Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

WWW - Medicaldelta.ro Cabinet Medical Tulcea Video Endoscopie EG-EC-Kp

Download as pdf or txt
Download as pdf or txt
You are on page 1of 48

OWNER’S MANUAL

VIDEO GI SCOPES

EG-290Kp, EC-380MKp,
EC-380MK2p
EC-380FKp
EC-380FK2p
EC-380LKp
IMPORTANT
Intended Use (Gastoroscope)
The Video Gastroscope is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the Upper
Gastrointestinal Tract. The Upper Gastrointestinal Tract includes, but is not restricted to, the organs, tissues, and subsystems: Esophagus,
Stomach, Duodenum and Small Bowel.
The instrument is introduced per orally when indications consistent with the requirement for the procedure are observed in Adult and Pediatric
patient populations.
Never use the endoscope for any purpose other than that for which it has been designed.
The video endoscope contained in this manual can only be used with PENTAX Video Processor, Model EPK-100p, EPK-700 and EPK-1000.

Intended Use (Colonoscopes)


These Video Colonoscopes are intended to provide optical visualization of (via video monitor), and therapeutic access to, the Lower
Gastrointestinal Tract. The Lower Gastrointestinal Tract includes, but is not restricted to, the organs, tissues, and subsystems: Large Bowel.
These instruments are introduced per rectally when indications consistent with the requirement for the procedure are observed in Adult and
Pediatric patient populations.
Never use these endoscopes for any purpose other than that for which they have been designed.
These video endoscopes contained in this manual can only be used with PENTAX Video Processors, Model EPK-100p, EPK-700 and EPK-
1000.

Notes
Read this manual before operating, and save this book for future reference. Failure to read and thoroughly understand the information
presented in this manual, as well as those developed for ancillary endoscopic equipment and accessories, may result in serious injury including
infection by cross contamination to the patient and/or user. Furthermore, failure to follow the instructions in this manual may result in damage
to, and/or malfunction of, the equipment.
This manual describes the recommended procedures for inspecting and preparing the equipment prior to its use and for the reprocessing and
maintenance of the equipment after its use. It does not describe how an actual procedure is to be performed, nor does it attempt to teach the
beginner the proper technique or any medical aspects regarding the use of the equipment.
It is the responsibility of each medical facility to ensure that only well educated and appropriately trained personnel, who are competent and
knowledgeable about the endoscopic equipment, antimicrobial agents/processes and hospital infection control protocol be involved in the use
and the reprocessing of these medical devices. Known risks and/or potential injuries associated with flexible endoscopic procedures include,
but are not limited to, the following: perforation, infection, hemorrhage, burns and electric shock.
Current infection control guidelines require that G.I.scopes and other semi-critical medical devices, that normally come into contact with intact
mucous membranes, such as in the gastrointestinal tract, must at least be high-level disinfected before patient use. Only the user can
determine if an instrument has undergone appropriate infection control procedures prior to each clinical use. It must be recognized that
infection control practices involve many complex and often controversial issues which are constantly evolving. PENTAX strongly recommends
that user remain informed of the latest federal and local regulations, and encourages users to follow infection control guidelines developed by
various organizations for health care professionals.
The text contained in this manual is common for various types/models of PENTAX endoscopes and users must carefully follow only those
sections and instructions pertaining to the specific instrument models appearing on the front cover.
If you have any questions regarding any of the information in this manual or concerns pertaining to the safety and/or use of this equipment,
please contact your local PENTAX representative.

Sterility Statement
The instruments identified in this instructional booklet are reusable medical devices. Since they are packaged non-sterile, they must be high-
level disinfected or sterilized BEFORE initial use. Prior to each subsequent procedure, they must be subjected to an appropriate cleaning and
either high-level disinfection or sterilization process.

Conventions
Throughout this manual, the following conventions will be used to indicate a potentially hazardous situation which, if not avoided;

WARNING : could result in death or serious injury.

CAUTION : may result in minor or moderate injury or property-damage.

NOTE : may result in property-damage. Also, advises owner/operator about important information on the
use of this equipment.

Prescription Statement
Federal (U.S.A) law restricts this device to sale by or on the order of a physician or other appropriately licensed medical professional.
TABLE OF CONTENTS
1. NOMENCLATURE AND FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1-1. VIDEO ENDOSCOPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1-2. ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1-3. VIDEO PROCESSOR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2. PREPARATION AND INSPECTION FOR USE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6


2-1. INSPECTION OF THE VIDEO PROCESSOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2-2. INSPECTION OF ENDOSCOPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2-3. PREPARATION JUST BEFORE INSERTION OF ENDOSCOPE . . . . . . . . . . . . . . . 13

3. DIRECTIONS FOR USE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14


3-1. PRETREATMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3-2. INSERTION AND WITHDRAWAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3-3. BIOPSY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3-4. LASER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3-5. ELECTRO-SURGERY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

4. CARE AFTER USE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20


4-1. ENDSCOPES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4-1-1INTERNAL CHANNELS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4-1-2PRE-CLEANING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4-1-3LEAKAGE TESTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4-1-4CLEANING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4-1-5HIGH-LEVEL DISINFECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4-1-6STERILIZATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4-2. ENDOSCOPIC ACCESSORY INSTRUMENT (EAIs) AND ENDOSCOPE
COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4-2-1CLEANING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4-2-2DISINFECTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4-2-3STERILIZATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4-3. WATER BOTTLE ASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

5. POST REPROCESSING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5-1. SERVICING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5-2. CARE AND MAINTENANCE TIPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
1. NOMENCLATURE AND FUNCTION
1-1. VIDEO ENDOSCOPE

AIR/WATER FEEDING VALVE OF-B121 RUBBER INLET SEAL


Covering of hole in the top of the valve delivers pressurized air. Covering Allows passage of accessories while
of the hole and fully depressing the valve delivers pressurized water. preventing escape of fluids and air.

SUCTION CONTROL VALVE (Part # OF-B120) INSTRUMENT


Depress to remove fluids or air through CHANNEL INLET
the instrument channel. For introduction of
CONTROL BODY
biopsy forceps and
other accessories.
“F” (FREEZE) BUTTON
Push to freeze an image.

U
L
“C” (COPY) BUTTON
Push to activate the hardcopy system that was selected

F
between “FILE” and “HARD COPY”.

R
D
MODEL DESIGNATION
RIGHT/LEFT DEFLECTION
CONTROL KNOB

UP/DOWN DEFLECTION LOCK


UP/DOWN DEFLECTION
When this lever is in the “F” position, turned clockwise, the bending section
CONTROL KNOB
moves freely. When turned counterclockwise, the bending section
becomes progressively more stabilized.

RIGHT/LEFT DEFLECTION LOCK


Functions similar to Up/Down lock
ENHANCE BUTTON
Push to activate the VCR for recording live procedures.

EPK-700 EPK-1000/100p

BUTTON Freeze Freeze UMBILICAL CABLE

BUTTON Hardcopy Hardcopy

BUTTON VCR Enhance

-2-
DISTAL END
(Refer to the inside rear cover of this manual)

BENDING SECTION

INSERTION TUBE

FEED BACK TERMINAL


Certain manufacturers’ electrosurgical units, such as older VENTING CONNECTOR
Olympus models, require the connection of a scope feedback Accepts “RED” ETO GAS Sterilization Venting cap.
cord (S-cord) to this feedback terminal. Also accepts Leakage Tester.
ETO GAS STERILIZATION VENTING
CAP RED OF-C5
AIR/WATER PORT
To connect feeding tube from water bottle Provides venting of endoscope interior to equalize
assembly. internal and external pressures. This cap must be
removed before immersion.
NOTE: See important separate section regarding the
use of this cap!

PVE CONNECTOR
Can be rotated within a
180° range. PVE SOAKING CAP OE-C9
This cap must be securely attached
before immersion.

ELECTRICAL CONTACTS
SUCTION NIPPLE
For attachment to external
suction source LIGHT GUIDE
Transmits light from light source to distal end of endoscope.

WARNING: Note:
Immediately after use, the metal light guide prong and the Ensure that the soaking cap has been securely
electrical contacts/pins of the endoscope may be HOT. attached (by properly rotating it) to prevent the cap from
To avoid burns, do not touch these areas immediately coming off during reprocessing. Failure to securely
after use. For safer handling after a procedure, grasp the attach the soaking cap can result in scope damage.
PVE connector housing.

-3-
1-2. ACCESSORIES

1) Cleaning Brush for Instrument Channel

WHITE BRISTLE FLEXIBLE SHAFT HANDLE

(CS6021SN)

2) Cleaning Brush for A/W Suction Valve Cylinder

( CS-C5S )

3) Biopsy Forceps

ENHANCED FLEXIBLE GRIP


FLEXIBLE SHAFT
PORTION

CUPS HANDLE
Pink handle defines autoclavable forceps

NOTE:
Depending upon country and/or local PENTAX distributor, each PENTAX endoscopic accessory may be
an optional accessory.

CAUTION:
Because of the effect that accessories used through the instrument channel of the endoscope can have
on the performance of the endoscope itself, it is strongly recommended that PENTAX accessories be
used with PENTAX endocopes. If a unique or highly specialized accessory is available from another
source, please contact PENTAX to arrange a test of its compatibility before using it through the PENTAX
endoscope.

NOTE:
Maximum outer diameter of an endoscopic accessory instrument must be at least 0.2 mm less than the
specified instrument channel diameter in PENTAX endoscopes. Working length of an endoscopic acces-
sory instrument may be approximately 30 cm longer than the endoscope working length.

NOTE:
For patient contact endoscopic accessories, follow the specific and detailed instructions on use, care and
maintenance supplied with each product.

-4-
1-3. VIDEO PROCESSOR

CAUTION:
Please refer to the instruction supplied with the processor.

NOTE:
Do not use the new PENTAX OS-H4 water bottle cap with the older OS-H2 water container/bottle.
Although the cap may appear to fit onto the bottle, air may escape resulting in insufficient pressure and
flow of air and water during the endoscopic procedure. Both the PENTAX water bottle cap and bottle
(container) are identified by their appropriate model designation. Ensure that an OS-H4 cap is used only
with the OS-H4 water container/bottle.
Do not overtighten the bottle cap. Overtighting can cause the bottle cap to break.

BRIGHTNESS CONTROLS DOCUMENTATION CONTROLS


- Hardcopy, VCR etc.
EPK-700 - Freeze

EPK-700
L
E CONTRO
BALANC
COLOR
ESS
BRIGHTN COPE
+5
LAMP
AUTO
RED
SERVICE MAN BLUE
0 FREEZE

AVE
CENT -5
XLUM

POEWR
P
AIR PUM ON

ON OFF
OFF

POWER SWITCH
I : ON
WATER BOTTLE O : OFF
AIR PUMP
SWITCH
INTERFACE SOCKET X-LUM
SWITCH

BRIGHTNESS CONTROLS
EPK-1000
AIR PUMP SWITCH

EPK-1000

POWER SWITCH
LAMP SWITCH I : ON
WATER BOTTLE O : OFF
INTERFACE SOCKET

EPK-100p BRIGHTNESS CONTROLS


AIR PUMP SWITCH

EPK-100p

POWER SWITCH
LAMP SWITCH I : ON
O : OFF
WATER BOTTLE INTERFACE SOCKET

-5-
2. PREPARATION AND INSPECTION FOR USE
Prior to use, the endoscope, video processor and endoscopic accessory instruments must be carefully inspected for
cleanliness and proper function to determine that they are appropriate for patient use:

NOTE:
PENTAX video endoscopes contained in this manual are only compatible with PENTAX video proces-
sors, model EPK-100p, EPK-700 and EPK-1000.

2-1. INSPECTION OF THE VIDEO PROCESSOR

Please refer to the Owner’s Manual of the specific model of PENTAX video processor for complete instructions.
1) Attach water bottle assembly, 2/3 filled with sterile water to the appropriate location on the left side of the
video processor.

WARNING:
The addition of defoaming agents to the water supply is NOT recommended. Due to their nature, these
silicone based agents cling tenaciously to surfaces. Unless they are rinsed very thoroughly, a “barrier”
could be created which could reduce the effectiveness of the disinfection/sterilization process. Addition-
ally, repeated use of such defoamers could eventually lead to residual silicone build up resulting in
equipment malfunction such as clogged air and/or water channels.

2) Set the drain lever on the water bottle assembly to the upright position labeled A/W (air/water).
3) Plug the processor into a properly grounded receptacle with the power switch in the OFF position.

EPK-100p
4) Make sure that the PENTAX PVE connector is aligned
with the interface socket on the front panel of the proc-
essor.
EPK-100p
5) Connect the endoscope to the interface socket on the
processor as illustrated. Rotate the lever of the inter-
face socket clockwise after insertion.

CAUTION
Failure to do so may create a “miscommunication”
air/water feeding tube between the scope and processor and may result
in the loss of the system’s automatic brightness
control and in excessive light emission from the
scope. Excessive light intensity can in turn
increase the amount of delivered thermal energy
causing heating of the scope tip, specifically the
illumination (LCB) cover glass and/or objects in
contact with the distal cover glass.

6) Connect the air/water feeding tube from the water bot-


tle assembly to the air/water port on the side of the
PVE connector.
7) Turn the processor and air pump to the “ON” position and check for proper functioning.
8) Prior to each procedure, check the endoscope image quality displayed on the monitor. Confirm that the image
quality, color, automatic brightness (iris) functions are acceptable as per the instructions provided with the
PENTAX video processor.

NOTE:
As indicated elsewhere in PENTAX product labeling, endoscopes particularly the quality of the endo-
scopic image should be checked prior to patient use.

-6-
2-2. INSPECTION OF ENDOSCOPE

NOTE:
Flexible endoscopes and other sophisticated medical instruments are constructed of special materials,
unique parts and intricate components with strict dimensional tolerances. Specialized assembly tech-
niques and application of specific sealants and/or adhesives are required to ensure the watertight integ-
rity and maintain the functionality of these devices. It is therefore imperative that endoscopes be routinely
checked to ensure that parts used in their construction are not loose, missing or compromised that could
otherwise negatively affect the functionality of these devices. Compromised or loose components could
result in device failure, scope damage (via fluid invasion) and/or in incomplete decontamination of used
instruments.
PENTAX recommends that prior to use endoscopes should be carefully inspected for their integrity and
checked for any "looseness" in the mating or joining of components including the following parts/areas:
• the channel inlet assembly (biopsy inlet port) (1)
• the suction nipple/connector (2)
• the air/water inlet port (3)
• the forward water jet connector (4)
• any valve cylinder (5)
• basically, any inlet or outlet port associated with an internal channel, an indirect patient contact portion
of the endoscope
• rubber strain relief along insertion tube and umbilical cable (rotate clockwise only to tighten)

GI Scope Non-GI Scope

One method to check for looseness is to lightly grip the exposed part, and while grasping the component
carefully attempt to move it in various directions. Use of a lint-free gauze while grasping metal parts is
recommended as a protection for one's fingers.
If any part/component remains loose (after attempting to tighten) and/or if there is any indication
or suspicion of an abnormality or outward signs of damage, do NOT use the endoscope. Contact
your local PENTAX service facility.

-7-
CAUTION:
If the endoscope is intended to be clinically used after testing of individual scope functions (suction, air/
water delivery, water jet, etc.) without further reprocessing, the following precaution should be exercised.
Use “fresh” distilled or sterile water during individual scope function tests to avoid recontamination of the
previously reprocessed instrument by waterborne microorganisms. Tap water, especially that which may
be left idle and uncovered for a prolonged period of time, should not be used during any inspection/test-
ing of the endoscope.

Before proceeding with inspection of individual functions, PENTAX Endoscopes should be tested for the integrity
of their water-tight design (example: tear in the instrument channel). This test is described in another section of
this manual entitled: “Leakage Testing.”
ST
1) Inspection of the Insertion Tube
TE 19.6

a) Check the entire surface of the insertion tube for


D
AN

29.4
GER

9.8
LEKAGE TESTER
SHA-P2

0
kPa
39.2
abnormal conditions such as dents, wrinkles or bite
marks. Any indentation of the flexible shaft of the
endoscopes can cause damage to the internal mech-
anisms of the endoscopes.
b) Similarly, check the condition of the umbilical
cable for outward signs of damage such as buck-
Leakage Tester ling, crush marks, etc.

CAUTION:
To avoid further damage to the endoscope or the possibility of malfunction during a procedure, do not
use an endoscope with any abnormalities or outward signs of damage.

c) Make sure that the entire endoscope is clean and has been subjected to either a high-level disinfection or
sterilization process before each patient use.

WARNING:
All instruments must be reprocessed prior to first time use, after any repairs/service and before every
patient use. When utilizing chemo-thermal processes for reprocessing the endoscope, the instrument
should be allowed to return to room temperature prior to use and/or further handling.

NOTE:
• The distal end of the endoscope must be protected against damage from impact. Never apply excess
force such as twisting, or severe bending to the flexible portion of the endoscope.
• During pre-use inspection, ensure that the distal objective lens and the illumination (LCB) cover glass
are clean and no residues are present on these distal surfaces. If not, crisp images can NOT be dis-
played. Wipe with a gauze or the like moistened with an enzymatic detergent solution.

2) Inspection of Deflection Controls and Locks


a) Slowly manipulate the Up/Down and the Right/Left
control knobs to see that they function smoothly. Be
RIGHT-LEFT
certain that a full and appropriate range of deflec-
tion is possible.
b) Engage the deflection locks to be certain that the
position of the deflected tip can be stabilized.
UP-DOWN

NOTE:
The deflection locks are of the friction type and the degree of friction is adjustable. The degree of locking
friction depends on the rotational position of the lever.

-8-
CAUTION:
ANY lack of smooth operation of the deflection controls may be an early indication of internal damage to
and/or part(s) failure within the endoscope’s angulation system. To avoid the possibility of further endo-
scope damage or the potential for malfunction of the angulation system, do NOT use the endoscope if the
angulation mechanism does not operate properly.
Prior to use ensure that the deflection controls can rotate smoothly, that there is no grinding or excess
friction within the angulation system and that the distal bending section bends freely and smoothly.
NEVER APPLY EXCESSIVE FORCE TO THE DEFLECTION CONTROLS!
When an endoscope exhibits excessive “knob play” or if angulation is lost in any direction, do NOT use
the instrument. Excessive “knob play” can be defined as rotating of the angulation control knob(s) in any
one direction for more than 30 degrees without any corresponding distal tip deflection. The examples
above are indications that service is required to avoid more serious problems with the angulation control
system, including angle or pulley cable/wire breakage and/or the possibility of a “frozen” distal bending
section.
A “frozen” bending section can make instrument extraction from a patient more difficult.

3) Inspection of Air/Water Feeding


OF-B121 a) Prior to use, the air/water feeding valve (OF-B121)
should be inspected. Remove the valve from the
control body and make sure that the black rubber
check-valve (OE-C14) in good condition is prop-
erly attached to the bottom of the metal valve stem.
Don’t let the skirt
fold under b) If the rubber check-valve is missing or not attached
properly, correctly reposition the check-valve by
turning it several times on the air/water valve stem.
OE-C14 For proper positioning, there should be no clear-
ance (gap) between the rubber check valve and the
metal air/water valve stem.

WARNING:
A worn or damaged check-valve should be replaced with a new one which has already been subjected
to a high-level disinfection or sterilization procedure (Check valve set, model OE-C15 is optionally avail-
able). For repeated use, always ensure that the check-valve has already been reprocessed. A dam-
aged, worn or missing check-valve could create continous air flow or excessive air insufflation and result
in potential patient injury such as pneumatic perforation. If the check-valve is not attached properly, it
could fall off during the endoscopic procedure and result in potential injury to the patient.

CAUTION:
If there is any doubt as to the suitability for use of the rubber check-valve, replace with a fully reprocessed
one prior to each procedure. A set of 10 pieces of the OE-C14 check-valve are available in a package as
model OE-C15.

c) To confirm that the check-valve is attached properly, visually inspect the check-valve and install the air/
water valve into the air/water cylinder on the control body. Proceed to test the air delivery function.
d) Connect the scope to the video processor. Turn air pump “ON” to desired pressure setting.
Place the scope distal tip into sterile water and confirm that no air bubbles exit the distal air nozzle.

WARNING:
If air bubbles are observed during the test, the rubber check-valve MUST BE REPLACED. Repeat the
test procedure with a new check-valve (OE-C14).

-9-
e) To inspect air delivery, cover the hole at the top of
the air/water valve and confirm that air flows freely
from the air/water nozzle at the scope disital tip.
Action
f) By depressing the air/water feeding valve, the water
delivery system is activated. Water should flow in a
steady stream from the air/water nozzle at the distal
tip of the endoscope. (This may take several sec-
Result Air Water onds on the initial attempt.) USE STERILE
– Feeding Injection
WATER ONLY.
g) Release the air/water valve to determine if the valve
freely returns to its OFF (neutral) position and
OF-B121 delivery of water (and air) ceases.
h) If the air/water valve OF-B121 does not move
smoothly, remove the valve and apply a small
amount of silicone oil lubricant onto all the
O-rings. Remove/wipe off excessive lubricant with
a soft gauze.
Do not use excessive silicone oil.
O Ring : OF-B143

NOTE:
Excessive silicone oil (lubricant) should be avoided to prevent occlusion of the internal air or water chan-
nels/nozzles and potential impairment of the normally clear endoscopic image.

i) If air and/or water do not flow properly, NEVER attempt to clear the air or water nozzles with a needle or
any other sharp object. Instead, the following steps should be followed.
1 Disconnect the endoscope from the video processor.
2 Remove the air/water feeding valve.
3 Using a cotton tipped applicator and alcohol, clean the valve recess (receptacle) in the control body thor-
oughly to remove any debris. Do NOT attempt to insert the applicator into the small openings within the
valve receptacle as the cotton or applicator could become lodged within these openings and cause channel
blockage.
4 • Following the section of this manual for cleaning the air and water channel with detergent, flush deter-
gent through both the air and water channel.
Then flush the air several times to force any residual solution out of the channel.
• Remove the adapters and install the air/water feeding valve.
(Alternate) By leaving the air/water valve in the cylinder instead of the adapter, one may direct pressured
fluid (or air) independently to either channel to expel debris from and/or more forcefully flush solution
into either the air or water channel. This should not be attempted on a completely clogged/blocked air or
water channel/nozzle.

NOTE:
If blockage of the line is encountered, avoid use of excessive pressure to prevent scope damage.

5 Test for normal delivery of air and water. It may be necessary to repeat the above procedure if normal air
and water delivery is still not available.

NOTE:
Do NOT apply excessive force in an attempt to unblock a clogged channel as the channel and/or brush
could become damaged. Whenever channel damage is suspected, the endoscope should be leak tested.
If repeated attempts to flush the air/water system are unsuccessful, do not attempt to use the endoscope
on a patient. Contact the PENTAX Service Department.

NOTE:
Prior to clinical use, it is important that the entire air channel system be dry. Failure to thoroughly dry the
air system could result in an unclear or blurry image caused by very fine droplets of moisture being swept
over and/or onto the objective lens at the distal end of the scope.

- 10 -
CAUTION:
The instrument, Air/Water channel systems are made of stainless steel, poly phenylene oxide and fluo-
rine-contained polymers. When any fluids are used with this scope, please read carefully and follow all
instructions in the manual supplied with the fluids for use and pay special attention to any reactions with
the materials identified in the intended fluid path. Only the user can determine if the fluids are appropriate
for patient use.

DEPRESS 4) Inspection of Suction Mechanism


a) Prior to use suction control valve (OF-B120)
should be inspected. Remove the valve from the
control body and make sure that the rubber portion
SUCTION is not damaged or worn.
TUBE
SUCTION
NIPPLE

WARNING:
A worn or damaged valve and/or O-ring should be replaced with a new one. The entire valve mecha-
nism should be subjected to a high-level disinfection or sterilization procedure prior to use (O-ring set,
model OF-B127, is optionally available). Failure to do so could result in continuous aspiration which in
certain clinical situation can suction tissue into the distal channel opening at the scope tip and/or create
a loss of insufflated air via the suction system.
A compromised valve could also result in the potential for reflux or spit-back of patient fluids that may
present infection control risks.

b) Connect suction tubing from an external suction source to the suction nipple located on the PVE Connec-
tor at the end of the umbilical cable. Place the distal tip of the endoscope in a basin of sterile water and
depress the suction control valve. Water should be rapidly aspirated into the suction system collection con-
tainer.
c) Release the suction control valve to determine if the valve freely returns to its OFF position and the aspira-
tion of water ceases.

CAUTION:
If the instrument is to be used immediately after the inspection, use only “fresh” distilled or sterile water.
To avoid recontamination of a previously reprocessed endoscope, avoid use of idle/uncovered tap water.

NOTE:
A rubber inlet seal in good condition must be on the instrument channel inlet to prevent the loss of suction
and a risk of cross contamination to the end user due to the potential for reflux (spit-back) of patient flu-
ids. Worn seal will result in leakage and should be replaced. To ensure maximum performance of these
sealing mechanisms, consider replacing the rubber inlet seal with a new fully reprocessed one for each
procedure.

- 11 -
OF-B120 d) If the suction valve does not move smoothly,
remove the valve from the suction cylinder on the
control body of the endoscope. Apply a small
amount of silicone oil lubricant, OF-Z11, onto rub-
ber part and rubber O-rings. Remove/wipe off
excess lubricant with a soft gauze. Do not use
excessive silicone oil.
O-Ring Silicone oil 5) Inspection of Biopsy Forceps and Instrument Channel
OF-B127
a) Make sure there are no kinks in the flexible shaft of
the biopsy forceps.
CLOSE
OPEN
b) The jaws of the forceps must be free of any residual
debris. Any debris must be cleaned from the for-
ceps before they are used. Use only sterile forceps.
c) The handle mechanism on the forceps should be
operated to open and close the jaws. This mecha-
nism should operate freely.
d) Close and inspect the jaws of the forceps to make
sure the cups are in proper alignment. If the forceps
has a spike, the spike must be completely straight
and fully within the cups.

WARNING:
The use of any forceps or accessory that shows any sign of damage or difficulty of operation must be
avoided. Any malfunction of a forceps or accessory during a patient procedure could result in serious
injury to the patient. Also, the use of damaged forceps or accessories may result in serious and costly
damage to the endoscope.

e) Any accessory should be slowly inserted through the instrument channel inlet with the endoscope in a
straight position. There should be no resistance encountered. If resistance is encountered, do not attempt
to introduce the accessory further. The instrument channel may be damaged and the endoscope should not
be used. Contact the PENTAX Service Department.

WARNING:
All patient contact accessories must be thoroughly cleaned and subjected to an appropriate high-level
disinfection or sterilization process before being used for the first time and subsequently after each clin-
ical use.

CAUTION:
The instrument, Air/Water channel systems are made of stainless steel, poly phenylene oxide and fluo-
rine-contained polymers. When any fluids are used with this scope, please read carefully and follow all
instructions in the manual supplied with the fluids for use and pay special attention to any reactions with
the materials identified in the intended fluid path. Only the user can determine if the fluids are appropriate
for patient use.

- 12 -
2-3. PREPARATION JUST BEFORE INSERTION OF ENDOSCOPE

WARNING:
Every endoscope should be properly disinfected or sterilized before being used for the first time. The
endoscope should have been properly cleaned, disinfected or sterilized after any previous use and after
being returned for any repairs/service.
Refer to the companion manual describing in detail PENTAX reprocessing instructions.

WARNING:
Current infection control guidelines require that endoscopes and their patient contact accessories either
be sterilized or at the least be subjected to high-level disinfection. Accessories which ENTER STERILE
TISSUE or THE VASCULAR SYSTEM must be sterilized before patient use. It is recommended that any
accessory intended for use in the biliary tract be subjected to an appropriate sterilization process. Only
the user can determine if an instrument has undergone appropriate infection control procedures prior to
each clinical use.

1) Gently wipe the insertion tube with a gauze moistened with 70% alcohol.
2) Gently clean the objective lens with a cotton-tip applicator moistened with 70% alcohol. A lens cleaner (anti-
fogging agent) may also be applied via gauze or other applicator.
3) Check the endoscopic image and confirm that it is of acceptable quality for clinical use. Refer also to the
owner’s manual supplied with the PENTAX video processor for inspection of the image quality.
4) (SCOPES TO BE INTRODUCED TRANSORALLY)
Prior to trans-oral insertion of the endoscope, place a bite-block (mouthpiece) into the patient’s mouth to pro-
tect the endoscope from damage during the procedure. Failure to do so can result in scratches, tears and/or
crushing of the insertion portion of the endoscope.
Place a bite block on the insertion tube to protect the insertion tube before the scope is introduced.
5) Apply a medical grade water soluble lubricant to the insertion tube. Do not use petroleum based lubricants.

NOTE:
• The objective lens and the light guide must be kept free of the lubricant and excess lens cleaner.
• Prior to a procedure, remove debris or secreta from observation area as much as possible.

WARNING:
Never drop this equipment or subject it to severe impact as it could compromise the functionality and/or
safety of the unit. Should this equipment to be mishandled or dropped, do not use it. Return it to an
authorized PENTAX service facility for inspection or repair.

- 13 -
3. DIRECTIONS FOR USE

WARNING:
This instrument should only be used by physicians who have thoroughly studied all the characteristics of
this instrument and who are familiar with the proper techniques of endoscopy. During the procedure,
always wear protective garments such as gloves, gowns, face masks, etc. to minimize the risk of cross
contamination.

3-1. PRETREATMENT

1) The patient should be prepared in your normal endoscopy regimen.

3-2. INSERTION AND WITHDRAWAL

1) Slowly insert the scope under direct vision.


2) (SCOPES TO BE INTRODUCED TRANSORALLY)
When the distal end of the scope is passed through the
pharynx, the patient should be gently biting down on the
bite block to maintain the bite block’s position during the
BITE BLOCK
procedure.
3) Adjust the intensity of the video processor to obtain a
brightness level suitable for observation.

WARNING:
The light emission from the endoscope could cause thermal injury. To minimize the risk, use only the
minimum amount of brightness and avoid close stationary viewing and unnecessary prolonged use.

4) The deflection controls should be used as needed to position the scope. The deflection of the tip should be
done under direct vision in a gentle and deliberate manner.

CAUTION:
ANY lack of smooth operation of the deflection controls may be an early indication of internal damage to
and/or part(s) failure within the endoscope’s angulation system. To avoid the potential for malfunction of
the angulation system, do NOT use the endoscope if the angulation mechanism does not operate prop-
erly.
Ensure that the deflection controls can rotate smoothly, that there is no grinding or excess friction within
the angulation system and that the distal bending section bends freely and smoothly.
NEVER APPLY EXCESSIVE FORCE TO THE DEFLECTION CONTROLS!
If during a procedure angulation is lost in any direction such as when “cables snap” (broken pulley wire,
broken angle wire, etc.), do NOT continue to use the instrument and do NOT rotate the deflection con-
trols. Should the angulation system fail for any reason, stop the procedure, release the lock lever and
carefully withdraw the endoscope under direct visualization.
The examples above are indications that service is required to avoid more serious problems with the
angulation control system, including the possibility of a “frozen” distal bending section.
A “frozen” bending section can make instrument extraction from a patient more difficult.

5) Insufflation can be controlled by the combined use of the air valve or CO2 valve to increase the amount of
insufflation and the suction control to decrease the level of insufflation. The user may elect to use CO2 deliv-
ery in lieu of air delivery by using the CO2 gas adapter (OF-G11) in combination with a special CO2/water
valve.

- 14 -
WARNING:
It must be recognized that variations in air flow (pressure and volume) for patient insufflation may exist
from one manufacturer’s equipment (processor and/or scope type) to another. It is, therefore, important
to closely monitor the patient at all times and to aspirate excessive air to prevent overinsufflation and
potential pneumatic perforation.

6) Mucous or other debris may be aspirated via the instrument channel to improve visualization. Do not attempt
to aspirate solid materials.

NOTE:
Continuing use of the light guide with sticky debris might cause steam because debris is deprived of
moisture by heat. If steam is found on the light guide during a procedure, stop it immediately, and with-
draw the scope carefully from a patient.

7) The objective lens may be cleaned during the procedure by alternately using the air/water and suction control
valves.

NOTE:
Should debris on the objective lens be difficult to clean, one can temporarily use the HIGH air pressure
setting on the processor and simultaneously press the air/water and suction control valves. Return air
pressure setting to original selection before proceeding.

8) Image capture, hard copy documentation, video recording, etc. may be carried out as necessary.
9) Before withdrawing the scope, trapped air should be suctioned to reduce patient discomfort.
10) When attempting to withdraw the scope, return the deflection locking levers to their free position. Always
withdraw the scope under direct visualization.

UP/DOWN DEFLECTION LOCK


free position RIGHT/LEFT WARNING:
(lock released) DEFLECTION LOCK
FREE POSITION If for any reason, the image is lost due to power
(LOCK RELEASED) shortage, lamp or processor failure, etc. the
deflection locking levers should be released, the
scope tip should be straightened to its neutral
position, and the insertion tube should be care-
fully and slowly withdrawn from the patient.
LOCK POSITION
LOCK POSITION

- 15 -
3-3. BIOPSY

CAUTION:
For ALL types of endoscopic accessory instruments, always maintain a view of the accessory during
advancement, use and withdrawal of the device.

1) Insert the forceps through the slit in the rubber inlet seal. Be certain to hold the forceps handle in such a way
to ensure that the jaws of the forceps are in a fully closed position during insertion.

NOTE:
When the cups are first passed through the inlet
m seal, a temporary resistance will be encountered.
5c Hold the shaft tightly at about 5cm from the cups
and push it through.

NOTE:
During insertion, if the forceps are found hard to advance further due to resistance, decrease the deflec-
tion of the bending section to a level suitable for smooth insertion and insert the forceps again.

CAUTION:
Never apply excessive pressure when introducing
any accessory since the instrument channel may
be damaged. Malfunction of the scope as well as
costly repairs may result.

2) When a portion of the cups of the forceps becomes


visible in the viewing field, carefully advance the for-
ceps onto the target area.
3) Open the forceps cups and advance the forceps against
the target area. Carefully squeeze the forceps handle to
close the cups and obtain a specimen within the cups.
Always maintain a view of accessory during advance-
ment.
4) Withdraw the forceps slowly with the cups closed.

NOTE:
Because of the effect accessories used in the instrument channel of the endoscope can have on the per-
formance of the endoscope itself, it is strongly recommended that only PENTAX accessories be used
with PENTAX endoscopes. If a unique or highly specialized accessory is available from another source,
please contact PENTAX to arrange a test of its compatibility before using it through the PENTAX endo-
scope.

- 16 -
3-4. LASER

Laser equipment should only be used by physicians who have thoroughly studied all the characteristics of the
equipment and who are familiar with the proper techniques of endoscopic laser therapy. The user must carefully
read and follow all instructions in the operating manual supplied with the Laser equipment. The Laser equipment
should be carefully and thoroughly inspected and calibrated. Only the user can determine if the condition of the
Laser equipment is suitable.

WARNING:
The PENTAX endoscopes identified in this manual are compatible with Nd:YAG laser and Diode Laser
(wavelength 800 nm ~ 1000 nm) only. Do not use these endoscopes with other types of laser such as
KTP, He-Cd, or Excimer Laser Systems. It could result serious injury to the patient.

1) The user has the option of using a non-explosive gas


CO2 GAS
ADAPTER for insufflation. Non-explosive gas from a pressure-
OF-G11 regulated and flow-rate controlled source can be con-
nected to the provided or optionally available gas
adapter, Model OF-G11, as illustrated.
CO2 GAS
CYLINDER
NOTE:
Never connect an unregulated source of gas to the
PENTAX endoscope.
PROCES-
SOR
LIGHT GUIDE PLUG

2) The gas adapter, which can be secured to the air/water port on the PVE connector, has a luer receptacle to
accept tubing from an external source of non-explosive gas. As long as the air/water feeding tube from a
PENTAX water bottle assembly is connected to the gas adapter and the air pump in the video processor is
turned OFF, non-explosive gas can be delivered.

NOTE:
Set the pressure below 49 kPa (0,5 kg/cm2, 7,1 PSI) and the flow rate at about 1 liter/min.

CAUTION:
Open the valve of the CO2 gas cylinder only AFTER turning off the pump switch of the processor. Failure
to do so will apply excessive pressure to the processor and can cause damage to the air pump.

3) Flow of gas from the nozzle at the distal end of the scope can be checked by placing the tip of the scope under
water and covering the hole on the top of the CO2/water valve. The flow rate of gas should be no greater than
the rate of air delivery when the air feeding valve on the control head of the scope is covered.
4) The water delivery system is activated by depressing the CO2/water feeding valve.
5) The operator and assistant(s) should wear surgical gloves to avoid burns during use of Laser equipment.

NOTE:
It should be noted that as long as the valve of the CO2 gas cylinder is OPEN and the hole at the top of the
A/W feeding valve is NOT covered, CO2 gas will constantly be vented through the A/W valve into the
room. To reduce excessive CO2 concentrations, it is, therefore, recommended to close the CO2 gas cylin-
der valve, work in a well ventilated room, and use air delivery whenever possible during examinations
which are lengthy or in very confined quarters. As an alternative, the optionally available gas/water valve,
Model OF-B130, may be used in place of the standards air/water valve. OF-B130 is a closed two-stage
valve mechanism. Pressing the first stage delivers CO2 gas and fully depressing the second stage acti-
vates water delivery.
When using the OF-B130 valve, there will be no venting of CO2 gas into the room.
Replace OF-B130 with the air/water valve OF-B121 after using the CO2 gas.

- 17 -
NOTE:
One may choose to leave the OF-G11 adapter attached to the endoscope during conventional air insuf-
flation using the standard air/water valve. However, the luer sideport of the OF-G11 must be capped.
Similarly, for normal water delivery, the air pump must be turned ON and the plastic luer lock cap must be
secured to the OF-G11 adapter.

6) The laser probe should be introduced through the endoscope in the same manner as described for biopsy for-
ceps in section 3-3.
7) The position of the active portion of the laser probe should always be clearly visualized before Laser equip-
ment is activated.
8) It should be recognized that a variety of factors can effect the quality of the video endoscope image during
laser use. Intensity of the aiming beam, high power setting of the laser, close distance of laser fiber to scope
tip, excessive tissue burning, can each adversely influence image quality. To obtain optimum results, it is rec-
ommended that the power settings of the aiming beam and laser be adjusted to minimal levels capable of
achieving the desired clinical effect.
9) Follow standard hospital protocol regarding safe-use of lasers, including the wearing of safety eyewear.

WARNING:
Prior to activation of the laser, make sure that the laser fiber exits the distal channel opening of the
endoscope. Failure to do so can result in scope damage and possible ignition/burning of the endoscope.

NOTE:
It is normal for the aiming light to appear white in the video scope image.

SMEAR 10) Should the distal tip of the endoscope be moved closer
than 20 mm from the irradiated tissue surface, the aim-
IRRADIATE
ing beam may create a “smear” in the image as shown
AREA at left. If this smear affect becomes too severe and dis-
torts the visual field, the intensity of the aiming light
should be decreased.
11) When activating the laser at high power (about 100 W
for Yag Laser and 60 W for Diode Laser.) and/or if the
endoscope tip comes to within 10 mm of the irradiated
PROBE
tissue, flare may appear at the corners of the image as
FLARE shown at left.

WARNING:
Activation of the laser at high power settings may
cause patient injury or thermal damage of the
endoscopes tip.
Avoid use of high power.

PROBE

- 18 -
3-5. ELECTRO-SURGERY

WARNING:
Please refer to the operating manual provided with the electrosurgical unit. Electrosurgical systems may
be of the floating type (BF type, CF-type) or non-floating (B type). To avoid patient and user burn, use
only the floating type ESU (such as ERBOTOM ICC 200) /accessory systems. Do not use the non-float-
ing (B type) electrosurgical systems. The electrosurgical generator and any electrosurgical accessory
should be carefully and thoroughly inspected. Only the user can determine if the condition of the electro-
surgical generator and the electro-surgical accessory are suitable.

1) The user has the option of using a non-explosive gas


CO2 GAS
ADAPTER
for insufflation. Non-explosive gas from a pressure-
OF-G11 regulated and flow-rate controlled source can be con-
nected to the provided or optionally available gas
adapter, Model OF-G11, as described for Laser in sec-
CO2 GAS tion 3-4.
CYLINDER 2) The electrosurgical accessories should be introduced
through the endoscope in the same manner as
described for biopsy forceps in section 3-3.
LIGHT
SOURCE
LIGHT GUIDE PLUG

WARNING:
To avoid patient and user burn, follow the instruction below before electrosurgical energy is delivered.
1) Use only the electrosurgical generator with the floating grounding type (BF or CF Type). Do not use
the non-floating (B type) electrosurgical systems.
2) Wear rubber gloves and face masks.
3) The position of the target area, the insulated distal portion of the electrosurgical accessory and the
active portion of the electrosurgical accessory, should be visible.
4) The active portion of the electrosurgical accessory should not touch the surrounding tissue directly or
via fluids.
5) The head of any lesion such as polyp should not touch the surrounding tissue directly or via fluids.
6) Physicians and assisting personnel should avoid contact with the patient while high frequency
energy is delivered.
7) Electro-surgical energy should be delivered for as short a time period as necessary to accomplish
the desired clinical effect.
8) Select a high frequency output power setting suitable for the particular intended procedure in order to
avoid thermal invasion of the tissue or insufficient coagulation resulting in excessive bleeding.
9) To avoid the risk of thermal injury, use only insulated accessories.
Never use non-insulated devices while performing endoscopic electrosurgical procedures.

NOTE: (In combination with PENTAX


Video Processor, Model EPK-1000/
EPK-100p)
to video processor It should be recognized that the use of electro-
surgical accessory devices employing high
frequency current may interfere with the normal
endoscopic image and this interference is not indi-
to endoscope cative of a malfunction of the video endoscope
system. Pentax has developed an earth cable,
model OL-Z3 intended to reduce potential RF inter-
ference and electronic noise that may appear in the
endoscope image when using electro-surgical
devices. Ensure that cable OL-Z3 is correctly con-
nected between the endoscope and video processor
as described in the instructions provided with the
OL-Z3.

- 19 -
4. CARE AFTER USE
IMPORTANT INSTRUCTIONS
Cleaning-Disinfection-Sterilization: PENTAX Endoscopes
To maintain maximum performance and a long service life of the Endoscope, proper care after each proce-
dure is extremely important. Immediately after the completion of a procedure, the Endoscope and its com-
ponents should be thoroughly and carefully cleaned. If the Endoscope or its components are left uncleaned
for some time after use, dried blood, mucous, contrast material or other patient debris may cause damage
to the instrument (component) or may interfere with the ability of the user to properly reprocess the endo-
scope or component.

NOTE:
This owner’s manual contains detailed recommendations on the manual reprocessing of PENTAX
endoscopes using PENTAX supplied cleaning/disinfecting adapters. Automated endoscope
reprocessors (AER) may provide a means of reprocessing flexible endoscopes, including PENTAX
instruments. Only those Automated Endoscope Reprocessors (AERs) should be used whose
manufacturers provide device-specific instructions and have validation data that support each AER claim
with respect to PENTAX model instruments.
AER manufacturers should be consulted for their specific claims including but not necessarily limited to
a) the ability of the AER to provide a cleaned and high-level disinfected (or sterilized) endoscope and
scope components (ex. valves),
b) the identification of any special feature area (internal channel) or scope component that can not be
reprocessed and therefore requires manual reprocessing,
c) the microbial quality of the rinse water,
d) the inclusion of an “automated” alcohol rinse cycle,
e) the inclusion of a terminal drying cycle that removes the majority of water/fluid within scope channels,
f) maintenance procedures for water filter replacement and/or decontamination of the filtration system to
ensure the microbial claim of water, etc.
g) compliance with local regulations and/or guidelines

NOTE (U.S.A):
The instructions for manual reprocessing of PENTAX endoscope contained in this manual are consistent
with reprocessing guideline, developed by medical professional organizations (ex. SGNA, ASGE, APIC)
and/or national consensus groups (including ASTM).

Summary of PENTAX Reprocessing Instructions


PENTAX reprocessing instructions can be broken down into seven basic steps.
1) Internal Channels Section 4-1-1
2) Pre-Cleaning (in the Examination Room) Section 4-1-2
3) Leakage Testing (in a designated Decontamination Area) Section 4-1-3
4) Cleaning (in a designated Decontamination Area) Section 4-1-4
5) High-Level Disinfection or sterilization (in a designated Decontamination Area) Section 4-1-5, 4-1-6
6) Endoscopic Accessory Instruments (EAIs) and Endoscope Components Section 4-2
7) Post Reprocessing Section 5

- 20 -
4-1. ENDSCOPES

4-1-1 INTERNAL CHANNELS


The following internal schematics have been provided as a service to help users better understand the intricate con-
struction of PENTAX endoscopes. Knowledge of the various internal channels and tubes within an instrument and
their relation to each other allows one to care for and reprocess the endoscope more easily and with greater confi-
dence.
Much time and effort has been expended into designing endoscopes and their cleaning/disinfecting components
so that reprocessing of the instruments before each patient use can be effectively and efficiently performed by ei-
ther manual methods or automated processes.
Connectors on all PENTAX cleaning/disinfecting adapters and scope inlet ports incorporate standard size luer-
lock and/or luer-slip fittings to easily accommodate reprocessing devices or systems available from other manu-
facturers.
As can be seen from these internal schematics, the PENTAX cleaning system promotes efficient unidirectional
flow of solution beginning from connections at the light guide plug, traveling up tubings in the umbilical cable to
the valve cylinders in the control body, passing through the channels in the insertion tube and finally exiting noz-
zles or channel openings at the distal tip of the scope.
The elimination of multiple branching of channels, combined with a direct and straightforward pathway for solu-
tions to travel maximizes flow efficiency and ensures contact of detergent/disinfectant/sterilant with all internally
exposed channel surfaces.

WARNING: (in the U.S.A. or other countries adhering to FDA regulations)


It is imperative that flexible endoscopes and other semi-critical devices be reprocessed using at least
high-level disinfection with a legally marketed sterilant/disinfectant. Only legally marketed endoscope
automated reprocessing devices/systems whose device specific claims have been validated by the AER
manufacturer and/or anti-microbial agents which have been tested and found to be compatible by
PENTAX should be used with PENTAX products.
Generally speaking, “2%” and “3.2%” alkaline glutaraldehyde solutions which have been FDA cleared
with High-Level Disinfection and/or Sterilization claims are recommended. It should be noted that the
actual percentage of active ingredient (glutaraldehyde) in these solutions, as per their product label,
may vary from the generic and traditional terms “2% glutaraldehyde” and/or “3.2% glutaraldehyde”.
For specific brands of compatible disinfectants/sterilants, please contact your local PENTAX service
facility or sales representative. Please also refer to the inside front cover of this manual regarding infec-
tion control.

- 21 -
(1) INTERNAL CHANNEL

Channel Inlet Air/Water Feeding Valve


Rubber Inlet Seal
Water Feeding Channel
Suction
Control Valve
Air Feeding Channel
Air/Water Nozzle

Instrument Channel

Air Water Feeding


Pump Tube

Air Feeding Tube

Water Bottle Suction Tube

Suction Source

Video Processor

The illustration above shows the actual routes taken by air, water, suction and forward water jet through single
biopsy/instrument channel PENTAX Videoscopes with forward optics.

- 22 -
(2) REPROCESSING SYSTEM

Channel Inlet A/W Instrument Channel Cleaning Adapter OF-B115


Rubber Inlet Seal
Water Feeding Channel
Air Feeding Channel
Air/Water Nozzle

Instrument Channel

Luer-Lock Syringe
with
Cleaning/Disinfecting
Solution

OF-G17
A/W Channel
Cleaning Adapter Water Feeding Tube

Air Feeding Tube

Suction Tube
Suction Nipple
PVE Soaking Cap ETO Cap
(ATTACH) (REMOVE)

Luer-Slip Syringe with


Cleaning/Disinfecting
Solution

CAUTION:
BEFORE IMMERSING:
a) The ‘Red’ ETO gas sterilization venting cap must be taken OFF.
b) The PVE Soaking cap should be securely ON the electrical contacts.

To reprocess a PENTAX single-channel videoscope, first an enzymatic detergent and then a high-level disinfectant
or sterilant must be exposed to all internal lumens as well as to all external instrument surfaces and scope components
(air/water valve, suction valve, etc.).
Exposure times of detergent and disinfectant/sterilant must be strictly adhered to.

Please note that all solution entrance ports and flow pathways are illustrated above.

NOTE:
Prior to exposure of all internal channels to an enzymatic detergent and high-level disinfectant (or steri-
lant), all brushable channels should be manually cleaned with cleaning brushes.

- 23 -
4-1-2 PRE-CLEANING

WARNING: The importance of meticulous mechanical cleaning of the endoscope and


its removable components cannot be overemphasized.
Prior to disinfection or sterilization, all instruments and components must be scrupulously cleaned.
Failure to do so could result in incomplete or ineffective disinfection and sterilization.
During the reprocessing process, always wear protective garments such as gloves, gowns, face masks,
etc. to minimize the risk of cross contamination.

1) Immediately after removing the endoscope from the


patient, gently wipe all debris from the insertion tube
with a gauze or the like moistened with an enzymatic
detergent solution.
2) Place the distal end of the endoscope into detergent
solution and aspirate through the channel for 5 ~
10 seconds. Alternate aspiration of solution and air sev-
eral times to create agitation for better pre-cleaning.
3) Set the lever on the water bottle to the drain position.
With the air pump of the video processor ON and set
to the HIGHEST pressure setting (if applicable),
depress the air/water valve of the scope fully until all
water has been discharged from the scope. Alternate
covering of the hole in the valve and depressing the
valve to forcefully expel mucous, debris, etc. which
may have entered the air and water nozzles.
4) Place removable scope components in enzymatic
detergent solution to pre-soak.

4-1-3 LEAKAGE TESTING


The PENTAX Leakage Tester allows for simple two (2) stage testing of the watertight integrity of the PENTAX
Endoscopes. Air pressure is delivered by means of a hand operated bulb, eliminating the need for any electro-me-
chanical devices.

DRY TEST, STAGE I


BEFORE IMMERSION, PENTAX Endoscopes should be tested for any major loss of integrity in their watertight
construction (example: major tear in the instrument channel). All valves, inlet seals, distal rubber tips and other
removable components should be detached from the scope prior to leak testing.
(Be certain this 1) Secure the Leakage Tester connector to the venting
connection is
TES
T
secure.)
connector on the scope light guide. The PVE soaking
0.2

cap should be securely attached to/over the electrical


D
ANGER

0.1 0.3

Pressure 0 0.4

Release
kg/cm2

contacts. The Leakage Tester connector and the vent-


Valve PVE
Soaking Cap ing connector on the scope light guide MUST be dry
before connecting. Proper connection will require
Leakage alignment of the venting connector pin and clockwise
Tester Connector
Hand Bulb rotation of the Leakage Tester connector.
2) Turn the gauge faceplate to ‘zero’ the pressure indica-
GREEN ZONE tor.
ST
3) Pressurize the scope by pumping the hand bulb until
TE 19.6
the indicator on the gauge is in the GREEN zone.
D
AN

RED ZONE DO NOT pressurize into the RED zone, as it may


29.4
GER

Faceplate 9.8
LEKAGE TESTER

rotates SHA-P2
cause serious damage to the scope.
39.2
0
kPa

- 24 -
NOTE:
During the leak test procedure, the insertion tube of the scope should be flexed in various positions and
the distal bending section should be fully angulated in all directions to confirm the absence of a leak.

4) Observe the gauge pressure to determine if the indicator remains in the GREEN zone. If the indicator drops
from the GREEN zone rapidly, a major leak may be indicated.

NOTE:
Be certain that the pressure release valve on the handle of the Leakage Tester has been tightened.
DO NOT IMMERSE the entire instrument if the gauge indicator does not remain in the GREEN zone.
Instead, contact your PENTAX Service facility.

WET TEST, STAGE II


After determining the absence of any major leak in Stage I testing, PENTAX Endoscopes may be immersed to test
for loss of integrity in their watertight construction.

PVE Soaking Cap


CAUTION:
BEFORE IMMERSING:
a) The ‘Red’ ETO gas sterilization venting cap
must be taken OFF.
b) The PVE Soaking cap should be securely
ETO Cap ON the electrical contacts.

1) With the Leakage Tester securely attached to the scope and the scope pressurized with the gauge indicator in
the GREEN zone, and the PVE Soaking Cap securely attached over the electrical contacts plug on the umbil-
ical cable of the scope, the entire scope with ALL valves and inlet seals removed may be immersed in clean
water.

NOTE:
Only the Leakage Tester connector and a small portion of its tubing should be immersed. NEVER
immerse the entire Leakage Tester.

2) Observe the instrument carefully while fully angulat-


ing the distal tip of the scope. A few bubbles may
occur initially from recessed areas of the scope. This is
normal. If a continuous stream of bubbles is observed
from the same spot, a leak is indicated. Immediately
remove the scope from the water.
DO NOT use the scope.
3) After removing the scope from water, release the air pressure within the scope by opening the pressure
release valve on the handle of the Leakage Tester. After the gauge indicates ‘zero’, disconnect the Leakage
Tester from the scope.

NOTE:
NEVER connect or disconnect the Leakage Tester under water. This will cause leakage of water into the
scope and Leakage Tester.
4) • If leakage was discovered in step (2), thoroughly dry the instrument and contact your PENTAX Service
Center.
• If no leakage was discovered in step (2), you may proceed with cleaning and disinfection of the scope as
described in this manual.

- 25 -
4-1-4 CLEANING
1) Prepare a basin with warm water and a mild enzymatic detergent per detergent manufacturer’s instructions.
The solutions must be enzymatic detergents or other cleaning agents specially formulated to clean flexible
endoscopes. For specific brands of compatible solutions, please contact your local PENTAX service facility
or sales representative.

CAUTION:
BEFORE IMMERSING:
a) The ‘Red’ ETO gas sterilization venting cap
PVE must be taken OFF.
Soaking Cap ETO Cap b) The PVE Soaking cap should be securely
ON the electrical contacts.

WARNING:
Immediately after use, the metal light guide prong and the electrical contacts/pins of the endoscope may
be HOT. To avoid burns, do not touch these areas immediately after use. For safer handling after a pro-
cedure, grasp the PVE connector housing.

1 Suction 2 air/water CAUTION:


control feeding valve The use of an enzymatic detergent immediately
valve
3 Inlet seal after each procedure to dissolve and remove
organic contaminants and proteinaceous debris
is essential to the care and maintenance of the
endoscope from the standpoints of infection
4 ETO Cap
control and functionality.

2) Immerse the endoscope and its components in fresh


detergent solution. Then thoroughly (but gently) wash
5 PVE Soaking Cap
the entire surface of the endoscope and its compo-
1 Suction Control Valve
2 Air/Water Feeding Valve
nents.
3 Rubber Inlet Seal Make sure that the recessed areas such as the channel
4 ETO venting cap openings, valve cylinders, etc are brushed clean using
5 Soaking cap
the provided or similarly effective cleaning brushes.

CAUTION:
While fully immersed, manipulate valve mechanisms and inject detergent via syringe into/through
removable scope components. This will remove entrapped air bubbles that could inhibit contact of solu-
tion with component surfaces & provide for better exposure of surfaces to detergent.

3) Allow all items to soak in an enzymatic solution for a time period recommended by the manufacturer of the
enzymatic detergent.

NOTE:
• Do not squeeze or severely bend the insertion tube.
• Do not use any abrasive materials.
• Be careful to avoid damage to the distal lenses.

- 26 -
4) Manual Cleaning of Suction System by Brush

5 Suction Valve

Channel Inlet

Rubber Inlet Seal


4 Suction Cylinder

3 Instrument Channel

2 Suction Channel

1 Suction Tube

Suction Nipple

Fig. Place Brush Entrance Exit Section

1 Suction Tube CS6021SN Suction Nipple Suction Cylinder 4-1

2 Suction Channel CS6021SN Suction Cylinder — 4-2

3 Instrument Channel CS6021SN Channel Inlet Distal Tip 4-3


CS6015ST (Ch. ø <2.4 mm)

4 Suction Cylinder CS-C5S Suction Cylinder — 4-4

5 Suction Valve CS-C5S Suction Valve — 4-5

- 27 -
Cleaning Brush CS6021SN A variety of special brushes have been provided to mechan-
3 2 ically brush clean the entire suction/instrument channels
and tubes. Whenever possible, the entire endoscope should
be immersed in detergent solution during the remainder of
the cleaning procedure.
Brush clean the entire instrument/suction channel system:
Instrument Channel Selector
1 Using the cleaning brush provided, insert the brush
into the opening of the suction nipple and gently pass
1 the brush until it appears in the suction control valve
receptacle. This will clean the suction tube within
the light guide/umbilical cable.
Then gently withdraw the brush. Repeat at least 3
times until it is visibly clean.
(CS-C5S)
2 Next, insert the brush into the opening at the bottom
of the suction control valve receptacle (cylinder) on
4
the control head and gently advance until resistance
is felt or, in case of two-channel scopes, until the
brush enters the cylinder of the instrument channel
selector. DO NOT USE EXCESS FORCE.
Then gently withdraw the brush. Repeat at least 3
times until it is visibly clean.
OF-B120
Depress 5 NOTE:
CS-C5S Be sure to inspect the bottom of the suction control
valve receptacle on the control head for any
debris.

3 Insert the brush into the instrument channel inlet and


gently advance the brush until it exits the distal end
of the scope. Clean debris off the brush and then gen-
tly withdraw the brush.
Repeat several times ensuring that only a clean brush
is introduced into the channel each time.
4 Using the large bristle of the specially designed
cleaning brush (CS-C5S), scrub clean the surfaces
inside the suction control valve receptacle on the
control head. Do not insert the brush excessively.
5 Scrub all internal and external surfaces of the suction
valve OF-B120 using the smaller side of the cleaning
brush (CS-C5S).

CAUTION:
A worn or damaged valve and/or O-ring should be replaced with a new one. The entire valve mechanism
should be subjected to a high-level disinfection or sterilization procedure prior to use (O-ring set, model
OF-B127, is optionally available). Failure to do so could result in continuous aspiration which in certain
clinical situations can suction tissue into the distal channel opening at the scope tip and/or create a loss
of insufflated air via the suction system.
A compromised valve could also result in the potential for reflux or spit-back of patient fluids that may
present infection control risks.

NOTE:
Brushing of all internal channels is not a substitute for exposure to an appropriate cleaning solution. Man-
ual cleaning by brush complements and augments the cleaning effectiveness of chemical cleaning (i.e.
enzymatic detergent).

- 28 -
5) Chemical Cleaning by Detergent Solution
Channel Adapters/Components to be attached Cleaning solution Rinse water
Air/Water Channel OF-B115 50 mL or more 100 mL or more
OF-G17
Biopsy/Suction Channel OF-B115 100 mL or more 150 mL or more
Rubber Inlet Seal
The rubber A/W Instrument channel cleaning adapter
(OF-B115) should be attached to the air/water and suc-
tion cylinders. This adapter caps (seals) off the air/wa-
OF-B115
ter and suction cylinders to allow unidirectional flow of
F solution through these delivery/aspiration systems.
C
Please note that the symbols on OF-B115 show a full
circle ( ) and circle with notch ( ) which represent
the shape of the respective cylinders for proper attach-
ment. The notched symbol should align with the suc-
tion cylinder and the circle symbol, the air/water
cylinder.

NOTE:
It is imperative that the OF-B115 channel cleaning adapter be securely attached to their respective valve
cylinders. Failure to properly match and secure the cleaning adapter could result in ineffective and incom-
plete reprocessing.

a) For Air/Water Cleaning


Air/Water Channel Air/water channel cleaning adapter, model OF-G17,
Cleaning Adapter (OF-G17) OF-B115
has a standard ANSI luer lock connector to which a
syringe or other device should be attached. Connect
OF-G17 to the air/water port on the PVE connector/
plug. Fresh detergent should be flushed through this
connector and will simultaneously flow through
both the air and water channels within the scope.
Please refer to the internal schematics.
Provided the enzymatic detergent is allowed remain in contact with the internal channel surfaces for the recom-
mended exposure time, the enzymatic solution should dissolve and clean any remaining debris within these chan-
nels.
(Alternate) By leaving the air/water valve in the cylinder instead of the OF-B115 adapter, one may direct pressured
fluid (or air) independently to either channel to expel debris from and/or more forcefully flush solution into either
the air or water channel. This should not be attempted on a completely clogged/blocked air or water channel/noz-
zle.

NOTE:
If blockage of the line is encountered, avoid use of excessive pressure to prevent scope damage.

b) For Biopsy/Suction Cleaning


1 Install a rubber inlet seal to the instrument channel inlet prior to injecting cleaning solution into the
suction system.
2 The suction nipple located on the PVE connector has a standard luer slip fitting to which a 50mL
syringe or other device may be attached. Fresh detergent solution should be flushed through the entire
instrument/suction channel system.
As an alternative, solution can be drawn into the instrument channels by attaching tubing from an aspirator to the
suction nipple, as long as the aspirator is turned on, detergent solution can be suctioned through the scope.

WARNING:
The enzymatic detergent solution should remain in contact with ALL internal channels and external
scope surfaces for the time period recommended by the manufacturer of the detergent.

- 29 -
WARNING:
It is important that ALL internal channels (air, water, instrument, etc.), external scope surtaces and com-
ponents be thoroughly rinsed with clean water to remove residual detergent solution.

6) Prior to rinsing, attach the previously removed adapters and components, and purge all internal channels with
air (using a syringe) to expel residual detergent solution out from each channel.
7) Using clean water, immerse the entire endoscope as well as all removed components and thoroughly rinse all
items.
8) With all cleaning adapters, including irrigation tube, still attached to the endoscope, flush all previously air
purged channels with 200 mL clean water. All internal channels must be thoroughly rinsed to remove resid-
ual detergent and debris. Then remove the endoscope and its components from the solution.
9) Rinse water remaining within the channels should be purged using air to prevent dilution and/or adulteration
of antimicrobial agents to be used in the subsequent disinfection or sterilization process.

NOTE:
70% alcohol for disinfection followed by compressed air, not greater than 165 kPa (1.69 kg/cm2, 24 PSI),
may be used to facilitate drying.

10) Gently dry all external surfaces of the endoscope with a soft gauze or the like. Dry the objective lens with a
cotton tip applicator. Do not put tension on the insertion tube on the endoscope while drying since the outer
cover of the bending section may be excessively stretched.

WARNING:
Prior to disinfection or sterilization, it is imperative that any solutions previously used in the cleaning
process be thoroughly rinsed and dried. Failure to do so, could result in ineffective or incomplete disin-
fection and sterilization.

CAUTION:
Never subject the endoscope to ultrasonic cleaning methods employing high-frequency ultrasound.

- 30 -
4-1-5 HIGH-LEVEL DISINFECTION

Channel Adapters/Components to be attached Disinfecting solution Rinse water


Air/Water Channel OF-B115 50 mL or more 100 mL or more
OF-G17
Biopsy/Suction Channel OF-B115 100 mL or more 150 mL or more
Rubber Inlet Seal
Before any attempt is made to disinfect the endoscope, the complete cleaning procedure described elsewhere in
this manual must have been completed. Prior to high-level disinfection, the end user should confirm the minimum ef-
fective concentration (MEC) of reused disinfectant as per the manufacturer’s instructions. Before complete immersion
in any disinfecting solution the endoscope should have been “Leak Tested” as described elsewhere in this manual.

WARNING:
It is imperative that flexible endoscopes and other semi-critical devices be reprocessed using at least
high-level disinfection with a legally marketed sterilant/disinfectant. Only legally marketed endoscope
automated reprocessing devices/systems whose device specific claims have been validated by the AER
manufacturer and/or anti-microbial agents which have been tested and found to be compatible by PEN-
TAX should be used with PENTAX products.
Generally speaking, “2%” and “3.2%” alkaline glutaraldehyde solutions which have been FDA cleared with
High-Level Disinfection and/or Sterilization claims are recommended. It should be noted that the actual per-
centage of active ingredient (glutaraldehyde) in these solutions, as per their product label, may vary from
the generic and traditional terms “2% glutaraldehyde” and/or “3.2% glutaraldehyde”.
For specific brands of compatible disinfectants/sterilants, please contact your local PENTAX service facility
or sales representative. Please also refer to the inside front cover of this manual regarding infection control.

CAUTION:
PVE Soaking Cap BEFORE IMMERSING, confirm that
a) The ‘Red’ ETO gas sterilization venting cap is
taken OFF.
b) The PVE Soaking cap is securely ON the elec-
ETO Cap trical contacts.

1) The air/water/instrument channel disinfecting adapters


(OF-B115)
consisting of two separate components OF-B115 and
OF-G17 should already be installed on the endoscope
from the previous cleaning procedure.
(OF-G17)
1 Model number OF-G17 incorporates a standard
ANSI luer lock connector to which a syringe or other
device should be attached. Fresh (or reused actively
effective) disinfecting solution should be flushed
through this connector and will simultaneously flow
through both the air and water channels of the scope.
(Please refer to the internal schematics)
2 After the entire scope is completely immersed, and the air and water channels have been filled with disin-
fecting solution, Model Number OF-G17 should be removed.
3 Adhere to proper exposure times for the disinfectant.
4 Confirm that a rubber inlet seal is already attached to each channel inlet during the next step.
5 The suction nipple located on the PVE connector incorporates a standard luer slip fitting to which a 50mL
syringe (or other device) can be attached. Fresh (or reused actively effective) disinfecting solution should
be flushed through or drawn into the entire suction system.

2) As long as the entire instrument is immersed in disin-


DC fecting solution, the cleaning adapters, the syringes
used in previous steps and the rubber inlet seal should
remain removed while the instrument remains fully
under the disinfecting solution. The removal of com-
ponent parts and cleaning adapters from the endoscope
will eliminate the risk of mated surfaces not being
Never remove the soaking cap. exposed to the liquid chemical germicide.

- 31 -
3) While fully immersed, manipulate valve mechanisms and inject disinfectant via syringe into/through remov-
able scope components. This will remove entrapped air bubbles that could inhibit contact of solution with
component surfaces and provide for better exposure of surfaces to germicide. Make sure disinfectant is
injected into/through the rubber inlet seal. The endoscope’s component parts, including all valves should
remain in contact with the disinfecting solution for the time period recommended by the manufacturer of the
solution and accepted by the user as appropriate to accomplish the desired clinical effect.
4) After the endoscope and its component parts have been in contact with the disinfecting solution for an appro-
priate time, flush all channels with air to purge remaining disinfectant, then remove the scope and its compo-
nents from the solution. Thoroughly rinse the entire endoscope and all its components with sterile water.
5) a)A 20 mL syringe filled with sterile water should be attached to adapter OF-G17 to flush disinfecting solu-
tion from the air and water channels of the scope. Repeat it four times. Fill a syringe with air and flush
through the air and water channels several times to force any residual water out of the tubing and nozzles.
Dry thoroughly.
b)With the A/W instrument channel cleaning adapter OF-B115 attached, rinse the entire suction system, in-
cluding both instrument channel (s) with sterilie water (200 mL or more). Flush air through the instrument
channel(s) several times to remove residual water. Dry thoroughly.

NOTE:
Ideally, all final rinses should be made with sterile water or bacteria-free water whose microbial quality
has been confirmed via monitoring. After water rinsing, 70% alcohol should be flushed through all chan-
nels, followed by compressed air, not greater than 165 kPa (24 PSI) to facilitate drying. Channel cleaning
adapters should be used for alcohol flushing/rinsing and forced air drying. The outer surfaces of the
endoscopes can be dried by gently wiping with a sterile gauze or lint-free cloth saturated with alcohol.
Regardless of the liquid chemical germicide (sterilant or high-level disinfectant) and/or the quality of the
rinse water used, a dry instrument accomplished by a final alcohol rinse followed by forced air is essential
to prevent bacteria colonization and/or infections associated with waterborne microorganisms. Such
infections are more likely to occur when wet/contaminated instruments are used on patients whose
immune systems are compromised or suppressed or when these devices are used in anatomical areas
considered sterile and/or susceptible to these organisms.

6) Following an alcohol rinse, the following steps may be performed to aid in the drying process.
a) Remove the air/water/instrument channel cleaning adapters, OF-B115, air/water/channel adapter OF-G17,
reinstall the previously reprocessed scope components.
b) Attach the scope to an external suction source, and aspirate air through the channel of the scope to remove
any residual alcohol and to air dry channel surfaces.

A/W
c) Remove the soaking cap and attach the scope to the
processor with the air pump turned ON to its
DRAIN
HIGHEST pressure setting and the drain lever of
the water bottle set in the DRAIN position, depress
the air/water valve of the scope fully until all alco-
hol has been discharged from the scope. Thor-
oughly drain all alcohol from the air channel as
well by covering the hole in the air/water valve.
Repeat until no moisture or alcohol is seen exiting
the scope or distal tip.
70% alcohol should be flushed through all chan-
nels, followed by compressed air, not greater
than 165 kPa (1,69 kg/cm2, 24 PSI), to facilitate
drying.
7) Gently dry all external surfaces of the Endoscope with a soft gauze or the like. Do not put tension on the
insertion tube while drying since the outer cover of the bending section may be excessively stretched. Dry the
objective lens with a cotton-tip applicator.

WARNING:
If the endoscope is to be stored after reprocessing, detach removable valves, components, etc. All
channels should be completely dry before storage.

- 32 -
4-1-6 STERILIZATION
Before any attempt is made to sterilize the endoscope, the complete cleaning procedure as described elsewhere in
this manual must have been completed.

CAUTION:
NEVER place the endoscope in a steam autoclave!! NEVER subject the endoscope to ultrasonic clean-
ing methods employing high-frequency ultrasound!! Follow provided ETO gas sterilization Parameters.

A) Ethylene Oxide Gas Sterilization (Recommended)


Ethylene Oxide (ETO) Gas Sterilization can be performed on these Endoscopes, provided the following special
instructions, which may differ from other endoscopes, are followed to ensure the proper performance of the instru-
ment. Adhere to the sterilization manufacturer’s instructions and always use a biological indicator.
1) The endoscope must first have been properly cleaned
and thoroughly dried according to the instructions in
this manual and each of the component parts such as
air/water valve, suction control valve, rubber inlet seal,
etc. should be removed.
a)

b)

CAUTION:
Prior to placing these Endoscopes in a Gas Sterilizer or Aeration Chamber:
a) The ‘Red’ ETO gas sterilization venting cap MUST be “ON” securely.
b) The PVE soaking cap should be “OFF” the electrical contacts.
This is opposite of the immersion instructions.

WARNING:
Failure to thoroughly dry all surface areas could result in incomplete or ineffective sterilization. Moisture
could prevent contact of the ETO gas with the actual contaminated surfaces.

2) The following parameters for Ethylene Oxide Gas Sterilization are proposed.
20:80 ETO/CO2 10:90 ETO/HCFC
Temperature: 55 °C (131 °F) 55 °C (131 °F)
Relative Humidity: 50% 50%
Vacuum: 71 kPa 71 kPa
Pressure (Start): 69 kPa 97 kPa
EO Concentration: 450 mg/L 600 mg/L
Pre-Conditioning: 1 hour 1 hour
Gas Exposure Time: 5 hours 5 hours
Aeration: 12 hours at 55 °C (131 °F) 12 hours at 55 °C (131 °F)
3) Following ETO Gas Sterilization, aeration time of 72 hours at room temperature is required.
4) Aeration Chamber: To shorten the aeration time to 12 hours, an aeration chamber may be used, provided the
temperature does not exceed 55 °C (131 °F).

CAUTION:
Prior to placing these Endoscopes in an aeration chamber the ‘RED’ ETO Gas Sterilization Venting Cap
MUST be “ON” securely and the PVE soaking Cap should be “OFF” the electrical contacts.

B) Other Methods of Sterilization


Other types of cleaning, disinfection and/or sterilization systems/processes are available for the reprocessing
of medical devices. However, due to the heat sensitive nature and/or the specific biocompatible materials used
in the construction of flexible endoscopes, some of these marketed systems/processes/ solutions could have
detrimental effects on flexible endoscopes.
To avoid the potential for instrument damage and/or patient injury, confirm the compatibility of such reproc-
essing systems/solutions with your local PENTAX dealer prior to use with any PENTAX products.
Prior to using other methods, confirm specific claims of any sterilization methods/processes and ensure man-
ufacturer of such processes has performed microbiological studies that support its claims of achieving sterili-
zation of those specific flexible endoscopes.

- 33 -
4-2. ENDOSCOPIC ACCESSORY INSTRUMENT (EAIs) AND ENDOSCOPE COMPONENTS

4-2-1 CLEANING

NOTE:
Not all manufacturers of automated endoscope reprocessors (AERs) make specific claims nor provide
special instructions for reprocessing all of the removable scope components that are integral to the safe
and effective operation of flexible endoscopes. Therefore, should the AER manufacturer’s instructions
not specifically address reprocessing of any particular scope component (air/water valve, suction valve,
irrigation valve, inlet seal, irrigation tube, check-valve, selector mechanism, etc.) in the AER, then those
components must be reprocessed manually as described in this instructions. Prior to use, check with
each AER manufacturer as to their specific claims with respect to reprocessing individual endoscope
components.

1) Reuseable accessories such as forceps should be cleaned immediately after each use since dried blood,
mucous, or other debris may cause damage to the instrument and render the mechanism inoperable, or may
interfere with the ability of the user to reprocess the device.
2) Place the forceps in a basin with enzymatic detergent solution being careful not to tightly coil or kink the
flexible shaft.
3) Clean the handle and flexible shaft by gently wiping with a soft gauze or the like. The biopsy cups and pivot
pin area should be carefully and gently cleaned with a soft brush. Removal components such as air/water,
suction valve, etc. should be manipulated and detergent injected directly into/onto components surfaces.
4) Rinse all residual detergent from the forceps by immersing the entire forceps under clean water and manipu-
lating the handle and cup mechanism.
5) Ultrasonic cleaning of forceps and similar accessories is then recommended, provided the manufacturer’s
instructions and the parameters below are followed:
Frequency Range 44 kHz ± 6%
Time 5 minutes

DO NOT use caustic or abrasive solutions in the ultrasonic cleaner.

CAUTION:
NEVER use ultrasonic cleaning methods with high-frequency ultrasound on the endoscope itself.

NOTE:
It is imperative that ultrasonic cleaning of the biopsy forceps, other endoscopic accessory instruments
and other components be performed PRIOR to steam sterilization. Only those PENTAX accessories
identified by their pink colored handle, labeled as being autoclavable or identified below may be sub-
jected to steam autoclaving.

6) After cleaning and thorough rinsing, the endoscopic accessory instruments and components should be gently
dried using a soft gauze or the like. Avoid tight coiling or kinking and do not put tension on the flexible shaft
of the forceps and similar endoscopic accessory instruments.

NOTE:
All detergent must be removed from the inner mechanism of the forceps and individual scope compo-
nents. Any detergent that remains after the water evaporates may cause increased friction that may
render the mechanism inoperable. Residual detergent may also interfere in the subsequent sterilization
process.

NOTE:
Other PENTAX reusable accessories (channel cleaning adapters, cleaning brushes, bite block, etc.) and
scope components (rubber inlet seals, air/water and suction control valves, etc. not specifically identified
previously should be cleaned in a similar manner as above. Ultrasonic cleaning methods are recom-
mended for accessories and scope components whose entire surfaces are not easily accessible by man-
ual cleaning. Heavily soiled components such as valve mechanisms, inlet seals, etc. should be
ultrasonically cleaned prior to subsequent high-level disinfection or sterilization. For the specific items,
please refer to the separate operation manual.

- 34 -
NOTE:
The following endoscopic accessory instruments and scope components may be subjected to ultrasonic
cleaning methods:
• PENTAX biopsy forceps • PENTAX A/W feeding valve OF-B121
• PENTAX cleaning brushes with pink handle • PENTAX check-valve OE-C14
• PENTAX bite block OF-Z5 • PENTAX A/W instrument channel cleaning
• PENTAX water bottle OS-H4 adapter OF-B115
• PENTAX suction control valve OF-B120 • PENTAX rubber inlet seal OF-B173

- 35 -
4-2-2 DISINFECTION

WARNING:
Current infection control guidelines require that biopsy forceps and similar endoscopic accessory instru-
ments which enter sterile tissue or the vascular system or break the mucosal barrier must be sterilized
before each patient use. It is recommended that any endoscopic accessory instruments intended for use
in the biliary tract be subjected to an appropriate sterilization process. For patient contact endoscopic
accessories, follow the specific and detailed reprocessing instructions supplied with each product.

Before any attempt is made to disinfect endoscopic accessory instruments and/or scope components such as bite
block, air/water and suction control valve, brushes, etc. the complete cleaning procedure as described elsewhere
in this manual, must have been completed. Heavily soiled components such as valve mechanisms, inlet seals, etc.
should be ultrasonically cleaned prior to subsequent high-level disinfection.
1) The entire accessory or component should be immersed in disinfecting solution.
2) Accessory and component surfaces should remain in contact with the disinfecting solution for the time period
recommended by the manufacturer of the solution and accepted by the user as appropriate. To ensure better
contact, manipulate components such as valves while injecting disinfectant into/onto components surfaces.
3) After the item has been in contact with the disinfecting solution for the appropriate amount of time, remove it
from the solution.
4) Rinse all residual disinfecting solution from the accessory/component by immersing it under sterile water.
5) After thoroughly rinsing, the items should be gently dried using a soft gauze or the like. Compressed air may
also be used to facilitate drying.

NOTE:
Ideally, all final rinses should be made with sterile water or bacteria-free water whose microbial quality
has been confirmed via monitoring. After water rinsing, 70% alcohol for disinfection should be flushed
through tubing of the endoscopic accessory instrument, followed by compressed air, not greater than
165 kPa (24 PSI) to facilitate drying. The outer surfaces can be dried by gently wiping with a sterile gauze
or lint-free cloth saturated with alcohol.
Regardless of the quality of the rinse water used, a dry instrument accomplished by a final alcohol rinse
followed by forced air is essential to prevent bacteria colonization and/or infections associated with
waterborne microorganisms. Such infections are more likely to occur when wet/contaminated instru-
ments are used on patients whose immune systems are compromised or suppressed or when these
devices are used in anatomical areas considered sterile and/or susceptible to these organisms.

WARNING:
The addition of defoaming agents to the water supply and/or automated reprocessing system is NOT
recommended. Due to their nature, these silicone based agents cling tenaciously to surfaces. Unless
they are rinsed very thoroughly, a “barrier” could be created which could reduce the effectiveness of the
disinfection/sterilization process. Additionally, repeated use of such defoamers could eventually lead to
residual silicone build up resulting in equipment malfunction such as clogged air and water channels.

- 36 -
4-2-3 STERILIZATION

WARNING:
Current infection control guidelines require that biopsy forceps and similar accessories which enter ster-
ile tissue or vascular system or break the mucosal barrier must be sterilized before each patient use.
It is recommended that any endoscopic accessary instruments intended for use in the biliary tract be
subjected to an appropriate sterilization process. For patient contact endoscopic accessories, follow the
specific and detailed reprocessing instructions supplied with each product.

Before any attempt is made to sterilize the accessories and/or individual scope components, the complete cleaning
procedure as described elsewhere in this manual must have been completed. Heavily soiled components such as
valve mechanisms, inlet seals, etc. should be ultrasonically cleaned prior to subsequent sterilization.

CAUTION:
Use only the type of packaging material and package configuration as recommended by the manufac-
turer of the sterilizer. Use appropriate heat process indicators and/or biological monitors as recom-
mended by the manufacturer of the sterilizer.

NOTE:
The following sterilization parameters are only valid with sterilization equipment that is properly main-
tained and calibrated.

• Steam Sterilization (Autoclaving) Recommended

NOTE:
The PENTAX accessories and scope components listed on page 28 may be subjected to Steam Auto-
claving:

1) Prior to steam sterilization, all autoclavable endoscopic accessary instruments and endoscopic components
previously identified should be thoroughly cleaned using manual and ultrasonic cleaning methods as
described elsewhere in this manual.
2) Autoclaving can then be performed under the following conditions:

Sterilizer Type: Prevacuum


Temperature: 132 ~ 135 °C (270 ~ 275 °F)
Time: 5 minutes

CAUTION:
Never place the endoscope in a steam autoclave nor subject it to high-frequency ultrasonic cleaning
methods!

• ETO GAS sterilization


1) ETO Gas Sterilization can be performed on these accessories and/or components, provided they have first
been properly cleaned and thoroughly dried.
2) Following ETO GAS Sterilization, aeration is required.

NOTE:
For ethylene oxide sterilization of PENTAX accessories and endoscope components, follow the same
ETO parameters as for PENTAX endoscopes.

- 37 -
4-3. WATER BOTTLE ASSEMBLY

Each water bottle assembly should be cleaned and sterilized at least on a daily basis or more frequently depending
upon the patient and/or type of endoscopic procedure. As with all endoscopic accessories, prior to sterilization,
water bottle assemblies must be scrupulously cleaned. Failure to do so could result in incomplete or ineffective
sterilization.

WARNING:
The addition of defoaming agents to the water supply is NOT recommended. Due to their nature, these
silicone based agents cling tenaciously to surfaces. Unless they are rinsed very thoroughly, a “barrier”
could be created which could reduce the effectiveness of the disinfection/sterilization process. Addition-
ally, repeated use of such defoamers could eventually lead to residual silicone build up resulting in
equipment malfunction such as clogged air and/or water channels.

NOTE:
Take care in handling the water bottle. DO NOT carry the bottle by the A/W Connector or the hose. When
the bottle cap has been removed from the bottle, be careful in handling the water feeding stem.

I. Cleaning of Water Bottle

Reprocessing At-A-Glance (2 options)

Autoclave

Option A Manual Ultrasonic Rinse and Steam


Cooling
(Recommended) Pre-Cleaning Cleaning Dry Sterilize

ETO Sterilization

Manual Ultrasonic Rinse and ETO


Option B Aeration
Pre-Cleaning Cleaning Dry Sterilize

1) After use, the entire bottle assembly (water container, cap and tubing) should be washed with clean
running water and a dampened gauze or scrub brush.
2) Ultrasonic cleaning of the entire water bottle assembly is then recommended to access difficult to
reach areas, provided the operating frequency is 44 kHz ± 6% for a period of at least 5 minutes.
3) After washing with a cleaning solution, all surfaces of the water bottle assembly should be thoroughly
rinsed with clean water and dried. Gauze or cloth can be used to wipe dry most surfaces. Compressed
air and 70% alcohol should be used to facilitate drying of hard to reach areas.

II. Sterilization of Water Bottle


Before any attempt is made to sterilize the water bottle assembly, the complete cleaning procedure described
above must have been completed. Failure to scrupulously clean the water bottle prior to the sterilizing process,
could result in incomplete or ineffective sterilization.

CAUTION:
Use only the type of packaging material and package configuration as recommended by the manufac-
turer of the sterilizer. Use appropriate heat process indicators and/or biological monitors as recom-
mended by the manufacturer of the sterilizer.

NOTE:
The following sterilization parameters are only valid with sterilization equipment that is properly main-
tained and calibrated.

- 38 -
Option A: Steam Sterilization (Autoclaving)
PENTAX water bottle assemblies, Model OS-H4 have been designed to withstand high-pressure steam steri-
lization procedures providing the following parameters are adhered to:
Sterilizer Type : Prevacuum Temperature : 132 ~ 135 °C (270 ~ 275 °F) Time : 5 minutes
During steam sterilization, ensure that the cap and tubing section have been removed from the water bottle
container.
Make sure that the drain lever on the water bottle cap has been set on the A/W position (upright).

Option B: Ethylene-Oxide Gas Sterilization


1) Ethylene Oxide (ETO) gas sterilization can be performed on PENTAX water bottles, provided they have first
been properly cleaned and thoroughly dried.

WARNING:
Failure to thoroughly dry all surface areas could result in incomplete or ineffective sterilization. Moisture
could prevent contact of the ETO gas with the actual contaminated surfaces.

2) The parameters below are proposed for ETO sterilization.

20:80 ETO/CO2 10:90 ETO/HCFC


Temperature: 55 °C (131 °F) 55 °C (131 °F)
Relative Humidity: 50% 50%
Vacuum: 71 kPa 71 kPa
Pressure (Start): 69 kPa 97 kPa
EO Concentration: 450 mg/L 600 mg/L
Pre-Conditioning: 1 hour 1 hour
Gas Exposure Time: 5 hours 5 hours
Aeration: 12 hours at 55 °C (131 °F) 12 hours at 55 °C (131 °F)
3) Following ETO sterilization, aeration time of 72 hours is required. To shorten the aeration time to 12 hours,
an aeration chamber may be used, provided the temperature does not exceed 55 °C (131 °F).

Caution:
To avoid disconnection and/or bursting of the internal tubing, always set the lever to the Air/Water posi-
tion (upright) and use less than 165 kPa (1.69 kg/cm2, 24 PSI) air pressure during forced air drying.

NOTE:
Do not use the new PENTAX OS-H4 water bottle cap with the older OS-H2 water container/bottle.
Although the cap may appear to fit onto the bottle, air may escape resulting in insufficient pressure and
flow of air and water during the endoscopic procedure. Both the PENTAX water bottle cap and bottle
(container) are identified by their appropriate model designation. Ensure that an OS-H4 cap is used only
with the OS-H4 water container/bottle. Do not overtighten the bottle cap. Overtighting can cause the bot-
tle cap to break.

- 39 -
5. POST REPROCESSING
1) Following reprocessing, the endoscope may either be reused or placed in storage.

CAUTION:
When utilizing chemo-thermal processes for reprocessing PENTAX endoscopes, the instruments should
be allowed to return to room temperature prior to use and/or further handling.

2) Prior to reuse, ensure that instrument has been properly inspected and fully prepared for the next clinical pro-
cedure.
3) Prior to storage, ensure that all internal channels, scope components, instrument surfaces and accessories are
thoroughly dry.
4) A cotton tipped applicator moistened with 70% alcohol for disinfection may be used to carefully remove any
films or residues left upon the lens surfaces, such as the distal objective lens.
5) The endoscope should be hung in a clean, dry, well-ventilated storage cabinet at room temperature. The
insertion tube and umbilical cable should be hung and kept as straight as possible during storage.

CAUTION:
• Make sure that all removable components such as the air/water valve, suction valve, forward water jet
components, suction selector mechanisms, rubber inlet seal(s), and rubber distal tip are detached
from the scope. This will allow for better air circulation throughout the internal channels to ensure
thorough drying.
• Never store the endoscope, its components and accessories in the carrying case, as this type of dark,
humid and unventilated environment is conducive to bacteria colonization, which increases the risk of
cross contamination. These cases are intended for transportation of the instrument, not storage.
• Never store the endoscope in areas of high humidity, high temperatures or in direct exposure to
sunlight or X-rays.
• Avoid storage of the scope in cabinets, which have any sharp edges, exposed nails/screws, etc.
Contact with sharp objects can puncture, scratch or otherwise damage the endoscope.

5-1. SERVICING

Prior to returning any instrument for repair to PENTAX, the instrument should first undergo appropriate reproc-
essing/decontamination procedures for the purpose of infection control. Check with your local PENTAX service
facility for more details.
1) All instruments requiring repair should be shipped in the original carrying case with appropriate packing
along with comments describing the instrument damage and complaint.
2) A repair purchase order number, contact name and phone number of the individual responsible for authoriz-
ing repairs, as well as shipping address should be included.
3) The “Red” ETO gas sterilization venting cap should be attached to the instrument if it will be shipped by air
freight.
4) Any accessories and/or scope component potentially related to the scope damage or complaint should also be
returned with the endoscope.
5) Soaking caps should also be returned with the scope to check/confirm the integrity of their watertight seal.
6) After servicing, all endoscopes must be reprocessed prior to patient use.

WARNING:
Instrument repairs should only be performed by an authorized PENTAX service facility. PENTAX
assumes no liability for any patient/user injury, instrument damage or malfunction, or REPROCESSING
FAILURE due to repairs made by unauthorized personnel.

7) At the end of life cycle do not discard this instrument in general waste.
Contact your local PENTAX distributor for correct disposal and recycling.

- 40 -
5-2. CARE AND MAINTENANCE TIPS

Flexible endoscopes have been an invaluable tool in the medical community’s armamentarium to successfully di-
agnose and treat a wide variety of illnesses in patients for several decades. Perhaps due to their longevity and pro-
gressive design changes over the years which have simplified their use, flexible endoscopes have been somewhat
taken for granted and have erroneously not been considered highly technological medical devices.

In fact, current generation flexible endoscopes although easier to clinically use, are much more sophisticated than
ever. Special reprocessing instructions must be followed to ensure the instruments are patient ready and patient
safe. Special care and handling must be exercised and practiced to prevent instrument malfunction and prolong the
reliability of the endoscope.

The burden of responsibility to ensure safe and reliably functioning instruments is left in the hands of the health-
care professionals who actually care for and reprocess flexible endoscopes.

Naturally, equipment manufacturers share in this responsibility and tremendous efforts have been made in design-
ing instruments which could be reprocessed and maintained as easy as possible. However, due to the nature of their
use and application, flexible endoscopes must be subjected to special cleaning procedures, followed by a disinfec-
tion or sterilization process after each and every patient use.

To highlight and simplify, what may appear to some as being complicated maintenance and reprocessing instruc-
tions, PENTAX strongly recommends the users review the following suggestions and advice on the care and main-
tenance of your PENTAX flexible endoscopes.

These tips, particularly those involving scope reprocessing should not be construed as “shortcuts” and are not in-
tended as substitute directions for complete instructions found elsewhere in the owner’s manual.

** When utilizing chemo-thermal processes for reprocessing PENTAX endoscopes, the instruments should be
allowed to return to room temperature prior to use and/or further handling.
** Make sure that the air/water/instrument channel cleaning adapter is securely attached to the top of the air/water
and suction valve cylinders.
** Avoid soaking of the endoscope with accessories (forceps, injection or aspiration needles, etc.) or any sharp
edged objects which could inadvertently scratch or cut the distal bending section sheath. (Subsequent flexing
back and forth of the rubber sheath could eventually stretch the scratched rubber until a pinhole and leak de-
velops.)
** Exposure to a compatible enzymatic detergent is essential to thorough cleaning of all surfaces of the endo-
scope. Rinsing and drying after cleaning is imperative to prevent dilution and/or adulteration of the disinfect-
ant/sterilant.
** Do not reuse disposable endoscope accessory instruments (EAIs) or scope component intended for single pa-
tient or one time use.
** Do not expose the endoscope or endoscope accessory instruments (EAIs) or scope component to harsh chem-
ical solutions. Strictly adhere to exposure times recommended by the manufacturers of compatible solutions.
** Not all manufacturers of automated endoscope reprocessors (AERs) make specific claims nor provide special
instructions for reprocessing all of the removable scope components that are integral to the safe and effective
operation of flexible endoscopes. Therefore, should the AER manufacturer’s instructions not specifcally ad-
dress reprocessing of any particular scope component (air/water valve, suction valve, inlet seal, etc.) in the
AER, then those components must be reprocessed manually as descrived in PENTAX instructions/labeling.
Prior to use, check with each AER manufacturer as to their specific claims with respect to reprocessing indi-
vidual endoscope components.
** Avoid contact of any flexible portion of the endoscope with any sharp edge objects (bed frames, table top cor-
ners, sink drains, accessories hanging in storage cabinets, etc.) at any time during the handling, reprocessing
or storage of the endoscope.

- 41 -
** Avoid stretching of the bending section rubber sheath at the distal portion of the scope. During mechanical
cleaning with a dampened gauze, do not use excessive force. A gentle back and forth wiping motion should
be sufficient to remove gross debris. Subsequent soaking in an enzymatic detergent will clean the remainder
of debris.
** Disinfectants and sterilants are toxic substances by nature. All residual solution must be thoroughly rinsed and
dried prior to each patient use.
** The key to preventing clogged air water channels/nozzles is to immediately flush the channels with either air
pressure or fluid/detergent right after removal from the patient. This should be followed by brushing of each
accessible channel and then soaking with an enzymatic detergent.
** Avoid attempting to remove or unscrew scope components which should not be removed. Parts such as the
distal portion of the light guide plug, any inlet ports and any rubber strain reliefs on either the insertion tube
or umbilical cable are essential to the watertight integrity of the instrument. Removal or loosening of these
components and subsequent immersion could lead to fluid invasion into the endoscope and/or compromise
effective reprocessing.
** Check for any sharp edges on all surfaces of an automated cleaning/reprocessing unit which may come in con-
tact with an endoscope. Some units may have sharp edged wire mesh filters and baskets or inlet/outlet ports
which could damage your scope.
** Do NOT overtighten the cap to the water bottle assembly. The metal pipe at the top of the PENTAX water
bottle assembly functions as an inlet port for air from the light source. This inlet pipe should not be used as a
leverage tool to tighten the cap to the water container. Overtightening could cause the plastic cap to crack.
** Do NOT introduce air bubbles into the scope’s internal channels during flushing of cleaning and/or disinfect-
ing/sterilizing solutions as these bubbles could interfere in the effectiveness of the disinfection/sterilization
process.
** Do not store the endoscope, its components or accessories in the carrying case as this type of dark, humid and
unventilated environment is conducive to bacteria colonization which increases the risk of cross-contamina-
tion.
** Prior to each use, check the condition of all endoscope accessary instruments.
Do NOT use any accessories with kinked or bent flexible shafts.
Do NOT use forceps with misaligned cups and/or bent needles/spikes.
Do NOT use aspiration or injection needles which are not retractable or whose sharp tips can not be protected.
Do NOT use cleaning brushes without smooth or rounded distal tips.
Do NOT use instruments with exceptionally long rigid sections or whose outer diameter restricts pasage
through the instrument channel/channel inlet.
Use of any of the above accessories could result in channel damage and costly repairs.
** Verification of the effective level of glutaraldehyde (via test strips or similar methods) is recommended to en-
sure potency of glutaraldehyde to achieve high-level disinfection and/or sterilization.

WARNING:
Instrument repairs should only be performed by an authorized PENTAX service facility. PENTAX
assumes no liability for any patient/user injury, instrument damage or malfunctions, or REPROCESS-
ING FAILURE due to repairs made by unauthorized personnel.

- 42 -
MEMO

- 43 -
MEMO

- 44 -
SPECIFICATIONS

EC-380MK2p/ EC-380FKp/
EG-290Kp EC-380LKp
EC-380MKp EC-380FK2p
Direction of View Forward
Field of View 140°
Depth of Field 5 ~ 100 mm 3 ~ 100 mm
Tip Deflection Up - Down 210° ~ 120° 180° ~ 180°
Right - Left 120° ~ 120° 160° ~ 160°
Rigid Distal Diameter ø 9.8 mm ø 13.4 mm
Insertion Tube Diameter ø 9.8 mm ø 13.2 mm
Diameter of Instrument Channel ø 2.8 mm ø 3.8 mm
Insertion Tube Working Length 1,050 mm 1,300 mm 1,500 mm 1,700 mm
Total Length 1,373 mm 1,623 mm 1,823 mm 2,023 mm
Operation envi- Ambient temperature 10 ~ 40 °C
ronment Relative humidity 30 ~ 85%
Air pressure 700 ~ 1060 hPa
Storage Ambient temperature -20 ~ 60 °C
environment Relative humidity 0 ~ 85%
Air pressure 700 ~ 1060 hPa
Degree of protection against electric shock BF type

NOTE: Specifications are subject to change without prior notice and without any obligation on the part of the man-
ufacturer.

DISTAL END

EG-290Kp EC-380MKp/MK2p/FKp/FK2p/LKp
OBJECTIVE LENS OBJECTIVE LENS
LIGHT GUIDE
LIGHT GUIDE LIGHT
GUIDE LIGHT GUIDE

AIR/WATER
INSTRUMENT AIR/WATER NOZZLE
CHANNEL INSTRUMENT
NOZZLE CHANNEL

- 45 -
FOR EUROPEAN COUNTRIES
DECLARATION OF CONFORMITY
We, PENTAX Corporation
2-36-9, Maeno-cho, Itabashi-ku, 174-8639 Tokyo, Japan
declare under our sole responsibility, that the product:
Product Name : VIDEO GI SCOPES
0197
Model Number(S) : EG-290Kp, EC-380MK2p, EC-380FKp, EC-380FK2p,
EC-380LKp
conforms to the applicable provisions of the Medical Devices Directive 93/42/EEC.
This declaration is made on the basis of: EC quality system approval by TÜV Rheinland
No. 0197 in accordance with Annex II of this Directive.

PENTAX Europe GmbH


Importer into EEA

PENTAX Corporation
Manufacturer

NOTICE
These instruments are used with Class B Medical Equipment (specified EN55011) and are intended for hospital or health
care districts.
Together, these endoscopes and the compatible processor comply with EN 60601-1-2 (2002) for EU, IEC 60601-1-2
(2001) for other countries.
When used in clinical or residential areas near radio and TV receiver units, these instruments may be subjected to radio
interference.
To avoid and resolve adverse electromagnetic effects, do NOT operate these instruments near the RF energy equipment.

PENTAX Corporation Explanations of the symbols used in the


2-36-9, Maeno-cho, Itabashi-ku, operating manual, on the devices or on the
174-8639 Tokyo, Japan packagings.
Tel. : ++81-3-3960 5155
Fax : ++81-3-5392 6724
PENTAX Europe GmbH Symbol for "MANUFACTURER"
Julius-Vosseler-Strasse 104,
22527 Hamburg, Deutschland
Tel. : ++49-40-56 192 0
Fax : ++49-40-56 042 13 Symbol for "DATE OF MANUFACTURE"
PENTAX U.K. Ltd
Pentax House, Heron Drive,
Langley SLOUGH SL3 8PN, Great Britain Symbol for "AUTHORISED
Tel. : ++44-1-75 37 92 792 REPRESENTATIVE"
Fax : ++44-1-75 37 92 794
PENTAX France S.A.S SN Symbol for "SERIAL NUMBER"
112, quai de Bezons In the case of endoscopes the serial number
B.P. 204 is either on the control body or the light
95106 ARGENTEUIL CEDEX, France guide connector. The serial number always
Tel. : ++33-1-30 25 75 75 has 6 digits preceded by a letter.
Fax : ++33-1-30 25 75 76
www.pentax.fr
Symbol for "BATCH CODE"
PENTAX Benelux
Spinveld 25,
4815 HR Breda, The Netherlands
Tel. : ++31-76-531 30 31
Fax : ++31-76-531 30 00
Specifications are subject to change without Our representative in your area:
notice and without any obligation on the part
of the manufacturer.
MTM 01/09/05/87930XX

Printed in Germany.
R00

You might also like