Ba LabSwift E 003529 - 03
Ba LabSwift E 003529 - 03
Ba LabSwift E 003529 - 03
Operating Instructions
www.novasina.com
1. Introduction ..................................................................................................................................... 4
1.1 Starting out ................................................................................................................................... 4
1.2. Important Notes ............................................................................................................................ 4
2. General Product Description ........................................................................................................... 5
2.1. System Overview ........................................................................................................................ 5
2.2. System Architecture .................................................................................................................. 6
2.3. Overview ..................................................................................................................................... 6
2.4. Standard scope of delivery ........................................................................................................ 8
2.5. Optional accessories ................................................................................................................. 8
3. Putting Into Operation ...................................................................................................................
3.1. Packaging / Installation ............................................................................................................
3.2. Putting into Operation ..............................................................................................................
3.3. Quick Introduction ....................................................................................................................
9
9
9
9
21
21
22
23
23
25
25
25
25
9. Maintenance ....................................................................................................................................
9.1. Cleaning of the instrument ........................................................................................................
9.2. Cleaning of the CM-2 aw-cell ...................................................................................................
9.3. Cleaning of the IR sensor .........................................................................................................
9.4. Periodic maintenance and test .................................................................................................
9.5. Replaceing protection filter .......................................................................................................
9.6. Changing of the CM-2 cell ........................................................................................................
26
26
27
27
27
27
29
www.novasina.com
English
INDEX
1.
Introduction
1.1 Starting out
English
www.novasina.com
www.novasina.com
English
2.
English
2.3. Overview
2.3.1. Function keys
Each function key has 3 markings. On top the measurement mode function
(standard operation). The central mark shows the configuration menu function.
The lower, green symbol shows the function, which can be selected by pressing
the key during the measurement mode (approx. 1s).
Menu mode
-
www.novasina.com
English
Menu mode
-
Menu mode
-
Actual
Symbol appears, when the actual measurement is displayed.
The symbol is diplayed flashing.
Stable
Symbol appears, when the stable measurement value is displayed.
SD rec
The logger function is activated. The SD-card must not be removed.
The symbol flashes if the SD-card is not disposable.
A LabSwift-aw is used as a stand-alone device. The power supply for the
LabSwift-aw is done by an external power cable or by an optional Lithium-Ion
battery. This allows a versatile employment with continuous operation as well
as in places without electric supply.
The LabSwift-aw stores measurement data and protocols on a SD-card.
These data may be readout and processed on a PC or Notebook by an optional
Windows based software.
www.novasina.com
English
Novasina LabSwift-aw
Aw- measurement instrument with accurate sample chamber. This instrument
stands out by its new design and its intuitive handling. Using the integrated SDcard, measurement results can be stored and processed.
Standard accessories :
Carrying case
Power supply EU or USA
3 pcs various humidity standards : SAL-T 11, 58, 84%
40 pcs standardised disposable sample dishes
Operating manual in English and German
Factory calibration certificate
SD-card 2 GB
Tension ring
5 pcs pre-filter (white)
Lithium-Ion Battery
For the LabSwift-aw device a Lithium-Ion battery is available. This allows a
mobile use of the instrument for more than 20 hours at places without electric
supply.
The battery can, if ordered, be delivered already mounted inside the LabSwiftaw or can be upgraded later by a Novasina distributor.
The later battery assembly may only be made by a Novasina distributor.
www.novasina.com
Check if the local mains voltage matches with the identification plate of the
supplied external power supply and plug it to the socket.
Plug the external power supply cable to the instrument socket on the back
of the LabSwift-aw.
Afterwards you can switch on the instrument by pushing the right function
key. A start display (self test) appears shortly and the software version is
displayed on the upper display line.
Note:
The LabSwift-aw is very efficient and requires only little electrical power. The
power supply fits with the latest norms (fulfills energy star level IV). Thats why
the instrument can be switched on permanently and kept ready for measurements
at any time. For accurate aw-value measurements the instrument should be
in a temperature equilibrium. For this reason the LabSwift-aw should be only
switched off, when no measurements are performed for a longer period.
www.novasina.com
English
English
Fill the sample dish with the sample. Make sure that the sample dish is well
filled but never above the upper rim. Do not grout in no case the sample
material. As bigger the overall surface, as faster and more accurate is the aw
detection. Put the sample dish in the measurement chamber. Close the
chamber by pushing it down until the fastener engages. You can now read the
actual value in aw / %RH or C / F on the display.
The measurement is started by pushing the Start/Stop button. When the
device is in the Autostart mode, the measurement (analysis) starts after
closing the measurement chamber. With that the analysis function starts and
the flashing display shows ANALIZING.
As long as it is flashing alternately (Analizing/ ...C) the aw-value and the
temperature are not stable. The result can be read at earliest when the
stable-value is displayed. Depending on the configuration also an acoustic
signal may sound. The analysis function can be optimised (see 5.4.).
During the analysis the display shows permanently the current aw-value and on
the 2nd display line alternately the sample temperature and stability setting. As
soon as the analysis function has terminated, the stable value is displayed.
Likewise the stability indicator arrows are displayed. Should the measurement
chamber be open or not completely closed, the analysis function will either not
start or will be interrupted immediately.
10
www.novasina.com
English
MODE functions :
S
A
F
0
Stability time
The stability is shown, when during the stability time the aw-value variation is
lower than +/- 0.001 aw.
Note :
When the measurement is started the last set stability parameter is taken over.
Note:
Thanks to the use of modern technology, the power consumption is extremely
low. Therefore we recommend to always keep the LabSwift-aw switched on
when connected to the power supply.
www.novasina.com
11
4.
English
Note:
The instrument can be configured, in order to start ANALYZE
automatically after closing the cover. See also chapter 4.2.2.
Simultaneously with the beginning of the measurement also the LOG function
on the SD card get started.
o
Note:
An ongoing measurement is stopped together with the LOG function as soon
as the measurement chamber is opened.
12
www.novasina.com
Button [Menu] (push button for a longer time): Switch off the device
o
With this function all SD-card data are closed and the system
is shut down.
Note:
The instrument should not be switched off by just pulling out the power plug,
as otherwise problems with the SD-card (file system) may happen.
Note :
Do not take out the SD-card whilst the SD symbol is displayed.
www.novasina.com
13
English
English
MODE
S (slow) :
Mode for samples with slow humidity exchange. Stability time is fixed at 6 min
setting.
A (average) :
Mode for samples with average humidity exchange. Stability time is fixed at
4 min setting.
F (fast) :
Mode for samples with fast humidity exchange. Stability time is fixed at 2 min
setting.
0
Mode for manual stability factor setting. The observation time can be set
between 1 and 30 min. under the submenu OSBTIME.
No mode defined. When the measurement starts the user is asked about the
mode to be used.
Note:
If the Autostart function is activated, the user is asked about the wanted
stability mode immediately after closing the measurement chamber.
OBSTIME
If mode 0 is selected, the stability time can be set manually. The Observation
time time range is 1.... 30 min.
AUTOST
If the Autostart function is selected, the measurement starts immediately after
closing the measurement chamber. No other button has to be pushed. The
lastly used stability mode is kept.
Exception see Note under Mode .
BEEPDUR
Setting of the duration of the acoustic signal once the stability has been
reached.
Acoustig signal (beep) at stability :
Duration of the acoustic signal (beep) : 0...10 min; 0 = OFF
14
www.novasina.com
EXIT
CAL XX
The deviation between the actual measurement value and the chosen reference
value (=calibration point) is displayed. The instrument defines automatically
the used humidity reference as soon as the menu is activated. The temperature
influence on the humidity reference is also taken into account. With the buttons
[up] or [down] other calibration points can be selected. If a value has not been
calibrated, the display flashes. If then Enter is pressed, the value is calibrated
after a query.
Note:
If a password is set, it will be asked prior to the calibration.
CAL CLR
Under this menu point single or all calibration points can be cancelled:
CLR Cxx
ALL
no
SET PSW
It is possible to protect the calibration values stored on the sensor by a 4-digit
number code.
With setting 0 0 0 0 the password is deactivated.
Attention:
A forgotten password can only be reset by Novasina!
Factory sensor setting: 8808
EXIT
Quit the submenu Calibration
www.novasina.com
15
English
CONTRA
For an optimal display reading the contrast can be adapted individually.
LC-Display contrast:
0....9
UNIT
The LabSwift-aw can display the measured values (results) in various units.
UNIT H
Humidity / AW: aw-value or relative humidity in % RH
UNIT T
Temperature:
C or F
EXIT
Quit the submenu Display settings
S NUMB
Free number setting (format 0000-9999) as start number and for generation of
a data number in order to match the data to the correspondent measurement.
After each measurement this number is increased by 1.
S INT
The data storage interval on the SD-card can be selected in minutes and
seconds (0:00). For standard measurements and analysis with the Novalog
MC software a 10 seconds (0:10) interval is recommended. For data import to
Excel, longer interval periods might be wiser in order to reduce the data
quantity.
EXIT
Quits the submenu SD-card
www.novasina.com
5.
Note :
The warm-up time mainly depends on the present humidity and the ambient
temperature. Generally this time is 2 to 3 minutes.
During the warm-up stage the measurement can be already started. But
there will be no output of measurement data until the warm-up is finished.
Fill the sample cup with the material to be measured. The sample cup must
be completely clean and dry (rub dry with tissue paper if necessary)!
Ideally, always use a fresh cup and dispose of it after use.
Important!
Fill the sample cup to approximately 2/3 with the product to be measured. Do
not overfill the sample cups ! Bulkier material must be suitably crushed or
manually cut into small pieces. Products consisting of several layers, or an
outer coating, must also be crushed (e.g. confectionery products). Prolonged
hand contact with the product must be avoided (falsification of measured
value!). A grinder, if used, must not warm up the product.
While placing the sample in the measuring chamber, open the cover and
place the sample dish carefully inside. Be aware about powder samples
that the product will not contact the measuring head or measuring cell.
Note:
Do not move the instrument while a sample is in it, to avoid overflow.
www.novasina.com
17
Important!
English
Tip :
If the sample contains high concentrations of glycol, glycerine or different
acids, it would make sense to let open the measuring chamber during hours.
In this way the sensor can recover.
18
www.novasina.com
Tip :
English
Let the LabSwift-aw running all the time. In this way you safe time during the
warm-up of a restart. The energy consumption is very low.
Sensors that are not used should be stored in a dust free, neutral
atmosphere at room temperature and ambient humidity.
Never open or take the measuring cell apart or clean the sensor with any
chemical or mechanical parts! The sensor itself cannot be cleaned. All
attempt will expire the warranty.
www.novasina.com
19
"Actual" protocol
"Stable" protocol
20
www.novasina.com
In the menu item STAB you can either select the fix mode S, A
and F or a selectable stability time with the mode O.
English
Some experienced data of the factors can be found in the table below :
Examples of products
fruits, jams, tomato sauce
chocolate, salmon, pharmaceuticals
dried meat, bakery products
cheese
butter, diary products
Stability setting
F
A
S
S
S
6. Calibration
6.1. Factory Calibration, Delivery Status
Every LabSwift-aw , as well as every new replacement sensor CM-2 is
delivered already factory calibrated at 6 points: 0.11, 0.33, 0.58, 0.75, 0.84 and
0.90 with a factory certificate. A LabSwift-aw has its full accuracy after the
exchange of a new sensor, without calibration by the customer. Nevertheless
for the best performance of your instrument we recommend you to verify the
sensor after the mounting. For the verification and a possible calibration of the
LabSwift-aw there are 3 SAL-T salt tablets included. If required, up to 3
additional SAL-T standards are available as accessories. These multiple-use
standards generate defined and reproducible humidity- respectively aw- values.
Please consult the enclosed data sheet for the correct handling and the visual
self-control of the standards. Novasina recommends a periodic verification of
the system and if necessary a recalibration. You dont need to calibrate any
point of the instrument over the whole aw-range each time. If you measure your
product only in small ranges, it will be enough to control the instrument within
the measurement range (minimum 2 -3 aw-values).
www.novasina.com
21
Note :
English
Define a standard quality process for your aw-measurements and start first
with shorter verification intervals. Afterwards you increase it according to the
results and deviation. Possible deviations occur by contamination, pollution,
vibration or general aging of the sensor.
Generally a recalibration at several aw-value points is simple to handle thanks
to the unique Novasina SAL-T humidity standards.
Please place always at first the SAL-T 75 or SAL-T 58 into the chamber. After
closing the measuring chamber the analysis starts by pushing the function key
Start/Stop.Set the stability parameter to 5 minutes (MODE S or MODE 5)
or wait for at least 45 minutes before you carry out a calibration at this point.
This time is necessary to reach the equilibrium in humidity and get a high
precision. Afterwards you can proceed:
The calibration function is seleceted in the menu: * CALIB -> CAL xx.
The system automatically shows you the reference value, which is nearby the
calibration point. Check if the displayed value effectively corresponds to the
inserted SAL-T salt in the chamber. If necessary use the button [up] or [down]
to select the correct reference.
22
www.novasina.com
Novasina offers additional SAL-T humidity standards which are not included
in the delivery. Further information are provided by your official Novasina
dealer or on www.novasina.com .
Factory setting
The measuring sensor can be protected under the menu point SET PSW by
a 4-digit numerical code. This code is directly stored on the sensor.
If the password is set, it will be required before the changing of the calibration
data or for the modification (clear) of the calibration values.
To deactivate the password protection the settings can be set on 0000.
TIP:
Store the password on a secured place!
Attention :
www.novasina.com
23
English
We recommend to start a new calibration with the SAL-T 75%. After that go
up to 84% and 90% and than forward with 58%, 33% and finally11%. This
procedure is reliable.
English
Note:
Tip :
By a periodic formatting of the SD-card on a PC, the card will get the full
capacity again.
Sample number
Note:
The logger will be automatically stopped after opening the measuring chamber,
after the storage or more than 60000 data sets (Excel capacity). The removing
of the SD-card will also stop the logger.
24
www.novasina.com
Protocol-files :
Stable and actual protocols are saved as XXXXPRO.txt file. The stable
protocol is generated automatically as soon as the stability is reached. If
required the user can generate additional (actual) protocols. For that purpose
the [Start/Stop]-key has to be pressed longer time, whereby the actual protocol
is generated.
The last lines of the protocol contain an electronic signature by what the PC
program NovaLog MC is able to prove the authenticity of the protocol. A
maximum of 255 protocols can be output in the same protocol file!
Data format :
Detailed information concerning the data format are available on demand at
Novasina AG.
8.
The battery
To the LabSwift-aw instrument there is a Lithium-Ion battery available. This
allows the operation of the instrument at places without power supply for over
20 hours.
8.1. Operation
If the battery is low the information LOW BATT is displayed and the
instrument gives 5 short beep. In this state there is normally enough time to
finish the current measurement. Certainly the charging can be directly started
by connecting the line adapter.
Is the battery totally discharged, the message LOW BATT will be displayed
for 2 seconds followed by a permanent beep and the instrument is switched off.
Note:
Do not let the instrument in this status (with discharged battery) standing
around for longer time. The battery might be damaged.
8.2. Charging
The charging can be done while the instrument is switched on or off. Ideally the
charging is made when the instrurment is switched off to prevent a heating of
the chamber temperature. The charge of battery is indicated with a 2-colour
LED on the backside of the LabSwift-aw :
red
green
www.novasina.com
25
English
LOG-files:
Akku LED
Stays the LED switched-off during the charging of the battery, there could be
the following reasons:
English
Warning :
!! By ignoring one of the remarks below there can
be the risk of fire or explosion !!
9. Maintenance
9.1. Cleaning of the instrument
Use for the cleaning of the whole instrument a soft and slightly humid cloth. If
it is necessary to clean the table of the measuring chamber with a damp cloth,
please consider the kind of you cleaning agent. It should not contain any
aggressive substances. If possible use water as cleaining agent. Air the
measurement chamber afterwards. For the other parts of the instrument you
can use cleaning cloth as available on the market. Make sure that no liquid
goes inside the instrument. Also the connectors on the backside has to be clear
of wetness.
Caution !
Before cleaning, switch off the LabSwift-aw and disconnect it from the
mains.
26
www.novasina.com
Please do NEVER clean the sensor CM-2 itself. The sensor is a very
sensitive element with a chemical filter protection on the top. Every
cleaning with chemical or mechanical means (touch, blow by air pressure, etc.) will destroy the measuring sensor inevitably. Possible warranty
demands expire instantly.
Verify the instrument monthly at the beginning. Increase the interval step
by step dependent on the correspondent results.
Check before and after the measurement of problematic samples the awrange with the nearest SAL-T.
www.novasina.com
27
English
Novasina provides filters for the protection of the measuring sensor against
alcohol, aromes, acids, base, etc. Please read our special information sheet
or contact your local Novasina dealer.
English
2 ppm
2 ppm
3 ppm
300 ppm
eVC-21 filter :
Chemical filter : protects against hydrogen peroxide, acid gases, diluted
acetic and formic acid, sulfur dioxide, chlorine.
eVC-26 filter :
Chemical filter : protects against a multitude or aggressive substances,
e.g. nitrogen oxides, amines, aldehyde solvents, aromatic hydrocarbons,
oil vapours, fine dust particles.
28
www.novasina.com
Switch off the instrument and disconnect the mains from the system.
Open the instrument cover and remove the 4 screws of the cover plate.
The cover plate is still connected to the housing cover by a ribbon cable.
Do not unfix it!
Afterwards you can unplug the connecting PCB. Therfore hold the PCB at
the long side and lift it carefully.
Remove now the two screws, which fix the actual sensor in the measuring
head and lift the sensor unit completely out of the head.
Tip:
Exchange at the same time the white sinter filter in front of the sensor. Press
the sinter filter downwards and insert a new one.
Insert now the new purchased measuring sensor in the same place and fix
it with the screws.
Warning !
-
Never press with fingers or any other item on the top of the measuring cell,
this might damage the sensor protection filter and leads to the damage of
the sensor
Always remove the measuring sensor during the exchange of the sinter
filter. Otherwise there may be the risk that the sensor get dampaged
during the removal or mounting of the filter.
www.novasina.com
29
English
Note:
General :
Supply
Line adapter*
:
:
:
:
:
Permissible ambient :
Operating temperature
Humidity range
Dimensions (instrument)
Weight (instrument)
Protection class
Temperature measurement :
Measuring principle
Measuring range
Accuracy
Resolution
:
:
:
:
Humidity measurement :
Measuring principle
:
Measuring range
:
Measuring accuracy after :
5-point-calibration
Resolution
:
Repeatability
:
Resistive-elektrolytic sensor
0.003 ... 1.00 aw
+/- 0.01aw (within the calibration range)
between 15 ... 30 C
0.001aw (0.1%rh)
+/- 0.003 aw
SD-card :
Type
File system
:
:
SD / SDHC
FAT-16 (recommended) / FAT-32
:
:
:
:
:
Display :
Type
Dimensions
Battery (Ooption) :
Type
Charging time
Running time
Instrument standards :
Tested CE standards
30
www.novasina.com
10.2.
Bez. des
Standards
Novasina
Typ
EU Gef.
Klasse
Farbe
des
Chemische
Symbole
Salz
Literatur
Referenz
15C
20C
25C
30C
SAL-T / 4
SC-4
weiss
CsF (>25%)
4.3
3.8
3.4
3.0
SAL-T / 6
SC-6
Xn
weiss
LiBr
6.9
6.6
6.4
6.2
SAL-T / 11
SC-11
Xn
weiss
LiCl
11.3
11.3
11.3
11.3
SAL-T / 33
SC-33
--
blau
MgCl2-6H2O
33.3
33.1
32.8
32.4
SAL-T / 53
SC-53
grn
54.4
52.9
51.4
SAL-T / 75
SC-75
--
violett
NaCl
75.6
75.5
75.3
75.1
A/B
SAL-T / 90
SC-90
weiss
Ba(Cl)2-2H2O
90.9
90.5
90.1
89.9
SAL-T / 97
SC-97
--
gelb
K2SO4 (>20%)
97.9
97.6
97.3
97.0
SAL-T / 98
SC-98
T+/N
orange
K2Cr2O7
98.5
98.2
98.0
98.0
Lit :
11. Troubleshooting
11.1. What is wrong when ....?
www.novasina.com
31
Please write down the error message carefully. Switch the instrument off and
restart. If the error message recurs, please follow the instructions on the table
and/or contact your Novasina representative
(www.novasina.com)
Error report
Error description
Needed action
DEW
ERROR
MCARD ERROR
PASSW ERROR
RANGE ERROR
SPACE ERROR
TEMP
ERROR
XX SENSOR
ERROR
NO SENS
FATAL
32
www.novasina.com
English
www.novasina.com
33
Notizen/Notes/Notes
English
34
www.novasina.com
English
Notizen/Notes/Notes
www.novasina.com
35
The Novasina team would like to thank you for your confidence and hope you will enjoy
and obtain successful measurements with the new LabSwift-aw for high performance
quality measurement of the water activity and its manifold options. You have now a
powerful water activity meter with outstanding options, simple handling and best
reproducebility for your special application in quality measurements.
In case of any further questions, Novasina and its agencies will be happy to advise you at
any time.
your Novasina team
Manufacturer :
Novasina AG, Neuheimstrasse 12, 8853 Lachen, Switzerland
Telephone +41-55-642-67-67, Fax +41-55-642-67-70, e-mail: info@novasina.ch, www.novasina.com