Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Mass Flow Versus Volumetric Flow PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Mass Air Flow Sensors Mass Flow Versus Volumetric Flow

WARNING
MISUSE OF DOCUMENTATION

Application Sheet

Equation 1 Using the Ideal Gas Law, PV = nRT , solve for Volume (V): or V = nRT
P

The information presented in this application sheet is for reference only. Do not use this document as product installation information. Complete installation, operation, and maintenance information is provided in the instructions supplied with each product. Failure to comply with these instructions could result in death or serious injury.
This application note describes the difference between mass flow in terms of volumetric flow at standard conditions (760 Torr and 0C) and volumetric flow at nonstandard conditions. Mass flow is a dynamic mass/rate unit measured in grams/minute. By referencing a volumetric flow to a standard temperature and pressure, an exact mass flow (g/min) can be calculated from volumetric flow. It is common in the industry to specify mass flow in terms of volumetric flow at standard (reference) conditions. In accordance with these standards, Honeywell mass flow sensors are specified as having volumetric flow at calibration reference conditions of 760 Torr and 0C. This translates to a specific mass flow rate. 3 For example, 200 cm /min (volumetric flow) of nitrogen at standard conditions of temperature and pressure calculates to 0.2498 g/min mass flow. The following formulae are used to find true mass flow in g/min from volumetric flow (Q): I. FINDING TRUE MASS FLOW (g/min) FROM VOLUMETRIC FLOW (Q) DEFINITIONS P = Pressure 3 V = Volume (cm ) n = Number of moles of gas R = Gas constant .0821 (liters atm/mole K) 3 or 82.1 (cm atm/mole K) T = Absolute temperature in Kelvin (K) 3 = Gas density (g/cm ) m = Mass in grams (g)  = Mass flow (g/min) m Q = Volumetric flow Qs = Volumetric flow at standard conditions (sccm)

Equation 2 m Gas density is defined as: = V Equation 3 Substitute Equation 1 into Equation 2 to redefine gas density as: =
mP nR T

Equation 4 Mass flow is equal to density times volumetric flow  = Q rate: m Equation 5 Redefine mass flow using gas density as derived from the Ideal Gas Law. Substitute Equation 3 into
 = Equation 4: m mP nR T

Example 1 3 Assume a volumetric flow rate of Q = 200 cm /min of nitrogen (N2) at standard pressure of 760 Torr and pressure of 0C, and solve for true mass flow (g/min): Given: 3 Q = 200 cm /min m = 28.0134 grams in 1 mole of N2 n = 1 mole P = 1 atm (760 Torr) 3 R = 82.1 (cm 1 atm)/(mole K) T = 273.13K(0C)  = .2498 (g/min) Answer: m II. FINDING VOLUMETRIC FLOW (Q) FROM TRUE MASS FLOW (g/min) Microbridge products are specified in standard volumetric flow (Qs) such as standard cubic centimeters per minute (sccm) or standard liters per minute (slpm) which can be translated into true mass flow as indicated above. The microbridge sensor is a mass flow device rather than a volumetric one. At a constant mass flow, the microbridge device will give the same output voltage even if there are temperature or pressure changes. Because the microbridge sensor senses mass flow, confusion may result when mass

MICRO SWITCH Sensing and Control

Mass Flow Versus Volumetric Flow


flow sensors are used with volumetric devices, such as rotometers or pith-ball indicators. Accurate mass flow calculations for volumetric devices require consideration of both temperature and pressure ranges. At varying temperatures and pressures, these other volumetric devices indicate different flow rates than those indicated by microbridge sensors. Simple calculations can be used to show the relationship between mass flow and nonstandard volumetric flow. An AWM3100V with a mass flow rate of .2498 g/min (200 sccm) at the same pressure of 760 Torr (1.0 atm) but at a different temperature, 25C, has a 5 VDC output voltage, indicating a standard flow rate(Qs) of 200 sccm. The rotometer, however, would indicate a nonstandard volumetric flow rate, (Q). Use Equation 5 to rearrange the formula for the volumetric flow value to calculate the rotometer nonstandard volumetric flow rate. Equation 6 Q = n R T
mP

Application Sheet

 m

Use the following given values to calculate volumetric flow rate (Q). Multiply the R value by 3 1000 to convert the number to cm : Given:  = .2498 (g/min) m m = 28.0134 grams in 1 mole of N2 n = 1 mole P = 1.000 atm (760 Torr) 3 R = 82.1 (cm 1 atm)/(mole K) T = 273.13K(0C) + 25C = 298.13 K 3 Answer: Q = 218.26 cm /min In this example, the standard volumetric flow rate 3 (Qs) is 200 cm /min while nonstandard volumetric 3 flow rate increases to 218.26 cm /min.

This increase reflects the fact that as temperature increases, gas expands, placing more distance between gas molecules. More distance between molecules means less mass in a given volume as temperature increases. If mass flow is kept constant, and temperature increases, volume flow increases to pass the same amount of mass (molecules) across the sensor. (see Figures 1 and 2).

Figure 1

Figure 2

2 Honeywell MICRO SWITCH Division

For application help: call 1-800-537-6945

Mass Flow Versus Volumetric Flow


III. FINDING VOLUMETRIC FLOW (Qx) FROM STANDARD VOLUMETRIC FLOW (QS): Nonstandard volumetric flow can be found with standard volumetric flow using the ratio of temperature and pressure at referenced conditions (760 Torr, 0C) versus X conditions of temperature and pressure. This method of determining volumetric flow eliminates the use of gas density values at reference conditions (760 Torr, 0C) versus X conditions of temperature and pressure. FURTHER DEFINITIONS Qx = Volumetric flow at X conditions of pressure and temperature Qs = Volumetric flow at standard conditions of 760 Torr (1 atm) and 0C Tx = Temperature at X conditions in Kelvin (K) Ts = Temperature at standard conditions in Kelvin (K) Px = Pressure at X conditions in Kelvin (K) Ps = Pressure at standard conditions in Kelvin (K) If mass flow is held constant over temperature and pressure, then the following is true:

Application Sheet

 =m  m s x
That is,

 m s mass flow, at standard conditions is equal to

 mass flow at nonstandard X conditions of m x temperature and pressure.


Therefore,
mPs mPx Qs Qx = nRTs nRTx

Equation 7: Solving for Qx yields: Ps Tx Qx = Qs Px Ts Equation 7 Equation 7 calculates volumetric flow (Qx) at X conditions from volumetric flow (Qs) at reference conditions of 760 Torr and 0C.

Given: Qs = 200 sccm Ps = 1 Torr or 1 atm Px = 1 Torr or 1 atm Ts = 273.13K (0C) Tx = 298.13K (25C) Ps Tx 3 = 218.3 cm /min Answer: Qx = Qs Px Ts

For application help: call 1-800-537-6945

Honeywell MICRO SWITCH Division 3

Mass Flow Versus Volumetric Flow

Application Sheet

WARRANTY and REMEDY Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace without charge those items it finds defective. The foregoing is Buyers sole remedy and is in lieu of all other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. Specifications may change at any time without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use. While we provide application assistance, personally, through our literature, and the Honeywell website, it is up to the customer to determine the suitability of the product in the application. For application assistance, current specifications, pricing or name of the nearest Authorized Distributor, contact a nearby sales office. Or call: 1-800-537-6945 USA 1-800-737-3360 Canada 1-815-235-6847 International FAX 1-815-235-6545 INTERNET www.honeywell.com/sensing Info.sc@honeywell.com

Sensing and Control Honeywell Inc. 11 West Spring Street Freeport, Illinois 61032
P rinted w ith S oy Ink on 50%R ecycled Paper

www.honeywell.com

008043-1-EN IL50 GLO 1298 Printed in USA

You might also like