Pulse Tube Cryocooler-Seminar

Download as pdf or txt
Download as pdf or txt
You are on page 1of 26
At a glance
Powered by AI
The key takeaways are that cryocoolers are used to achieve low temperatures and can be classified as recuperative or regenerative. Common cryocoolers discussed are Stirling, GM, Joule-Thomson, and pulse tube cryocoolers.

The different types of cryocoolers discussed are Joule-Thomson, Stirling, GM, and pulse tube cryocoolers. Joule-Thomson and Brayton cryocoolers are recuperative while Stirling and GM cryocoolers are regenerative.

The working principle of a pulse tube cryocooler involves alternately applying high and low pressure at one end of a hollow tube closed at the other end. This causes compression and expansion of the gas, resulting in heating at the closed end and cooling at the open end through repeated pressure pulsing.

Pulse Tube Cryocooler

By, Utsav rao (U10ME112) B.Tech. IV Mechanical Engineering department, SVNIT. Guided by, Dr. H.B Naik Professor Mechanical Engineering department, SVNIT.

Introduction
Cryogenic is a science to achieve low temperature Temperature below 123 Kelvin are generally consider in cryogenic range The cryogenic system whose power of cooling is in few watts are consider as cryocooler, where as the system whose cooling power is kilowatts is know as cryorefregerator Recuperative and Regenerative are the two basic classification of cryocooler

Joules thomson and byrayton cryocooler are recuperative type where as Stirling and GM cryocooler are regenerative type Cryocoolers are use in cooling of infrared sensor in missile guidance system, liquefaction of gases, SQUID (Super Conducting Quantum Interference Device), Super conductor etc.

Closed Cycle Cryocoolers

Like heat engine cryocooler make use of a woking substance which undergoes a thermodynamic cycle of operation. In close cycle cryocooler, the working substance absorbs heat Qc from a sample at a cold temperature at Tc and rejects heat Qw at a higher temperature Tw. In this process, the external work done on the system is W. The ratio W/Qc is a measure of effectiveness of the refrigerator.

For ideal carnot cycle COP is given as,

Multi stage cryocooler

Pulse Tube Cryocooler

Reliability of small cryocooler was a problem since long time which was under study from many years. One of the approaches to increase reliability is to eliminate mechanical moving component from the cryocooler. The Stirling type cryocooler have two moving component a piston and a gas displacer, where as the GM type cryocooler has one moving component i.e. displacer. Cooling effect at one end of a hollow tube with a pulsating pressure at the other end was first observed by Gifford and Longsworth, after that in 1963 they discover new type of cryocooler in which moving component was replaced with gas piston. This is known as pulse tube cryocooler.

The working of Pulse tube cryocoolers can be understood with the help of above Figure Consider a thin walled tube closed at one end and other end open. High and low pressure is alternatively applied at the open end. When high pressure is applied the gas boundary moves from left to right end of tube as the pressure inside increases from Pl to Ph. As the work is done on gas due to compression its temperature increases. Since gas in contact with the wall the closed end gets heated.

Similarly when low pressure is applied at closer end the pressure inside tube decreases from Ph to Pl so the gas boundary moves from right to left which leads to expansion of gas, so gas gets cooled so the wall at the open end cools down. Repeated compression and expansion or pressure pulsing cause heating effect on closed end and cooling effect at the open end of tube. When the heat exchanger are added at both end, the effect is now substantiated, since warm end heat exchanger (WHE) ensure that the gas is maintained at some steady temperature after compression. Hence on expansion, the cold end heat exchanger shows cooling effect. To convert cooling effect to useful refrigeration, a highly efficient regenerator is introduced. The WHE is now circulated with cooling water. By this, the heat of compression at the end of high pressure Ph, will be removed from WHE, causing the gas to come back to ambient temperature. Now the gas gets cooled on expansion and the gas will gradually cool both CHE and regenerator.The next batch of incoming gas takes in the cold storage in the regenerator and hence the temperature of the gas entering the pulse tube is slightly less than that of the previous cycle. Thus the CHE is gradually cooled to lower and lower temperature.

To attain the lowest possible temperature, one should use helium gas as a working fluid.The pressure variation produced at the inlet of the pulse tube cryocooler varies sinusoidally as This pressure oscillation can be produced by using a compressor and a rotary valve. The pressure wave produces a sinusoidal oscillation in the velocity v of the gas in the pulse tube. Let the velocity be given as.

The enthalpy flow at any point in the pulse tube can be averaged over one period of the pressure cycle and is given by

The average enthalpy is proportional to , where is the phase difference between pressure and velocity. The average enthalpy flow is a maximum when the pressure and velocity field at any point are in phase. If heat needs to be extracted from the cold end of the pulse tube, then the velocity and pressure oscillations must be in phase at this end.

Types of Pulse tube Cryocooler


Basic pulse tube cryocooler (BPTC). Pulse tube refrigeration systems can be classified as either a Stirling type or a GM type according to the method of pressurization and expansion as shown in Figure (a) and (b). Stirling type cryocooler works on frequency of comperessor which is generally 10-120 Hz and produce high cooling power but its performance is limited at low temperature

GM type cryocooler works on low frequency i.e 1-5 Hz. There is rotarary valve between compressor and Pulse tube. As due to valve there is friction loss in it so cooling power achieve in it less then stirling type but low temperature below 10 kelvin can be achieved through it.

Orifice Pulse Tube Cryocooler (OPTC).


The Orifice Pulse Tube Cryocooler is a modified version of basic pulse tube cryocooler. This modification is made by including an orifice valve and a surge volume at the warm end of the BPTC. Additional components create an advantage of in-phase relationship between the mass flow and the pressure within the pulse tube to enhance the heat transport mechanism. But the mass flow through the regenerator is increases leading to degradation of regenerator performance. This drawback is removed by adding a second orifice i.e. double inlet PTR.

Double Inlet Pulse Tube Cryocooler (DIPTC).


In the DIPTR the hot end of the pulse tube is connected with the entrance (hot end) of the regenerator by an orifice adjusted to an optimal value . The double inlet is a bypass for the regenerator and hence reduces the cooling power. In addition, the valve is a dissipative device, which leads to a deterioration of the performance. However, both these disadvantages are overcome by the fact that the double inlet reduces the dissipation in the regenerator. As a result, the performance of the overall system is improved significantly.

Multi-stage type PTC.


It is really impossible to achieve very low temperature in a single stage of PTC. So one PTC can be used to precool the other. For temperatures below 30K, it turns out to be advantageous to split the system in two i.e. double-stage PTR. In this arrangement the hot end of the second tube is connected to the cold end of the first stage. Three-stage PTCs have also been introduced. With a three-stage PTC, 1.78K has been reached using He3 as the working fluid.

Components

Pulse tube cryocooler


The Schematic and photograph of the two stage Pulse Tube refrigerator is shown below. The cold end of the first stage regenerator forms the warm end of the second stage regenerator. The warm ends of the Pulse Tubes and that of the first stage regenerator are mounted to the top flange of the system. The cold and warm end heat exchangers of the first and second stages are designed such that it leads to the best performance of Pulse Tube Cryocooler.

Rotary valve
The rotary valve is one of the critical components of a GM type pulse tube. It is used to switch high and low pressure from a helium compressor to the pulse tube system. The high and low pressure of helium compressor are connected to the rotary valve

The

rotary valve has a rulon part which is made to rotate with the help of a synchronous motor against an aluminium block with predefined passages connecting the high and low pressures from the helium compressor. The rotational frequency of the synchronous motor is controlled using an inverter drive.

Helium Compressor
The main function of the compressor is to supply gas pressurization and depressurization in the closed chamber. Electrical power is applied to the compressor where this electrical work is converted into the mechanical energy associated with sinusoidal pressure waves, resulting in gas motion. In an ideal compressor, the electrical power provided to the compressor must be equal to PdV, where the integration is performed over an entire cycle, P is the compressor pressure, and f is the compressor frequency. In an actual system, however, the above-mentioned power (the PdV power) is always less than the actual measured electrical power due to the associated irreversibilities.

Orifice valve
The role of either the inertance tube or the orifice valve is to appropriately adjust the phase difference between the mass flow rateand the pressure. By controlling the orifice diameter or the inertance tube diameter and length, the desired phase relationship can be obtained. In general, the orifice valve is a needle valve.

Temperature measurement Silicon diode. Silicon Diodes are the best choice for generalpurpose cryogenic use. The cryogenic temperature sensors are interchangeable (they follow a standard curve). Silicon Diodes are easy to instrument, and are used in a wide variety of cryogenic applications, such as cryo-coolers, laboratory cryogenics, cryo-gas production, and space satellites. Germanium resistor temperature sensor. Germanium RTDs have the highest accuracy, reproducibility, and sensitivity from 0.05 K to 100 K. They are resistant to ionizing radiation, but are not recommended for use in magnetic fields. Germanium RTDs are used mostly in research settings when the best accuracy and sensitivity are required. Germanium and Ruthenium Oxide are the only two cryogenic temperature sensors that can be used below 100 mK.

Thermocouples. Thermocouples can be used over an extremely wide range and in harsh environmental conditions, and follow a standard response curve. Less accurate than other cryogenic temperature sensors, special techniques must be employed when using thermocouples to approach temperature accuracies of 1% of temperature. Thermocouples are used for their small size, extremely wide temperature range (exceeding high temperature limits of Platinum RTDs), and simple temperature measurement methodology. Platinum resistor temperature sensor. Platinum RTDs are an industry standard. They follow an industry standard curve from 73 K to 873 K with good sensitivity over the whole range. Platinum RTDs can also be used down to 14 K. Because of their high reproducibility, they are used in many precision metrology applications. Platinum RTDs are inexpensive and require simple instrumentation. They are widely used in cryogenic applications at liquid nitrogen temperatures or greater.

Capacitive

sensors. Capacitance sensors are ideally suited for use as temperature control sensors in strong magnetic fields because they exhibit virtually no magnetic field dependence. Small variations in the capacitance/temperature curves occur upon thermal cycling. It is recommended that temperature in zero field be measured with another cryogenic temperature sensor, and that the capacitance sensor be employed as a control element only.

Regenerator Materials
The regenerator is the most important component in pulse tube refrigerator. Its function is to absorb the heat from the incoming gas during the forward stroke, and deliver that heat back to the gas during the return stroke. Ideally, PTC regenerators with no pressure drop and a heat exchanger effectiveness of 100% are desired, in order to achieve the maximum enthalpy flow in the pulse tube. The performances of the real regenerators are of course far from ideal. Stainless steel wire screens, compunds of rarer earth material such as Er3Ni and HoCu2 are usually selected as the regenerator packing material, since they offer higher heat transfer areas, low pressure drop, high heat capacity, and low thermal conductivity.

Generally in first stage regenerator lead (Pb) and wire mesh is used where as in second stage regenerator compound of rarer earth materials are used

Discussion and conclusion

Cooling through Pulse tube cryocooler is one of the best way because of its high reliability compared to G-M cryocooler and sterling cryocooler. It does not require big infrastructure in compared to other system. Rare earth material regenerator such as Er3Ni plays an important role to achieve low temperature with the use of Helium as a working fluid. Optimization of the pulse tube cryocooler is a big challenge. In optimization mass flow rate of the working fluid is to be set to achieve low temperature work is going on in the field of increasing of its efficiency and optimizing it with the help of CFD analysis. Future scope is there in the field of thermo acoustically driven pulse tube cryocooler in which the pressure pulse is generated in thermo acoustic prime mover. Solar energy can also be used to dive the thermo acoustic prime mover.