Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Monitor Signos Vitales Datascope Spectrum or de Mindray

Download as pdf or txt
Download as pdf or txt
You are on page 1of 152

iMEC12/iMEC10/iMEC8

Patient Monitor

Service Manual

Intellectual Property Statement


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rent, adaptation and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

, and

are the registered trademarks or trademarks owned

by Mindray in China and other countries. All other trademarks that appear in this manual are used only for editorial purposes without the intention of improperly using them. They are the property of their respective owners. Contents of this manual are subject to changes without prior notice.

Revision History
This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change. Contents of this manual are subject to change without prior notice. Revision number: 3.0 Release time: May 2012

2011 - 2012 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.

Preface
Manual Purpose
This manual provides detailed information about the assembling, dissembling, testing and troubleshooting of the equipment to support effective troubleshooting and repair. It is not intended to be a comprehensive, in-depth explanation of the product architecture or technical implementation. Observance of the manual is a prerequisite for proper equipment maintenance and prevents equipment damage and personnel injury. This manual is based on the maximum configuration; therefore, some contents may not apply to your monitor. If you have any question, please contact our Customer Service Department.

Intended Audience
This manual is for biomedical engineers, authorized technicians or service representatives responsible for troubleshooting, repairing and maintaining the patient monitors.

Abbreviations
Abbreviations used in this manual are: CMS PCB central monitoring system printed circuit board

Passwords
A password may be required to access different modes within the monitor. The passwords are listed below: User maintenance: Factory maintenance: Demo mode: Configuration mode: 888888 (User adjustable) 332888 2088 315666 (User adjustable)

II

Table of Contents
1 Safety ................................................................................................................................. 1-1 1.1 Safety Information .......................................................................................................... 1-1 1.1.1 Dangers .............................................................................................................. 1-2 1.1.2 Warnings............................................................................................................. 1-2 1.1.3 Cautions ............................................................................................................. 1-2 1.1.4 Notes .................................................................................................................. 1-3 1.2 Equipment Symbols ........................................................................................................ 1-3 2 Theory of Operation ........................................................................................................ 2-5 2.1 Introduction..................................................................................................................... 2-5 2.2 System Connections ........................................................................................................ 2-6 2.2.1 Installation Support ............................................................................................ 2-6 2.2.2 Connectors for Peripheral Devices..................................................................... 2-7 2.3 Main Unit ........................................................................................................................ 2-8 2.4 Front Housing Assembly................................................................................................. 2-9 2.4.1 Main Board......................................................................................................... 2-9 2.4.2 Keypad and the Navigation Knob .................................................................... 2-10 2.4.3 Power On/Off and LED Indicator Board...........................................................2-11 2.4.4 Alarm Lamp Board............................................................................................2-11 2.4.5 Display ..............................................................................................................2-11 2.4.6 Touchscreen and Touchscreen Control Board ...................................................2-11 2.5 Rear Housing Assembly.................................................................................................2-11 2.5.1 AC/DC Power Board........................................................................................ 2-12 2.5.2 Interface Board (DC/DC, Power Management Circuit) ................................... 2-12 2.5.3 Battery Back Board .......................................................................................... 2-13 2.5.4 Parameter Module ............................................................................................ 2-14 2.5.5 Multi-parameter Module .................................................................................. 2-14 2.5.6 SpO2 Board....................................................................................................... 2-15 2.5.7 C.O. Module..................................................................................................... 2-15 2.5.8 CO2 Module ..................................................................................................... 2-15 2.5.9 Recorder ........................................................................................................... 2-15 2.6 Speaker.......................................................................................................................... 2-16 3 Unpacking and Installation ............................................................................................. 3-1 3.1 Unpacking the Equipment............................................................................................... 3-1 3.2 Preparation for Installation.............................................................................................. 3-2 3.2.1 Preparation for Installation Site.......................................................................... 3-2 3.2.2 Electrical Requirements ..................................................................................... 3-3 3.2.3 Monitor Installation............................................................................................ 3-3 3.2.4 Preparation for Power on.................................................................................... 3-3 4 Hardware and Software Upgrade................................................................................... 4-1 4.1 Introduction..................................................................................................................... 4-1 4.2 Hardware Upgrade .......................................................................................................... 4-1 4.2.1 Upgrading the 1G SD Storage Card ................................................................... 4-1 4.2.2 Upgrading Touchscreen...................................................................................... 4-2 4.2.3 Upgrading Wi-Fi ................................................................................................ 4-3 4.2.4 Upgrading Recorder ........................................................................................... 4-4 4.2.5 Upgrading CO2 ................................................................................................... 4-5 4.2.6 Upgrading C.O. .................................................................................................. 4-9 4.2.7 Upgrading IBP ................................................................................................. 4-10 4.2.8 Upgrading VGA/USB/Multi-functional Interface.............................................4-11 4.3 Software Upgrade.......................................................................................................... 4-12

4.3.1 Installing Mindray Patient Monitor Software Upgrade Tool (PN: G-110-00049300) ............................................................................................................................. 4-13 4.3.2 Software Upgrade Procedure ........................................................................... 4-15 5 Testing and Maintenance................................................................................................. 5-1 5.1 Introduction..................................................................................................................... 5-1 5.1.1 Test Equipment................................................................................................... 5-1 5.1.2 Test Report ......................................................................................................... 5-1 5.1.3 Preventative Maintenance .................................................................................. 5-2 5.1.4 Recommended Frequency .................................................................................. 5-2 5.2 Visual Inspection ............................................................................................................. 5-3 5.3 Power-on Test.................................................................................................................. 5-3 5.4 Module Performance Tests.............................................................................................. 5-3 5.4.1 ECG Test and Calibration................................................................................... 5-3 5.4.2 Resp Test ............................................................................................................ 5-4 5.4.3 SpO2 Test............................................................................................................ 5-5 5.4.4 NIBP Test and Calibration.................................................................................. 5-5 5.4.5 Temp Test ........................................................................................................... 5-8 5.4.6 IBP Test and Calibration .................................................................................... 5-9 5.4.7 C.O. Test............................................................................................................5-11 5.4.8 Mainstream CO2 Tests.......................................................................................5-11 5.4.9 Sidestream and Microstream CO2Tests and Calibration .................................. 5-13 5.5 Nurse Call Function Test............................................................................................... 5-15 5.6 Analog Output Performance Test .................................................................................. 5-16 5.7 Electric Safety Tests ...................................................................................................... 5-16 5.8 Touchscreen calibration................................................................................................. 5-16 5.9 Recorder check.............................................................................................................. 5-17 5.10 Battery Check.............................................................................................................. 5-17 5.11 Factory Maintenance ................................................................................................... 5-18 5.11.1 Accessing Factory Maintenance Menu........................................................... 5-18 5.11.2 Drawing Waves .............................................................................................. 5-18 5.11.3 Recorder ......................................................................................................... 5-18 5.11.4 Software Version ............................................................................................ 5-19 5.11.5 Monitor information recording....................................................................... 5-20 5.11.6 Calibrate NIBP ............................................................................................... 5-21 6 Corrective Action ............................................................................................................. 6-1 6.1 Introduction..................................................................................................................... 6-1 6.2 Part Replacement ............................................................................................................ 6-1 6.3 Checking Patient Monitor Status..................................................................................... 6-1 6.4 Checking Software Version............................................................................................. 6-2 6.5 Checking Technical Alarms ............................................................................................ 6-2 6.6 Troubleshooting Guide.................................................................................................... 6-2 6.6.1 Power On/Off Failure......................................................................................... 6-2 6.6.2 Display Failures ................................................................................................. 6-3 6.6.3 Alarm Lamp Failures.......................................................................................... 6-4 6.6.4 Button and Knob Failures .................................................................................. 6-4 6.6.5 Sound Failures.................................................................................................... 6-4 6.6.6 Power Supply Failures ....................................................................................... 6-5 6.6.7 Recorder Failures ............................................................................................... 6-6 6.6.8 Output Interface Failure ..................................................................................... 6-6 6.6.9 SD Card Failure.................................................................................................. 6-7 6.6.10 Wired Network Related Problems.................................................................... 6-7 6.6.11 Wi-Fi Related Problems ................................................................................... 6-8 6.6.12 Software Upgrade Problems............................................................................. 6-8 6.6.13 Technical Alarm Messages............................................................................... 6-8

7 Disassembly and Repair .................................................................................................. 7-1 7.1 Tools Required ................................................................................................................ 7-1 7.2 Preparations for Disassembly.......................................................................................... 7-1 7.3 Disassembling the Main Unit.......................................................................................... 7-2 7.3.1 Separating the Front and Rear Half of the Monitor............................................ 7-2 7.3.2 Removing the Parameter Connector Panel Assembly ........................................ 7-3 7.3.3 Disassembling the Parameter Module Assembly ............................................... 7-4 7.3.4 Removing the Battery Compartment Assembly ................................................. 7-7 7.3.5 Removing the Power Management and Interface Board.................................... 7-8 7.3.6 Removing the Recorder (Optional) .................................................................... 7-9 7.3.7 Checking Waterproof Strips before Reassembling............................................. 7-9 7.4 Disassembling the Front Housing Assembly................................................................. 7-10 7.4.1 Removing Touchscreen Control Board (Optional)........................................... 7-10 7.4.2 Removing Antenna (Optional) ......................................................................... 7-10 7.4.3 Removing SD Card (Optional)..........................................................................7-11 7.4.4 Removing the Main Board ............................................................................... 7-12 7.4.5 Removing the Touchscreen (Optional)............................................................. 7-13 7.4.6 Disassembling the Screen................................................................................. 7-14 7.4.7 Removing the Keypad...................................................................................... 7-15 7.4.8 Removing the Encoder ..................................................................................... 7-16 7.4.9 Removing the Alarm Lamp Board ................................................................... 7-16 7.4.10 Removing the ON/OFF Key Board................................................................ 7-16 7.5 Disassembling the Battery Compartment Assembly ..................................................... 7-17 7.5.1 Removing the Power Board ............................................................................. 7-17 7.5.2 Removing the Battery Interface Board Assembly ............................................ 7-17 7.6 Disassembling the Parameter Module Assembly .......................................................... 7-18 7.6.1 Removing the Mindray M02C CO2 Module (Optional)................................... 7-18 7.6.2 Removing the Mainstream CO2 Module (Optional) ........................................ 7-19 7.6.3 Removing the Microstream CO2 module (Optional)........................................ 7-19 7.6.4 Removing the C.O. Module (Optional)............................................................ 7-20 7.6.5 Removing the SpO2 Board ............................................................................... 7-20 7.6.6 Removing the Multi-parameter Module........................................................... 7-20 7.6.7 Removing the Pump and Valve ........................................................................ 7-21 7.7 Removing the Parameter Connector Assembly............................................................. 7-21 7.7.1 Removing the Mainstream CO2 Signal Cable (Optional) ................................ 7-21 7.7.2 Removing the Microstream CO2 Connector Assembly (Optional) .................. 7-22 7.7.3 Removing the Watertrap Base Assembly (Optional)........................................ 7-22 7.7.4 Removing the Cables from the Parameter Connector Assembly ..................... 7-23 7.7.5 Removing the Recorder.................................................................................... 7-24 8 Parts .................................................................................................................................. 8-1 8.1 Introduction..................................................................................................................... 8-1 8.2 iMEC8............................................................................................................................. 8-2 8.2.1 Main unit ............................................................................................................ 8-2 8.2.2 Front Housing Assembly (Nakescreen).............................................................. 8-3 8.2.3 Front Housing Assembly (Touchscreen) ............................................................ 8-5 8.2.4 Rear Housing (Network/VGA/USB/Anolog Input & Output) ........................... 8-7 8.2.5 Rear Housing Assembly ..................................................................................... 8-9 8.2.6 Battery Compartment Assembly ...................................................................... 8-10 8.2.7 Parameter module assembly..............................................................................8-11 8.2.8 Parameter connector panel assembly ............................................................... 8-13 8.2.9 Mindray Sidestream CO2 Module Kit .............................................................. 8-15 8.2.10 Oridion Microstream CO2 Module Kit........................................................... 8-16 8.2.11 Capnostat Mainstream CO2 module kit.......................................................... 8-17 8.2.12 TR6F Recorder Kit......................................................................................... 8-18 8.2.13 Battery charge board assembly (12 inch) ....................................................... 8-19 8.3 iMEC10......................................................................................................................... 8-20

8.3.1 Main unit .......................................................................................................... 8-20 8.3.2 Front Panel Assembly (10 Inch/Touchscreen/Wi-Fi+ 1G SD Card) ................ 8-21 8.3.3 Fujitsu Touchscreen Kit (10 Inch LCD)........................................................... 8-23 8.3.4 Front Housing Assembly (10 Inch) .................................................................. 8-24 8.3.5 Tianma LCD display kit (10 inch).................................................................... 8-25 8.3.6 Rear Housing Assembly ................................................................................... 8-26 8.3.7 Rear Housing Assembly ................................................................................... 8-28 8.3.8 Battery Compartment Assembly ...................................................................... 8-29 8.3.9 Parameter Module Assembly ........................................................................... 8-30 8.3.10 Parameter Connector Assembly ..................................................................... 8-30 8.3.11 Mindray Sidestream CO2 Module Kit ............................................................ 8-30 8.3.12 Oridion Microstream CO2 Module Kit........................................................... 8-30 8.3.13 Capnostat Mainstream CO2 Module Kit ........................................................ 8-30 8.3.14 TR6F Recorder Kit......................................................................................... 8-31 8.3.15 Battery Interface Board Assembly ................................................................. 8-31 8.4 iMEC12......................................................................................................................... 8-31 8.4.1 Main Unit ......................................................................................................... 8-31 8.4.2 Front Panel Assembly (12 Inch/Touchscreen/Wi-Fi+ 1G SD Card) ................ 8-31 8.4.3 Fujitsu Touchscreen Kit (12 Inch).................................................................... 8-33 8.4.4 Front Housing Assembly (12 Inch) .................................................................. 8-34 8.4.5 Tianma LCD Display Kit (12 Inch).................................................................. 8-35 8.4.6 Rear Housing Assembly ................................................................................... 8-36 8.4.7 Rear Housing Subassembly.............................................................................. 8-36 8.4.8 Battery Compartment Assembly ...................................................................... 8-36 8.4.9 Parameter Module Assembly ........................................................................... 8-36 8.4.10 Parameter Connector Assembly ..................................................................... 8-36 8.4.11 Mindray Sidestream CO2 Module Kit ............................................................ 8-37 8.4.12 Oridion Microstream CO2 Module Kit........................................................... 8-37 8.4.13 Capnostat Mainstream CO2 Module Kit ........................................................ 8-37 8.4.14 TR6F Recorder Kit......................................................................................... 8-37 8.4.15 Battery Interface Board Assembly ................................................................. 8-37 8.5 Other Maintainable Parts............................................................................................... 8-37 A Electrical Safety Inspection ........................................................................................... A-1

1 Safety
1.1 Safety Information
DANGER

Indicates an imminent hazard that, if not avoided, will result in death or serious injury.

WARNING

Indicates a potential hazard or unsafe practice that, if not avoided, will result in death or serious injury.

CAUTION

Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.

NOTE

Provides application tips or other useful information to ensure that you get the most from your product.

1-1

1.1.1 Dangers
There are no dangers that refer to the product in general. Specific Danger statements may be given in the respective sections of this manual.

1.1.2 Warnings WARNING


All installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel. There is high voltage inside the equipment. Never disassemble the equipment before it is disconnected from the AC power source or the battery. When you disassemble/reassemble a parameter module, a patient leakage current test must be performed before it is used again for monitoring. The equipment must be connected to a properly installed power outlet with protective earth contacts only. If the installation does not provide for a protective earth conductor, disconnect it from the power line and operate it on battery power, if possible. Disposal of the packaging material should observe the applicable waste control regulations. and keeping it out of childrens reach.

1.1.3 Cautions CAUTION


Make sure that no electromagnetic radiation interferes with the performance of the equipment when preparing to carry out performance tests. Mobile phone, X-ray equipment or MRI devices are a possible source of interference as they may emit higher levels of electromagnetic radiation. Before connecting the receiver to the power line, check that the voltage and frequency ratings of the power line are the same as those indicated on the units label or in this manual. Protect the equipment from damage caused by drop, impact, strong vibration or other mechanical force during servicing.

1-2

1.1.4 Notes NOTE


Refer to Operation Manual for detailed operation and other information.

1.2 Equipment Symbols


Attention: Consult accompanying documents (this manual). Danger: High-voltage Alternating currentAC Silencing Alarms Freeze waveforms NIBP Equipotential Grounding Terminal USB connector Gas outlet Manufacture date Power ON/OFF Serial number Battery Pausing Alarms Record Enter the main menu VGA output

Multifunctional output Network interface Inserted direction

The product bears CE mark indicating its conformity with the provisions of the Council Directive 93/42/EEC concerning medical devices and fulfils the essential requirements of Annex I of this directive. ESD warning symbol for electrostatic sensitive devices. Type CF applied part. The unit displaying this symbol contains an F-type isolated (floating) patient part providing a high degree of protection against shock, and is suitable for use during defibrillation. Type BF applied part. Defibrillator-proof protection against electric shock.

1-3

FOR YOUR NOTES

1-4

2 Theory of Operation
2.1 Introduction
This patient monitor is designed to monitor a fixed set of physiological parameters including ECG, respiration (Resp), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), and carbon dioxide (CO2). The patient monitor also: Provides audible and visual alarm indications in case of patient or equipment problems. Enables displaying, reviewing, storing and transferring of real-time data and wave. Incorporates multiple input devices such as buttons, knob, and touchscreen. Enables program upgrade over the network. Integrates the information of other devices, which include but are not restricted to defibrillator.

2-5

The above figure shows a system consists of the iMEC patient monitor and its peripheral devices. The iMEC patient monitor: Can be used for monitoring the physiological parameters, giving alarms and reviewing patient data, etc. Supports recorder. Supports nurse call signal, synchronization defibrillation signal, and analog output signal. Supports Wi-Fi module, wired network, remote view, and communication with the HyperVisor Central Monitoring System. Supports a secondary display. Supports external AC power source and an internal battery. Supports clinical data acquisition, which has two ways: by SD card and by USB drive. The system software should support data output function, for SD card is a built-in device.

2.2 System Connections


2.2.1 Installation Support
The patient monitor can be mounted on a wall bracket or on a trolley support. The wall bracket or trolley support can be ordered optionally. Each type of mounting bracket is delivered with a complete set of mounting hardware and instructions. Refer to the documentation delivered with the mounting hardware for instructions on assembling mounts.

CAUTION

Use mounting brackets we supply or approve. If other compatible mounting bracket is used, be sure it can be safely used on the patient monitor. The mounting bracket should be installed by our qualified service personnel, or engineers who have adequate knowledge on it. If other mounting solution is used, the installation personnel and the customer should verify if it can be safely used on the patient monitor, and the customer assume the responsibility for any risk resulting from that.

2-6

2.2.2 Connectors for Peripheral Devices


On the back of the patient monitor you can find all connectors for peripheral devices.

1 4

iMEC Rear View

1. 2. 3. 4. 5. 6.

AC power input: used to connect an AC power source (100 to 240 VAC, 50/60 Hz). Equipotential terminal: used to connect the equipotential terminal of other equipments, eliminating potential difference between different pieces of equipment. VGA connector: used to connect a secondary display. Network connector: an RJ45 connector, used to connect an Ethernet network or a PC. Multifunctional connector: used to output analog signals and defibrillator synchronization signals. General USB connector: used to connect any USB-compatible peripheral device.

2-7

2.3 Main Unit


The main unit of the patient monitor consists of two parts: Front housing assembly: main board, keypad board assembly (knob), display, touchscreen, alarm lamp board, power switch, and indicator board. Rear housing assembly: power board (AC/DC), battery pad, interface board, recorder, speaker, multi-parameter module which includes three types of SpO2 stacking board, CO2 module, and IBP/C.O. module. The following figure shows the main unit architecture of the patient monitor.

2-8

2.4 Front Housing Assembly


Alarm lamp board Display bracket

Navigation knob Keypad Main board Touchscreen control board

Power on/off board

2.4.1 Main Board


The main board is located at the center of the front housing assembly's LCD metal sheet. It is the heart of the patient monitor. It provides communication, display, and data storage functions, including: Display drive and backlight control Wired and wireless network Data storage (DDR, NOR flash, and SD card) Printing Serial ports: there are 7 serial ports, respectively for multi-parameter module, C.O./IBP module, sidestream/mainstream CO2 module, recorder, power board, and key board. The seventh serial board is for future use. Connection with touchscreen control board Audio drive EEPROM drive USB

2-9

The main board has the following interfaces to connect other PCBAs: Display signal interface Backlight control interface Keypad interface Alarm lamp interface Wi-Fi interface SD storage card interface Interface board interface (for external output, module communication, power supply and etc.)

2.4.2 Keypad and the Navigation Knob


The keypad and navigation knob are located at the right side of the front housing assembly. The navigation knob can be rotated both clockwise and counterclockwise. It can also be pressed down. The navigation knob is connected with the keypad. The keypad has the following functions: It scans and detects the input of the six keys; It integrates the power on/off hardkey; It integrates the rotary encoder; It connects and forwards the status of power switch and LED board to the main board; and, It communicates with the main board through a serial port. The following figure shows the connection of keypad:

2-10

2.4.3 Power On/Off and LED Indicator Board


The power on/off and LED indicator board is connected with the keypad. It controls the patient monitor's power switch. It has 3 groups of indicators which are used to indicate the status of AC mains, battery, as well as the monitor's power on/off status.

2.4.4 Alarm Lamp Board


The alarm lamp board is located at the top of the front housing. It has two-color indicators, red and yellow. The alarm lamp board directly connects the main board through a cable. It is controlled directly by the main board.

2.4.5 Display
iMEC8, iMEC10 and iMEC12 respectively support 8-inch, 10-inch, and 12-inch LCDs. The display connects the main board through a display signal cable and a backlight cable. The interface cables for the three types of patient monitors are different and cannot be mixed. The reliability of display signal cable is relatively poor. If display problem occurs, consider first to replace the display signal cable.

2.4.6 Touchscreen and Touchscreen Control Board


The fully configured iMEC patient monitor supports touchscreen. The touchscreen can be controlled by touch and can be calibrated. The touchscreen is connected with the touchscreen control board and the main board. The touchscreen control board drives the touchscreen and implements communication with the patient monitor. The touchscreen control board communicates with the main board via an I2C interface and connects the touchscreen via FPC.

2.5 Rear Housing Assembly

2-11

The rear housing assembly mainly includes power management system, multi-parameter collecting and processing system, recorder, and speaker. The power management system consists of three key components: AC/DC power board, DC/DC and power management, battery and battery back board. The structure is shown below:

Brown circuits are placed on the connector board. The three boards are introduced respectively in the following:

2.5.1 AC/DC Power Board


The AC/DC power board is located at the metal sheet of battery compartment at the lower part of the patient monitor. It transforms the input AC (100 to 240V AC, 50/60 Hz) into 15.2 V bus voltage VIN, which is the power source for all voltages in the patient monitor. The AC/DC power board supplies power for all modules through the two-pin power cord connecting to the interface board.

2.5.2 Interface Board (DC/DC, Power Management Circuit)


The interface board mainly performs DC/DC conversion, power management, and transmission of external connector signals: DC-DC: Output 12 V (10.8 to 12.6V), 5V +/- 5% Power management and interface DC/DC conversion circuit Single lithium battery charging and discharging management Power on/off control: no power-on delay, and power-off delay Battery type detection (current high-capacity and low capacity lithium batteries) Battery on-site detection Battery voltage (capacity) detection AC or battery power source detection Temperature check 2-12

Communication with main board External connector signal transfer (RJ45, USB, multi-functional connector, and VGA)

The interface board is the signal center of the whole system. All boards in rear housing assembly needs to communicate with main board through the transfer of interface board. The interface definitions are: Multi-parameter module connector (27 pin) CO2 connector (5 pin) M03B-C.O. module connector (4 pin) Recorder (6 pin) AC-DC power board (2 pin) Battery back board (24 pin) Speaker (2 pin) External connector (directly welt on the board): network, USB, VGA, multi-functional connector Main board connector (225 pin) Reserved serial port

All connectors are designed with mechanical barriers to prevent mis-operations in production or maintenance. The interface board has two types of configuration: Configuration Type Standard (support patient monitors of standard configuration) Extended (support patient monitors other than standard ones) Common Requirement Difference No other ports RJ45 USB, VGA, multi-functional output

2.5.3 Battery Back Board


The iMEC patient monitor consumes low power thanks to the better system design. One lithium battery is capable to supply the power for the whole system. The battery back board transmits the signal of battery capacity and status into the internal system. The patient monitor only supports lithium battery. The on-site detection of lithium battery is achieved by the resistance inside the battery. A new battery provides two types of information by 2 pins: Battery on-site information Battery low/high capacity information The patient monitor supports 11.1 V lithium chargeable battery. The battery is located at the base of the patient monitor. The battery power is input into power module through battery interface board, and then power module processes and forwards the power into each components of the patient monitor. Two types of batteries are supported: Current high-end lithium battery 4500 mAh New developed low capacity lithium battery 2600 mAh The battery is designed with mechanical barriers for correct type and insertion. The patient monitor requires a battery of correct type to be inserted in correct direction. Otherwise, the battery cannot be loaded.

2-13

2.5.4 Parameter Module


The parameter module is a relatively independent system, which consists of a multiparameter module and a SpO2 board, as shown below:

2.5.5 Multi-parameter Module


The multi-parameter module, as well as the NIBP pump and electromagnetic valve, are located on the bracket of the multi-parameter module, which is fixed to the rear housing with four screws. The multi-parameter module is connected to the connector board with a 27 pin cable to supply power for the parameter module, communicate with the main board, and realize analog output and Defib sync output. It adopts the DSP processor as the core. For ECG applications, the self-developed Mindray ECG_ASIC chip is used to realize 3/5/12-lead ECG, and Resp (I/II lead selectable) measurement, which greatly simplifies the sampling and control of ECG signals for the DSP. As for NIBP, IBP, and Temp measurement, the high-precision - ADC is used to realize a direct sampling of corresponding parameter signals. Mindray SpO2 measurement is realized by a tacking connector. The Nellcor and Masimo SpO2 boards are also compatible with the multi-parameter module, being connected through a BTB socket. Supports 2-channel 3-/5-lead ECG monitoring (and 12-lead in future), and I/II lead Resp monitoring Provides power for and communicates with Mindray/Nellcor/Masimo SpO2 board Supports 2-channel Temp monitoring Supports 2-channel IBP monitoring Supports NIBP monitoring Processes all algorithms and communicates with the main board via UART Realizes analog output (four channels, one for ECG, two for IBP, and the other for Defib sync output) Isolates the parameter modules from the earth Isolates ECG from other parameters

2-14

2.5.6 SpO2 Board

The multi-parameter module supports the independently developed Mindray SpO2 board, which provides SpO2 measurement as good as other boards but with smaller size and lower consumption. It is also compatible with Nellcor NELL-1 SpO2 board and Masimo MS-2013 SpO2 board. SpO2 board is used to collect SpO2 signals transmitted by the multi-parameter module and process SpO2 algorithms.

2.5.7 C.O. Module


The C.O. module is located on the back of the bracket of the multi-parameter module. facing the multi-parameter module with the bracket in between. The current on-line M03B module is used for C.O. measurement. The C.O. module is an independent subsystem. It is connected to the main board through the serial port and interface board.

2.5.8 CO2 Module

The CO2 module is also located on the back of the back of the bracket of the multi-parameter module, next to the C.O. module. The CO2 module is an independent subsystem. It is connected to the main board through the serial port and interface board. The socket for CO2 module on the interface board is 5-pin, different from that for C.O. module, which avoids mis-operation. Three gas modules, which can be selected in corresponding menu, are supported on the patient monitor: M02C sidestream module Capnostat mainstream CO2 module Oridion microstream CO2 module

2.5.9 Recorder
The recorder receives data from the main board and then sends them to the thermal print head for printing. The recorder has a hard key (starting/stopping recordings) and a green LED on its front. It is connected with the main board. The following diagram shows its operating principle:

2-15

Module Power interface Power module CPU Signal interface Motor drive circuit Key&LED interface

Description Introduces a DC from the main board. Converts the input power into voltages that fit each module and then forwards them to each module. Control the communications between modules. Control the communication between the main board and the recorder CPU. Receives the control signals from the CPU and then forwards them to the step engines. Includes a button and an LED which are directly controlled by the CPU.

2.6 Speaker
The speaker provides sound for alarms, key strokes, heartbeats and pulse, and allows PITCH TONE and multi-level tone modulation. It is connected with the main board and is directly driven by the main board.

2-16

3 Unpacking and Installation


3.1 Unpacking the Equipment
Open the package and take out the packing list. Check that all the articles included in the packing list are available and the quantity and specification are correct. Make sure that: All the optional parts purchased by the customer shall also be checked. Notify the supplier if provided components are not correct as compared to the packing list. In case of damage during transportation, keep the packing material and notify the supplier immediately. Keep the packing material till new equipment is accepted.

The following pictures show the patient monitor and accessory packing.

Accessorie Operator's Manual and Packing List

Main Unit

Main unit packing

Accessory packing

3-1

3.2 Preparation for Installation


3.2.1 Preparation for Installation Site
1. 2. 3. Ensure that the site meets all safety, environmental and power requirements. Check that required power sockets are available. Check that a network connector is available if the patient monitor needs to be connected to network.

Environmental Requirements
To avoid explosion hazard, do not use the equipment in the presence of flammable anesthetics, vapors or liquids. The environment where the patient monitor will be used should be reasonably free from vibration, dust and corrosive substances. If these conditions are not met, the system may not function normally. The environmental specification is as follows: Main unit Item Operating environment Storage environment Temperature () 0 to 40 -20 to 60 Relative humidity (noncondensing) 15% to 95% 10% to 95% Altitude (kPa) 57.0 to 107.4 16.0 to 107.4

Mainstream CO2 module Item Operating environment Storage environment Sidestream CO2 module Item Operating environment Storage environment Temperature () 5 to 40 -20 to 60 Relative humidity (noncondensing) 15% to 95% 10% to 95% Altitude (kPa) 57.3 to 105.3 57.3 to 105.3 Temperature () 0 to 40 -20 to 60 Relative humidity (noncondensing) 15% to 90% 10% to 90% Altitude (kPa) 57.0 to 107.4 53.3 to 107.4

Microstream CO2 module Item Operating environment Storage environment Temperature () 0 to 40 -20 to 60 Relative humidity (noncondensing) 15% to 95% 10% to 95% Altitude (kPa) 57.3 to 105.3 57.3 to 105.3

3-2

NOTE

The environmental specifications of unspecified parameters are the same as those of the main unit.

3.2.2 Electrical Requirements


Check cables and power cords. Make sure that: 1. 2. Check that the system cables, power cords, and power plugs are not damaged, and pins are not loose. In case of any damage, remove it from use. The insulation of patient cables and leadwires is not damaged, and connectors are not loose.

WARNING

Only power sockets with protective grounding can be used.

Line voltage Current Frequency

100 to 240V AC 1.1 to 0.5 A 50/60 Hz

3.2.3 Monitor Installation


Refer to iMEC12/iMEC10/iMEC8 Patient Monitor Operator's Manual.

3.2.4 Preparation for Power on


1. Before you start to make measurements, check the patient monitor for any mechanical damage and make sure that all external cables, plug-ins and accessories are properly connected. Plug the power cord into the AC power source. If you run the patient monitor on battery power, ensure that the battery is sufficiently charged. Press the power on/off switch on the monitor's front.

2. 3.

3-3

FOR YOUR NOTES

3-4

4 Hardware and Software Upgrade


4.1 Introduction
iMEC patient monitors supports upgrade of parameter modules, functional assemblies, and software (with network upgrade tool).

NOTE

Before disassembling the equipment, if necessary for the upgrade, be sure to eliminate the static charges first. When disassembling the parts labeled with staticsensitive symbols, make sure you are wearing electrostatic discharge protection such as antistatic wristband or gloves to avoid damaging the equipment. Properly connect and route the cables and wires when reassembling the equipment to avoid short circuit. Select appropriate screws to assemble the equipment. If unfit screws are tightened by force, the equipment may be damaged and the screws or part may fall off during use, causing unpredictable equipment damage or human injury. Follow correct sequence to disassembly the equipment. Otherwise, the equipment may be damaged permanently. Disconnect all the cables before disassembling any parts. Be careful not to damage any cables or connectors. Place removed screws and disassembled parts properly, preventing them from being lost or contaminated.

4.2 Hardware Upgrade


4.2.1 Upgrading the 1G SD Storage Card
4.2.1.1 Upgrade Package
Upgrade package Monitor config. before upgrade Main board (standard configuration) Main board (standard configuration) Main board (standard configuration) Main board (full configuration) Description of upgrade package 1G SD storage card upgrade kit (iMEC8) 1G SD storage card upgrade kit (iMEC10) 1G SD storage card upgrade kit (iMEC12) 1G SD storage card upgrade kit (without main board) 4-1 PN of upgrade package 801-6302-00008-00 801-6303-00006-00 801-6301-00012-00 801-6301-00026-00

1G SD storage card

4.2.1.2 Upgrade Method


1. 2. Select the correct upgrade kit based on the product model and configuration. For monitors with fully configured main board, select the upgrade kit without main board. Separate the front and rear housing as described in 7.3.1 Separating the Front and Rear Half of the Monitor. For monitors with standard configuration, the main board shall also be disassembled as described in 7.4.4 Removing the Main Board. Install the main board (only for monitors with standard configuration), SD storage card, and screws in the upgrade kit into the front housing assembly as described in the subchapters of 7.4 Disassembling the Front Housing Assembly. For iMEC8 patient monitors, the shield cover and its insulation sheet shall be installed on the main board. Re-assembly the equipment correctly.

3.

4.

When the hardware is assembled correctly as per the above procedure, connect the equipment with AC power, turn on the equipment, and select [Main Menu] [Maintenance] [Factory Maintenance >>] enter the required password [Device Config >>] to access the device configuration menu. Select [Storage Card] and close the device configuration menu. Exit factory maintenance and turn off the equipment. Then, restart the equipment, wait for a few seconds, and select [Main Menu] [Review]. If [Full Closure] is displayed in the menu, the SD card is upgraded successfully.

4.2.2 Upgrading Touchscreen


4.2.2.1 Upgrade Package
Upgrade package Monitor config. before upgrade Standard Configuration Standard Configuration Standard Configuration Description of upgrade package PN of upgrade package 801-6302-00009-00 801-6303-00007-00 801-6301-00013-00

Touchscreen

Touchscreen upgrade kit (iMEC8, with no silicon keys) Touchscreen upgrade kit (iMEC10, with no silicon keys) Touchscreen upgrade kit (iMEC12, with no silicon keys)

4.2.2.2 Upgrade Method


1. Separate the front and rear housing as described in 7.3.1 Separating the Front and Rear Half of the Monitor. Then take out the silicon keys from the original front housing and install them onto the front housing assembly in the upgrade kit as described in the subchapters of 7.4 Disassembling the Front Housing Assembly. Re-assembly the equipment correctly.

2.

4-2

When the hardware is assembled correctly as per the above procedure, connect the equipment with AC power, turn on the equipment, and select [Main Menu] [Maintenance] [Factory Maintenance >>] enter the required password [Device Config >>] to access the device configuration menu. Select [Touchscreen] and close the device configuration menu. Exit factory maintenance and turn off the equipment. Then restart the equipment, wait for a few seconds, and select [Main Menu] [Maintenance] [Cal. Touchscreen] to calibrate the touchscreen. When the calibration is done, if the message "Screen Calibration Completed!" is displayed, the touchscreen is successfully upgraded on this equipment.

4.2.3 Upgrading Wi-Fi


4.2.3.1 Upgrade Package
Upgrade package Monitor config. before upgrade Main board (standard configuration) Main board (standard configuration) Main board (standard configuration) Main board (full configuration) Description of upgrade package Wi-Fi upgrade kit (iMEC8) Wi-Fi upgrade kit (iMEC10) Wi-Fi upgrade kit (iMEC12) Wi-Fi upgrade kit (without main board) PN of upgrade package 801-6302-00010-00 801-6303-00008-00 801-6301-00014-00 801-6301-00027-00

Wi-Fi

4.2.3.2 Upgrade Method


1. 2. Select the correct upgrade kit based on the product model and configuration. For monitors with fully configured main board, select the upgrade kit without main board. Separate the front and rear housing as described in 7.3.1 Separating the Front and Rear Half of the Monitor. For monitors with standard configuration, the main board shall also be disassembled as described in 7.4.4 Removing the Main Board. Install the main board (only for monitors with standard configuration), the antenna, and the Cyberlink (Wi-Fi) module as described in 7.4.2Removing Antenna (Optional), and then manage the cables properly. For iMEC8 patient monitors, the shield cover and its insulation sheet shall be installed on the main board. Re-assembly the equipment correctly.

3.

4.

When the hardware is assembled correctly as per the above procedure, connect the equipment with AC power, turn on the equipment, and select [Main Menu] [Maintenance] [User Maintenance >>] enter the required password [Network Setup >>] [Network Type >>]. If you can select [WLAN] in the menu, the Wi-Fi function is upgraded successfully.

4-3

If Wi-Fi module function test is needed, configure the wireless network setting on the patient monitor and connect to a nearby wireless network as described in the Network Connection chapter in iMEC12/iMEC10/iMEC8 Patient Monitor Operator's Manual, and make sure the connection succeeds.

4.2.4 Upgrading Recorder


4.2.4.1 Upgrade Package
Upgrade package Recorder Monitor config. before upgrade / / / Description of upgrade package Recorder upgrade kit (iMEC8) Recorder upgrade kit (iMEC8, with tapping screws) Recorder upgrade kit (iMEC10/12) PN of upgrade package 801-6302-00011-00 801-6302-00014-00 801-6301-00015-00

4.2.4.2 Upgrade Method


1. Select the correct upgrade kit based on the product model and configuration. For iMEC8 patient monitors, the upgrade kits shall be selected based on the screw type on the original rear panel.

iMEC12/iMEC10
2. 3. Separate the front and rear housing as described in 7.3.1 Separating the Front and Rear Half of the Monitor. Fix the recorder bracket onto the rear housing with two M36 screws, as shown below; for rear housing with no hot pressed nuts, use two PT38 tapping screws to fix the recorder bracket. Screw, Phillips M36, with washer

Recorder bracket 4. Insert one end of the recorder cable into the J10 socket on the interface board and the other end to the recorder. Then fix the recorder onto the rear housing with two M36 screws as shown below: 4-4

Recorder cable 5. Re-assembly the equipment correctly.

Screw, Phillips M36, with washer

iMEC8
2. Separate the front and rear housing as described in 7.3.1 Separating the Front and Rear Half of the Monitor, and then take out the parameter module and interface panel as described in step 1 in 7.3.2 Removing the Parameter Connector Panel Assembly. Fix the recorder bracket, connect the recorder cable and install the recorder as described in steps 3 and 4 in iMEC12/iMEC10. Install the parameter module and interface panel, and reassemble the equipment correctly.

3. 4.

When the hardware is assembled correctly as per the above procedure, connect the equipment with AC power, turn on the equipment, and select [Main Menu] [Maintenance] [Factory Maintenance >>] enter the required password, and then in the maintenance menu, select [Recorder >>] and change the setting to [ON]. Exit factory maintenance and turn off the equipment. Then restart the equipment, wait for a few seconds, install the papers, and perform printing test as per 5.9 Recorder check.

4.2.5 Upgrading CO2


4.2.5.1 Upgrade Package
Upgrade package CO2 Monitor config. before upgrade / / / Description of upgrade package Mindray sidestream CO2 upgrade kit Capnostat mainstream CO2 upgrade kit Oridion microstream CO2 upgrade kit PN of upgrade package 801-6301-00016-00 801-6301-00017-00 801-6301-00018-00

4-5

4.2.5.2 Upgrade Method Upgrading Mindray Sidestream CO2


1. Remove the parameter connector assembly and parameter module assembly as described in 7.3.2 Removing the Parameter Connector Panel Assembly, and then split the original labels on the parameter connector assembly. Install the watertrap socket onto the parameter connector assembly as described in 7.7.3 Removing the Watertrap Base Assembly (Optional). Fix the gas outlet nozzle onto the parameter connector assembly with an M5 nut, as shown below:

2. 3.

M5 nut

CO2 outlet nozzle 4. 5. Install the M02C module into the parameter module assembly as described in 7.6.1 Removing the Mindray M02C CO2 Module (Optional). Connect the parameter connector assembly and gas pipes of corresponding parameter modules as described in 7.3.3 Disassembling the Parameter Module Assembly. Make sure the pipes are inserted in position. Re-assembly the equipment correctly. Select appropriate parameter labels from the upgrade kit and paste them onto the parameter connector assembly based on the configuration of the serviced equipment.

6. 7.

8. Install the AC inlet hook as shown in the figure below. Make sure the direction is correct.

AC Inlet Hook

4-6

Upgrading Capnostat Mainstream CO2


1. Remove the parameter connector assembly and parameter module assembly as described in 7.3.2 Removing the Parameter Connector Panel Assembly, and then split the original labels on the parameter connector assembly. Install the mainstream CO2 interface board to the parameter connector assembly and fix in position as described in 7.7.1 Removing the Mainstream CO2 Signal Cable (Optional). Fix the right end of the mainstream CO2 bracket into the positioning pillar in the parameter board bracket, and then fix the bracket for mainstream CO2 onto the parameter module bracket with two M36 screws. Insert the 4-pin plug of the cable between the interface board and mainstream converter board into the J1 socket on the 6301 mainstream CO2 insulation power board.

2.

3.

M36 screw, with washer

Fix the right end of the mainstream CO2 bracket into the positioning pillar in the parameter board bracket

Cable between the interface board and mainstream converter board 4. 5. 6. Connect the mainstream signal cable of the parameter connector assembly to the J3 socket as described in 7.3.3 Disassembling the Parameter Module Assembly. Re-assembly the equipment correctly. Select appropriate parameter labels from the upgrade kit and paste them onto the parameter connector assembly based on the configuration of the serviced equipment.

7. Install the AC inlet hook as shown in the figure below. Make sure the direction is correct.

AC Inlet Hook

4-7

Upgrading Oridion Microstream CO2


1. Remove the parameter connector assembly and parameter module assembly as described in 7.3.2 Removing the Parameter Connector Panel Assembly, and then split the original labels on the parameter connector assembly.

2.

Assemble the microstream CO2 connector assembly to the parameter connector assembly as described in 7.7.2 Removing the Microstream CO2 Connector Assembly (Optional). Make sure it is assembled in position. Fix the gas outlet nozzle onto the parameter connector assembly with an M5 nut, as shown below:

3.

M5 nut

CO2 outlet nozzle 4. 5. Install the microstream CO2 module into the parameter module assembly as described in 7.6.3 Removing the Microstream CO2 module (Optional). Connect the parameter connector assembly and parameter module pipe as described in 7.3.3 Disassembling the Parameter Module Assembly. Make sure the pipe is inserted in position. Then install the microstream CO2 connector into the microstream CO2 connector assembly. Re-assembly the equipment correctly. Select appropriate parameter labels from the upgrade kit and paste them onto the parameter connector assembly based on the configuration of the serviced equipment.

6. 7.

8. Install the AC inlet hook as shown in the figure below. Make sure the direction is correct.

AC Inlet Hook

4-8

When the hardware is assembled correctly as per the above procedure, connect the equipment with AC power, turn on the equipment, and select [Main Menu] [Maintenance] [Factory Maintenance >>] enter the required password [Device Config >>] to access the device configuration menu. Select the upgraded CO2 module in the CO2 module menu, close the device configuration menu, exit factory maintenance, and turn off the equipment. Then restart the equipment, wait for a few minutes, and perform CO2 module test as described in 5.4 Module Performance Tests.

4.2.6 Upgrading C.O.


4.2.6.1 Upgrade Package
Upgrade package C.O. Monitor config. before upgrade / Description of upgrade package C.O. upgrade kit PN of upgrade package 801-6301-00019-00

4.2.6.2 Upgrade Method


1. Remove the parameter connector assembly and parameter module assembly as described in 7.3.2 Removing the Parameter Connector Panel Assembly, and then split the original labels on the parameter connector assembly. Assemble the C.O. interface signal cable to the correct position on the parameter connector assembly as described in 7.6.4 Removing the C.O. Module (Optional). Insert the bottom of the C.O. board into the slot of the parameter board bracket and fix the C.O. board onto the bracket with a M36 screw. Then connect the cable between the interface board and M03B module through the P1 socket on the C.O. board.

2. 3.

M36 screw, with washer C.O. Board (M03B) Cable between interface board and M03B module

Insert the bottom of the C.O. board into the slot of the parameter board bracket 4. 5. 6. Insert the C.O. signal cable into the correct socket on the C.O. board as described in 7.3.3 Disassembling the Parameter Module Assembly. Re-assembly the equipment correctly. Select appropriate parameter labels from the upgrade kit and paste them onto the parameter connector assembly based on the configuration of the serviced equipment. 4-9

When the hardware is assembled correctly as per the above procedure, connect the equipment with AC power, turn on the equipment, and select [Main Menu] [Maintenance] [Factory Maintenance >>] enter the required password [Device Config >>] to access the device configuration menu. Select C.O. module and close the device configuration menu. Exit factory maintenance and turn off the equipment. Then restart the equipment, wait for a few minutes, and perform C.O. module test as described in 5.4.7 C.O. Test.

4.2.7 Upgrading IBP


4.2.7.1 Upgrade Package
Upgrade package Monitor config. before upgrade / / / Description of upgrade package IBP upgrade kit (for monitors with Masimo SpO2 IBP upgrade kit (for monitors with Mindray SpO2 IBP upgrade kit (for monitors with Nellcor SpO2 PN of upgrade package 801-6301-00020-00 801-6301-00021-00 801-6301-00022-00

IBP

4.2.7.2 Upgrade Method


1. 2. 3. Remove the multi-parameter module from the parameter module assembly as described in 7.6.6 Removing the Multi-parameter Module. Install the multi-parameter module in the upgrade kit, and then reassemble the parameter module assembly correctly. Install the parameter connector assembly in the upgrade kit, and then select appropriate parameter labels from the upgrade kit and paste them onto the parameter connector assembly based on the configuration of the serviced equipment. Re-assembly the equipment correctly.

4.

When the hardware is assembled correctly as per the above procedure, connect the equipment with AC power, turn on the equipment, and select [Main Menu] [Maintenance] [Factory Maintenance >>] enter the required password [Device Config >>] to access the device configuration menu. Select IBP1/IBP2 module and close the device configuration menu. Exit factory maintenance and turn off the equipment. Then restart the equipment, wait for a few minutes, and perform IBP module test as described in 5.4.6 IBP Test and Calibration.

4-10

4.2.8 Upgrading VGA/USB/Multi-functional Interface


4.2.8.1 Upgrade Package
Upgrade package Monitor config. before upgrade Main board (standard configuration) Main board (standard configuration) Main board (standard configuration) Main board (full configuration) Main board (full configuration) Description of upgrade package VGA/USB/Multi-functional interface upgrade kit (iMEC8) VGA/USB/Multi-functional interface upgrade kit (iMEC10) VGA/USB/Multi-functional interface upgrade kit (iMEC12) VGA/USB/Multi-functional interface upgrade kit (iMEC8, without main board) VGA/USB/Multi-functional interface upgrade kit (iMEC12/10, without main board) PN of upgrade package 801-6302-00012-00 801-6303-00009-00 801-6301-00023-00 801-6302-00015-00 801-6301-00028-00

VGA/USB/ Multifunctional interface

4.2.8.2 Upgrade Method


1. 2. Select the correct upgrade kit based on the configuration of the serviced equipment. For monitors with full configuration, select upgrade kits with no main boards. Remove the multi-parameter module from the parameter module assembly as described in 7.6.6 Removing the Multi-parameter Module, and then remove the power management and interface board on the rear housing. For monitors with standard configuration, the main board on the front housing shall also be removed. Install the multi-parameter module, main board (only for monitors with standard configuration), and power management and interface board. For iMEC8 patient monitors, the shield cover and its insulation sheet shall be installed on the main board, and for iMEC10/12 patient monitors, the grounding sheet shall also be installed.

3.

Shield cover for main board

Grounding sheet

(iMEC8) 4. Re-assembly the equipment correctly.

(iMEC10/iMEC12)

4-11

5.

Replace the original IO label on the rear housing with the one (with full configuration) in the upgrade kit.

IO label

When the hardware is assembled correctly as per the above procedure, connect the equipment with AC power, turn on the equipment, and perform test on the nurse call and analog output functions as described in chapters 5.5 Nurse Call Function Test and 5.6 Analog Output Performance Test. Connect to a VGA display. If the monitoring information is correctly displayed on the VGA display, the VGA function is upgraded successfully. Connect a USB device to the patient monitor through the USB connector. If the USB icon is displayed at the left corner of the screen, the USB function is upgraded successfully. Access the configuration management menu on the patient monitor and select [Main Menu] [Maintenance >>] [Manage Configuration >>] enter the required password [User Maintenance], and then [Export Config.]. When the export is finished, the message "Configurations exported successfully!" shall be displayed. Select [Import Configuration]. When the import is finished, the message "Configurations imported successfully!" shall be displayed. If so, the USB function is upgraded successfully.

4.3 Software Upgrade


You can upgrade system software and module software by installing and running the Mindray Patient Monitor Software Upgrade Tool (PN: G-110-000493-00) on a PC with Windows operating system. Connect the monitor to be upgraded and a PC running the upgrade tool to the same network, or directly connect the monitor and the PC via a crossover network cable. Then configure the IP address of the PC. You can upgrade the following software:

4-12

Software Boot program System software System software package Language lab BMP resource file LINUX platform software

PN 110-001792-00 110-001851-00 110-001795-00 110-001796-00 110-001847-00 110-001848-00 110-001861-00 110-001839-00

Description BIOS program System software Multilingual library Icon library Welcome LOGO LINUX kernel (including drive) MO software for power management board DSP (BF512) software for M51A V2.0 module Software for 9008 SpO2 module STM32 software for M02C module Software for M03B module (download online)

Module software

110-000539-00 110-001838-00 M03B-30-86661

All programs except the module software will be cleared during the upgrade of LINUX kernel (including drive). Therefore, remember to upgrade the system software and boot program etc. after the LINUX kernel (including drive) is upgraded.

4.3.1 Installing Mindray Patient Monitor Software Upgrade Tool (PN: G-110-000493-00)
1. Find the installation program installation. 2. Select language. and double click it to start

3.

Click [] and the following screen is displayed. Click [Next] to go to the next step.

4-13

4.

Enter User Name, Company name, and Serial Number "26582640". Then click [Next].

5.

Specify the destination folder for installing this program. Then select [Next].

4-14

6.

Select Program Folder. Then select [Next].

7.

Click [Finish] to complete installation.

4.3.2 Software Upgrade Procedure


1. 2. 3. 4. Connect the patient monitor to a PC through a network cable. Set the IP to 77.77.1.XX and subnet mask to 255.255.255.0. Run Mindray Patient Monitor Software Upgrade Tool on the PC and set Machine to [iMEC]. On the Mindray Patient Monitor Software Upgrade Tool screen, select [Select Package] and then the packages you want to upgrade. Then select [Start]. Start up the patient monitor and press the [Silence] + [Main Menu] hardkeys quickly to enter upgrade mode. Determine if the software is upgraded successfully according to the prompt message of the patient monitor when finishing upgrading the software. 4-15

Software upgrade is taken in the sequence of LINUX kernel (including drive) system software package (including system software, language library, BMP resource files, and boot program) Module software. When software upgrade is finished, restart the patient monitor and check if the software is correctly upgrade. For details of software upgrade, please refer to the Instructions for Use and Help of Mindray Patient Monitor Software Upgrade Tool or consult our service personnel.

CAUTION

Disconnect the equipment from the patient and make sure the important data are saved before upgrade. Do not shut down or power off the equipment when upgrading the boot software. Otherwise, it may cause the equipment to break down. Software upgrade should be performed by qualified service personnel only.

NOTE

Make sure the version of the upgrade package is what you desired. To obtain the latest upgrade package, please contact Mindray Customer Service Department.

4-16

5 Testing and Maintenance


5.1 Introduction
To ensure the patient monitor always functions normally, qualified service personnel should perform regular inspection, maintenance and test. This chapter provides a checklist of the testing procedures for the patient monitor with recommended test equipments and frequency. The service personnel should perform the testing and maintenance procedures as required and use appropriate test equipments. The testing procedures provided in this chapter are intended to verify that the patient monitor meets the performance specifications. If the patient monitor or a module fails to perform as specified in any test, repairs or replacements must be done to correct the problem. If you have any question, contact our Customer Service Department.

CAUTION

All tests should be performed by qualified service personnel only. Care should be taken to change the settings in [User Maintenance] and [Factory Maintenance] menus to avoid loss of data. Service personnel should acquaint themselves with the test tools and make sure that test tools and cables are applicable.

5.1.1 Test Equipment


See the following sections.

5.1.2 Test Report


Upon the completion of the tests, the table of preventative maintenance test reports and the table of maintenance test reports in this chapter should be kept properly.

5-1

5.1.3 Preventative Maintenance


Below are preventative maintenance tests which need to be performed on the monitor. The recommended frequency of preventative maintenance is at least once per year. See the following sections for detailed maintenance procedures. Visual inspection NIBP test and calibration CO2 test and calibration

5.1.4 Recommended Frequency


Check/Maintenance Item Visual inspection Power-on test Performance Test Calibration Frequency 1. When first installed or reinstalled. 1. When first installed or reinstalled. 2. Following any repairs or replacement of any main unit parts.

ECG test

Resp performance test SpO2 test NIBP test Pressure check Leakage test Calibration TEMP test IBP test Performance test Pressure calibration

1. If the user suspects that the measurement is incorrect. 2. Following any repairs or replacement of relevant module. 3. At least once every two years. Note: At least once a year is recommended for NIBP and CO2.

C.O. test Mainstream CO2 test Leakage test Sidestream and Microstream CO2 Performance test tests Calibration Nurse call function test Analog output performance test Electric Safety Tests Refer to A Electrical Safety Inspection.

If the user suspects that the nurse call or analog output does not work well. 1. Following any repair or replacement. 2. After the monitor drops. 3. At least once every two years. 1. When the touchscreen appears abnormal. 2. After the touchscreen is replaced. Following any repair or replacement of the recorder. 5-2

Touchscreen calibration Recorder check

Functionality test Battery check Performance test

1. When first installed. 2. Whenever a battery is replaced. Once every six months or if the battery run time reduced significantly.

5.2 Visual Inspection


Inspect the equipment for obvious signs of damage. The test is passed if the equipment has no obvious signs of damage. Following these guidelines when inspecting the equipment: Carefully inspect the case, display screen, buttons, and knob for obvious signs of damage. Inspect the power cord, wall mount, and accessories for obvious signs of damage. Inspect all external connections for loose connectors. bent pins or frayed cables. Inspect all connectors on the equipment for loose connectors or bent pins. Make sure that safety labels and data plates on the equipment are clearly legible.

5.3 Power-on Test


This test is to verify that the patient monitor can power up correctly. The test is passed if the patient monitor starts up by following this procedure: 1. 2. Insert the battery in the battery compartment and connect the patient monitor to the AC mains. The AC mains LED and battery LED light. Press the power on/off switch to switch on the patient monitor.

The monitor performs self-test as soon as the monitor is powered on. During the self-test, the alarm lamp turns yellow and red, and then turns off; the monitor gives a beep. This indicates that the visual and audible alarm indicators operate properly.

5.4 Module Performance Tests


5.4.1 ECG Test and Calibration
ECG Performance Test
Tool required: Fluke Medsim 300B patient simulator recommended

5-3

1. 2. 3. 4. 5.

Connect the patient simulator with the ECG connector using an ECG cable. Set the patient simulator as follows: ECG sinus shythm, HR = 80 bpm with the amplitude as 1mV. Check the ECG waves are displayed correctly without noise and the displayed HR value is within 80 1 bpm. Disconnect each of the leads in turn and observe the corresponding lead off message displayed on the screen. Set the simulator outputs paced signals and set [Paced] to [Yes] on the monitor. Check the pace pulse marks on the monitor screen.

ECG Calibration
Tool required: Vernier caliper

1. 2. 3. 4. 5.

Select the ECG parameter window or waveform area and set [Filter] to [Diagnostic]. Select [Main Menu] [Maintenance >>]. Select [Calibrate ECG]. A square wave appears on the screen and the message [ECG calibrating] is displayed. Compare the amplitude of the square wave with the wave scale. The difference should be within 5%. After completing the calibration, select [Stop Calibrating ECG].

You can print out the square wave and wave scale and then measure the difference between them if necessary.

5.4.2 Resp Test


Tool required: Fluke Medsim 300B patient simulator recommended

1. 2.

Connect the patient simulator to the module using a non ESU-proof cable and set lead II as the respiration lead. Configure the simulator as follows: lead II as the respiration lead, base impedance line as 1500 , data impedance as 0.5 , and respiration rate as 40 rpm.

5-4

3.

Check the Resp wave is displayed without any distortion and the displayed Resp value is within 40 2 rpm.

5.4.3 SpO2 Test


Tool required: None

1. Connect SpO2 sensor to the SpO2 connector of the monitor. Set [Patient Cat.] to [Adu] and [PR Source] to [SpO2]. 2. 3. 4. Apply the SpO2 sensor to your ring finger (assume that you stay healthy). Check the Pleth wave and PR reading on the screen and make sure that the displayed SpO2 is within 95 and 100. Remove the SpO2 sensor from your finger and make sure that an alarm of SpO2 Sensor Off is triggered.

Measurement validation: The accuracy of Mindray SpO2 board has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, and only about two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-oximeter measurements.

NOTE

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor. However, it can be used to demonstrate that a particular pulse oximeter monitor reproduces a calibration curve that has been independently demonstrated to fulfill a particular accuracy specification.

5.4.4 NIBP Test and Calibration


NIBP Leakage Test

NOTE

Perform NIBP leakage test before any other NIBP concerned test and calibration.

5-5

Tools required: NIBP cuff for adult patient Appropriate tubing Cylinder

Follow this procedure to perform the leakage test: 1. 2. 3. Set [Patient Cat.] to [Adu]. Connect the NIBP cuff to the NIBP connector on the monitor. Apply the cuff to the cylinder as shown below.

Monitor Connector for NIBP cuff Air tubing

Cylinder

Cuff

4.

Select [Main Menu] [Maintenance >>] [NIBP Leakage Test]. The message [Leakage Testing] is displayed in the NIBP parameter area.

5. The cuff automatically deflates after 20 s, which means NIBP leakage test is completed. 6. If no message is displayed in the NIBP parameter area, it indicates that the system has no leakage. If the message [NIBP Pneumatic Leak] is displayed, it indicates that the system may have a leakage. In this case, check if all connections are good and the cuff and tubing have no leakage. Perform the test again after making sure all connections are good and the cuff and tubing have no leakage.

You can either perform a manual leakage test: 1. 2. 3. 4. Perform steps 1 to 4 in the NIBP Accuracy Test section. Raise the pressure in the rigid vessel to 250 mmHg with the balloon pump. Then, wait for 5 seconds to let the measured values becoming stable. Record the current pressure value, and meanwhile use a time counter to count the time. Then, record the pressure value after 60 s. Compare the two pressure values and make sure the difference should not be greater than 6 mmHg.

5-6

NIBP Accuracy Test


Tools required: T-shape connector Appropriate tubing Balloon pump Rigid Vessel with volume 500 25 ml Reference manometer (calibrated with accuracy equal to or better than 0.75 mmHg)

Follow this procedure to perform the accuracy test: 1. Connect the equipments as shown below.

Monitor Connector for NIBP cuff

Appropriate tubing

Standard sphygmomanomet er

Balloon pump 2.

Rigid Vessel

Before inflation, check that the reading of the manometer is 0. If not, turn off the balloon pump to let the whole airway open to the atmosphere. Turn on the balloon pump after the reading is 0. Select [Main Menu] [Maintenance >>] [NIBP Accuracy Test]. Check the manometer values and the monitor values. Both should be 0 mmHg. Raise the pressure in the rigid vessel to 50 mmHg with the balloon pump. Then, wait for 10 seconds to let the measured values become stable. Compare the manometer values with the monitor values. The difference between the manometer and displayed values should be 3 mmHg. If it is greater than 3 mmHg, calibrate the monitor as described in NIBP Calibration Raise the pressure in the rigid vessel to 200 mmHg with the balloon pump. Then, wait for 10 seconds to let the measured values become stable. Repeat step 6.

3. 4. 5. 6.

7.

NOTE

You can use an NIBP simulator to replace the balloon pump and the reference manometer to perform the test.

5-7

You can use an appropriate cylinder and a cuff instead of the rigid vessel.

NIBP Calibration
Tools required: T-shape connector Appropriate tubing Balloon pump Metal Vessel with volume 500 25 ml Reference manometer (calibrated, with accuracy equal to or better than 0.75 mmHg)

1. Perform steps 1 to 4 in the NIBP Accuracy Test section. 2. 3. Select [Main Menu] [Maintenance >>] [Factory Maintenance >>] enter the required password [Calibrate NIBP >>]. In [NIBP Measurement Circuit], set [NIBP Pressure] to 150 mmHg. Raise the pump pressure to 150 mmHg. After the pressure value is stabilized, select the [Calibrate] button to start calibration. In [Overpressure Protection Circuit], set [Patient Cat.] to [Adu/Ped]. Raise the pressure to 330 mmHg. After the pressure value is stabilized, select [Calibrate] to start calibration. In [Overpressure Protection Circuit], set [Patient Cat.] to [Neo]. Raise the pressure to 165 mmHg. After the pressure value is stabilized, select [Calibrate] to start calibration. All calibration results are displayed in the [Calibrate NIBP] menu. If the calibration fails, check the test system for leakage and perform another calibration.

4.

5.

6.

5.4.5 Temp Test


Tool required: Resistance box (with accuracy above 0.1 )

1. 2. 3. 4.

Connect the two pins of any Temp connector in the monitor to the two ends of the resistance box using 2 wires. Set the resistance box to 1354.9 (corresponding temperature is 37 ). Verify each Temp channel of the monitor and make sure that the displayed value is within 37 0.1 . Repeat steps 1 to 3 and verify another temperature channel.

5-8

5.4.6 IBP Test and Calibration


IBP Performance Test
Tools required: Medsim300B patient simulator, or MPS450, or equivalent equipments IBP adapter cable for test (P/N 00-002199-00 for Medsim 300B, P/N 00-002198-00, for MPS450)

1. 2. 3. 4. 5. 6.

Connect the patient simulator through the monitor's IBP connector. Make the patient simulator outputs 0 mmHg to an IBP channel. Select IBP Zero in the IBP setup Menu to make a zero calibration. Set the patient simulator as P (static) 200 mmHg. The displayed value should be within 200 4 mmHg. If the error is beyond 4 mmHg, calibrate the pressure module. If the IBP module was calibrated with a dedicated reusable IBP sensor, check the calibration together with this IBP sensor. Let the patient simulator output 120/80 mmHg ART signal and 120/0 mmHg LV signal to the IBP channel and check that the IBP wave is displayed correctly. Repeat the steps above to calibrate other IBP channels.

7. 8.

IBP Pressure Calibration


Method 1: Tools required: Medsim300B patient simulator, or MPS450, or equivalent equipments IBP adapter cable for test (P/N 00-002199-00 for Medsim 300B, P/N 00-002198-00, for MPS450)

1. 2. 3. 4. 5. 6.

Connect the patient simulator through the monitor's IBP connector. Set the patient simulator to 0 for the desired IBP channel. Select IBP Zero in the IBP setup Menu to make a zero calibration. Set the patient simulator as P (static) 200 mmHg. Select [Main Menu] [Maintenance >>] [Cal. IBP Press. >>]. In the [Cal. IBP Press.] menu, set the calibration value to 200 mmHg. Select the [Calibrate] button next to the desired IBP channel to start calibration.

5-9

7.

If the calibration is completed successfully, the message [Calibration Completed!] will be displayed. Otherwise, a corresponding message will be displayed.

Method 2: Tools required: Standard sphygmomanometer Balloon pump Appropriate tubing T-shape connector

1.

Connect the 3-way stopcock, the sphygmomanometer and the balloon pump through a T-shape connector, as shown below.

Pressure transducer

Stopcock

Adapter cable for using disposable transducers T-shape connector

IBP module

Sphygmomanometer

2. 3. 4. 5. 6.

Vent the transducer to the atmospheric pressure by turning on the 3-way stopcock to the air. Zero the transducer, and then open the stopcock to the sphygmomanometer. Select [Main Menu] [Maintenance >>] [Cal. IBP Press. >>]. In the [Cal. IBP Press.] menu, configure the IBP calibration value. Inflate using the balloon pump until the reading of sphygmomanometer approximates the preset calibration value. Adjust the preset calibration value until it equals to the reading of the sphygmomanometer. Select the [Calibrate] button next to the desired IBP channel to start calibration.

When the calibration is completed, the message [Calibration Completed!] will be displayed. Otherwise, a corresponding message will be displayed.

5-10

5.4.7 C.O. Test


Tools required: Medsim300B patient simulator, or MPS450, or equivalent equipments C.O. adapter box ( CI-3 module/cable, P/N: 3010-0289 for 300B; P/N: 5180500 for MPS450) C.O. trunk cable (P/N: 0010-21-42716)

1. 2. 3.

Connect the patient simulator to the C.O. connector on the patient monitor using a C.O. trunk cable and a C.O. adapter box. Set the blood pressure (BT) to 37 on the patient simulator and check that the temperature value displayed on the monitor is within 37 0.2 . In the [C.O. Setup] menu on the patient monitor, set [Auto TI] to [Off], [Manual TI ()] to 2 , and [Comp. Const.] to 0.542. Select [C.O. Measure] to enter the C.O. measurement window. Select [Start] in the C.O. measurement window to start C.O. measurement. On the patient simulator, set C.O. to 5 L/min and wait for 3 to 10 seconds.

4. 5.

6. Verify that the C.O. value displayed on the monitor is 5 0.25 L/min.

5.4.8 Mainstream CO2 Tests NOTE


before performing CO2 tests, select [Main Menu] [Maintenance >>] [User Maintenance] enter the required password [Maintain CO2], and make sure that the setting of [Barometric Pressure] is correct.

Tools required: A steel gas cylinder with 6 0.05% CO2 and balance gas N2 A steel gas cylinder with 100% N2 T-shape connector Appropriate tubing Flowmeter

5-11

1. 2.

Connect the sensor. Wait for 10 seconds until the CO2 warmup is finished and then in the [CO2 Setup] menu, select [Start Zero Cal.] to start a zero calibration. If the calibration fails, the prompt message [CO2 Zero Failed] is displayed. Otherwise, the baseline of waveform recovers to zero. In the [CO2 Setup] menu, set [Apnea Delay] to 10 s. Blow to the CO2 sensor to generate a CO2 waveform and then place the sensor in the air. Check if the alarm message [CO2 Apnea] is displayed on the screen. Connect the test system as follows:

3. 4. 5.

Relief valve N2 cylinder T-shape connector

Flowmeter Mainstream sensor

T-shape connector Relief valve CO2 cylinder

Monitor

6. Adjust the power supply and turn on/off 3-way valves to ensure that only one cylinder is connected to the Mainstream CO2 sensor via the 3-way valves at one time. 7. Adjust the relief valve and make sure the flowmeter reading is stable and within 2 and 5 L/min. 8. Switch between the two cylinders to connect Mainstream CO2 sensor at an intervals of 6 to 10s and check if the displayed CO2 value is within 45 2 mmHg.

5-12

5.4.9 Sidestream and Microstream CO2Tests and Calibration


Leakage Test
1. 2. Connect the CO2 module with the patient module. Wait for 10 seconds until CO2 warmup is finished, and then use your hand or other objects to completely block the gas inlet of the module or watertrap. The sidestream and microstream CO2 modules will behave as follows:

Sidestream: the alarm message [CO2 Filter Line Err] is displayed on the screen after 3 seconds. Block the gas inlet for another 60 seconds, and select [User Maintenance >>] enter the required password [Maintain CO2 >>] [Calibrate CO2 >>] and check that the current CO2 flow is less than 10ml/min. If the alarm message continues, it indicates that the module does not leak. Microstream: The alarm message [CO2 Purging] is displayed on the screen after 3 seconds. Block the gas inlet for another 40 seconds. If alarm message [CO2 Filter Line Err] is displayed, it indicates that the module does not leak.

Accuracy Test
Tools required: A steel gas cylinder with 6 0.05% CO2 and balance gas N2 T-shape connector Appropriate tubing Flowmeter

1. 2. 3. 4.

Connect the CO2 module with the patient module. Wait for 10 minutes until the CO2 warmup is finished, and check the airway for leakage and perform a leakage test as well to make sure the airway has no leakage. Select [User Maintenance >>] enter the required password [Maintain CO2 >>] [Calibrate CO2 >>]. Connect the test system as follows:

5-13

Flowmeter

Sampling line Relief valve T-shape connector Gas bottle Monitor

5. 6.

Adjust the relief valve and make sure the flowmeter reading is stable and within 10 and 50 L/min. Check the realtime CO2 value is within 6 0.3 in the [Calibrate CO2] menu.

Calibration
Tools required: A steel gas cylinder with 6 0.05% CO2 and balance gas N2 T-shape connector Appropriate tubing

1. 2. 3.

Make sure that the sidestream CO2 module or microstream CO2 module has been warmed up or started up. Check the airway for leakage and perform a leakage test as well to make sure the airway has no leakage. Select [Main Menu] [Maintenance >>] [User Maintenance >>] enter the required password [Maintain CO2 >>] [Calibrate CO2 >>] to enter the [Calibrate CO2] menu. Then select [Zero]. After the zero calibration is finished successfully, connect the equipment as follows:

4. 5.

5-14

Flowmeter

Sampling line Relief valve T-shape connector Gas bottle Monitor

6. 7. 8.

Adjust the relief valve and make sure the flowmeter reading is stable and within 10 and 50 L/min. In the [Calibrate CO2] menu, enter the vented CO2 concentration in the [CO2] field. In the [Calibrate CO2], the measured CO2 concentration is displayed. After the measured CO2 concentration becomes stable, select [Calibrate CO2] to calibrate the CO2 module. If the calibration is finished successfully, the message [Calibration Completed!] is displayed in the [Calibrate CO2] menu. If the calibration failed, the message [Calibration Failed] is displayed. In this case, perform another calibration.

9.

5.5 Nurse Call Function Test


Tool required: Oscilloscope

1. 2. 3.

Connect the nurse call cable to the Multifunctional Connector of the patient monitor. Enter [Demo] mode. Then select [Main Menu] [Maintenance >>] [User Maintenance >>] enter the required password [Others >>]. In the [Others >>] menu, select [Nurse Call Setup >>] and then in the [Nurse Call Setup] menu, select all optional [Alm Lev] and [Alm Cat.], and set [Contact Type] to [Normally Open]. In [Nurse Call Setup >>] menu, set [Signal Type] to [Pulse]. Make the monitor to generate an alarm and check that the oscillograph displays positive pulses of 1s width when there is an alarm.

4.

5-15

5.

In [Nurse Call Setup >>] menu, set [Signal Type] to [Continuous]. Make the monitor to generate an alarm and check that the oscillograph outputs continuous high level when there is an alarm.

5.6 Analog Output Performance Test


Tools required: Medsim300B patient simulator, or MPS450, or equivalent equipments Oscilloscope

Connect the patient simulator to the monitor using an ECG or IBP cable and connect the oscillograph to the Multifunctional Connector of the patient monitor. Verify that the waves displayed on the oscillograph are identical with those displayed on the monitor.

5.7 Electric Safety Tests


See A Electrical Safety Inspection for electrical safety tests.

5.8 Touchscreen calibration


Tool required: None

1. 2. 3. 4.

Select [Main Menu] [Maintenance >>] [Cal. Touchscreen]. The symbol will appear at different positions of the screen. symbol.

Select, in turn, the central point of the

After the calibration is completed, the message [Screen Calibration Completed!] is displayed. Select [Ok] to confirm the completion of the calibration.

5-16

5.9 Recorder check


Tool required: None

1. 2.

Print ECG waveforms. The recorder should print correctly and the printout should be clear. Set the recorder to some problems such as out of paper, etc. the patient monitor should give corresponding prompt messages. After the problem is removed, the recorder should be able to work correctly. Switch automatic alarm recording for each parameter ON and then set each parameters limit outside set alarm limits. Corresponding alarm recordings should be triggered when parameter alarms occur.

3.

5.10 Battery Check


Tool required: None

Functional Test
1. 2. If the patient monitor is installed with a battery, remove the battery first. Verify that the patient monitor works correctly when running powered form an AC source.

3. Insert the battery per the procedures provided in the Operators Manual. 4. Remove the AC power cord and verify that the patient monitor still works correctly.

Performance Test
Perform the test by referring to the Battery chapter in the Operators Manual and verify the operating time of the battery meets the product specification.

5-17

5.11 Factory Maintenance


5.11.1 Accessing Factory Maintenance Menu
Select [Main Menu] [Maintenance >>] [Factory Maintenance >>] enter the required password to access the [Factory Maintenance] menu.

5.11.2 Drawing Waves


There are two methods to draw waves: [Color] and [Mono]. Color: selecting Color will have smoother waveforms. Mono: selecting Mono will have a wider viewing angle.

5.11.3 Recorder
To enable/disable the recorder, select [Recorder] and toggle between [On] and [Off].

NOTE

The recorder is disabled if [Recorder] is set to [Off] in the [Factory Maintenance] menu.

5-18

5.11.4 Software Version


Selecting the [Software Version >>] will show software version information, as shown below:

1 2

3 4

5-19

In the above figure, 1. 2. 3. 4. System software version Power software version SpO2 software version (0.0 indicates that the monitor is equipped with Nellcor or Masimo SpO2 board) MPM software version

5.11.5 Monitor information recording


Selecting [Monitor Information >>] will show the status of the patient monitor as shown below:

5-20

NOTE

If the main board is replaced, you need to check the label on the main unit and reconfigure the serial number of the patient monitor.

5.11.6 Calibrate NIBP


For details, refer to section 5.4.4 NIBP Test and Calibration. Maintenance and Test Report (See the above sections for detailed test procedures and contents) Customer name Customer address Servicing person Servicing company Equipment under test (EUT) Model of EUT SN of EUT Hardware version Software version Test equipment

Model/No.

Effective date of calibration

Test Content Visual Inspection The case, display screen, buttons, knob, power cord, wall mount, and accessories have no obvious signs of damage. The external connecting cables are not frayed and the connector pins are not loose and bent. The external connectors are not loose or their pins are not bent. The safety labels and data plate are clearly legible. Power-on test The power-on test is passed. The power indicator and alarm system work correctly and the monitor start up properly. Performance test ECG performance test and calibration ECG waves are displayed correctly without noise and the HR value is within 80 1 bpm. 5-21

Test Record

Test Result (Pass/Fail)

ECG Lead Off alarm behaves correctly. Paced signals are detected and pace pulse marks are displayed when [Paced] is set to [Yes]. The difference between the amplitude of the ECG calibration square wave and that of the wave scale is not greater than 5%. RESP test The Resp wave is not distorted and the Resp value is within 40 2 rpm. SpO2 test Measure SpO2 on a healthy persons finger and a Pleth wave and PR value are displayed. The displayed SpO2 value is within 95% and 100%. NIBP test and calibration The difference is within 3 mm when 0, 50 or 200 mmHg is set for NIBP accuracy test. There is no leakage with NIBP, or the manual leakage test result does not exceed 6 mmHg/min. Temp test The value displayed for each Temp channel of the monitor is within 37 0.1 C. IBP test and calibration The static pressure value displayed for each IBP channel is within 200 4 mmHg. The ART and LV waves for each IBP channel are displayed correctly. C.O. test The TB value displayed on the monitor is within 37 0.2 C. The displayed C.O. value is within 5 0.25 L/min. Mainstream CO2 tests The mainstream CO2 is zeroed successfully and the waveform baseline recovers to zero. CO2 Apnea alarm behaves correctly. The displayed CO2 value is within 6.0 0.3. Mainstream CO2 tests Block the gas inlet of the module or watertrap. The sidestreamCO2 flowrate is slower than 10ml/min and an alarm of CO2 Filter Line Err is given. It indicates that there is no leakage. The displayed CO2 value is within 6.0 0.3. Mainstream CO2 tests Block the gas inlet of the module or watertrap for 40s. An alarm of CO2 Filter Line Err is given. It indicates that there is no leakage. The displayed CO2 value is within 6.0 0.3. Nurse Call Function Test When an alarm is reported on the patient monitor, a nurse call is sent out though the cable. 5-22

Analog output performance test The waves displayed on the oscillograph are identical with those displayed on the monitor. Electric safety tests Refer to A Electrical Safety Inspection. All the electrical safety tests should be passed. Touchscreen calibration The touchscreen is calibrated successfully. Recorder check The recorder can print ECG waves correctly and the printout is clear. Set the recorder to some problems such as out of paper, etc. the patient monitor gives corresponding prompt messages. After the problem is removed, the recorder is able to work correctly. Automatic alarm recording for each parameter functions correctly when parameter alarms occur. Battery check The monitor can operates correctly from battery power when an AC power failure accidentally occurs. The patient monitor can operate independently on a single battery. Test conclusion (Yes No) Test date:

Tested by:

5-23

FOR YOUR NOTES

5-24

6 Corrective Action
6.1 Introduction
In this chapter, patient monitor problems are listed along with possible causes and recommended corrective actions. Refer to the tables to check the patient monitor, identify and eliminate the troubles. The troubles we list here are frequently arisen difficulties and the actions we recommend can correct most problems, but not all of them. For more information on troubleshooting, contact our Customer Service Department.

6.2 Part Replacement


Printed circuit boards (PCBs), major parts and components in the patient monitor are replaceable. Once you isolate a PCB you suspect defective, follow the instructions in 7 Disassembly and Repair to replace the PCB with a known good one and check that the trouble disappears or the patient monitor passes all performance tests. Defective PCB can be sent to us for repair. If the trouble remains, exchange the replacement PCB with the original suspicious PCB and continue troubleshooting as directed in this chapter. To obtain information on replacement parts or order them, refer to 8 Parts.

6.3 Checking Patient Monitor Status


Some troubleshooting tasks may require you to identify the hardware version and status of your monitor. To check equipment status, 1. 2. Select [Main Menu][Maintenance >>][Monitor Information >>]. Then you can view the information on system start time, self check, etc. Select [Main Menu][Maintenance >>][Factory Maintenance >>]enter the required password[Monitor Information >>]. You can also view the information on the monitors current status.

6-1

6.4 Checking Software Version


Some troubleshooting may involve software compatibility. Thus it requires you to know your monitor configuration and software version. For detailed information on version compatibility, please contact our Customer Service Department. To view information on the system configuration and system software version, 1. 2. Select [Main Menu][Maintenance >>][Software Version >>]. You can also view the information on system software version and module software version. Select [Main Menu][Maintenance >>][Factory Maintenance >>]enter the required password[Software Version>>]. You can also view the information on system software version and module software version.

6.5 Checking Technical Alarms


Before troubleshooting the patient monitor, check for technical alarm messages. If an alarm message is presented, eliminate the technical alarm first. For detailed information on technical alarm message, possible cause and corrective action, refer to the patient monitors Operation Manual.

6.6 Troubleshooting Guide


6.6.1 Power On/Off Failure
Symptoms The patient monitor fails to start. The AC power indicator, battery indicator, or startup LED does not light or fails. Possible Cause AC mains not connected or battery too low Power supply protection Cable defective Corrective Action Check that AC mains is properly connected or battery capacity is sufficient. Refer to 6.6.6 Power Supply Failures. 1. Check the power switch and the cables between the indicator board and keypad, keypad and main board, power board and interface board, and interface board and main board and see if they are properly connected. 2. Check that the cables and connectors are not damaged. Replace the power and LED indicator board. Replace the power board. Replace the connector board. Replace the main board.

Power and LED indicator board defective Power board defective The connector board failed. The main board failed.

6-2

6.6.2 Display Failures


Symptoms The display is blank or black. Possible Cause Cable defective Corrective Action 1. Check if the cable between the display and main board and the backlight cable are correctly connected. 2. Check that the cables and connectors are not damaged. Replace the main board. Replace the display. Replace the main board, or upgrade the main board with the upgrade software. 1. Check if the cable between the display and main board and the backlight cable are correctly connected. 2. Check that the cables and connectors are not damaged. 1. Check that the cable between the secondary display and the monitor is correctly connected. 2. Check that the cables and connectors are not damaged. Replace the connector board . Replace the main board. Check if there is a symbol displayed above the [Main Menu] quickkey. If yes, press and hold the [Main Menu] quickkey for more than 3 seconds to enable the touchscreen. Check that the cables between the touchscreen and touchscreen control board, and touchscreen control board and main board are properly connected. 2. Check that the cables and connectors are not loose. Replace the touchscreen control board Replace the touchscreen. Replace the main board. Calibrate the touchscreen.

Images overlapped or distorted

Main board defective Display defective Main board error Cable defective

Snows or flashing specks on the display. The display does not function.

Cable defective

Touchscreen does not respond.

The connector board failed. The main board failed. Touchscreen disabled

Cable defective

Touch position invalid

Touchscreen control board defective Touchscreen defective The main board failed. Touchscreen not calibrated

6-3

6.6.3 Alarm Lamp Failures


Symptoms The alarm lamp is not lit, extinguished, or the alarm lamp illuminates abnormally. Possible Cause Cable defective Corrective Action 1. Check that the cable between the alarm lamp board and main board is correctly connected. 2. Check that the cables and connectors are not damaged. Replace the alarm lamp board Replace the main board.

Alarm lamp board defective The main board failed.

6.6.4 Button and Knob Failures


Symptoms Buttons do not work Possible Cause Cable defective Corrective Action 1. Check that the cable between the keypad board and main board is correctly connected. 2. Check that the cables and connectors are not damaged. Replace the keypad board. Check that the cables between the knob and keypad, and keypad and main board are properly connected. 2. Check that the cables and connectors are not damaged. Replace the knob encoder. Replace the keypad board.

Knob does not work

Keypad board failure Cable defective

Knob failure Keypad board failure

6.6.5 Sound Failures


Symptoms No hardkey or knob sound, or hardkey or knob sound abnormal Possible Cause The key volume is set to zero. Cable defective Corrective Action Select [Main Menu] [Screen Setup >>] [Key Volume >>] and adjust the key volume to appropriate level. 1. Check that the cable between the speaker and interface board is properly connected. 2. Check that the cables and connectors are not damaged. Replace the speaker. Replace the main board. Replace the connector board.

Speaker defective The main board failed. The connector board failed.

6-4

No alarm sound or alarm sound abnormal

The alarm sound is set to zero.

Cable defective

Speaker defective The main board failed. The connector board failed.

Select [Main Menu] [Maintenance >>] [User Maintenance >>] enter the required password [Alarm Setup >>] and set the [Minimum Alarm Volume] to appropriate level. Select [Alarm Setup] on the main menu to adjust the alarm volume. 1. Check that the cable between the speaker and interface board is properly connected. 2. Check that the cables and connectors are not damaged. Replace the speaker. Replace the main board. Replace the connector board.

6.6.6 Power Supply Failures


Symptoms Different battery voltages. Battery cannot be fully charged. Battery cannot be recharged. Battery on-site detection error. Possible Cause Battery defective Cable defective Corrective Action Replace the battery. 1. Check that the cable between the battery interface board and interface board is properly connected. 2. Check that the cables and connectors are not damaged. Replace the battery interface board. Replace the interface board. 1. Check that the cables between power board and interface board, and AC inlet and power board are properly connected. 2. Check that the cables and connectors are not damaged. Replace the power board. 1. Turn off the patient monitor and then restart it. 2. If the problem remains, disconnect the AC mains for 5 s and reconnect it. Then restart the patient monitor. 3. If the patient monitor cannot be restarted, replace the power board.

No voltage output from the main lead or the voltage output abnormal.

Battery interface board defective The interface board failed. Cable defective

No +12 V output

No +5.0 V output

Power board defective 1. Power supply protection 2. Power board defective.

NOTE

When the battery module has a failure, it may cause problems to other components, In this case, troubleshoot the battery module per the procedure described in the table above.

6-5

Components of the main unit are powered by the power module. In the event that a component malfunctions, check if the operating voltage is correct.

6.6.7 Recorder Failures


Symptoms No printout Possible Cause Recorder module disabled Corrective Action 1. Check if the recorder status indicator lights. 2. If yes, enable the module in the [Factory Maintenance] menu. Otherwise, check for other possible causes. Re-install the paper roll. 1. Check that the cable between the recorder and main board is correctly connected. 2. Check that the cables and connectors are not damaged. Check if the power module outputs 5 V DC and 12 V DC correctly. Replace the recorder. Stop the recorder and re-install the paper roll. 1. Check the thermal print head and the paper roller for foreign matter. 2. Clean the thermal print head with an appropriate clean solution. Replace the thermal print head. Replace the recorder.

Paper reversed Cable defective

Poor print quality or paper not feeding properly

Recorder power supply failure. Recorder defective Paper roll not properly installed Print head dirty

Print head failure Recorder defective

6.6.8 Output Interface Failure


Symptoms Possible Cause Cable defective No analog out signal The connector board failed. Multi-parameter board failure Cable defective The connector board failed. The main board failed. Corrective Action 1. Check that the cable between the parameter board and interface board is properly connected. 2. Check that the cables and connectors are not damaged. Replace the connector board. Replace the multi-parameter board 1. Check that the cable between the interface board and main board is correctly connected. 2. Check that the cables and connectors are not damaged. Replace the connector board. Replace the main board.

USB device does not function (assume that the peripheral devices are good) Unable to use the USB devices USB drive data transfer failure

6-6

Improper setup

Select [Main Menu] [Maintenance >>] [User Maintenance >>] enter the required password [Others >>] and set [Data Transfer Method] to [USB Device].

6.6.9 SD Card Failure


Symptoms Possible Cause Abnormal patient admitting SD card full; unavailable for more patient data SD card fails The main board failed. Corrective Action Admit the patient properly. Delete garbage patient data, remove the related alarm, and readmit the patient. Format the SD card on the patient monitor. Then restart the monitor. If the problem remains, replace the SD card. Replace the main board.

Fails to review archived patient data

6.6.10 Wired Network Related Problems


Symptoms Unable to connect the wired network Possible Cause Incorrect LAN cable connection Incorrect IP address configuration Cable defective Corrective Action Check LAN cable connection. LAN cable shall not be longer than 50 m. Check for IP address conflict. If yes, reconfigure the IP address. 1. Check that the cable between the interface board and main board is correctly connected. 2. Check that the cables and connectors are not damaged. Replace the connector board. Replace the main board. Check LAN cable connection. LAN cable shall not be longer than 50 m.

The monitor is frequently off line or disconnects from the network. The patient monitor is connected to a LAN but cannot view other patients under the View Others mode

The connector board failed. The main board failed. Incorrect LAN cable connection

Incorrect LAN cable connection Excessive requests for viewing the patient monitor at the same time Incorrect IP address configuration

Check LAN cable connection. LAN cable shall not be longer than 50 m. A patient monitor can only be viewed by 4 other patient monitors at the same time under the View Others mode. The excessive view requests system will be ignored. Check for IP address conflict. If yes, reconfigure the IP address.

6-7

6.6.11 Wi-Fi Related Problems


Symptoms The monitor is frequently off line or disconnects from the Wi-Fi network. Possible Cause The Wi-Fi signal is unstable in the operating area. The monitor's Wi-Fi antenna is detached or not properly connected to the Wi-Fi module. Incorrect IP address configuration The Wi-Fi signal is unstable in the operating area. The monitor's Wi-Fi antenna is detached or not connected to the WiFi module. Main board defective Corrective Action Check the signal quality of the hospital Wi-Fi network. Disassemble the monitor and fix the Wi-Fi antenna.

Unable to connect the Wi-Fi network.

Check for IP address conflict. If yes, reconfigure the IP address. Check the signal quality of the hospital Wi-Fi network. Fix the Wi-Fi antenna.

Replace the main board.

6.6.12 Software Upgrade Problems


Symptoms Boot file upgrade fails Possible Cause Power failure or unintended power off during boot file upgrade Corrective Action Return the main board to factory for repair. 1. Check the network connector on the patient monitor. 2. Make sure that the hub or switch runs normally. Check that net twines are of the right type and have been connected correctly. Select package according to system requirement. Upgrade package shall be .pkg files. Configure a fixed IP address for the patient monitor. We recommend not to upgrade a program when the patient monitor is connected to a network with multiple PCs. Upgrade the power management software again and then power cycle the patient monitor.

Incorrect network connection

Program upgrade fails

Wrong upgrade package has been downloaded Incorrect IP address configuration

Battery abnormal after upgrading the power management program

Fails to power cycle the patient monitor after upgrading the power management program

6.6.13 Technical Alarm Messages


Refer to the Operator's Manual. 6-8

7 Disassembly and Repair


7.1 Tools Required
To disassemble and replace the parts and components, the following tools may be required: Philips screwdrivers Tweezers Sharp nose pliers Clamp

7.2 Preparations for Disassembly


Before disassembling the equipment, finish the following preparations: Stop patient monitoring, turn off the equipment, and disconnect all the accessories and peripheral devices. Disconnect the AC power source and remove the battery.

WARNING

Before disassembling the equipment, be sure to eliminate the static charges first. When disassembling the parts labeled with static-sensitive symbols, make sure you are wearing electrostatic discharge protection such as antistatic wristband or gloves to avoid damaging the equipment. Properly connect and route the cables and wires when reassembling the equipment to avoid short circuit. Select appropriate screws to assemble the equipment. If unfit screws are tightened by force, the equipment may be damaged and the screws or part may fall off during use, causing unpredictable equipment damage or human injury. Follow correct sequence to disassembly the equipment. Otherwise, the equipment may be damaged permanently. Be sure to disconnect all the cables before disassembling any parts. Be sure not to damage any cables or connectors. Be sure to place removed screws and disassembled parts properly, preventing them from being lost or contaminated. Place the screws and parts from the same module together to facilitate reassembling.

7-1

To reassemble the equipment, first assemble the assemblies, and then the main unit. Carefully route the cables. Make sure that the waterproof material is properly applied during reassembling.

7.3 Disassembling the Main Unit


NOTE

The recorder can be disassembled separately. To disassemble the equipment, place the equipment on a work surface free from foreign material, avoiding damaging the antiglare screen, touchscreen and the knob. Be careful not to break the two cotters on the front ends of rear housing. All the operations should be performed by qualified service personnel only. Make sure to put on the insulating gloves during service operations. Operations relating to optional parts may not apply to your equipment.

7.3.1 Separating the Front and Rear Half of the Monitor


1. Stand the equipment on the work surface. Loose and unscrew the five M410 screws as shown below:

7-2

2.

Separate the front housing assembly and rear housing assembly with caution. Disconnect the cable between the main board and keypad and then take off the front housing.

Cable between main board and interface board (iMEC10/12) (iMEC8)

NOTE

When reassembling iMEC10/12 patient monitors, the cable between the main board and interface board shall be placed downward; when reassembling iMEC8 patient monitors, the cable between main board and interface board shall be placed upward.

7.3.2 Removing the Parameter Connector Panel Assembly


1. For equipments with standard configuration, the parameter connector assembly can be disassembled separately, as shown below: Disconnect the multi-parameter signal cable, NIBP tube, and ECG signal cable, and then take out the parameter connector assembly along the rear housing. NIBP pipe Multi-parameter signal cable

Multi-parameter signal cable NIBP pipe ECG Parameter connector assembly 2.

Parameter connector assembly (iMEC10/12)

ECG signal cable

(iMEC8)

For equipments with optional C.O. and CO2 modules, remove the parameter module assembly as described in 7.3.3Disassembling the Parameter Module Assembly. 7-3

7.3.3 Disassembling the Parameter Module Assembly


1. Lay the equipment on the work surface as shown below, and unscrew the four M410 screws:

(iMEC10/12) 2.

(iMEC8)

Disconnect the multi-parameter signal cable, NIBP pipe, ECG signal cable, the cable between the parameter board and interface board, CO2 communication cable (optional), and the cable between the interface board and M03B module (optional) respectively, as shown below: Multi-parameter signal cable Cable between parameter board and interface board

NIBP pipe

ECG signal cable (iMEC10/12)

CO2 communication cable (optional)

Cable between interface board and M03B module (optional)

7-4

For iMEC8 patient monitors, the power management and interface board is located below the parameter module assembly. So, the correct sequence is: take out the parameter module assembly, and disconnect the CO2 communication cable (optional) and the cable between interface board M03B module (optional). Multi-parameter signal cable

NIBP pipe

Cable between parameter board and interface board

ECG signal cable Cable between interface board and M03B module (optional) (iMEC8)

CO2 communication cable (optional) 3.

For equipments with standard configuration, the parameter module assembly can be removed directly.

Parameter connector assembly (iMEC10/12)

Parameter module assembly (iMEC8)

4. For equipments with non-standard configurations, take the parameter connector assembly and parameter module assembly out from the rear housing and then separate the two assemblies as per the actual configuration:

7-5

For those configured with C.O. function: disconnect the C.O. signal cable, and then separate the parameter connector assembly and parameter module assembly, as shown below:

C.O. signal cable

For those configured with mainstream CO2: disconnect the mainstream CO2 signal cable, and then separate the parameter connector assembly and parameter module assembly, as shown below:

Mainstream CO2 signal cable

For those configured with microstream CO2: disconnect the CO2 pipe, take the microstream CO2 connector assembly out from the parameter connector assembly, and then separate the parameter connector assembly and parameter module assembly, as shown below:

Microstream CO2 connector assembly

CO2 outlet nozzle

7-6

For those configured with Mindray M02C CO2: disconnect the CO2 pipe, pipe 1, pipe 2, and the cable between the interface board and M02C module respectively, and then separate the parameter connector assembly and the parameter module assembly, as shown below: CO2 outlet nozzle

Communication cable between interface board and M02C module Pipe 2

Pipe 1

NOTE

All pipes should be assembled properly without occlusion during reassembly. For equipments configured with optional C.O. or CO2 modules, be sure to remove the parameter connector assembly and parameter module assembly together from the rear housing and then separate the two module. For those with standard configuration, remove the parameter module separately.

7.3.4 Removing the Battery Compartment Assembly


iMEC10/12 Loose and unscrew the four M36 screws as shown below. Disconnect the cable for battery backboard and cable for AC/DC module, and then remove the battery compartment assembly. Cable for AC/DC module

Cable for battery backboard

7-7

iMEC8
Loose and unscrew the M36 screw and then the three M38 screws as shown below. Disconnect the cable for battery backboard and cable for AC/DC module, and then remove the battery compartment assembly. Cable for AC/DC module Cable for battery backboard

7.3.5 Removing the Power Management and Interface Board


iMEC10/12
Unscrew the four M410 screws indicated below and unplug the speaker cable, and then remove the power management and interface board.

Speaker cable

iMEC8
Unscrew the three M410 screws and then the M47+10-5 screw indicated below and unplug the speaker cable, and then remove the power management and interface board.

Speaker cable

7-8

7.3.6 Removing the Recorder (Optional)


Unscrew the two M36 screws and disconnect the recorder cable to remove the recorder.

NOTE

The recorder can be disassembled separately.

7.3.7 Checking Waterproof Strips before Reassembling WARNING


Before reassembling the equipment, make sure that the waterproof material on the rear housing assembly and power base assembly is stuck to the proper places.

1.

Check that the waterproof strip is properly stuck on the battery socket.

Interface board gasket

2.

Check that the white waterproof strip is properly stuck on the front housing assembly. White waterproof strip

7-9

7.4 Disassembling the Front Housing Assembly


NOTE

To disassemble the equipment, place the equipment on a work surface free from foreign material, avoiding damaging the antiglare screen, touchscreen and the knob. Remember to install the screen support pad properly during reassembly. Operations relating to optional parts may not apply to your equipment. Position the touchscreen properly with the flexible cable facing down.

7.4.1 Removing Touchscreen Control Board (Optional)


Loose and unscrew the two M38 screws as shown below. Disconnect the touchscreen cable and cable for touchscreen control board, and then remove the touchscreen control board.

Touchscreen cable

Cable for touchscreen control board

7.4.2 Removing Antenna (Optional)


1. Take out the two antennas on the front panel from the slots as shown below: Antenna Antenna Antenna

(iMEC10)

(iMEC12)

(iMEC8)

7-10

2. Remove the antennas from the Wi-Fi module PCBA.

3.

Push the clamps aside to remove the Wi-Fi module.

7.4.3 Removing SD Card (Optional)


Unscrew the M36 screw and press the SD card as indicated below to remove the SD card:

SD card

7-11

7.4.4 Removing the Main Board


Disconnect the alarm lamp cable, cable for display backlight, display cable, and the cable between the main control board and keypad respectively. Unscrew the four M38 screws and take out the main board, as shown below: Alarm lamp cable

Cable for screen backlight Screen cable

Cable between the main board and keypad (iMEC12) Alarm lamp cable

Cable for screen backlight Cable between the main board and keypad Screen cable

(iMEC10) Alarm lamp cable

Cable for screen backlight Screen cable

Cable between the main board and keypad

(iMEC8)

7-12

NOTE

There is an additional shield cover on the front housing of iMEC8 patient monitors. But the disassembling method is the same with that of iMEC10/12 patient monitors.

7.4.5 Removing the Touchscreen (Optional)


iMEC10/12
Unscrew the ten M38 screws indicated below. Take out the touchscreen bracket and then the touchscreen.

Touchscreen

iMEC8
Unscrew the eight M38 screws indicated below. Take out the touchscreen bracket and then the touchscreen.

To

7-13

7.4.6 Disassembling the Screen


iMEC10/12
1. Unscrew the ten M38 screws indicated below. Then take out the screen bracket.

2.

Unscrew the four M38 screws indicated below to remove the screen.

iMEC8
1. Unscrew the eight M38 screws indicated below. Then take out the screen bracket.

7-14

2.

Unscrew the four M2.56 screws indicated below to remove the screen.

7.4.7 Removing the Keypad


iMEC10/12
Unplug the encoder cable and unscrew the five PT38 screws indicated below. Take out the keypad.

Encoder cable

iMEC8
Unplug the encoder cable and unscrew the five PT38 screws indicated below. Take out the keypad.

Encoder cable

7-15

7.4.8 Removing the Encoder


Poke the encoder knob out from the slot and loose the nut with a sharp nose plier. Take out the encoder.

Encoder nut

Knob

7.4.9 Removing the Alarm Lamp Board


Unscrew the two PT38 screws indicated below and take out the alarm lamp board.

7.4.10 Removing the ON/OFF Key Board


Unscrew the two PT38 screws indicated below and take out the ON/OFF key board.

7-16

7.5 Disassembling the Battery Compartment Assembly


7.5.1 Removing the Power Board
Disconnect the cable for AC input receptacle. Remove the three M36 screws indicated below and remove the power board. Cable for AC input receptacle

NOTE

Follow the instructions and unscrew the correct screws.

7.5.2 Removing the Battery Interface Board Assembly


iMEC10/12
Unscrew the two M3 nuts with a sharp nose plier and remove the battery interface board assembly.

7-17

iMEC8
Unscrew the two M36 screws with a screwdriver and then take out the battery interface board.

7.6 Disassembling the Parameter Module Assembly


NOTE

Perform disassembly with exercise care to protect the boards on both sides of the module from damage. Be sure to reassemble the fast valve (red) and slow vale (blue) in correct sequence. The slow valve should be connected to the reducer-equipped end of the pipe. Operations relating to optional parts, like C.O. and CO2, may not apply to your equipment.

7.6.1 Removing the Mindray M02C CO2 Module (Optional)


Unscrew the four M38 screws indicated below and then take out the Mindray M02C CO2 module.

7-18

7.6.2 Removing the Mainstream CO2 Module (Optional)


Unscrew the three M36 screws indicated below and then take out the mainstream CO2 board.

7.6.3 Removing the Microstream CO2 module (Optional)


1. Unscrew the four M36 screws indicated below and then take out the microstream CO2 module.

2.

Then unscrew the three M36 screws and then take out the microstream converter, as shown below:

7-19

7.6.4 Removing the C.O. Module (Optional)


Unscrew the M36 screw indicated below and take out the C.O. module.

7.6.5 Removing the SpO2 Board


Unscrew the two M36 screws indicated below and take out the SpO2 board.

(Nellcor)

(Masimo)

7.6.6 Removing the Multi-parameter Module


Disconnect the pipe from the module. Then unscrew the four M36 screws and take out the multi-parameter module, as shown below:

7-20

7.6.7 Removing the Pump and Valve


1. Clip the cable holders indicated below with a clamp and take out the pump.

2.

Push the four clamps indicated below aside and take out the valve.

7.7 Removing the Parameter Connector Assembly


7.7.1 Removing the Mainstream CO2 Signal Cable (Optional)
1. Push the four clamps aside as shown below and take out the mainstream CO2 interface board.

Mainstream CO2 interface board

7-21

2.

Screw the connector of the mainstream signal cable clockwise with a sharp nose plier. Pull the connector up to remove the mainstream signal cable.

Mainstream CO2 signal cable

7.7.2 Removing the Microstream CO2 Connector Assembly (Optional)


Push the four clamps aside as shown below and take out the microstream CO2 interface assembly.

Microstream CO2 connector assembly

7.7.3 Removing the Watertrap Base Assembly (Optional)


1. Unscrew the two screws of the watertrap indicated below with a screwdriver and remove the watertrap base assembly.

Screw of the watertrap 7-22

2.

Unscrew the four PT26 screws indicated below and take out the watertrap base bracket.

7.7.4 Removing the Cables from the Parameter Connector Assembly


To remove the cables, loose the connectors clockwise with a sharp nose plier; To reassemble the cables, screw the connectors counter-clockwise to the correct sockets as shown below: ECG

SpO2 Temp 1 Temp 2 IBP 1 IBP 2

C.O.

Make sure that the connector is installed in place.

7-23

7.7.5 Removing the Recorder


1. Push the two clamps aside and take out the recorder drive board, as shown below:

Clamps

Recorder drive board

2.

Unscrew the two PT26 screws and disconnect the cable between the recorder drive board and recorder keypad and the flexible cable, as shown below. Then remove the thermal print head and recorder drive board. Cross recessed pan head self-tapping screw PT26

Thermal print head

Flexible cable

Cable between the recorder drive board and recorder keypad

3.

Unscrew the two PT26 screw indicated below and take out the keypad.

Keypad

Cross recessed pan head self-tapping screw PT26

7-24

8 Parts
8.1 Introduction
This section contains the exploded views and parts lists of the main unit. It helps the engineer to identify the parts during disassembling the patient monitor and replacing the parts. This manual is based on the maximum configuration. Your equipment may not have same parts and the quantity of the screws or stacking sleeves etc. may be different with those included in the parts lists. Hardware architecture of the main unit is shown below:

NOTE

The part number listed in the Parts List is only for checking the FRU part number which is also included in the Parts List. Please provide the FRU parts number if you want to purchase the spare parts.

8-1

8.2 iMEC8
8.2.1 Main unit
8.2.1.1 Exploded View

8.2.1.2 Parts List


SN 1 2. / 3. 4. / / PN / Description Front panel subassembly (8 Inch/WiFi+ 1G SD card) Rear housing of the main unit (8 inch/Network/VGA/USB/Anolog input & output) iMEC12 label (Chinese) Cross recessed pan head screw GB/T818-2000 M410 FRU part number / / / / / / Remarks / /

8-2

8.2.2 Front Housing Assembly (Nakescreen)


8.2.2.1 Exploded View

8.2.2.2 Parts List


SN 1 2. Front housing (8 inch) 043-001965-01 Front housing (8 inch) 3. / 6301 Antenna attached PCB 8-3 801-6302-00016-00 115-010845-00 801-6302-00005-00 PN / Description Knob FRU part number 043-001790-00 Remarks / For monitors with machine screws For monitors with tapping screws /

4. 5. 6.

/ / / / / / / / /

Antenna, 2400-2500MHz 2.15dBI IPEX-II interface Encoder Key (8 inch, Chinese) Key (8 inch, English) Key (8 inch, Italian) Key (8 inch-Spanish) Key (8 inch, Portuguese) Key (8 inch, French) Key (8 inch, German) Keypad PCBA, 8.4 inch Screw, self-tapping, PT38 Encoder cable PCB shielding insulator Screw, Pan head with washer, Phillips M36 PCB-shielding 6301 Main board PCBA (Full configuration ) 6301 Main board PCBA (Standard configuration) Screw, pan head Phillips, M36 Display bracket (8 inch nakescreen) Truly LED display, 8.4" AU LED, TFT, 8.4" Screw, pan head Phillips, M2.56 Long gasket, 8 nakescreen Short gasket, 8 nakescreen Power button 6301 On/Off key board PCBA 6301 alarm board PCBA Alarm lamp gasket Alarm lamp

0010-30-43089 049-000334-00 049-000350-00 049-000380-00 049-000381-00 049-000382-00 049-000383-00 049-000384-00 051-000877-00 / 9200-21-10460 / / / 801-6302-00001-00

/ As configured

7. 8. 9. 10. 11. 12. 13.

/ / / / / / 051-000829-00 051-000964-00

/ / / Optional / Optional As configured

801-6302-00002-00 / / 021-000078-00 021-000060-00 / / / 6802-20-66691 051-000880-00 801-6301-00005-00 / / / / As configured/No t mutually compatible nor exchangeable / / / / / / / /

14. 15. 16.

/ / / /

17. 18. 19. 20. 21. 22. 23. 24.

/ / / / / 051-000879-01 / /

8-4

8.2.3 Front Housing Assembly (Touchscreen)


8.2.3.1 Exploded View

8.2.3.2 Parts List


SN 1 2. PN / 043-001965-01 Description Knob Front housing (8 inch) Front housing (8 inch) 3. 4. 5. / / / 6301 Antenna attached PCB Antenna Knob 8-5 FRU part number 043-001790-00 801-6302-00006-00 801-6302-00017-00 115-010845-00 0010-30-43089 Remarks / For monitors with machine screws For monitors with tapping screws / /

6.

/ / / / / / / / / / / / / 051-000829-00 / / / 021-000078-00 021-000060-00

Key (8 inch, Chinese) Key (8 inch, English) Key (8 inch, Italian) Key (8 inch-Spanish) Key (8 inch, Portuguese) Key (8 inch, French) Key (8 inch, German) 6301 keypad PCBA 8.4 inch Screw, self-tapping, PT38 Encoder cable Screw, Pan head with washer, Phillips M36 PCB shielding insulator PCB-shielding Main board PCBA Screw, pan head Phillips, M36 Touchscreen control board PCBA Screen bracket (8 inch touchscreen) Truly LED display, 8.4" AU LED, TFT, 8.4" Screw, Phillips, M2.56 Long pressing strip for 8 inch display bracket Short pressing strip for 8 inch display bracket Touchscreen, resistive-type, 8.4" 4-line Long gasket, 8 touchscreen Short gasket, 8 touchscreen Power button 6301 On/Off key board PCBA 6301 alarm board PCBA Alarm lamp gasket Alarm lamp

049-000334-00 049-000350-00 049-000380-00 049-000381-00 049-000382-00 049-000383-00 049-000384-00 051-000877-00 / 9200-21-10460 / / / 801-6302-00001-00 / 051-000881-00 / 801-6302-00013-00 801-6302-00003-00 / / / 021-000058-00 / / 6802-20-66691-51 051-000880-00 801-6301-00005-00 / /

7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17.

/ / / / Optional Optional / / / / As configured/No t mutually compatible nor exchangeable / / / / / / / / / / /

18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28.

/ / / / / / / / 051-000879-01 / /

8-6

8.2.4 Rear Housing (Network/VGA/USB/Anolog Input & Output)


8.2.4.1 Exploded View

8.2.4.2 Parts List


SN 1 2. 3. 4. 5. 6. PN / / / / / / 051-000850-00 051-000965-00 7. / / Description Rear Housing Assembly (8 Inch) IO label (all) TR6F recorder Recorder door Screw, M38,Cross recessed countersunk head, GB/T819.1-2000 Battery Compartment Assembly (8 inch) PCBA,6301 Power management/interface board 6301 Power management and interface board PCBA (with standard configuration) Parameter module assembly (ECG+Temp+NIBP+Mindray SpO2) Parameter module assembly (ECG+Temp+NIBP+Nellcor SpO2) 8-7 FRU part number 115-010391-00 / TR6F-30-67306 9211-20-87402 / 115-010390-00 801-6301-00001-00 801-6301-00002-00 / / As configu red Remar ks / / / / / / As configu red

/ / / / 8. 9. 10. 11. / / / / / / / / / 12. /

Parameter module assembly (ECG+Temp+NIBP+Masimo SpO2) Parameter module assembly (ECG+Temp+NIBP+Mindray SpO2+IBP) Parameter module assembly (ECG+Temp+NIBP+Nellcor SpO2+IBP) Parameter module assembly (ECG+Temp+NIBP+Mindray SpO2, analog output) Cross recessed pan head self-tapping screw PT 414 Interface board gasket Cross recessed pan head self-tapping screw PT38 Parameter connector panel assembly (ECG+Temp+NIBP+Mindray SpO2) Parameter connector panel assembly (ECG+Temp+NIBP+Nellcor SpO2) Parameter connector panel assembly (ECG+Temp+NIBP+Masimo SpO2) Parameter connector panel assembly (ECG+Temp+NIBP+Mindray SpO2+IBP) Parameter connector panel assembly (ECG+Temp+NIBP+Nellcor SpO2+IBP) Parameter connector panel assembly (ECG+Temp+NIBP+Masimo SpO2+IBP) Thermal conductive gel for AC board

/ / / / / / / As configu red

/ / / 115-010355-00 115-010356-00 115-010357-00 115-010358-00 115-010359-00 115-010360-00

8-8

8.2.5 Rear Housing Assembly


8.2.5.1 Exploded View

8.2.5.2 Parts List


SN 1 2. 3. 4. 5. 6. 7. 8. 9. PN 043-001793-00 043-001966-01 / / / / / / / Description Handle (8 inch) Rear housing (8 inch) Buffer cushion, Bumpon SJ5302 Battery door subassembly (8 inch) Rubber feet Hook bracket Cross recessed pan head self-tapping screw PT38 Handle waterproof pad Handle bracket FRU part number 801-6302-00007-00 115-010391-00 / 801-6302-00004-00 / / / / / Remark s / / / / / / / / /

8-9

10. 11. 12. 13.

/ / / /

Speaker, 2W, 4ohm, 60mm Gasket, speaker pressure plate and spacer Gasket, speaker spacer Cross recessed pan head self-tapping screw PT38

020-000015-00 / / /

/ / / /

8.2.6 Battery Compartment Assembly


8.2.6.1 Exploded View

8.2.6.2 Parts List


SN 1 2. 3. PN / / / Description Main bracket (8 inch) Bracket insulator Grounding terminal FRU part number / / / Rema rks / / /

8-10

4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14.

/ / / / / / / / / / /

Serrated lock washers external teeth 6 Cable for AC input receptacle Battery interface board M48 combined screw Screen sponge, D-shape, 12.79.5mm Spring, EMI Battery Spring (8 inch) Power board insulator 6301 AC/DC power PCBA Screw, Phillips M36, with washer Knob, battery latch

/ 009-002107-00 051-000891-00 / / / / / 051-000824-00 / /

/ / / / / / / / / / /

8.2.7 Parameter module assembly


8.2.7.1 Exploded View

8-11

8.2.7.2 Parts List


SN 1 2. 3. PN / / / / 051-000950-00 051-000949-00 4. 5. 6. / / / / 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. / / / / / / / / / / / / / / / / / / / / / / Description Parameter PCBA bracket Hexagon plastic nut, M30.5P, PC MEC Multi-parameter board PCBA (complete) MEC Multi-parameter board PCBA (standard) MEC Multi-parameter PCBA (China) MEC Multi-parameter PCBA (China) Screw, self-tapping, PT38 Hexagonal plate isolator Nellcor SpO2 board Masimo MS-2013 SpO2 board (For Shenzhen Mindray only) 9008 V2.0 SpO2 board PCBA Screw, pan head Phillips, M36 Connector, Elbow, 200Barb, 3/32"ID, White Nylon NIBP pipe 2 NIBP pipe 7 Plastic connector Airway tube 2 NIBP pipe 5 NIBP pipe 3 Filter, inline filter 43um 1/8 I.D.Tubing NIBP pipe 6 Connector, Tee, 200Barb, 3/32"ID, White Nylon NIBP pipe 4 Pump, 12V DC, with 120 mm wire and connector Cable holder, CHS-4150mm, Nylon C.O./IBP (M03B) module Mindray sidestream CO2 module kit Capnostat Mainstream CO2 module kit Ordion Microstream CO2 module kit 630F reducer Airway tube Gas valve, fast, 12VDC 300mmHg FRU part number / / 051-000950-00 051-000949-00 801-6301-00006-00 801-6301-00007-00 / / 0671-00-0102-01 040-001149-00 051-000943-00 / / / / / / / / / / / / 082-000862-00 / M03B-30-26064 / / / / / 082-000861-00 / / / / / / / / / / / / / / / / / / / Optional As configured / / / As configured (China) Remarks / / As configured (international)

8-12

8.2.8 Parameter connector panel assembly


8.2.8.1 Exploded View

8.2.8.2 Parts List


SN 1 PN / / / / / / / / Description Parameter label (standard) Label (standard+IBP+C.O.+sidestream/microstream CO2) Label (standard+IBP) Label (standard+C.O.) Label (standard+C.O.+sidestream/microstream CO2) Label (standard+mainstream CO2) Label (standard+C.O.+sidestream/microstream CO2) Label (standard+C.O.+mainstream CO2) 8-13 FRU part number 047-005546-00 047-005695-00 047-005756-00 047-005757-00 047-005758-00 047-005759-00 047-005760-00 047-005761-00 Remarks As configured

/ / / / 2. 3. 4. 5. / / / / / / 6. 7. 8. 9. 10. 11. 12. / / / / / / / / / / / /

Label (standard+IBP+C.O.) Label (standard+IBP+sidestream/microstream CO2) Label (standard+IBP+mainstream CO2) Label (standard+IBP+C.O.+mainstream CO2) NIBP pedestal (hexagon) Steal ball, 2.5 Gas outlet Microstream CO2 connector assembly Watertrap pedestal assembly (M02C) Mainstream CO2 interface board Parameter connector assembly Nut, Stainless Steel M5 GB6170 Silicon jacket NIBP external pedestal ECG signal cable C.O. signal cable Multi-parameter signal cable, with IBP, Mindray Multi-parameter signal cable (Nellcor SpO2+Temp+IBP) Multi-parameter signal cable (Masimo SpO2+Temp+IBP) Multi-parameter signal cable (Mindray SpO2+Temp) Multi-parameter signal cable (Nellcor SpO2+Temp) Multi-parameter signal cable (Masimo SpO2+Temp)

047-005762-00 047-005763-00 047-005764-00 047-005765-00 / / / 9201-30-35923 115-010540-00 / / / / / 009-001678-00 009-001993-00 009-001679-00 009-001680-00 009-001681-00 009-002123-00 009-002124-00 009-002125-00 / / / / / Optional As configured / / / As configured

8-14

8.2.9 Mindray Sidestream CO2 Module Kit


8.2.9.1 Exploded View

8.2.9.2 Parts List


SN 1 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. PN / / / / / / / / / / / / / / / Description CO2 module unit (M02C) Cross recessed pan head self-tapping screw PT310, stainless steel AG airway sampling line (T=10) AG airway sampling line (T=50, U-type) Silicon tube (T=15) AG airway sampling line (T=50) Air filter GB8368-98 AG airway sampling line (T=90) Cross recessed pan head self-tapping screw PT2.06 AG watertrap bracket AG watertrap receptacle AG airway sampling line (T=30) AG airway sampling line (T=110) Moisture exchanger (Nafion Tube for Mindray CO2) AC Inlet Hook 8-15 FRU part number 115-010077-00 / / / / / / / / / 115-010540-00 / / / 9211-20-87369 Rema rks / / / / / / / / / / / / / / /

8.2.10 Oridion Microstream CO2 Module Kit


8.2.10.1 Exploded View

8.2.10.2 Parts List


SN 1 2. 3. 4. 5. 6. PN / / / / / / Description 6301 Microstream converter PCBA Microstream CO2 bracket Microstream CO2 module, MiniMediCO2-9 Cross recessed pan head screw GB/T818-2000 M36 Cross recessed pan head self-tapping screw PT38 AC Inlet Hook FRU part number 051-000892-00 / 0010-10-42559 / / 9211-20-87369 Remarks / / / / / /

8-16

8.2.11 Capnostat Mainstream CO2 module kit


8.2.11.1 Exploded View

8.2.11.2 Parts List


SN 1 2. 3. 4. 5. 6. PN / / / / / / Description Mainstream CO2 bracket Power isolation board PCBA of 6301 mainstream CO2 Screw, Pan head with washer, Phillips M36 Screw, self-tapping, PT38 AC Inlet Hook Mainstream CO2 interface board FRU part number / 051-001038-00 / / 9211-20-87369 / Remarks / / / / / /

8-17

8.2.12 TR6F Recorder Kit


8.2.12.1 Exploded View

8.2.12.2 Parts List


SN 1 2. 3. 4. PN / / / / Description Recorder bracket Cross recessed pan head selftapping screw PT38 TR6F recorder Screw, Phillips M36, with washer FRU part number / / TR6F-30-67306 / Remarks / / / /

8-18

8.2.13 Battery charge board assembly (12 inch)


8.2.13.1 Exploded View

8.2.13.2 Parts List


SN 1 2. 3. 4. PN / / / / Description 6301 battery interface PCBA Battery spring Screw, Pan head with washer, Phillips M38 M3 nut with spring washer FRU part number 051-000891-00 / / / Remarks / / / /

8-19

8.3 iMEC10
8.3.1 Main unit
8.3.1.1 Exploded View

8.3.1.2 Parts List


SN 1 2. 3. 4. PN Description Front panel assembly (12 Inch) Front panel assembly (10 Inch) Rear housing assembly (12 inch) Screw, pan head Phillips, M410 Main unit label FRU part number / / / / / / / / Remarks As configured

/ / / /

8-20

8.3.2 Front Panel Assembly (10 Inch/Touchscreen/Wi-Fi+ 1G SD Card)


8.3.2.1 Exploded View

8.3.2.2 Parts List


SN 1 2. 3. 4. 5. 6. 7. 8. 9. 10. PN / / / 051-000829-00 / / / / / / Description Touchscreen control board PCBA Screw, pan head Phillips, M36 Screw, Pan head with washer, Phillips M38 Main board PCBA Screw, self-tapping, PT38 Fujitsu touch screen kit (10 Inch LCD) Screw, pan head Phillips, M26 Backlight board insulator Inverter, 12VDC/500Vrms 6mA signal output Touch panel, resistive-type, 10.4" 4-line 8-21 FRU part number 051-000881-00 / / 801-6303-00001-00 / / / / 022-000070-00 021-000005-00 Remarks / / / / / / / / / /

11. 12. 13.

/ / / / / / / / /

DMC touchscreen position pad 6301 keypad PCBA 12.1 inch Key (12 inch, Chinese) Key (12 inch, English) Key (12 inch, Italian) Key (12 inch, Spanish) Key (12 inch, Portuguese) Key (12 inch, French) Key (12 inch, German) Encoder board Knob Front housing (10 inch)

/ 051-000878-00 049-000335-00 049-000351-00 049-000385-00 049-000386-00 049-000387-00 049-000388-00 049-000389-00 0010-30-43089 043-001790-00 801-6303-00005-00 801-6303-00012-00 115-010845-00 / / 801-6301-00005-00 6802-20-66691-51 051-000880-00 / / /

/ / As configured

14. 15. 16.

/ /

/ / For monitors with machine screws For monitors with tapping screws / / / / / / / / /

043-001973-01 Front housing (10 inch) 17. 18. 19. 20. 21. 22. 23. 24. 25. / / / 051-000879-01 / / / / / Antenna and cable assembly Alarm lamp shade Alarm lamp gasket Alarm lamp board PCBA Power button 6301 On/Off key board PCBA FUJI touchscreen position pad (10 inch) 10 inch touchscreen long gasket 10 inch touchscreen short gasket

8-22

8.3.3 Fujitsu Touchscreen Kit (10 Inch LCD)


8.3.3.1 Exploded View

8.3.3.2 Parts List


SN 1 2. 3. PN / / 021-000014-00 021-000062-00 4. 5. 6. / / / Description Screen bracket 2 (10 inch touchscreen) 10 inch display buffer cushion (nakescreen) LCD, TFT10.4'', 800600 LCD, TFT, 10.4", 800*600, LED-BackLight Touchscreen pressing strip 4 (10 inch) Touchscreen pressing strip 3 (10 inch) Screw, pan head Phillips, M26 FRU part number / / 801-6303-00010-00 801-6303-00003-00 / / / / / / Remarks / / Main Backup As configured/Not mutually compatible nor exchangeable

8-23

8.3.4 Front Housing Assembly (10 Inch)


8.3.4.1 Exploded View

8.3.4.2 Parts List


SN 1 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. PN / 051-000964-00 / / / / / / / / / / Description Screw, Pan head with washer, Phillips M38 6301 Main board PCBA (Standard configuration) Tianma LCD display kit (10 inch) Screw, self-tapping, PT38 Screw, pan head Phillips, M26 Backlight board insulator Inverter, 12VDC/500Vrms 6mA signal output 10 inch nakescreen long gasket 10 inch nakescreen short gasket 6301 keypad PCBA 12.1 inch Key (12 inch, Chinese) Key (12 inch, English) 8-24 FRU part number / 801-6303-00002-00 / / / / 022-000070-00 / / 051-000878-00 049-000335-00 049-000351-00 Remarks / / / / / / / / / / As configured

/ / / / / 12. 13. /

Key (12 inch, Italian) Key (12 inch, Spanish) Key (12 inch, Portuguese) Key (12 inch, French) Key (12 inch, German) Encoder board Front housing (10 inch)

049-000385-00 049-000386-00 049-000387-00 049-000388-00 049-000389-00 0010-30-43089 801-6303-00004-00 801-6303-00011-00 043-001790-00 / / 801-6301-00005-00 6802-20-66691-51 051-000880-00 / For monitors with machine screws For monitors with tapping screws / / / / / /

043-001973-01 Front housing (10 inch) 14. 15. 16. 17. 18. 19. / / / 051-000879-01 Knob Alarm lamp shade Alarm lamp gasket Alarm lamp board PCBA Power button 6301 On/Off key board PCBA

8.3.5 Tianma LCD display kit (10 inch)


8.3.5.1 Exploded View

8-25

8.3.5.2 Parts List


SN 1 2. 3. PN / / 021-000014-00 Description LCD display bracket 1 (10 inch touchscreen) 10 inch display buffer cushion (nakescreen) LCD, TFT10.4'', 800600 (Tianma) FRU part number / / 801-6303-00013-00 Remarks / / As configure d/Not mutually compatibl e nor exchange able /

LCD, TFT 10.4", 800600AU

021-000062-00

4.

Screw, pan head Phillips, M26

8.3.6 Rear Housing Assembly


8.3.6.1 Exploded View

8-26

8.3.6.2 Parts List


SN 1 2. 3. 4. PN / / / / / / / / / 5. / / / / / / 6. 7. 8. 9. 10. 11. / / / / / 051-000850-00 051-000965-00 12. 13. 14. / / / / Description Rear Housing Assembly (12 Inch) Interface board gasket Screw, self-tapping, PT414 Parameter module assembly (ECG+Temp+NIBP+Mindray SpO2) Parameter module assembly (ECG+Temp+NIBP+Nellcor SpO2) Parameter module assembly (ECG+Temp+NIBP+Masimo SpO2) Parameter module assembly (ECG+Temp+NIBP+Mindray SpO2+IBP) Parameter module assembly (ECG+Temp+NIBP+Nellcor SpO2+IBP) Parameter module assembly (ECG+Temp+NIBP+Mindray SpO2+analog output) Parameter connector assembly (ECG+Temp+NIBP+Mindray SpO2) Parameter connector assembly (ECG+Temp+NIBP+Nellcor SpO2) Parameter connector assembly (ECG+Temp+NIBP+Masimo SpO2) Parameter connector assembly (ECG+Temp+NIBP+Mindray SpO2+IBP) Parameter connector assembly (ECG+Temp+NIBP+Nellcor SpO2+IBP) Parameter connector panel assembly (ECG+Temp+NIBP+Masimo SpO2+IBP) Spring, EMI Earth bracket Screw, self-tapping, PT38 Battery compartment subassembly (12 inch) Recorder bracket (12 inch) PCBA,6301 Power management/interface board 6301 Interface board PCBA (STD) Screw, Pan head with washer, Phillips M36 TR6F recorder Recorder door IO label (all) IO label 8-27 FRU part number 115-010374-00 / / / / / / / / 115-010355-00 115-010356-00 115-010357-00 115-010358-00 115-010359-00 115-010360-00 / Optional / / / As configured / / / As configured As configured Remarks / / / As configured

/ / / 115-010373-00 / 801-6303-00001-00 801-6303-00002-00 / TR6F-30-67306 9211-20-87402 / /

8.3.7 Rear Housing Assembly


8.3.7.1 Exploded View

8.3.7.2 Parts List


SN 1 2. 3. 4. 5. 6. 7. PN 043-001970-01 / 043-001800-00 / / / / Description Rear housing (12 inch) Buffer washer, Bumpon SJ5302 white and transparent Handle (12 inch) Handle waterproof pad Handle bracket Hook bracket Screw, FT38, self-tapping FRU part number 115-010374-00 / 801-6301-00011-00 / / / / Remarks / / / / / / /

8-28

8. 9. 10. 11. 12. 13. 14.

/ / / / / / /

Speaker, 2W, 4ohm, 60mm Gasket, speaker pressure plate and spacer Screw, self-tapping, PT38 Gasket, speaker spacer Rubber feet Battery door (12 inch) Battery door belt

020-000015-00 / / / / 115-010376-00

/ / / / / /

8.3.8 Battery Compartment Assembly


8.3.8.1 Exploded View

8.3.8.2 Parts List


SN 1 2. 3. PN / / / Description Main bracket (12 inch) Knob, battery latch Battery spring washer FRU part number / / / Remarks / / /

8-29

4. 5. 6. 7. 8. 9. 10. 11. 12.

/ / / / / / / / /

Serrated lock washers external teeth 6 Grounding terminal Cable for AC input receptacle Battery interface board M3 nut with spring washer Screw, Pan head with washer, Phillips M36 6301 AC/DC power PCBA M48 combined screw Spring, EMI

/ / 009-002107-00 051-000891-00 / / 051-000824-00 / /

/ / / / / / / / /

8.3.9 Parameter Module Assembly


For the exploded view and parts list of the parameter module assembly, refer to Section 8.2.7 Parameter module assembly.

8.3.10 Parameter Connector Assembly


For the exploded view and parts list of the parameter connector assembly, refer to Section 8.2.8 Parameter connector panel assembly.

8.3.11 Mindray Sidestream CO2 Module Kit


For the exploded view and parts list of Mindray sidestream CO2 module assembly, refer to Section 8.2.9 Mindray Sidestream CO2 Module Kit.

8.3.12 Oridion Microstream CO2 Module Kit


For the exploded view and parts list of Oridion microstream CO2 module assembly, refer to Section 8.2.10 Oridion Microstream CO2 Module Kit.

8.3.13 Capnostat Mainstream CO2 Module Kit


For the exploded view and parts list of Capnostat mainstream CO2 module assembly, refer to Section 8.2.11 Capnostat Mainstream CO2 module kit.

8-30

8.3.14 TR6F Recorder Kit


For the exploded view and parts list of the TR6F recorder kit, refer to Section 8.2.12 TR6F Recorder Kit.

8.3.15 Battery Interface Board Assembly


For the exploded view and parts list of the battery interface board assembly, refer to Section 8.2.13 Battery charge board assembly (12 inch).

8.4 iMEC12
8.4.1 Main Unit
For the exploded view and parts list of the main unit, refer to Section 8.3.1 Main unit.

8.4.2 Front Panel Assembly (12 Inch/Touchscreen/Wi-Fi+ 1G SD Card)


8.4.2.1 Exploded View

8-31

8.4.2.2 Parts List


SN 1 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. PN / / / 051-000829-00 / / / / / / / / / / / / / 12. 13. 14. Front housing (12 inch) 043-001969-01 Front housing (12 inch) 15. 16. 17. 18. 19. 20. 21. 22. / / / / 051-000879-01 / / / Knob Antenna and cable assembly Alarm lamp shade Alarm lamp gasket Alarm lamp board PCBA FUJI 12" Touchscreen long position pad Power button 6301 On/Off key board PCBA 801-6301-00030-00 043-001790-00 115-010845-00 / / 801-6301-00005-00 / 6802-20-66691-51 051-000880-00 801-6301-00010-00 / / Description Screw, pan head Phillips, M36 Touchscreen control board PCBA Screw, Pan head with washer, Phillips M38 Main board PCBA Screw, self-tapping, PT38 Fujitsu touchscreen kit (12 Inch) Touchscreen, resistive-type, 12.1" 4-line 12 inch touchscreen long gasket 12 inch touchscreen short gasket 6301 keypad PCBA 12.1 inch Key (12 inch, Chinese) Key (12 inch, English) Key (12 inch, Italian) Key (12 inch, Spanish) Key (12 inch, Portuguese) Key (12 inch, French) Key (12 inch, German) FUJI 12 Touchscreen long position pad Encoder board FRU part number / 051-000881-00 / 801-6302-00001-00 / / 021-000059-00 / / 051-000878-00 049-000335-00 049-000351-00 049-000385-00 049-000386-00 049-000387-00 049-000388-00 049-000389-00 / 0010-30-43089 / / For monitors with machine screws For monitors with tapping screws / / / / / / / / Remarks / / / / / / / / / / As configured

8-32

8.4.3 Fujitsu Touchscreen Kit (12 Inch)


8.4.3.1 Exploded View

8.4.3.2 Parts List


SN 1 2. 3. PN / / 021-000083-00 0010-10-42633 021-000061-00 4. 5. 6. / / / Description Screen bracket (12 inch touchscreen) 12 inch AU display buffer cushion (touchscreen) LCD, TFT, 12.1", 800600, 3.3V, LED-BackLight LCD, TFT, 12.1", 800600, 3.3 V, LED-BackLight LCD, TFT, 12.1", 800*600, LED-BackLight Short pressing strip 4 for 12 inch display bracket Screw, pan head Phillips, M38 Short pressing strip 3 for 12 inch display bracket 8-33 FRU part number / / 801-6301-00025-00 (Tianma) 801-6301-00024-00 (AU) 801-6301-00008-00 (AU) / / / Remarks / / Main Backu p Backu p / / / As configure d/Not mutually compatib le nor exchange able

8.4.4 Front Housing Assembly (12 Inch)


8.4.4.1 Exploded View

8.4.4.2 Parts List


SN 1 2. 3. 4. 5. 6. 7. 8. PN / 051-000964-00 / / / / / / / / / Description Screw, Pan head with washer, Phillips M38 6301 Main board PCBA (Standard configuration) Tianma LCD display kit (12 inch) Screw, self-tapping, PT38 12 inch nakescreen long gasket 12 inch nakescreen short gasket 6301 keypad PCBA 12.1 inch Key (12 inch, Chinese) Key (12 inch, English) Key (12 inch, Italian) Key (12 inch, Spanish) FRU part number / 801-6301-00004-00 / / / / 051-000878-00 049-000335-00 049-000351-00 049-000385-00 049-000386-00 Remarks / / / / / / / As configured

8-34

/ / 9. 10. 11. 043-001969-01 / /

Key (12 inch, Portuguese) Key (12 inch, French) Key (12 inch, German) Encoder board Knob Front housing (12 inch) Front housing (12 inch)

049-000387-00 049-000388-00 049-000389-00 0010-30-43089 043-001790-00 801-6301-00009-00 801-6301-00029-00 / / 801-6301-00005-00 6802-20-66691-51 051-000880-00 / / For monitors with machine screws For monitors with tapping screws / / / / /

12. 13. 14. 15. 16.

/ / 051-000879-01 / /

Alarm lamp shade Alarm lamp gasket Alarm lamp board PCBA Power button 6301 On/Off key board PCBA

8.4.5 Tianma LCD Display Kit (12 Inch)


8.4.5.1 Exploded View

8-35

8.4.5.2 Parts List


SN 1 2. 3. PN / / Description Display bracket (12 inch nakescreen) Nakescreen supporter LCD, TFT, 12.1", 800600, 3.3V, LED-BackLight LCD, TFT, 12.1", 800600, 3.3 V, LED-BackLight LCD, TFT, 12.1", 800*600, LED-BackLight Screw, pan head Phillips, M38 FRU part number / / 021-000083-00 0010-10-42633 021-000061-00 / Remarks / / Main Backup Backup / As configured/Not mutually compatible nor exchangeable

4.

8.4.6 Rear Housing Assembly


For the exploded view and parts list of the rear housing assembly, refer to Section 8.4.6 Rear Housing Assembly.

8.4.7 Rear Housing Subassembly


For the exploded view and parts list of the rear housing subassembly, refer to Section 8.3.7 Rear Housing Assembly.

8.4.8 Battery Compartment Assembly


For the exploded view and parts list of the battery compartment assembly, refer to Section 8.3.8 Battery Compartment Assembly.

8.4.9 Parameter Module Assembly


For the exploded view and parts list of the parameter module assembly, refer to Section 8.2.7 Parameter module assembly.

8.4.10 Parameter Connector Assembly


For the exploded view and parts list of the parameter connector assembly, refer to Section 8.2.8 Parameter connector panel assembly.

8-36

8.4.11 Mindray Sidestream CO2 Module Kit


For the exploded view and parts list of Mindray sidestream CO2 module assembly, refer to Section 8.2.9 Mindray Sidestream CO2 Module Kit.

8.4.12 Oridion Microstream CO2 Module Kit


For the exploded view and parts list of Oridion microstream CO2 module assembly, refer to Section 8.2.10 Oridion Microstream CO2 Module Kit.

8.4.13 Capnostat Mainstream CO2 Module Kit


For the exploded view and parts list of Capnostat mainstream CO2 module assembly, refer to Section 8.2.11 Capnostat Mainstream CO2 module kit.

8.4.14 TR6F Recorder Kit


For the exploded view and parts list of the TR6F recorder kit, refer to Section 8.2.12 TR6F Recorder Kit.

8.4.15 Battery Interface Board Assembly


For the exploded view and parts list of the battery interface board assembly, refer to Section 8.2.13 Battery charge board assembly (12 inch).

8.5 Other Maintainable Parts


No. 1 2. 3. 4. 5. 6. 7. 8. PN / / / / / / / / Description Mainstream CO2 signal cable M02C sidestream communication cable Cable between interface board and C.O. module Cable between the interface board and mainstream converter board Cable between the interface board and microstream converter board Recorder cable Cable for AC/DC module Cable for touchscreen control board FRU part number 009-001994-00 009-001995-00 009-001996-00 009-001997-00 009-001998-00 009-002000-00 009-002108-00 009-002109-00 Remarks / / / / / / / /

8-37

9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28.

/ / / / / / / / / / / / / / / / / / / / /

8.4" AU display cable Display signal cable (Truly, 8") Cable for display signal (AU 10") Cable for display signal (AU 12") 8" AU display backlight cable 10.4" AU display backlight cable Cable for backlight of 12.1" display (AU) Alarm lamp cable Cable between main board and keypad board Cable between keypad and power board Cable between parameter board and interface board Cable between main board and interface board Cable for ON/OFF key board (12") Cable for battery backboard Cable between inverter and main board (10") Display signal cable (Tianma, 10") Encoder cable TFT display backboard cable Silicon, white Main board PCBA (full configuration, 12 inch) Cyberlink module PCBA

009-002110-00 009-002468-00 009-002111-00 009-002112-00 009-002113-00 009-002114-00 009-002115-00 009-002116-00 009-002117-00 009-002118-00 009-002119-00 009-002126-00 009-002127-00 009-002132-00 009-002469-00 009-002471-00 9200-21-10460 8000-21-10239 M6G-020015-801-6301-00003-00 051-000811-00

As configure d / / / / / / / / / / / / / / / / / / /

8-38

A Electrical Safety Inspection


The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program. They are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or the operator. Additional tests may be required according to local regulations. All tests can be performed using commercially available safety analyzer test equipment. These procedures assume the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer. Other popular testers complying with IEC 60601-1 used in Europe such as Fluke, Metron, or Gerb may require modifications to the procedure. Follow the instructions of the analyzer manufacturer. The consistent use of a safety analyzer as a routine step in closing a repair or upgrade is emphasized as a mandatory step if an approved agency status is to be maintained. The safety analyzer also proves to be an excellent troubleshooting tool to detect abnormalities of line voltage and grounding, as well as total current loads.

A.1 Power Cord Plug


A.1.1 The Power Plug
Test Item The power plug pins The plug body The power plug The strain relief The power plug Acceptance Criteria No broken or bent pin. No discolored pins. No physical damage to the plug body. No physical damage to the strain relief. No plug warmth for device in use. No loose connections. No physical damage to the cord. No deterioration to the cord. For devices with detachable power cords, inspect the connection at the device. For devices with non-detachable power cords, inspect the strain relief at the device.

The power cord

A-1

A.2 Device Enclosure and Accessories


A.2.1 Visual Inspection
Test Item Acceptance Criteria No physical damage to the enclosure and accessories. No physical damage to meters, switches, connectors, etc. No residue of fluid spillage (e.g., water, coffee, chemicals, etc.). No loose or missing parts (e.g., knobs, dials, terminals, etc.).

The enclosure and accessories

A.2.2 Contextual Inspection


Test Item The enclosure and accessories Acceptance Criteria No unusual noises (e.g., a rattle inside the case). No unusual smells (e.g., burning or smoky smells, particularly from ventilation holes). No taped notes that may suggest device deficiencies or operator concerns.

A.3 Device Labeling


Check the labels provided by the manufacturer or the healthcare facility are present and legible. Main unit label Integrated warning labels

A-2

A.4 Earth Leakage Test


Run an Earth Leakage test on the device being tested before performing any other leakage tests. Leakage current is measured the following ways: Earth Leakage Current, leakage current measured through DUT outlet Earth Earth Leakage Current AP-EARTH (ALL Applied Parts connected to Earth), leakage current measured through DUT outlet Earth

There is no need to attach a test lead; the 601PRO automatically connects the measuring device internally.

To Perform the Test


1. 2. 3. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front panel outlet, and turn on the device. Attach the device's applied parts to the 601PRO applied part terminals if applicable. Press shortcut key 4.The Earth Leakage test appears on the display, and the test begins immediately:

4.

SOFT KEY 1 toggles the DUT outlet Polarity from Normal to Off to Reverse. SOFT KEY 2 toggles the DUT outlet from Earth to No Earth. SOFT KEY 3 toggles the DUT outlet from L2 to No L2. SOFT KEY 4 toggles the AP to Earth to No AP to Earth. Press the print data key at any time to generate a printout of the latest measurement.

A-3

In Case of Failure
Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead. Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation. If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.

LIMITS
For USA,

300 A in Normal Condition 1000 A in Single Fault Condition

For other countries:


500 A in Normal Condition 1000 A in Single Fault Condition

A.5 Patient Leakage Current


Patient leakage currents are measured between a selected applied part and mains earth. All measurements may have either a true RMS or a DC-only response.

Preparation
Perform a calibration from the Mains on Applied Part menu. The following outlet conditions apply when performing this test: Normal Polarity, Earth Open, Outlet ON Normal Polarity, L2 Open, Outlet ON Normal Polarity, Outlet ON Reversed Polarity, Outlet ON

Reversed Polarity, Earth Open, Outlet ON Reversed Polarity, L2 Open, Outlet ON

A-4

WARNING

If all of the applied parts correspond to the instrument type, the applied parts will be tied together and one reading will be taken. If any of the applied parts differ from the instrument type, all applied parts will be tested individually, based on the type of applied part. This applies to Auto and Step modes only.

To Perform the Test


1. 2. 3. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front panel outlet, and turn on the device. Attach the applied parts to the 601PRO's applied part terminals. Press shortcut key 6. The Patient Leakage test is displayed, and the test begins immediately.

4. 5. 6. Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part leakage current. Modify the configuration of the front panel outlet by pressing the appropriate SOFT KEY on the 601PRO. Press the print data key at any time to generate a printout of the latest measurement.

NOTE

If the current test standard being used does not include Patient Leakage DC readings, or the DC option is not enabled, then DC readings will not be available through the APPLIED PART SOFT KEY selections. Refer to Chapter 8, Standards and Principles.

A-5

In Case of Failure
Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead. Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation. If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.

LIMITS
For CF applied parts

10A in Normal Condition 50A in Single Fault Condition

For BF applied parts


10A DC,100A AC in Normal Condition 50A DC, 500A AC in Single Fault Condition

A.6 Mains on Applied Part Leakage


The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage, through a limiting resistance, to selected applied part terminals. Current measurements are then taken between the selected applied part and earth. Measurements are taken with the test voltage (110% of mains) to applied parts in the normal and reverse polarity conditions as indicated on the display.

The following outlet conditions apply when performing the Mains on Applied Part test. Normal Polarity; Reversed Polarity

A-6

Preparation
To perform a calibration from the Mains on Applied Part test, press CAL (SOFT KEY 2). 1. 2. Disconnect ALL patient leads, test leads, and DUT outlet connections. Press CAL to begin calibration, as shown:

If the calibration fails, the previously stored readings will be used until a passing calibration has occurred. Also, the esc/stop key has no effect during calibration. 3. When the calibration is finished, the Mains on Applied Part test will reappear.

WARNING

A 2-beep-per-second signal indicates high voltage present at the applied part terminals while a calibration is being performed. High voltage is present at applied part terminals while measurements are being taken.

To Perform the Test


1. 2. 3. 4. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601 Attach the applied parts to the 601PRO applied part terminals. Attach the red terminal lead to a conductive part on the DUT enclosure. Press shortcut key 7. The Mains on Applied Part test is displayed.

A-7

5. 6. 7.

Select the desired outlet configuration and applied part to test using the appropriate SOFT KEYS: Press START TEST (SOFT KEY 1) to begin the test. Press the print data key to generate a printout of the latest measurement.

NOTE

If all of the applied parts correspond to the instrument type, the applied parts will be tied together and one reading will be taken. If any of the applied parts differ from the instrument type, all applied parts will be tested individually, based on the type of applied part. This applies to Auto and Step modes only.

In Case of Failure
Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead. Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation. If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.

LIMITS
For CF applied parts: 50A For BF applied parts: 5000A

A-8

A.7 Patient Auxiliary Current


Patient Auxiliary currents are measured between any selected ECG jack and the remaining selected ECG jacks. All measurements may have either a true RMS or a DC-only response.

Preparation
1. 2. 3. 4. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front panel outlet, and turn on the device. Attach the patient leads to the 601PRO ECG jacks. Define the Lead Types from the View Settings Option (refer to: Lead Type Definitions in Section 5 of this chapter). Press shortcut key 8. The Patient Auxiliary Current test is displayed, and the test begins immediately. Display values are continuously updated until another test is selected.

5. 6. 7. 8.

Press SOFT KEYS 1-4 to select leakage tests Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part leakage current: Modify the configuration of the front panel outlet by pressing the appropriate SOFT KEY on the 601PRO: Press the print data key at any time to generate a printout of the latest measurement.

NOTE

If the current test standard being used does not include Patient Auxiliary Current DC readings, or the DC option is not enabled, then DC readings will not be available through the APPLIED PART SOFT KEY selections.

A-9

In Case of Failure
Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead. Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation. If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.

LIMITS
For CF applied parts,

10A in Normal Condition 50A in Single Fault Condition

For BF applied parts,


10A DC, 100A AC in Normal Condition 50A DC, 500A AC in Single Fault Condition

A.8 Functional test


For functional test items, please refer to relevant functional tests in Chapter 3 Testing and Maintenance.

A-10

ELECTRICAL SAFETY INSPECTION FORM (American Version)


Overall assessment: Scheduled inspection Unopened repair type Opened repair type, not modify the power part including transformer or patient circuit board Opened repair type, modify the power part including transformer or patient circuit board Location: Equipment: Manufacturer: Measurement equipment /SN: INSPECTION AND TESTING 1 Power Cord Plug 2 Device Enclosure and Accessories 3 Device Labeling Normal condition(NC) Earth 4 Single Fault Leakage condition(SFC) Normal condition(NC) Patient 5* Leakage Single Fault Current condition(SFC) 6* Mains on Applied Part Leakage Normal condition(NC) 7* Patient Auxiliary Current Single Fault condition(SFC) Model: Test item: 1, 2, 3, 8 Test item: 1, 2, 3, 8 Test item: 1, 2, 3, 4,8 Test item: 1, 2, 3, 4, 5, 6, 7, 8

Technician: Control Number: SN: Date of Calibration: Pass/Fail Comments

____A ____A ____A ____A

____A ____A ____A

Max: NC: 300A SFC: 1000A Max: CF applied part: NC:10A, SFC: 50A BF applied part: NC:100A, SFC: 500A Max: CF applied part: 50A BF applied part: 5000A Max: CF applied part: NC:10A, SFC: 50A BF applied part: NC:100A, SFC: 500A

Functional test (parameters tested): Note: The test items marked * are needed only for incoming inspections and after repairs or modifications that may have affected lead leakage [NFPA 99 (2005)8.5.2.1.3]. Deficiency /Note: Name/ Signature: __________________________ Date:_____________________________

A-11

ELECTRICAL SAFETY INSPECTION FORM (International version)


Overall assessment: Scheduled inspection Unopened repair type Opened repair type, not modify the power part including transformer or patient circuit board Opened repair type, modify the power part including transformer or patient circuit board Location: Equipment: Manufacturer: Measurement equipment /SN: INSPECTION AND TESTING 1 Power Cord Plug 2 Device Enclosure and Accessories 3 Device Labeling Normal condition(NC) Earth 4 Single Fault Leakage condition(SFC) Normal condition(NC) Patient 5* Leakage Single Fault Current condition(SFC) 6* Mains on Applied Part Leakage Normal condition(NC) 7* Patient Auxiliary Current Single Fault condition(SFC) Model: Test item: 1, 2, 3, 8 Test item: 1, 2, 3, 8 Test item: 1, 2, 3, 4,8 Test item: 1, 2, 3, 4, 5, 6, 7, 8

Technician: Control Number: SN: Date of Calibration: Pass/Fail Comments

____A ____A ____A ____A

____A ____A ____A

Max NC:500A SFC:1000A Max CF applied part NC:10A, SFC: 50A BF applied part NC:100A, SFC: 500A Max: CF applied part: 50A BF applied part: 5000A Max: CF applied part: NC:10A, SFC: 50A BF applied part: NC:100A, SFC: 500A

Functional test (parameters tested): Note: The test items marked * are needed only for incoming inspections and after repairs or modifications that may have affected lead leakage [NFPA 99 (2005)8.5.2.1.3]. Deficiency /Note: Name /Signature: __________________________ Date:__________________________

A-12

PN: 046-002371-00 (3.0)

You might also like