PH and Acidity in Milk
PH and Acidity in Milk
PH and Acidity in Milk
DOC316.52.93099
1.
Introduction
Principle
Electrodes:
Titrant:
NaOH 0.1 eq/L. Use a commercial solution or dissolve 4.00 g of sodium hydroxide in 1 L of
CO2-free boiled water
Standard for titrant calibration: Dihydrate oxalic acid, molar weight = 126.07 g/mol
pH standards:
Deionized water
4.
V sample = 30 mL
Syringe volume = 10 mL
1.5
4.5
50
2
9.5
28.5
315
13
The syringe is allowed to refill once during titration, so the range is widened:
Veq (mL)
Acidity in dg/L of lactic acid
Acidity in mL of 0.1N NaOH/L
Acidity in S-H
1.5
4.5
50
2
19
57.0
630
25
Page 1 of 6
04/2015, Edition 2
4.3. Settings
4.3.1. pH in milk
Name
Probe
Recommended probe
Stirring
Stirring speed
Stirring time
pH measurement
Stirring speed
Measured parameter
Stability criteria
Stability max time
Default parameter
Unit
PHC725
15
15
[%]
[s]
[%]
[pH]
[pH/min]
[s]
0.05
120
Default parameter
Unit
Milk
30
0
35
[mL]
[mL]
[mL]
NaOH
0.1000
Syringe 1
[eq/L]
PHC725
no
30
20
0
20
True
0
0.2
1.5
8.4
Veq
True
Acidity (g/L lactic acid)
True
Acidity (dg/L lactic acid)
2 decimals
0
60
0
60
Acidity (mL of 0.1N NaOH /L)
2 decimals
0
670
0
670
Soxlet-Henkel acidity
[s]
[%]
[pH]
[mL]
[mL]
[s]
[mL]
[mL]
[pH]
[dg/L]
[dg/L]
[dg/L]
[dg/L]
[mL/L]
[mL/L]
[mL/L]
[mL/L]
Page 2 of 6
04/2015, Edition 2
R5 resolution
R5 min
R5 max
R5 QC min
R5 QC max
2 decimals
0
27
0
27
[S-H]
[S-H]
[S-H]
[S-H]
Titration procedure
5.1. Leveling
To use this method, an external pump is required. All elements (probes, tubes from the titrator and the tube
from the external pump) have to be well installed on the probe holder. The beaker has to contain a level of
sample higher than the position of the tube of the external pump. When the beaker is attached to the probe
holder, this method allows the system to automatically remove the excess sample by a defined pump working
time, and always keep the same sample volume before launching the analysis.
In order to define this volume, autoleveling calibration sequence has to be previously executed (see section 9.3
Autoleveling calibration).
When this option is active, the working time of the external pump must be set (default 30 s). The minimum
working time must allow the pump to be removing air during the last few seconds of the external pump
activation.
Note: Do not forget to re-edit the sample amount with the expected value when deactivating the leveling
method.
5.2. Titration
Rinse the pH probe with deionized water. If leveling is disabled, use a pipette to collect precisely 30 mL of
sample.
Pour the sample into the 50 mL polypropylene beaker. Put in a magnetic stir bar, dip the probe and the delivery
tip in the solution. Start the application.
At the end of the titration, a first window displays the result. A second window displays the titration curve and
the equivalent point coordinates.
After the titration, there are two possibilities:
6.
Replicate the sample. This is used to study the repeatability by analyzing several samples
successively. At the end of each titration, a window displays the average value, the standard
deviation (SD) and the relative standard deviation (RSD in %).
Analyze a new sample. Another titration can be started but no Standard Deviation and RSD
value will be made.
Results
6.1. Result calculation
Ctitrant (eq/L) FX
Vsample (mL) 0.1
Page 3 of 6
04/2015, Edition 2
Acidity (S-H) =
Ctitrant (eq/L) FX
Vsample (mL) 0.25
The factor FX (1 by default) has to be changed if the titrant unit is changed, for example:
Titrant titer unit
Factor FX
eq/L
1
meq/L
0.001
Parameter
15.60
0.08
0.5
43
75
Unit
[dg/L of lactic acid]
[dg/L of lactic acid]
[%]
[s]
[s]
7.
Recommendations
Always rinse the pH probe and the delivery tip between measurements.
Refill the electrode regularly with KCl saturated solution to maintain the level of internal solution around 1 cm
(0.4 inches) below the refill hole.
8.
Bibliography
9.
Appendices
9.1. Electrode calibration
For the precision of the titration, it is recommended to calibrate the probe with pH standards at the same
temperature as the samples. For calibration, the pH is compensated at 25 C.
Pour a sufficient amount of the first buffer into a 50 mL beaker and add a magnetic stir bar. Place it on the
titrator under the probe holder and dip the probe into the beaker. Start the electrode calibration sequence.
Rinse the probe between each buffer. Repeat this operation for each buffer (at least two buffers are
recommended).
At the end of the calibration, the results of the slope and offset are displayed and the user can accept or reject
this result.
Page 4 of 6
04/2015, Edition 2
Setting
25
0.05
300
97
102
44
74
Unit
[%]
[pH/min]
[s]
[%]
[%]
[mV]
[mV]
Default parameter
NaOH
0.1000
Syringe 1
Oxalic acid
30
25
40
126.07
25
2
8
True
0
0.08
0.8
8.55
Titer
4 decimals
0.09
0.11
Unit
[eq/L]
[mg]
[mg]
[mg]
[g/mol]
[%]
[pH]
[mL]
[mL]
[s]
[mL]
[mL]
[pH]
[eq/L]
[eq/L]
Page 5 of 6
04/2015, Edition 2
Autoleveling calibration uses the same settings as for a titrant calibration (see section 9.2 Titrant calibration).
Default settings for autoleveling calibration
Name
Sample
Amount min
Amount max
Autoleveling calibration
Solution name
Concentration
Resolution
Default parameter
Unit
0
35
[mL]
[mL]
Oxalic Acid
0.007
3 decimals
[mol/L]
Page 6 of 6
04/2015, Edition 2