M4 M6 Q Series Manual
M4 M6 Q Series Manual
M4 M6 Q Series Manual
OPERATING MANUAL
12/14/13
Q Series with Temperature display and Tank pressure control
Bottle detection system, Evac. Purge
TABLE OF CONTENTS
GENERAL ...................................................................................................3
CONTROL PANEL .....................................................................................3
SAFE OPERATION OF THE BOTTLING MACHINE .............................5
PROCEDURE FOR CLEANING AND SANITIZING ...............................5
OPERATING SUMMARY .........................................................................7
SANITIZE ........................................................................................8
BOTTLING ......................................................................................8
AUTO CLEANING .......................................................................13
CLEANING ...................................................................................15
MANUAL OVERRIDE .............................................................................16
ADJUSTMENTS TO THE MACHINE ....................................................18
BOTTLE INDEXING & FILL HEAD ALIGNMENT ..................18
CAP FEED .....................................................................................18
COMPUTER .............................................................................................19
TROUBLESHOOTING .............................................................................20
PRESSURE SENSOR CHECK .................................................................22
LUBRICATION & MAINTENANCE.......................................................22
FILL MANIFOLD & TUBING..................................................................23
BOTTLE DETECTION SYSTEM 25
BOTTLE INDEXING SPEED ADJUST....26
WARNING:
AS WITH ALL MECHANICAL EQUIPMENT, CARE MUST BE TAKEN
TO AVOID INJURY FROM MOVING PARTS WHICH OPERATE WITH
CONSIDERABLE FORCE AND WITHOUT WARNING. MEHEEN
MANUFACTURING SHALL NOT BE HELD LIABLE FOR INJURY OR
DAMAGE RESULTING FROM THE INAPPROPRIATE USE OF THIS
MACHINE OR FROM NORMAL OPERATION WITHOUT REGARD TO
NORMAL SAFETY CONSIDERATIONS. MEHEEN MANUFACTURING
SHALL BE HELD HARMLESS IN THE CASE OF INJURY RELATED TO
THE USE OF THIS MACHINE REGARDLESS OF THE
CIRCUMSTANCES AND OPERATING PROCEDURES USED.
GENERAL
The Merlin automated bottling machine is designed with quality and portability in mind. An
onboard computer controls all automated functions including bottle fill levels and pressures.
Standard automated features are sanitizing, cleaning, filling and crowning.
The Merlin is specifically designed for bottling carbonated beverages under cold counter pressure
conditions directly from the bottling tank. Generally optimum bottling conditions are cold (30340F) and 2.0-3.0 volumes dissolved CO2.
A compressed air source that can provide 100psi + is required to operate Merlin as well as 110
volt AC electrical power and a regulated CO2 source. Meheen Manufacturing suggests the use of
an oil sump, reciprocating type air compressor with a min of 15scfm@90psi rating, DO NOT use
a home style direct drive oil-less compressor. Regulated air pressure on the machine itself needs
to be 95-100psi. Minimum air hose size is 3/8 inside diameter for a 50 hose or less and larger
for longer runs. Also all devices restricting air flow, such as regulators and moisture traps should
be removed between the air compressor and the bottling machine.
CONTROL PANEL
The control panel of the Merlin bottling machine consists of the operator interface shown above
and a large red mushroom switch which is located near the bottom center of the panel. The
opening screen gives information about the Merlin machine, such as Brewery Name, Serial#,
interface and computer programs versions. This screen appears each time the machine is turned
on. To access the main operating screen, simply touch the wizard and the screen will change to
the main operating screen.
3
ON/OFF: Main power to the machine is turned on by pulling the red mushroom button out,
and turned off by pushing it in.
MAIN SCREEN: The main operating screen shown below allows you to choose the function
needed. From any of the operating screens the Reset key can be pressed and the Main button
with the wizard pictured on it and you will be returned to the main operating screen.
Auto Filling, Clean, Sanitize & Manual Override: These operations are entered from
the Main screen show above by pressing the desired button. When any of these buttons are
pressed the screen will change to that operation. Note: The reset key must be pressed before
starting any operational mode except the Manual override functions.
RESET: The RESET button is used any time the machine has been stopped, paused, or about
to be started in a new mode. The RESET always returns the machine to its starting conditions for
each mode. RESET must be used each time the machine enters a new operation before the
machine will start any program functions and before pressing the Main screen button.
CAUTION: If the bottling machine is paused and rapidly RESET during a time when the
machine has pressure in the bottle (i.e., filling), pressure in the bottles will be released
rapidly without warning. Rapid release of pressure can be hazardous due to glass
breakage and liquid spray.
4
7.
4. Prior to bottling, sanitize using an appropriate no-rinse sanitizer such as Chlorine Dioxide
(ClO2) or Peracetic Acid (PAA). Operate the machine in the Sanitize mode.
To remove beer stone (Calcium Oxalate) deposits, the use of an acid product such as Acid Brite
No.2 (Phosphoric/Nitric blend) is recommended. Run 1 ounce (30ml) of acid per gallon of warm
(1200F) water for a minimum of 2-4 minutes prior to CELL-R-MASTER.
OPERATIONAL SUMMARY
SANITIZING:
1.
Mix a sufficient amount of sanitizing solution (about 5 gallons) to keep the pump
submersed during sanitation. Most sanitizing solutions which require less than 2 minutes
of contact time to be effective can be used.
2.
Connect air pressure to the bottling machine before loading caps or connecting the
sanitizing pump to the machine.
3.
4.
Connect the hose to be used for bottling between the sanitizing pump and the beer
distribution manifold on the bottling machine. Be sure to clamp both ends of the hose
securely.
5.
Disconnect the 5/32" liquid line pressure sensor hose (small tubing) from the pressure
sensor located in the bottom of the control panel. To disconnect the hose, depress the
collar of the tube fitting and pull the hose out. Place the end of the hose you just
disconnected below the machine into a bucket or drain. Sanitizing solution will flow
through this line when the pump is started.
6.
Place three rows of four bottles on the bottle runway at the base of the bottle feed ramp.
When the bottling machine is started it will automatically index the bottles to the filling
and crowning stations.
7.
Press the RESET button on the interface and the System is Reset message will appear
at the bottom of the screen.
8.
Place the sanitizing pump into the bucket of sanitizing solution and plug it in.
9.
Press the Sanitize Off/On button to begin the sanitizing cycle. The button will change
color from red to green and the green RUN light will illuminate at the top of the screen.
A message in the bottom message bar will have the word Sanitizing displayed. The
machine will automatically perform two 1 minute sanitizing cycles then stop Illuminating
the PAUSE light and display in the message box Sanitizing Complete.
CAUTION: The bottle filling head will come up without warning to index the next row
of bottles. Keep hands and face away from moving parts at all times.
10.
Bottle caps in the cap feed ramps will automatically feed into the capping head each time
the capping head cycles to the up position. The cap feed ramps hold enough caps for 2-3
cases of bottles. More caps can be added to the cap feed ramps at any time.
11.
The machine automatically stops at the end of the second sanitizing cycle with the filling
heads down. The amber pause light in the center of the interface will be lit when the
cycle is complete and the message display will read Sanitizing Complete. Press the
Sanitize button to turn the program off. The button will turn Red and the message display
will show System is Reset. Do not remove bottles filled with sanitizing solution from
the bottling machine as they will be needed for weight to provide proper bottle feeding
when starting the automatic bottling cycle. When the Sanitize cycle is complete, unplug
the pump to stop solution still being pumped through the liquid pressure sensor hose.
You may now press the Main button with the wizards image to return to the Main
Operation menu.
12.
Allow the liquid pressure sensor hose from the beer distribution to drain before
connecting to the control panel. After the pressure sensor hose is drained reinstall it on
the bottling machine by pushing the hose firmly into the fittings on the sensor and beer
manifold. Note: The tubing is often difficult to push into the fittings. You should feel
the tube slide into the fitting and seat when it is installed properly.
BOTTLING:
1.
Connect the hose from the distribution manifold of the machine used for the sanitizing
cycle to the product dispensing tank. Be sure to clamp both ends of the hose securely.
Open the dispensing tank valve very slowly when first charging the line. If the tank valve
is opened too fast, severe damage may occur to the beer line pressure sensor.
2.
Connect the CO2 supply to the CO2 inlet on the top of the bottling machine and adjust the
pressure to 25psi. This CO2 source is used to purge the bottles, bring them to counter
pressure and pulse the full bottles. DO NOT EXCEED 40 psi to avoid damage to
pressure sensors on the machine.
3.
Adjust the CO2 pressure applied to the head space of the dispensing tank 1-2 psi above
the tank equilibrium pressure or an adequate level to dispense the product into the bottles.
4.
Place 4-5 rows of bottles on the sloped bottle feed ramp to begin the bottling process.
The full bottles from the sanitizing cycle should remain on the machine until enough full
bottles have accumulated on the machine to keep the bottle feeding system operating
properly. As full bottles accumulate on the exit portion of the machine more bottles can
be added to the bottle feed ramp. The full bottles provide the needed friction for the
bottle gravity feed system to operate properly, and keep the bottles being indexed from
sliding too far and misfeeding. Once the machine is operating in the bottling mode, at
least 6 rows of full bottles must be maintained on the exit portion of the machine.
Auto Filling:
5. The Auto Filling screen is shown above and consists of three slide bars, two pressure meters,
manifold temperature display and Auto Off/On, Rest, Pulse, Evac/Purge, Liquid psi Control
buttons and Liquid psi arrow key adjustment with pressure display. The first time the
9
machine is turned on, the Fill Sensor will be in the Auto mode and the slide for the Manual
mode will always default to 10. This is quite sensitive for fill detection at low filling
speeds. When the Fill Sensor is in the Auto mode, it will automatically adjust to changing
pouring conditions and is not adjustable by the operator. Should the Auto Fill Detection fail
to work properly, it may be changed to Manual adjustment by pressing the selector button at
the top of the Fill Sensor. The color of the button will change from green to blue and the
text will change to Manual indicating the mode change for Fill Detection. As filling speeds
are increased in the Manual mode, the Fill Sensor will need to be raised to a higher number
to ensure the bottles are completely filled. The Low Pressure slide is used to adjust how
much pressure is in the bottles when the filling heads are raised, or the pressure at which the
Pulse will occur. It is generally recommended to operate the Low Pressure as low as
possible and still maintain a 1 second or less time from when the filling valve closes until
the filling heads are raised from the bottles. If the filling heads remain in the bottle more
than 1 second after the filling valve has closed, you need to raise the Low Pressure until the
1 second has been met. The two pressure bars simply display incoming product pressure to
the machine and the pressure in the bottle. It can be used for adjustments to the machine or
trouble shooting any problems such as counter pressure fail or fill detection issues.
6. The Merlin control panel also displays temperature in the product manifold located at the
top of the machine. This temperature display simply lets the operator observe the incoming
temperature of the product and can be used to help with determining proper pressures in the
liquid tank.
7. A button labeled Liquid psi Control is located in the upper right hand portion of the
display. This button turns on and off the controls to adjust CO2 dispensing pressure in the
head space of the tank based on liquid manifold pressure at the machine. This function is
optional and intended to help resolve dispensing pressure drops that can occur using manual
style regulators. To utilize this feature, an independent CO2 source, different from the one
used for the bottling machine, is connected to the
control valve inlet (see photo). The inlet pressure
should be set at 30-35psi from the CO2 source. A
hose will need to be connected between the
control valve and head space of the dispensing
tank.
How Liquid psi Control (LpC) works, is when
connected to a CO2 source and the head space of
the dispensing tank and the system is activated by
pressing the Liquid psi Control button. The
computer initially records the static liquid pressure
at the manifold and displays that pressure as the
set point above the adjusting arrow keys. Using
the arrow keys you can adjust this set point up or
down as needed. Each cycle, before the product
valve opens the computer, will sample the
manifold pressure and turn the CO2 control valve
on or off as needed.
10
NOTE: This system is limited to a maximum control pressure of 14.9 psi and only operates
when the filling process is running. It does not function when the machine is paused.
8. Press the RESET button on the operator interface and the message display should read
System is Reset. Press Auto Filling to start the machine in its automatic filling and
capping mode. The message display will display Automatic Filling, the RUN light will be
illuminated and the Auto Filling button will change from red to green. To adjust the Low
Pressure and Fill Sensor simply move the slide bars to the desired amount. Adjusting the
Fill Sensor to a higher value makes it less sensitive and a lower value makes it more
sensitive.
Pre-Purge is the default operation of the machine, where the atmosphere in the bottles is
purged with CO2 before counter pressure and filling. Additional purge time can be
programmed using the arrow keys at the bottom of the screen. Press this button again and
the operation changes to Evac/Purge, drawing the atmosphere from the bottles then
purging before counter pressure and filling. Press this button again and the machine
performs a double evacuation of the bottles, then counter pressures for filling.
11
6.
The bottle fill rate is adjusted by using the off-gas flow regulator valve located just below
and to the left of the control panel. Closing this valve slows the flow of liquid into the
bottle, while opening the regulator increases the rate of fill. The fill rate should be slow
enough that the liquid does not foam while the bottles are filling.
7.
Once the machine is operating, you can adjust the fill sensitivity as needed to get
consistent fills, and the Low Pressure as needed to control foaming along with turning the
PULSE on or off.
8.
To pause the machine at any time during automatic filling, press the Auto Filling button
and the machine will stop in the safe shut down configuration, releasing pressure from the
bottles with the filling head down. When the Reset button is pressed the machine will
reset and the filling heads will go up. Once the machine has been reset you can start
another Auto Filling operation by pressing that button or press the Main button with the
wizards image to go to the main menu.
12
AUTO CLEANING:
1.
Make sure the air and power are still connected to the machine and enough filled rows of
bottles remain on the machine for proper feeding. Also leave at least 3 rows of empty
bottles on the bottle feed ramp.
2.
Connect the product hose to the sanitizing pump ensuring the other end is connected to
the product manifold on the bottling machine.
3.
Immerse the sanitizing pump in an adequate volume of clean water or cleaning solution.
Most cleaning solutions are acceptable with the exception of acid based materials that can
damage the internal parts of valves used in the off-gas system. Remember you will be
filling 2 rows of bottles and washing through the off-gas system.
4.
Disconnect the 5/32" liquid pressure sensor hose (small tubing) from the pressure sensor
located in the bottom of the control panel. To disconnect the hose, depress the collar of
the tube fitting and pull the hose out. Place the end of the hose you just disconnected
below the machine into a bucket or drain. Cleaning solution will flow through this line
when the pump is started in the next step.
13
5.
Disconnect the CO2 purge and pulse hoses from the top of the machine and place the ends
in a bucket or to a drain. Cleaning solution or water will flow through these hoses when
the sanitizing pump is plugged in.
6.
7.
Be certain you have at least 3 rows of empty bottles on the machine and press the Clean
button. The machine will begin the program and Cleaning will appear in the message
display and the Clean button will change from red to green. The RUN light will also be
illuminated.
8.
During the cleaning cycle it is acceptable to hose off the machine and bottling area. Care
should be taken to keep water from splashing on the control panel and electrical devices.
9.
The cleaning cycle will fill 2 rows of bottles with solution flowing through the CO2 purge
and pulse hoses removed in step 5 as well as from the valves under the machine. At the
end of the second cycle the machine will stop with the filling and capping heads down
and the operator interface will display the following instruction Reconnect hoses
and/Reset to continue. Reconnect the CO2 purge and pulse hoses and press the Blow
Out button. The machine will index one more row of bottles, the filling and capping
heads will come down and CO2 will flow from the system for 15 seconds to complete the
cleaning. When cleaning is complete the fill head and capping head will remain down
and Cleaning Complete will be displayed in the message display. The filling heads can
be raised by pressing the Clean button to turn off the cycle and pressing the Reset button.
10.
When the cleaning cycle is complete, unplug the pump, as cleaning liquid will continue to
flow through the pressure sensor hose until the pump is unplugged. Leave the beer line
pressure sensor hose disconnected from the pressure sensor and hang vertically to
promote draining.
11.
During the cleaning cycle the off-gas regulating system is flushed and blown out with
CO2 and should be clean. No other cleaning of this system should be required.
12.
When the cleaning cycle is complete, be sure to remove all bottle caps from the cap
feeding ramps before disconnecting the air supply. If caps are not removed from the
machine when the air source is disconnected, they will all fall into the capping head and
jam the cap feeding mechanism. Also pull rubber bottle seals down or remove them and
clean behind them so that no moisture remains behind the seals.
13.
After the cleaning cycle is complete, it is recommended that excess moisture and debris
be blown off the machine using compressed air. The bottler is now ready to be stored for
the next use. Never use high pressure air, gas or city water to blow out the fill
manifold while the 5/32 pressure sensor hose is connected to the bottom of the
control panel. Damage to the pressure sensor due to over pressurization is not
warranted by Meheen Manufacturing.
14
CLEANING:
Auto cleaning does a good job of rinsing the internal parts of the bottling machine and leaves it
relatively clean; however, we recommend extra cleaning be performed on a regular basis to
ensure good quality of your beverages and to avoid contamination. For this cleaning the filling
head, distribution manifold and fill tubes should be removed from the machine and washed in a
warm caustic or PBW solution which will remove protein and other contaminates from internal
surfaces. The frequency of this type of cleaning will depend on the beverages and amount of use,
as well as the cleaning requirements of your facility.
1. With the machine turned off, and all product lines, air, CO2 and electricity disconnected,
disconnect the pressure sensor hose and the clamps and hoses connected to the bottom of the
manifold. Remove the liquid manifold by removing the two bolts that secure it to the frame.
2. Loosen or remove the screws for the tubing support clamps on the front of the filling head
carriage. Then, disconnect the tubing from the front of the filling head by pressing in on the
fitting collar while pulling the tubes out. Remove the filling head from the carriage by
removing the two bolts that secure it to the carriage.
3. Remove the silicone filling tubes from the stainless steel filling tubes and discard them.
Remove the filling tubes from the filling head carefully to avoid bending them. Remove
rubber bottle seals.
4. Place the manifold, filling head and filling tubes in a container with the cleaning solution.
Cleaning solution may be circulated with the sanitizing pump provided the operating
temperatures and cleaning solution used is compatible with the pump specifications. If the
sanitizing pump is used to circulate cleaners, be cautious and take all safety precautions to
avoid being splashed by hot cleaners.
5. After cleaning is complete, rinse all parts thoroughly with water to remove all cleaner and
reassemble the machine. Always flush the inner surface of the crowning heads thoroughly
with water to remove any residual beverage deposits which may cause contamination.
15
Manual Override:
A Manual Override system is incorporated into the control system of the Merlin bottling
machine, which allows you to operate any function on the machine without operating an
automated cycle. Manual overrides are very useful for adjusting the machine for new bottles,
trouble shooting components and verifying the operation of sensors. The overrides are accessed
from the Main Menu by pressing the Manual Override button.
From the Manual Override screen shown above you can actuate any of the items on the machine
by simply pressing the labeled buttons. The activated button will turn green while pressed
indicating it is activated. The corresponding item on the machine should respond while
activated. The Bottle Detection Controls button will load that page on the screen, see page 23.
At the right side of the screen is the Bottle and Liquid pressure meters which will indicate the
pressure applied to each of the sensors in the control panel. These displays are very useful to
verify proper operation of each of the sensors which are critical for proper operation of the
bottling machine.
At the bottom of the screen the Rows of 10,000 and Rows of bottles are displayed. By
multiplying the number displayed in the left counter by 10,000 and adding it to the counter on the
right you can determine the total rows of bottles filled.
16
Place bottles on the machine so that empty bottles with be indexed under the filling head when
the machine is started. Have compressed air and CO2 connected to the machine as you normally
would and press the Start Leak Test button to begin the leak test. Bottles will be indexed under
the filling heads and pressurized with CO2 and you will see the pressure rise at the right of the
screen. A leak test is performed by bringing the bottles up to a certain pressure and monitoring if
pressure is lost or gained over time. If pressure is lost or gained at such a rate that a leak is
detected the leak test will be terminated, the red FAIL light will flash and a text message will be
displayed at the bottom of the screen to identify the problem. This test can determine if a CO2
valve or other valve is leaking into the system causing the pressure to increase as well as pressure
leaking out of the system. Pressure increasing can be CO2 leaking past one of the valves or
possibly compressed air entering the system through a failed off-gas valve. Pressure leaking out
of the system can be any mechanical connection, seal or valve and is the most common type of
leak.
[To return the machine to normal operation, press the Start Leak Test button to turn it off and
press reset, then the Main button to return to the main menu. Once a leak test is stopped, it
cannot be restarted except from the Manual Override screen.]
17
To adjust the travel of the bottle indexing, remove the end caps from the box tube under
the bottle feed ramp which holds the air cylinder and loosen the two bolts located inside
the box tube. The air cylinder can now be slid forward or backward to adjust the travel of
the bottles so they are centered under the capping head, and the bottles under the filling
head are centered in the bottle locator with the air cylinder fully extended. When the
required location of the air cylinder has been found, tighten the two bolts for the air
cylinder.
CAP FEED:
The cap feeding mechanism should not require adjustment unless a bottle size change is made
which may require new riser blocks be installed. Contact Meheen Manufacturing regarding any
bottle size changes.
NOTE 1: For crowns to feed properly, the crown feed shoots must be kept dry during
operation and the crowning head must be fully in the up position when crowns are
dropped. If the crown shoots become wet they may feed two crowns rather than one,
resulting in a crown jam.
NOTE 2: The bottom face of the bottle ejection plunger should be flush with the bottom
of the stainless steel crowning bar to locate the crown and keep it in the proper
orientation. The location of the plunger is set by the ejection air cylinders located on top
of the capping bar and are not adjustable.
18
COMPUTER:
The bottling machine computer program is designed to perform the bottling process based on
certain events taking place in a certain fashion. Items of primary concern to the computer are the
detection of filled bottles and proper bottle pressures. When the computer detects a pressure or
fill problem it performs a predetermined response to identify the problem and stop the machine
safely. When the machine experiences no or low counter pressure, or it takes in excess of 40
seconds to fill a row of bottles, the machine performs a safe shut down routine. This means any
items which make pressure are shut off and the Snift valve opens to relieve pressure from the
bottles. During a safe shut down the filling heads will remain in the down position until the
machine is reset, at which time the fill heads will return to the up position. It should also be
noted that any time the machine is changed from running a program to a paused condition the
computer performs the safe shut down routine.
The following is a list of outputs to the computer. Pressure sensors are analog and do not show
on this list. Outputs are listed from left to right; the first module is 0-17 (12-17 are not used).
OUTPUT
0
1
2
3
4
5
6
7
10
11
12
FUNCTION
TUBING SIZE AND COLOR ON MACHINE
Speed valve
5/32 Yellow
Snift valve
5/32 Orange
Purge/Vent valve
5/32 Blue
CO2 pulse
5/32 Red
Cap drop
1/4 Green
Bottle index
1/4 Blue
Fill head
1/4 Gray
Fill valve
1/4 Yellow
Crowner
3/8 Red
Educator
5/32 Black
Tank Head space CO2 Valve mounted to right side of control panel
Product sensor
5/32 Natural
Bottle sensor
5/32 Blue
Temperature RTD, located in product manifold, uses wire connection.
The following are descriptions of possible problems with the pressure or fill detection and the
resulting actions by the computer. This section should be very useful for determining which
parameters need to be adjusted and for troubleshooting the machine when a part failure is
suspected. The description of the problem is given, then a possible resolution to the problem is
suggested.
1) Fill detection is not set sensitive enough to trip, indicating to the computer that the bottles are
not full. The machine starts normally by indexing bottles, indexing caps into the capping
head, capping head puts caps on bottles, the fill head stays down, and the fill valve remains
open until the computer times out (about 40 seconds). When time out occurs, the computer
performs a safe shut down and displays the Filling Time Out message.
Corrective Action: Increase the sensitivity of the fill detector by decreasing its numeric value.
19
2)
Fill detection is set so sensitive that it trips early, indicating to the computer that the
bottles are full even when they are only partially filled. The machine starts normally by
indexing bottles, capping head puts caps on bottles, fill head remains down. The fill
valve opens, then closes, as the fill valve closes the Snift valve opens and releases the
pressure from the bottles. The fill head pulls up quickly and a new row of bottles are
indexed.
Corrective Action: Reduce the sensitivity of the fill detector by increasing its value.
TROUBLESHOOTING:
1.
2.
3.
C)
CO2 valve is leaking into the off-gas tubing of the filling head and the
valve should be cleaned or replaced.
Liquid pressure sensor has been damaged by over pressurization. Check
pressure sensor values using a separate gauge and regulator and the
manual override feature to compare the sensor value against the gauge.
(See sensor check section of this manual.)
Liquid in bottle sensor and line. Drain liquid from line and wick liquid
from sensor using a paper towel with the tube fitting removed. (DO NOT
USE COMPRESSED AIR FOR CLEANING).
20
4.
6.
8.
Check bottle indexing and fill head alignment per instructions in this
manual.
Check with bottle manufacture for possible changes in bottle dimensions
or bottles improperly tempered.
7.
The most common cause of this problem is a power surge of the electrical
circuit where the bottling machine is plugged. If you experience a CPU
fault, shut off the machine for 1 minute, then turn the power back on. The
computer will automatically reboot the program. Repeated power surges
can cause permanent damage to the computer and should be corrected.
NOTE: If electrical surges are experienced, Meheen Manufacturing
recommends installing a good quality surge suppresser.
21
9.
22
23
4. Place the end of the tubing protruding from the top of the pinch valve tube guide fully on the
tubes of the manifold. Grasp the silicone tubes from below the pinch valve and pull them
down while pushing the liquid manifold down using a back and forth motion. When fully
down you may still have a small gap between the manifold and pinch valve, this is normal.
Above, round style manifold with RTD temperature probe. Machines are shipped with the
control panel removed and the temperature probe removed from the manifold. The probe is
spring loaded and simply pushes in to connect to the manifold similar to the pneumatic tubing
connections on the machine. The tip of the RTD probe must be in firm contact with the SS in the
manifold to provide accurate readings. Uses nylon ties to secure the manifold to the tube guide
as shown.
24
25
26