Operating manual
eng
General Information
Spectroquant photometers
Contents
1 Photometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1
1.2
Photometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Photometers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2
Basic Principle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.1 Spectroquant Cell Tests . . . . . . . . . . . . . . . . . . . .
2.1.2 Spectroquant Reagent Tests . . . . . . . . . . . . . . . . .
Notes for Practical Use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.1 Measuring Range . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.2 Influence of pH . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.3 Influence of Temperature . . . . . . . . . . . . . . . . . . . .
2.2.4 Time Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.5 Influence of Foreign Substances. . . . . . . . . . . . . . .
2.2.6 Dosing of Reagents . . . . . . . . . . . . . . . . . . . . . . . .
2.2.7 Shelf-life of the Reagents . . . . . . . . . . . . . . . . . . . .
5
5
6
6
6
7
7
8
8
10
10
10
11
11
12
3 Sample Preparation. . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.1
3.2
3.3
3.4
3.5
3.6
Taking Samples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preliminary Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dilution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Filtration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Homogenization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Decomposition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12
13
13
14
15
15
4 Pipetting System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5 Analytical Quality Assurance (AQA) . . . . . . . . . . . . . . . 18
5.1
5.2
5.3
5.4
18
19
20
20
21
21
22
22
23
Release 06/2014
Spectroquant photometers
1 Photometers
1.1 Photometry
When a beam of light is transmitted through a colored solution, then this
beam loses its intensity, in other words a part of the light is absorbed by
the solution. Depending on the substance in question, this absorption
occurs at specific wavelengths.
Monochromators (e. g. narrow-band interference filters, lattices) are used
to select the wavelength from the total spectrum of a tungsten-halogen
lamp (VIS spectrum), a deuterium lamp (UV spectrum) or, respectively, a
xenon lamp.
The intensity of the absorption can be characterized using the transmittance T (or, respectively, T in percent).
T = I/I 0
I 0 = Initial intensity of the light
I = Intensity of the transmitted light
If the light is not absorbed at all by a solution, then this solution has a
transmittance of 100 %; a complete absorption of the light in the solution
means 0 % transmittance.
The measure generally used for the absorption of light is the absorbance
(A), since this correlates directly with the concentration of the absorbing
substance. The following connection exists between absorbance and
transmittance:
A = log T
Experiments by BOUGUER (16981758) and LAMBERT (17281777)
showed that the absorbance is dependent on the thickness of the absorbing layer of the cell used. The relationship between the absorbance
and the concentration of the analyte in question was discovered by BEER
(18251863). The combination of these two natural laws led to the derivation of Lambert-Beers law, which can be described in the form of the following equation:
A=
Release 06/2014
c d
Spectroquant photometers
1 Photometers
1.2 The Photometers
The photometers that belong to the Spectroquant Analysis System differ
from conventional photometers in the following important aspects:
All cells formats used are automatically identified and the correct measuring range is selected automatically.
The measurement value can be immediately read off from the display
in the desired form.
The method for the test kits (Cell Tests and reagent tests) belonging to
the Spectroquant analysis system is automatically selected via the
scanning of the bar code.
For technical data and instructions for use please refer to the section
Function description or can also be found on the internet.
Release 06/2014
Spectroquant photometers
Leakproof cap
Risk phrases
Special cell in
optical quality
Additional reagent(s)
Certain cell tests, e. g. COD or nitrite, already contain all necessary reagents in the cells, and the sample must merely be added with a pipette.
In other tests, however for reasons of chemical compatibility it is necessary to separate the test into two or three different reagent mixtures. In
such cases, besides the sample a metered reagent must also be added.
Release 06/2014
Spectroquant photometers
Below the specified measuring range, either a different cell or else another
procedure must be used. The lower limit of the measuring range either
takes the form of nonlinearity of the calibration curve, as shown in the
figure, or else is given by the method detection limit. The method detection limit of an analytical method is the lowest concentration of the analyte in question that can be measured quantitatively with a defined degree
of probability (e. g. 99 %).
The upper limit of the measuring range is the point at which the linear
correlation between the concentration and the absorbance ends. In such a
case the sample must be diluted accordingly so that it lies ideally in the
middle of the effective range (least-error measurement).
In photometry it is conventional practice to measure against the reagent
blank value. Here the analysis is carried out blind, i.e. without any analyte added. Instead of the sample volume, the corresponding quantity of
distilled or DI water is used. This reagent blank value is prestored in the
photometers belonging to the Spectroquant analysis system, which
means that - due to the high batch reproducibility - it is possible to dispense with a separate measurement of the reagent blank. At the lower
limit of the measuring range, the accuracy of the determination can be
enhanced by performing the measurement against a separately prepared
reagent blank.
Absorbance
Concentration
In some cases the intensity of the color of the solution and thus the absorbance can drop again when very high concentrations of the analyte
are present (see package insert).
Release 06/2014
Spectroquant photometers
Absorbance
The temperature of the sample solution and the reagents may have an
effect on the color reaction and thus on the measurement result. The typical temperature course is illustrated in the figure.
Attention! After thermic decomposition procedures, the determination of COD or total contents of nitrogen, phosphorus, or metal, a
sufficient waiting time must be allowed for to permit the solution
cool to room temperature.
10
20
30
Temperature (C)
40
30
Reaction time (minutes)
60
The reaction time specified in the working instructions refers to the period
of time from the addition of the last reagent until the actual measurement.
In addition, the package inserts for the individual test kits also state the
time interval in which the measurement value does not change. The maximum time interval is 60 minutes; this time should not be exceeded, even in
the case of stable color reactions.
Release 06/2014
Absorbance
Most of the color reactions require a certain time to reach the maximum
color intensity. The solid curve in the figure at the right gives a schematic
impression of a typical time course. The behavior of relatively instable color reactions with time is shown by the dotted curve.
Spectroquant photometers
A positive-displacement pipette should be used for larger quantities of liquid or for the exact dosage of smaller reagent quantities. In these cases
the reagent bottles are not fitted with a dropper insert.
Solid substances are dosed either with the dose-metering cap or with
microspoons that are integrated into the screw cap of the respective reagent bottle. The dose-metering cap is used for solid reagents or reagent
mixtures that are free-flowing.
In all other cases the substances are dosed with the microspoon.
In this case it is necessary to add only level microspoonfuls. To this end
the spoon must be drawn over the brim of the reagent bottle.
At the first use replace the black screw cap of the reagent bottle by the
dose-metering cap.
Hold the reagent bottle vertically and, at each dosage, press the slide all
the way into the dose-metering cap. Before each dosage ensure that the
slide is completely retracted.
Reclose the reagent bottle with the black screw cap at the
end of the measurement series, since the function of the reagent is impaired by the absorption of atmospheric moisture.
m
10
Release 06/2014
Spectroquant photometers
3 Sample Preparation
Sample preparation covers all the steps necessary before the actual analysis can be performed.
Release 06/2014
Preservation
+2 to +5 C max. 24 h or
18 C max. 14 days
analyze immediately, only in exceptional case
+2 to +5 C max. 6 h
short-term storage, no preservation;
with nitric acid to pH 1, max. 4 weeks
short-term storage, no preservation;
with nitric acid to pH 1, max. 4 weeks
11
Spectroquant photometers
3 Sample Preparation
3.2 Preliminary Tests
Correct measurement results can be obtained only within the measuring
range specified for each individual parameter. When dealing with sample
solutions of an unknown concentration, it is advisable to establish whether
the sample concentration is indeed within the specified measuring range,
ideally roughly in the middle of the range.
Preliminary tests enhance the analytical reliability and make the determination of the necessary dilution ratios in the case of high concentrations
easier. MQuant TM Test Strips are very well suited for preliminary tests.
3.3 Dilution
Dilution of samples is necessary for two reasons:
The following auxiliaries are absolute prerequisites for the dilution of the
sample:
Volumetric flasks of varying sizes (e.g. 50, 100 and 200 ml)
Positive-displacement pipette
Distilled or DI water.
Only dilutions carried out with these auxiliary products are of sufficient reliability in the area of trace analysis, to which photometry belongs (for the
simplified procedure see page 14).
An important aspect here is that once the volumetric flask has been filled
up to the mark with distilled water the flask is closed and the contents are
thoroughly mixed.
The dilution factor (DF) resulting from the dilution procedure is calculated
as follows:
Final volume (total volume)
DF =
Initial volume (sample volume)
The analytical result is subsequently multiplied by the dilution factor.
A calculation can be dispensed with when the dilution is programmed into
the photometer. The dilution number (see the table on page 14) is
entered and the measurement value is subsequently calculated correctly
and immediately displayed.
12
Release 06/2014
Spectroquant photometers
3 Sample Preparation
All dilutions should be made in such a way that the measurement value
lies in the middle of the measuring range. As a rule, the dilution factor
should never be higher than 100. In the event that yet higher dilutions
become necessary all the same, then this must be done in two separate
steps.
Example
Step 1:
Step 2:
The dilution factor for the total dilution is calculated by multiplying the
individual dilutions:
DF total = DF1 x DF2 = 100 x 20 = 2000, dilution number 1+1999
Simplified procedure
Dilutions up to 1:10 can also be prepared without volumetric flasks in a
glass beaker, measuring the volumes of the sample and the dilution water
using a previously calibrated positive-displacement pipette (see table for
instructions).
Desired
dilution
Volume of
sample
[ml]
Volume of
distilled water
[ml]
Dilution
factor
Dilution
number
1:2
1:3
1:4
1:5
1:10
5
5
2
2
1
5
10
6
8
9
2
3
4
5
10
1+1
1+2
1+3
1+4
1+9
3.4 Filtration
Strongly turbid samples require pretreatment before they can determined
in a photometer, since the effect of turbidity can result in considerable
variations in the measurement values and in false-high readings. Care
must be taken here to ensure that the substance to be determined is not
contained in the suspended material, in which case a sample decomposition must be carried out.
Compounds that always occur in dissolved form (for example ammonium,
nitrate, nitrite, chlorine, chloride, cyanide, fluoride, orthophosphate, and
sulfate) permit a previous filtration, even when the sample solution is
strongly turbid.
Weak turbidity is eliminated by the automatic turbidity-correction
feature built into the photometer (see Function description, Device set-up/
Correction function); in such cases it is not necessary to filter the sample
before analysis.
As a measure to distinguish between dissolved and undissolved waterborne substances, the water sample can be filtered through a simple
paper filter. Following the recommendations stated in the reference methods, membrane filters with a pore size of 0.45 m are required for fine
filtration.
Release 06/2014
13
Spectroquant photometers
3 Sample Preparation
Procedure for microfiltration
3.5 Homogenization
As a measure to ensure that a representative sample can be taken in the
presence of suspended matter in the water sample in question, for certain
parameters - e. g. COD and the total content of heavy metals - the sample
must be homogenized. This must be carried out using a high-speed blender (2 minutes at 5000 20 000 rpm and taking the sample while stirring.
3.6 Decomposition
Ion
Complex
Solid
substance
14
Release 06/2014
Spectroquant photometers
3 Sample Preparation
The manner in which the sample is pretreated enables the three proportions to be distinguished from each other. This can be illustrated using
a copper-containing wastewater sample as an example.
Example
Filtration
Decomposition
Decomposition
Filtration
Total content
Dissolved proportion Dissolved proportion
Solid Substances
Cu(OH) 2
Complexes Cu-EDTA Complexes Cu-EDTA
Ions Cu 2+
Ions Cu 2+
Ions Cu 2+
Result A
Proportion:
Ionogenic
Complex
Solid Substances
Total content
Result B
Result C
=C
= BC
= AB
=A
Decomposition converts the substance to be determined into an analyzable form. In most cases, decomposition agents take the form of acids
in combination with oxidizing agents; in exceptional cases (e. g. in the
determination of total nitrogen) an alkaline decomposition is more effective. The type of decomposition procedure used depends on the analyte to
be determined and the sample matrix.
Determination of
Total phosphorus*
Total chromium*
[= sum of chromate and chromium(III)]
Total metal
[= sum of free and complex-bound metal]
Total nitrogen*
Crack Set 20
* The decomposition reagents are already contained in the packs of the respective cell tests.
** Decomposition cells are included in the pack; empty cells are required for the decomposition for
Crack Sets 10 and 20.
Release 06/2014
15
Spectroquant photometers
3 Sample Preparation
In the event that the sample to be analyzed is a highly contaminated material (high proportion of organic substances) or water-insoluble samples,
decomposition using concentrated acids and other agents is indispensible.
Corresponding examples from the collection of applications for real samples are available on request.
The necessity for decomposition can be checked according to the following diagram:
Decomposition
Procedure
Procedure
Measurement
Measurement
Result A
Result B
Decomposition
necessary
No
A and B
idential?
Yes
No decomposition
necessary
4 Pipetting System
Positive-displacement pipettes permit
Pipettes of varying volumes and also ones with a fixed volume are available.
Sources of error and hints on how to avoid them:
Closely follow the instructions for use contained with the pipette in
question.
Check the pipetted volumes
a) by weighing using analytical scales (weighing accuracy 1 mg),
1 ml of water at 20 C = 1.000 g 1 mg
b) using Spectroquant PipeCheck;
this is a photometric check of the pipette, and scales are not
necessary (see section AQA).
16
Release 06/2014
Spectroquant photometers
These errors have effects on both the accuracy and precision of the
results obtained.
Release 06/2014
17
Spectroquant photometers
Chargenzertifikat
Lot Certificate / Certificado del lote
Spectroquant CSB-Kvettentest
Messbereich
Measuring Range / Intervalo de medida
HC119527
Verwendbarkeit
Expiry date / Fecha de caducidad
Sollwert
Messergebnis
Target value
Result / Resultado
Valor nominal
(Standard / Patrn)
(Standard / Patrn)
mg/l
mg/l
CSB/COD/DQO
CSB/COD/DQO
4,0
4,7
8,0
31.10.2012
8,7
12,0
11,8
16,0
16,4
20,0
19,9
340 nm
24,0
24,1
Fr. Brandner
28,0
28,4
32,0
32,4
20.09.2011
36,0
36,3
1145600001_HC119527_EN
40,0
40,2
Sollwert
Target value
Valor nominal
1,00 0,03
Chargenwert
Lot value
Valor del lote
0,99
Calibration function:
The calculated function must agree, within specified tolerances, with the
function electronically stored in the photometer.
n = 10
0,4
Reag.blindwert / Reagent blank / Valor en blanco del react +/- Tolerance / Tolerancia
Vertrauensbereich (95% Wahrscheinlichkeit)
Confidential interval (P=95%) / Intervalo de confianza (95 % de probabilidad)
Verfahrensstandardabweichung
Standard Deviation of the Method / Desviacin estndar del procedimiento
Verfahrensvariationskoeffizient
Variation Coefficient of the Method / Coeficiente de variacin del procedimiento
1,410 0,020 A
1,403 A
1,0 mg/l
0,8 mg/l
0,3 mg/l
1,4%
2,5 %
40,0
Messergebnis / Result / Resultado (mg/l)
Confidence interval:
Maximum deviation from the desired value over the entire measuring
range; every measurement value can be affected by this deviation; this
parameter is a measure for the accuracy.
35,0
30,0
25,0
20,0
15,0
10,0
5,0
0,0
0,0
5,0
10,0
15,0
20,0
25,0
30,0
35,0
40,0
Qualittskontrolle
Quality control / Control de calidad
Checking the
working equipment
Chemischer Sauerstoffbedarf
2 ml Probelsung
in ein Reaktionskvette geben
Mischen
Pipette
Test kit
Mischen
Kvette wird hei,
am Verschluss
anfassen
Abkhlen auf
Raumtemperatur
(mind. 30 min)
mind. 10 min
abkhlen
14 mm
im Thermoreaktor
erhitzen
148 C, 120 min
Messen
9
C
Photometer
Thermoreactor
Checking the
handling
operations
Suspend the bottom
sediment in the cell by
swirling.
18
Release 06/2014
Spectroquant photometers
Chemischer Sauerstoffbedarf
2 ml Probelsung
in ein Reaktionskvette geben
Mischen
Abkhlen auf
Raumtemperatur
(mind. 30 min)
Mischen
Kvette wird hei,
am Verschluss
anfassen
mind. 10 min
abkhlen
Messen
14 mm
im Thermoreaktor
erhitzen
148 C, 120 min
9
C
19
Spectroquant photometers
20
Release 06/2014
Spectroquant photometers
Release 06/2014
21
Spectroquant photometers
Accuracy: poor
Precision: poor
Major errors have been made!
Accuracy: good
Precision: poor
Calculation of the mean values from at least three or better even more
parallel determinations yields an approximation of the true value.
Accuracy: poor
Precision: good
The high degree of precision mistakenly indicates a correct value!
Accuracy: good
Precision: good
The ideal objective!
22
Release 06/2014
eng
Description of Function
06/2014
General instructions
General instructions
Symbols used
indicates notes that you must read for
your own safety, the safety of others and to
protect your meter from being damaged.
indicates notes that draw your attention to
special features.
Scope of delivery
Photometer
Power pack
Product documentation
Warranty
The designated meter is covered by a warranty of 2
years from the date of purchase. The meter warranty
extends to manufacturing faults that are determined
within the period of warranty. The warranty excludes
components that are replaced during maintenance,
such as batteries, accumulators, lamps etc.
The warranty claim extends to restoring the meter to
readiness for use but not, however, to any further claim
for damages. Improper handling or unauthorized
opening of the instrument invalidates any warranty
claim.
To ascertain the warranty liability, return the meter and
proof of purchase together with the date of purchase
freight paid or prepaid.
Contents
1. Overview . . . . . . . . . . . . . . . . . . . . . . . 29
1.1
1.2
1.3
1.4
2. Safety . . . . . . . . . . . . . . . . . . . . . . . . . . .31
2.1
2.2
Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . 32
General instructions . . . . . . . . . . . . . . . . . . . . . 32
2.2.1 Labeling of notes . . . . . . . . . . . . . . . . . 32
2.2.2 Dangers of disregarding the safety
instructions . . . . . . . . . . . . . . . . . . . . . .32
2.2.3 Qualification of the personnel . . . . . . . . 32
2.2.4 Technical state of the meter . . . . . . . . .32
3. Commissioning . . . . . . . . . . . . . . . . . . 34
3.1
3.2
5. Measuring the
Absorbance/Transmission
5.1
5.2
5.3
5.4
5.5
. . . . . . 39
Switching to the Absorbance/ Transmission
measuring mode . . . . . . . . . . . . . . . . . . . . . . . 39
Measuring the absorbance or transmission
39
Measuring using cell tests . . . . . . . . . . . . . . . . 40
Measuring using reagent tests . . . . . . . . . . . . 41
Measuring using tests without barcode . . . . . . 41
6. Documentation . . . . . . . . . . . . . . . . . . 42
6.1
6.2
6.3
6.4
7. Method Parameters . . . . . . . . . . . . . . . 48
7.1
7.2
7.3
Citation form . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.1.1 Changing the citation form . . . . . . . . . . 49
7.1.2 Performing a difference measurement . 50
Selecting the unit . . . . . . . . . . . . . . . . . . . . . . . 52
Entering the dilution . . . . . . . . . . . . . . . . . . . . . 53
Activating AQA . . . . . . . . . . . . . . . . . . . . . . . . 54
25
Contents
8.2
8.3
9. Kinetics . . . . . . . . . . . . . . . . . . . . . . . . .70
10. Correction functions . . . . . . . . . . . . . .71
10.1 Blank value . . . . . . . . . . . . . . . . . . . . . . . . . . .71
10.1.1 Activating the blank value
measurement
. . . . . . . . . . . . . . . .72
10.1.2 Measuring the blank value . . . . . . . . . .73
10.1.3 Erasing blank values . . . . . . . . . . . . . .73
10.1.4 Recalling blank values . . . . . . . . . . . . .74
10.2 Reference absorbance . . . . . . . . . . . . . . . . . .75
10.3 Turbidity correction . . . . . . . . . . . . . . . . . . . . .77
26
Contents
16. Maintenance, cleaning, disposal . . . 93
16.1 Maintenance - Changing the lamp . . . . . . . . . . 93
16.2 Cleaning - Actions to take if a cell is broken . . 93
16.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
27
Contents
28
1. Overview
Overview
a Display
b Menu call/Enter key
c Scroll key
d Absorbance measurement key
e Concentration measurement key
f Notch for cell alignment
g Round cell shaft
h Storage space for analysis regulations (short
form)
i Cover with integrated
on/off switch
j Rectangular cell shaft
k Keypad: numeric keypad,
cursor function, erase,
comma
29
1. Overview
1.3 Short manual
The short manual lists all of the steps necessary to
determine the concentration of a sample and to activate
AQA2 at a glance.
or
30
4.
1. Overview
1.4 Selecting and calling up the menu items
To switch on the photometer, open the cover.
Press
1.
3 or 28.
1.
3 or 28
called up using 1.
selected using
Selection lists:
Changes to the settings are accepted after confirmation by pressing
1
Scroll with 3 or 28.
Pressing
Character input:
via the numeric keypad or by using
3,
1.
31
2. Safety
Safety
32
2. Safety
It is the responsibility of the operator to continuously
observe the overall technical condition (externally
recognizable deficits and damage as well as alterations to the operational behavior) of the meter.
If safe operation is no longer possible, the equipment
must be taken out of service and secured against
inadvertent operation.
33
3. Commissioning
The photometer operates at an environmental temperature of +5 C to +40 C. During transport from cold to
warm surroundings, condensation can form resulting
in the malfunction of the meter.
Commissioning
Battery operation
Charge the battery for approx. 5 hours before the
initial commissioning. To do this:
After approx.
5 s:
34
4.
conc. + abs.
Example
35
After approx.
2 s:
36
After approx.
2 s:
3 or 28,
1.
37
After approx.
2 s:
38
1.
1.
In the setup menu, call up the abs./trm. %
submenu.
transmission
5.
Measuring mode, absorbance
39
3 or 28.
1.
40
3 or 28.
3 or 28
1.
41
6. Documentation
Documentation
1.
1.
42
6. Documentation
6.1 Resetting the number of the measured value
Call up the no. of meas. value submenu.
yes
3 or 28
1.
off
1.
43
6. Documentation
Measuring using the activated "ID number" function
or
1. Confirmation without
number
for method all measured values of a special
method
AQA all measurements of a specific method monitored with AQA.
44
1.
6. Documentation
Selecting "total"
Select the output medium:
to display
3 or 28
1.
3 or 28
45
6. Documentation
Selecting "with I. D. no."
Enter the ID number
Default:
The last ID number entered is preselected.
Confirm each time with
1.
3 or 28
to display
to printer/PC (serial interface).
Select the menu item with
Confirm with
3 or 28
Selecting "AQA"
Input the method
Default:
The last method selected.
Confirm with
46
6. Documentation
Memory download to display
Each data record appears individually on the display
beginning with the data record just measured.
The display shows:
Scroll with
citation
meas. value
unit
Where necessary,
AQA ID, e.g. AQA2.
3 or 28.
1.
Sample printout:
003
002
001
14541
14541
14729
10.02.98
10.02.98
03.02.98
11:56:33
11:54:21
18:30:53
t
t
*
80
70
0.3
mg/l
mg/l
mg/l
COD
COD
PO4-P
3 or 28
Start the download with 1.
47
7. Method Parameters
Method Parameters
unit
dilution
To switch on the photometer, open the cover.
Press
1.
1.
3 or 28
Call up the parameter by pressing 1.
48
7. Method Parameters
7.1 Citation form
7.1.1 Changing the citation form
Example:
Change the citation form from "NH4-N" to "NH4".
Call up the citation submenu.
3, scroll to NH
Confirm with 1.
Using
49
7. Method Parameters
7.1.2 Performing a difference measurement
Difference measuring is possible for some methods
(e.g. Iron II/III, Ca-/Mg Hardness).
For more information on this, see part, "Analysis
specifications".
Example:
Determination of iron (II) and iron (III).
Enter method 106
Confirm with
1.
Using
4.
Start the 1st measurement by inserting cell 1.
50
7. Method Parameters
After approx.
2 s:
1.
After approx.
2 s:
1.
51
7. Method Parameters
7.2 Selecting the unit
The preset unit is "mg/l".
It can be changed to "mmol/l".
Call up the unit submenu.
3 scroll to mmol/l
Confirm with 1.
Using
52
7. Method Parameters
7.3 Entering the dilution
Diluting a sample with distilled water enables the
measuring range to be extended.
The photometer automatically includes any dilution
factor entered when calculating the measured value.
Values from 0 to 99 can be entered as dilution factors.
Example:
Enter dilution 1:20 (i.e. 1 part sample + 19 parts distilled
water).
Call up the dilution submenu.
or
1.
53
8
Analytical Quality Assur-
standard solutions
The total system monitoring (AQA2) is a
method-specific check using standard solutions.
If this is performed successfully, it also includes
photometer monitoring (AQA1).
See also part "General information" for further
information on Analytical Quality Assurance
(AQA).
1.
Call up the meter setup submenu.
The meter setup submenu appears with the AQA functions menu item preselected.
Confirm with
1.
54
1.
off
n weeks
n measurements
Confirm with
1.
meter
system
The menu item, meter, only appears after the
corresponding PhotoCheck standards have
been input (see section 8.2.1).
55
Locking of the method for concentration measurements (as long as the locking is active).
Setting ranges:
Photometer monitoring (AQA1):
1 to 52 weeks (default: 12 weeks) or
1 to 9999 measurements (default: 1500)
Monitoring of the total system using standard solutions (AQA2):
1 to 52 weeks (default: 4 weeks) or
1 to 9999 measurements (default: 100)
With the n measurements setting, a difference
measurement (see section 7.1.2) is counted
as one measurement only.
1 three times
56
1.
1 three times
1.
1.
57
58
1.
3 or 28
1.
1.
1.
1.
1.
59
Select between
input
Input the theoretical value (absorbance) from the lot
certificate of Spectroquant PhotoCheck
output
Print/display theoretical values
erase
Erase theoretical values.
The erase and output menu items only appear
after at least one standard has been input.
60
3 or 28
1.
1.
1.
61
3 or 28
Confirm with 1 to start the download.
Select with
AQA1
13:19
AQA interval
12 weeks
test sol.
445-1
unit
A
theor. val.
0.200
tolerance
0.020
AQA date
26.08.97
3 or 28
62
1.
standards.
Press
After approx.
1 s:
After approx.
4 s:
After approx.
1 s:
63
After approx.
3 s:
Error elimination:
1. Repeat the measurement (insert
the cell again)
2. If necessary, perform a zero adjustment and repeat the check
3. Exchange the test solution (each
packet contains two identical test
solutions)
4. Use a new Spectroquant PhotoCheck packet
5. Quit and have the photometer
checked in the factory
The absorbance test is terminated if an error message occurs and the meter is not released.
On switching on, the warning message "AQA interval expired" appears until the AQA was
successfully performed or the AQA mode was switched off.
Example: Report output
AQA check meter
26.08.97
operator:
AQA1
10:23
AQA interval
AQA check AQA1
12 weeks
ok
L check
ok
test sol.
445-1
64
meas.
value
0.211
unit
theor. val.
tolerance
result
0.200
0.020
ok
Select between
input
Enter standards
output
Print/display standards
erase
Erase standards.
Displays the last selected method.
Select the method with
Confirm with
3 or 28
65
1.
1.
1.
66
1.
1.
to display
to printer/PC (serial interface).
3 or 28
Confirm with 1 to start the download.
Select with
AQA2
13:57
system locked
on
method
14554
14555
unit
mg/l
mg/l
theor. val.
2.00
5000
tolerance
0.20
400
AQA date
24.08.97
26.08.97
67
1.
3 or
28
Erase with
1.
or
68
After approx.
2 s:
AQA2
11:02
AQA interval
method
14554
4 weeks
meas.
value
1.95
unit
theor. val.
tolerance
result
mg/l
2.00
0.20
ok
AQA2
11:02
AQA interval
method
14554
100 measurements
meas.
value
1.95
unit
theor. val.
tolerance
result
mg/l
2.00
0.20
ok
69
9. Kinetics
9
Kinetics
The kinetics function enables the tracking of timedependent changes in concentration or absorbance (for
one selected wavelength) by repeating measurements
over an adjustable time interval (setting possibility 00:05
to 60:00 min, smallest interval 5 seconds).
The kinetic measured values are stored in the memory
in the same way as the measured values from single
measurements (older measured values are overwritten). The maximum number of measuring cycles of
a kinetic measurement is 1000 measurements
(until the memory is full of kinetic measured values). The
current measured value number appears in the header
of the display. When the memory is full, the meter aborts
the kinetic measurement. The starting values of the
kinetic are not overwritten!
The output of kinetic measured values after the termination of the measurement is performed via the menu
item, download memory.
To switch on the photometer, open the cover.
Press
1.
1.
70
10 Correction functions
1.
reference absorbance
turbidity correct.
Confirm with
1.
71
Press
mode.
The value measured against the prepared blank solution is displayed.
3 or 28
1.
72
After approx.
2 s:
menu opens.
The erase blank value menu item first appears
after a blank value has been measured.
Select between
all
Erase all stored blank values
single
Erase individual stored blank value
73
3 or 28
Erase the displayed blank value with 1.
3 or 28
1.
28
Confirm with
1.
74
1.
3 or 28
3 or
3 or 28
1.
75
3 key.
mode.
Insert the measuring cell.
After approx.
2 s:
76
Press
3 or 28
1.
mode.
Insert the measuring cell.
After approx.
2 s:
77
11 Zero adjustment
percussion, transport
if the ambient temperature changed by more than
5 C since the last zero adjustment
at least every six months.
78
Press
3 or
28.
After approx.
2 s:
79
12 User-defined methods
Press
3 or
28.
80
3 or
28.
3,
1.
Example:
1.
3,
1.
81
1.
0.01
0.1
1
10
100
1.
1.
1.
1.
82
1.
round
10 mm
20 mm
50 mm
3
Confirm with 1.
Select with
Example:
Display and automatic printout of code no. and characteristics as a list.
3
Return with 1.
Scroll with
Character
4
3
5
5
9
12
5
5
5
Example
CEME
301
TEST1
690nm
mmol/l
C6H5OH
0.009
2.12
0.1
22.3
2
5
(=round)
0.1
Example:
CEME 301 Test1 690nm mmol/l C6H5OH 0.009 2.12 0.1 22.3 14 0.1 <CR>
83
3 or
28
Start the print with
1.
1.
301
TEST1
690 nm
mmol/l
C6H5OH
0.1
+0.009
+2.12
0.1
22.3
round
84
3 or 28.
1.
Selecting "all"
Call up the erase menu item with
Confirm with
3 or 28
1.
Selecting "single"
Select a method, e.g. TEST1:
Input the method number (301) via the numeric
keypad
or
by scrolling with
Confirm with
1.
3 or 28
1.
85
13 Meter Setup
select language
set date/time
Performing a meter reset
system info
Deutsch (German)
English
Franais (French)
Italiano (Italian)
Portugus (Portuguese)
Polski (Polish)
Dansk (Danish)
Svenska (Swedish)
Espaol (Spanish)
Nederlands (Dutch)
Indonesia (Indonesian)
Cetina (Czech)
Magyar (Hungarian)
Russkij (Russian)
Trke (Turkish)
Brasil (Brasilian)
1
Press the 1 key again:
Confirm with
86
1.
13.3 Reset
It is possible to reset the photometer to its factory
settings (delivery state) in single steps. The reset total
function resets all settings and blank values at once.
Select between
total
1.
87
Sample display
88
You will always find the latest method data for your
photometer on the Internet. A method update contains
all new test sets and methods respectively. Additionally,
minor modifications of already existing methods are
transferred with it. With a method update, you receive all
new methods and, at the same time, can easily and
conveniently update all method data.
All method data are reloaded into the photometer with the update. The old method data are
overwritten by this.
Click on the "Search meter" button.
The program automatically recognizes the
connected photometer. Another "Update
Method Data" window appears.
Click on the "Start" button to start the method
download. The process takes approx. 3
minutes. You can terminate it at any time by
clicking on the "Cancel" button. In this case,
however, the download has to be carried out
once again completely so that the photometer
can save the method data and is operative.
89
Error messages
Message
No meter found
Meaning
Connection PC - photometer out of order or not
available
Photometer not recognized
90
Remedy
Tightly connect the cable to the
serial interfaces of the PC and
photometer.
Use the correct cable
Select the photometer manually
15
RS 232 C interface
Printer
Interface cable
Communication software.
15.1 Principle course of the remote control
String to meter
S <CR>
Command xx (see 15.2
command list)
.
.
.
CLOC <CR>
Operating mode
Remote (remote control)
Remote (remote control)
Concentration measurement
Function
Begin communication
Switchover to normal operation (concentration measurement)
Reads out stored measured values;
[anz] = number of the measured values to be output
Measurement and transmission of the concentration value with date/time;
[MMM] = method number (e.g. 086 for method 14729)
Measurement and transmission of the absorbance value for the wavelength;
[LLL] = wavelength
Measurement and transmission of the sample blank value;
[MMM] = method number
Erase measured sample blank values; [MMM] = method number
Input user-defined methods (see chapter chapter 12 USER-DEFINED METHODS)
Output user-defined methods; [MMM] = method number
91
10
unit
12
citation
AQA ID (AQA2/AQA1)
dilution factor
4
4
Notes:
Data fields are separated by spaces.
Character set: IBM, code page 437
Meaning of the special characters:
!
Q = AQA measurement
4800
8
1
none
Hardware
15 m
Photometer
9-pin socket
1
2
3
4
5
6
7
8
9
92
Computer
9-pin socket
4
3
2
1 and 6
5
4
8
7
25 pin plug
20
2
3
6
7
20
5
4
-
Printer
with RS 232 C interface
TXD
RXD
SG
DTR (if not available:
short-circuit CTS and RTS )
-
3:
Press
93
Measuring instrument
Dispose of the measuring instrument as electronic
waste at an appropriate collection point. It is illegal to
dispose of it in household refuse.
Within the European Union, the batteries are removed
at a specialized treatment center at the instrument's end
of life. The instruments are taken to one of those
specialized treatment centers via the recycling system
set up for this purpose.
94
17 Technical Data
Optical
measuring
principle
Light source
Receiver
Optical filters
Photometric
reproducibility
0.001 A at 1.000 A
Photometric
resolution
0.001 A
Warm-up time
none
Measuring time
approx. 2 s
Types of
measurement
Permanently stored
Drift correction
Retrofitting of
new methods
User-defined
methods
max. 10
Kinetics
Cell recognition
automatic
Self-Check
Time/Date
Dimensions
Weight
Guidelines and
norms used
Power pack
FRIWO FW 7555O/15
Friwo Part. No. 1822367
Input: 100 ... 240 V ~ /
50 ... 60 Hz / 400 mA
Output: 15 V DC / 1 A
Backup
battery
Battery
(optional)
Climatic class
Annual mean: 75 %
30 days/year: 95 %
other days: 85 %
Test certificate
CE
95
Display
Connections
Digital interface
Power supply 2-pin socket to connect the plugin power supply unit
Data storage
Equipment Statement
Note: This equipment has been tested and found
to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable
protection against harmful interference when the
equipment is operated in a commercial
environment. This equipment generates, uses,
and can radiate radio frequency energy and, if not
installed and used in accordance with the
instruction manual, may cause harmful
interference to radio communications. Operation
of this equipment in a residential area is likely to
cause harmful interference in which case the user
will be required to correct the interference at his
own expense.
AQA2 values:
none
(theoretical values and tolerances of all
methods are set to default values
96
18. What to do if
18 What to do if
Connect the photometer to the line power via the power pack.
In the case of battery operation: Battery empty, charging required (approx. 5h);
line operation is possible without restrictions during charging time.
Battery nearly empty. Charging required (see chapter 3 COMMISSIONING).
appears
Date/time is lost when
switched off
The backup battery of the real time clock is empty and has to be replaced.
Send the photometer to the service department for this.
Password forgotten
The connected printer is off line. Switch on the printer or pull out the interface
cable.
Error messages:
remove cell
The message remove cell appears on the display although no cell is inserted.
Clean the cell shaft with a damp, lint-free cloth.
If the error message still appears, return the photometer to the service department.
lamp defective
no zero adjustment
No zero adjustment is stored in the meter for the cell. Perform zero adjustment
(see chapter chapter 11 ZERO ADJUSTMENT).
cell error
The rectangular cell is inserted incorrectly, or two cells are inserted. Insert the
cell correctly.
cell invalid
A cell type that is not allowed for the selected method was selected, e. g. a round
cell for the reagent test.
method invalid
No data is stored in the photometer for the selected method. Update method
data (see chapter chapter 14 UPDATING METHOD DATA).
wrong method
During a difference measurement, the method was changed between the first
and second measurement. During a difference measurement, the method
must remain identical.
E_0
97
eng
Analytical Procedures
Appendices
Spectroquant photometers
Contents
Table
Analytical Procedures
Appendix 1
Appendix 2
Solutions
Appendix 3
Solutions
Method
No.
Determination
Total Range
Method
208
040
020
031
101758
Indicator reaction
100594
114825
114739
114558
114544
114559
114752
100683
100683
Application
100675
101747
100687
100826
114839
Application
Chromazurole S
Chromazurole S
Indophenol blue
Indophenol blue
Indophenol blue
Indophenol blue
Indophenol blue
Indophenol blue
Indophenol blue
Brilliant green
Oxidation to chloride
Ag-DDTC
Modification of Winkler method
Azomethine H
Rosocyanine
3,3-Dimethylnaphtidine
100605
114834
101745
100858
114815
114815
114730
114897
114897
101804
101807
100595
100597
S-DPD
Cadion derivative
Cadion derivative
Phthalein purple
Glyoxal-bis-hydroxyanil
Glyoxal-bis-hydroxyanil
Iron(III)-thiocyanat
Iron(III)-thiocyanat
Iron(III)-thiocyanat
Iron(III)-thiocyanat
Iron(III)-thiocyanat
S-DPD
S-DPD
100598
100602
100599
S-DPD
S-DPD
S-DPD
100086/100087/
100088
100608
114552
114552
DPD
114758
114560
101796
014
114540
105
114895
093
114690
023
114541
094
114691
024
114555
S-DPD
Diphenylcarbazide
Peroxodisulfate oxidation,
diphenylcarbazide
Diphenylcarbazide
Inherent color
Chromosulfuric acid oxidation,
chromate determination
Chromosulfuric acid oxidation,
chromate determination
Chromosulfuric acid oxidation,
chromate determination
Chromosulfuric acid oxidation,
chromate determination
Chromosulfuric acid oxidation,
chromate determination
Chromosulfuric acid oxidation,
chromium(III) determination
Chromosulfuric acid oxidation,
chromium(III) determination
Chromosulfuric acid oxidation,
chromium(III) determination
211
196
043
104
051
052
053
054
155
163
130
156
132
157
164
046
195
146
067
183
165
042
125
095
110
063
218
219
141
142
143
145
144
194
149
039
039
Determination
Total Range
Method
209
101797
109772
138
109773
220
117058
221
117059
015
CO445
CO525
Measurement at 525 nm
CO620
Measurement at 620 nm
032
Color (445)
(spectral absorption coefficient)
Color (525)
(spectral absorption coefficient)
Color (620)
(spectral absorption coefficient)
Color Hazen*
137
CU340
179
Color Hazen*
CU445
026
027
083
228
044
099
128
198
147
033
021
037
106
Platinum-cobalt-Standard Method,
measurement at 340 nm
Platinum-cobalt-Standard Method,
measurement at 445 nm
Cuprizone
Cuprizone
Inherent color
Barbituric acid and
pyridinecarboxylic acid
Barbituric acid and
pyridinecarboxylic acid
Citric acid, barbituric acid,
and pyridinecarboxylic acid
Barbituric acid and
pyridinecarboxylic acid
Citric acid, barbituric acid,
and pyridinecarboxylic acid
Triazine derivative
Alizarin complexone
Alizarin complexone
Alizarin complexone
Alizarin complexone
Alizarin complexone
Alizarin complexone
SPADNS
Chromotropic acid
Chromotropic acid
Rhodamine B
038
161
Iron Test
Iron Test*
114761
100796
066
160
158
159
184
114833
109717
100815
100816
101739
061
078
075
075
109
109
210
076
124
215
216
166
167
217
028
091
045
4-Dimethylaminobenzaldehyde
Titanyl sulfate
Titanyl sulfate
Phenanthroline derivative
S-DPD
Measurement at 340 nm
Measurement at 445 nm
Triazine
2,2-Dipyridyl
Triazine
1,10-Phenanthroline
PAR
PAR
Phthalein purple
Formaldoxime
PAN
Determination
019
226
135
175
206
185
017
018
057
059
030
107
151
060
139
072
140
227
035
197
036
068
153
Manganese Test*
Manganese Test*
Mercury in water and wastewater
Molybdenum Cell Test
Molybdenum Test
Monochloramine Test
Nickel Cell Test*
Nickel Test*
Nickel Baths
Nitrate Cell Test*
Nitrate Cell Test*
Nitrate Cell Test*
Nitrate Cell Test*
Nitrate Test*
Nitrate Test*
Nitrate Cell Test in seawater*
Nitrate Test in seawater*
Nitrate Test
Nitrite Cell Test*
Nitrite Cell Test*
Nitrite Test*
Nitrogen (total) Cell Test
Nitrogen (total) Cell Test*
114770
101846
Application
100860
119252
101632
114554
114785
108
092
207
148
133
186
073
176
177
212
055
055
213
086
086
152
214
214
056
162
069
070
134
103
150
098
079
081
169
171
225
047
168
229
064
154
082
065
224
230
080
Total Range
Method
114542
114563
114764
100614
114773
109713
114556
114942
101842
114547
100609
114776
114537
100613
114763
10 150 mg/l N
114694
119251
100607
Application
101744
114551
100856
100856
100474
114543
114543
100475
114729
114729
100616
100673
100673
114848
100798
114546
114842
Application
114562
100615
114683
114794
114794
100857
100857
101813
114831
100885
Formaldoxime
PAN
Michlers ketone
Bromopyrogallol red
Mercaptoacetic acid
Indophenol blue
Dimethylglyoxime
Dimethylglyoxime
Inherent color
Nitrospectral
2,6-Dimethylphenol
2,6-Dimethylphenol
2,6-Dimethylphenol
Nitrospectral
2,6-Dimethylphenol
Resorcine
Resorcine
Benzoic acid derivative
Griess reaction
Iron(II) ethylenediammonium sulfate
Griess reaction
Peroxodisulfate oxidation, nitrospectral
Peroxodisulfate oxidation,
2,6-dimethylphenol
Peroxodisulfate oxidation,
2,6-dimethylphenol
Modification of Winkler method
FerroZine
S-DPD
Thio-Michlers ketone
Phenol red
MBTH
Aminoantipyrine
Aminoantipyrine, by extraction
Phosphomolybdenum blue
Phosphomolybdenum blue
Peroxodisulfate oxidation,
phosphomolybdenum blue
Phosphomolybdenum blue
Phosphomolybdenum blue
Peroxodisulfate oxidation,
phosphomolybdenum blue
Phosphomolybdenum blue
Phosphomolybdenum blue
Peroxodisulfate oxidation,
phosphomolybdenum blue
Phosphomolybdenum blue
Phosphomolybdenum blue
Vanadatomolybdate
Vanadatomolybdate
o-Phenylendiamine
Kalignost, turbidimetric
Kalignost, turbidimetric
Phthalein purple
Silicomolybdenum blue
Silicomolybdenum blue
Molybdatosilicate
Molybdatosilicate
Silicomolybdenum blue
Eosine / 1,10-phenanthroline
indirectly as chloride
102532
114548
100617
114564
114791
101812
102537
114779
Bariumsulfate, turbidimetric
Bariumsulfate, turbidimetric
Bariumsulfate, turbidimetric
Bariumsulfate, turbidimetric
Tannin
Bariumsulfate, turbidimetric
Bariumsulfate, turbidimetric
Dimethyl-p-phenylendiamine
Determination
071
127
187
087
231
Total Range
Method
114394
114394
101746
114697
Ellmans reagent
Ellmans reagent
Ellmans reagent
Methylene blue
102552
Methylene blue
192
193
182
100
172
173
178
077
191
222
223
174
074
041
Disulfine blue
TBPE
1 100 FAU
50 3000 mg/l HOAc
50 3000 mg/l CH3COOH
50 3000 mg/l CH3COOH
0.025 1.000 mg/l Zn
0.20 5.00 mg/l Zn
0.05 2.50 mg/l Zn
Measurement at 550 nm
Esterification
Esterification
Esterification
PAR
PAR
Cl-PAN
Pyrocatechol violet
Peroxodisulfate oxidation, indicator
Peroxodisulfate oxidation, indicator
Phthalein purple
Release 06/2014
101758
Cell Test
8.00 mmol/l
400
mg/l CaCO3
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a sodium hydroxide
solution 0.1 mol/l, Cat.No. 109141, can be used after
diluting accordingly (see section Standard solutions).
Aluminium
100594
Cell Test
Measuring
range:
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
aluminium standard solution Certipur, Cat.No. 119770,
concentration 1000 mg/l Al can be used after diluting
accordingly.
114825
Aluminium
Test
Measuring
range:
10-mm cell
20-mm cell
50-mm cell
Reaction time:
2 minutes
Important:
Quality assurance:
114739
Ammonium
Cell Test
Measuring
range:
0.01
Reaction time:
15 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 50, Cat.No. 114695, or the
Standard solution for photometric applications, CRM,
Cat.No. 125022 and 125023.
Ready-for-use ammonium standard solution Certipur,
Cat.No. 119812, concentration 1000 mg/l NH4+, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 50) is highly recommended.
114558
Ammonium
Cell Test
Measuring
range:
Reaction time:
15 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10, Cat.No. 114676, or the
Standard solution for photometric applications, CRM,
Cat.No. 125022, 125023, 125024, and 125025.
Ready-for-use ammonium standard solution Certipur,
Cat.No. 119812, concentration 1000 mg/l NH4+, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
114544
Ammonium
Cell Test
Measuring
range:
Reaction time:
15 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 20, Cat.No. 114675, or the
Standard solution for photometric applications, CRM,
Cat.No. 125023, 125024, 125025, and 125026.
Ready-for-use ammonium standard solution Certipur,
Cat.No. 119812, concentration 1000 mg/l NH4+, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 20) is highly recommended.
114559
Ammonium
Cell Test
Measuring
range:
Reaction time:
15 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 70, Cat.No. 114689, or the
Standard solution for photometric applications, CRM,
Cat.No. 125025, 125026, and 125027.
Ready-for-use ammonium standard solution Certipur,
Cat.No. 119812, concentration 1000 mg/l NH4+, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 70) is highly recommended.
114752
Ammonium
Test
Measuring
10-mm cell
range:
10-mm cell
20-mm cell
20-mm cell
50-mm cell
50-mm cell
Shake vigorously to
dissolve the solid
substance.
Reaction time:
5 minutes
Reaction time:
5 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 50, Cat.No. 114695, or the
Standard solution for photometric applications, CRM,
Cat.No. 125022, 125023, and 125024.
100683
Ammonium
Test
Measuring range:
10-mm cell
10-mm cell
10-mm cell
10-mm cell
Reaction time:
15 minutes
Shake vigorously to
dissolve the solid
substance.
Important:
Very high ammonium concentrations in the sample
produce turquoise-colored solutions (measurement solution should be yellow-green to green) and false-low readings are yielded. In such cases the sample must be diluted (plausibility check).
Release 06/2014 - Spectroquant photometer NOVA 60
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 70, Cat.No. 114689, or the
Standard solution for photometric applications, CRM,
Cat.No. 125025, 125026, and 125027.
Ready-for-use ammonium standard solution Certipur,
Cat.No. 119812, concentration 1000 mg/l NH4+, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 70) is highly recommended.
Application
10-mm cell
Add approx. 1.5 g of alu- Shake the cell vigorousminium chloride hexa- ly to dissolve the solid
substance.
hydrate extra pure
(Cat.No. 101084), close
the cell with the screw
cap.
Add 2 drops of
reagent 1, close the
cell with the screw cap,
and mix.
Reaction time:
3 minutes
Add 2 drops of
reagent 2, close the
cell with the screw cap,
and mix.
Add 2 drops of
reagent 3, close the
cell with the screw cap,
and mix.
Aspirate the clear upper Transfer the solution into Place the cell into the
phase from the tube with a rectangular cell.
cell compartment.
pipette.
Select method no. 130.
Reaction time:
2 minutes
Note:
Important:
AOX
100675
Cell Test
Run 3 separate
1-ml portions of
AOX-2 through the
column. Discard the
wash solution.
Run 3 separate
1-ml portions of
AOX-3 through the
column. Discard the
wash solution.
Sample enrichment:
AOX
100675
Cell Test
Digestion:
Determination:
Pipette 0.20 ml of
AOX-1K into a reaction
cell, and mix.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) Spectroquant AOX
Standard, Cat.No. 100680, concentration 0.2 2.0 mg/l
can be used.
101747
Arsenic
Test
Measuring
range:
10-mm cell
20-mm cell
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
arsenic standard solution Certipur, Cat.No. 119773,
concentration 1000 mg/l As can be used after diluting
accordingly.
BOD
100687
Cell Test
Measuring
range:
Measurement of inital
oxygen concentration
= Result 1
(measurement sample)
= Result 1
(blank)
Check the pH of the
sample, specified range:
pH 6 8.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Add 10 drops of
BOD-3K, reclose,
and mix.
Determination:
Measurement of final
oxygen concentration
= Result 2
(measurement sample)
= Result 2
(blank)
After incubation, use
one bottle of pretreated
sample and one of
inoculated nutrientsalt solution for the
measurement of the final
oxygen concentration.
Calculation:
BOD of measurement sample:
Result 1 Result 2 (measurement sample) = A in mg/l
BOD of blank:
Result 1 Result 2 (blank) = B in mg/l
BOD of original sample in mg/l = A dilution factor B
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) Spectroquant BOD
Standard (acc. to EN 1899), Cat.No. 100718, can be used.
100826
Boron
Cell Test
Measuring
range:
Reaction time:
60 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
boron standard solution Certipur, Cat.No. 119500,
concentration 1000 mg/l B can also be used after diluting
accordingly.
114839
Boron
Test
Measuring
10-mm cell
range:
Reaction time:
12 minutes
Reaction time:
2 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
boron standard solution Certipur, Cat.No. 119500,
concentration 1000 mg/l B can also be used after diluting
accordingly.
Application
50-mm cell
Attention!
The measurement is carried out at 550 nm in a 50-mm rectangular cell against a blank, prepared
from distilled water (Water for analysis EMSURE, Cat.No. 116754, is recommended) and the
reagents in an analogous manner.
20 ml
Evaporate 200 ml of
sample solution in a
glass beaker almost to
dryness on the hob.
Add 0.10 ml of
reagent 1 with pipette
and mix.
Add 0.20 ml of
reagent 2 with pipette
and mix.
Reaction time:
30 minutes
Important:
The exact composition and preparation of the reagents 1
and 2 used are given in the corresponding application,
which also includes further information on the method
employed. This application can be downloaded directly at
www.analytical-test-kits.com.
100605
Bromine
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Pipette 10 ml of the
sample into a test tube.
Shake vigorously to
dissolve the solid
substance.
Reaction time:
1 minute
Important:
Quality assurance:
114834
Cadmium
Cell Test
Measuring
range:
Reaction time:
2 minutes
Important:
Quality assurance:
101745
Cadmium
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Add 10 ml of the sample Add 0.20 ml of Cd-2 with Add 1 level green
with pipette and mix.
pipette and mix.
microspoon of Cd-3
and dissolve the solid
substance.
Reaction time:
2 minutes
Important:
Quality assurance:
100858
Calcium
Cell Test
Measuring
range:
10 250 mg/l Ca
14 350 mg/l CaO
25 624 mg/l CaCO3
Expression of results also possible in mmol/l.
Reaction time:
exactly 3 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a freshly prepared
standard solution can be used (see section Standard
solutions).
114815
Calcium
Test
Measuring
range:
10 160
mg/l Ca
5 80
mg/l Ca
14 224
mg/l CaO
25 400
mg/l CaCO3
10-mm cell
7 112
mg/l CaO
12 200
mg/l CaCO3
20-mm cell
10-mm cell
Reaction time:
8 minutes, measure
immediately.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
calcium standard solution Certipur, Cat.No. 119778,
concentration 1000 mg/l Ca, can be used after
diluting accordingly.
114730
Chloride
Cell Test
Measuring
5 125 mg/l Cl
range:
Pipette 0.50 ml of
Add 1.0 ml of the sample
with pipette, close with
Cl-1K into a reaction
cell, close with the screw the screw cap, and mix.
cap, and mix.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) we recommended
to use Spectroquant CombiCheck 10 and 20, Cat.Nos.
114676 and 114675.
Ready-for-use chloride standard solution Certipur,
Cat.No. 119897, concentration 1000 mg/l Cl , can also be
used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck) is highly recommended.
114897
Chloride
Test
Measuring
10 250 mg/l Cl
10-mm cell
range:
10-mm cell
Reaction time:
1 minute
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) we recommended
to use Spectroquant CombiCheck 60, Cat.No. 114696.
Ready-for-use chloride standard solution Certipur,
Cat.No. 119897, concentration 1000 mg/l Cl , can also be
used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 60) is highly recommended.
101804
Chloride
Cell Test
Measuring
range:
Pipette 10 ml of the
sample into a reaction
cell, close with the
screw cap, and mix.
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use chloride
standard solution Certipur, Cat.No. 119897, concentration 1000 mg/l Cl-, can be used after diluting accordingly.
101807
Chloride
Test
Measuring
range:
50-mm cell
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use chloride
standard solution Certipur, Cat.No. 119897, concentration 1000 mg/l Cl-, can be used after diluting accordingly.
Chlorine
100595
Cell Test
Measuring
range:
Reaction time:
1 minute
Important:
Quality assurance:
Measuring
range:
Chlorine
100597
Cell Test
Reaction time:
1 minute
Important:
Quality assurance:
Chlorine
100598
Test
Measuring
0.05 6.00
mg/l Cl2
10-mm cell
range:
0.02 3.00
mg/l Cl2
20-mm cell
50-mm cell
Pipette 10 ml of the
sample into a test tube.
Shake vigorously to
dissolve the solid
substance.
Reaction time:
1 minute
Important:
Quality assurance:
Chlorine
100602
Test
Measuring
0.05 6.00
mg/l Cl2
10-mm cell
range:
0.02 3.00
mg/l Cl2
20-mm cell
50-mm cell
Pipette 10 ml of the
sample into a test tube.
Reaction time:
1 minute
Shake vigorously to
dissolve the solid
substance.
Important:
Quality assurance:
Chlorine
100599
Test
Measuring
0.05 6.00
mg/l Cl2
10-mm cell
range:
0.02 3.00
mg/l Cl2
20-mm cell
50-mm cell
Expression of results also possible in mmol/l and also in free Cl2 [Cl2 (f)], combined Cl2 [Cl2 (b)], and
total Cl2 [Cl2 (t)].
Determination of free chlorine
Pipette 10 ml of the
sample into a test tube.
Shake vigorously to
dissolve the solid
substance.
Reaction time:
1 minute
Quality assurance:
Measuring
range:
100086/100087/
100088
Cell Test
Reaction time:
1 minute
Important:
Quality assurance:
Measuring
range:
100086/100087/
100088
Test
50-mm cell
Expression of results also possible in mmol/l and also in free Cl2 [Cl2 (f)], combined Cl2 [Cl2 (b)], and
total Cl2 [Cl2 (t)].
Reaction time:
1 minute
Quality assurance:
Chlorine Dioxide
100608
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Pipette 10 ml of the
sample into a test tube.
Reaction time:
2 minutes
Shake vigorously to
dissolve the solid
substance.
Reaction time:
1 minute
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a freshly prepared
standard solution can be used (see section Standard
solutions).
Chromate
114552
Determination of chromium(VI)
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
chromate standard solution Certipur, Cat.No. 119780,
concentration 1000 mg/l CrO 42, can be used after diluting
accordingly.
114552
Chromate
Determination of total chromium
Cell Test
range:
Pipette 10 ml of the
sample into an empty
round cell (Empty cells,
Cat.No. 114724).
Reaction time:
1 minute
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
chromate standard solution Certipur, Cat.No. 119780,
concentration 1000 mg/l CrO 42, can be used after diluting
accordingly.
Chromate
114758
Determination of chromium(VI)
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Reaction time:
1 minute
Important:
Quality assurance:
20 400
g/l CrO3
10-mm cell
range:
10 200
g/l CrO3
20-mm cell
50-mm cell
COD
114560
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 50, Cat.No. 114695, or the
Standard solution for photometric applications, CRM,
Cat.No. 125028.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 50) is highly recommended.
COD
101796
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 50, Cat.No. 114695, or the
Standard solution for photometric applications, CRM,
Cat.No. 125028.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 50) is highly recommended.
COD
114540
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10, Cat.No. 114676, or the
Standard solution for photometric applications, CRM,
Cat.No. 125029.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
COD
114895
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 60, Cat.No. 114696, or the
Standard solution for photometric applications, CRM,
Cat.No. 125029 and 125030.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 60) is highly recommended.
COD
114690
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 60, Cat.No. 114696, or the
Standard solution for photometric applications, CRM,
Cat.No. 125029, 125030, and 125031.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 60) is highly recommended.
COD
114541
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 20, Cat.No. 114675, or the
Standard solution for photometric applications, CRM,
Cat.No. 125029, 125030, 125031, and 125032.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 20) is highly recommended.
COD
114691
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents,measurement device, and handling) we recommended to use
Spectroquant CombiCheck 80, Cat.No. 114738, or the
Standard solution for photometric applications, CRM,
Cat.No. 125031, 125032, and 125033.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 80) is highly recommended.
COD
114555
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 70, Cat.No. 114689, or the
Standard solution for photometric applications, CRM,
Cat.No. 125032, 125033, and 125034.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 70) is highly recommended.
COD
101797
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
the Standard solution for photometric applications, CRM,
Cat.No. 125034 and 125035.
COD (Hg-free)
109772
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
the Standard solution for photometric applications, CRM,
Cat.No. 125028 and 125029.
COD (Hg-free)
109773
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
the Standard solution for photometric applications, CRM,
Cat.No. 125029, 125030, 125031, and 125032.
COD
117058
Cell Test
16-mm cell
Tmax: 45 C
Chloride depletion:
Fill 6 - 7 g each of
Sodalime with indicator (Cat. No. 106733)
into two absorption
tubes (Cat. No. 115955).
COD
117058
Cell Test
Determination:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a COD/chloride standard
solution must be prepared from Potassium hydrogen
phthalate, Cat.No. 102400 and Sodium chloride, Cat.No.
106404 (see section Standard solutions).
COD
117059
Cell Test
16-mm cell
Tmax: 45 C
Chloride depletion:
Fill 6 - 7 g each of
Sodalime with indicator (Cat. No. 106733)
into two absorption
tubes (Cat. No. 115955).
COD
117059
Cell Test
Determination:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a COD/chloride standard
solution must be prepared from Potassium hydrogen
phthalate, Cat.No. 102400 and Sodium chloride, Cat.No.
106404 (see section Standard solutions).
Color
(Spectral Absorption Coefficient)
analogous to EN ISO 7887
Measuring
range:
445 nm
50-mm cell
-1
525 nm
50-mm cell
-1
620 nm
10-mm cell
-1
0.3 125.0 m
620 nm
20-mm cell
620 nm
50-mm cell
0.1 50.0 m
1
250
Notes:
Filtered sample =
true color.
Unfiltered sample =
apparent color.
Color Hazen
(Platinum-Cobalt Standard Method)
analogous to APHA 2120B, DIN EN ISO 6271-2, Water Research Vol. 30, No. 11, 2771-2775, 1996
Measuring 1 - 500
mg/l Pt/Co
1 - 500
mg/l Pt
1 - 500
Hazen 1 - 500
CU
range:
mg/l Pt/Co
1 - 250
mg/l Pt
1 - 250
Hazen 1 - 250
1 - 250
340 nm
10-mm cell
CU
340 nm
20-mm cell
340 nm
50-mm cell
Notes:
Filtered sample =
true color.
Unfiltered sample =
apparent color.
Quality assurance:
To check the measurement system (measurement device,
handling) ready-for-use Platinum Cobalt Color Reference
Solution (Hazen 500) Certipur, Cat.No. 100246, concentration 500 mg/l Pt, can be used after diluting accordingly.
Color Hazen
(Platinum-Cobalt Standard Method)
analogous to APHA 2120B, DIN EN ISO 6271-2, Water Research Vol. 30, No. 11, 2771-2775, 1996
1 - 1000 mg/l Pt
1 - 1000 Hazen
1 - 1000 CU
445 nm
50-mm cell
Notes:
Filtered sample =
true color.
Unfiltered sample =
apparent color.
Quality assurance:
To check the measurement system (measurement device,
handling) ready-for-use Platinum Cobalt Color Reference
Solution (Hazen 500) Certipur, Cat.No. 100246, concentration 500 mg/l Pt, can be used.
114553
Copper
Cell Test
Measuring
range:
Reaction time:
5 minutes
Important:
Quality assurance:
114767
Copper
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Reaction time:
5 minutes
Important:
Quality assurance:
10-mm cell
20-mm cell
50-mm cell
Pipette 25 ml of the
sample into a 100-ml
volumetric flask,
fill to the mark with
distilled water and
mix thoroughly.
Cyanide
102531
Cell Test
Measuring
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
cyanide standard solution Certipur, Cat.No. 119533,
concentration 1000 mg/l CN , can be used after diluting
accordingly.
Cyanide
114561
Cell Test
Measuring
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
cyanide standard solution Certipur, Cat.No. 119533,
concentration 1000 mg/l CN , can be used after diluting
accordingly.
Cyanide
114561
Cell Test
Measuring
range:
Expression of results also possible in mmol/l and cyanide readily liberated [CN(v)].
Pipette 10 ml of the
sample into an empty
round cell (Empty cells,
Cat.No. 114724).
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
cyanide standard solution Certipur, Cat.No. 119533,
concentration 1000 mg/l CN , can be used after diluting
accordingly.
Measuring
range:
Cyanide
109701
Test
10-mm cell
20-mm cell
50-mm cell
Reaction time:
10 minutes
Note:
Quality assurance:
Important:
To measure in the 50-mm cell, the sample volume and
the volume of the reagents CN-3 and CN-4 have to be
doubled for each.
Alternatively, the semi-microcell, Cat.No. 173502, can be
used.
Release 06/2014 - Spectroquant photometer NOVA 60
Cyanide
109701
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Expression of results also possible in mmol/l and cyanide readily liberated [CN(v)].
Reaction time:
10 minutes
Note:
Important:
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
cyanide standard solution Certipur, Cat.No. 119533,
concentration 1000 mg/l CN , can be used after diluting
accordingly.
119253
Cyanuric Acid
Test
Measuring
range:
20-mm cell
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a cyanuric acid standard
solution must be prepared from Cyanuric acid, Cat.No.
820358 (see section Standard solutions).
114557
Fluoride
Cell Test
Measuring
Round cell
range:
Reaction time:
5 minutes
Fluoride sensitive
Use the same preparation procedure as above, but
add 10 ml of sample instead of 5.0 ml. Prepare an
own blank by using 10 ml of distilled water and all
reagents. For measurement transfer the solution into
a 50-mm cell. Configure the photometer prior for
blank-measurement. Select method F sens in the
menu (method no. 124).
Important:
Quality assurance:
Fluoride
100809
Cell Test
Measuring
Round cell
range:
50-mm cell
Reaction time:
15 minutes
Important:
Quality assurance:
100809
Fluoride
Cell Test
Pipette 10 ml of the
sample into a reaction
cell, close with the screw
cap, and mix.
Pipette 10 ml of distilled
water into a second reaction cell, close with the
screw cap, and mix.
(Blank)
Reaction time:
15 minutes
Important:
Quality assurance:
114598
Fluoride
Test
Measuring range: 0.10 2.00 mg/l F
10-mm cell
10-mm cell
Reaction time:
5 minutes
Important:
Quality assurance:
Fluoride
100822
Test
Reaction time:
1 minute
Important:
Quality assurance:
114500
Formaldehyde
Cell Test
Measuring
range:
Add 1 level green micro- Shake the cell vigorousspoon of HCHO-1K into ly to dissolve the solid
a reaction cell, close
substance.
with the screw cap.
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a formaldehyde
standard solution must be prepared from Formaldehyde
solution 37%, Cat.No. 104003 (see section Standard
solutions).
114678
Formaldehyde
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Pipette 4.5 ml of
HCHO-1 into an empty
round cell (Empty cells,
Cat.No. 114724).
Reaction time:
5 minutes
Note:
Quality assurance:
114821
Gold
Test
Measuring
10-mm cell
range:
Add 2 drops of Au-1 and Add 4 drops of Au-2 and Add 6 drops of Au-3 and
mix.
mix.
mix.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
gold standard solution Certipur, Cat.No. 170216,
concentration 1000 mg/l Au, can be used after diluting
accordingly.
109711
Hydrazine
Test
Measuring
10-mm cell
range:
0.01 1.00
mg/l N2H4
20-mm cell
50-mm cell
Reaction time:
5 minutes
Important:
Quality assurance:
Hydrogen Peroxide
114731
Cell Test
Measuring
Round cell
range:
50-mm cell
Pipette 10 ml of the
sample into a reaction
cell, close with the
screw cap, and mix.
Reaction time:
2 minutes
Pipette 10 ml of the
sample into a reaction
cell, close with the
screw cap, and mix.
Reaction time:
2 minutes
Select method
H2O2 sens in the menu
(method no. 128).
Important:
Quality assurance:
Hydrogen Peroxide
118789
Test
Measuring
10-mm cell
range:
20-mm cell
Pipette 0.50 ml of
H2O2-1 into a test tube.
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a hydrogenperoxide
standard solution must be prepared from Perhydrol
30% H2O2 GR, Cat.No. 107209 (see section Standard
solutions).
100606
Iodine
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Pipette 10 ml of the
sample into a test tube.
Shake vigorously to
dissolve the solid
substance.
Reaction time:
1 minute
Important:
Quality assurance:
Measuring
0.05 3.00
range:
0.03 1.50
340 nm
20-mm cell
0.010 0.600
340 nm
50-mm cell
340 nm
10-mm cell
Measuring
1.0 50.0
445 nm
10-mm cell
range:
0.5 25.0
445 nm
20-mm cell
0.2 10.0
445 nm
50-mm cell
114549
Iron
Cell Test
Measuring
range:
Reaction time:
3 minutes
Important:
Quality assurance:
Iron
114896
Cell Test
Measuring
range:
Reaction time:
5 minutes
Important:
Quality assurance:
114761
Iron
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Important:
Quality assurance:
Iron
100796
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Determination of iron(II)
Calculation of iron(III)
Same preparation as
discribed above. After
adding of Fe-2 continue
with Fe-3.
Reaction time:
10 minutes,
then measure.
Important:
Quality assurance:
114833
Lead
Cell Test
Measuring
range:
Result A
Result B
= mg/l Pb
Important:
Quality assurance:
109717
Lead
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Important:
Quality assurance:
100815
Magnesium
Cell Test
Measuring
range:
Reaction time:
exactly 3 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a freshly prepared
standard solution can be used (see section Standard
solutions).
100816
Manganese
Cell Test
Measuring
range:
Reaction time:
5 minutes
Reaction time:
2 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) we recommended
to use Spectroquant CombiCheck 30, Cat.No. 114677.
Ready-for-use manganese standard solution Certipur,
Cat.No. 119789, concentration 1000 mg/l Mn, can also be
used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 30) is highly recommended.
101739
Manganese
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Reaction time:
10 minutes
Important:
Quality assurance:
114770
Manganese
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Reaction time:
2 minutes
Reaction time:
2 minutes
Important:
Quality assurance:
101846
Manganese
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Important:
Quality assurance:
Application
50-mm cell
Add 1.0 ml of
reagent 1 with pipette
and mix.
Add 1.5 ml of
reagent 2 with pipette
and mix.
Reaction time:
5 minutes
Important:
The exact composition and preparation of the reagents 1
and 2 used are given in the corresponding application,
which also includes further information on the method
employed. This application can be downloaded directly at
www.analytical-test-kits.com.
100860
Molybdenum
Cell Test
Measuring
range:
Place 2 drops of
Mo-1K into a reaction
cell and mix.
Add 10 ml of the
sample with pipette,
close the cell with the
screw cap.
Reaction time:
2 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a ready-for-use
molybdenum standard solution Certipur, Cat.No. 170227,
concentration 1000 mg/l Mo, can be used after diluting
accordingly.
119252
Molybdenum
Test
Measuring
20-mm cell
range:
20-mm cell
20-mm cell
Pipette 10 ml of the
sample into into an
empty round cell (Empty
cells, Cat.No. 114724).
Reaction time:
5 minutes, measure
immediately.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a ready-for-use
molybdenum standard solution Certipur, Cat.No. 170227,
concentration 1000 mg/l Mo, can be used after diluting
accordingly.
101632
Monochloramine
Test
Measuring
0.05 1.98
mg/l NH2Cl-N
10-mm cell
range:
20-mm cell
50-mm cell
Pipette 10 ml of the
sample into a test tube.
Reaction time:
10 minutes
Reaction time:
5 minutes
Important:
Quality assurance:
114554
Nickel
Cell Test
Measuring
range:
Reaction time:
2 minutes
Important:
Quality assurance:
114785
Nickel
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Reaction time:
1 minute
Reaction time:
2 minutes
Important:
Quality assurance:
10 120 g/l Ni
10-mm cell
20-mm cell
50-mm cell
114542
Nitrate
Cell Test
Measuring
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 20, Cat.No. 114675, or the
Standard solution for photometric applications, CRM,
Cat.No. 125037 and 125038.
Ready-for-use nitrate standard solution Certipur,
Cat.No. 119811, concentration 1000 mg/l NO3, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 20) is highly recommended.
Release 06/2014 - Spectroquant photometer NOVA 60
114563
Nitrate
Cell Test
Measuring
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 20, Cat.No. 114675, or the
Standard solution for photometric applications, CRM,
Cat.No. 125037 and 125038.
Ready-for-use nitrate standard solution Certipur,
Cat.No. 119811, concentration 1000 mg/l NO3, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 20) is highly recommended.
Release 06/2014 - Spectroquant photometer NOVA 60
114764
Nitrate
Cell Test
Measuring
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 80, Cat.No. 114738, or the
Standard solution for photometric applications, CRM,
Cat.No. 125037, 125038, and 125039.
Ready-for-use nitrate standard solution Certipur,
Cat.No. 119811, concentration 1000 mg/l NO3, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 80) is highly recommended.
Release 06/2014 - Spectroquant photometer NOVA 60
100614
Nitrate
Cell Test
Measuring
range:
Pipette 1.0 ml of
NO3-1K into a reaction
cell, do not mix.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use nitrate standard solution Certipur, Cat.No. 119811, concentration
1000 mg/l NO3, can be used after diluting accordingly as
well as the Standard solution for photometric applications,
CRM, Cat.No. 125039 and 125040.
114773
Nitrate
Test
Measuring
range:
Reaction time:
10 minutes
Note:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10 and 20, Cat.No. 114676
and 114675, or the Standard solution for photometric
applications, CRM, Cat.No. 125036, 125037, and 125038.
Ready-for-use nitrate standard solution Certipur,
Cat.No. 119811, concentration 1000 mg/l NO3, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck) is highly recommended.
109713
Nitrate
Test
Measuring
10-mm cell
range:
20-mm cell
0.4
50-mm cell
Reaction time:
10 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10 and 20, Cat.No. 114676
and 114675, or the Standard solution for photometric
applications, CRM, Cat.No. 125036, 125037, and 125038.
Note:
Empty cells with screw caps, Cat.No. 114724 are
recommended for the preparation. These cells can be
sealed with the screw caps, thus enabling a hazard-free
mixing of the sample.
Release 06/2014 - Spectroquant photometer NOVA 60
Measuring
range:
Nitrate
114556
in seawater
Cell Test
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10, Cat.No. 114676, or the
Standard solution for photometric applications, CRM,
Cat.No. 125036 and 125037.
Ready-for-use nitrate standard solution Certipur,
Cat.No. 119811, concentration 1000 mg/l NO3, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
Release 06/2014 - Spectroquant photometer NOVA 60
Nitrate
114942
in seawater
Test
Measuring
range:
10-mm cell
Immediately add
1.5 ml of NO3-2 with
pipette.
Reaction time:
15 minutes
Reaction time:
60 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 20, Cat.No. 114675, or the
Standard solution for photometric applications, CRM,
Cat.No. 125036, 125037, and 125038.
Ready-for-use nitrate standard solution Certipur,
Cat.No. 119811, concentration 1000 mg/l NO3, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 20) is highly recommended.
101842
Nitrate
Test
Measuring
range:
Pipette 10 ml of the
sample into a test tube
(Flat-bottomed tubes,
Cat.No. 114902).
50-mm cell
Reaction time:
5 minutes, measure
immediately.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a ready-for-use
nitrate standard solution Certipur, Cat.No. 119811,
concentration 1000 mg/l NO3-, can be used after diluting
accordingly.
114547
Nitrite
Cell Test
Measuring
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use nitrite standard solution Certipur, Cat.No. 119899, concentration
1000 mg/l NO2, can be used after diluting accordingly as
well as the Standard solution for photometric applications,
CRM, Cat.No. 125041.
100609
Nitrite
Cell Test
Measuring
range:
Reaction time:
20 minutes, measure
immediately.
Do not shake or swirl
the cell before the
measurement.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use nitrite standard solution Certipur, Cat.No. 119899, concentration
1000 mg/l NO2, can be used after diluting accordingly as
well as the Standard solution for photometric applications,
CRM, Cat.No. 125042.
114776
Nitrite
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Reaction time:
10 minutes
Shake vigorously to
dissolve the solid
substance.
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use nitrite standard solution Certipur, Cat.No. 119899, concentration
1000 mg/l NO2, can be used after diluting accordingly as
well as the Standard solution for photometric applications,
CRM, Cat.No. 125041.
Nitrogen (total)
114537
Cell Test
Measuring
range:
Pipette 10 ml of the
sample into an empty
round cell (Empty cells,
Cat.No. 114724).
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 50, Cat.No. 114695, or the
Standard solution for photometric applications, CRM,
Cat.No. 125043 and 125044.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 50) is highly recommended.
Nitrogen (total)
100613
Cell Test
Measuring
range:
Pipette 10 ml of the
sample into an empty
round cell (Empty cells,
Cat.No. 114724).
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 50, Cat.No. 114695, or the
Standard solution for photometric applications, CRM,
Cat.No. 125043 and 125044.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 50) is highly recommended.
Nitrogen (total)
114763
Cell Test
Measuring
10 150 mg/l N
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 70, Cat.No. 114689, or the
Standard solution for photometric applications, CRM,
Cat.No. 125044 and 125045.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 70) is highly recommended.
114694
Oxygen
Cell Test
Measuring
range:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a oxygen standard solution
must be prepared (application see the website).
Oxygen Scavengers
119251
Test
20-mm cell
* N,N-diethylenhydroxylamine
0.027 0.666 mg/l Carbohy*
20-mm cell
* carbohydrazide
0.05 1.31 mg/l Hydro*
20-mm cell
* hydroquinone
0.08 1.95 mg/l ISA*
20-mm cell
* isoascorbic acid
0.09 2.17 mg/l MEKO*
20-mm cell
* methylethylketoxime
Pipette 10 ml of the
Add 1 powder pack of
sample into into an
Oxyscav 1 and close
empty round cell (Empty with the screw cap.
cells, Cat.No. 114724).
Add 0.20 ml of
Oxyscav 2 with pipette,
close with the screw
cap, and mix.
Reaction time:
10 minutes, protect
from light in the process, measure immediately.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a oxygen scavengers standard solution must be prepared from N,N-diethylhydroxylamine, Cat.No. 818473 (see section Standard solutions).
100607
Ozone
Test
Measuring
0.05 4.00
mg/l O3
10-mm cell
range:
0.02 2.00
mg/l O3
20-mm cell
50-mm cell
Pipette 10 ml of the
sample into a test tube.
Reaction time:
1 minute
Shake vigorously to
dissolve the solid
substance.
Important:
Quality assurance:
Application
10-mm cell
Add 0.20 ml of
reagent 2 with pipette,
close the cell with the
screw cap, and mix.
Aspirate the organicTransfer the dried soclear upper phase from lution into a rectangular
the tube with pipette and cell.
dry over sodium sulfate
anhydrous (Cat.No.
106649).
Note:
Important:
101744
pH
Cell Test
Measuring range:
Pipette 10 ml of the
sample into a round cell.
pH 6.4 8.8
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) buffer solution
pH 7.00 Certipur, Cat.No. 109407, can be used.
114551
Phenol
Cell Test
Measuring
range:
Pipette 10 ml of the
sample into a reaction
cell, close with the screw
cap, and mix.
Reaction time:
1 minute
Important:
Quality assurance:
100856
Phenol
Test
Measuring
20-mm cell
range:
Attention!
The measurement is carried out in a 20-mm rectangular cell against a blank, prepared from distilled water
(Water for analysis EMSURE, Cat.No. 116754, is recommended) and the reagents in an analogous manner.
Reaction time:
30 minutes
(protected from light)
Add 10 ml of chloroform
with pipette, close separation funnel.
100856
Phenol
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Pipette 10 ml of the
sample into a test tube.
Shake vigorously
to dissolve the solid
substance.
Shake vigorously
to dissolve the solid
substance.
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a phenole standard
solution must be prepared from Phenol GR, Cat.No.
100206 (see section Standard solutions).
Phosphate
100474
Determination of orthophosphate
Cell Test
Measuring
range:
Reaction time:
5 minutes
Important:
Quality assurance:
Phosphate
114543
Determination of orthophosphate
Cell Test
Measuring
range:
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) we recommended
to use Spectroquant CombiCheck 10, Cat.No. 114676.
Ready-for-use phosphate standard solution Certipur,
Cat.No. 119898, concentration 1000 mg/l PO43, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
114543
Phosphate
Determination of total phosphorus
Cell Test
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10, Cat.No. 114676, or the
Standard solution for photometric applications, CRM,
Cat.No. 125046 and 125047.
Ready-for-use phosphate standard solution Certipur,
Cat.No. 119898, concentration 1000 mg/l PO43, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
Phosphate
100475
Determination of orthophosphate
Cell Test
Measuring
range:
Reaction time:
5 minutes
Important:
Quality assurance:
Phosphate
114729
Determination of orthophosphate
Cell Test
Measuring
range:
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) we recommended
to use Spectroquant CombiCheck 20 and 80, Cat.Nos.
114675 and 114738.
Ready-for-use phosphate standard solution Certipur,
Cat.No. 119898, concentration 1000 mg/l PO43, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck) is highly recommended.
114729
Phosphate
Determination of total phosphorus
Cell Test
range:
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 20 and 80, Cat.No. 114675
and 114738, or as well as the Standard solution for photometric applications, CRM, Cat.No. 125047 and 125048.
Ready-for-use phosphate standard solution Certipur,
Cat.No. 119898, concentration 1000 mg/l PO43, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck) is highly recommended.
Measuring
range:
Phosphate
100616
Determination of orthophosphate
Cell Test
229
mg/l P2O5
Reaction time:
5 minutes
Important:
Quality assurance:
Measuring
range:
Phosphate
100673
Determination of orthophosphate
Cell Test
229
mg/l P2O5
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
phosphate standard solution Certipur, Cat.No. 119898,
concentration 1000 mg/l PO43, can be used after diluting
accordingly.
100673
Phosphate
Determination of total phosphorus
Cell Test
229
mg/l P2O5
Expression of results also possible in mmol/l and also in P total ( P), and P org* [P(o)].
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use phosphate
standard solution Certipur, Cat.No. 119898, concentration 1000 mg/l PO43, can be used after diluting accordingly as well as the Standard solution for photometric applications, CRM, Cat.No. 125047, 125048, and 125049.
Phosphate
114848
Determination of orthophosphate
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Reaction time:
5 minutes
Shake vigorously to
dissolve the solid
substance.
Important:
Quality assurance:
Phosphate
100798
Determination of orthophosphate
Test
Measuring
range:
Shake vigorously to
dissolve the solid
substance.
Reaction time:
5 minutes
10-mm cell
Important:
Quality assurance:
Phosphate
114546
Determination of orthophosphate
Cell Test
Measuring
range:
Important:
Quality assurance:
Phosphate
114842
Determination of orthophosphate
Test
Measuring
10-mm cell
range:
20-mm cell
Add 1.2 ml of PO4-1 with Transfer the solution into Select method with
piette and mix.
a corresponding cell.
AutoSelector.
Important:
Quality assurance:
Application
Measuring range:
10-mm cell
Attention!
The measurement is carried out at 690 nm in a 10-mm rectangular cell against a blank, prepared
from distilled water (Water for analysis EMSURE, Cat.No. 116754, is recommended) and the
reagents in an analogous manner.
Add 0.50 ml of
reagent 2 with pipette,
close the cell with the
screw cap, and mix.
Note:
Important:
114562
Potassium
Cell Test
Measuring
range:
Add 1 level blue microShake the cell vigorousspoon of K-2K, close the ly to dissolve the solid
cell with the screw cap.
substance.
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
potassium standard solution Certipur, Cat.No. 170230,
concentration 1000 mg/l K, can be used after diluting
accordingly.
100615
Potassium
Cell Test
Measuring
30 300 mg/l K
range:
Add 1 level blue microShake the cell vigorousspoon of K-2K, close the ly to dissolve the solid
cell with the screw cap.
substance.
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
potassium standard solution Certipur, Cat.No. 170230,
concentration 1000 mg/l K, can be used after diluting
accordingly.
114683
Residual Hardness
Cell Test
Measuring
range:
Measuring
0.070 0.700 d
range:
0.087 0.874 e
0.12 1.25 f
Reaction time:
10 minutes, measure
immediately.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
calcium standard solution Certipur, Cat.No. 119778,
concentration 1000 mg/l Ca, can be used after diluting
accordingly. (Pay attention to pH value!)
114794
Test
Measuring
range:
0.1
5.00 mg/l Si
10-mm cell
20-mm cell
50-mm cell
Reaction time:
10 minutes
Reaction time:
3 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use silicon standard solution Certipur, Cat.No. 170236, concentration
1000 mg/l Si, can be used after diluting accordingly
(Attention! Do not store standard solutions in glass
vessels - see section "Standard solutions").
100857
Test
Measuring
10-mm cell
range:
11 1070
5 500
10-mm cell
mg/l SiO2
mg/l Si
Reaction time:
2 minutes
Reaction time:
2 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use silicon standard solution Certipur, Cat.No. 170236, concentration
1000 mg/l Si, can be used after diluting accordingly
(Attention! Do not store standard solutions in glass
vessels - see section "Standard solutions").
101813
Test
Measuring
range:
50-mm cell
Pipette 10 ml of the
sample into a plastic
vessel (Flat-bottomed
tubes, Cat.No. 117988).
Pipette 10 ml of distilled
water (Water Ultrapur,
Cat.No. 101262, is
recommended) into a
second plastic vessel
(Flat-bottomed tubes,
Cat.No. 117988).
(Blank)
Reaction time:
5 minutes
Reaction time:
5 minutes
Important:
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use silicon standard solution Certipur, Cat.No. 170236, concentration
1000 mg/l Si, can be used after diluting accordingly
(Attention! Do not store standard solutions in glass
vessels - see section "Standard solutions").
114831
Silver
Test
Measuring
10-mm cell
range:
20-mm cell
Reaction time:
5 minutes
Pipette 10 ml of the
sample into an empty
round cell (Empty cells,
Cat.No. 114724).
Important:
Very high silver concentrations in the sample produce turbid solutions (measurement solution
should be clear). In such cases the sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents,measurement device, and handling) readyfor-use silver standard solution Certipur, Cat.No. 119797, concentration 1000 mg/l Ag, can be
used after diluting accordingly.
Sodium
100885
in nutrient solutions
Cell Test
Measuring
10 300 mg/l Na
range:
Reaction time:
1 minute
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-for-use
chloride standard solution Certipur, Cat.No. 119897,
concentration 1000 mg/l Cl- (corresponds to 649 mg/l Na),
can be used after diluting accordingly (see section Standard solutions).
102532
Sulfate
Cell Test
Measuring
range:
Reaction time:
2 minutes, measure
immediately.
Pipette 10 ml of the
sample into a reaction
cell, close with the screw
cap, and mix.
Add 1 level green micro- Shake the cell vigorousspoon of SO4-1K, close ly to dissolve the solid
the cell with the screw
substance.
cap.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use sulfate
standard solution Certipur, Cat.No. 119813, concentration 1000 mg/l SO42-, can be used after diluting accordingly.
114548
Sulfate
Cell Test
Measuring
range:
Reaction time:
2 minutes, measure
immediately.
Add 1 level green micro- Shake the cell vigorousspoon of SO4-1K, close ly to dissolve the solid
the cell with the screw
substance.
cap.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10, Cat.No. 114676, or the
Standard solution for photometric applications, CRM,
Cat.No. 125050 and 125051.
Ready-for-use sulfate standard solution Certipur,
Cat.No. 119813, concentration 1000 mg/l SO42, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
Release 06/2014 - Spectroquant photometer NOVA 60
100617
Sulfate
Cell Test
Measuring
range:
Reaction time:
2 minutes, measure
immediately.
Add 1 level green micro- Shake the cell vigorousspoon of SO4-1K, close ly to dissolve the solid
the cell with the screw
substance.
cap.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10, Cat.No. 114676, or the
Standard solution for photometric applications, CRM,
Cat.No. 125051 and 125052.
Ready-for-use sulfate standard solution Certipur,
Cat.No. 119813, concentration 1000 mg/l SO42, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
Release 06/2014 - Spectroquant photometer NOVA 60
114564
Sulfate
Cell Test
Measuring
range:
Reaction time:
2 minutes, measure
immediately.
Add 1 level green micro- Shake the cell vigorousspoon of SO4-1K, close ly to dissolve the solid
the cell with the screw
substance.
cap.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 20, Cat.No. 114675, or the
Standard solution for photometric applications, CRM,
Cat.No. 125051, 125052 and 125053.
Ready-for-use sulfate standard solution Certipur,
Cat.No. 119813, concentration 1000 mg/l SO42, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 20) is highly recommended.
Release 06/2014 - Spectroquant photometer NOVA 60
114791
Sulfate
Test
Measuring
range:
10-mm cell
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10, Cat.No. 114676, or the
Standard solution for photometric applications, CRM,
Cat.No. 125050 and 125051.
Ready-for-use sulfate standard solution Certipur,
Cat.No. 119813, concentration 1000 mg/l SO42, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
101812
Sulfate
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Add 10 ml of the sample Add 1 level green microwith pipette and mix.
spoon of SO4-2.
Reaction time:
2 minutes, measure
immediately.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) ready-for-use sulfate
standard solution Certipur, Cat.No. 119813, concentration 1000 mg/l SO42-, can be used after diluting accordingly.
102537
Sulfate
Test
Measuring
10-mm cell
range:
Reaction time:
2 minutes, measure
immediately.
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) we recommended to use
Spectroquant CombiCheck 10, Cat.No. 114676, or the
Standard solution for photometric applications, CRM,
Cat.No. 125050 and 125051.
Ready-for-use sulfate standard solution Certipur,
Cat.No. 119813, concentration 1000 mg/l SO42, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
Release 06/2014 - Spectroquant photometer NOVA 60
114779
Sulfide
Test
Measuring
10-mm cell
range:
20-mm cell
50-mm cell
Reaction time:
1 minute
Important:
Quality assurance:
114394
Sulfite
Cell Test
Measuring
Round cell
range:
50-mm cell
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a sulfite standard
solution must be prepared from sodium sulfite GR,
Cat.No. 106657 (see section Standard solutions).
114394
Sulfite
Cell Test
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a sulfite standard
solution must be prepared from sodium sulfite GR,
Cat.No. 106657 (see section Standard solutions).
101746
Sulfite
Test
Measuring
10-mm cell
range:
10-mm cell
Shake vigorously
to dissolve the solid
substance.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a sulfite standard
solution must be prepared from sodium sulfite GR,
Cat.No. 106657 (see section Standard solutions).
Surfactants (anionic)
114697
Cell Test
Measuring
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a surfactants
standard solution must be prepared from dodecane-1sulfonic acid sodium salt GR, Cat.No. 112146 (see section
Standard solutions).
Surfactants (anionic)
102552
Cell Test
Measuring
range:
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a surfactants
standard solution must be prepared from dodecane-1sulfonic acid sodium salt GR, Cat.No. 112146 (see section
Standard solutions).
Surfactants (cationic)
101764
Cell Test
Measuring
range:
Reaction time:
5 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a surfactants
standard solution must be prepared from N-cetyl-N,N,Ntrimethylammonium bromide, Cat.No. 102342 (see section
Standard solutions).
Surfactants (nonionic)
101787
Cell Test
Measuring
range:
Reaction time:
2 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a surfactants
standard solution must be prepared from Triton X-100,
Cat.No. 112298 (see section Standard solutions).
Suspended Solids
Measuring range:
Homogenize 500 ml of
Transfer the solution into
sample for 2 minutes in a a cell.
mixer running at high
speed.
20-mm cell
114622
Tin
Cell Test
Measuring
range:
Reaction time:
15 minutes
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a tin standard solution must
be prepared from ready-for-use tin standard solution
Certipur, Cat.No. 170242, concentration 1000 mg/l Sn
(see section Standard solutions).
Measuring range:
TOC
114878
Cell Test
Place 25 ml of the
sample into a suitable
glass vessel.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a TOC standard
solution Certipur, Cat.No. 109017, concentration
1000 mg/l TOC, can be used after diluting accordingly.
Measuring range:
TOC
114879
Cell Test
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a TOC standard
solution Certipur, Cat.No. 109017, concentration
1000 mg/l TOC, can be used after diluting accordingly.
Measuring
range:
Total Hardness
100961
Cell Test
215 mg/l Ca
Measuring
0.7 30.1 d
range:
0.9 37.6 e
1.2 53.7 f
Reaction time:
3 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a freshly prepared
standard solution can be used (see section Standard
solutions).
Measuring
range:
Total Hardness
100961
Cell Test
Reaction time:
3 minutes
After pressing
enter, the individual
measuring values for
Ca- and Mg-hardness
are shown on the
display.
Press enter,
remove the cell.
Turbidity
analogous to EN ISO 7027
Measuring range:
1 100 FAU
550 nm
50-mm cell
101763
Cell Test
Measuring
range:
Add 0.50 ml of the samp- Heat the cell in the therle with pipette, close with moreactor at 100 C for
the screw cap, and mix.
10 minutes. Then cool to
room temperature under
running water.
Reaction time:
10 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a standard solution
must be prepared from sodium acetate anhydrous,
Cat.No. 106268 (see section Standard solutions).
101749
Cell Test
Measuring
range:
Reaction time:
1 minute
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a standard solution
must be prepared from sodium acetate anhydrous,
Cat.No. 106268 (see section Standard solutions).
101809
Test
Measuring
range:
Reaction time:
1 minute
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a standard solution
must be prepared from sodium acetate anhydrous,
Cat.No. 106268 (see section Standard solutions).
100861
Zinc
Cell Test
Measuring
range:
Pipette 10 ml of sample
into a glass vessel.
Reaction time:
15 minutes
Important:
Quality assurance:
114566
Zinc
Cell Test
Measuring
range:
Reaction time:
15 minutes
Important:
Quality assurance:
114832
Zinc
Test
Measuring
10-mm cell
range:
Reaction time:
3 minutes
Aspirate the clear upper Transfer the solution into Leave to stand for
3 minutes.
phase from the tube with a cell.
pipette.
Important:
For the determination of total zinc a pretreatment
with Crack Set 10C, Cat.No. 114688, or Crack Set 10,
Cat.No. 114687, and thermoreactor is necessary.
Result can be expressed as sum of zinc ( Zn).
Quality assurance:
Select method with
AutoSelector.
Cat. No.
Seawater
20
10
5
20
20
20
10
20
0.4
10
20
10
20
10
1
1
2
20
10
10
10
10
10
20
20
5
10
15
20
10
20
20
10
20
20
5
10
10
10
2
20
20
10
0.5
0.5
10
10
10
10
10
20
20
5
15
20
20
20
20
20
10
20
20
20
10
1
1
1
10
1
0.1
0.05
0.05
10
10
10
10
10
no
no
10
10
10
10
10
10
yes
10
10
10
no
yes
no
no
no
no
no
no
no
no
no
no
no
yes
yes
yes
yes
no
no
no
yes
no
no
yes
yes 2)
no
no
yes
1
10
0.4
0.4
0.4
0.4
0.4
0.4
0.4
1.0
10
0
0
35
35
15
15
10
10
10
10
10
20
0.05
5
5
10
10
10
10
10
10
10
20
10
20
10
20
10
10
10
10
15
15
10
10
10
10
10
20
0.05
0
0
20
10
10
10
10
10
10
20
10
20
10
20
10
10
10
10
15
15
10
10
10
10
10
20
0.001
10
10
5
no
yes
no
no
yes
no
20
20
0.1
10
20
5
5
20
1
10
20
5
2
20
5
10
20
5
This test kit is also suitable for testing seawater after the addition of sodium hydroxide solution (see package insert).
distill beforehand analogous APHA 4500-F- B
Release 06/2014
Na2 SO4
no
yes
yes
no
yes
yes
yes
no 1)
yes
no
no
yes
yes
no
no
no
no
no
yes
no
yes
yes
no
no
no
no
no
no
no
Cat. No.
Seawater
114761
100796
114833
109717
100815
100816
101739
114770
101846
100860
119252
101632
114554
114785
114542
114563
114764
100614
114773
109713
114556
114942
101842
114547
100609
114776
114537
100613
114763
114694
119251
100607
101744
114551
100856
100474
yes
yes
no
no
yes
no
no
yes
no
no
no
no
no
no
no
no
no
no
no
no
yes
yes
no
yes
yes
yes
no
no
no
no
no
no
yes
yes
yes
yes
114543
yes
114543
100475
10
20
20
0.4
0.2
0.5
2
0.4
0.2
20
20
0.001
20
20
20
0.5
0.2
2
10
10
20
20
5
Na2 SO4
20
20
20
5
2
20
25
20
25
20
10
20
20
20
20
20
10
20
20
10
20
20
1
15
1
20
5
20
5
5
20
20
20
20
20
20
20
20
20
20
20
0.001
15
15
15
10
10
20
1
10
15
20
10
10
10
no
yes
1
20
10
20
10
20
114729
yes
20
20
20
114729
100616
yes
yes
5
20
20
20
20
20
100673
yes
20
20
20
100673
114848
100798
114546
114842
114562
100615
114683
114794
100857
101813
114831
100885
102532
114548
100617
114564
114791
101812
102537
114779
114394
101746
114697
102552
yes
yes
yes
yes
yes
yes
yes
no
yes
no
no
no
no
no
yes
yes
yes
no
no
yes
no
no
no
no
no
20
5
15
20
20
20
20
0.01
5
5
0.5
0
2
10
10
10
0.2
2
10
0.5
20
20
0.1
0.1
20
10
20
20
20
20
20
0.01
10
10
1
1
10
0.007
0.1
0.1
0.5
0.2
0.007
0.015
1
20
20
0.01
0.01
20
10
10
20
20
20
20
0.01
5
2.5
0.2
5
1
1
20
20
10
10
Release 06/2014
Release 06/2014
Cat. No.
Seawater
Na2 SO4
101764
101787
114622
114878
114879
100961
no
no
yes
no
no
no
0.1
2
20
0.5
5
2
0.1
5
20
10
20
2
20
2
20
10
20
1
101763
no
20
20
10
101749
101809
100861
114566
114832
no
no
no
no
no
20
20
20
10
5
20
20
20
10
15
10
10
1
10
15
Evaluation
as
CombiCheck,
Cat. No.
Confidence interval
Spec. value max.
for the
working
standard
tolerance
Ready-to-use
standard
solution,
Cat. No.
OH
Al
Al
NH4-N
5.00 mmol/l*
50*
250*
0.25 mg/l*
0.75 mg/l
1.00 mg/l
NH4-N
NH4-N
NH4-N
NH4-N
NH4-N
AOX
As
O2
B
B
Br2
Cd
Cd
Ca
Ca
Ca
1.00 mg/l*
0.050 mg/l*
210 mg/l
1.00 mg/l*
0.400 mg/l*
5,00 mg/l*
0.500 mg/l
0.250 mg/l
75 mg/l*
80 mg/l*
2.00 mg/l*
0.50 mmol/l
0.03 mg/l
0.08 mg/l
0.10 mg/l
125022
125023
0.30 mg/l
125022
125023
125024
125025
1.0 mg/l
125023
125024
125025
125026
5.0 mg/l
125025
125026
125027
0.10 mg/l
125022
125023
125024
5.0 mg/l
125025
125026
125027
0.10 mg/l
0.005 mg/l
20 mg/l
0.15 mg/l
0.040 mg/l
0,50 mg/l
0.060 mg/l
0.010 mg/l
7 mg/l
8 mg/l
0.20 mg/l
Cl
10 mg/l
6 mg/l
13 mg/l
0.13 mg/l
0.8 mg/l
0.25 mg/l
0.30 mg/l
0.30 mg/l
0.30 mg/l
0.30 mg/l
0.30 mg/l
119897
Cl
60 mg/l
25 mg/l
125 mg/l
12.5 mg/l*
7.5 mg/l*
2.50 mg/l*
3.00 mg/l*
3.00 mg/l*
3.00 mg/l*
3.00 mg/l*
3.00 mg/l*
119897
119897
119897
see prep. instr.
see prep. instr.
see prep. instr.
see prep. instr.
see prep. instr.
3.00 mg/l*
0.30 mg/l
0.500 mg/l*
0.050 mg/l
0.012 mg/l
0.04 mg/l
0.012 mg/l
0.04 mg/l
0.07 mg/l
0.13 mg/l
0.04 mg/l
0.07 mg/l
0.13 mg/l
0.4 mg/l
0.13 mg/l
0.4 mg/l
1.2 mg/l
0.012 mg/l
0.04 mg/l
0.07 mg/l
0.13 mg/l
0.4 mg/l
1.2 mg/l
119812
119812
119812
119812
119812
119812
100680
119773
100718
119500
119500
see prep. instr.
119777
119777
see prep. instr.
119778
119778
3.00 mg/l*
0.30 mg/l
0.500 mg/l*
5.00 mg/l*
1.00 mg/l*
1.00 mg/l*
20.0 mg/l
20.0 mg/l
80 mg/l
250 mg/l
0.050 mg/l
0.50 mg/l
0.10 mg/l
0.10 mg/l
4.0 mg/l
2.0 mg/l
12 mg/l
20 mg/l
COD
25 mg/l
COD
75 mg/l
125028
125028
125029
125029
125030
125029
125030
125031
125029
125030
125031
125032
20.0 mg/l
20.0 mg/l
100 mg/l
100 mg/l
200 mg/l
100 mg/l
200 mg/l
400 mg/l
100 mg/l
200 mg/l
400 mg/l
1000 mg/l
0.7 mg/l
0.7 mg/l
3 mg/l
3 mg/l
4 mg/l
3 mg/l
4 mg/l
5 mg/l
3 mg/l
4 mg/l
5 mg/l
11 mg/l
Evaluation
as
CombiCheck,
Cat. No.
COD
150 mg/l
COD
400 mg/l
COD
50000 mg/l*
5000 mg/l
COD
80 mg/l*
12 mg/l
COD
750 mg/l*
75 mg/l
COD
COD
Pt/Co
(Hazen)
Pt/Co
(Hazen)
Cu
Cu
CN
CN
CN
Cyan Acid
F
F
F
30.0 mg/l*
1500 mg/l*
250 mg/l*
3.0 mg/l
150 mg/l
500 mg/l
2.00 mg/l
2.00 mg/l
0.250 mg/l*
0.250 mg/l*
0.250 mg/l*
80 mg/l*
0.75 mg/l*
0.75 mg/l*
1.00 mg/l*
10.0 mg/l*
1.00 mg/l*
5.00 mg/l*
4.50 mg/l*
6.0 mg/l*
Color Hazen
Copper Cell Test, 114553
Copper Test, 114767
Cyanide Cell Test, 102531
Cyanide Cell Test, 114561
Cyanide Test, 109701
Cyanuric Acid Test, 119253
Fluoride Cell Test, 114557
Fluoride Cell Test, 100809
Fluoride Test, 114598
Confidence interval
Spec. value max.
for the
working
standard
tolerance
400 mg/l
1000 mg/l
2000 mg/l
1000 mg/l
2000 mg/l
8000 mg/l
8000 mg/l
50 000 mg/l
20.0 mg/l
100 mg/l
100 mg/l
200 mg/l
400 mg/l
1000 mg/l
5 mg/l
11 mg/l
32 mg/l
11 mg/l
32 mg/l
68 mg/l
68 mg/l
894 mg/l
0.7 mg/l
3 mg/l
3 mg/l
4 mg/l
5 mg/l
11 mg/l
Ready-to-use
standard
solution,
Cat. No.
100246
0.20 mg/l
0.20 mg/l
0.030 mg/l
0.030 mg/l
0.030 mg/l
10 mg/l
0.08 mg/l
0.08 mg/l
0.15 mg/l
1.2 mg/l
0.15 mg/l
0.50 mg/l
0.50 mg/l
0.6 mg/l
119786
119786
119533
119533
119533
see prep. instr.
119814
119814
119814
119814
see prep. instr.
see prep. instr.
170216
1.00 mg/l*
10.0 mg/l*
0.10 mg/l
1.0 mg/l
2.00 mg/l*
5.00 mg/l*
1.00 mg/l
25.0 mg/l*
1.00 mg/l
1.00 mg/l
2.00 mg/l
2.00 mg/l
40.0 mg/l*
1.00 mg/l
1.00 mg/l*
1.00 mg/l
1.00 mg/l*
0.50 mg/l*
25.0 mg/l*
5.00 mg/l*
2.00 mg/l
2.00 mg/l
9.0 mg/l
0.20 mg/l
0.50 mg/l
0.15 mg/l
2.5 mg/l
0.15 mg/l
0.15 mg/l
0.20 mg/l
0.20 mg/l
4.0 mg/l
0.15 mg/l
0.10 mg/l
0.15 mg/l
0.10 mg/l
0.05 mg/l
2.5 mg/l
0.50 mg/l
0.20 mg/l
0.20 mg/l
0.9 mg/l
125037
125038
125037
125038
125037
125038
125039
125039
125040
125036
125037
125038
NO3-N
0.9 mg/l
NO3-N
2.5 mg/l
NO3-N
10 mg/l
NO3-N
100 mg/l*
0.9 mg/l
2.50 mg/l
15.0 mg/l
2.50 mg/l
15.0 mg/l
2.50 mg/l
15.0 mg/l
40.0 mg/l
40.0 mg/l
200 mg/l
0.500 mg/l
2.50 mg/l
15.0 mg/l
0.06 mg/l
0.4 mg/l
0.06 mg/l
0.4 mg/l
0.06 mg/l
0.4 mg/l
1.0 mg/l
1.0 mg/l
5 mg/l
0.05 mg/l
0.06 mg/l
0.4 mg/l
119811
119811
119811
119811
119811
Release 06/2014
Evaluation
as
CombiCheck,
Cat. No.
Confidence interval
Spec. value max.
for the
working
standard
tolerance
NO3-N
0.9 mg/l
NO3-N
0.25 mg/l
NO3-N
0.9 mg/l
1.5 mg/l
0.030 mg/l
5 mg/l
0.05 mg/l
0.7 mg/l
0.7 mg/l
7 mg/l
O2
DEHA
0.250 mg/l*
0.6 mg/l
0.030 mg/l
O3
pH
C6H5OH
C6H5OH
PO4-P
PO4-P
2.00 mg/l*
7.0
1.25 mg/l*
2.50 mg/l*
0.80 mg/l
0.80 mg/l
0.20 mg/l
0.2
0.13 mg/l
0.25 mg/l
0.08 mg/l
0.08 mg/l
PO4-P
PO4-P
PO4-P
PO4-P
15.0 mg/l
8.0 mg/l
15.0 mg/l
8.0 mg/l
50.0 mg/l*
50.0 mg/l*
1.0 mg/l
0.7 mg/l
1.0 mg/l
0.7 mg/l
5.0 mg/l
5.0 mg/l
PO4-P
PO4-P
PO4-P
PO4-P
K
K
Ca
0.80 mg/l
50.0 mg/l*
15.0 mg/l*
15.0 mg/l*
25.0 mg/l*
150 mg/l*
2.50 mg/l*
0.08 mg/l
5.0 mg/l
1.0 mg/l
1.0 mg/l
4.0 mg/l
15 mg/l
0.30 mg/l
SiO2
10.0 mg/l*
0.300 mg/l*
45.0 mg/l*
0.50 mg/l*
5.0 mg/l
125041
125042
125041
125043
125044
125043
125044
125044
125045
125046
125047
125047
125048
125047
125048
125049
SiO2
SiO2
Ag
Na
SO4
SO4
5.00 mg/l*
0.750 mg/l*
50.0 mg/l*
0.1000 mg/l*
1.50 mg/l*
100 mg/l*
25.0 mg/l*
CombiCheck 10, 114676 100 mg/l
0.50 mg/l
0.075 mg/l
5.0 mg/l
0.0100 mg/l
0.20 mg/l
10 mg/l
3.0 mg/l
15 mg/l
SO4
15 mg/l
SO4
75 mg/l
SO4
15 mg/l
SO4
SO4
5.00 mg/l*
CombiCheck 10, 114676 100 mg/l
0.50 mg/l
15 mg/l
S
SO3
SO3
MBAS
0.75 mg/l*
12.5 mg/l*
30.0 mg/l*
1.00 mg/l*
0.08 mg/l
1.5 mg/l
1.0 mg/l
0.20 mg/l
125050
125051
125051
125052
125051
125052
125053
125050
125051
125050
125051
MBAS
1.00 mg/l*
0.20 mg/l
0.500 mg/l
2.50 mg/l
15.0 mg/l
0.500 mg/l
2.50 mg/l
0.500 mg/l
2.50 mg/l
15.0 mg/l
0.05 mg/l
0.06 mg/l
0.4 mg/l
0.05 mg/l
0.06 mg/l
0.05 mg/l
0.06 mg/l
0.4 mg/l
Ready-to-use
standard
solution,
Cat. No.
119811
119811
119811
119811
119899
119899
119899
see prep. instr.
see prep. instr.
see prep. instr.
see the website
see prep. instr.
see prep. instr.
109407
see prep. instr.
see prep. instr.
119898
119898
119898
119898
119898
119898
119898
119898
119898
119898
170230
170230
119778
170236
170236
170236
119797
see prep. instr.
119813
40 mg/l
125 mg/l
125 mg/l
400 mg/l
125 mg/l
400 mg/l
800 mg/l
40 mg/l
125 mg/l
6 mg/l
6 mg/l
6 mg/l
20 mg/l
6 mg/l
20 mg/l
27 mg/l
6 mg/l
6 mg/l
40 mg/l
125 mg/l
6 mg/l
6 mg/l
119813
119813
119813
119813
119813
119813
see prep. instr.
see prep. instr.
see prep. instr.
see prep. instr.
see prep. instr.
Evaluation
as
CombiCheck,
Cat. No.
Confidence interval
Spec. value max.
for the
working
standard
tolerance
Ready-to-use
standard
solution,
Cat. No.
1.00 mg/l*
0.10 mg/l
4.00 mg/l*
0.40 mg/l
1.25 mg/l*
40.0 mg/l*
400 mg/l*
75 mg/l*
1500 mg/l*
0.13 mg/l
3.0 mg/l
30 mg/l
7 mg/l
80 mg/l
1500 mg/l*
80 mg/l
1500 mg/l*
80 mg/l
0.050 mg/l
0.40 mg/l
0.20 mg/l
119806
119806
119806
0.500 mg/l*
CombiCheck 40, 114692 2.00 mg/l
1.25 mg/l*
Release 06/2014
Reagents required:
1.09141.1000 Sodium hydroxide solution
0.1 mol/l
Titripur
1.16754.9010 Water for
analysis
EMSURE
Stability:
When stored in a cool place (refrigerator), the diluted investigational
solutions remain stable for one week.
Release 06/2014
Reagents required:
1.02404.0100 Potassium
iodate, volum.
standard
1.05043.0250 Potassium iodide
for analysis
EMSURE
1.09072.1000 Sulfuric acid
0.5 mol/l
Titripur
1.09136.1000 Sodium hydroxide solution
2 mol/l
Titripur
1.16754.9010 Water for
analysis
EMSURE
Reagents required:
1.02121.0500 Calcium nitrate
tetrahydrate
for analysis
EMSURE
1.16754.9010 Water for
analysis
EMSURE
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
Stability:
The standard solution of 1000 mg/l remains stable for one week. The
diluted standard solutions (investigational concentrations) remain stable
for one day.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l and the diluted standard solutions (investigational concentrations) remain stable for one day.
Note:
This is a standard solution that can be prepared particularly rapidly and
easily.
Reagents required:
1.10888.0250 Dichloroisocyanuric acid
sodium salt dihydrate GR for
analysis
1.16754.9010 Water for
analysis
EMSURE
Release 06/2014
Release 06/2014
Reagents required:
1.02404.0100 Potassium
iodate, volum.
standard
1.05043.0250 Potassium iodide
for analysis
EMSURE
1.09072.1000 Sulfuric acid
0.5 mol/l
Titripur
1.09136.1000 Sodium hydroxide solution
2 mol/l
Titripur
1.16754.9010 Water for
analysis
EMSURE
Reagents required:
1.00316.1000 Hydrochloric
acid 25 %
for analysis
EMSURE
1.05614.9025 Sodium hypochlorite solution
techn. approx.
13% active
chlorine
1.09147.1000 Sodium thiosulfate solution
0.1 mol/l
Titripur
1.05043.0250 Potassium
iodide GR for
analysis
1.05445.0500 Zinc iodidestarch solution
GR for analysis
1.16754.9010 Water for
analysis
EMSURE
Stability:
When stored in a cool place (refrigerator), a standard solution remains
stable for approx. one week. The diluted standard solutions (investigational concentrations) are stable for approx. 2 hours.
Note:
This is a standard solution that is absolutely necessary for the preparation of the monochloramine standard.
Reagents required:
1.02426.0250 Chloramine T
trihydrate GR
for analysis
1.16754.9010 Water for
analysis
EMSURE
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l and the diluted standard solutions (investigational concentrations) remain stable for one day.
Release 06/2014
Reagents required:
1.02404.0100 Potassium
iodate, volum.
standard
1.05043.0250 Potassium iodide
for analysis
EMSURE
1.09072.1000 Sulfuric acid
0.5 mol/l
Titripur
1.09136.1000 Sodium hydroxide solution
2 mol/l
Titripur
1.16754.9010 Water for
analysis
EMSURE
Stability:
The KlO3 stock solution remains stable for 4 weeks when stored in a
cool place (refrigerator). The KlO3/Kl standard solution can be used for
5 hours when stored in a cool place (refrigerator). The diluted chlorine
dioxide standard solution is not stable and must be used immediately.
Release 06/2014
Reagents required:
1.02400.0080 Potassium
hydrogen
phthalate GR
for analysis,
volum. standard
1.16754.9010 Water for
analysis
EMSURE
Reagents required:
1.02400.0080 Potassium
hydrogen
phthalate GR
for analysis,
volum. standard
1.06404.0500 Sodium chloride
for analysis
EMSURE
1.16754.9010 Water for
analysis
EMSURE
Reagents required:
8.20358.0005 Cyanuric acid
for synthesis
1.16754.9010 Water for
analysis
EMSURE
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l and the diluted standard solutions (investigational concentrations) remain stable for one day.
Release 06/2014
37.525
Release 06/2014
Reagents required:
1.04003.1000 Formaldehyde
solution min.
37% GR for
analysis
1.09099.1000 Iodine solution
0.05 mol/l
Titripur
1.09147.1000 Sodium thiosulfate solution
0.1 mol/l
Titripur
1.09137.1000 Sodium hydroxide solution
1 mol/l
Titripur
1.00716.1000 Sulfuric acid
25% for analysis
EMSURE
1.05445.0500 Zinc iodidestarch solution
GR for analysis
1.16754.9010 Water for
analysis
EMSURE
Reagents required:
1.04603.0100 Hydrazinium
sulfate GR
for analysis
1.16754.9010 Water for
analysis
EMSURE
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with oxygen-low distilled water.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l and the diluted standard solutions (investigational concentrations) remain stable for one day.
Reagents required:
1.09122.1000 Potassium
permanganate
solution
0.02 mol/l
Titripur
1.07209.0250 Perhydrol 30%
for analysis
EMSURE
1.00716.1000 Sulfuric acid
25% for analysis
EMSURE
1.16754.9010 Water for
analysis
EMSURE
Release 06/2014
Reagents required:
1.02404.0100 Potassium
iodate, volum.
standard
1.05043.0250 Potassium iodide
for analysis
EMSURE
1.09072.1000 Sulfuric acid
0.5 mol/l
Titripur
1.09136.1000 Sodium hydroxide solution
2 mol/l
Titripur
1.16754.9010 Water for
analysis
EMSURE
Stability:
The KlO3 stock solution remains stable for 4 weeks when stored in a
cool place (refrigerator). The KlO3/Kl standard solution can be used for
5 hours when stored in a cool place (refrigerator). The diluted iodine
standard solution is not stable and must be used immediately.
Release 06/2014
Reagents required:
1.05853.0500 Magnesium
nitrate hexahydrate
for analysis
EMSURE
1.16754.9010 Water for
analysis
EMSURE
Reagents required:
Chlorine standard solution
100 mg/l Cl2
Preparation see Standard solution of
free chlorine with hypochlorite solution
(standard solution that is absolutely necessary for the preparation of the
monochloramine standard)
Ammonium standard
solution 10 mg/l NH4-N
Preparation with Ammonium standard
solution Certipur,
Cat.No. 1.19812.0500, 1000 mg/l NH4 =
= 777 mg/l NH4-N
Reagents required:
1.04201.0100 Glycine GR for
analysis
1.16754.9010 Water for
analysis
EMSURE
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l remains stable for one week. The diluted standard solutions
(investigational concentrations) must be used immediately.
Reagents required:
8.18473.0050 N,N-Diethylhydroxylamine for
synthesis
1.16754.9010 Water for
analysis
EMSURE
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l and the diluted standard solutions (investigational concentrations) remain stable for one day.
Release 06/2014
Reagents required:
1.02404.0100 Potassium
iodate, volum.
standard
1.05043.0250 Potassium iodide
for analysis
EMSURE
1.09072.1000 Sulfuric acid
0.5 mol/l
Titripur
1.09136.1000 Sodium hydroxide solution
2 mol/l
Titripur
1.16754.9010 Water for
analysis
EMSURE
Stability:
The KlO3 stock solution remains stable for 4 weeks when stored in a
cool place (refrigerator). The KlO3/Kl standard solution can be used for
5 hours when stored in a cool place (refrigerator). The diluted ozone
standard solution is not stable and must be used immediately.
Release 06/2014
Reagents required:
1.00206.0250 Phenol GR for
analysis
1.16754.9010 Water for
analysis
EMSURE
Reagents required:
1.70236.0100 Silicone
standard
solution
Certipur
1.16754.9010 Water for
analysis
EMSURE
Example:
Mix 4.675 ml of silicon standard solution (1000 mg/l Si) with distilled
water in a calibrated or conformity-checked 1000-ml volumetric flask and
make up to the mark with distilled water.
The standard solution prepared according to this procedure has a concentration of 10.00 mg/l SiO2.
After its preparation, the solution must be immediately transferred to a
clean polyethylene vessel for further storage.
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
After its preparation, the solution with the desired working concentration
must be immediately transferred to a clean polyethylene vessel for
further storage.
Stability:
The diluted standard solutions (investigational concentrations) remain
stable - depending on the respective concentration - for one day to
approximately six months.
Reagents required:
1.19897.0500 Chloride
standard
solution
Certipur
1.16754.9010 Water for
analysis
EMSURE
Stability:
When stored in a cool place (refrigerator), the diluted standard solutions
(investigational concentrations) remain stable for one month.
Release 06/2014
64.13
Release 06/2014
Reagents required:
1.09099.1000
1.09147.1000
1.00716.1000
1.05445.0500
1.16754.9010
Sodium sulfide
hydrate approx.
60 % GR for
analysis
Iodine solution
0.05 mol/l
Titripur
Sodium thiosulfate solution
0.1 mol/l
Titripur
Sulfuric acid
25% for analysis
EMSURE
Zinc iodidestarch solution
GR for analysis
Water for
analysis
EMSURE
80.06
Reagents required:
1.06657.0500 Sodium sulfite
anhydrous
for analysis
EMSURE
1.08418.0100 Titriplex III GR
for analysis
1.09099.1000 Iodine solution
0.05 mol/l
Titripur
1.09147.1000 Sodium thiosulfate solution
0.1 mol/l
Titripur
1.00316.1000 Hydrochloric
acid 25 % GR
for analysis
EMSURE
1.05445.0500 Zinc iodidestarch solution
GR for analysis
1.09461.1000 Buffer solution
pH 9.00
Certipur
1.16754.9010 Water for
analysis
EMSURE
Stability:
When stored in a cool place (refrigerator), the stock solution of approx.
1000 mg/l remains stable for at most one day. The diluted standard solutions (investigational concentrations) must be used immediately.
Release 06/2014
Reagents required:
1.12146.0005 Sodium 1-dodecanesulfonate
1.16754.9010 Water for
analysis
EMSURE
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l remains stable for one month. The diluted standard solutions
(investigational concentrations) must be used immediately.
Reagents required:
1.02342.0100 N-cetyl-N,N,Ntrimethylammonium bromide
GR for analysis
1.16754.9010 Water for
analysis
EMSURE
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l remains stable for one month. The diluted standard solutions
(investigational concentrations) must be used immediately.
Release 06/2014
Reagents required:
1.12298.0101 Triton X-100
1.16754.9010 Water for
analysis
EMSURE
Reagents required:
1.70242.0100 Tin standard
solution
Certipur
1.09057.1000 Hydrochloric
acid 1 mol/l
Titripur
1.16754.9010 Water for
analysis
EMSURE
Stability:
The tin standard solution 100 mg/l remains stable for 30 minutes. The
diluted standard solutions (investigational concentrations) must be used
immediately.
Reagents required:
1.02121.0500 Calcium nitrate
tetrahydrate
for analysis
EMSURE
1.16754.9010 Water GR for
analysis
Further investigational concentrations may be prepared from this standard solution by diluting accordingly with distilled water.
Stability:
The standard solution of 1000 mg/l remains stable for one week. The
diluted standard solutions (investigational concentrations) remain stable
for one day.
Reagents required:
1.06268.0250 Sodium acetate
anhydrous
for analysis
EMSURE
1.16754.9010 Water GR for
analysis
Stability:
When stored in a cool place (refrigerator), the standard solution remains
stable for one week.
Release 06/2014