TECNO-GAZ S.p.a.

EUROPA B XP AND EUROPA B XP 24L


TECHNICAL MANUAL
DXBT004 REV_16

THIS MANUAL MUST BE INTEGRATED WITH THE INFORMATION OF


THE USER MANUAL

0434
Council Directive 93/42/CEE

INDEX

MAIN COMMANDS

TECHNICAL SETUP

CONTROL MENU

CHECK-UP OF LEVEL PROBES

CHECK AND CLEAN THE VACUUM PUMP

TEST OF THE PERIPHERALS CONNECTED TO CBL606

DIAGRAMS
o PNEUMATIC DIAGRAM
o ELECTRIC DIAGRAM

FUNCTIONAL COMPONENTS

ERROR MESSAGES

CYCLE DIAGRAMS

EXPLODED DRAWING

INDEX

QUADRO COMANDI / MAIN COMMANDS

QUADRO COMANDI / MAIN COMMANDS

TECHNICAL SETUP
In this setup you could do the following:
1.
2.
3.
4.
5.

change the serial number of the unit


change the kind of printer
change temperature and pressure offsets
change the customer number
set the technical transmission

You could enter in this setup with the following procedure:


1. switch-on the unit with the main green button;
2. during the switch-on keep pushed B-START and push at least 5 times B-PUMP until you see on the
display the serial number of the unit;
3. release B-START and B-PUMP.
To go through the various page of the list above you have to use B-START.
1. change the serial number of the unit
B-DOOR selects the number of the serial number that you want to change; the number selected blinks;
B-SELECT increment the number selected;
B-PUMP decrement the number selected.
To SAVE the new serial number keep pushed for at least 10 seconds the 3 button B-DOOR, B-SELECT
and B-PUMP.

TECHNICAL SETUP

2. change the kind of printer


with B-SELECT or B-PUMP you could select the following:

PRINTER 4050
Only for the unit with the internal printer.
4050 is the name of the internal printer.
This is an old model and on the side of the printer you could find the label HTP-4050.
The code of this printer is DNVA061.
The code of the paper for this printer is DNVA063.
The printer is like the one that you see on figure 1.

Figure 1

TECHNICAL SETUP

PRINTER 9050
Only for the unit with the internal printer.
9050 is the name of the internal printer.
This is the new version of the model 4050 and on the printer you could find the label HTPN-9050.
The code of this printer is DXGA061.
The code of the paper for this printer is DNVA063.
The printer is like the one that you see on figure 1.
This setting is compatible with our new Custom printer:
on this printer you could find the label CUSTOM PLUS-M2B.
The code of this printer is DXBA361.
The code of the paper for this printer is DNVA063.
The printer is like the one that you see on figure 2.

Figure 2

TECHNICAL SETUP

EXTERNAL PRINTER (STAMP. ESTERNA)


This setting is used to connect an external printer to the unit.
The external printer is connected to the serial port of the unit.
If the serial port is in the front panel then the connector of this port is round, otherwise is rectangular
like the COM port on PC computer.
The external printer must have the following parameters:
a. BAUD RATE = 600 bps
b. STOP BITS = 2
c. PARITY = NONE
d. BITS PER CHR = 7
Typically we refer to the following model of external printer:
MODEL DP1012 (you could find the printer model on the bottom side of the printer)
In this model you see 4 buttons on the top.
The first button is an empty square, the second is like a @, the third is an 1 and the fourth is a 0.
You could enter in the setup by pressing together the button number 1 and the button number 3.
Here you could advance by pressing the button number 2 and you could change the selected option
by pressing the button number 1.
Then you have to press together the buttons number 1 and the button number 2 to save and exit from
setup.
MODEL DP24S4N (you could find the printer model on the bottom side of the printer)
In this model you have 2 buttons on the top (named PRINT and FEED) and 1 switch on the back.
To enter in the setup push together the 2 buttons on the top and then switch on the printer by the
switch on the back.
Now use the PRINT button to change the selected option and the FEED button to select the next
option.
Exit from setup by switching off the printer with the switch on the back.

NO PRINTER (STAMP. ASSENTE)


Use this setting if you dont have any of the above or if you want to record cycle data only with a PC.

This setting is automatically saved.

TECHNICAL SETUP

3. change temperature and pressure offset


To modify/save the offset of the pressure/temperature sensor (buttons that you need to use) please refer
to the chapter CONTROL MENU.
4. change customer number
This page is used to change a global setting of the unit and it is only for manufacturer setup.
Please dont change this number if you dont know his function.
With the software release 2.56 or newer:
if you have an Europa B xp with standard European cycles then ensure that this number is 10;
if you have an Europa B xp 24 with standard European cycles then ensure that this number is 11;
otherwise ask to the manufacturer.
This setting is automatically saved.
5. set the technical transmission
On the LCD you see Tx TECNICO on the first line;
in the second line, with B-SELECT and B-PUMP, you could select the following:

PRESENTE
The electronic board will send on the serial port of the unit status information and values of T1, T2 and
P1 every second when the unit is in the standard user mode.
You need a special software to acquire this information.

ASSENTE
The electronic board will send on the serial port the information that are normally printed during a
cycle in a codified format.
You need a special software to acquire this information.

TECHNICAL SETUP

CONTROL MENU
Attention: in this device it is not recommended to do the temperature probes calibration if the device is younger
than 1 years old.
This device has 2 temperature probes into the chamber (TEMP1 and TEMP2) and 1 pressure transducer
(PRESSURE1).
For the temperature probes calibration in this autoclave it is not necessary to work on the electronic board.
NOTE: it is not possible to adjust the probe offset when the display shows OPEN DOOR.
Instead it is possible to do that in every phase of sterilization cycle.

CONTROL MENU

For temperature probes offset regulation it is necessary go into the control menu:
1. Push and hold B-DOOR
2. Push and release at least 5 time B-SELECT until the information showed on display change (you can
see for example the picture above)
3. Release B-DOOR
If in the second line from the bottom the display shows T1 then you are in the control menu (you can see for
example the picture above)
At this point with B-SELECT and B-PUMP is possible to increase and decrease the T1 offset; the offset value
is showed on the last line of display.
T1 is the temperature probe called TEMP1 in the list of functional components.
It is possible to save the value set by keeping pushed B-DOOR and B-PUMP at the same time until you listen
the acoustic signal.
It is possible to switch between T2 and P1 with B-DOOR.
T2 is the second temperature probe and it is called TEMP2 in the list of the functional components.
P1 is the pressure transducer and it is called PRESSURE1 in the list of the functional components.
If you want to exit from the menu you have to push and hold pushed B-DOOR.

CONTROL MENU

CHECK-UP OF LEVEL PROBES


The level probes are two, one on the clean water tank (TANK1) that we will call LEVEL1 and the other one in
the used water tank (TANK2) that we will call LEVEL2:
LEVEL1 this shows the minimum and the maximum level of the clean water tank
LEVEL2 this shows the maximum filling of the used water tank
To check if the probes work correctly its necessary to remove the cover of the tank and, with the autoclave
turned on, you have to flaw the craft on the auction: then, looking at the front panel you have to verify, that the
signal of the level is showed. Clean the crafts before reassembling the cover of the tank.
Acting on the craft of LEVEL1 they will be had to turn on the LED indicative following: LEVEL MIN and LEVEL
MAX, respectively when the craft is found down at the auction on the probe of level and the summit.
Acting on the craft of the LEVEL2 will be had to turn on the LED indicator WASTE LEVEL when the craft is
found to the summit of the auction probe.
For color wires connection of LEVEL1 please look at the picture in the next page.

CHECK-UP OF LEVEL PROBES

WHITE

BROWN

ORANGE

GREEN

VIOLET

BLACK

CHECK-UP OF LEVEL PROBES

CHECK AND CLEAN THE VACUUM PUMP


CPGM020
The failure vacuum to reach the fixed values compromise the sterilisation cycle. Its important the periodically
check of the vacuum pump, to guarantee the correct operation of the autoclave. Dont make maintenance on
vacuum pump under warranty (1 year). To do maintenance proceed as follow:
1. Unscrew the fixing screw and take out the pump
2. Place the vacuum pump on a good and clean work table
3. Proceed to open the pump unscrewing the four screw of both head (eight screw at all)

4. Lift at the same time the two heads of the pump and separate them: the heads are jointed by a collector
inserted by pressure
5. Open the two heads by unscrewing the two internal screw (four screw at all)
Its important dont invert the two heads when reassembling them

CHECK AND CLEAN THE VACUUM PUMP

6. The cleaning of the diaghpram must be done with non-chemical agent products (eg. Alchool)
Verify the diaghpram condition by view
Its better to replace the seal and the two orings

7. Clean the metal head and the intermediate plate with the same procedure as the diaghpram
Remove completely the impurities to avoid any malfunction

CHECK AND CLEAN THE VACUUM PUMP

8. Place the intermediate plate on the pump (everyone on the original position and direction)
9. Insert the seals into the appropriate seat (the seals are all of the same type)

10. Joint the metal heads with the appropriate collector and slowly place on the pump

11. Tighten the two central screws without forced it


12. Tighten the four external screws by forced it

CHECK AND CLEAN THE VACUUM PUMP

13. Closed at the end the heads by tighten the two central screw, blocked it but not forced, to avoid to damage
the intermediate plates: a high force can broken them (see detail photo)

14. Verify the operating of the pump, by connect it to a pressure transducer (or a manometer) and feeding for
five seconds (no more).
The pump must reach a vacuum degree of -0.96 / -0.98 bar: a correct holding means the right operation
of the pump
In a first time there is lack of pressure of 0.02 / 0.04 bar, afterwards the pressure is aligned with the
holding parameter. In case of a little but constantly lack of pressure, its necessary to repeat the cleaning
and a new control procedure
15. Final assembling of the pump on the autoclave

CHECK AND CLEAN THE VACUUM PUMP

Spare parts for vacuum pump:

CPGM027 - CPGM028

CPGM023

CPGM043

CHECK AND CLEAN THE VACUUM PUMP

CPGM025

CPGM024

CHECK AND CLEAN THE VACUUM PUMP

TEST OF THE PERIPHERALS CONNECTED TO CBL606


With the CBL606 its possible to power up one by one all the peripherals connected to this board without
following the sterilization cycle.
Attention: during the test, for example, if you power up the heating element then this will be always on until you
dont stop it.
We suggest you to execute the test only with the door opened to avoid that the unit can go in
pressure.
In order to execute the test proceed as follows.
Insert on the board CBL606 a JUMPER in J3 position.
The J3 position in the figure below is between the 2 red lines.

TEST PERIFERICHE CONNESSE A CBL606

Switch on the autoclave with main green button.


And use the B-START until you read on the display the string USCITA.
Then use B-PUMP in order to increase the number indicated on the display.
To each number corresponds the activation of one peripheral connected to the board;
You could find the number-peripheral correspondence in the list that follows:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.

HEAT1
not connected
EV7
EV5
EV2
it closes relay "the direction" of BLOCK1 on CBL606
BLOCK1
not connected
it opens relay "the direction" of BLOCK1 on CBL606
BLOCK1
EV4
PUMP1
PUMP2
FAN1
EV3
not connected
EV6
not connected
BUZZER on CBL606
not connected

TEST PERIFERICHE CONNESSE A CBL606

SCHEMI / DIAGRAMS

SCHEMI / DIAGRAMS

SCHEMA PNEUMATICO / PNEUMATIC DIAGRAM

SCHEMA ELETTRICO / ELECTRIC DIAGRAMS

SCHEMA ELETTRICO / ELECTRIC DIAGRAMS

FUNCTIONAL COMPONENTS
Exploded
Name
drawing
53 CBL606B
8 CBL607
83C
83A
81
82
84
83B

37
89
87
76
5
72
71A
71B
73
54
58A
58B
62A
62B
66
67
64
57
40

EV2
EV3
EV4
EV5
EV6
EV7
FILTER1
FILTER2
FILTER3
PRESSURE1
VALVE1
VALVE2
MAINS
P230
FUSE1
FUSE2
PEXT
FUSE3
TANK1
TANK2
LEVEL1
LEVEL2
COOLER1
FAN1
PUMP1
PUMP2
VESSEL1

27 BLOCK1
39 SWITCH1
29 SWITCH2
43
41A
41B
42
75D
75C
75A
75B
34A
34B
34C
74
4
6

HEAT1
TEMP1
TEMP2
TEMP3
OUT-1
OUT-2
OUT-3
OUT-4
OUT-A
IN-B
IN-C
RS232
PANEL1
PRINTER1

FUNCTIONAL COMPONENTS

Description
Power board (main electronic board)
Display board (electronic board to control the display, the main
commands and to read water levels)
Solenoid valve typically close (for connections 1/4 gas)
Solenoid valve typically close (for connections 1/4 gas)
Solenoid valve typically close (for connections 1/8 gas)
Solenoid valve typically open (for connections 1/4 gas)
3 way solenoid valve (for connections 1/4 gas)
Solenoid valve typically closed (for connections 1/4 gas)
Brass filter
Plastic filter
Bacteriological filter
Pressure sensor
Non return valve
Safety valve
Mains switch (green)
Mains input 230 Vac
Fuse 12 A
Fuse 12 A
Mains auxiliary output (for deionizer connection)
Fuse 500 mA
Clean water tank
Waste water tank
Level probe for the clean water tank
Level probe for the waste water tank
Cooler
Fan
Load pump
Vacuum pump
Vessel
Door electromagnet (used to advance the pin to block the
handle)
Door open/close switch (push button that is pushed when the
user close the door)
Door electromagnet switch (switch activated when the
electromagnet pin advance into the handle)
Chamber heat
Temperature probe 1
Temperature probe 2
Resetting manual thermostat heater element
Posterior tap to remove water from TANK2
Posterior vent output for TANK2
Posterior tap to remove water from TANK1
Posterior vent output for TANK1
Frontal output to remove water from TANK2
Frontal input to charge clean water with PUMP1 on TANK1
Frontal input to manual charge clean water on TANK1
RS232 port (to connect the computer)
Front panel (see main commands)
Printer

ERROR MESSAGES
E200
The cycle has been started with the LEVEL MIN LED on.
You could have this error at the beginning of the cycle.
1.
2.
3.
4.

add water into TANK1


check LEVEL1
check the electrical connections between LEVEL1 and CBL607
check or replace CBL607

E250
The cycle has been started with the LEVEL MAX LED off and
the autoclave is set with water charge by demineralizer.
You could have this error at the beginning of the cycle.
1. refill TANK1
check the tap in the back of the autoclave
(if the water clean system is connected then the tap must be open)
check if the water is going into the water clean system
check the installation of the water clean system
2. check if there is 230V on PEXT when the LED LEVEL_MAX is off (if there is no voltage then the problem
could be the board CBL606, FUSE3 or the connection between the 3 object above)
3. check LEVEL1
4. check the electrical connections between LEVEL1 and CBL607
5. check or replace the board CBL607
E300
The cycle has been started with the WASTE LEVEL LED on.
You could have this error at the beginning of the cycle.
1. remove water from TANK2
2. check LEVEL2
3. check the electrical connections between LEVEL2 and CBL607

ERROR MESSAGES

E401
The cycle has been started with the door open.
You could have this error at the beginning of the cycle.
Close the door and restart the cycle.
If the door isn't open then SWITCH1 doesnt feel the handle of the door.
1. check SWITCH1
2. check the electrical connections between SWITCH1 and CBL606:
the problem could be a false contact on the cable connector for the electronic board
3. check the status of the cables
4. check the position of SWITCH1:
you could have problem if the switch is too much within the metallic support panel

E451
The cycle has been started with the door closed but the door is not blocked.
You could have this error at the beginning of the cycle.
1. check BLOCK1:
refer to chapter test of the peripherals connected to CBL606
2. check the electrical connections between BLOCK1 and CBL606
3. check SWITCH2:
its possible that SWITCH2 hasn't told to CBL606 that
the pin of the electromagnet BLOCK1 has blocked the handle
4. check the electrical connections between SWITCH2 and CBL606:
the problem could be a false contact on the cable connector for the electronic board
5. its possible that the pin of the electromagnet BLOCK1 couldn't go into the handle
due to a non correct alignment between the handle hole and the BLOCK1 pin:
regulate the door as showed in the user manual:
if the door is hard to close then the handle heel keep in distance from the metallic panel
check the position of the electromagnet BLOCK1 on its support
check the position of the door arm
6. check the position of SWITCH2:
its possible that the crank of SWITCH2 hasn't heard the movement of BLOCK1 pin
due to a non correct alignment between the pin and the crank

E461
The vacuum test has been started with the temperature in the chamber greater than 40C.
You could have this error at the beginning of the cycle.
Wait that temperature in the chamber fall under 40C.

ERROR MESSAGES

A001
The user has stopped the cycle (by pushing the button B-START).
If this alarm comes out every time you start a cycle
(after about 10 seconds from the beginning of the cycle)
then its possible that you have a problem on the front panel buttons.
A011
CBL606 doesn't communicate with CBL607.
Check the connections between the 2 electronic boards:
the problem could be the cable that connects the 2 board;
in other case we have observed that the problem is the electronic display board CBL607.
A101
The unit hasn't reached the prefixed vacuum.
You could have this alarm during T0, T2, T4 or T10.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

clean or replace the gasket of the door


clean the external border of the chamber
adjust the door as explained in the user manual: the handle may close too soft
if you find a lot of water in the chamber after the alarm then the problem could be on EV4
check EV6 that must open and close alternatively while PUMP2 is working
check EV2 and EV7
check if EV5 is closed:
you could put a finger on FILTER3 and look if it sucks air
(you could feel the air flow only if the leakage is not very small)
check if COOLER1 has a small hole
check the pipes, the connections and the links near the chamber
check if the stoppers/nuts over the chamber are well closed
check the silicon gaskets on TEMP1 / TEMP2
clean PUMP2

Typically each vacuum phase is about 4 or 5 minutes.


If this phase takes more times then you could have one of the problem above.

ERROR MESSAGES

A111 / A121
Loose of pressure during the vacuum test.
You could have the alarm A111 during T11 and the alarm A121 during T12.
1.
2.
3.
4.
5.
6.

clean or replace the gasket of the door


clean the external border of the chamber
adjust the door as explained in the user manual: the handle may close too soft
if you find a lot of water in the chamber after the alarm then the problem could be on EV4
check EV2 and EV3
check if EV5 is closed:
you could put a finger on FILTER3 and look if it sucks air
(you could feel the air flow only if the leakage is not very small)
7. check the pipes, the connections and the links near the chamber
8. check if the stoppers/nuts over the chamber are well closed
9. check the silicon gaskets on TEMP1 / TEMP2
It could be useful to find small hole or broken connection that make fail the vacuum phase with some special
products like the one that you see in the picture below:

During the cycle you can put a lot of this product on a suspect link or part: if you have a small hole you will see
that the foam will be sucked (during the vacuum phase) or sprayed (if the chamber is in pressure).

ERROR MESSAGES

A401
SWITCH1 has indicated to the electronic board that the door has been opened during the cycle or
SWITCH2 has indicated that the pin of the electromagnet has been removed from the handle.
You could have this alarm during the whole cycle.
1. check SWITCH1 and SWITCH2
2. check the electrical connections between SWITCH1, SWITCH2 and CBL606:
the problem could be a false contact on the cable connector for the electronic board
3. check the status of the cables
4. check the alignment of SWITCH2 on the electromagnet BLOCK1:
the switch could be in a critical position
5. check the position of SWITCH1:
you could have problem if the switch is too much within the metallic support panel
A501
Loose of pressure during the sterilization phase.
You could have this alarm during T6.
If you have leakage during the sterilization then the pressure decrease and
if the pressure goes under about 1.8 Bar during a 134 cycle or under 0.8 Bar during a 121 cycle then you have
this alarm.
Note that all the solenoid valve are closed during the sterilization so
if a solenoid valve doesnt close well then you could have a loose of pressure.
Also check the gasket of the door,
the links and the connections on the chamber and
the gasket of the temperature probes.
Check also the quantity of water that goes into the chamber (as explained in the alarm A651).
This alarm message could be related to A651, A661 and A751.
A551 / A552
There is a problem with TEMP1 / TEMP2.
The problem could be one of the temperature probe or the electronic board.
TEMP1 is connected in the position marked with T1 on the board CBL606 and
TEMP2 is connected in the position marked with T2.
Change the position of the 2 probe connectors on the board CBL606,
reset the alarm and look if it change: from A551 to A552 or from A552 to A551.
If the alarm doesnt change then the problem is the board CBL606.
Otherwise the problem is the temperature probe.
Check also the status of the probe connectors / cables.

ERROR MESSAGES

A641
High pressure in the chamber.
You could have this alarm during T5 or T6.
If the pressure in the chamber, during the preheating phase or the sterilization phase, is over about 2.4Bar
then the unit will show this alarm.
This problem could be related to the status of EV3 and EV6 or
to a problem on the electronic board CBL606.
1.
2.
3.
4.

check the electrical connections between EV3, EV6 and CBL606


clean or replace EV3 and EV6
replace the electronic board CBL606
check the status of PRESSURE1

A651
The temperature in the chamber is over the maximum value for the selected cycle.
You could have this alarm during T1, T3, T5 or T6.
The unit shows this alarm code
during the sterilization phase when the temperature read by TEMP1 or TEMP2
goes over 125C for the cycle 121 or goes over 138C for the cycle 134 or
during the preheating phases when the temperature
goes over 110C and the pressure is less than 0.3 Bar.
Step 1
If the pressure rises quickly after the vacuum phase then
check the status of the solenoid valves connected to the chamber: EV2 and EV3.
During the preheating phase EV7 is open to discharge the water in the COOLER1,
so if you have a problem on EV2 or EV3 that remains open for some reasons
then water or steam will escape from the chamber and youll have the overheating alarm.
In this case the problem could be a solenoid valve, the electronic board or a cable.
Step 2
If you have this alarm during the sterilization phase then run a cycle and look the values of TEMP1 and
TEMP2 in the control menu: they must be quite closed near the sterilization phase and during this phase;
if they arent then try to adjust the offset of TEMP1 or the offset of TEMP2 as explained in the example below:
Example 1
If you run a cycle 134 and you find in the control menu the following value:
T1 = 134.2C, T2 = 135.0C and P1 = 2.1 Bar
then increase the offset of TEMP1 of about 0.8C
Example 2
If you run a cycle 134 and you find in the control menu the following value:
T1 = 135.0C, T2 = 134.1C and P1 = 2.1 Bar
then increase the offset of TEMP2 of about 0.9C

ERROR MESSAGES

Example 3
If you run a cycle 134 and you find in the control menu the following value:
T1 = 135.0C, T2 = 134.1C and P1 = 2.0 Bar
then decrease the offset of TEMP1 of about 0.9C
Note: typically the value of PRESSURE1 is correct and doesnt need any adjustment.
To understand the example above you have to refer to the following temperature-pressure diagrams:
125

124

123

122

121

120
0,9

1,1

1,2

1,3

1,4

2,2

2,3

2,4

Bar

138

137

136

135

134

133
1,9

2,1
Bar

ERROR MESSAGES

Step 3
The cause could be also that the right quantity of water after the vacuum phase isnt came into the chamber;
for example by a closed pipe from TANK1 to the chamber trough EV4.
1. replace TANK1 with a graduated tank and
measure the quantity of water that goes into the chamber;
if the chamber is 24 liters then you have to measure:
about 0.35 liters after the first 2 vacuum phases and
about 0.60 liters after the last vacuum phase;
otherwise you have to measure:
about 0.20 liters after the first 2 vacuum phases and
about 0.45 liters after the last vacuum phase
2. if the quantities above are correct then
check the hole in TANK1 that could be closed by some impurities
check the pipe from TANK1 to FILTER2 that could be closed
remove the water in TANK1 by OUT-3 until turn-on the LED of the minimum clean water level;
remove all the remaining water in TANK1 by OUT-3 and put it in a graduated can;
if the water is less than 0.8 liters then you have to replace TANK1
3. if they arent correct then
check and clean EV4
check FILTER2
check the filter in the chamber
check the connection and the link from EV4 to the chamber
ask to the manufacturer if you need an upgraded microcontroller
Step 4
This alarm could be also caused by a small loose of steam during the cycle:
try to adjust the door as indicated in the user manual
(if the door closes too softly then steam could came out)
check and clean the solenoid valve connected to the chamber: EV2, EV3 and EV4
check the connections and the links near the chamber
(they could have a small hole, check it with a small mirror)
check the gasket of TEMP1 and TEMP2
(during the cycle put a small mirror in the back of the chamber near TEMP1 and TEMP2)

ERROR MESSAGES

Step 5
This alarm could be also caused by a problem on temperature reading:
replace the electronic board CBL606
replace TEMP1 and TEMP2
If you replace the temperature probes make attention to the silicon gasket Figure 1.
Every time you remove a temperature probe,
remove also the old silicon gasket in the chamber and replace it.
When you mount the new probe dont close too much the screw in Figure 1
because you could damage the silicon gasket:
put 2 fingers on the probe (from the internal side of the chamber) and
try to push and pull the probe, if its fixed then you have closed enough the screw.

Figure 1

Figure 2

Figure 3

On the Figure 4 you see the silicon gasket that comes out from the hole:
this is ok and its another signal that the probe has been well fixed.

Figure 4

ERROR MESSAGES

Step 6
This alarm could be also caused by an arrangement phenomenon of the heating element
over the chamber that will generate a non correct conduction of the heat:
cut the top of the insulation band in 4 points
(near the screws that fix the heating band) and screw them more

check the position of the heating band:


if it is over the weddings of the chamber top link then
unscrew the fixing screws and move the heating band outside
replace the heating band

A661
The temperature probes show different values during the sterilization phase:
the difference between TEMP1 and TEMP2 is more than 2C.
You could have this alarm during T6.
This case is like A651 (follow the step 2, 3, 4, 5 and 6)
A701
The pressure in the chamber doesnt rise correctly.
You could have this alarm during T1, T3, T5.
1. check the thermo switch of the heating band:
this element could have opened by an over temperature problem
2. check if you have 230V on the contacts for the heating band in the CBL606
during the preheating phase: if not then replace the CBL606
3. the heating band could be broken: replace it

ERROR MESSAGES

10

ATTENTION
The A701 is shown when the unit take too much time to make the preheating phase, for example because the
heating band isnt working.
If the heating band isnt working then could be damaged the electronic board so the switch on the board
doesnt close (in this case you have to replace the CBL606).
Also if the heating band isnt working then its possible that in a previous cycle the unit is gone in over
temperature and the thermoswitch has opened.
The unit could be gone in over temperature for some different case:
- no water in the chamber after the vacuum phase;
- water is escaped during the preheating (for example an EV doesnt close)
- the switch on the electronic board that control the heating band is broken and the heating band is always
on
So, when you have A701, its necessary also to keep in note the information above.
A751
The temperature in the chamber is under the minimum value for the selected cycle.
You could have this alarm during T6.
The unit shows this alarm code
during the sterilization phase when the temperature read by TEMP1 or TEMP2
goes under 121C for the cycle 121 or goes over 134C for the cycle 134.
This case is like A651 (follow the step 2, 3, 4, 5 and 6)
A761
The pressure sensor doesnt work or there are problems with the connections
1. check the connections between PRESSURE1 and CBL606
2. replace PRESSURE1
3. replace CBL606
A801
The pressure in the chamber doesnt fall correctly
after the preheating phases or after the sterilization phase.
You could have this alarm when the unit discharges the pressure in the chamber.
1. check and clean EV2 and EV7:
it's possible that some impurity has gone into the hydraulic circuit
so one of the 2 solenoid valve above could be blocked
2. check the cable and the connection between EV2, EV7 and CBL606
3. clean FILTER1
4. check and clean the filter in the chamber:
its possible that something blocks the flow of water / steam
in the bottom hole / filter of the chamber
5. replace CBL606:
its possible that the electronic doesnt send the command to open the 2 valve above

ERROR MESSAGES

11

ATTENTION
If the unit isnt running a cycle and the pressure is increasing (for example because there is a lot of steam in
the chamber and the unit is hot) then, over a certain level of pressure, the electronic board open the valve EV2
and EV7 to discharge the pressure. If the pressure doesnt fall then the unit will show A801.
A901
Power has been removed from the unit during a cycle.
Its possible that you have a problem on the power line or
someone has disconnected the power cable or switched-off the unit during a cycle.
1. reset the alarm and restart the cycle
2. check the status of the cables and the connections
3. check the status of the main green button

ERROR MESSAGES

12

CYCLE DIAGRAMS

PLOT 1

Pressure diagram when we push the key B-START for a cycle of 121C or 134C

PLOT 2

Pressure diagram when we push the key B-START for a vacuum test

CYCLE DIAGRAMS

DISEGNO ESPLOSO / EXPLODED DRAWING


EUROPA B XP

DISEGNO ESPLOSO / EXPLODED DRAWING

Rev_0

DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW

Rev_1

DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW

Rev_3

DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW

Rev_1
15/06/2005
Sostituiti raccordi per carico e scarico acqua
Cambia anche il pannello frontale in quanto cambiano i fori di fissaggio
Replaced link for charge and discharge water
Replaced also the front panel because its changed the position of the hole for the link above
Rev_3
20/03/2007
Sostituito raccordo per carico acqua
Eliminato:
n1 DANA111
n1 CM20011
n1 CM05028
Replaced link for water charging.
Deleted:
n1 DANA111
n1 CM20011
n1 CM05028

DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW

Rev_0

DISEGNO ESPLOSO PARTE 1 / EXPLODED DRAWING PART 1

Rev_1

DISEGNO ESPLOSO PARTE 1 / EXPLODED DRAWING PART 1

Rev_1
06/06/06
Sostituiti DHYA035, DXBA332 Rev_10 con DXBA335+CEST012 (TEMP3) e DXBA332 Rev_11
Replaced DHYA035 , DXBA332 Rev_10 with DXBA335+CEST012 and DXBA332 Rev_11

DISEGNO ESPLOSO PARTE 1 / EXPLODED DRAWING PART 1

Rev_0

DISEGNO ESPLOSO PARTE 2 / EXPLODED DRAWING PART 2

Rev_1

DISEGNO ESPLOSO PARTE 2 / EXPLODED DRAWING PART 2

Rev_1
25/07/2005
Sostituite sonde CESL006 con CESL009
Replaced level probe CESL006 with CESL009

DISEGNO ESPLOSO PARTE 2 / EXPLODED DRAWING PART 2

Rev_0

DISEGNO ESPLOSO PARTE 3 / EXPLODED DRAWING PART 3

Rev_0

DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4

Rev_1

DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4

Rev_2

DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4

Rev_1
15/06/2005
Sostituiti raccordi in plastica CPRP003 con i raccordi in metallo DANA111
Sostituito pannello frontale DXBA307 con il DXBA304
Replaced plastic link CPRP003 with metallic connection DANA111
Replaced front panel DXBA307 with DXBA304
Rev_2
20/03/2007
Sostituito raccordo per carico acqua
Eliminato:
n1 DANA111
n1 CM20011
n1 CM05028
Replaced link for water charging.
Deleted:
n1 DANA111
n1 CM20011
n1 CM05028

DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4

DISEGNO ESPLOSO / EXPLODED DRAWING


EUROPA B XP 24L

DISEGNO ESPLOSO / EXPLODED DRAWING

Rev_0
DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW

Rev_1
DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW

Rev_2
DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW

Rev_1
08/06/06
Cambiate viti di fissaggio DXBA326 e DXBA323
Changed screw to fix the DXBA326 and the DXBA323

Rev_2
20/03/2007
Sostituito raccordo per carico acqua
Eliminato:
n1 DANA111
n1 CM20011
n1 CM05028
Replaced link for water charging.
Deleted:
n1 DANA111
n1 CM20011
n1 CM05028

DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW

Rev_0
DISEGNO ESPLOSO PARTE 1 / EXPLODED DRAWING PART 1

Rev_0

DISEGNO ESPLOSO PARTE 2 / EXPLODED DRAWING PART 2

Rev_0
DISEGNO ESPLOSO PARTE 3 / EXPLODED DRAWING PART 3

Rev_0

DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4

Rev_1

DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4

Rev_2
DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4

Rev_1
08/06/06
Cambiate viti di fissaggio DXBA326 e DXBA323
Changed screw to fix the DXBA326 and the DXBA323

Rev_2
20/03/2007
Sostituito raccordo per carico acqua
Eliminato:
n1 DANA111
n1 CM20011
n1 CM05028
Replaced link for water charging.
Deleted:
n1 DANA111
n1 CM20011
n1 CM05028

DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4