765 Dosimat: Instructions For Use
765 Dosimat: Instructions For Use
765 Dosimat: Instructions For Use
765 Dosimat
Instructions for Use
8.765.1023
2005.06 ti/dm
Content
Content
1 Overview ...................................................................................................................2
5 Appendix .................................................................................................................51
5.1 Technical specifications............................................................................................ 51
5.2 Warranty and certificates .......................................................................................... 53
5.2.1 Warranty ............................................................................................................. 53
5.2.2 Certificate of Conformity and System Validation................................................ 54
5.3 Connection of a balance........................................................................................... 56
5.4 Connection of a printer ............................................................................................. 56
5.5 Continuous dosing with two Dosimats ..................................................................... 57
5.6 Scope of delivery and ordering designations ........................................................... 59
Index...........................................................................................................................61
765 Dosimat
Overview
Explanation of symbols:
765 Dosimat
1
Overview
1 Overview
1 Exchange unit
Normally the models with automatic cock changeover.
Note:
Choose the volume of the exchange unit in such a way that a volume between
10...100% of the nominal volume is expelled.
2 Display
The 16 digit display shows all important information:
DOS 3.456 ml Mode (DOS = dosing) and dosed volume. Dosimat is in
stand-by position
DOS ↑ 3.456 ml The piston is moving upwards.
DOS ↓ 3.456 ml The piston is moving downwards.
DOS → 3.456 ml The cock is turned to the right.
DOS ← 3.456 ml The cock is turned to the left.
The display of the status ↑ resp.↓ are specially important for very slow dosings
where movements of the piston cannot be clearly identified.
765 Dosimat
2
Overview
9 Power connection
In power supply systems, in which strong HF interferences (transients) are su-
perimposed on the power voltage, the 765 Dosimat should be connected via an
additional power line filter, e.g. METROHM 615 model.
The main cables supplied with the instrument are three-core and equipped with
a plug with an earthing pin. If a different plug has to be fitted, the yellow/green
lead must be connected to the protective earth. Each break in the earthing inside
or outside the instrument can make it a hazard.
When the instrument is opened or if parts of it are removed, certain components
may be live if the instrument is connected to the power line. The power cable
must therefore always be unplugged when certain adjustments are made or
parts replaced.
10 Power switch
Switching on and off Dosimat. The 765 Dosimat is equipped with a non-volatile
memory, i.e. set parameters remain in the working memory if the Dosimat is
switched off and on.
11 Earthing socket
The 765 Dosimat must be grounded correctly and effectively, if necessary
through the separate earthing socket.
15 Identification plate
Indication of model, series and serial number.
765 Dosimat
3
2.1. Keypad
2.1 Keypad
1 2 3
MODE
MODE Selection of mode.
0 . –
765 Dosimat
4
2.1 Keypad
• On entering a negative number, key in minus sign first; <-> is not a "change of sign"
key!
• Changeover between first functions (blank, factor etc.) and digits is done
automatically.
• Terminate parameter entries with <ENTER>.
• Some keys are organized as inquiry drums, i.e. pressing these keys several times, dis-
play shows new inquiries. A new value is stored or a new feature is selected with
<ENTER>. The program then returns to the initial state, the inquiry drum is left. Enter-
ing an inquiry drum, that inquiry, where the drum has been left last time, is displayed
first.
• The Dosimat works with a resolution of 10'000 pulses per burette cylinder volume.
10 mL .001 .010 1 µL
20 mL .002 .020 2 µL
50 mL .005 .050 5 µL
If a volume value is entered which can not be dosed exactly with the exchange unit on
the Dosimat, the value is rounded off to the next possible one and stored accordingly.
The smallest possible increments for 1 and 5 mL cylinders are not displayed. However,
they are dosed and the display is rounded.
• Key <CLEAR> sets parameters to "OFF".
765 Dosimat
5
2.2. Modes
2.2 Modes
MODE The modes are selected by the inquiry drum <MODE> and
confirmed with <ENTER>.
DOS: DOSing
Dosimat is dosing as long as <GO> is pressed. Result calculation
can be activated.
DIS R: DISpensing, Repetitive
Dosimat is dosing a stored dispensing volume if <GO> is pressed,
the burette cylinder is refilled and display reset to 0.000 mL.
DIS C: DISpensing, Cumulative
Dosimat is dosing a stored dispensing volume if <GO> is pressed,
the dispensed volume (V-DIS) remains displayed.
PIP: PIPetting
Aspirating and subsequent expelling of a stored pipetting volume.
DIL: DILuting
Aspirating a stored pipetting volume and subsequent expelling of the
pipetting and diluting volume.
CNT D: CoNTent Dispenser
Preparation of solutions with preselected content.
765 Dosimat
6
2.2 Modes
Example:
Selection of mode "DIS C", cumulative dispensing.
Press <MODE>.
Display shows that mode which has been selected last with key <MODE>, e.g.
DOS .
All modes which are loaded into the working memory by key <MODE> are equipped
with a set of standard parameters:
765 Dosimat
7
2.2. Modes
Calculation values
BLANK
7 b = 0. ml Blank value
Input range: 0...±999.999 mL
FACTOR
8 f = 1. Factor
Input range: 0... ±1E33
SMPL
s = 1. Sample size
9 Input range: 0... ±1E33
Input manually or on-line with balance, see page 56
UNIT
4 unit Unit
Input range: ppm, %, g, mg, g/L, mg/L, mol, mol/L, mL, L,
/pc, none
Result calculation
If one of the calculation values (blank, factor, smpl) is not set to its standard value, a
result is calculated on filling of the Dosimat according to formula:
(dosed volume–blank)*factor
Result = ————————————————
smpl
The result is recalculated on each entry of a calculation value (blank, factor, smpl).
Pressing <CLEAR> display shows the dosed volume in mL. To start a new dosing, press
<GO> twice, pressing once resets the volume in display to 0.000 mL.
Example of print-out
765 Dosimat
8
2.2 Modes
Tandem Dosing
Mode DIS C is suitable for continuous dosing with 2 Dosimats, see page 57.
765 Dosimat
9
2.2. Modes
If the pipetting volume is changed, a new preparation step is always carried out.
Sequence of PIP
Note
765 Dosimat
10
2.2 Modes
• A new air bubble is built with every preparation step ("prep.") e.g. its volume increases.
If you wish to keep the volume of the air bubble expel it in mode DOS before changing
V-PIP.
• For best pipetting results we recommend exchange units with volumes ≤ 20 mL and
pipetting equipment 6.5611.000.
• The aspirating and expelling rates should not be higher than 20 mL/min.
• Hold tubing tip in an angle of app. 45° to the vessel wall during pipetting. Just the
same as you do with glass pipettes!
• The vessel, containing the liquid you want to pipette should stand on the same level as
the vessel into which you are going to expel the liquid in order to ascertain app. the
same level of the pipetting tubing during work.
765 Dosimat
11
2.2. Modes
Sequence of DIL
Notes
• If you wish to change V-PIP, it is best to change it during filling in the preparation step,
i.e. when display shows DIL ↓ prep..
• If V-PIP is changed at another time, a new preparation step is carried out, which
changes the volume of the air bubble. The first dilution after such a change could be
erroneous and should be discarded. Another possibility is to expel the air bubble in
mode DOS and to start Mode DIL from the beginning. V-DIL can be changed at any
time without a new preparation step.
765 Dosimat
12
2.2 Modes
Content entries which can be implemented in the CNT D mode are summarized below
and designated with a bold frame
Sequence of CNT D
↓
s 1 g Input of the weight manually or via a balance,
↓ see page 56.
Input <ENTER>
↓
add V XXX.XXX ml The calculated volume is displayed and expelled
↓ ↓ with <GO>.
<GO> <CLEAR> With <CLEAR> the values for "cnt" and/or "s" can
↓ be changed. This gives an idea of the
dosing V approximate weighing.
Pressing <MODE>, the mode can be changed.
765 Dosimat
13
2.2. Modes
The formulae for calculation of the volume to be dispensed "add V" are shown in the
following table, with
mol/L f·s·103
Amount-of-substance cnt·M
concentration mmol/L f·s·106
cnt·M
g/L f·s·103
Mass concentration cnt
mg/L f·s·106
cnt
% f·s(102-cnt)
Mass fraction cnt·dens
ppm f·s(106-cnt)
cnt dens
mol/kg s·103
Molality cnt·M·dens
mmol/kg s·106
cnt·M·dens
Application of factor f
n*M(A) m*M(B)
f = ———— or f = ————
M(AnBm) M(AnBm)
765 Dosimat
14
2.2 Modes
The following table shows factors for the most common ionic standards:
765 Dosimat
15
2.2. Modes
Since the volume of the solvent V0 is dispensed in the operational method of the CNT D
mode, higher concentrations require a correction factor which takes the difference
between V0 and V (volume of the solution) into consideration:
V0
f = —–
V
This factor can be determined with the Dosimat in the DOS mode:
A solution of the desired concentration is prepared in the conventional manner in a
volumetric flask by dispensing the solvent with the aid of the Dosimat up to the mark of
the flask (V0). If the volume V of the volumetric flask is inputted in the calculation
parameter "s", the factor f is calculated directly by the Dosimat and appears on the
display.
The factor f determined in this manner holds for the appropriate substance/solvent pair in
the measured concentration range with the possibility of linear extrapolations up to
concentrations of ca. 1 mol/L.
765 Dosimat
16
2.3 User memory
Up to 10 modes, complete with their user selected specific parameters, can be stored in
the user memory.
<MODE> <STORE>
<RECALL>
<RECALL><X><ENTER>
Loading a mode from the user memory into the working
memory.
765 Dosimat
17
2.4. Special settings
Special settings can be executed by pressing keys <0> and <—> simultaneously. The
display shows Sys. Soft-Reset . Press key <0> and keep it pressed until special key 0..6
appears in the display.
Press key 1...6. Pressing key <CLEAR> once, leads back to the blinking display
special key 0..6 and pressing key <CLEAR> again leads to the corresponding mode in
the working memory.
Pressing key <GO>, the next inquiry is displayed, with <ENTER> the setting is stored.
<1> baud rate 9.6 K Setting of baud-rate: 110, 150, 300, 600, 1.2K, 2.4K,
4.8K, 9.6K, 19.2K: Press <GO> until the right
number is displayed and store with <ENTER>.
Several quantities are set to a fixed value: data bit
= 7, parity = even, stop bit = 1.
<3> auto fill on Automatical refilling in mode DOS if more than one
burette volume has been expelled: on = yes; off =
no. Press <GO> until the right answer is displayed
and store with <ENTER>.
<4> send RS232 off Data transfer in mode DOS: on = yes, off = no.
Press <GO> until the right answer is displayed and
store with <ENTER>.
<6> RS232 HSHK full Choice of handshakes for data transfer to RS232:
full, none.
"full" is used for connections with Metrohm
instruments.
"none" can be used e.g. for data transfer via a
Windows terminal.
See also page 45.
765 Dosimat
18
3.1 Special messages and error messages
cylinder empty! The Dosimat is set to "auto fill off" and one burette volume has
been expelled in mode DOS.
Exit: <FILL>
error 1 Check sum error in PROM.
Remedy: Call Metrohm-Service
error 2 RAM-check: Error in on-chip-RAM.
Remedy: Call Metrohm-Service
error 3 RAM-check: Error in off-chip-RAM.
Remedy: Call Metrohm-Service
error 4 RAM-check: Error in on- and off-Chip-RAM.
Remedy: Call Metrohm-Service
error 5 Check sum error in off-chip-RAM.
Remedy: RAM has to be re-initialized: Switch Dosimat off.
Press <FILL> during switching it on again. Display shows
"RAM init.". Press <GO> . Display shows "RAM init. passed".
<CLEAR> leads to basic program.
Note: Stored user modes will be cleared on re-initializing of the
RAM and standard mode DOS is loaded into the working
memory.
INF In mode DOS, a result has been calculated with s=0 or the
calculated result is larger than the range of numbers which can
be displayed (1 E+39).
Exit: <CLEAR>
NaN (Not a number). In mode DOS, a result has been calculated
with s=0 and f=0.
Exit: <CLEAR>
no exch. unit! Exchange unit is not (properly) mounted.
Exit: Mount exchange unit properly.
Note: The filling rate is set to maximum.
V> XXXX ml In mode CNT D the volume to be dosed is >999.999 mL.
Exit: <CLEAR> and enter new weight.
V< XXXX ml In mode CNT D the volume to be dosed is smaller than the
smallest possible increment which can be dosed with the ex-
change unit mounted.
Exit: <CLEAR> and enter new weight.
765 Dosimat
19
3.1. Special messages and error messages
volume <resol.! The volume to be expelled is smaller than the resolution of the
burette with the exchange unit mounted on the Dosimat.
Exit: Change volume to a value which can be expelled with
the exchange unit mounted on the Dosimat
or
mount an exchange unit where the volume can be
expelled.
V-LIM reached! Security volume is reached.
Exit: <FILL>.
V-PIP > V(B) The stored pipetting volume is higher than the burette volume
of the exchange unit mounted on the Dosimat.
Exit: Change volume to a value which can be expelled with
the exchange unit mounted on the Dosimat
or
mount an exchange unit where the volume can be
expelled.
765 Dosimat
20
3.2 Diagnosis
3.2 Diagnosis
3.2.1 General
The 765 Dosimat is a very precise and dependable feeding instrument.
Thanks to its rugged construction, it is highly unlikely that external mechani-
cal or electrical influences will have any adverse effect on its functions.
Although a fault in the instrument can not be excluded with certainty, the
possibility is greater that malfunctions are caused by wrong operation or
handling, through improper connections and the operation with third-party
devices.
Whatever the case, it is always advisable to localise the fault with the diag-
nostic tests, which can be performed quickly and simply. The customer
need call Metrohm service only when the instrument really has a fault. Fur-
ther, he can use the results of the specific diagnostic function to provide the
service engineer with much more precise information.
In the case of inquiries, always quote the serial number on model plate
(see page 3), the program version (see page 18) and if applicable the error
message.
Procedure
The diagnostic menu listed in section 3.2.2 shows all components for which
detailed instructions (diagnostic steps) are available for checking the func-
tionality.
In the case of a possible malfunction, we advise you to perform either the
corresponding diagnostic step or all diagnostic steps as a routine check on
the instrument.
The reactions of the Dosimat to the instructions must be compared with the
descriptions in the diagnostic step. If the Dosimat do not show the ex-
pected reaction ("No" case), the appropriate diagnostic step must be re-
peated to exclude operating errors. However, it is highly probable that re-
peated wrong reactions indicate a malfunction.
Equipment required:
3.496.0070 Dummy-exchange unit (or exchange units if possible with different cylinder
volumes).
6.2107.000 Push-button cable or ordinary test lead with 4 mm banana plugs
6.2149.000 Keyboard
3.496.8360 Test plug
-.---.---- Stop watch or watch with second hand.
-.---.---- Voltmeter (perhaps a calibrated recorder)
765 Dosimat
21
3.2. Diagnosis
3.2.2 Summary
3.2.4 Diagnosis of cylinder code ..............................................23
3.2.5 Diagnosis of key board....................................................23
3.2.6 Diagnosis of display ........................................................23
3.2.7 Diagnosis analog output .................................................24
3.2.8 Diagnosis of digital timer .................................................25
3.2.9 Diagnosis of analog timer................................................26
3.2.10 Diagnosis of external inputs, outputs ..............................26
3.2.11 RAM-test ..........................................................................27
3.2.12 Diagnosis of spindle zero and cock changeover............27
3.2.13 Diagnosis of spindle drive ...............................................27
1. Power off.
2. Remove exchange unit.
3. Disconnect all cables at rear, except mains cable and keyboard.
4. Power on and simultaneously press key <9> and keep pressed until
diagn. key 0...9
5. The individual inquiries of the several diagnostic steps are accessed using the key
<0> ... <9>, exit is with the <CLEAR> key.
1)
This diagnostic step is reserved for the service engineer and therefore will not be described within this document.
765 Dosimat
22
3.2 Diagnosis
3. <GO>
no exch. unit!
5. Check whether the displayed ml-Code (xx ml) corresponds to the exchange unit.
Various exchange units can be inserted to verify their ml-code.
If an exchange unit is coded incorrectly or if the code switches are inoperative, the display shows: E 90: ..
no code!
6. <CLEAR>
diagn. Key 0...9
3. <GO>
key: rate
The display requests to press the key <RATE> on the keyboard 6.2149.000.
5. <CLEAR>
diagn. key 0...9
765 Dosimat
23
3.2. Diagnosis
2. <2>
display test
3. <GO>
Characters are generated for an optical check of the display:
key: FILL }
The display requests to press the key <FILL> on the front of the Dosimat.
5. <CLEAR>
diagn. key 0...9
Fig. 1
To carry out this diagnosis step the optional analog output has to be fitted in the Dosi-
mat (fitted as a standard with versions 2.765.0020, 2.765.0040)
765 Dosimat
24
3.2 Diagnosis
3. <3>
analog output
4. <GO>
V-out = 0.000 V
Measuring instrument reads 0 V (tolerance ±6 mV). Take also into account the tolerance of the measuring
instrument!
5. <GO>
V-out = 1.000 V
Measuring instrument reads +1.000 V (tolerance ±6 mV + tolerance of point 4.)
6. <GO>
V-ramp 1...2
In this test sequence the Dosimat produces at the analog output a triangle voltage. Two different rising or fal-
ling times can be selected:
1: rising or falling time = 48 ms (reserved for the service engineer)
2: rising or falling time = 40 s
7. <2>, <GO>
V-ramp = 40s ↑/↓
Uout [V]
0 t [s]
40 80
V-ramp = 1...2
8. <CLEAR>
diagn. key 0...9
The digital timer is that part of the electronic circuit in the dosimat which is responsable
for the digital spindle speed rate.
765 Dosimat
25
3.2. Diagnosis
3. <GO>
timer dig.
The frequency of the digital timer is measured during 1.5 s. If no fault is found, the following appears:
timer dig. o.k.
4. <CLEAR>
diagn. key 0...9
The analog timer is that part of the electronic circuit in the dosimat which is responsable
for the analog spindle speed rate (adjustable with knob 'dV/dt').
4. <GO>
timer ana.
The frequency of the analog timer is measured during 1.5 s. If no fault is found, the following appears:
timer ana. o.k.
5. <CLEAR>
diagn. key 0...9
This test is necessary only if the 765 Dosimat is interconnected with other instruments via
terminal A. Moreover, this test requires the test plug 3.496.8360 which is a facility for the
repair service. However, it can also be ordered by customers with the above order num-
ber.
3. <6>
extern in/output
4. <GO>
The test runs automatically. If no fault is found, the following appears:
ext. in/out o.k.
765 Dosimat
26
3.2 Diagnosis
5. <CLEAR>
diagn. key 0...9
3.2.11 RAM-test
3. <GO>
The test runs automatically. If no fault is found, the following appears:
RAM TEST passed
4. <CLEAR>
diagn. key 0...9
1. Power off.
2. Insert exchange unit.
3. Power on.
Dosimat fills.
0.2 - 0.6 mm
Fig. 2 Fig. 3
1. Insert exchange unit and put the burette tip into a collecting receptacle.
2. Connect push-button cable 6.2107.000 (if available).
765 Dosimat
27
3.2. Diagnosis
5. <3>
auto fill ???
└─┬─┘
check whether display reads 'on' or 'off' (make note!)
7. <ENTER>, <CLEAR>
The dotted pattern is displayed, afterwards the display changes to the mode used last before starting the di-
agnosis.
Dosimat fills.
10. <RATE>
↑ OFF ml/min
12. <CLEAR>
↓ OFF ml/min
16. Measure the spindle height [can be performed only if dummy exchange unit
3.496.0070 is fitted, or the exchange unit removed and the locking switch (in the
right-hand hole) carefully actuated by means of a screw driver].
765 Dosimat
28
3.2 Diagnosis
17. Actuate <FILL> and simultaneously take the time until the dosimat is in 'ready'
position again.
Filling time: one cock cycle 1s
filling 18 ... 20 s
General rules:
Spindle and cock must move in regular speed (observe sound!)
In the filling position the cock coupling must turn the lever of the exchange unit blamelessly to the left stop
(almost without play and without jamming).
21. <VOLUME>
V-DIS 1. ml
22. Depending on the exchange unit used, enter the volume as below:
1 ml: 0.02 ml
5 ml: 0.1 ml
10 ml: 0.2 ml
20 ml: 0.4 ml
50 ml: 1 ml
23. <ENTER>
24. <GO> (depress briefly) and with a stop watch take the time until the cock starts
turning.
The time must be 14 ... 24 s.
25. Reset the parameter if under 5. the reading was auto fill = on .
765 Dosimat
29
3.3. RAM-initialisation
3.3 RAM-initialisation
In rare cases, it is possible that major interference signals such as line spikes and light-
ning can have an adverse influence on the contents of the data memory. If the contents
of the data memory are undefined, this is indicated after “power on“ with ‘error 5'. The
keyboard is then blocked, no entering is possible until the RAM is initialised again.
4. <GO>
RAM init. passed
5. <CLEAR>
DOS 0.000 ml
Dosimat fills.
The RAM-initialisation deletes the data present in the User-Memory and also those for the special functions
and overwrites them with the standard data below:
The User-Memory is loaded with the standard modes.
Memory 0: Mode DOS
1: DIS R
2: DIS C
3: PIP *
4: DIL *
5: DOS
6: DIS R
7: DIS C
8: PIP *
9: DIL *
The working memory is loaded with the standard mode DOS. The special functions of the dosimat are set to
the following values:
Analog output scale: 1 stroke per 1000 mV
RS 232 sending: off
Baudrate: 9600 Baud
auto. filling: on
765 Dosimat
30
3.4 Releasing a locked spindle
edge of bench
Fig. 4
6. Remove spindle from mechanical stop by turning the large gear wheel. (In case
that the motor is inoperative, position spindle by hand to zero position.)
765 Dosimat
31
4.1 General
4.1 General
The Dosimat offers an extensive remote control. Data transmission occurs via an interface
according to RS 232 C in half duplex procedure.
765 Dosimat
33
4.2. Control commands
765 Dosimat
34
4.2 Control commands
765 Dosimat
35
4.2. Control commands
Information byte 1:
Bit Function
0 Cylinder code
1 Cylinder code
2 Cylinder code
3 1 =no exch. unit
4 1 = New exchange unit
5 1 = Ready
6 1 =V-LIM reached
Cylinder code:
Bit 1 ml 5 ml 10 ml 20 ml 50 ml
0 0 1 1 1 1
1 1 0 1 0 1
2 1 0 1 1 0
765 Dosimat
36
4.2 Control commands
Information byte 2:
Bit Function
0 1= Wrong command code
1 1 = Parameter corrected to its limit value
2 1 = Repeat command in the READY state
3 1 = Cylinder empty
4 1 = Remote control on
5 1 = Data transfer on (send RS232 on)
6 Reserve
765 Dosimat
37
4.2. Control commands
765 Dosimat
38
4.2 Control commands
765 Dosimat
39
4.2. Control commands
The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the
volume of the cylinder:
Cylinder Volume min Volume max
ml ml/min ml/min
1 0.001 0.900
5 0.001 4.9009
10 0.001 9.800
20 0.002 19.700
50 0.005 49.500
If a value is entered which is too high or too low, resp. , the Dosimat corrects it
automatically to volume max or volume min, resp. and sets bit 1 of information byte 2 to 1 (see
4.2.7).
765 Dosimat
41
4.2. Control commands
765 Dosimat
42
4.2 Control commands
The binary value is transmitted in 4 bytes, where only the low-order nibble of a byte
contains a 4 bit information.
1st byte: 20 ... 23
2nd byte: 24 ... 27
3rd byte: 28 ... 211
4th byte: 212 ... 215
5th byte: 'CR'
6th byte: 'LF'
Note: bytes 1...4 may have the value of 'CR' and 'LF' (0DH, 0AH)!
765 Dosimat
43
4.2. Control commands
765 Dosimat
44
4.3 Handshake and other properties
Dosimat as Receiver :
Dosimat external device
DTR DTR
DCD DCD
RxD RxD
Char 0 Char 1 Char n
Time
Notes:
• The DTR has to be active before the external device can transmit.
• The DTR has to be inactive before setting the DCD.
Dosimat as Transmitter:
RTS RTS
CTS CTS
DSR DSR
DCD DCD
TxD TxD
Char 0 Char 1 Char n
Time
Notes:
• During the whole transmission, the DCD has to be inactive (no duplex service).
• The CTS has to be inactive before the Dosimat sends 1 character.
765 Dosimat
45
4.3. Handshake and other properties
Only a shielded data cable (for example, METROHM D.104.0201) may be used to con-
nect the Dosimat with external devices. The cable shield must be properly grounded on
both instruments (pay attention to current loops; always ground in a star-head formation).
Only plugs with sufficient shielding may be used (for example, METROHM K.210.0381
with K.210.9045).
765 Dosimat
46
4.4 Pin assignment of the RS232 / I/O socket
A18 Ready = L
VCE = 30 V, IC = 20 mA
A16 Limit volume reached = L
VCE = 30 V, IC = 20 mA
+5 V
A17 tp tp ≥ 1 ms
Fill = L; during filling
+5 V
A20 tp tp ≥ 1 ms
Feed = L; during dosing
+5 V
0 mV ... 1000 mV
A21 IL ≤ 5 mA
A11 Analog Ground
+12 V A8 IL ≤ 30 mA
+5 V A3 IL ≤ 80 mA
0V A4/5
-12 V A2 IL ≤ 30 mA
765 Dosimat
47
4.4. Pin assignment of the RS232 / I/O socket
A14
and
Request to Send (RTS) B8
ON condition: Titrino is ready to send data.
A23
and
Clear to Send (CTS) B5
ON condition: Remote station is ready to receive data.
A4/5
and
Signal ground (GND) B1
A19
and
B4
Data Terminal Ready (DTR)
ON condition: Instrument is ready to receive data.
A1
and
Data Set Ready (DSR) B3
ON condition: Communication channel is connected.
A24
and
Received Line Signal Detector (RLSD) B2
ON condition: The connected device is ready to send data.
Protective earthing
Direct connection from cable plug to the protective ground of the in-
strument.
765 Dosimat
48
4.4 Pin assignment of the RS232 / I/O socket
Important: The pin numbers are not used according to the RS232
standard version. Therefore do not plug in standard RS232 cables!
765 Dosimat
49
4.4. Pin assignment of the RS232 / I/O socket
765 Dosimat
50
5.1 Technical specifications
5 Appendix
Modes
DOS Dosing, with the ability to calculate a result from the dosed
volume
DIS R Repetitive Dispensing
DIS C Cumulative Dispensing
PIP Pipetting
DIL Diluting
CNT D Preparation of solutions with preselected content
Data outputs
digital Data transfer interface according to RS 232 C
analogue
Representation Volume as voltage signal: 0 ... 1000 mV = 1, 2 ... 10 burette
cylinder volumes (selectable)
Resolution 1 mV
765 Dosimat
51
5.1. Technical specifications
Material
Cabinet Polybutylene terephthalate (PBTP)
Key cover Polycarbonate (PC)
Power supply
Voltage 100, 117, 220, 230/240 V ± 10% (adjustable)
Frequency 50... 60 Hz
Consumption 15 VA
Fuse Thermal fuse (100°C)
Dimensions
Dosimat with exchange unit
Width 150 mm
Height 450 mm
Depth 275 mm
Titrating stand with exchange unit
Width 240 mm
Height 450 mm
Depth 275 mm
Weight
Dosimat with keyboard app. 3.2 kg
as titrating stand app. 4 kg
765 Dosimat
52
5.2 Warranty and certificates
5.2.1 Warranty
The warranty regarding our products is limited to rectification free of
charge in our workshops of defects that can be proved to be due to
material, design or manufacturing faults which appear within 12
months from the day of delivery. Transport costs are chargeable to
the purchaser.
For day and night operation, the warranty is valid for 6 months.
Glass breakage in the case of electrodes or other glass parts is not
covered by the warranty. Checks which are not a result of material or
manufacturing faults are also charged during the warranty period.
For parts of outside manufacture insofar as these constitute an ap-
preciable part of our instrument, the warranty stipulations of the
manufacturer in question apply.
With regard to the guarantee of accuracy, the technical specifica-
tions in the Instructions for Use are authoritative.
Concerning defects in material, construction or design as well as the
absence of guaranteed features, the purchaser has no rights or
claims except those mentioned above.
If damage of the packaging is evident on receipt of a consignment
or if the goods show signs of transport damage after unpacking, the
carrier must be informed immediately and a written damage report
demanded. Lack of an official damage report releases METROHM
from any liability to pay compensation.
If any instruments and parts have to be returned, the original pack-
aging should be used if at all possible. This applies above all to in-
struments, electrodes, burette cylinders and PTFE pistons. Before
embedding in wood shavings or similar material, the parts must be
packed in a dustproof package (for instruments, use of a plastic bag
is imperative). If open assemblies are enclosed in the scope of de-
livery that are sensitive to electromagnetic voltages (e.g. data inter-
faces etc.) these must be returned in the associated original protec-
tive packaging (e.g. conductive protective bag). (Exception: assem-
blies with built-in voltage source belong in a non-conductive protec-
tive packaging). For damage which arises as a result of non-
compliance with these instructions, no warranty responsibility what-
soever will be accepted by METROHM.
765 Dosimat
53
5.2. Warranty and certificates
This is to certify the conformity to the standard specifications for electrical appliances and
accessories, as well as to the standard specifications for security and to system validation
issued by the manufacturing company.
This Metrohm instrument has been built and has undergone final type testing according
to the standards:
Electromagnetic compatibility: Emission
EN50081-1/92, EN55022/class B,
EN55011/class B Generic emission
Electromagnetic compatibility: Immunity
EN50082-1/92 Immunity
IEC1000-4-2/95 (level 4), NAMUR/93 Static discharge
IEC801-3, ENV50140/93+ENV50204/93 (level 2) Radiated rf electromag.field immunity
IEC801-4, IEC1000-4-4/95 (level 3) El.fast transient requirements
IEC801-5, IEC1000-4-5/95 (level 2/3) "Surges" immunity
NAMUR/93 Paragr. 3.2.2., IEC1000-4-11/94 Voltage dips, short interruptions
Security specifications
IEC1010 class1, EN61010 class1, UL3101-1, EN60947:IP31
Metrohm Ltd. is holder of the SQS certificate of the quality system ISO 9001 for quality
assurance in design/development, production, installation and servicing.
765 Dosimat
54
5.2 Warranty and certificates
The company Metrohm AG, Herisau, Switzerland, certifies herewith, that the following
instrument:
765 Dosimat
meets the CE mark requirements of EU Directives 89/336/EWG and 72/23/EWG.
Source of specifications:
Description of apparatus:
765 Dosimat
55
5.3. Connection of a balance
Connecting cable:
Sartorius, MP 8, YDO01... 3.980.3380
Mettler AE, Option 011 3.980.3370 + cable by Mettler: Hand switch ME
42500 or foot switch ME 46278
Mettler AT (15 pin) Cable by Mettler: ME 33640
Mettler, AM/PM: Cable by Mettler: ME 33640 + T-Adapter ME47473 +
Hand switch ME 42500 or foot switch ME 46278
Mettler, AB/AG (LC-RS25): 3.980.3370 + cable by Mettler: LC-RS25
(Adjustments 7 5 2)
Mettler, PG: 3.980.3370 + 6.2125.130
AND Cable on demand
Precisa Cable on demand
Note
• Select the correct balance on the Dosimat (special key 5, see page 18).
• Select the baud rate of the Dosimat such that it corresponds with that of the balance
(special key 1, see page 18).
• Select handshake "none" (special key 6, see page 18).
• Weights can be transferred to parameter "s" of the Dosimat with the aid of a transfer
key on the balance:
- In DOS mode with display s = .
- In CNT D mode with display s g (substance size).
• If another instrument is to be connected to socket A of the Dosimat in addition to the
balance, the 6.2125.000 double plug is available.
• The balance must transmit in mode "g".
Connecting cable:
Seiko DPU-414 6.2124.060 + 6.2125.130
Citizen iDP562 6.2124.070
Epson LQ, FX with interface #8148 6.2124.070
Epson LX-300 6.2124.070
Printer:
Seiko DPU-414, 230 V 2.141.0014
Seiko DPU-414, 120 V 2.141.0015
Citizen iDP562, 230 V 2.140.0024
Citizen iDP562, 115 V 2.140.0025
765 Dosimat
56
5.5 Continuous dosing with two Dosimats
with
V1 cylinder volume on Dosimat 1 (start Dosimat)
V2 cylinder volume on Dosimat 2
765 Dosimat
57
5.5. Continuous dosing with two Dosimats
Example:
Continuous dosing of 55 mL. Exchange units with volumes of 20 mL and 10 mL.
Key:
End volume V = 55 mL
Exchange unit 1 V1 = 20 mL
Exchange unit 2 V2 = 10 mL
55
1. N = Int———— = 1
20+10
2. VR = 55 - 1*(20+10) = 25
3. 25 (VR) is larger than 20 (V1), i.e. Dosimat 2 is the final burette.
4. V-LIM for Dosimat 2 is: 55-20*(1+1) = 15
Settings:
Dosimat 1 (start Dosimat) Dosimat 2 (final Dosimat)
20 mL Exchange unit 10 mL Exchange unit
V-DIS = 20 mL V-DIS = 10 mL
V-LIM = OFF V-LIM = 15 mL
rate ↑ = 25.5 mL/min rate ↑ = 25.5 mL/min
rate ↓ = 60 mL/min rate ↓ = 30 mL/min
The rate ↑ of 25.5 mL/min is the highest possible value for the predefined configuration.
765 Dosimat
58
5.6 Scope of delivery and ordering designations
765 Dosimat
59
5.6. Scope of delivery and ordering designations
Options
Accessories to separate order and on payment of extra charge:
Cables
Cable 765 Dosimat — 682, 686, or 678 Titroprocessor....................................... 3.980.3090
Cable 2 x 765 Dosimat — 682, 686, or 678 Titroprocessor ................................. 3.980.3100
Cable 765 Dosimat — 670 Titroprocessor or 646 VA-Processor......................... 6.2124.010
Cable 765 Dosimat — 586 Labograph ................................................................ 3.980.3170
Cable 765 Dosimat — Dosimat 765 .................................................................... 3.980.3140
Double plug for 25 pin socket.............................................................................. 6.2125.000
Balance connection
Cable 765 Dosimat — Sartorius, MP 8, YDO01................................................... 3.980.3380
Cable 765 Dosimat — Mettler AE, Option 011 .................................................... 3.980.3370
Cable 765 Dosimat — Mettler AT (15 pin) ........................................from Mettler: ME 33640
Cable 765 Dosimat — Mettler, AM/PM ........................ from Mettler: ME 33640 + ME47473
Cable 765 Dosimat — Mettler, AB/AG (LC-RS25) ........3.980.3370 + from Mettler LC-RS25
Cable 765 Dosimat — Mettler, PG................................................ 3.980.3370 + 6.2125.130
Printer connection
Printer Seiko DPU-414, 230 V .............................................................................. 2.141.0014
Printer Seiko DPU-414, 120 V .............................................................................. 2.141.0015
Printer Citizen iDP562, 230 V................................................................................ 2.140.0024
Printer Citizen iDP562, 115 V................................................................................ 2.140.0025
Cable 765 Dosimat — Seiko DPU-414 ......................................... 6.2124.060 + 6.2125.130
Cable 765 Dosimat — Citizen iDP562 ................................................................. 6.2124.070
765 Dosimat
60
Index
Index
Keys are marked with < >, display texts are in bold characters, and pages concerning
the green part are printed in italic.
A Diluting ............................................... 11
Accessories ........................................59 Diluting volume .................................. 11
add V ..................................13 DIS C .................................................... 9
Analogue output ................................. 2 DIS R .................................................... 9
Number of VB ............................... 18 Dispensing ........................................... 9
Analog rate .......................................2, 6 Dispensing volume .............................. 9
Arrows...............................................2, 6 Disturbances .................................... 19ff
auto fill ..................................18 DOS...................................................... 7
Dosing................................................ 7ff
B Dosing key ..................................... 3, 59
b ....................................8 Dosing rate........................................... 6
balance ..................................18 Dropping funnel ................................. 57
Balance
Connection .................................. 56 E
Selection ...................................... 18 Earthing ................................................ 3
baud rate ..................................18 Electronic dropping funnel ................. 57
<BLANK> ............................................8 <ENTER>............................................ 4
error X ................................. 19
C Error messages................................ 19ff
Cables ................................................60 <EXP> ................................................ 4
Calculations ..........................................8 Exchange unit .................................... 60
CE sign ...............................................55 Expelling rate........................................ 6
Certificate............................................54
<CLEAR> ............................................2 F
cnt ..................................13 f = ............................. 8, 12
CNT D .................................................12 <FACTOR>......................................... 8
Connection Factor ................................................... 8
Balance........................................ 56 <FILL> ................................................ 2
Printer........................................... 56 Filling rate............................................. 6
Power............................................. 3 Formula
Stirrer ...............................................3 for add V (CNT D).........................14
Content .............................................12ff for result calculation (DOS) ............. 8
Continuous dosing .............................57
Cumulative dispensing .........................9 G
cylinder empty ! ..................................19 <GO> ................................................. 2
D H
Data Handshake................................... 18, 45
Bit................................................. 46
Input............................................. 4ff I
Output.....................................8, 37ff INF ................................. 19
Printing....................................8, 37ff Initializing............................................ 30
Transfer.......................................37ff ISO quality system ............................. 54
Transfer rate...................................18
Delivery ...............................................59 K
dens. ..................................13 Keypad ................................................. 4
Diagnose ..........................................21ff
Digital rate ............................................6 L
DIL ......................................................11 Limit volume ..................................... 7, 9
765 Dosimat
61
Index
N T
NaN ................................. 19 Tandem dosing.................................. 57
no exch.unit! ................................. 19 Technical specifications .................... 51
O U
Ordering designations ....................... 59 unit ................................. 13
<UNIT>......................................... 8, 13
P User memory ..................................... 17
Parity (data transfer)........................... 46
Pin assignment................................... 47 V
PIP ...................................................... 10 V > XXXX ml .................................
19
Pipetting ..........................................10ff V < XXXX ml .................................
19
Equipment ....................................60 #V(B) 1000 mV ................................. 18
Volume .................................... 10, 11 V-DIL ................................. 11
Power V-DIS ................................... 9
Connection .....................................3 V-LIM ............................... 7, 9
Switch .............................................3 V-LIM reached! ................................. 20
Voltage ............................................ 3 <VOLUME> ........................................ 4
Printer connection .......................... 8, 56 volume<resol.! ................................. 20
Print-out................................................ 8 V-PIP ........................... 10, 11
Program number ................................ 18
PULSE-Mode...................................... 38 W
Warranty............................................. 53
R Working modes.................................. 6ff
RAM-Initializing................................... 30
<RATE> .............................................. 6
<RECALL>........................................ 17
Rate ...................................................... 6
Remote control................................. 37ff
Repetitive dispensing........................... 9
Report................................................... 8
Result
Calculations....................................8
Output ............................................. 9
RS232 HSHK ................................. 18
RS232
Interface........................................46
Handshake ...................................45
Control via .................................. 37ff
Parameter ................................ 18, 46
S
s ................................. 13
s = ................................... 8
Scope of delivery................................ 59
Security volume................................ 7, 9
send RS 232 ................................. 18
special key 0..6 ................................. 18
765 Dosimat
62