Service Manual Europa B
Service Manual Europa B
Service Manual Europa B
0434
Council Directive 93/42/CEE
INDEX
MAIN COMMANDS
TECHNICAL SETUP
CONTROL MENU
DIAGRAMS
o PNEUMATIC DIAGRAM
o ELECTRIC DIAGRAM
FUNCTIONAL COMPONENTS
ERROR MESSAGES
CYCLE DIAGRAMS
EXPLODED DRAWING
INDEX
TECHNICAL SETUP
In this setup you could do the following:
1.
2.
3.
4.
5.
TECHNICAL SETUP
PRINTER 4050
Only for the unit with the internal printer.
4050 is the name of the internal printer.
This is an old model and on the side of the printer you could find the label HTP-4050.
The code of this printer is DNVA061.
The code of the paper for this printer is DNVA063.
The printer is like the one that you see on figure 1.
Figure 1
TECHNICAL SETUP
PRINTER 9050
Only for the unit with the internal printer.
9050 is the name of the internal printer.
This is the new version of the model 4050 and on the printer you could find the label HTPN-9050.
The code of this printer is DXGA061.
The code of the paper for this printer is DNVA063.
The printer is like the one that you see on figure 1.
This setting is compatible with our new Custom printer:
on this printer you could find the label CUSTOM PLUS-M2B.
The code of this printer is DXBA361.
The code of the paper for this printer is DNVA063.
The printer is like the one that you see on figure 2.
Figure 2
TECHNICAL SETUP
TECHNICAL SETUP
PRESENTE
The electronic board will send on the serial port of the unit status information and values of T1, T2 and
P1 every second when the unit is in the standard user mode.
You need a special software to acquire this information.
ASSENTE
The electronic board will send on the serial port the information that are normally printed during a
cycle in a codified format.
You need a special software to acquire this information.
TECHNICAL SETUP
CONTROL MENU
Attention: in this device it is not recommended to do the temperature probes calibration if the device is younger
than 1 years old.
This device has 2 temperature probes into the chamber (TEMP1 and TEMP2) and 1 pressure transducer
(PRESSURE1).
For the temperature probes calibration in this autoclave it is not necessary to work on the electronic board.
NOTE: it is not possible to adjust the probe offset when the display shows OPEN DOOR.
Instead it is possible to do that in every phase of sterilization cycle.
CONTROL MENU
For temperature probes offset regulation it is necessary go into the control menu:
1. Push and hold B-DOOR
2. Push and release at least 5 time B-SELECT until the information showed on display change (you can
see for example the picture above)
3. Release B-DOOR
If in the second line from the bottom the display shows T1 then you are in the control menu (you can see for
example the picture above)
At this point with B-SELECT and B-PUMP is possible to increase and decrease the T1 offset; the offset value
is showed on the last line of display.
T1 is the temperature probe called TEMP1 in the list of functional components.
It is possible to save the value set by keeping pushed B-DOOR and B-PUMP at the same time until you listen
the acoustic signal.
It is possible to switch between T2 and P1 with B-DOOR.
T2 is the second temperature probe and it is called TEMP2 in the list of the functional components.
P1 is the pressure transducer and it is called PRESSURE1 in the list of the functional components.
If you want to exit from the menu you have to push and hold pushed B-DOOR.
CONTROL MENU
WHITE
BROWN
ORANGE
GREEN
VIOLET
BLACK
4. Lift at the same time the two heads of the pump and separate them: the heads are jointed by a collector
inserted by pressure
5. Open the two heads by unscrewing the two internal screw (four screw at all)
Its important dont invert the two heads when reassembling them
6. The cleaning of the diaghpram must be done with non-chemical agent products (eg. Alchool)
Verify the diaghpram condition by view
Its better to replace the seal and the two orings
7. Clean the metal head and the intermediate plate with the same procedure as the diaghpram
Remove completely the impurities to avoid any malfunction
8. Place the intermediate plate on the pump (everyone on the original position and direction)
9. Insert the seals into the appropriate seat (the seals are all of the same type)
10. Joint the metal heads with the appropriate collector and slowly place on the pump
13. Closed at the end the heads by tighten the two central screw, blocked it but not forced, to avoid to damage
the intermediate plates: a high force can broken them (see detail photo)
14. Verify the operating of the pump, by connect it to a pressure transducer (or a manometer) and feeding for
five seconds (no more).
The pump must reach a vacuum degree of -0.96 / -0.98 bar: a correct holding means the right operation
of the pump
In a first time there is lack of pressure of 0.02 / 0.04 bar, afterwards the pressure is aligned with the
holding parameter. In case of a little but constantly lack of pressure, its necessary to repeat the cleaning
and a new control procedure
15. Final assembling of the pump on the autoclave
CPGM027 - CPGM028
CPGM023
CPGM043
CPGM025
CPGM024
HEAT1
not connected
EV7
EV5
EV2
it closes relay "the direction" of BLOCK1 on CBL606
BLOCK1
not connected
it opens relay "the direction" of BLOCK1 on CBL606
BLOCK1
EV4
PUMP1
PUMP2
FAN1
EV3
not connected
EV6
not connected
BUZZER on CBL606
not connected
SCHEMI / DIAGRAMS
SCHEMI / DIAGRAMS
FUNCTIONAL COMPONENTS
Exploded
Name
drawing
53 CBL606B
8 CBL607
83C
83A
81
82
84
83B
37
89
87
76
5
72
71A
71B
73
54
58A
58B
62A
62B
66
67
64
57
40
EV2
EV3
EV4
EV5
EV6
EV7
FILTER1
FILTER2
FILTER3
PRESSURE1
VALVE1
VALVE2
MAINS
P230
FUSE1
FUSE2
PEXT
FUSE3
TANK1
TANK2
LEVEL1
LEVEL2
COOLER1
FAN1
PUMP1
PUMP2
VESSEL1
27 BLOCK1
39 SWITCH1
29 SWITCH2
43
41A
41B
42
75D
75C
75A
75B
34A
34B
34C
74
4
6
HEAT1
TEMP1
TEMP2
TEMP3
OUT-1
OUT-2
OUT-3
OUT-4
OUT-A
IN-B
IN-C
RS232
PANEL1
PRINTER1
FUNCTIONAL COMPONENTS
Description
Power board (main electronic board)
Display board (electronic board to control the display, the main
commands and to read water levels)
Solenoid valve typically close (for connections 1/4 gas)
Solenoid valve typically close (for connections 1/4 gas)
Solenoid valve typically close (for connections 1/8 gas)
Solenoid valve typically open (for connections 1/4 gas)
3 way solenoid valve (for connections 1/4 gas)
Solenoid valve typically closed (for connections 1/4 gas)
Brass filter
Plastic filter
Bacteriological filter
Pressure sensor
Non return valve
Safety valve
Mains switch (green)
Mains input 230 Vac
Fuse 12 A
Fuse 12 A
Mains auxiliary output (for deionizer connection)
Fuse 500 mA
Clean water tank
Waste water tank
Level probe for the clean water tank
Level probe for the waste water tank
Cooler
Fan
Load pump
Vacuum pump
Vessel
Door electromagnet (used to advance the pin to block the
handle)
Door open/close switch (push button that is pushed when the
user close the door)
Door electromagnet switch (switch activated when the
electromagnet pin advance into the handle)
Chamber heat
Temperature probe 1
Temperature probe 2
Resetting manual thermostat heater element
Posterior tap to remove water from TANK2
Posterior vent output for TANK2
Posterior tap to remove water from TANK1
Posterior vent output for TANK1
Frontal output to remove water from TANK2
Frontal input to charge clean water with PUMP1 on TANK1
Frontal input to manual charge clean water on TANK1
RS232 port (to connect the computer)
Front panel (see main commands)
Printer
ERROR MESSAGES
E200
The cycle has been started with the LEVEL MIN LED on.
You could have this error at the beginning of the cycle.
1.
2.
3.
4.
E250
The cycle has been started with the LEVEL MAX LED off and
the autoclave is set with water charge by demineralizer.
You could have this error at the beginning of the cycle.
1. refill TANK1
check the tap in the back of the autoclave
(if the water clean system is connected then the tap must be open)
check if the water is going into the water clean system
check the installation of the water clean system
2. check if there is 230V on PEXT when the LED LEVEL_MAX is off (if there is no voltage then the problem
could be the board CBL606, FUSE3 or the connection between the 3 object above)
3. check LEVEL1
4. check the electrical connections between LEVEL1 and CBL607
5. check or replace the board CBL607
E300
The cycle has been started with the WASTE LEVEL LED on.
You could have this error at the beginning of the cycle.
1. remove water from TANK2
2. check LEVEL2
3. check the electrical connections between LEVEL2 and CBL607
ERROR MESSAGES
E401
The cycle has been started with the door open.
You could have this error at the beginning of the cycle.
Close the door and restart the cycle.
If the door isn't open then SWITCH1 doesnt feel the handle of the door.
1. check SWITCH1
2. check the electrical connections between SWITCH1 and CBL606:
the problem could be a false contact on the cable connector for the electronic board
3. check the status of the cables
4. check the position of SWITCH1:
you could have problem if the switch is too much within the metallic support panel
E451
The cycle has been started with the door closed but the door is not blocked.
You could have this error at the beginning of the cycle.
1. check BLOCK1:
refer to chapter test of the peripherals connected to CBL606
2. check the electrical connections between BLOCK1 and CBL606
3. check SWITCH2:
its possible that SWITCH2 hasn't told to CBL606 that
the pin of the electromagnet BLOCK1 has blocked the handle
4. check the electrical connections between SWITCH2 and CBL606:
the problem could be a false contact on the cable connector for the electronic board
5. its possible that the pin of the electromagnet BLOCK1 couldn't go into the handle
due to a non correct alignment between the handle hole and the BLOCK1 pin:
regulate the door as showed in the user manual:
if the door is hard to close then the handle heel keep in distance from the metallic panel
check the position of the electromagnet BLOCK1 on its support
check the position of the door arm
6. check the position of SWITCH2:
its possible that the crank of SWITCH2 hasn't heard the movement of BLOCK1 pin
due to a non correct alignment between the pin and the crank
E461
The vacuum test has been started with the temperature in the chamber greater than 40C.
You could have this error at the beginning of the cycle.
Wait that temperature in the chamber fall under 40C.
ERROR MESSAGES
A001
The user has stopped the cycle (by pushing the button B-START).
If this alarm comes out every time you start a cycle
(after about 10 seconds from the beginning of the cycle)
then its possible that you have a problem on the front panel buttons.
A011
CBL606 doesn't communicate with CBL607.
Check the connections between the 2 electronic boards:
the problem could be the cable that connects the 2 board;
in other case we have observed that the problem is the electronic display board CBL607.
A101
The unit hasn't reached the prefixed vacuum.
You could have this alarm during T0, T2, T4 or T10.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
ERROR MESSAGES
A111 / A121
Loose of pressure during the vacuum test.
You could have the alarm A111 during T11 and the alarm A121 during T12.
1.
2.
3.
4.
5.
6.
During the cycle you can put a lot of this product on a suspect link or part: if you have a small hole you will see
that the foam will be sucked (during the vacuum phase) or sprayed (if the chamber is in pressure).
ERROR MESSAGES
A401
SWITCH1 has indicated to the electronic board that the door has been opened during the cycle or
SWITCH2 has indicated that the pin of the electromagnet has been removed from the handle.
You could have this alarm during the whole cycle.
1. check SWITCH1 and SWITCH2
2. check the electrical connections between SWITCH1, SWITCH2 and CBL606:
the problem could be a false contact on the cable connector for the electronic board
3. check the status of the cables
4. check the alignment of SWITCH2 on the electromagnet BLOCK1:
the switch could be in a critical position
5. check the position of SWITCH1:
you could have problem if the switch is too much within the metallic support panel
A501
Loose of pressure during the sterilization phase.
You could have this alarm during T6.
If you have leakage during the sterilization then the pressure decrease and
if the pressure goes under about 1.8 Bar during a 134 cycle or under 0.8 Bar during a 121 cycle then you have
this alarm.
Note that all the solenoid valve are closed during the sterilization so
if a solenoid valve doesnt close well then you could have a loose of pressure.
Also check the gasket of the door,
the links and the connections on the chamber and
the gasket of the temperature probes.
Check also the quantity of water that goes into the chamber (as explained in the alarm A651).
This alarm message could be related to A651, A661 and A751.
A551 / A552
There is a problem with TEMP1 / TEMP2.
The problem could be one of the temperature probe or the electronic board.
TEMP1 is connected in the position marked with T1 on the board CBL606 and
TEMP2 is connected in the position marked with T2.
Change the position of the 2 probe connectors on the board CBL606,
reset the alarm and look if it change: from A551 to A552 or from A552 to A551.
If the alarm doesnt change then the problem is the board CBL606.
Otherwise the problem is the temperature probe.
Check also the status of the probe connectors / cables.
ERROR MESSAGES
A641
High pressure in the chamber.
You could have this alarm during T5 or T6.
If the pressure in the chamber, during the preheating phase or the sterilization phase, is over about 2.4Bar
then the unit will show this alarm.
This problem could be related to the status of EV3 and EV6 or
to a problem on the electronic board CBL606.
1.
2.
3.
4.
A651
The temperature in the chamber is over the maximum value for the selected cycle.
You could have this alarm during T1, T3, T5 or T6.
The unit shows this alarm code
during the sterilization phase when the temperature read by TEMP1 or TEMP2
goes over 125C for the cycle 121 or goes over 138C for the cycle 134 or
during the preheating phases when the temperature
goes over 110C and the pressure is less than 0.3 Bar.
Step 1
If the pressure rises quickly after the vacuum phase then
check the status of the solenoid valves connected to the chamber: EV2 and EV3.
During the preheating phase EV7 is open to discharge the water in the COOLER1,
so if you have a problem on EV2 or EV3 that remains open for some reasons
then water or steam will escape from the chamber and youll have the overheating alarm.
In this case the problem could be a solenoid valve, the electronic board or a cable.
Step 2
If you have this alarm during the sterilization phase then run a cycle and look the values of TEMP1 and
TEMP2 in the control menu: they must be quite closed near the sterilization phase and during this phase;
if they arent then try to adjust the offset of TEMP1 or the offset of TEMP2 as explained in the example below:
Example 1
If you run a cycle 134 and you find in the control menu the following value:
T1 = 134.2C, T2 = 135.0C and P1 = 2.1 Bar
then increase the offset of TEMP1 of about 0.8C
Example 2
If you run a cycle 134 and you find in the control menu the following value:
T1 = 135.0C, T2 = 134.1C and P1 = 2.1 Bar
then increase the offset of TEMP2 of about 0.9C
ERROR MESSAGES
Example 3
If you run a cycle 134 and you find in the control menu the following value:
T1 = 135.0C, T2 = 134.1C and P1 = 2.0 Bar
then decrease the offset of TEMP1 of about 0.9C
Note: typically the value of PRESSURE1 is correct and doesnt need any adjustment.
To understand the example above you have to refer to the following temperature-pressure diagrams:
125
124
123
122
121
120
0,9
1,1
1,2
1,3
1,4
2,2
2,3
2,4
Bar
138
137
136
135
134
133
1,9
2,1
Bar
ERROR MESSAGES
Step 3
The cause could be also that the right quantity of water after the vacuum phase isnt came into the chamber;
for example by a closed pipe from TANK1 to the chamber trough EV4.
1. replace TANK1 with a graduated tank and
measure the quantity of water that goes into the chamber;
if the chamber is 24 liters then you have to measure:
about 0.35 liters after the first 2 vacuum phases and
about 0.60 liters after the last vacuum phase;
otherwise you have to measure:
about 0.20 liters after the first 2 vacuum phases and
about 0.45 liters after the last vacuum phase
2. if the quantities above are correct then
check the hole in TANK1 that could be closed by some impurities
check the pipe from TANK1 to FILTER2 that could be closed
remove the water in TANK1 by OUT-3 until turn-on the LED of the minimum clean water level;
remove all the remaining water in TANK1 by OUT-3 and put it in a graduated can;
if the water is less than 0.8 liters then you have to replace TANK1
3. if they arent correct then
check and clean EV4
check FILTER2
check the filter in the chamber
check the connection and the link from EV4 to the chamber
ask to the manufacturer if you need an upgraded microcontroller
Step 4
This alarm could be also caused by a small loose of steam during the cycle:
try to adjust the door as indicated in the user manual
(if the door closes too softly then steam could came out)
check and clean the solenoid valve connected to the chamber: EV2, EV3 and EV4
check the connections and the links near the chamber
(they could have a small hole, check it with a small mirror)
check the gasket of TEMP1 and TEMP2
(during the cycle put a small mirror in the back of the chamber near TEMP1 and TEMP2)
ERROR MESSAGES
Step 5
This alarm could be also caused by a problem on temperature reading:
replace the electronic board CBL606
replace TEMP1 and TEMP2
If you replace the temperature probes make attention to the silicon gasket Figure 1.
Every time you remove a temperature probe,
remove also the old silicon gasket in the chamber and replace it.
When you mount the new probe dont close too much the screw in Figure 1
because you could damage the silicon gasket:
put 2 fingers on the probe (from the internal side of the chamber) and
try to push and pull the probe, if its fixed then you have closed enough the screw.
Figure 1
Figure 2
Figure 3
On the Figure 4 you see the silicon gasket that comes out from the hole:
this is ok and its another signal that the probe has been well fixed.
Figure 4
ERROR MESSAGES
Step 6
This alarm could be also caused by an arrangement phenomenon of the heating element
over the chamber that will generate a non correct conduction of the heat:
cut the top of the insulation band in 4 points
(near the screws that fix the heating band) and screw them more
A661
The temperature probes show different values during the sterilization phase:
the difference between TEMP1 and TEMP2 is more than 2C.
You could have this alarm during T6.
This case is like A651 (follow the step 2, 3, 4, 5 and 6)
A701
The pressure in the chamber doesnt rise correctly.
You could have this alarm during T1, T3, T5.
1. check the thermo switch of the heating band:
this element could have opened by an over temperature problem
2. check if you have 230V on the contacts for the heating band in the CBL606
during the preheating phase: if not then replace the CBL606
3. the heating band could be broken: replace it
ERROR MESSAGES
10
ATTENTION
The A701 is shown when the unit take too much time to make the preheating phase, for example because the
heating band isnt working.
If the heating band isnt working then could be damaged the electronic board so the switch on the board
doesnt close (in this case you have to replace the CBL606).
Also if the heating band isnt working then its possible that in a previous cycle the unit is gone in over
temperature and the thermoswitch has opened.
The unit could be gone in over temperature for some different case:
- no water in the chamber after the vacuum phase;
- water is escaped during the preheating (for example an EV doesnt close)
- the switch on the electronic board that control the heating band is broken and the heating band is always
on
So, when you have A701, its necessary also to keep in note the information above.
A751
The temperature in the chamber is under the minimum value for the selected cycle.
You could have this alarm during T6.
The unit shows this alarm code
during the sterilization phase when the temperature read by TEMP1 or TEMP2
goes under 121C for the cycle 121 or goes over 134C for the cycle 134.
This case is like A651 (follow the step 2, 3, 4, 5 and 6)
A761
The pressure sensor doesnt work or there are problems with the connections
1. check the connections between PRESSURE1 and CBL606
2. replace PRESSURE1
3. replace CBL606
A801
The pressure in the chamber doesnt fall correctly
after the preheating phases or after the sterilization phase.
You could have this alarm when the unit discharges the pressure in the chamber.
1. check and clean EV2 and EV7:
it's possible that some impurity has gone into the hydraulic circuit
so one of the 2 solenoid valve above could be blocked
2. check the cable and the connection between EV2, EV7 and CBL606
3. clean FILTER1
4. check and clean the filter in the chamber:
its possible that something blocks the flow of water / steam
in the bottom hole / filter of the chamber
5. replace CBL606:
its possible that the electronic doesnt send the command to open the 2 valve above
ERROR MESSAGES
11
ATTENTION
If the unit isnt running a cycle and the pressure is increasing (for example because there is a lot of steam in
the chamber and the unit is hot) then, over a certain level of pressure, the electronic board open the valve EV2
and EV7 to discharge the pressure. If the pressure doesnt fall then the unit will show A801.
A901
Power has been removed from the unit during a cycle.
Its possible that you have a problem on the power line or
someone has disconnected the power cable or switched-off the unit during a cycle.
1. reset the alarm and restart the cycle
2. check the status of the cables and the connections
3. check the status of the main green button
ERROR MESSAGES
12
CYCLE DIAGRAMS
PLOT 1
Pressure diagram when we push the key B-START for a cycle of 121C or 134C
PLOT 2
Pressure diagram when we push the key B-START for a vacuum test
CYCLE DIAGRAMS
Rev_0
Rev_1
Rev_3
Rev_1
15/06/2005
Sostituiti raccordi per carico e scarico acqua
Cambia anche il pannello frontale in quanto cambiano i fori di fissaggio
Replaced link for charge and discharge water
Replaced also the front panel because its changed the position of the hole for the link above
Rev_3
20/03/2007
Sostituito raccordo per carico acqua
Eliminato:
n1 DANA111
n1 CM20011
n1 CM05028
Replaced link for water charging.
Deleted:
n1 DANA111
n1 CM20011
n1 CM05028
Rev_0
Rev_1
Rev_1
06/06/06
Sostituiti DHYA035, DXBA332 Rev_10 con DXBA335+CEST012 (TEMP3) e DXBA332 Rev_11
Replaced DHYA035 , DXBA332 Rev_10 with DXBA335+CEST012 and DXBA332 Rev_11
Rev_0
Rev_1
Rev_1
25/07/2005
Sostituite sonde CESL006 con CESL009
Replaced level probe CESL006 with CESL009
Rev_0
Rev_0
Rev_1
Rev_2
Rev_1
15/06/2005
Sostituiti raccordi in plastica CPRP003 con i raccordi in metallo DANA111
Sostituito pannello frontale DXBA307 con il DXBA304
Replaced plastic link CPRP003 with metallic connection DANA111
Replaced front panel DXBA307 with DXBA304
Rev_2
20/03/2007
Sostituito raccordo per carico acqua
Eliminato:
n1 DANA111
n1 CM20011
n1 CM05028
Replaced link for water charging.
Deleted:
n1 DANA111
n1 CM20011
n1 CM05028
Rev_0
DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW
Rev_1
DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW
Rev_2
DISEGNO ESPLOSO COMPLESSIVO / EXPLODED DRAWING GLOBAL VIEW
Rev_1
08/06/06
Cambiate viti di fissaggio DXBA326 e DXBA323
Changed screw to fix the DXBA326 and the DXBA323
Rev_2
20/03/2007
Sostituito raccordo per carico acqua
Eliminato:
n1 DANA111
n1 CM20011
n1 CM05028
Replaced link for water charging.
Deleted:
n1 DANA111
n1 CM20011
n1 CM05028
Rev_0
DISEGNO ESPLOSO PARTE 1 / EXPLODED DRAWING PART 1
Rev_0
Rev_0
DISEGNO ESPLOSO PARTE 3 / EXPLODED DRAWING PART 3
Rev_0
Rev_1
Rev_2
DISEGNO ESPLOSO PARTE 4 / EXPLODED DRAWING PART 4
Rev_1
08/06/06
Cambiate viti di fissaggio DXBA326 e DXBA323
Changed screw to fix the DXBA326 and the DXBA323
Rev_2
20/03/2007
Sostituito raccordo per carico acqua
Eliminato:
n1 DANA111
n1 CM20011
n1 CM05028
Replaced link for water charging.
Deleted:
n1 DANA111
n1 CM20011
n1 CM05028