119
providing the radiation power, in which case it has a marked advantage over grat-
ing spectroscopy both in accuracy and time. It is only to be undertaken on variable
sources whenthe frequencyspectrum
6
of the variationdoes not include the sampling
frequency or, more prudently, when a separate detector monitors the moment-to-
moment intensity uctuations of the source. But beware: this latter precaution does
not work if the spectral content varies frommoment to moment. It is the sort of thing
which happens in atmospheric research for example, by the selective absorption of
sunlight in the water vapour of occasional passing clouds.
To summarize: none of the techniques involved in Fourier-multiplex spectrom-
etry is to be undertaken lightly and for this reason the description of Fourier spec-
trometers is curtailed here.
6
The frequency spectrum is not to be confused with the infra-red spectrum: it refers to the frequencies (measured
in Hz) present in the moment-to-moment variation of the source strength.
12
Detectors
In spectrography, where spatial resolution is required, there is a clear subdivision
in detection technology between photochemical and photoelectric methods. So far
as chemical methods are concerned they possess fewadvantages over photoelectric
methods except possibly for observations at remote sites where a reliable supply
of electricity is uncertain or unavailable. They comprise the reduction of silver
halide salts to colloidal silver in a solid emulsion coated on to glass or on to a
transparent exible substrate.
1
The detective quantum efciency (DQE) of silver
halide crystals is in the region of 10
4
10
3
, the dynamic range is in the region
of 100:1 and as photodetectors they are severely non-linear in response, especially
where long recording times greater than 100 s are involved. ISOratings, usually
quoted in the range of 12 for high-resolution ne-grain emulsions, to 2000 at the
higher end of the range for severely coarse high- high-sensitivity emulsions,
2
fall rapidly at exposures longer than 1030 s. The effect, known as reciprocity
failure, is the notion that the blackening produced depends solely on the product
of intensity and time and is a necessary assumption if spectrophotometry is to be
done.
This sensitivity range may be compared with a liquid-nitrogen cooled CCD
detector with a DQE in the region of 0.5 and strict linearity over a dynamic range
of 25 000:1, a sensitivity broadly equivalent to an ISO of 3 10
6
.
3
Liquid-nitrogen
cooled CCD detectors require far more careful screening from light than silver
halide emulsions. An ordinary photograph with a cooled CCD camera and an F,2
lens in a typical photographers darkroom will give a clear, detailed picture of the
interior in a few seconds exposure.
1
Gold salts can also be made to follow the same photoselective reduction process.
2
Sometimes known as soot-and-whitewash emulsions.
3
No strict equivalence is possible because of the non-linearities and different dynamic ranges involved, and this
estimate is made by comparing the exposures at a shutter speed of 0.02 s needed for photographs of similar
quality.
120
12.1 Silver halide photography 121
12.1 Silver halide photography
Emulsions on glass plates have become comparatively rare in spectrography and it is
now the custom to rely on sheet lm. Spectrographic plates have traditionally been
large, 75 mm 250 mm being typical, and the emulsions coating them have been
specialised to cope with the peculiar requirements of spectrography. Chief among
these is the need for uniform sensitivity across the whole spectral range covered by
the instrument and generally instruments which deal chiey with infra-red radiation
are not required to cope simultaneously with the visible and UV regions.
12.1.1 Infra-red emulsions
Emulsions are available from photographic material manufacturers which will
record radiation to 1.2 m in the infra-red. These have a comparatively coarse
grain so that a spectrograph with high dispersion is needed for high-resolution
spectrography. The roll lm, sheet lm or plates must be kept in a refrigerator until
shortly before they are to be exposed, to avoid themfogging fromthermal radiation.
Condensation is then a problem especially in tropical regions, and the installation
of a drying agent such as silica gel in the spectrograph is advisable well before the
plates are introduced. Appropriate development conditions are required afterwards
and the manufacturers instructions must be followed closely for satisfactory re-
sults. If any sort of photometric precision is intended, the technique is an exacting
and unforgiving one.
12.1.2 Visible and near UV emulsions
For the visible region and near UV, so-called astrographic emulsions are available
which were developed initially for astronomical research and which possess rea-
sonably ne grain and uniform sensitivity throughout the visible region. The point
which must be emphasised here is that if spectrophotometry is to be attempted,
it is important that all the plates or sheet lms come from the same batch, that
they be developed and xed under rigidly identical conditions of time, temperature
and agitation, and preferably with the same batch and conditions of mixing of the
developer. De-ionised water is recommended but is not of overriding importance.
4
This is provided that the water is free from hostile cations, i.e. that it has a pH
higher than 7.
The bugbear of reciprocity failure at long exposure times has been the subject of
a great deal of experiment, chiey with the use of reducing agents to remove any
4
The author was advised by a manufacturer that, on an eclipse expedition to a tropical island (Manuae, Cook
Islands, 30 May 1965), their plates could if necessary be washed in sea-water, the nal clearance washing being
deferred until a later date.
122 Detectors
trace of free oxygen. Hydrogen bathing and soaking in a solution of ammonia have
been claimed to be effective, but following the development of the CCD, they are
no longer worth considering seriously.
With experience, sufcient uniformity can be reached with the silver halides that
spectrophotometry to 1% is possible. This seems to be a statistical limit, such that
there is little improvement on it even with large amounts of repetitive data.
Emulsions can be sensitised to the far UV by suitable phosphorescent materials
such as uorescein, and sensitised emulsions are available commercially. For the
extreme UV (XUV) below 2000
A Ilford Q plates are to be preferred. These
are manufactured, chiey for the nuclear industry, without the clear top-coating
of emulsion, so that the silver halide crystals actually reach the surface. Such
emulsions, on thin glass plates, can be used in vacuum spectrographs and can be
bent on a mandrel or in a plateholder to t a 500 mm radius Rowland circle. The
emulsion surface is sensitive to touch and so the plates require careful handling
in the darkroom when they are loaded. As usual with silver halide emulsions,
different grades are available with different crystal sizes and corresponding different
sensitivities.
Included in the paraphernalia of traditional silver halide spectrophotometry must
be a plate measuring machine or travelling microscope, and a densitometer is re-
quired if accurate photometry is to be done. Calibration exposures must also be
made, although the method of the common calibration curve is a useful alterna-
tive.
5
12.2 Elementary electronic detectors
12.2.1 Photocells and photomultipliers
In the age of the monochromator the photomultiplier led the way for 50 years. It
is a vacuum tube and comprises typically a photocathode deposited on an end-
window a thin lm of silver-silver oxide overcoated with caesium, antimony
or other substances with a low work-function, to act as a photoelectron emitting
surface, and a succession of dynodes: electrodes with a slatted structure similar
to a Venetian blind and held at successively increasing positive voltages to attract
electrons sufciently strongly to provoke secondary emission and thus to act as
current ampliers. A single photoelectron expelled from the photocathode will
provoke a cascade through anything up to 13 successive dynodes until a pulse of
10
7
electrons arrives at the anode. The pulse may have a duration of substantially
less than a nanosecond
6
allowing high pulse rates to be collected. The pulses can
5
See Appendix 4.
6
Bearing in mind that the speed of light is 1 foot per nanosecond, this is not so surprising.
12.3 Detectors with spatial resolution 123
be collected as a photocurrent or, especially if the radiation rate is very low, can
be amplied and counted as individual pulses in a digital recording system. Either
way, the DQE of the device is in the region of 0.150.35. Photomultipliers must be
cooled, usually to 80
.
Briey, the Schmidt camera uses the condition for zero coma in a spherical
mirror given by Eq. (4.8): the pupil must be at z = 2 f , that is, at the centre
of curvature of the mirror. At this pupil a zero-power corrector plate gured to a
fourth-order surface curvature is placed to remove the spherical aberration, which
at this high numerical aperture would be severe. Symmetry precludes astigmatism
(Eq. 4.9) but eld curvature must be accepted as the price to pay for removal
of the other aberrations at such a high numerical aperture, and herein lies the
problem with spectrography. In a large Schmidt camera a photographic plate can
be bent over a mandrel to accommodate the eld curvature and in a small one the
emulsion can be coated on to a spherical glass substrate; or a eld-attening optical
system can be placed in the appropriate position. But a CCD detector requires such
a bulk of electronics and refrigeration that a totally unacceptable obstruction to
the beam would be incurred and it is this which makes the Schmidt conguration
impracticable for CCDspectrography. Ahigh numerical aperture depends therefore
on a focal reducer followed by a good quality camera lens of low focal ratio.
10
13.5 Scattered light and bafing
The purpose of including bafing in the design is the reduction of scattered light.
10
For most spectrographic purposes a standard double-Gauss 35 mmF,2 camera lens, of which there are probably
a hundred designs on the market, is adequate.
13.5 Scattered light and bafing 135
13.5.1 Causes of scattering
Scattering comes chiey fromuneven edges to the input slit, fromdust and scratches
on any of the optical surfaces, from imperfections in the grating rulings and from
light coming through the entry slit which reaches any of the walls of the interior of
the spectrograph. Bafes are simply opaque materials in the form of walls or tubes
which intercept and absorb unwanted light. They may be inside the spectrograph
or outside or both.
Decient optics are of course a matter of quality control by the manufacturer,
and large, thick lenses of glass or quartz are more likely to have striae, stones, sleaks
and scratches than small ones. Mirrors are superior in this respect but still must be
manufactured, coated and chosen with care. Dust particles on optical surfaces are a
matter of cleanliness and accessibility. Gratings, by the nature of their construction,
scatter more than most optical elements and for this reason prisms are still to be
considered when scattered light is more of an obstacle to research than lack of high
resolution.
13.5.2 Scattering from edges
The edges of the entry slit were mentioned above. Any edge which can be seen
from the grating is suspect, because of the possibility of Fresnel diffraction effects.
The sharp edge of a lens-hood is a particular culprit even in a camera, and may
scatter enough light on to the lens surface to be re-scattered into the interior. For
true bafing, several such edges must shield the lens, each one shielding from the
last: alternatively, a trumpet-shaped bafe will show a smooth horizon rather than
a sharp edge to the input optics. The problem is particularly acute in space-borne
spectrographs where it is desired to acquire stellar spectra from anywhere within
90
of the Sun.
13.5.3 External bafes
The principle to followwhen judging where to incorporate bafes is that the grating
must see nothing except light from the external condenser, and that light entering
through the slit can go nowhere except to the grating.
An external bafe in addition to the internal bafe may help to achieve this.
There is an image of the grating formed by the collimating mirror and the input
Fabry lens. It is situated before the entry-slit/Fabry-lens assembly. There may well
be a condensing lens there, but whether there is or no, this is the place to put a real
stop, its aperture cut to the shape of the grating, inserted in the optical train and
136 Auxiliary optics
tilted to match the image of the tilted grating. The tilt is not large, bearing in mind
that the longitudinal demagnication is the square of the lateral demagnication of
the grating image which it covers. It must follow then that light which enters the
spectrograph through this stop, unless scattered by the input slit, must fall on to the
grating. Similarly, if there is a re-imaging of the spectrum, for example through a
focal reducer, there may be another opportunity to insert a made-to-measure stop
at the image of the grating on the output side, although this will coincide with the
camera entry pupil.
13.6 Absorption cells
Absorption spectra of liquids and solids require only the simplest of absorption
cells, typically with paths from a few micrometres to a few centimetres, and such
cells are normally placed just before the entry slit, sometimes covering only one
half of it so that comparison spectra come through the other half. If this method is
adopted, then the designer must be careful to see that the whole slit length receives
identical illumination and it is the light source and not the condenser which must
be imaged on to the grating by the Fabry lens. Strictly speaking, the light should
be collimated, but in practice a low numerical aperture, equivalent to F,12, is
adequate to give the accuracy normally required. The marginal rays in the cell
differ in length by less than one part in a thousand from the paraxial ray.
13.6.1 White cells
Serious problems only begin to arise when the absorption is very low so that long
paths are required.
In the White cell,
11
long path absorption is achieved through the periscopic
principle, in which a beam of light is sent down a tube with relays of lenses which
achieve a useful eldof viewwhile conningthe light tothe interior. Inanabsorption
cell the lenses are replaced by mirrors, and absorption path-lengths of tens of metres
can be achieved in a 1-metre long tube.
Three concave spherical mirrors are employed, all of the same radius of curvature
(Fig. 13.4). Light is admitted through an entry slit and is reected off mirror 1 to a
focus at a point on the surface of mirror 2. From mirror 2 it is returned to the centre
of mirror 3 and thence to a focus at another point on mirror 2. The process continues
until a nal focus is reached outside mirror 2, whence it goes to the spectrograph.
Figure 13.4 shows the chief-ray as it passes through the cell with seven reections.
11
J. V. White, J. Opt. Soc. Am., 32 (1942), 285.
13.7 Fibre optical input 137
1
6
5
2
7
4
3
8
entry slit
exit slit
Figure 13.4 The White cell. This is a device for measuring feeble absorption coef-
cients where long path-lengths are required. The device uses multiple reections
between three mirrors and depending on the reectivity of the mirrors, the number
of reections may be as high as 30 in the visible and several hundred in the FIR
where reection coefcients are higher.
The actual number will depend on the tilt of the two smaller mirrors, which may
be adjustable.
The number of reections available depends on the reection coefcients of
the mirror coatings. An untarnished silver coating has a reection coefcient
in the region of 0.92 and transmits 10% of the incident light after 27 reec-
tions. Higher reection coefcients and longer paths are possible over restricted
wavelength ranges with multilayer coatings. Gold coatings in the infra-red, with
reection coefcients in the region of 0.980.99, will allow several hundred
reections.
13.7 Fibre optical input
The salient facts which concern us in spectrography are that multiple bre bundles
are used rather than single bres, and that a circular luminous object, such as a star
image, can be converted to a line image suitable for the input slit of a spectrograph.
A single bre in a bundle may have a diameter anywhere from 1 m to 25 m,
10 m being typical, and the output beam from the bundle emerges at a focal ratio
in the region of F,3. The illumination is Gaussian rather than uniform, but should
be converted by a lens to F,12 so that all the emerging light will be accepted
by the collimator. The apparent slit width will then be four times greater than the
diameter of the bres at the output end and the apparent slit length will also be four
138 Auxiliary optics
times greater than the length of the output end of the bre. Alow-power microscope
objective may be a suitable converter, focusing the bre-bundle end on to the entry
slit. A Fabry lens at the slit will be needed to image the converter pupil on to the
grating. Alternatively a virtual magnied image of the bre end may act as the entry
slit, in which case aplanatic lenses are indicated.
14
Optical design
Le mieux est lennemi du bien.
(The best is the enemy of the good.)
Voltaire
14.1 First steps
Once the task for the spectrograph has been dened, a suitable type may be chosen
and a catalogue search made for a possible manufactured model. Many factors
affect the decision. The rst of course is tness for purpose. Time and cost of
manufacture may be factors and the facilities available for local or in-house design
and construction are primary considerations. A decision on whether to buy or to
design and build a dedicated instrument must rest on such an appreciation.
It is a mistake when doing fundamental or academic research to construct a
more elaborate, high-performance or expensive instrument than the immediate task
demands, possibly in the hope or expectation that it may prove useful for some other
investigation at some later date. In the authors experience this is almost never the
case and the chief result is delay and unnecessary expense. In the long term the
instrument is a sad relic, cannibalised of its optical components and left to decay
in the attic with its ingenious mechanisms and precision micrometers.
14.2 Initial layout
Once the spectrograph type has been chosen and its main parameters such as wave-
length range, resolution, type of detector, number of resolved elements, etendue etc.
have been decided, a sketch can be made, and the traditional back-of-an-envelope is
as good a place as any for this. Manufactured mirrors and lenses must be considered
rst before taking a decision to have optical parts made, and it may well be dis-
covered that the design can be adapted to use bought-in elements. Almost certainly
139
140 Optical design
the optical components will come from a makers catalogue. Rarely a bespoke item
must be ordered from a specialist manufacturer and if this is so, optical manufac-
turers are usually happy to let the designer have lists of their tool radii, of suitable
glasses at their disposal, with their optical properties and the costs of manufacture.
Catalogues of optical, opto-mechanical and opto-electronic components must be
available to the designer, quite probably from the Internet.
14.3 The drawing board
The next stage is to compute the Gaussian optics and to make the outline optical
drawing, and this will determine the sizes and positions of the optical components.
When making the rst sketch it is both clearer and more instructive to lay
out all optical components in position as if they were transmitting light rather
than reecting it. Consider all the mirrors as if they were lenses and the grating
as if it were a transmission grating.
The x- and y-scales on the diagram need not be the same. The ray paths are not
affected and the positions of stops and pupils will be unaltered. A CAD (computer
aided design) program may lend some precision to the drawing, but the old ruler,
pencil and compasses is probably adequate and may well be more convenient at
this stage.
If the instrument is to be of the
CzernyTurner or EbertFastie type, this is
the time to choose input- and output-mirror off-axis angles for the chief-rays. It
is a simple matter to write a short computer program which, given the zero-coma
wavelength and the sumof the two off-axis angles, will compute, on the assumption
of parallel input and output chief-rays, the and angles, the grating tilt and the
meridional coma for rays on either side of the ZMC direction. If the instrument is
to have a Littrow conguration, the off-axis angle should be chosen so as to allow
adequate spacing for the various components with room for the mirror and grating
cells, auxiliary mirror mounts, slits and CCD/plateholders. A computer ray trace
will determine whether the aberrations are under adequate control. Assume at this
stage that all mirrors will have a thickness equal to one-sixth of the diameter, and
that all positive lenses will have a rim thickness of at least 2 mm. Grating blank
sizes are to be found in the manufacturers catalogue.
14.4 Computer ray tracing
This is the point to emphasise once again that one cannot design an optical
instrument on a computer with a ray-tracing program.
The design must be done rst and set up on the ray tracer for renements to be
made. Remember that the Seidel aberration calculations are not exact and that small
14.5 Renement of the optical design 141
changes in the dispositions of the optical surfaces can be made by the program to
minimise the aberrations, reducing them sometimes to zero.
A typical ray-tracing program invites the operator to insert the x-, y- and z-
coordinates of the vertex of each surface. The z-axis is normally the optic axis. It
then requires the pitch and tilt angles of the surface, its curvature and excentricity,
whether it is reecting or transmitting, whether it is a grating and if so the number
of rulings per millimetre. Refractive indices will generally follow each vertex,
referring to the material following.
Next, the x-, y- and z-coordinates of the launch points of each ray are required
together with the sines of the angles which each one makes with the z-axis in the
xz and yz planes, and the wavelength associated with the ray.
Bear in mind that the outer 3 mm at the rim of a mirror cannot be relied on, as
there is more often than not a turn-down as the result of the polishing action. If
the manufacturer is supplying a 150 mm diameter mirror, it is prudent to assume
that the aperture is 140 mmwhen doing the optical design. In any case the retaining
plate which holds the mirror in its cell may well have such an overlap at the edge.
The program will trace the ray through the system with an accuracy far beyond
anything that can be measured in practice. It can be asked to adjust any (or all)
of the input parameters until the ray arrives at a denite given goal, usually either
the x- and y-coordinates of a point on the last surface or, if the output is to be a
collimated beam ready for a bought-in camera lens, the sines of the angles at which
the ray leaves the last surface. A multiplicity of rays, 20 or 30 perhaps, can be
treated simultaneously.
14.5 Renement of the optical design
Thus, when the outline design appears satisfactory sufcient room left for baf-
es, detector optics and fore-optics unlikely to interfere with each other etc. the
optical train can be set up in the ray-tracing program as outlined above. This is the
design stage which formerly was done by hand using seven-gure logarithm and
trigonometrical tables, when an experienced ray tracer could trace a ray through
eight or nine surfaces in 20 minutes and an experienced lens designer could make
an intelligent guess as to which parameter needed changing next before repeating
the process. A modern ray-tracing program will trace 30 rays through 20 surfaces
in as long as it takes to press the start key on the computer keyboard.
There is a sufcient variety of these ray-tracing programs on the market that it is
pointless to particularise too much. No great elaboration is required here, as focal
ratios are unlikely to be belowF,8, the limit at which simple achromatic doublets
provide adequate eld and resolution for a spectrograph. One aims generally at a
spatial resolution in the neighbourhood of 35 mdepending on the type of detector
142 Optical design
and the resolving power required. A simple geometrical ray-tracing program such
as BEAM 3 from Stellar Software
1
is adequate although it is strictly geometrical
and does not include renements such as diffraction effects at the image points.
In the event that a manufactured camera lens is to be used for the camera,
2
then
the spectrograph designers aim may be for parallel ray bundles at each computed
wavelength at the nal output pupil of the spectrograph.
14.5.1 Initial alignment
Once the optical elements are put in their places, a single ray, the chief-ray, at
the central wavelength is launched from the centre of the entry slit to the vertex
of the rst (collimating) mirror or lens. The mirror is given the tilt required by the
outline design and the grating (or prism rst surface) position may be adjusted by
the program to receive the ray at its centre.
The grating tilt will already be known from the outline design and this will have
already been entered in the ray-tracing program. Alternatively, if the zero-coma
wavelength has been decided, the grating can be tilted by the program to reect the
chosen wavelength to the vertex of the next surface.
The position of the second mirror is then adjusted to receive the diffracted chief-
ray at its vertex and its tilt is then adjusted by the program to send the ray either
towards the detector centre or possibly to be made parallel to the input ray. In the
latter case the detector vertex may be adjusted by the program to receive the
chief-ray.
Exact positions are not set rmly at this stage. The distance from the entry slit
to the M1 mirror of a
CzernyTurner mounting is found by tracing parallel rays
backwards, starting from the corners of the grating and nding the place where the
marginal rays would meet to forma comatic image of a virtual point at innity. This
is where the entry slit must go, and the procedure takes care of the eld curvature
of the M1 mirror and avoids any focus defect. Rays launched from a point source
at the slit centre towards the M1 mirror then arrive at the grating with the proper
amount of coma.
Procedures like this set out the initial dispositions of the major components, and
ray bundles of six or eight marginal rays may be launched from the same point on
the entry slit and adjusted (again by the program, which can adjust the sines of the
launched rays) to arrive initially at eight equally spaced points around the margin
of the iris of the system usually the grating. These are then traced on through the
system so that the nal image quality can be assessed.
1
Stellar Software, PO Box 10183, Berkeley CA 94709, USA.
2
Recall the advice in Section 12.4 regarding radioactive elements in low focal-ratio camera lenses for CCD
cameras.
14.5 Renement of the optical design 143
This is the stage where such items as coma, astigmatism and eld curvature can
actually be measured (as opposed to being calculated) and necessary adjustments
made to the positions and tilts of the various components to ensure compliance with
the original design goals.
The process is a lengthy one and experiments on the computer are generally
needed to nd the optimum positions. The departures from the positions calculated
by Seidel theory will not be large a small percentage probably but they will be
signicant, and the nal aberrations may well be substantially less than the original
estimates.
14.5.2 Elaboration of the design
Once optimum positions have been found for imaging a point source on the entry
slit, the length of the slit can be added as a newcondition and the effects of spectrum
line curvature can be measured. It is at this stage that compromises will be needed
to secure the best image and to discover the spectral range over which the whole
spectrograph will operate satisfactorily.
When the process is complete and a satisfactory optical design has been reached,
small changes should be made in the ray-tracing program to the positions on the
optical layout, to determine the tolerances that may be allowed in positioning the
various components. This in turn determines the accuracy which must be required
in the workshop when the instrument is nally constructed. More importantly, it
may well reveal that the manufacturing tolerances are too tight and that the design
cannot be realised in practice. This sort of defect occurs, for example, when a ray
is incident on a refracting surface at a large angle, and where a small displacement
of the surface vertex along the optic axis would entail a large change in the height
of incidence and consequently the ruin of all that follows. Considerations like this
play a signicant part in the design of camera lenses for example.
14.5.3 Local minima
A warning is appropriate at this point a general warning in fact to everyone using
a ray-tracing program. The program varies the positions, tilts, curvatures etc. as
it searches for a minimum departure from the goals set by the programmer a
focus perhaps or parallelism of rays in a collimated beam. The initial xed points
of the program may be the launch points of the rays and the refractive indices
of the transmitting elements. The variables are the bendings, excentricities, vertex
positions, tilt and pitch of some or all of the optical components (some, taken froma
manufacturers catalogue, may be taken as xed, except in position). Vertices can be
ganged together so that a lens or a group of lenses can be moved together as one unit.
144 Optical design
It is inevitable, with such a multiplicity of variables, even when only two or three
are allowed to vary at one time, that many local minima will be found such that any
small change in any one of the variables will produce a worse result. In practice a
great deal of experiment will be needed to nd the ultimate best arrangement of
the components, and the nal result, which may sometimes be surprising, must be
tested with a large variety of possible ray-input parameters.
Metaphorically, the climber is climbing a mountain in a fog, and must be certain
that she has actually climbed the mountain and is not standing on the summit of
one of the foothills.
The design found by computing the Seidel aberrations will almost certainly
need minor corrections. The at-eld condition of Eq. (4.12) will not give an
exact position for the grating and a small movement of the grating or the M2
mirror, of several per cent of the M2 focal length, may be required to produce a
eld at enough for the CCD detector. Computer experiment is the answer, and
it is instructive to plot, for example, a rough graph of the way the eld curvature
actually changes with M2 position.
14.6 Requirements of a ray-tracing program
The obvious function of a ray-tracing program is to allow you to make small
variations in such variables as you choose and to re-compute the rays to see whether
the variation moves them towards or away from a goal which you have chosen.
Several variables may be chosen at one time and this is the chief cause of the
programs nding local minima.
There are other useful functions of the programbeyond this. It should be possible
to compute spot diagrams at the nal focus or at any other intermediate point in
order to check the contributions of the various aberrations. In a CCD spectrograph
a square or rectangular iris at the position of a pixel or a column of pixels should
measure the proportion of randomly launched rays which enter the target area, a
potent measure of the efciency of the optical design.
There are other, more subtle variables which can usefully be displayed. Apart
from drawing the ray diagram, an xy plotting program should be available to
make graphs of one variable against another. For instance the so-called Htan U
plot,
3
in which an oblique parallel bundle of rays at some angle with the optic axis
is traced and a graph made of the intersection height, H, above the z-axis at the
image plane, against some other variable such as tanU
, the points at which serious lens design begins. Seidel theory is adequate to
give the iterative ray tracer a place to begin, and when low focal ratios are needed,
photographic lenses can be made to look at the collimated beams which emerge
from F,12 optical systems. The spectrograph designer will rarely have to deal with
anything more complex than the design of an air-spaced achromatic doublet. It
is the proper positioning of simple elements along the optic axis which chiey
gives renement, resulting eventually in a high resolution and etendue over a wide
spectral range.
14.6.1 Worked example
This is a 1-metre focal length
CzernyTurner CCD spectrograph intended for night
airglow spectroscopy, with angles chosen to remove tangential coma at 5000
A in
second order. The spectrograph was intended to photograph the spectrum between
4000
A and 6000
A in tranches of 500
A at a resolution of 0.5
A, turning the grat-
ing for each new spectrum and moving the focal reducer on a sliding carriage if
necessary to refocus.
A Ramsden-type focal reducer was added to put the spectrum on to a CCD with
1152 325 pixels of 25 m width covering 28.8 mm. The plano-convex lenses are
of BK7 glass and have focal lengths of 497 mm (eld lens) and 429 mm (eye lens).
The focal reducer requires a longitudinal movement of a few millimetres to bring
all sections of the visible region into focus. As it is a tranche of 500
A can be
photographed at any one time.
It was designed using the BEAM 3 program mentioned earlier.
Table 14.1 shows the layout of the ray-tracing program for the
CzernyTurner
spectrograph with a focal reducer made to order, taking surface radii from the
manufacturers catalogue. The specication was that the grating ruled width was
120 mm, the (paraboloidal) mirror focal lengths were 1000 mm and the nal CCD
camera lens was to be an F,1.8 50 mm focal length 35 mm camera lens. The design
requirement then was for parallel ray bundles to emerge fromthe focal reducer with
a diameter of 25 mm.
T
a
b
l
e
1
4
.
1
14.6 Requirements of a ray-tracing program 147
0
0
200
200
400
200
Figure 14.1 The optical layout of the
CzernyTurner spectrograph described in
the text. The focal reducer is a simple Ramsden-eyepiece design, that is to say, with
plano-convex lenses having the plane surfaces outwards. This serves to collimate
the emerging spectrum at a diameter of 23.4 mm and simultaneously images the
grating on to the nal image-plane where the pupil of the CCD camera lens would
be placed. The camera lens (not shown) may typically be a 35 mm camera lens
with a focal ratio of F,2 or less, and its optical quality is likely to determine the
ultimate resolving power of the instrument.
The initial layout had the angles appropriate to zero coma at 5000
A, with the
mirrors at z = 1000 and the grating at 2 f ,3 from the M2 mirror vertex, as required
by the Seidel at-eld condition. Rays were launched from z = 2.65 for minimum
coma at the M1 mirror, and the nal positions for M2 and the following optics were
achieved by the requirement for a at eld at the CCD, that is, for the emerging ray
bundles at 5000
A (paraxial) and 5250
A (marginal) to be parallel, and in particular
to have the x-direction cosines all as closely as possible equal to each other. The
y-direction cosines which describe the astigmatism were not subjected to the
iteration process. The question marks in the delimiter columns following the lens
vertices were instructions to the iterator to vary the vertex positions of the focal
reducer to minimise the aberrations. The letters d and f , in the column below each,
were instructions to move the second lens surface with the rst to keep the lens
thickness constant.
Figure 14.1 shows the layout of the resulting spectrograph. In the nal ver-
sion the ray bundles have diameters of 23.4 mm. This compromise came from
the best choice of available tool radii when the lenses for the focal reducer were
ordered.
Figure 14.2a shows the spot diagrams of the nal collimated output beams at
5000.00
A and 5000.15
A, showing the eventual resolution at the zero-coma wave-
length. Figure 14.2b shows a similar pair at = 5250
A. The 1000 plotted points
are of random rays from the same launch point at z = 2.56; x = 0 and are of U
f
148 Optical design
0.002
(a)
0.000 V
f
0.002
0.0000 0.0002
U
f
0.002
(b)
0.000
0.2362 0.2364
U
f
0.2366 0.2368
V
f
0.002
Figure 14.2 Spot diagrams of images in this spectrograph at (a) 5000
A and
5000.15
A at the centre of the eld and (b) 5250
A and 5250.15
A at the long-
wavelength margin. The astigmatism increases at the margin but the spectral res-
olution is not signicantly compromised. Notice that the units are radians, the
inclination to the optic axis of the outgoing rays, and that the vertical and horizon-
tal scales are different. These gures must be multiplied by the focal length of the
CCD camera lens to give the actual sizes of the spots on the CCD surface. With a
50 mm camera lens for example, the widths would be 7 m and the length at the
eld centre would be 150 m.
and V
f
, the angles
4
(in radians) which they make with the optic axis. These must
be multiplied by 50 to show the spot positions on the CCD. An angle of 0.0002
radian corresponds to 10 m at the CCD surface.
4
The sines of the angles in fact, but clearly the difference is negligible.
14.6 Requirements of a ray-tracing program 149
0.04
0.02
0.0000
U
f
0.0005
V
f
0.00
Figure 14.3 This is the spot diagram of rays from the top half of a straight entry
slit 20 mm long. The horizontal scale is magnied to 100 times the vertical. Line
curvature is now dominant and it may be necessary, before co-adding rows of
pixels, to shift (in the computer) the upper and lower rows sideways in order to
straighten the image and give the best spectral resolution. Care is needed here
because the line curvature varies with wavelength. There is room in this data-
reduction process for considerable sophisticated software writing, since each row
of pixels is, in effect, a separate spectrum with a slightly different dispersion.
Figure 14.3 shows the spot diagramfor = 5000
Awhen a nite (10 mm) entry-
slit length is included. Nine rays have been launched in two annular bundles from
x = 0 mm and from x = 10 mm, both directed to arrive at the same places on the
grating. A thousand rays are then launched randomly from all points with x = 0,
z = 2.56 and y between 0 and 10, to arrive at the same points on the grating.
There is little change in the magnitude of the aberrations but the line curvature
described in Eq. (8.8) is apparent. As in the other spot diagrams, notice the different
vertical and horizontal scales, which here show that with a 50 mm focal length
camera lens the line images are 3 mm or 120 pixels long but that there is still less
than 1 pixel width of curvature.
The nal layout of the instrument was the result of many hours of investigation,
and the position in particular of the M2 mirror is the position which determines the
at eld at the CCD. It is some distance from the position demanded by Eq. (4.12)
but takes account of the eld curvature added by the focal reducer. In the end the
resolving power of 30 000 achieved was better than the CCD resolution, and a
CCD with a 10 m pixel width would have been useable.
With a sufciently bright source, the CCD camera could have been replaced by
a traditional 35 mm camera with ne grain lm.
15
Mechanical design and construction
Mechanical motions, automatic adjustments and electro-mechanical monitoring
and position sensors are so diverse, and the range of transducers and electric motors
for remote focusing, wavelength shifting and measuring so vast that there is no
point here in listing them or criticising their relative merits. So much depends on
the specication of the instrument and the capability of the available workshops that
it must be left to the builder to decide how the instrument shall be controlled. The
simplest possible mechanical construction is shown here handles to turn to adjust
grating turntables, capstan-bar mirror adjustments and so on, and the improvement
of these by other means is clearly up to the designer and the size of his purse. This
chapter therefore is conned to those matters which are peculiar to spectrograph
construction.
15.1 The optical layout
The positions of the optical components will have been decided by the nal version
of the design coming from the ray-tracing program. The position tolerance for the
mirrors is 0.5 mm on both x- and z-axes, the small errors being taken up when
focusing.
It remains to dress them in appropriate mountings and assemble the mountings
either on to an optical table or into a tube or a space-frame.
The art of constructing a scientic instrument is different from that of engineer-
ing mechanisms. The loadings are generally lighter and the precision as high as
that of a machine tool or an internal combustion engine, and different criteria apply.
Kinematic mounting, for example, may involve high point-loading with the conse-
quent risk of fretting corrosion but yields a precision of fractions of a micrometre
where necessary. It is suitable for optical instruments where there is no sliding
motion between parts.
150
15.1 The optical layout 151
15.1.1 Kinematic mounting
Briey, the principle is that to x the relative positions of two solid objects, six
restraints are required, three of translation and three of rotation. The common
example is that of a surveyors cube-corner reector which is to stand on a xed
table. The cube-corner mounting has three steel balls press-tted into holes in its
base to act as feet and the table has three radial grooves, each of which xes a ball
by making contact with it in two places. The device can be removed and replaced
with a precision limited only by any particles of dust which may have landed in
a groove at a point of contact; in other words, to a fraction of a wavelength of
light. Occasionally a hole-slot-and-plane arrangement is used instead to give the
six constraints, although the trihedral hole required for this is more difcult to make.
The penalty for this precision is the high point-loading at the contacts between
the surfaces, and the method fails if there is a weight or contact force sufcient to
distort the material signicantly. It also fails when heavily loaded sliding contacts
are involved. Dust in the air consists largely of micrometre-sized crystals of SiO
2
and these embed themselves in the softer of two surfaces and abrade the harder.
This is a classical example of the apparent anomaly that if two surfaces of unequal
hardness rub together it is the harder one which wears away. If the two surfaces are
of the same material, the frictional heat generated at the point of contact may be
enough to weld them together. It that case more robust methods of support are used
and a ball-race, for example, will provide adequate precision with rolling friction
for a grating spectrograph.
There is little in the literature on kinematic mounting and Braddicks classic
work was published over 50 years ago.
1
The principles however are simple and still
leave much room for inventive design.
For any optical instrument the order of design is important. In a
CzernyTurner
spectrograph for example the largest item is likely to be the M2 mirror, and the
mounting for this will determine the height of the optic axis above the baseplate,
or the internal diameter of the tube, or the datum surface if a space-frame is to be
used. From this the designs of the other mirror mounts, the slit and detector holders
and the grating turntable will follow.
15.1.2 The main frame
There are broadlyfour ways of constructinga spectrographas illustratedinFig. 15.1.
(1) A light alloy casting, forming an optical bench or table with raised pads at appropriate
places to be milled at to hold the optical components.
1
H. J. J. Braddick, The Physics of Experimental Method (London: Chapman & Hall, 1954).
152 Mechanical design and construction
(a)
(b)
(c)
Figure 15.1 Various methods of constructionfor spectrographs. For manypurposes
the plain optical table (a) is adequate but may prove too heavy for instruments much
greater than 1 m focal length. The cover, which must of course be removable, may
be of sheet metal, wood or GRP, depending on the skills available. The tube (b),
which is its own shroud, is appropriate to large instruments of 3 m focal length
and is also very convenient for small, portable spectrographs of 30 cm focal
length. The Serrurier truss (c) allows maintenance of the optical alignment when
the direction of g is changing as on a large astronomical telescope.
(2) An optical table of sheet aluminiumwith holes bored and threaded at appropriate places
to hold the optical components. It can be a box girder and, if very large and heavy, may
be self-standing, resting its Airy points on two pillars with the optical table vertical for
greater stiffness. The optical components are then cantilevered out sideways on their
holders. (This, incidentally, takes up much less space in a crowded laboratory.)
The covering shroud may be of metal, wood or GRPdepending on the skills available.
(3) An open-frame lattice to which the optical component-holders are bolted, with a
wooden,
2
GRP or sheet-aluminium shroud. If the instrument is to be attached to an
astronomical telescope the Serrurier truss type of space-frame (Fig. 15.1c) is worth
considering.
(4) A tube, drawn if of small (-20 cm) diameter, but probably rolled and welded if larger,
with shrouded tables, plates or boxes at the ends to hold the optical components.
The choice depends on the size, the application and the quantity to be constructed.
Access is always important for focus adjustment and cleaning, and large, light-tight,
hinged doors at appropriate places should be part of the design, as are access holes
for screwdrivers, spanners and capstan bars for ne adjustment while the instrument
is operating.
2
The author once used this type of construction with the optical shroud doubling as the packing case into which
everything was packed for transport to a remote observing site.
15.1 The optical layout 153
The optical table is probably the simplest method of construction for spectro-
graphs up to 1 m focal length. The mounting positions on the table are set by the
optical design and from this the screw-hole positions follow. Threaded holes are to
be preferred to through-bolting for ease of access and assembly.
Space-frame construction may be considered for large instruments, with 23 m
focal length mirrors. This is because optical tables with adequate stiffness become
unacceptably heavy and unwieldy and one must consider the feasibility of moving
them from room to room and from building to building. Space-frames are more
transportable although equally bulky, and the dimensions of corridors and goods-
lifts
3
which may be needed to move them about must be measured. Spectrographs
with focal lengths much greater than 1 m are not generally transportable. If they
are to have carrying handles, then the limit is about 0.5 m.
When designing a space-frame, the criterion is, again, stiffness. Because of
interference with the optical path, internal cross-bracing is scarcely possible but
Warren-girder construction of welded 25 mm square-section steel tubing with solid
12.5 mm light alloy end-plates will answer the purpose. Mirror mounts similarly
will act as cross-braces if designed into the frame. There is much to be said for a
vertical optical table if a
CzernyTurner mounting is specied. The space-frame is
comparatively stiffer and both sides of the optical table are accessible in the event
that electronic equipment is situated there. If this arrangement is chosen, the input
side should be below the detector, so that the grating tilt, facing the input side, is
downwards, reducing the amount of dust falling on the grating surface.
The principle of the Serrurier truss is that the frame is allowed to sag but in such a
waythat the ends remainparallel toeachother andthe optic axis is not compromised.
Such an arrangement is worth considering if there is a weight constraint and the
spectrograph is to be mounted on an astronomical telescope where the direction of
gravity is varying with respect to the instrument during the observation. Elaborate
civil-engineering design is not called for and the rule is: if in doubt make it stiffer.
A large laboratory- or Coud e-focus instrument, by contrast, is not expected to
move and once it is set in position the optical alignment can be completed in the
knowledge that it will remain aligned until the desired measurements are made.
15.1.3 Mirror and lens holders
The construction of these is governed by the need to avoid any sort of pressure or
loadingwhichmight cause distortionof the optical surface, especiallyas the result of
a change of temperature. Small mirrors, of diameter less than 50 mmcan be glued
to a backing plate with an epoxy resin, provided that a single small area of contact
3
Freight-elevators if you are an American reader!
154 Mechanical design and construction
Figure 15.2 The mirror holder. For mirrors of intermediate size, in the region of
150400 mm diameter, a simple semi-kinematic mounting is adequate. The rim
rests on two of three cylindrical bearings covered by thin polythene sheaths, and is
held by retainers of berylliumcopper screwed to the ends of the bearings, exerting
minimum pressure on the mirror.
Figure 15.3 The mirror cell. This is suitable for smaller mirrors of a fewcentimetres
diameter. The mirror rests on a polythene pad and is retained by a counter-cell held
in place by screws (not threaded as lens counter-cells sometimes are).
(5 mmdiameter) is sufcient to carry the weight. Multiple attachment points must
be avoided because of the bending which may result from differential expansion.
Mirrors and lenses larger than this require either a cell, or quasi-kinematic support.
Figure 15.2 illustrates a suitable mounting for mirrors of 150400 mm diameter.
Mirrors smaller than this may be contained in a mirror cell such as that in Fig. 15.3,
15.1 The optical layout 155
screw for
altitude alignment
transverse hole for
capstan-bar to make
azimuthal alignment
Figure 15.4 The mirror mount. The principle is that the mirror can be rotated in
azimuth and altitude almost exactly about its vertex and, once adjusted satisfactor-
ily, is xed in place by a single bolt. The same mounting is suitable for gratings
when no rotation is needed to change the wavelength range.
where they are backed by a thin polythene lm and held in place by a counter-cell
xed by screws which are not over-tightened.
The thin polythene lm is used to avoid high local pressure, which may damage
brittle materials like glass. The accuracy with which a mirror is positioned is not
high by engineering standards, and an error in centration
4
of 0.1 mm is generally
acceptable when focal lengths are of the order of 1 m.
15.1.4 Mirror adjustments
The mirrors require adjustment and an altazimuth set of axes is convenient and
practicable. In Fig. 15.4, the mirror holder may be attached to an L-bracket by
means of a horizontal berylliumcopper strip-spring clamped in place with a keeper,
and the L-bracket is rotated about a vertical axis which passes through the mirror
vertex. To ensure this is a simple matter of design and the rotation is secured by a
journal bearing a simple drilled hole in the baseplate with a brass dowel passing
through a similar hole in the L-bracket. This arrangement works well for mirrors
4
Centration here means the accuracy with which the optic axis of a lens or the vertex of a paraboloidal mirror
coincides with the centre of the circular rim.
156 Mechanical design and construction
up to 400 mm diameter. The L-bracket is clamped in place with a bolt tting a
threaded hole in the baseplate through an arc in the L-bracket base. A washer at the
bolt head is vital here to retain the accurate positioning of the azimuth as the bolt
is tightened.
The optical alignments are then carried out as follows: the tilt about the horizontal
axis is governed by a brass screw with a 1 mm pitch bearing against the rear of the
mirror cell and held in contact either by the force of gravity or if necessary by a
retaining spring. The adjustment about the vertical axis is by a silver-steel capstan
bar inserted into a hole at the elbow of the L-bracket. The capstan bar should be
long enough to allow the positioning of the mirror to 3
of arc a comfortable
tolerance in practice. Multiple attachment of the L-bracket to the baseplate is not
necessary, provided of course that the single bolt is big enough.
If the instrument is subject to shaking or vibration a more solid xing is needed.
In this case three pairs of push-pull screws will hold plates against plates essentially
as a solid, and the holding of the mirror in its cell becomes a more difcult task when
allowance is made for differential thermal expansion and the brittleness of glass and
fused quartz. An oversized cell may be used with a gap around the mirror edge
approximately three times the thickness of the berylliumcopper leaf-springs which
are inserted to keep the mirror centration exact. A front-plate, also of beryllium
copper, with a 3 mmoverlap at the edge, will do no more than obscure the turn-down
at the edge of the mirror. These extra precautions are only needed if a very large
temperature range between midday and midnight in the desert for example is
expected.
For small mirrors, of less than 50 mm diameter, the vertical axis is sometimes
carried on a berylliumcopper strip-spring, but this foregoes the opportunity to have
the axis of rotation through the vertex, and requires a holding spring to locate the
azimuth adjustment.
15.1.5 Grating holders
Much depends on the portability of the nished spectrograph. In the case of a large,
immobile instrument with a horizontal optical table, the grating may be held free as
in Fig. 15.5, resting on a lower pad of stiff plastic such as polythene, and constrained
by a light alloy bar which rests on an upper polythene pad and in turn is held in
place by capstan-head screws. The rear face of the grating may be insulated from
its mount by a thin sheet of polythene.
If a cell is required it may be fabricated, or milled from a single block of light
alloy, with relieving holes drilled at the corners as in Fig. 15.6. Again space should
be left for insulating polythene sheets and if necessary PTFE clamping screws
will hold it immobile. Gratings are generally formed on quartz blanks and the low
15.1 The optical layout 157
Figure 15.5 The grating cell. The grating can stand on a shelf at the appropri-
ate height above the table, and is retained in place by a bar exerting minimum
downward pressure through two capstan screws.
cover plate
Figure 15.6 An alternative grating cell, in which the grating is fully enclosed
and retained by a face-plate of light alloy with an appropriate mask-shape as in
Fig. 8.5.
158 Mechanical design and construction
thermal expansion of these must be borne in mind. The grating may be held in
its cell by a face-plate which restrains the corners, exible enough that when the
screws are fastened it places only the necessary pressure on the grating, sufcient
to maintain it against temperature changes and to bear its weight when it is possibly
face-down at the Cassegrain focus of a telescope.
Optical adjustments are probably needed only in the initial alignment process,
in which case no elaborate rotation about the grating normal is called for. Rotating
the grating in its own plane to align the rulings can be done with copper or bronze
shims under the lower edge.
5
A metal shim, a sliver 25 m thick, under one end of
a 150 mm wide grating will tilt the rulings by 0.5
of arc.
15.1.6 The grating turntable
In a high-resolution spectrograph it would be unusual if the whole of the spectral
range were available to the detector without rotating the grating. If the adjustment
is infrequent the grating cell can be mounted on a circular brass pad which rotates
about a simple journal bearing in the form of a spindle lying in a hole in the optical
table. It is then held in position by a clamping screw, much like the provision of the
mirror holders.
If adjacent regions of the spectrum are to be photographed in rapid succession
a more elaborate arrangement is needed and a proper turntable must be used. The
standard method in this instance is to use two large-diameter, light-duty angular
contact bearings inopposition, heldinpermanent contact bymeans of a compression
spring as in Fig. 15.2. A simple worm-gear can be used for rotation, but there is
an advantage in using a sine-bar, as in Fig. 15.7. The valuable property of a sine-
bar motion is that the wavelength presented at a particular pixel varies linearly
with the angle turned by the screw. This in turn leads to convenient calibration
and re-setting. In a scanning monochromator, such a drive, usually motorised, is
regarded as virtually essential. Of various possibilities in practice, the double-
hinged version is to be preferred. The moving disc bearing against a ball engenders
wear and fretting corrosion with consequent loss of calibration and smoothness of
scan.
In the asymmetrical
CzernyTurner mounting the grating face must make an
angle withthe sine-bar towhichit is xedsince the gratingequation, Eq. (8.2),
which governs the relation between captive-nut position and grating tilt angle, t ,
can be rewritten as
n,a = 2 cos( +) sin(t + ). (15.1)
5
With good modern workshop practice this will almost certainly not be necessary.
15.1 The optical layout 159
Figure 15.7 The grating turntable. In most cases it is not possible to record the
entire spectrum in one exposure, and different regions are brought into view by
rotating the grating. It is important to be able to do so accurately and for this a grat-
ing turntable is required. A standard form is shown here with manual adjustment
which can obviously be replaced by a stepper-motor under computer control. The
sine-bar link ensures that the screw moves the central wavelength of the spectrum
linearly with rotation angle.
In the original Ebert mounting and were the same and + was called the
ebert angle.
Whichever method is used, there is backlash in the drive screw to be considered.
Although there exist sophisticated anti-backlash split-nuts and similar devices, a
light pressure applied to the turntable by a spring is adequate for most purposes.
15.1.7 Entry and exit slits
Many constructors have spent time, ingenuity and energy on manufacturing double-
openingslits, where bothjaws move apart symmetrically. This is probablya mistake.
Traditionally only one jaw moved and the other, xed jaw was the ducial mark
for measuring deviation angle in a prism spectroscope. An emission line spectrum
could then be observed with slits as wide as convenient to allow plenty of light to
enter the eye, and this in turn made for easy location of the xed edge with a pair
of cross-hairs at 45
3 1),2
of arc in a 1-metre
spectrograph. The ultimate resolution at 5000
A typically is 0.015
A, subtending
1
I () d = 1
5
Especially in XUV region where it is next to impossible.
16.4 Absorption measurements 171
S() I () C()
=
Figure 16.1 The principle of the equivalent width measurement. The true absorp-
tion line S(), when convoluted with the instrumental prole I (), produces a
shallower, broader absorption line in the recorded spectrum. Provided there is no
contribution to the recorded continuumby scattered light, the area of the absorption
line is the same as that in the original, and measures the total amount of continuum
radiation that has been absorbed. The equivalent width then is the width of an
idealised rectangular absorption feature of full depth with the same area.
and if the limits a and b are far removed from the absorption region we can replace
them as usual by .
The output from the spectrograph, which is the convolution of the two, is
_
S()I (
) d = C(
)
then
_
C(
) d
= S
because with
= y. d
= dy so that
_
S()I (
) d =
_
S() d
_
I (y) dy = S.
The equivalent width is thus dened independently of the spectrometer instru-
mental prole as L, where I ()L = S. In other words the area still measures
the amount of continuum power removed from the spectrum by an absorption line.
For those occasions where there is a need for the absorption line prole for
example where there are narrow deep cores to comparatively broad shallow lines,
showing perhaps that a cool, optically thick outer layer of stellar atmosphere or
an interstellar cloud has been traversed by the radiation there is no option but to
use a very high resolving power so that the slit width is narrow compared with the
half-width of the absorption line (which is not always well dened).
17
The alignment of a spectrograph
The method described here is suitable for the assembly and adjustment of a
Czerny
Turner type of spectrograph. Essentially similar methods are indicated for other
types and mountings.
Alignment is best carried out systematically after the instrument has been in-
stalled in its place of work or, if it is newly made, close to the workshop which
made it.
17.1 The optical alignment
The optical table has a number of holes drilled at the vertices of the mirrors and the
grating. Into these holes will normally t the vertical axle about which the angular
adjustments of the optical components are made. If a precision milling machine is
available, similar holes can be drilled along the optic path so that alignments can
be checked at any time.
An alignment tool (Fig. 17.1) should be made, preferably but not necessarily of
metal, which will stand on the optical table with a base extension which will t these
holes. It should carry a vertical disc with a 1 mmhole at the height of the optic axis.
1
Stage 1. A laser beam, adjusted to the height of the optic axis, is made to shine
through the centre of the entry slit so that it is parallel to the optical table surface.
Parallel here means that the light from the laser would be able to pass through the
hole in the alignment tool no matter where the latter is placed on the table.
Stage 2. The laser beamdirectionis nowadjusteduntil it shines towards the vertex
of the collimator (M1) mirror. The alignment tool, placed in the hole corresponding
to the vertical axle of the mirror mount, has the central hole at the mirror vertex
position. The tool can then be removed and the M1 mirror installed. The procedure
ensures that the beam is shining along the optic axis.
1
I am indebted to colleagues at Perkin-Elmer Ltd. for this method of optical alignment.
172
17.2 The focus 173
optic axis
base plate
Figure 17.1 The optical alignment tool. Its use depends on having holes for the
dowel drilled in the spectrograph baseplate along the optic axis, with x- and y-
coordinates accurate to 25 m. When a laser beam shines through the central
hole with maximum intensity, the alignment in a 1.5-metre spectrograph is good
to better than 1
of arc. The dowel holes should be close to (-25 mm) the mirror
and lens vertices and to the grating surface.
Stage 3. The alignment tool is next transferred to the grating table vertical axle
hole and the M1 collimator mirror is adjusted in both axes until the reected laser
beam passes through the hole in the alignment tool. The tool is then removed and
the grating with its mount is installed in its proper place.
Stage 4. The alignment tool is transferred to the vertical axle of the focusing (M2)
mirror and the grating adjusted in zero order so that the light passes through the
alignment tool. When this adjustment is satisfactory the grating should be rotated
until the rst-order reected beam passes through the hole in the alignment tool.
It is at this point that the grating may need shimming in case the rulings are not
parallel to the rotation axis. The adjustment is then checked by rotating the grating
so as to pass other orders through the alignment hole.
The M2 mirror and any auxiliary optics are installed, and the M2 mirror is
adjusted until the laser beam arrives satisfactorily at the pupil of the camera.
17.2 The focus
Stage 1. The grating is rotated on its table until zero order is reected back to the
input slit. A small (5 mm) 90
45
45
A) of discharge lamps
Hydrogen Mercury Sodium Helium Cadmium Zinc
6562.79 11287 11404.2 10830 6438.4696 6362.35
4861.33 10140 11382.4 7065.19 5085.82 5181.99
4340.47 5790.65 8194.81 5875.62 4799.91 4810.53
4101.74 5769.59 8183.27 4685.75 4687.15 4722.16
3970.07 5460.75 5895.92 4471.68 3610.51 4680.14
3889.06 4358.35 5889.95 4026.19 3466.20 3345.02
3835.40 4046.56 5688.22 3888.65 3403.65 3282.33
3797.91 3663.28 5682.66 3261.06
3770.64 3650.15 3303.00 2748.58
3750.16 2536.52 3302.32
Notes
(1) The cadmiumline at 6438.4696
Ais an old international wavelength standard. Being the
result of a singlet transition it is a particularly narrowline and eminently suitable for demon-
strating the normal Zeeman effect. It was subsequently replaced by the 6056.125
A line of
krypton-86 but cadmium lamps are still in common use and readily available.
(2) The zinc line at 5181.99
A is a convenient standard wavelength for measuring small
Doppler shifts in a moving source of the Mg triplet, a group of closely spaced Fraunhofer
lines in a solar or stellar spectrum.
(3) The HeNe laser line is at 6328.17
A.
Arc spectra
The spectrum of an iron arc is less convenient than a set of laboratory lamps and the
necessary apparatus is mostly to be found in older, well-equipped spectroscopic
laboratories. The iron arc in fact is the traditional source for the visible and near UV
region, with many hundreds of well-measured wavelengths. Sundry authors have
recommended a vertical arc about 12 mm long taking 5 amps from a 230-volt source and
with only the central 2 mm of the arc focused on to a horizontal slit. Welders dark glasses
are needed when adjusting it and it is a source of skin-damaging UV.
Alternatively, a hollow-cathode lamp will yield a similar spectrum but at much lower
intensity.
Passive wavelength standards
There are sundry methods of calibrating spectrographs which are independent of
electricity or sunlight and therefore suitable for eld-work at remote sites. Chief among
these is absorption spectroscopy using a tungsten lamp, possibly battery-powered, as a
light source.
One of the most useful calibrators for the visible region is an absorption cell preferably
of glass or quartz, with a diameter of a few centimetres and an optical absorption path
length of about 75 cm. It should be evacuated or possibly lled with an inert gas such as
helium, and should contain some crystals of iodine. At the equilibrium vapour pressure of
Wavelengths of spectral lines for calibration 181
Table A2.2 Solar Fraunhofer lines
Letter Wavelength,
A Element Source
A 7621 O Telluric absorption band
7594 O Telluric absorption band
B 6869.95 O Telluric absorption band
C 6562.79 hydrogen Balmer H
D 5895.92 Na Sodium resonance line
D 5889.95 Na Sodium resonance line
E 5269.54 Fe
b
1
5183.62 Mg Magnesium triplet
b
2
5172.70 Mg Magnesium triplet
b
3
5167.51 Fe
b
4
5167.33 Mg Magnesium triplet
F 4861.33 H Balmer H
G
4340.46 H Balmer H
G 4307.91 Fe
4307.74 Ca
g 4226.73 Ca
h 4101.74 H Balmer H
H 3968.47 Ca II Ionised calcium resonance
K 3933.67 Ca II Ionised calcium resonance
L 3820.43 Fe
M 3727.67 Fe
N 3581.21 Fe
O 3440.99 Fe
P 3361.21 Ti II
Q 3286.76 Fe
R 3181.28 Ca II
3179.33 Ca II
s
1
3100.67 Fe
3100.30 Fe
s
2
3099.94 Fe
T 3021.07 Fe
iodine at room temperature the electronic transition X
1
Y
+
g
B O
+
u
of I
2
gives a
well-known series of absorption bands.
2
These are sufciently deep and have a rotational
structure open enough that the narrow individual absorption lines give a large number of
accurate calibration points throughout the yellow and green. The amount of absorption
and therefore the depth of the lines depends markedly on the ambient temperature but the
wavelength calibration is very precise.
More compact, with some absorption lines in the visible but a large number in the UV,
is europium.
3
An absorption cell with an absorption path length a few millimetres long
2
I. J. McNaught, J. Chem. Educ., 57 (1980), 101.
3
G. Smith & F. S. Tomkins, Proc. Roy. Soc., A342 (1975), 149; Phil. Trans. Roy. Soc., A283 (1976), 345.
182 Appendix 2
lled with an aqueous solution of europium chloride is highly portable. The lines result
from atomic transitions betweeen energy levels of unlled inner electron shells in the
europium atom. The lines are not generally as narrow as those in the vapour phase but as
calibration marks are comparable with solar Fraunhofer lines.
Appendix 3
The evolution of a FabryPerot
interference spectrograph
This is to show the stages of evolution of a design for a FabryPerot spectrograph for
night airglow or astronomical applications.
(1) Astronomical considerations required that the resolution should be 1.2
A at
5183.6
A, the wavelength of the Mg triplet of solar Fraunhofer lines. The free spectral
range was to be about 20
A, set by the FWHM of the interference lter which was to act as
order-sorter. The available nesse was 40 and this constrained the etalon gap thickness to
be 28 m. The order at the centre of the ring system is then 56 m,0.5183.6 m = 108.
(2) The available CCD camera used a 50 mm F,1.4 Takumar lens and a CCD with
1152 325 pixels each 25 m square. The eld semi-angle was then 16
. A primary
requirement was for spatial resolution, so that both the sky and the FabryPerot fringes
were to be in focus on the CCD. It was also required that the angle subtended by the CCD
on the sky should not be greater than 5
.
(3) The etalon gap thickness determined the angular radii of the FabryPerot rings
(assuming m
0
= 108 at the centre) which are given by cos
n
= (m
0
n),m
0
with
n = 1. 2. 3 . . . so that cos
1
= 7.8
and 13.53
. Their
diameters on the CCD are at 50 mmtan
n
which are 0, 13.7 mm, 19.5 mm and 24.07 mm,
so that the CCD records approximately three orders of interference. This would be so even
if the order at the centre is not integer.
(4) Conservation of etendue then required a telescope with an aperture
1
32,5 times that
of the camera lens, i.e. 160 mm. The design then compromised with an available
plano-convex lens of 6
(152 mm) aperture and 1000 mm focal length. This would give
adequate resolution on the sky and the small spectral range meant that chromatic
aberration was no problem. It remained to collimate the image so that the sky could be
received at the CCD with the FabryPerot fringes superimposed on it. At the same time
this collimator had to image the objective on to the pupil of the camera. The result was the
conguration shown in Fig. A3.1a in which the image collimator needed is clearly
inconveniently large. This diagram, like the others, is not to scale, the vertical dimension
having been greatly exaggerated for clarity.
(5) The eld of the objective required by the camera is in the ratio (again conserving
etendue) of their pupil diameters, so that the eld semi-angle for the objective is
25,150 16
= 2.66
. The actual eld radius is then 46.5 mm, requiring the order-sorter
to have an effective diameter of 93 mm. In the actual instrument a standard 100 mm
diameter interference lter was used.
1
The ratio of the eld angles.
183
184 Appendix 3
etalon
CCD camera
image collimator objective
objective Chief-ray collimator Order-sorter image collimator CCD camera
objective
(a)
(c)
(b)
Figure A3.1 The evolution of a FabryPerot CCD spectrograph following the
precepts of Chapter 10. The vertical scale has been exaggerated for clarity. In
practice a 150 mm diameter objective would result in an instrument 2 m long.
(a) The primitive design. Objective collimator camera.
(b) The addition of a chief-ray collimator immediately to the left of the order-
sorting interference lter moves the exit pupil (and hence the CCD camera) in to
the F
1
focal plane of the image collimator.
(c) The splitting of the image collimator into two smaller lenses, with the same
focal length and P
1
principal plane as the original singlet lens, reduces the overall
diameter of the spectrograph to that of the objective. Ideally the rst member of the
pair should have a focal length which makes the marginal ray parallel to the optic
axis, but in practice one may compromise with the contents of a manufacturers
catalogue.
(6) It was a sine qua non that the order-sorter should have all chief-rays parallel to the
optic axis as they passed through it. This was achieved by inserting, just in front of the
focus of the 1000 mm objective lens, a chief-ray collimator lens of sufcient diameter
(100 mm) to cover the necessary 5
CzernyTurner, 80
Eagle, 94, 95, 99
EbertFastie, 77
Littrow prism, 60
grating, 78
PaschenRunge, 93
PellinBroca, 60, 61
Pfundt, 79
Rowland, 93
SeyaNamioka, 96, 97
JohnsonOnaka, 96
Wadsworth stigmatic, 94, 95
Connes, P., 4
Conrady, A. E., 28, 111, 133, 175
convolutions, 42
convolution algebra, 44
convolution theorem, 44
cosmic ray events (CREs), 124
Coud e focus, 41, 53
188
Index 189
Court` es, G., 132
Crookes, Sir William, 2
cubic splines, 169
curvature, astigmatic, 97
curvature, eld, 33, 177
curvature, surface, see surface curvature