ModuLab MTS
ModuLab MTS
the
difference
the
difference
The ModuLab XM MTS system is able to autosequence all of the above time domain and AC
techniques without switching cables, removing the
need to configure complicated systems using discrete
measurement units. The ModuLab XM MTS PC
software also controls a wide range of temperature
accessories in convenient, easy to configure
experiment sequences.
Power Curve
ottk
ch
ott S
Impe
danc
Xtreme Measurement
Control and Slave Modules
The ModuLab XM system has two intelligent Control
Modules:
Materials Core
Module
XM MAT 1MHz makes use of latest generation
high-technology hardware for accurate
waveform generation and data acquisition.
Fast auto-sequencing of I-V, ramp, P-E
and high-speed pulse techniques without
switching cables
Frequency Response
Analyzer
XM MFRA 1MHz is the most versatile
Frequency Response Analyzer available
today. It is fully compatible with all XM slave
modules for high voltage, high/low current and
sample/reference AC measurements, and is
able to auto-sequence time domain and AC
measurements without switching cables.
Single Sine Analysis
XM control modules
Measurements from highest conductivity to highest
impedance materials by appropriate choice of plug and
play slave option modules
No matter which modules are in operation, ModuLab
XM provides smooth analog voltage ramp waveforms
essential for many research applications
Selectable high-speed data acquisition up to 1 MS/s
available for all measurement techniques including
pulse and fast I-V
Xtreme Measurement
XM slave modules
Low Current
Option
Extends current measurement resolution
to 0.15 fA
Available for time domain and AC (with
FRA) tests
Combined with XM MHV100 option
provides ultra-high impedance
measurement
Sample/Reference
Option
2 A High Current
Option
XM MBST 2A extends XMs high current
to 2 Amps for accurate time-domain and
impedance tests of low impedance materials:
XM MBST 2A has low output impedance
useful for testing electronic materials in
their ON state
Combined with XM MHV100 provides
high power tests
Accessories
A wide range of accessories are available for use with the ModuLab XM MTS, including sample holders,
cryostats, furnaces and amplifiers.
Temperature Control
The 129610A cryostat provides fully integrated temperature related measurements of solid, liquid or powder materials samples over a wide range
of temperature (5 K to > 600 K) using liquid nitrogen or liquid helium cryogen. This system ensures that the sample is not exposed to the cryogen
vapor which could cause it to crack or swell, by positioning the sample in a separate chamber that heats or cools the sample via an exchange gas
(usually dry inert helium gas)
The ModuLab XM MTS is compatible with various furnace systems for very high temperature tests up to 1200 C. Other temperature systems may
be configured for special requirements.
Xtreme Measurement
Specifications
Control Module
Slave Modules
100 V High Voltage
Femto Ammeter
2 A Booster
XM MBST 2A
XM MAT 1MHz
XM MHV100
XM MFA
Single
Single
Single
Double
Measurement mode
2 or 4 terminal
2 or 4 terminal
2 or 4 terminal
2 or 4 terminal
I
MAT- Gen, VHi, VLo
Gen- I
MAT- VHi, VLo
4 x BNC to BNC
MAT cables
Triax to BNC
MAT cables
Yes
Yes
Yes
Yes
XM MAT 1MHz
XM MHV100
XM MFA
XM MBST 2A
64 MS/s
interpolated/filtered
Uses MAT
N/A
Uses MAT
8 V
100 V
N/A
8 V with MAT
20 V with MHV
150 V (< 3 V)
400 V ( 3 V)
2 mV (< 37 V)
5 mV ( 37 V)
N/A
1.5 mV
100 mA
100 mA
N/A
2 A
General
Floating
Generator Output - (Gen)
Maximum waveform generator sample rate
Maximum voltage (open-circuit load)
DC + peak AC (subject to slew rate limit)
Maximum voltage resolution
Maximum output current
Output impedance (nominal)
Applied voltage error (open-circuit load)
50
50
N/A
<1
0.2% setting
800V (< 3 V)
/ 2 mV ( 3 V)
0.2% setting
12.5 mV (< 37 V)
/ 35 mV ( 37 V)
N/A
0.2% setting +
10 mV (< 3 V)
/ 25 mV ( 3 V)
10 MV/s to 1 V/s
N/A
25 kVs
150 kV/s
N/A
25 kV/s
Uses MAT
1 s
Uses MAT
N/A
>10 V/s
>15 V/s
N/A
>10 V/s
XM MAT 1MHz
XM MHV100
XM MFA
XM MBST 2A
8 V
100 V
N/A
N/A
Voltage ranges
8, 3 (V)
300, 30, 3 mV
N/A
N/A
0.1% reading +
0.05% range +
0.1% reading +
0.05% range
N/A
N/A
100V
+ 100V
1 MS/s
N/A
N/A
N/A
Unlimited
N/A
N/A
N/A
XM MAT 1MHz
XM MHV100
XM MFA
XM MBST 2A
100mA
N/A
100mA
2 A
100 mA, 30 mA to
30 nA in decades
N/A
100 mA, 30 mA to
3 pA in decades
2 A plus MAT
ranges
1.5 pA
N/A
0.15 fA
1.5 pA (MAT)
0.1% reading +
+0.05% range +
100 pA
N/A
0.1%** reading +
+0.05% range +
30 fA
0.1% reading +
+0.05% range +
0.1 mA
XM MAT 1MHz
XM MHV100
XM MFA
XM MBST 2A
Four
N/A
N/A
N/A
8 V, 3 V to 3 mV
N/A
N/A
N/A
Maximum resolution
1 V
N/A
N/A
N/A
Yes
N/A
N/A
N/A
Voltage ranges
** MFA Femto Ammeter current measurement reading % error term is 0.2% for 300 pA range, 2% for 30 pA range and 5% for 3 pA range
Specifications XM MFRA
Frequency Response Analyzer
Impedance accuracy specification for MAT, Femto Ammeter, 2A booster and MREF
XM MFRA 1MHz
single
40 MS/s
Frequency range
10 Hz to 1 MHz
Frequency resolution
1 in 65,000,000
Frequency error
100 ppm
10 ms
Linear / logarithmic
All or selected
Accuracy (ratio)
0.1%, 0.1
Automatic
Specifications XM MREF
Sample / Reference Option
XM MREF
Single
Reference modes
Internal / External
Gen, I
*High Voltage option module provides up to 10X higher impedance than shown
*Femto ammeter and MREF is used for dielectrics (1 pF to 10 nF) and for high impedance
*2A booster is used for very low impedance measurements (sub 100 mohm)
*Faraday cage is recommended for dielectric and high impedance measurements
10 nF to 1 pF
(3 per decade)
MFRA Accuracy
*High Voltage option module provides up to 10X higher impedance than shown
*MREF is used for dielectrics (1 pF to 10 nF) and high impedance measurements
*Faraday cage is recommended for dielectric and high impedance measurements
USA
Europe
Visit our website for a complete list of our global offices and
authorized agents
solartron.info@ametek.com www.solartronanalytical.com
0714A