Isolating Transformer ES710 - E: Single-Phase Isolating Transformers For The Design of Medical IT Systems
Isolating Transformer ES710 - E: Single-Phase Isolating Transformers For The Design of Medical IT Systems
Isolating Transformer ES710 - E: Single-Phase Isolating Transformers For The Design of Medical IT Systems
ES710-E_D00151_02_D_XXEN/12.2015
Device features
Built-in temperature sensors acc. to
DIN 44081 (120 C)
Screen winding with brought-out
insulated connection terminal
Insulated mounting angles
Degree of protection, IP00
Protection class I
Reinforced insulation
Rated power
According to DIN VDE 0100-710 (VDE 0100-710), the rated power of the transformer shall
not be less than 3.15 kVA and shall not exceed 8 kVA.
Overload protection
When isolating transformers are used to form a medical IT system in accordance with
para. 710.512.1.6.2 of DIN VDE 0100-710 (VDE 0100-710), overload protection is not
permitted. In this case, short-circuit protection is required. That means, emphasis is
focused on the availability of the power supply; it is therefore essential to avoid disconnection on the occurrence of transient overload. The protection of isolating transformers
against overload and overtemperature can be realized by using monitoring devices in
accordance with para. 710.531.3.1. The appropriate fuses for short-circuit protection can
be selected from the table Technical data.
Standards
ES710-E isolating transformers comply with the device standards and the regulations
for installation: DIN EN 61558-1 (VDE 570-1), IEC 61558-1, DIN VDE 0100-710 (VDE 0100-710),
DIN EN 61558-2-15 (VDE 0570-2-15), IEC 61558-2-15, IEC 60364-7-710.
2 ES710-E_D00151_02_D_XXEN/12.2015
Nameplate
ES: single-phase isolating transformer
Nominal power
ES710/3150S-E
without = standard
S
= terminals on top side end
Wiring diagram
Terminal diagram
PA
ST
Br
Br
ST
FE
max. measure F
Connection properties
Type
Input terminals
flexible/rigid
Screen winding
flexible/rigid
Control terminals
flexible/rigid
Output terminals
flexible/rigid
ES710/3150-E
ES710/4000-E
ES710/5000-E
ES710/6300-E
ES710/8000-E
ES710/10000-E
16/25 mm
16/25 mm
16/25 mm
16/25 mm
16/25 mm
35/35 mm
16/25 mm
16/25 mm
16/25 mm
16/25 mm
16/25 mm
35/35 mm
4/6 mm
4/6 mm
4/6 mm
4/6 mm
4/6 mm
4/6 mm
16/25 mm
16/25 mm
16/25 mm
16/25 mm
16/25 mm
35/35 mm
ES710-E_D00151_02_D_XXEN/12.2015 3
S series
Dimension E is the depth incl. terminals
E F
F B
Ordering information
S series
Standard
Dimensions (mm)
A
240
280
280
280
280
320
280
280
280
280
280
320
B
230
200
210
225
240
280
180
150
160
175
190
233
C
325
370
370
370
370
420
370
420
420
420
420
440
D
200
240
240
240
240
270
240
240
240
240
240
270
E
180
150
160
175
190
330
290
290
290
290
290
333
4 ES710-E_D00151_02_D_XXEN/12.2015
F
145
115
125
140
155
193
145
115
125
140
155
193
G
11 x 28
11 x 28
11 x 28
11 x 28
11 x 28
13 x 18
11 x 28
11 x 28
11 x 28
11 x 28
11 x 28
13 x 18
Cu weight
(kg)
Weight
(kg)
Core
U/I
Type
Art. No.
11
14
16
18
24
30
11
14
16
18
24
30
42.5
52
62
62
75
85
42.5
52
62
62
75
85
180/78
210/63
210/73
210/88
210/103
240/83
180/93
210/63
210/73
210/88
210/103
240/83
ES710/3150-E
ES710/4000-E
ES710/5000-E
ES710/6300-E
ES710/8000-E
ES710/10000-E
ES710/3150S-E
ES710/4000S-E
ES710/5000S-E
ES710/6300S-E
ES710/8000S-E
ES710/10000S-E
B92090101
B92090102
B92090103
B92090104
B92090105
B92090106
B92090111
B92090112
B92090113
B92090114
B92090115
B92090116
Technical data
Type
Power/voltages/currents
Rated power
Rated frequency
Rated input voltage
Rated input current
Rated output voltage
Rated output current
Inrush current IE
Leakage current
No-load input current i0
No-load output voltage U0
Short-circuit voltage Uk ca.
3150 VA
5060 Hz
AC 230 V
14,2 A
AC 230/115 V
13.7 A
< 12 x n
0,5 mA
3 %
236 V
4.4 %
Environmental conditions
Ambient temperature
40 C
No-load temperature rise
22 C
Full-load temperature rise
55 C
Noise level (under no-load conditions and nominal load) 35 dB(A)
Other
Insulation classification
Degree of protection
Protection class
Core U/I
Recommended use when
used in accordance with DIN VDE 0100-710
Induction
Rprimary
Rsecondary
Efficiency
Documentation number: D00151
Loss
Fe loss (iron loss)
Cu loss (copper loss)
Heat dissipation loss
4000 VA
5060 Hz
AC 230 V
18 A
AC 230/115 V
17.4 A
< 12 x n
0,5 mA
3 %
233 V
4 %
5000 VA
5060 Hz
AC 230 V
22,5 A
AC 230/115 V
21.7 A
< 12 x n
0,5 mA
3 %
234 V
3.7 %
6300 VA
5060 Hz
AC 230 V
28,5 A
AC 230/115 V
27.4 A
< 12 x n
0,5 mA
3 %
235 V
2.8 %
8000 VA
5060 Hz
AC 230 V
36 A
AC 230/115 V
34.7 A
< 12 x n
0,5 mA
2,8 %
233 V
2.7 %
10000 VA
5060 Hz
AC 230 V
45.2 A
AC 230/115 V
43.5 A
< 12 x n
0,5 mA
3%
237 V
3.3 %
40 C
22 C
53 C
35 dB(A)
40 C
26 C
62 C
35 dB(A)
40 C
31 C
67 C
35 dB(A)
40 C
33 C
76 C
35 dB(A)
40 C
36 C
65 C
35 dB(A)
ta40/B
ta40/B
ta40/B
ta40/B
ta40/B
ta40/B
IP00 IP00 IP00 IP00 IP00 IP00
I I I I I I
180/78 210/63 210/73 210/88 210/103 240/83
25 A gL/gG
0.86 T
< 0.30
< 0.25
95 %
35 A gL/gG
0.94 T
< 0.23
< 0.17
96 %
50 A gL/gG
1 T
< 0.16
< 0.11
96 %
50 A gL/gG
1.05 T
< 0.10
< 0.08
96 %
63 A gL/gG
1 T
< 0.065
< 0.053
96 %
80 A gL/gG
1.09 T
< 0.071
< 0.045
96 %
< 44 W
< 143 W
< 187 W
< 40 W
< 164 W
< 204 W
< 48 W
< 159 W
< 207 W
< 56 W
< 178 W
< 234 W
< 69 W
< 204 W
< 273 W
< 76 W
< 310 W
< 386 W
ES710-E_D00151_02_D_XXEN/12.2015 5
BENDER Group