Noise Figure Measurement Using A Spectrum Analyzer and Signal Generator
Noise Figure Measurement Using A Spectrum Analyzer and Signal Generator
measurement is to be performed; adjust Sig Gen for a signal approximately 3 dB above the
noise as displayed on the S/A and input this Sig Gen reading into equation below.
Procedure, Detailed: With two markers set, one at the signal peak and the other riding the
output noise at approximately +- 4.5 MHz or so and adjust the Sig Gen for a 3 dB ratio
between the two cursors. This yields a 3 db S/N ratio and Sig Gen power = noise power.
Note: One marker indicates the DUT’s output noise power level, NO = FGkTOB and the other marker indicates
So + No for double the power (+ 3dB) indication or 2*NO of the noise-only marker.
(Equation derivation is from ref. 1 slide #36: G*Si = NF*G*k*TO*BW and NF = Si / k*TO*BW)
Procedure, Calculation: Using this equation input the Sig Gen reading to calculate DUT NF:
NF (in dB) = Si + 114 - 10*Log (BW) (where Si is in dBm and BW = S/A RBW in MHz)
Test Equipment:
• Sig Gen - Agilent E4437B
• Spec Ana - R&S FSEB 30
• Attenuator - MiniCircuits VAT-6
Detailed equipment set up, Low-Gain (~20 dB) DUT. This setup is suitable for measuring
the lower gain DUTs like the 438 product that exhibit on the order of 20 dB of gain. The
addition of an amplifier after the DUT assures us that the noise of the DUT will sufficiently
override the inherent noise floor of the Spectrum Analyzer.
Test Equipment:
• As above, plus
• Test amplifier - MiniCircuits model ZHL-42, 36 dB 700 – 4200 MHz, NF < 10 dB
Gain: 19 – 23 dB
NF: < 6 dB
Freq: 2400 – 2483 MHz
Also notice the Test Amp with the black heat sink fins fed by the DC power supply in the
lower right corner.
Screen shows as 3 dB
S/N ratio
References:
www.ieee.li/pdf/viewgraphs_mohr_noise.pdf