Sonda O2 Fuji
Sonda O2 Fuji
Sonda O2 Fuji
INZ-TN5ZFK8b-E
PREFACE
Thank you very much for your kind purchase of Fujis zirconia oxygen analyzer detector (Type ZFK 8).
Read this instruction manual carefully to get a thorough understanding of how this zirconia oxygen analyzer
works prior to installing, operating and maintaining the zirconia oxygen analyzer.
If abused, unnecessary troubles or failure could occur.
The specification of this zircoia oxygen analyzer may be subject to change without previous notice for improvements of the product.
Under no circumstances should this zirconia oxygen analyzer be modified without permission.
If any trouble should occur because of having been modified without permission, we wont be responsible
for it anyway.
This instruction manual should be kept in custody by a person who operates the zirconia oxygen analyzer
actually.
After reading this manual, it should always be kept in a place which allows the person who operates it to
refer to any time as requirfed.
A due consideration should be given so that this instruction manual is delivered to a final user certainly.
Manufacturer
Type
Date of manufacture
Product nationality
NOTE
It is prohibited to transfer a part or all of the contents of this manual
without permission.
The contents of this manual may be subject to change in future without previous notice
SAFETY PRECAUTIONS
Prior to operating this detector, read this Safety precautions carefully for correct use.
In the precautions shown here, important contents on safety are included. So, be sure to observe them. The safety
precautions have been ranked into DANGER and CAUTION.
DANGER:
CAUTION:
DANGER:
CAUTION:
Install this product in a place compatible with the conditions set forth in
instruction manual. The use at a place not conforming to the installation
conditions may result in an electric shock, a fire and malfunction.
When this product is mounted on a furnace which is under operation, take
utmost care with blow-out from the furnace. There is a risk of burn injury.
In the case of the wiring work, be careful not to drop foreign matters including
wire chips into the product. Otherwise, a fire, failure or malfunction may
result.
Connect a power source compatible with the specified rating. Connection of
power source different from the rating might cause a fire.
Before doing the wriring work, be sure to turn off the main power. Otherwise,
it results in getting an electric shock as the case may be.
Use proper wiring materials according to the rating of apparatus. If a wiring
material which is not bearable to the rating is employed, it might cause a fire.
PROHIBITION: Never do the work at a place where the product gets wet with water, such as
the rain.
A failure to observe this instruction may result in getting an electric shock or
failure.
ii
In case where combustible gas is contained in the measured gas, check the gas
composition and specifications carefully before using. Otherwise, the original
performance is not displayed, and there is a fear of explosion.
CAUTION:
Do the work in a condition where the main power has been turned off. If the
work is done while current is flowing, there is a fear of getting an electric
shock.
The operating temperature of the detector (tip of ceramic heater) is about 800
C and the surface temperature is also very high. So, never touch the detector
by bare hand.
Otherwise, there is a fear of getting a burn.
Before cleaninhg the flow guide tube, turn off the main power and cool the
tube down fully.
There is a fear of getting a burn.
Dont use any other renewal part than those designated by the maker. Otherwise, the original performance is not fully displayed and a accident or failure
may result.
Dispose of the renewal parts such as the maintenance parts as an incombustible.
PROHIBITION:
Never do the work at a place where rain water splases the product directly.
A failure to observe this instruction may result in getting an electric shock or
failure.
Other precaution
CAUTION:
For a failure which cannot be judged even if referring to the instruction manual,
be sure to ask the nearest dealer or Fuji adjustment serviceman for repair. If
disassembled carelessly, an accident or injury could result.
iii
CONTENTS
PREFACE ......................................................................................................................................... i
SAFETY PRECAUTIONS .............................................................................................................. ii
1.
Introduction .............................................................................................................................. 1
1.1
1.2
1.3
1.4
1.5
2.
Mounting .................................................................................................................................. 3
2.1
2.2
3.
8.
7.
6.
Wiring ..................................................................................................................................... 12
4.1
4.2
4.3
5.
Piping ....................................................................................................................................... 7
3.1
3.2
3.3
3.4
4.
Check .........................................................................................................................................16
Maintenance ..............................................................................................................................17
Standard output of detector .......................................................................................................19
Arrangement ..............................................................................................................................20
Troubleshooting...................................................................................................................... 21
Appendix ................................................................................................................................ 22
8.1
8.2
8.3
8.4
Specication ..............................................................................................................................22
Designation of type (code table) ...............................................................................................24
Device composition ..................................................................................................................25
Outline diagram (unit: mm) .......................................................................................................26
iv
1.
INTRODUCTION
1.1
High-oxygen partial
pressure side PH
O2
Nernsts equation
Solid electrolyte
Ionization
E=
2-2OO
Electrode
O2
RT
4F
PH(O2)
1n
PL(O2)
Molecularization Electromotive
force
E
: Electromotive force
PH(O2) : Oxygen partial pressure of a
reference gas (atmosphere)
PL(O2) : Oxygen partial pressure of a
measured gas
R
: Gas constant 8.3144[Jm ol-1K-1]
T
: Absolute temperature[K]
F
: Faraday constant 9.649104[cmol-1]
Low-oxygen partial
pressure side PL
Tree-phase interface:
a part where electrochemical
reactions occur.
Zirconia
Platinum
Oxygen molecule
1.2
Converter
Cable
(sensor output line, thermocouple output line,
heater power wire)
Flow of gas
1.3
CAUTIONS
The operating temperature of detector (tip of ceramic heater) is about 800C and the surface temperature is also very high. So, never touch it by bare hand. Otherwise, there is a fear of getting a burn.
Especially when a ceramic lter at the end of the detector is replaced, take utmost care.
(1) ZFK2
Furnace environment
(Gas to be measured)
Ceramic filter
Air environment
(Reference gas)
O-ring
Thermo sticker
Terminal box
Filter frame
Measured gas
Quartz filter
Ceramic heater
Heat insulating
material
Calibration
gas inlet
Caution in handling
Since the detector is made of porcelain of zirconia, there is a case where it breaks due to drop or
impact. So, be sufficiently careful.
The detector is unusable for a plant in which water droplets might run down inside a sampling pipe.
There is no need to remove M5-hexagon set screw of the terminal box unless otherwise a reference
gas inlet is used. Dont loosen the screw without thought since a waterproofing effect becomes zero.
1.4
Check of type
The name of type has been put in the specication name-plate. Make sure it is as ordered.
Refer to Item 8.2 Desingation of type.
1.5
1
2
3
4
5
6
7
8
9
Description
Detector
Instruction manual (this manual)
O-ring
Mounting screw, plain & spring washer (M5)
Thermo sticker
Ceramic filter
Flow guide tube
Heat insulating cover
Reference gas inlet
Classification
Q'ty
1 unit
1 copy
Accessories
Accessories
Accessories
Accessories
Specification goods
Specification goods
Specification goods
1 pc.
6 pcs each
1 sh.
1 sh.
As ordered
As ordered
As ordered
Remark
Confirmed in Item 1.3
INZ-TN5ZFK8-E
See Item 2.2.1.
See Item 2.2.1.
See Item 2.2.1.
See Item 6.2.3.
See Item 2.2.2 and 2.2.3.
See Item 2.2.4
See Item 3.2.
2.
MOUNTING
2.1
Mounting location
DANGER
This product has no explosion-proof specification. Dont use the product in an explosive gas
environment. If used, a serious trouble such as a fire or explosion might occur.
Install this product at a place compatible with the following conditions. The use of it at a place not
conforming the installation conditions specified in this manual could cause an electric shock, a fire
or incorrect operation.
CAUTIONS
2.2
Mounting method
CAUTIONS
When mounting the detector on a furnace which is under operation, take utmost care about the
blowout from the furnace; otherwise, there is a fear of getting a burn.
2.2.1
Caution in mounting
Up
Up
Down
Detector
Up
Down Down
thermo sticker
+45
-45
O-ring (accessory)
Detector
Gas flow
Detector flange
2
3
4
Wiring hole
Attach the O-ring (accessory, Viton P36) to the groove of the detector. Subsequently, attach the detector to the ange of the ow guide tube or ejector (type: ZTA) by xing the mounting screws (accessory,
M5) and the at and spring washers (accessories) to the three locations (at the opposing corners of six
mounting holes) (recommended tightening torque: 3.5 N m).
Attach the detector so that it is within the range of 45 to the horizontal plane.
Attach the detector so that the wiring hole is located at the bottom.
Set the temperature of the detector ange to 125C or less regardless of the measured gas temperature.
<How to check>
Complete the mounting, piping and wiring of the detector, make sure the color of thermo sticker 125 put
on the detector ange is not changing to red in the exhaust gas measuring condition (while current is owing to the detector and the plant is under running). (Usually, the color of thermo sticker is light pink.)
If the color has changed to red, it measns that the temperature of detector ange has been over 125C. So,
take the following steps:
(a) Change the existing flange packing to a thicker one.
(b) Use a longer mating flange.
(c) Mount the flow guide tube according to Item 2.2.2.
By taking the above steps, minimize heat transfer from the gas duct wall and lower the temperature.
The thermo sticker does not return to its originatl condition once it discolors. So, after taking the steps, reput a thermo sticker available as an accessory on the detector ange and make certain that it does not turn
red. (For the part No. for additional procurement of the thermo sticker, refer to Item 6.4.
2.2.2
The ange of ow tube has mounting holes at 8 locations. These holes are available for regulating an inow into the ow guide and mounting the tube correctly in the owing direction of gas and it is enough if
mounted at 4 locations.
(1) Direction of tongue and Partition plate of ow guide tube
1 When exhaust gas temperature is under 200C and gas owing velocity is low
As illustrated below, set the partition plate inside the ow guide tube at a right angle to the gas ow and
mount the tube so that the tongue turns to an upstream direction relative to the gas ow.
Packing (not included
in scope of supply)
Furnace wall
Partition plate
Tongue
Arrow mark of
flow guide tube
Gas flow
min.60mm
Mating flange (not included
in scope of supply)
15 (JIS5K-65A)
When exhaust gas temperature is 200C or higher and gas flowing velocity is fast
As illustrated below, till the partition plate inside the flow guide tube 45 to the gas flow and mount the
tube so that the tongue turns to a down-steam direction relative to gas flow.
Gas flow
Furnace wall
Gas flow
Tongue
Partition plate
Arrow mark of
flow guide tube
min.150mm
+
Inserting
angle
Down
Inserting
angle
2.2.3
Gas outlet
Gas flow
Gas flow
Gas flow
Gas flow
Up
+
Inserting
angle
Be careful not to block the gas outlet by the furnace wall or pipe and
keep the periphery of the outlet widely.
Set the inserting angle within a range of 0 to +45.
2.2.4
Down
For using at a cold distric, use a heat insulating cover (For the part No. for additional procurement of the heat
insulating cover, refer to Item 6.4).
Put M4-mounting screw of the heat insulating cover into M4-heat insulating cover xing tapped hole of
the teminal box with M4-screw ring put between the tapped hole and screw.
Do not use a heat insulating cover in the place where the temperature is high.
Terminal box
M4-Mounting screw
M4-Screw ring
M4-tapped hole
(for fixing heat insulating cover)
3.
PIPING
3.1
Main body
(calibration gas inlet)
(1) Coupling
(2) Nut
3.2
CAUTIONS
When used at an ordinary place (oxygen concentration : 20.6vol%), the reference gas is unnecessary.
So, dont remove a hexagonal plug of the mounting port of reference gas inlet unnecessarily since a
waterproofness becomes poor.
Refer to the previous section 3.1 Piping of calibration gas for the
installation method of piping.
Set one of the two reference gas ports as inlet and the other as outlet.
Install piping so that dust or water does not enter the outlet.
3.3
1 Packing
2 Reference gas inlet
3.4
Piping drawing
Reduction valve
(unnecessary when supply air pressure is
200 to 300kPa)
Solenoid valve
(not supplied)
Supply air
15ASGP
tube equivalent
or 10/8 PTFE
pipe or copper pipe
(not supplied)
15ASGP
tube equivalent
or 10/8 PTFE
pipe or copper pipe
(not supplied)
Rainproof cover
Detector
(ZFK8)
AC power
supply
Gas temperature.
600C max.
5
Sample gas outlet
Yellow
Red
*2
Ref. air
Blue
Rainproof flexible
conduit.
(Max. 20m)
White
Black
White
Gas temperature.
600C max.
Flow guide tube
for high particulate
Converter (ZKM)
1
Gas temperature.
600C max.
Flow guide for high
particulate with cover
O2
TC1
O2 input
Thermocouple
input
*4
6mm or 1/4 inch tube
for calibration gas
(not supplied)
6
AO
10
BLOW
11
12
13
TM-1
14
ZERO
GAS
Calibrating
HEATER
Blow
Fault
Heater
contact gas contact
contact
output
output
output
Alarm
Maintenance
contact
Calibrating gas
contact
output
contact output
output
Contact input
ALARM
TC2
DI1
DI2
DI3 DICOM
15
17
18
19
16
ZV
FAULT
Analog output
4 to 20mA DC
or 0 to 1V DC
Thermocouple
for combustion
control
or
20
SPAN
GAS
MAINTE
COM
21
22
23
24
25
SV
26
AC power
supply
PE
*3
N AC L
27
28
1
2
RS232C
or
RS485
(option)
TM-2
3
(Option)
Flowmeter
Valve
Reduction valve
(ZBD6)
MEASURE
ZERO
Standrad gas
for zero point (ZBM)
SPAN
Reduction valve
(ZBD6)
*1
Span gas (ZBM)
Flowmeter
*2 Ref. air
*1 Air supply
to Detector
(0.2 to 0.5L/min)
(not supplied)
Reduction valve
(unnecessary when
supply air pressure
is 50 to 100kPa)
Solenoid valve
(not supplied)
Reduction valve
(unnecessary when supply air pressure is
200 to 300kPa)
Supply air
15ASGP
tube equivalent
or 10/8 PTFE
pipe or copper pipe
(not supplied)
15ASGP
tube equivalent
or 10/8 PTFE
pipe or copper pipe
(not supplied)
Rainproof cover
Detector
(ZFK8)
AC power
supply
Gas temperature.
600C max.
1
5
3
2
*2
Ref. air
Blue
Yellow
Red
Rainproof flexible
conduit.
(Max. 20m)
White
Black
White
Gas temperature.
600C max.
Flow guide tube
for high particulate
Converter (ZKM)
1
Gas temperature.
600C max.
Flow guide for high
particulate with cover
O2
TC1
O2 input
Thermocouple
input
6
AO
Thermocouple
for combustion
control
Fault
contact
output
Alarm
contact
output
Contact input
ALARM
TC2
DI1
DI2
DI3 DICOM
15
17
18
19
16
10
11
12
13
TM-1
14
ZV
FAULT
Analog output
4 to 20mA DC
or 0 to 1V DC
*4
or
20
ZERO
BLOW
GAS
Blow Calibrating
contact gas contact
output
output
Maintenance
contact
output
Calibrating gas
contact output
22
23
AC power
supply
SPAN
GAS
MAINTE
COM
21
HEATER
Heater
24
25
PE
*3
N AC L
SV
26
27
28
1
2
RS232C
or
RS485
(option)
TM-2
3
AC power
supply
Solenoid valve
(not supplied)
Joint
Reduction valve
(ZBD6)
Standrad gas
for zero point (ZBM)
Flowmeter
(ZBD4)
(1 to 1.5L/min)
Reduction valve
(unnecessary when supply air pressure is 50 to 100kPa)
*2 Ref. air
*1 Air supply
to Detector
(0.2 to 0.5L/min)
(not supplied)
Solenoid valve
(not supplied)
Reduction vale
(ZBD6)
Span point
Note: *1 Standard gas or instrumentation air can be used in place of span gas.
*2 Instrument quality air or bottled air is available as reference air instead
of ambient air.
*3 Protective earth.
*4 Connect the shield of a exclusive cable with the ground terminal in
the converter.
*1
Standard gas for
span point (ZBM)
Flowmeter
(ZBD)
Ejector (5 to 10L/min)
Reduction valve
(unnecessary when supply air pressure
is 200 to 300kPa )
Joint
Supply air
1
2
4
5
Gas temperature.
1500C max.
Power supply
Solenoid valve
(not supplied)
Copper pipe
10/8mm
(not supplied)
Detector (ZFK8)
6
1
5
AC power
supply
3
2
4
White
Rainproof flexible
conduit.
(Max. 20m)
Black
White
Bule
Yellow
Red
O2
O2 input
TC1
6
AO
Thermocouple
input
Thermocouple
for combustion
control
10
11
12
13
TM-1
14
FAULT
BLOW
ZERO
GAS
Calibrating
HEATER
Blow
Fault
Heater
contact gas contact
contact
output
output
output
Alarm
Maintenance
contact
Calibrating gas
contact
output
contact output
output
Contact input
ALARM
TC2
DI1
DI2
DI3 DICOM
15
17
18
19
16
ZV
Analog output
4 to 20mA DC
or 0 to 1V DC
*4
or
20
SPAN
GAS
MAINTE
COM
21
22
23
24
25
SV
26
AC power
supply
*3
PE
N AC L
27
28
1
2
RS232C
or
RS485
(optein)
TM-2
3
(Option)
Flowmeter
Valve
AC power
supply
Reduction valve
(ZBD6)
MEASURE
ZERO
Standrad gas
for zero point (ZBM)
SPAN
Reduction valve
(ZBD6)
Note: *1 Standard gas or instrumentation air
Flowmeter
can be used in place of span gas.
*2 Instrument quality air or bottled air is
*2 Ref. air
available as reference air instead of
ambient air.
to Detector
*3 Protective earth.
*4 Connect the shield of a exclusive cable
(0.2 to 0.5L/min)
with the ground terminal in the converter.
(not supplied)
*1
Span gas (ZBM)
*1 Air supply
Reduction valve
(unnecessary when
supply air pressure
is 50 to 100kPa)
10
Ejector system
Ejector (ZTA)
Flowmeter
(ZBD)
Ejector (5 to 10L/min)
Reduction valve
(unnecessary when supply air pressure
is 200 to 300kPa )
Joint
Supply air
1
2
4
5
Gas temperature.
1500C max.
Power supply
Solenoid valve
(not supplied)
Copper pipe
10/8mm
(not supplied)
Detector (ZFK8)
6
1
5
AC power
supply
3
2
4
White
Rainproof flexible
conduit.
(Max. 20m)
Black
White
Bule
Yellow
Red
O2
O2 input
TC1
6
AO
Thermocouple
input
Thermocouple
for combustion
control
10
11
12
13
TM-1
14
BLOW
ZERO
GAS
Calibrating
HEATER
Blow
Fault
Heater
contact gas contact
contact
output
output
output
Alarm
Maintenance
contact
Calibrating gas
contact
output
contact output
output
Contact input
ALARM
TC2
DI1
DI2
DI3 DICOM
15
17
18
19
16
ZV
FAULT
Analog output
4 to 20mA DC
or 0 to 1V DC
*4
or
20
SPAN
GAS
MAINTE
COM
21
23
22
24
25
AC power
supply
*3
PE
N AC L
SV
26
27
28
1
2
RS232C
or
RS485
(optein)
TM-2
3
AC power
supply
(1 to 1.5L/min)
Joint
Solenoid valve
(not supplied)
Reduction
vale (ZBD6)
Standrad gas
for zero point (ZBM)
Flowmeter
(ZBD4)
Reduction valve (unnecessary when
supply air pressure is 50 to 100kPa)
*2 Ref. air
to Detector
*1 Air supply
Solenoid valve
(not supplied)
(0.2 to 0.5L/min)
(not supplied)
Note: *1 Standard gas or instrumentation air can be used in place of span gas.
*2 Instrument quality air or bottled air is available as reference air instead
of ambient air.
*3 Protective earth.
*4 Connect the shield of a exclusive cable with the ground terminal in
the converter.
11
Reduction
vale (ZBD6)
Span gas
*1 Standard gas for
span point (ZBM)
4.
WIRING
CAUTIONS
In the case of the wiring work, be careful not to drop foreign matters including wire chips inside the
product. Otherwise, this might cause a fire, failure or malfunction.
Connect a power source compatible with the rating. Connection of a power source not conforming
to the rating may cause a fire.
Before proceeding with the wiring work, be sure to turn off the main power supply. Otherwise, there
is a fear of getting an electric shock.
As the wiring material, use a proper one conforming to the rating of apparatus. The use of a wiring
material which is not bearable to the rating could cause a fire.
PROHIBITION
Under no circumstances be the work at a place where the product gets wet with water, such as the
rain. Otherwise, an electric shock or failure may result.
12
4.1
Before wiring
Put a cable (6 cores in all) connected between detector and converter into a piping tube for protecting the
cable. Also, put the cables for R thermocouple and element output away from the power cable to take a
noise preventive step.
When an exclusive cable is not used, use the following wire rods:
For heater (2 pcs.) ...............3A or more in rating
For R thermocouple ............Specied in JIS C1610-1995 (equivalent to RCA-2-G-0.75mm2-S2).
Recommended wire rod (at 20 C)
Element output
compensation conductor
0.75
For heater
Composition
30/0.18
1.1
1.14
0.6
1.0
24.4
1000
5
7
1.5
1500
40
For wiring of the detector, be sure to use a solderless terminal (for M3).
Recommended solderless terminal ............ Solderless terminal prescribed in JIS C 2805 (Nominal designation: R1.25-3)
4.2
CAUTIONS
Be careful not to mix up the wiring for thermocouple (four-core, white) and the wiring for heater
(two-core, white).
(1) Black
2 core
wire
Exclusive cable
for converter
Heater power
100 to 120VAC 50/60Hz
or 200 to 240VAC 50/60Hz
(1) Black
(2) White
(6) Blue
(3) Red
S+ H
6
1
5
S-
T+
3
2
T-
(4) White
(5) Yellow
Sensor output
(5) Yellow
(6) Blue
13
(2) White
Connect the protective grounding to one of the two terminals in the gure below. (Class D (Class 3)
grounding, grounding resistance: 100 or less)
T+
Note
3
made in japan
TK4J4664R0
S+ H
6
1
S-
T-
Grounding
terminal
Use the cable more than 0.75mm2 for main ground (earth) line.
For solderless terminal , doubly caulk the core and the sheath
separately.
The core should be poked from 0.5 to 1.0 mm.
The core should be so as not to
be seen or not to be come apart.
Grounding
terminal
Use
Tooth lock
washer
Double caulking
4.3
Mounting of conduit
Remove (2) packing holder nut,(3) packing and (4) bushing from M coupling of the main body.
Fit the (2) packing holder nut and (3) packing onto the conduit and t the end face of conduit into a groove
of the (4) bushing.
Insert the (4) bushing tted to the conduit into (1) M coupling of the main body and tighten with the (3)
packing and (2) packing holder nut for xing.
(2) Packing holder nut
(4) Bushing
Conduit
(4) Bushing
(3) Packing
(3) Packing
Conduit
14
5.
5.1
In case where combustible gas is contained in the measured gas, make sure of the gas composition
and specifications carefully before using this product. Otherwise, the original performance is not
displayed and there is a fear of explosion.
Start of operation
If the power switch of the converter is turned on after completion of wiring and piping work, the detector
starts its operation.
After warm-up at least 10 minutes, start the operation of the furnace.
After zero calibration and span calibration have nished, get to work on the measurement.
For the method of calibration, refer to each instruction manual of converters (ZKM, ZRM and ZRY).
When the converters (ZKM, ZRM and ZRY) are not used, run the zero and span calibration gases and
calibrate by converting the output in a stabilized state into oxygen concentration according to the standard
output table of converter in Item. 6.3.
If using the reference gas, supply it before the calibration. (ow rate: 0.2 to 0.5L/min)
5.2
Stop of operation
If the power is turned on in in a state of dew condensation, it leads to the failure of detector. Stop the operation following the procedures described below.
(1) In case of short-term shutdown (about 1 week) of furnace
Keep the power supply of the detector (converter) turned ON. This can prevent the detector from
getting dewed.
Also, note that if ON-OFF is repeated in a condition where the detector has dewed (according to the
furnace and ambient conditions), the detector might fail.
When the ejector (ZTA) is in use, stop the air supply to the ejector.
(2) In case of long-term shutdown of furnace
Turn off the power of the detector (converter) after the peripheral air of the detector inside the furnace
(especially, temperature and humidity) has become an air environment. Or, turn off the power after
taking the detector out of the furnace and leaving it as it is for 15 minutes or more.
When the ejector (ZTA) is in use, stop the air supply to the ejector.
15
6.
Do the work in a condition where the main power supply has been turned off.
If the work is done while current is flowing, there is a fear of getting an electric shock.
The operation temperature of the detector (tip of the ceramic heater) is about 800C and the surface
temperature is also very high. So, never touch it by bare hand. Otherwise, there is a fear of getting a
burn.
Before proceeding with the cleaning of the flow guide tube, turn off the main power and cool the tube
down fully and then, do the work. Otherwise, there is a fear of getting a burn.
Dont use other renewal parts than those designated by the maker. Otherwise, the original performance
is not displayed fully and an accident or failure could come about.
Dispose of the renewal parts including the maintenance parts as an incombustible.
PROHIBITION
6.1
Under no circumstances the work at a place where the product gets wet with water, such as the rain.
Otherwise, an electric shock or failure may result.
Check
Perform the check periodically for using the product always in good condition. Especially, perform the
checks shown in table below. Moreover, perform the periodic check at a time of checking the furnace or
every 6 months.
Daily check
Check items
Execution of zero & span calibration
Periodic check
16
6.2
Maintenance
The replacing intervals of sensor unit, ceramic lter and O-ring, and the maintenance periods of ow guide
tube and sampling probe differ depending on the working conditions including the components of measured
gas and the amount of dust.
The replacing intervals in a general conditions are shown below. Determine the replacing intervals in the
individual working condition with a period till a rst replacement after delivery and operation as a rough
standard.
Sensor unit ................................................ Yearly
Ceramic lter............................................ At 6 month interval
Flow guide tube ........................................ At 3 to 4 year interval
ZTA sampling probe ................................. At 3 to 4 year interval
O-ring ....................................................... Yearly
6.2.1
Check that the sensor unit complies with the specifications of the power supply you use.
Avoid replacing the sensor unit while the current is applied.
The temperature of the detector becomes high. Replace the sensor unit after the temperature is reduced sufficiently.
(1) Turn off the power for the detector. (Set the switch of the converter or the main power supply to OFF.)
(2) Check that the surface temperature of the detector is reduced sufciently.
(3) Remove the terminal box cover and then remove the terminal block cover in the terminal box (M3 screw,
two locations).
(4) Remove the six wires connected the terminal block of the sensor unit (M3 screws, six locations).
(5) Remove the screw with washer (M3 screws, four locations) fastening the sensor unit .
(6) Remove the sensor unit and o-ring from the detector.
(7) Attach the supplied O-ring of the replacement sensor unit to the groove for the O-ring at the detector.
(8) Attach the replacement sensor unit to the detector.
(Attach the sensor unit so that the rotational direction is as shown in the gure below.)
(9) Fasten the sensor unit with the supplied screw with washer (M3 screws, four locations).
(recommended tightening torque: 0.7 Nm)
(10) Connect six wires to the terminal block of the sensor unit and attach the terminal block cover.
(11) Attach the terminal box cover.
Groove for the O-ring
O-ring
Sensor unit
(1) Black
(6) Blue
(3) Red
S+ H
6
1
2 core
wire
Exclusive cable
for converter
4 core
wire
Heater power
100 to 120V AC 50/60Hz
or 200 to 240V AC 50/60Hz
(1) Black
(2) White
TK4J4664R0
Detector
made in japan
T+
5
S-
2
T-
(4) White
(5) Yellow
17
Grounding wire
screw: M4
(2) White
6.2.2
(1) After turning OFF the power of the detector, lower the surface
Reflector
temperature of the tip (at the ceramic lter side) by cooling down
fully with the air.
(2) After having been cooled down fully, remove the lter frame
from the detector, take the ceramic lter and reector off the lter
frame.
(3) Set a new ceramic lter and reector in the lter frame and t the
frame to the detector and then, tighten till the ceramic lter does
Ceramic filter
not move any longer. (Be careful then not to fail to set the reecFilter frame
tor in place. For the part No. for additional procurement of the
reector, refer to Item. 6.4).
6.2.3
6.2.4
CAUTIONS
The high temperature-use sampling probe (made of SIC) of the ejector (ZTA) is liable to break. So,
be careful not to cool abruptly by water (quenching) or apply undue force to the probe in the dust
removing work.
Remove dust sticking to the inside of the ow guide tube by using a metallic rod or screwdriver.
(Clean so that the tube is through at least about 3/4 part of the whole interior.)
18
6.3
Oxygen
concentration
(Vol%)
0.01
0.05
0.1
0.5
1.0
1.2
1.4
1.5
1.6
1.8
2.0
2.2
2.4
2.5
2.6
2.8
3.0
Detector output
(Unit: mV)
168.15
132.68
117.41
81.94
66.67
62.65
59.25
57.73
56.31
53.71
51.39
49.29
47.37
46.47
45.61
43.98
42.46
Oxygen
concentration
(Vol%)
3.5
4.0
4.5
5.0
5.5
6.0
6.5
7.0
7.5
8.0
8.5
9.0
9.5
10.0
11.0
12.0
13.0
Detector output
(Unit: mV)
39.06
36.12
33.52
31.20
29.10
27.18
25.42
23.79
22.26
20.84
19.51
18.25
17.06
15.92
13.82
11.91
10.14
19
Oxygen
concentration
(Vol%)
14.0
15.0
16.0
17.0
18.0
19.0
20.0
20.6
21.0
22.0
23.0
24.0
25.0
30.0
35.0
40.0
45.0
50.0
Detector output
(Unit: mV)
8.51
6.99
5.57
4.23
2.97
1.78
0.65
0.00
-0.43
-1.45
-2.43
-3.37
-4.27
-8.29
-11.68
-14.62
-17.22
-19.54
6.4
No.
Arrangement
Description
Classification
Ceramic filter
Consumable
Consumable
*ZZPZFK5-8552836
Viton
Spare parts
Spare parts
Additionally
procurement part
Additionally
procurement part
Additionally
procurement part
Additionally
procurement part
Additionally
*ZZPZFK5-TK7K1652P10
procurement part
6
7
10 Reflector
11 Heat insulating cover
Remark
*ZZPZFK5-TK7N6820C2
*ZZPZFK5-TK746983P1
*ZZPZFK5-TK7K1652P9
Gasket included.
(2 pieces are required for the
standard gas inlet/outlet.)
Gasket included.
(2 pieces are required for the
standard gas inlet/outlet.)
Additionally
*ZZPZFK5-TK7H6762P1
procurement part
Additionally
*ZZPZFK5-TK4E5339C1
procurement part
Classification
Inserting length
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
800mm
800mm
20
7.
TROUBLESHOOTING
CAUTIONS
If a failure should occur which cannot be judged even if referring to the operation manual, be sure to
ask the nearest dealer or Fuji adjustment serviceman for repair. If disassembled carelessly, there is a
fare of an accident or injury.
Troubles
Indication is fixed.
Indication response
is slow.
Probable causes
Clogging of ceramic filter
of detector and flow guide
tube interior
Leak from joint and airtightness of mounted part.
Deterioration of detector
Temperature alarm
continues coming on
despite 20 minutes
having elapsed after
turning on power.
Disconnection of cable
Error in wiring
Low supply voltage
Disconnection of thermocouple
Blow out of fuse of converters (ZKM, ZRM and
ZRY)
Disconnection of detector
heater
Looseness of flange
mounted part or defects
of O-ring or packing (not
included in scope of supply).
Deterioration of detector
Abnormality of detector
temperature
Change of ambient air
oxygen concentration of a
terminal box or very high
humidity
21
Remedies
Clean or exchange ceramic filter, if need be.
Increase amount of
exhaust gas inside flow
guide tube to be taken in.
Clean flow guide tube.
Replace cable.
Wire correctly.
Supply correct power.
Replace detector.
Exchange fuse(s).
(Refer to each instruction
manual of conver-ters
ZKM, ZRM and ZRY).
Replace detector.
Retighten mounting
screw.
Exchange O-ring.
Exchange packing
(not included in scope of
supply
Shield
Retighten calibration
gas joint.
Replace detector.
8.
APPENDIX
8.1
Specication
8.1.1
General
Measuring object:
Repeatability:
Within 0.5%FS
Linearity:
Within 2%FS
Response time:
Warmup time:
Analog output:
Power supply:
Rated voltage;
100 to 120V AC (operating voltage 90 to
132V AC)
200 to 240V AC (operating voltage 190 to
264V AC)
Rated frequency; 50/60Hz
8.1.2
8.1.3
Ejector (general-use):
Probe for guiding measured gas to detector
Flange; JIS10K 65A RF
Insertion length; 0.5, 0.75, 1, 1.5m
(according to customer's specication)
Operating temperature:
20 to +60C for detector
5 to +100C for ejector section
125C or less at detector ange surface with
power applied
Note2) Alarm selects just one as mentioned below (1) High (2)
Low (3) High and Lower (4) High-high (5) Low-low, it
turns ON while operating.
Note3) It is enable during blowdown when an option is specied, and it turns ON while operating.
Storage temperature:
Detector: 30 to +70C
Ejector: 10 to +100C
Structure:
Finish color:
Calculation reset
22
Heater OFF
Finish color:
Calibration start
Range change
Calibration method:
(a) Manual calibration with key operation
(b) Auto. calibration (option)
Calibration cycle; 00 day 00 hour to 99
days 23 hours
A function for blowing out dust with compressed air that has deposited in the ow
guide tube. Blowdown can be performed for
a predetermined time and at predetermined
intervals.
Blowdown cycle; 00 hour 00 minute to 99
hours 59 minutes
Output hold:
Cock(option):
Blowdown time;
Communication function:
RS232C (MODBUS) standard specication
RS485 (MODBUS)(option)
Combustion efciency display (option):
When you select this display, "rich mode
display" will be an simultaneous display.
This function calculates and displays combustion efciency from oxygen concentration and
measured gas temperature.
Thermocouple (R) is required for temperature
measurement.
Operating temperature:
20 to +55C
Operating humidity:
95% RH or less, non condensing
Storage temperature:
30 to +70C
Storage humidity:
Construction:
Material:
Aluminume case
23
8.2
8.2.1
Detector
1 2 3 4 5 6 7 8
Z F K 8 R
8.2.2
9 10 11 12 13
5 -
14 15 16
Power supply
Code symbols
100 to 120V AC
ZFK8YY15-0Y0YY-0YY
200 to 240V AC
ZFK8YY35-0Y0YY-0YY
Description
- 1
1
2
3
8.2.3
Power supply
100 to 120VAC 50/60Hz
200 to 240VAC 50/60Hz
1
3
1 2 3 4 5 6 7 8
Z T A
0Y 0
5A3
5A5
5A7
5A1
5B3
5B5
5B7
5B1
5C3
5C5
5C7
5C1
6D8
6E8
ZZZ
Y
A
Y
A
B
Filter spec.
Standard
J
E
C
1
2
Ejector
24
Description
Measured gas temperature
For high temperatures (+1500C max.)
General-use (+800C max.)
1
2
B
C
D
E
1
3
5
Power supply
100V/115V AC 50/60Hz
200V/220V AC 50/60Hz
230VAC 50/60Hz
8.2.4
Converter
(2) Exclusive cable for converter
(1) ZKM
1 2 3 4 5 6 7 8
Z KM
1 2 3 4 5 6 7 8
9 10 11 12
1 -
Z R Z KR
Description
Structure
Small case (IP66)
Large case (IP67)
Bench type
1
2
3
1
2
Optional Functions
None
Combustion efficiency display function Note4)
Blowdown
Auto calibration
Combustion efficiency display
+ Blowdown Note4)
Combustion efficiency display
+ Auto calibration Note4)
Blowdown + Auto calibration
Combustion efficiency display + Blowdown
+ Auto calibration Note4)
Y
1
2
3
4
5
6
7
0
1
2
Y
1
2
Cable length
6m
10m
15m
20m
30m
40m
50m
60m
70m
80m
90m
100m
6m
10m
15m
20m
Terminal treatment
None
One side (detector side)
Both sides
Display language
Japanese
English
Chinese
J
E
C
Conduit length
None
None
None
None
None
None
None
None
None
None
None
None
6m
10m
Note5
15m
20m
YA
YB
YC
YD
YE
YF
YG
YH
YJ
YK
YL
YM
AA
BB
CC
DD
Mounting bracket
None (Specify "None" when the bench
type is selected)
Mounting on panel surface
Pipe mounting
Types
For R thermocouple
Communication function
RS-232C
RS-485
1
2
Description
Connectable devices
For ZKM
Output signal
4 to 20mA DC
0 to 1V DC
Other
B
E
Z
1 -
Option
Option
With valve (Specify "None" when the bench
type or the auto calibration is selected)
With valve + flowmeter
8.3
Device composition
The device to be combined differ according to the conditions of the gas to be measured. Select the devices to be
combined with reference to the following table.
Application Temperature Gas Flow
Measured gas
DUST
Protection
cover
Note
Fuel; gas, oil
Fuel: coal
with blow down
Included low moisture
Included low moisture
with blow down
Included low moisture
with blow down
Included high moisture
with blow down
When a measurement gas has
a low water content with blow
down
When a measurement gas has
a low water content with blow
down
General-use 600C or
(boiler)
less
For
600C or
corrosive
less
gas (refuse
incinerator)
Sludge
no
yes
1m/s or
less
1m/s or
less
incinerator
General-use 800C or
(furnace)
less
1500C or
less
Device conguration
Detector type
Converter Ejector
type
type
ZFK8R
ZFK8R
5-5A
5-5C
ZKM
ZKM
ZFK8R
ZFK8R
5-5B
5-5C
ZKM
ZKM
ZFK8R
5-6D
ZKM
ZFK8R
5-6E
ZKM
ZFK8R
5-0Y0
ZKM
ZTA2
ZFK8R
5-0Y0
ZKM
ZTA1
25
8.4
Approx.62
Approx.130
67
5
6-6
Terminal box
60
Filter
85
Thermocouple
1
3
2
4
Black / White/ red / White/
Approx.30
Ref. gas inlet (as specified)
SUS316,for 6/4 tube
Or 1/4 inch tube
Heater
Element output
125.5
26
79
2-core wire
Ground-wire
Screw: M4
5
6
yellow / blue
4-core wire
Exclusive cable
Exclutive
Cable gland
20
(17.5)
(88.4)
T+
5
S-
45
TK4J4664R0
made in japan
40
S+ H
6
1
45
35
2
T-
12
Approx. 20
130
67
34
Gas inlet
20
ZFK8R
Gas outlet
6-M5
detector
mounting side
815
MTG. holes
Oxygen detector
26
Code 11th
L (m)
0.3
MASS
2.7
Approx.(kg)
5-5A3
5
7
1
4.1
4.8
L=
(as specified)
Approx. 40
Approx. L
130
3
67
67
60.5
Gas inlet
ZFK8R
6-M5
detector side
Gas outlet
Code 11th
L (m)
0.3
MASS
3.3
Approx.(kg)
5-5B3
5
7
1
5
6.1
7.6
L=
(as specified)
8-15
MTG. holes
Oxygen detector
Approx. L
18
Approx. 20
130
67
60.5
67
Gas inlet
6-M5
detector
mounting
side
ZFK8R
5-5C3
5
7
1
Gas outlet
Oxygen detector
27
Code 11th
L (m)
0.3
Mass
3.0
Approx.(kg)
815
MTG. holes
4.8
5.7
L=
(as specified)
18
0
1
45
8 - 19
MTG. hole
67
29
.5
185
Packing
ZFK8R
Gas outlet
Approx. L
5-6D3
5
7
1
50
'longate
hole
30
Code 11th
L (m)
0.3
Mass
4.5
Approx.(kg)
Gas flow
Tube (50A SCH40)
28
7.0
8.3
L=
(as specified)
45
1
18
0
67
2
9
.5
(175)
Packing
(Approx. L)
Approx. L+10
ZFK8R
5-6E3
5
7
1
30
Code 11th
L (m)
0.3
Mass
7.1
Approx.(kg)
Gas flow
50
'longate
hole
Gas outlet
(38)
29
L=
9.0 11.4 13.6 (as specified)
IN-SITU
DATA SHEET
This oxygen analyzer is used to continuously measure
oxygen concentration in combustible exhaust gas of industrial
boilers or furnaces, and is ideally suited for combustion management and control.
The analyzer system is comprised of the detector and converter coupled together as a complete system. Detector setting configuration includes the detector flow tube and detector sensor. The flow tube is inserted directly into the gas and
directs gas to the sensor for measurement. The converter
(ZKM) is comprised of the signal processor, input/output and
communications, display and system controls.
The converter is equipped with advanced functionality
such as performing the sensor diagnostics and sensor recovery function, so the detector can be used within long term
stability.
General-use detector
(ZFK8)
High-temperature detector
(ZTA)
FEATURES
1. Gas sampling device is unnecessary
For quick response, insert the detector directly into the
flue Gas sampling functions such as a gas aspirator and a
dehumidifier are not required.
2. Easy maintenance
The sensor equipped with the detector, has unit construction, it is easy to replace.
By separating the detector and the flow guide tube, filter
replacement is easy.
3. More reliable than sensor diagnosis, sensor recoverable function
Depending on the concentration of the measurement gas,
the power of the sensor might deteriorate. The equipment
includes sensor recovery function electronically, checking
the deterioration status of the sensor depletion.
Therefore, it has high reliability and long-lasting stability.
4. Safe and secure
System detects thermocouple break for heater control on
the sensor side. Safety functions of isolating power supply to the detector or isoralting power via external contact
input are also.
5. Easy operation
The operation and setting for the converter can be performed interactively, and available as English, Japanese or
Chinese for language display.
<IP66>
Converter (ZKM1)
<IP67>
Converter (ZKM2)
SPECIFICATIONS
General Specifications
Measuring object: Oxygen in noncombustible gas
Measuring method:
Directly insert type zirconia system
Measuring range: 0 to 2 setting range at option 2 in 50
vol% O2
(in 1 vol% O2 steps)
Repeatability:
Within 0.5%FS
Linearity:
Within 2%FS
Response time:
Within 4 to 7 sec, for 90% (from calibration gas inlet)
Warmup time:
More than 10 min
Analog output:
4 to 20mA DC (allowable load resistance
less than 500:) or 0 to 1V DC (output
resistance more than 100:)
Power supply:
Rated voltage;
100 to 120V AC (operating voltage 90
to 132V AC)
200 to 240V AC (operating voltage
190 to 264V AC)
Rated frequency; 50/60Hz
Power consumption:
Maximum 240VA (Detector: approx.
200VA, Converter: approx. 40VA)
Normal 70VA (Detector: approx. 50VA,
Converter: approx. 20VA)
EDSX3-137a
Date
chanical thermostat
N.O. (1a) contact, 200V AC, 2A
Converter specification (ZKM)
Concentration value indication:
Digital indication in 4 digits
Contact output signal:
(1) Contact specification; 6 points, 1a 250V AC/3A or 30V DC/3A
(2) Contact function;
Under maintenance
Under blowdown Note3)
Span calibrating gas
Zero calibration gas
Instrument anomalies Note1)
Alarm Note2)
Note1) The following Instrument errors (1) Thermocouples break (2) Sensor break (3) Temperature fault
(4) Calibration fault (5) Zero/span adjustment fault
(6) Output error turn the contact-ON
Note2) Alarm selects just one as mentioned below (1)
High (2) Low (3) Upper and Lower (4) High-high
(5) Low-low, it turns ON while operating.
Note3) Under blow down is available in case of option,
and it turns ON while operating.
Contact input signal:
(1) Contact specification; 3points (the following option)
ON; 0V (10mA or less), OFF; 5V
(2) Contact function;
External hold
Calculation reset
Heater OFF
Blow down (option)
Inhibition of calibration
Calibration start
Range change
Calibration method:
(a) Manual calibration with key operation
(b) Auto. calibration (option)
Calibration cycle; 00 day 00 hour to
99 days 23 hours
(c) All calibration
Calibration gas: Range settings
Zero gas; 0.010 to 25.00% O2
Span gas: 0.010 to 50.00% O2
Recommended calibration gas concentration
Zero gas; 0.25 to 2.0% O2
Span gas; 20.6 to 21.0% O2
(oxygen concentration in the air)
Blowdown:
A function for blowing out with compressed air dust that has deposited in
(option)
the flow guide tube. Blowdown can be
performed for a predetermined time and
at predetermined intervals.
Blowdown cycle; 00 hour 00 minute to
99 hours 59 minutes
Blowdown time; 0 minute 00 second
to 0 minutes 999
seconds
Output signal hold:
Output signal is held during calibration,
processing recoverable sensor, warmup, and blowdown. The hold function
can also be released.
Cock (option):
CODE SYMBOLS
(Detector)
4 5 6 7 8
ZFK 8 R
9 10 11 12 13
5 -
14 15
Description
- 1
1
2
Power supply
100 to 120VAC 50/60Hz
200 to 240VAC 50/60Hz
1
3
0Y 0
5A3
5A5
5A7
5A1
5B3
5B5
5B7
5B1
5C3
5C5
5C7
5C1
6D8
6E8
ZZZ
length
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
800mm
800mm
Protection cover
Without
With
Y
A
Y
A
B
Filter spec.
Standard
1
J
E
C
Electrical Safety:
Overvoltage category
; II power supply input
; I relay interfaces
(IEC1010-1)
External overcurrent protective device
; 10A
Equipment interfaces are safety
separated (SELV)
Power supply
Code symbols
100 to 120V AC
ZFK8YY15-0Y0YY-0Y
200 to 240V AC
ZFK8YY35-0Y0YY-0Y
ZFK, ZKM
(Ejector)
1 2 3 4 5 6 7 8
Z KM
1 2 3 4 5 6 7 8
9 10 11
Description
1 Construction
IP66
IP67
1
2
Optional Functions
None
Combustion efficiency display function Note4)
Blowdown
Auto calibration
Combustion efficiency indication + Blowdown
Combustion efficiency indication
+ Auto calibration
Blowdown + Auto calibration
Combustion efficiency indication + Blowdown
+ Auto calibration
Y
1
2
3
4
5
6
7
Display language
Japanese
English
Chinese
J
E
C
Y
1
2
(Exclusive-special cable)
1 2 3 4 5 6 7 8
Z R Z KR
Description
1 -
Connectable devices
For ZKM
Types
For R thermocouple
Conduit length
None
None
None
None
None
None
None
None
None
None
None
None
6m
10m
Note5
15m
20m
YA
YB
YC
YD
YE
YF
YG
YH
YJ
YK
YL
YM
AA
BB
CC
DD
0
1
2
Mounting bracket
Mounting on panel surface
Pipe mounting
1
2
Communication function
RS-232C
RS-485
1
2
Description
1
2
Output signal
4 to 20mA DC
0 to 1V DC
Other
B
E
Z
Power supply
100V/115V AC 50/60Hz
200V/220V AC 50/60Hz
230VAC 50/60Hz
SCOPE OF DELIVERY
Detector:
CAUTIONS
Cable length
6m
10m
15m
20m
30m
40m
50m
60m
70m
80m
90m
100m
6m
10m
15m
20m
Z T A
CONFIGURATION
Flow guide tube system
Rc1/4
(upper side blow port should be selected)
Blow-down, 200 to 300kPa
Solenoid valve
(not supplied)
Reduction valve
(unnecessary when supply air pressure is
200 to 300kPa
Supply air
15ASGP
tube equivalent
or 10/8 PTFE
pipe or copper pipe
(not supplied)
15ASGP
tube equivalent
or 10/8 PTFE
pipe or copper pipe
(not supplied)
Rainproof cover
Detector
(ZFK8)
Gas temperature.
600C max.
3 4
AC power
supply
6
Bule
Black
Red Yellow
White White
Converter (ZKM)
4-core
cable
2-core
cable
Thermocouple
input
O2 input
Gas temperature.
600C max.
Analog
output
Fault contact
output
Calibrating gas
Blow
contact
contact output
output
Heater
NC
10
11
12
13
Black White
4 to 20mA DC
or
0 to 1V DC
14
(TM1-1)
RS232C
or
RS485
(option)
Gas temperature.
600C max.
Contact input
Thermocouple
for combustion
control
(TM1-2)
C
15
16
17
18
20
21
22
23
24
25
*3
26
27
28
PE
AC power
supply
*2 Ref. air
19
Solenoid valve
(not supplied)
Reduction
valve (ZBD6)
Standrad gas
for zero point
(ZBM)
Flowmeter
(ZBD4)
(1 to 1.5L/min)
Reduction valve
(unnecessary when supply air pressure is 50 to 100kPa)
to Detector
*1 Air supply
(0.2 to 0.5L/min)
Solenoid valve
(not supplied)
Reduction vale
(ZBD6)
Span point
*1
Standard gas for
span point (ZBM)
Note: *1 Standard gas or instrumentation air can be used in place of span gas.
*2 Instrument quality air or bottled air is available as reference air instead of ambient air.
*3 Protective earth.
Flowmeter
(ZBD)
Ejector (5 to 10L/min)
Reduction valve
(unnecessary when supply air pressure
is 200 to 300kPa )
Joint
Supply air
1
2
4
5
Gas temperature.
1500C max.
Power supply
Solenoid valve
(not supplied)
Copper pipe
I10/I8mm
(not supplied)
Detector (ZFK8)
3 4
4-core
cable
2-core
cable
Converter (ZKM)
Thermocouple Analog
output
input
O2 input
Fault contact
output
Calibrating gas
Blow
contact
contact output
output
Heater
NC
10
11
12
13
Black White
14
(TM1-1)
1
(TM2)
RS232C
or
RS485
(optein)
3
Contact input
Thermocouple
for combustion
control
Alarm
contact
output
15
16
17
18
19
20
21
22
23
24
25
(TM1-)2
*3
26
27
28
PE
(1 to 1.5L/min)
Joint
Solenoid valve
(not supplied)
AC power
supply
Reduction
vale (ZBD6)
Standrad gas
for zero point (ZBM)
Flowmeter
(ZBD4)
Reduction valve (unnecessary when supply air pressure is
50 to 100kPa)
*2 Ref. air
*1 Air supply
to Detector
(0.2 to 0.5L/min)
Solenoid valve
(not supplied)
Reduction
vale (ZBD6)
*1 Standard gas for
span point (ZBM)
Note:
DEVICE CONFIGURATION
The device to be combined differ according to the conditions of the gas to be measured. Select the devices to be combined
with reference to the following table.
Measured gas
Application
General-use
600C or
(boiler)
less
Device configuration
DUST
Protection Note
cover
less
incinerator)
Less than 25g/Nm3 no
3
yes
General-use
800C or
(boiler)
less
1590C or
1m/s
Less than Less than 1g/Nm3
less
1m/s
Detector type
Converter Ejector
type
type
ZFK8R
5- A 5
-1
Fuel: coal
ZFK8R
5- C 5
-1
ZFK8R
5- B 5
-2
ZFK8R
5- C 5
ZFK8R
with blow down
Included high moisture ZFK8R
with blow down
SUS316 tube
ZFK8R
with blow down
ZFK8R
SIC tube
ZKM
ZKM
-2
ZKM
ZKM
5- D 6
-2
ZKM
5- E 6
-2
ZKM
5-0Y0
-1
ZKM
ZTA1
5-0Y0
-1
ZKM
ZTA2
OUTLINE DIAGRAM
(Unit:mm)
Mounting hole
2-9
2.5
Converter (ZKM1)
<IP66>
75
142
72
153
10
2.5
U bolts M8 (option)
170
239
200
Oxgen Analyzer
MODE
ESC
ENTER
POWER
Cock (option)
Rc 1/83
Mounting for 2B pipe
ZKM
Oxgen Analyzer
153
Mtg. Plate
Converter (ZKM2)
<IP67>
Mounting hole
4-9
195
72
20
U bolts M8 (option)
ESC
ENTER
POWER
312
220
MODE
Cock (option)
Rc 1/83
14
230
5
1
U bolts M8 (option)
95
5.7
39
4-G1/2
For power supply
For contact Input/output
For output
For communication
Thermocouple
input
O2 signal
input
Blow output
contact
250V AC/3A COM
25
30V DC/3A
Heater power
output
+ TC1
+ AO
FAULT
BLOW ZV
2
3
4
5
6
7
8
9
10 11
15 16 17 18 19 20 21 22 23 24
+ TC2 DI1 DI2 DI3 DI- ALARM
MAINTENANCE
COM
HEATER
12 13
25 26
COM SV
14
27
N
28
AC
L
SPAN
GAS
Thermocouple
inputtype R
Contact input
250V AC3A
30V DC3A
Alarm
contact output
250V AC/3A
30V DC/3A
Maintenance
output contact
250V AC/3A
30V DC/3A
Calibrating gas
contact output
250V AC/3A
30V DC/3A
40
40
40
ZERO
GAS
+ O2
1
Fault
contact output
contact capacity
Analog
250V AC/3A
output
4-20mA DC 30V DC/3A
or 0-1V DC
Power supply
input
100V AC to 120V or
200V AC to 240V
Detector (ZFK8)
Approx.132
I 80
Approx.62
Approx.130
I 67
5
6-I 6
Temp.Proof cover
(To order)
Terminal box
I 79
Heater
Element output
Thermocouple
60
125.5
I 26
I 85
1
3
2
4
Black / White/ red / White/
Approx.30
Filter
2-core wire
Ground-wire
Screw: M4
5
6
yellow / blue
4-core wire
Exclusive-special cable
Exclutive
Cable gland
3.5
20
(?17.5)
(88.4)
TK4J4664R0
5
S-
T-
45
made in japan
40
T+
S+ H
6
1
45
I 35
12
Approx. 20
130
67
34
Gas inlet
20
ZFK8R
Gas outlet
Oxygen detector
5-5A3
5
7
1
6-M5
detector side
815
MTG. holes
Cord 11th
L (m)
0.3
0.5
MASS
2.7
Approx.(kg)
3.3
0.75 1.0
4.1
4.8
L=
(to order)
Approx. L
18
Approx. 20
130
67
60.5
67
Gas inlet
6-M5
detector side
ZFK8R
5-5C3
5
7
1
Gas outlet
Oxygen detector
Cord 11th
L (m)
0.3
Mass
3.0
Approx.(kg)
3.8
4.8
815
MTG. holes
5
14
180
67
29.5
185
Packing
5-6D3
5
7
1
Gas outlet
800
50
ZFK8R
30
Cord 11th
L (m)
0.3
Mass
4.5
Approx.(kg)
Gas flow
Tube (50A SCH40)
10
7.0
8.3
L=
(to order)
5.7
L=
(to order)
180
14
8 - 19
MTG. holes
67
29
.5
(175)
50
Packing
ZFK8R
5-6E3
5
7
1
800
(790)
Gas outlet
30
Cord 11th
L (m)
0.3
Mass
7.1
Approx.(kg)
Gas flow
(38)
Approx. 40
Approx. L
130
3
67
67
60.5
Gas inlet
ZFK8R
Gas outlet
6-M5
detector side
Cord 11th
L (m)
0.3
MASS
3.3
Approx.(kg)
Oxygen detector
5-5B3
5
7
1
5
0.5 0.75
4.5 6.1
1.0
7.6
L=
(to order)
8- 15
MTG. holes
11
30
8
170
Gas inlet
217
25
105
190
Approx. L
140
(JIS 10K65ARF)
40
Detector
(ZFK8)
Ejector air
outlet
(Rc1/4)
35
Blow -down air inlet
(Rc1/4)
Ejector air inlet
(Rc1/4)
40
50
45
Cable gland
(A15C)
AC
power supply
AC100/110V
AC200/220V
AC230V
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Printed in Japan