4860 DS
4860 DS
4860 DS
FEATURES
Easy handling
Suitable for all Tettex (2767, 2765, 2711/22 and
ELECTRONIC DEVICE
The main components of this system are an input
amplifier and a guard-potential control (up to 1010 V),
an electronic divider and a power amplifier with the
output stage. The instrument is protected against
setting errors, overvoltages and short-circuits. Errors
message are indicated on the dot matrix display.
Upgradeable
to an automatic
combination with a type 2767
test
system
by
1/4
C2 Type (air)
[Serie/pF]
Divider ratio CR
1 ... 1010 V*
1:1*
1 100 kV
3370 NK/100/100
3330/9900
1:100
1 100 kV
3370 NK/100/200
3330/9900
1:100
2 200 kV
3370 NK/100/200
3330/10000 + 3330/9900
1:200
1 100 kV
3370 NK/50/300
3330/4950
1:100
3 300 kV
3370 NK/50/300
3330/10000 + 3330/4950
1:300
4 400 kV
3370 NK/50/400
3330/10000 + 3330/9950
1:400
1 100 kV
3370 NK/50/400
3330/4950
1:100
4 400 kV
3370 NK/50/400
3330/10000 + 3330/9950
1:400
6 600 kV
3370 NK/33.3/600
3330/10000 + 3330/9946
1:600
1 100 kV
3370 NK/33.3/600
3330/3296
1:100
6 600 kV
3370 NK/33.3/600
1:600
8 800 kV
3370 NK/50/800
1 100 kV
3370 NK/50/800
8 800 kV
3370 NK/50/800
1 1200 kV
3370 NK/20/1200
3330/10000 + 3330/9946
3 x 3330/10000 +
3330/9950
3330/4950
3 x 3330/10000 +
3330/9950
2 x 3330/10000 +
3330/3980
Type
Measurement
range (Uprim)
4861
4861/100
4861/200
4861/300
4861/400a
4861/400b
4861/600a
4861/600b
4861/800a
4861/800b
4861/1200b
2/4
1:800
1:100
1:800
1:1200
GENERAL DATA
Power supply
Dimensions (W x H x D)
Weight
230/115 V, 50/60 Hz
Power requirement
approx. 100 VA
20 ...+ 70 C
ERROR LIMITS
TECHNICAL SPECIFICATIONS
1010 V
Input resistance
> 10 G
Input capacitance
< 0.1 pF
45 65 Hz
In operating range
Test frequency
50 or 60 Hz 1%
At reference conditions
10 nF
250 V
5 VA
0.4 A
10 nF
Output
SYSTEM PARAMETERS
Rated primary voltage PV
Direct input range
With capacitive divider
Adjustment factors
0.010 ...1.010 kV
(0.010 ... 1.010) x
1/CR [kV]
5 ... 250 V
Adjustment factors
1 : 1 ... 1 : 10
Resolution
110
IMPORTANT!
1000 V
Capacitance values
Dimensions (L x W x H)
Weight
3/4
ORDER SPECIFICATION
OPTIONS
STANDARD SUPPLY
type 4861/1
Electronic device
Type 4861
type 5993
series 3370 NK
- RS 232C interface, 3m
type 5991
Air capacitors
series 3330
- RS 232C interface
(fibre optic data-link)
type 5992
no. 014235-00
no. 014107-00
Triax/coax adapter
for compressed gas capacitor
(for C/tan measurement)
no. 008140-72
Measuring cable, 10 m
Measuring cable, 1.5 m
no. 013685-00
no. 014108-00
no. 016351-00
Power cable
Power voltage
230 or 115 V,
50/60 Hz
IMPORTANT
type 4861/2
4/4