Fluke 789 Manual PDF
Fluke 789 Manual PDF
Fluke 789 Manual PDF
ProcessMeter
Users Manual
August 2002
2002 Fluke Corporation, All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.
Table of Contents
Title
Introduction....................................................................................................................
Contacting Fluke ...........................................................................................................
Safety Information .........................................................................................................
How to Get Started........................................................................................................
Getting Acquainted with the Meter ................................................................................
Measuring Electrical Parameters...................................................................................
Input Impedance.......................................................................................................
Ranges .....................................................................................................................
Testing Diodes..........................................................................................................
Displaying Minimum, Maximum, and Average..........................................................
Using AutoHold.........................................................................................................
Compensating for Test Lead Resistance..................................................................
Using the Current Output Functions ..............................................................................
Source Mode ............................................................................................................
Simulate Mode..........................................................................................................
Producing a Steady mA Output ................................................................................
Manually Stepping the mA Output ............................................................................
Page
1
1
2
5
6
18
18
18
18
19
19
20
20
20
22
24
25
789
Users Manual
Auto Ramping the mA Output ...................................................................................
Power-Up Options..........................................................................................................
Loop Power Supply Mode..............................................................................................
Battery Life.....................................................................................................................
Maintenance ..................................................................................................................
General Maintenance.....................................................................................................
Calibration .................................................................................................................
Replacing a Fuse ......................................................................................................
If the Meter does not Work ........................................................................................
Replacement Parts and Accessories .............................................................................
Specifications.................................................................................................................
ii
26
27
29
31
31
31
31
34
34
35
39
List of Tables
Table
Title
Page
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
4
7
9
11
11
13
16
25
26
26
28
31
iii
789
Users Manual
iv
List of Figures
Figure
Title
Page
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
5
6
8
10
12
15
21
23
29
30
33
36
789
Users Manual
vi
ProcessMeter
Introduction
WWarning
Read Safety Information before using the
meter.
The Fluke 789 ProcessMeter (referred to as the meter)
is a handheld, battery-operated tool for measuring
electrical parameters, supplying steady or ramping current
to test process instruments, and providing a > 24 V loop
power supply. It has all the features of a digital multimeter,
plus current output capability.
Contacting Fluke
To order accessories, receive operating assistance, or get
the location of the nearest Fluke distributor or Service
Center, call:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
789
Users Manual
WWarning
Safety Information
The meter complies with EN61010, ANSI/ISA S82.011994 and CAN/CSA C22.2 No. 1010.1-92 Overvoltage
Category III. Use the meter only as specified in this
manual, otherwise the protection provided by the meter
may be impaired.
A Warning identifies conditions and actions that pose
hazard(s) to the user; a Caution identifies conditions and
actions that may damage the meter or the equipment
under test.
International symbols used on the meter and in this
manual are explained in Table 1.
ProcessMeter
Safety Information
WCaution
To avoid possible damage to meter or to
equipment under test:
789
Users Manual
Table 1. International Symbols
Symbol
Meaning
Symbol
Meaning
Alternating current
Earth ground
Direct current
Fuse
Battery
Double insulated
Listed
950 Z
CAT III
N10140
ProcessMeter
How to Get Started
Display
789 PROCESSMETER
Pushbuttons
100%
MIN MAX
RANGE
HOLD
SpanCheck
%STEP
COARSE
FINE
0%
Rotary
Switch
REL
mV
Hz
mA
A
V
V
mA
mA
OFF
250
HART
LOOP POWER
mA
mA
COM
Input/Output
Jacks
anw014f.eps
789
Users Manual
mA
COM
anw001f.eps
ProcessMeter
Getting Acquainted with the Meter
Table 2. Input/Output Jacks
Item
Jack
Measurement Functions
Source Current
Function
Ac
mA d
C
D
COM
Simulate Transmitter
Function
789
Users Manual
mV
mA
A
mA
OFF
mA
250
HART
LOOP POWER
mA
anw002f.eps
ProcessMeter
Getting Acquainted with the Meter
Table 3. Rotary Function Switch Positions for Measurements
No.
Position
OFF
Function(s)
Meter off
Default:
Measure ac V
Frequency counter
C
$
Pushbutton Actions
Default:
Measure dc V
Frequency counter
D
%
Default:
Measure dc mV
Same as above
Frequency counter
E
&
Default: Measure
for continuity
(Blue) D test
'
Same as above, except there is only one range for each input
jack position, 30 mA or 1 A
789
Users Manual
mV
V
mA
A
mA
2
mA
OFF
3
250
HART
LOOP POWER
mA
anw008f.eps
10
ProcessMeter
Getting Acquainted with the Meter
Table 4. Rotary Function Switch Positions for mA Output
No.
A
Position
OUTPUT
Default Function
Test leads in
SOURCE:
Source 0 % mA
OUTPUT
0213
Pushbutton Actions
Test leads in
SOURCE:
Source repeating
0 % -100 %-0 %
slow ramp (0)
Test leads in SIMULATE:
Sink repeating
0 % -100 %-0 %
slow ramp (0)
Position
250
HART
LOOP POWER
mA
Default Function
Pushbutton Actions
11
789
Users Manual
100%
MIN MAX
RANGE
HOLD
SpanCheck
%STEP
COARSE
FINE
0%
REL
Hz
7
10
8
anw003f.eps
Figure 5. Pushbuttons
12
ProcessMeter
Getting Acquainted with the Meter
Table 6. Pushbuttons
No.
Pushbutton
Span Check
Function(s)
Toggles the backlight (low, high, and off)
Span Check
% STEP
"
Measuring: Selects a fixed range (hold for 1 second for auto range)
mA Output: Adjusts output up 0.1 mA
COARSE
FINE
13
789
Users Manual
Table 6. Pushbuttons (cont.)
No.
Pushbutton
FINE
Function(s)
Measuring: Toggles between frequency counter and voltage measurement functions
(BLUE)
(alternate
function)
COARSE
% STEP
W
14
ProcessMeter
Getting Acquainted with the Meter
10
11
12
5
13
4
3
14
15
2
1
17
16
anw004f.eps
15
789
Users Manual
Table 7. Display
No.
Element
% (Percentage display)
OUTPUT
Meaning
Shows the mA measured value or output level in %, in a 0-20 mA or 4-20 mA scale
(change scales with power-up option)
Lights when mA output (source or simulate) is active
Lights in continuity function
Lights when dangerous voltage is present
Numerals
IK
/
! MAX MINAVG
16
ProcessMeter
Getting Acquainted with the Meter
Table 7. Display (cont.)
Meaning
No.
Element
Auto Range
Manual Range
400100030
The number plus the unit and multiplier indicate the active range.
Show the input or output units and multipliers associated with the numerals
mV
02 13
One of these lights in mA ramping or step output (rotary function switch position
)0213):
0 - slow continuous 0 % - 100 % - 0 % ramping (40 seconds)
2 - fast continuous 0 % - 100 % - 0 % ramping (15 seconds)
1 - slow ramp in 25 % steps (15 seconds/step)
3 - fast ramp in 25 % steps (5 seconds/step)
250
HART
Loop Power
17
789
Users Manual
2.
3.
4.
Testing Diodes
Input Impedance
For the voltage measurement functions, input impedance
is 10 M. See "Specifications" for more information.
Ranges
A measurement range determines the highest value and
resolution at which the meter can measure. Most meter
measurement functions have more than one range (see
"Specifications").
2.
3.
4.
5.
18
ProcessMeter
Measuring Electrical Parameters
6.
Using AutoHold
Note
! Warning
To avoid possible electric shock, do not use
AutoHold to determine if dangerous voltage
is present. AutoHold will not capture
unstable or noisy readings.
Activate AutoHold to freeze the meters display on each
new stable reading (except in the frequency counter
mode). Press to activate AutoHold. This feature
allows measurements to be taken in situations in which it
is difficult to look at the display. The meter beeps and
updates the display with each new stable reading.
19
789
Users Manual
Compensating for Test Lead Resistance
Use the relative reading feature (Q on the display) to set
the present measurement as a relative zero. A common
use for this feature is to compensate for test lead
resistance when measuring ohms.
Select the measure function, touch the test leads
together, and then press 5. Until 5 is pressed
again, or the meter is switched to another measurement
or source function, the readings on the display will
subtract the lead resistance.
20
Source Mode
Source mode is selected automatically by inserting the
test leads into the SOURCE + and jacks as shown in
Figure 7. Use source mode whenever it is necessary to
supply current into a passive circuit such as a current
loop with no loop supply. Source mode depletes the
battery faster than simulate mode, so use simulate mode
whenever possible.
The display looks the same in source and simulate
modes. The way to tell which mode is in use is to see
which pair of output jacks is in use.
ProcessMeter
Using the Current Output Functions
789 PROCESSMETER
40
20
0
100%
MIN MAX
RANGE
SpanCheck
%STEP
COARSE
0%
REL
mV
HOLD
FINE
60
80
100
Hz
mA
A
V
V
mA
mA
OFF
250
mA HART
LOOP POWER
mA
COM
anw010f.eps
21
789
Users Manual
Simulate Mode
WCaution
Set the rotary function switch to one of the
mA output settings BEFORE connecting the
test leads to a current loop. Otherwise, a low
impedance from the other rotary function
switch positions could be presented to the
loop, causing up to 35 mA to flow in the loop.
Simulate mode is selected automatically by inserting the
test leads into the SIMULATE + and jacks as shown in
Figure 8. Simulate mode conserves battery life, so use it
instead of source mode whenever possible.
The display looks the same in source and simulate
modes. The way to tell which mode is in use is to see
which pair of output jacks is in use.
22
0 mA = 0 %, 20 mA = 100 %
2.
3.
ProcessMeter
Using the Current Output Functions
dc V
Power Supply
789 PROCESSMETER
40
COM
+24V
20
100%
MIN MAX
RANGE
HOLD
SpanCheck
%STEP
COARSE
FINE
0%
REL
mV
60
80
100
Hz
mA
A
V
V
mA
mA
OFF
250
HART
LOOP POWER
mA
mA
COM
anw011f.eps
23
789
Users Manual
Producing a Steady mA Output
When the rotary function switch is in the OUTPUT [ mA
position, and the OUTPUT jacks are connected to an
appropriate load, the meter produces a steady mA dc
output. The meter begins sourcing or simulating 0 %. Use
the pushbuttons to adjust the current as shown in Table 8.
Select either sourcing or simulating by choosing the
SOURCE or SIMULATE output jacks.
24
Note
The STEP pushbuttons described Table 9 are
available when the meter is producing a steady
mA output. The STEP pushbuttons go to the next
multiple of 25 %.
ProcessMeter
Using the Current Output Functions
Table 8. mA Output Adjust Pushbuttons
Pushbutton
X
"
Adjustment
Adjusts up 0.1 mA
COARSE
X
Adjusts up 0.001 mA
FINE
FINE
.
W
COARSE
5
W
Note
The COARSE and FINE adjustment
pushbuttons described in Table 8 are available
when manually stepping the mA output.
25
789
Users Manual
Table 9. mA Stepping Pushbuttons
Pushbutton
X
Span Check
Span Check
Adjustment
% STEP
% STEP
Sets to 0 % value
Step
4 to 20 mA
0 to 20 mA
0%
4.000 mA
0.000 mA
25 %
8.000 mA
5.000 mA
50 %
12.000 mA
10.000 mA
75 %
16.000 mA
15.000 mA
100 %
20.000 mA
20.000 mA
125 %
24.000 mA
120 %
24.000 mA
26
ProcessMeter
Power-Up Options
When the rotary function switch is in the OUTPUT
)0213 position, and the output jacks are
connected to an appropriate load, the meter produces a
continuously repeating 0 % - 100 % - 0 % ramp in a
choice of four ramp waveforms:
2
1
Power-Up Options
To select a power-up option, hold down the pushbutton
shown in Table 11 while turning the rotary function switch
from OFF to any on position. Wait 2 seconds before
releasing the pushbutton after powering up the meter.
The meter beeps to acknowledge the power-up option.
Only the setting for current span is retained when the
power is turned off. The other options have to be
repeated for each operating session.
Holding down more than one pushbutton can activate
more than one power-up option.
Note
At any time during auto ramping, the ramp can
be frozen simply by moving the rotary function
switch to the [ mA position. Then the COARSE,
FINE, and % STEP adjust pushbuttons can be
used to make adjustments.
27
789
Users Manual
Table 11. Power-Up Options
Option
Pushbutton
"
Disable beeper
Disable auto power-off
Default
Remembers
last setting
Enabled
Disables beeper
Enabled
Disabled
(Blue)
28
Action Taken
ProcessMeter
Loop Power Supply Mode
30
Voltage (V)
32
28
26
24
22
20
0
12
16
20
24
Current (mA)
Loop voltage w/ 250
29
789
Users Manual
789 PROCESSMETER
100%
MIN MAX
RANGE
HOLD
SpanCheck
%STEP
COARSE
FINE
0%
REL
mV
Hz
mA
A
TEST DC PWR
++
mA
OFF
mA
Red
250
HART
LOOP POWER
mA
mA
COM
Black
anw009f.eps
30
ProcessMeter
Battery Life
Maintenance
Battery Life
WWarning
To avoid false readings, which could lead to
possible electric shock or personal injury,
replace the battery as soon as the battery
indicator (B) appears.
Table 12 shows typical alkaline battery life. To preserve
battery life:
General Maintenance
Periodically wipe the case with a damp cloth and
detergent; do not use abrasives or solvents.
Calibration
Calibrate the meter once a year to ensure that it performs
according to its specifications. Contact a Fluke Service
Center for instructions.
Hours
140
140
10
31
789
Users Manual
Replacing the Batteries
2.
3.
4.
5.
6.
! Warning
To avoid electrical shock:
32
ProcessMeter
General Maintenance
F1
F2
anw037.eps
33
789
Users Manual
If a fuse is blown, replace it as follows. Refer to Figure 11
as necessary:
Replacing a Fuse
! Warning
To avoid personal injury or damage to the
meter, use only the specified replacement
fuse, 440 mA 1000 V fast-blow, Fluke PN
943121.
1.
Remove the test leads from the meter and turn the
meter OFF.
2.
3.
4.
5.
'.
1.
2.
Plug the black test lead into COM, and the red test
lead into the A c input.
3.
4.
5.
34
ProcessMeter
Replacement Parts and Accessories
If the meter still does not work, contact a Fluke Service
Center. If the meter is under warranty, it will be repaired
or replaced (at Flukes option) and returned at no charge.
See the Warranty on the back of the title page for terms.
If the warranty has lapsed, the meter will be repaired and
returned for a fixed fee. Contact a Fluke Service Center
for information and price.
Note
When servicing the meter, use only the
replacement parts specified here.
Replacement parts and some accessories are shown in
Figure 12 and listed in Table 13. Many more DMM
accessories are available from Fluke. For a catalog,
contact the nearest Fluke distributor.
To find out how to order parts or accessories use the
telephone numbers or addresses shown in "Contacting
Fluke".
35
789
Users Manual
17
30
18
16
8
4
15
5
29
26
14
27
13
28
19
25
24
20
23
12
21
22
11
10
anw005f.eps
36
ProcessMeter
Replacement Parts and Accessories
Table 13. Replacement Parts
Item Number
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Reference
Designator
Description
Quantity
MP14
Knob Assembly
658440
MP1
1622855
MP8
1623923
MP6
Keypad
1622951
MP5
Top Shield
1622924
MP47
674853
MP4
Contact Housing
1622913
MP28-31
RSOB Contact
1567683
! F1, F2
943121
H7,8
PCB Screw
832220
MP9
Bottom Shield
1675171
MP12
IR Lens
658697
MP40,41
1641965
MP7
Backlight/Bracket
1622960
P1
LCD Display
1883431
MP3
Mask
1622881
37
789
Users Manual
Table 13. Replacement Parts (continued)
Item Number
38
Reference
Designator
Description
Quantity
878983
Q
R
S
T
U
V
W
X
Y
Z
a
b
MP50
Shock Absorber
MP11
Bottom Case
659042
MP20
658382
BT1-4
376756
H1-2
948609
MP13
Tilt-Stand
659026
MP15
658424
MP2
1622870
MP46
Shock Absorber
674850
MP16-18
666435
MP19
666438
H3-6
Case Screws
1558745
MP21
Calibration Label
948674
MP22
Calibration Keypad
658689
Not shown
1274382
1 (set of 2)
Not shown
1670095
1 (set of 2)
Not shown
1627890
Not shown
1636493
ProcessMeter
Specifications
The standard specification interval is 1 year.
Specifications
All specifications apply from +18 C to +28 C unless
stated otherwise.
All specifications assume a 5-minute warm-up period.
Note
Counts refers to the number of increments or
decrements of the least significant digit.
DC Volts Measurement
Range (V dc)
Resolution
4.000
0.001 V
0.1 % + 1
40.00
0.01 V
0.1 % + 1
400.0
0.1 V
0.1 % + 1
1000
1V
0.1 % + 1
39
789
Users Manual
DC Millivolts Measurement
Range (mV dc)
400.0
Resolution
0. 1 mV
0.1 % + 2
AC Volts Measurement
Accuracy, (% of Reading + Counts)
Range (ac)
Resolution
50 Hz to 60 Hz
200 Hz to 500 Hz
400.0 mV
0.1 mV
0.7 % + 4
1.2 % + 4
7.0 % + 4
4.000 V
0.001 V
0.7 % + 2
1.2 % + 4
7.0 % + 4
40.00 V
0.01 V
0.7 % + 2
1.2 % + 4
7.0 % + 4
400.0 V
0.1 V
0.7 % + 2
1.2 % + 4
7.0 % + 4
1000 V
1V
0.7 % + 2
1.2 % + 4
7.0 % + 4
40
45 Hz to 200 Hz
ProcessMeter
Specifications
AC Current Measurement
Range
45 Hz to 2 kHz
Resolution
1.000 A (Note)
0.001 A
Typical Burden
Voltage
1.5 V/A
DC Current Measurement
Typical Burden
Voltage
Resolution
30.000 mA
0.001 mA
0.05 % + 2
14 mV/mA
1.000 A (Note)
0.001 A
0.2 % + 2
1.5 V/A
Range
41
789
Users Manual
Ohms Measurement
Range
Resolution
400.0
0. 1
220 A
0.2 % + 2
4.000 k
0.001 k
60 A
0.2 % + 1
40.00 k
0.01 k
6.0 A
0.2 % + 1
400.0 k
0.1 k
600 nA
0.2 % + 1
4.000 M
0.001 M
220 nA
0.35 % + 3
40.00 M
0.01 M
22 nA
2.5 % + 3
42
Measurement Current
ProcessMeter
Specifications
Frequency Counter Accuracy
Range
Resolution
199.99 Hz
0.01 Hz
0.005 % + 1
1999.9 Hz
0.1 Hz
0.005 % + 1
19.999 kHz
0.001 kHz
0.005 % + 1
DC
(approximate trigger level 5 % of full scale)
400 mV
150 mV
4V
1V
1V
40 V
3V
3V
400 V
30 V
30 V
1000 V
300 V
300 V
Input Range
43
789
Users Manual
Diode Test and Continuity Test
Simulate Mode:
General Specifications
DC Current Output
Source mode:
44
ProcessMeter
Specifications
Accuracy adders for use in RF Fields: In an RF field of
3 V/m, change the accuracy specifications as follows:
For AC Volts Measurement, add 0.25 % of range
For DC Current Measurement, 30.000 mA range,
add 0.14 % or range
For DC Current Output, add 0.32 % of span
Accuracy for all meter functions is not specified in RF
fields > 3 V/m.
Certifications:
Relative humidity: 95 % up to 30 C, 75 % up to 40 C,
45 % up to 50 C, and 35 % up to 55 C
Vibration: Random 2g, 5 to 500 Hz
Shock: 1 meter drop test
Safety: Complies with EN61010, ANSI/ISA S82.01-1994
and CAN/CSA C22.2 No. 1010.1-92 Overvoltage
Category III.
45
789
Users Manual
46