Conference Proceedings: Inhaled Nitric Oxide: Delivery Systems and Monitoring
Conference Proceedings: Inhaled Nitric Oxide: Delivery Systems and Monitoring
Conference Proceedings: Inhaled Nitric Oxide: Delivery Systems and Monitoring
Introduction
Excitement over the potential clinical applications of
inhaled nitric oxide (INO) has frequently been tempered
Richard D Branson RRT, Robert S Campbell RRT, and Jay A Johannigman MD are affiliated with the Division of Trauma and Critical Care,
Department of Surgery, University of Cincinnati, Cincinnati, Ohio. Dean
R Hess PhD RRT FAARC is affiliated with Respiratory Care Services,
Department of Anaesthesia and Critical Care, Massachusetts General
Hospital, Harvard Medical School, Boston, Massachusetts.
Correspondence: Richard D Branson RRT, Department of Surgery, University of Cincinnati, 231 Bethesda Avenue, Cincinnati, OH, 452670558. E-mail: Richard.Branson@UC.edu.
281
282
AND
MONITORING
AND
MONITORING
283
284
AND
MONITORING
NO, FIO2, minute volume, and total gas flow into the machine relative to minute volume. These findings support
the work of Nishimura et al,22 as the minute volume and
total gas flow conspire to alter the residence time. As
minute volume falls, gas in the inspiratory reservoir of the
Servo 900C, at the desired working pressure (40 70 cm
H2O), produces NO2 quickly with increasing NO and FIO2
concentrations.
Limiting NO2 production should be paramount in the
design of any NO delivery system. Previous work suggests
that limiting residence time and using the lowest effective
NO dose are 2 simple techniques toward achieving this
goal. When high FIO2 and high NO concentrations are
required, mixing NO with N2, as opposed to compressed
air, can also be useful when delivering gases prior to the
ventilator.22
Use of Absorbers/Scrubbers to Reduce NO2
Numerous investigators have used carbon dioxide (CO2)
absorbers to scrub NO2 from the inspiratory limb of the
ventilator circuit.8,15,26,42 47 These investigations are difficult to compare because of the differences in experimental
protocols and types of absorbents. Absorbents commonly
used in anesthesia for CO2 removal vary in volume and
use materials including calcium dihydroxide, sodium hydroxide, potassium hydroxide, potassium permanganate,
and other color indicators. The components of these materials should be recorded by investigators to allow comparisons.
Stenqvist et al42 evaluated the effectiveness of soda lime
to absorb NO2, and reported that 3 ppm NO2 decreased to
less than 1 ppm after passing through the absorber. These
authors also found that the NO2 absorbing properties of
soda lime lasted approximately 72 hours. Westfelt et al
found that a soda lime absorber reduced mean NO2 values
from 4.7 ppm to 1.8 ppm.15
Weiman et al43 compared 3 kinds of absorbers, and
found NO2 absorption rates of 15% for Sodasorb, 24% for
Dragersorb, and 34% for Sofnolime. The material with the
highest potassium hydroxide (KOH) content had the greatest absorption properties. These investigators also tested 2
special preparations with KOH levels of 3.0% weight per
weight and 7.3% weight per weight. They found that NO2
removal rates increased to 44% and 47%, respectively.
The presence of excess KOH may lead to caking of the
absorbent when humidified gases are used. In fact, it has
been reported that humidification reduces NO2 absorption
by soda lime,43 but this may be due to an increase in NO2
production because of the added dead space and residence
time created by the humidification chamber. Ishibe et al
suggested that soda lime (Sodasorb and Wako lime-A)
was capable of completely eliminating NO2 concentration
up to 40 ppm from the inspiratory limb.44 They also found
AND
MONITORING
Fig. 1. Time required to generate 2 ppm NO2 using rate constant published by Nishimura et al.22 Top. NO concentrations up to 80 ppm.
Note that at a high NO concentration very little time is required to generate significant NO2. Bottom: NO concentrations in the clinically
relevant range up to 20 ppm. Note that when the NO concentration is less than 10 ppm a relatively long time is required to generate
significant NO2 concentrations. (From Reference 22, with permission.)
285
AND
MONITORING
Comparison of Nitrogen Dioxide and Nitric Oxide Absorption Properties of Different Materials from Published Studies
Absorber Product
(Reference #)
Sofnol B.P.8
Q-Sorb42
Q-Sorb42
Q-Sorb42
Q-Sorb42
Sofnolime45
Sofnolime45
Sofnolime50
Soda lime51
Sofnolime51
Sofnolime51
Intersorb51
Soda Sorb44,52
Soda Sorb44,52
Wako-Lime A44
Wako-Lime A44
Charcoal Filter 63353
Soda lime54
Soda lime54
Soda lime54
Q-Sorb15
Charcoal15
Charcoal 1 Permasorb48
Ca-A zeolite46
Sofnolime46
Sofnolime46
ABEK HgCONO-P347
Charcoal47
DragerSorb47
Absorption
percentage
NO
NO2
NO
12%
40
20
40
NS
40
40
40
80
47
42
4043
35
10
35
10
80
60
60
60
100
100
120
70
70
70
5570
86
86
12%
0.8
0.5
0.5
35
,1
,1
,1
NS
33.5
3
3.1
4.1
31
6.1
30
10
2.8
2.8
2.8
4.7
4.7
NS
5
5
5
912
17.6
17.6
NS
NS
15
NS
NS
80
4
1
18
100
13
8
12
97
21
90
. 99
. 95
135
135
7
41
100
. 98
. 90
10
99
0
0
FIO2
Indicator
Duration of
Exposure
NS
0.5
1.0
0.5
NS
0.78
0.78
0.78
0.85
0.7
0.7
0.7
0.02
0.02
0.02
0.02
NS
0.8
0.8
0.8
0.9
0.9
0.86
NS
NS
NS
1.0
1.0
1.0
Green-Brown
White-Violet
White-Violet
White-Violet
White-Violet
Green-Brown
Pink-White
White-Violet
White-Violet
Green-Brown
Pink-White
White-Violet
White-Violet
White-Violet
White-Violet
White-Violet
NA
Green-Brown
Pink-White
White-Violet
White-Violet
NA
NA
NA
Green-Brown
Green-Brown
NA
NA
Pink-White
3 min
NS
NS
NS
NS
20 min
20 min
20 min
NS
30 min
30 min
30 min
20 min
20 min
20 min
20 min
NS
NS
NS
NS
72 hrs
72 hrs
12 hrs
24 hrs
1 hr
24 hrs
170 hrs
NS
1015 min
NO2
NS
87.5
100
100
98.6
80
3
0
100
100
84
100
100
100
100
100
. 99
100
. 60
. 60
62
77
100
. 98
. 90
50
95
0
17
FIO2 5 fraction of inspired oxygen; NS 5 not specified; NA 5 not applicable. (Modified from Reference 49, with permission.)
286
5 960 ft3
5 5,760 ft3/h
3 NO concentration
5 (163,123 L/h)
3 Flow rate
250 L/h
5 Ambient concentration
Room dimensions
3 Air exchanges per hour
Note that the ambient concentration will be very low for the NO doses typically
used clinically (ie, # 20 ppm). NO 5 nitric oxide; ppm 5 parts per million.
(From Reference 11, with permission.)
AND
MONITORING
situations, when expiratory gases must be scavenged, expired gas can be scrubbed from the gas stream. This is
accomplished by directing gases through a canister of potassium permanganate (Purafil) and charcoal to remove
both NO and NO2.48
Ideal NO Delivery System
Early INO delivery systems were constructed from available equipment and used in a number of investigational
protocols involving patients and experimental animals (Fig.
3). Some of these homemade systems have functioned
well and others may pose hazards due to unstable and
unpredictable NO delivery and NO2 production. Several
problems can complicate NO delivery if not recognized by
the system architects. Mixing of gas in the inspiratory limb
can be incomplete and NO concentrations greater than the
calculated dose can be administered. Streaming of NO in
the inspiratory limb can occur because NOs density and
viscosity are different than the other inspired gases. The
calculated NO concentration may also vary with changes
in mode of ventilation, inspiratory flow pattern, minute
ventilation, or FIO2. Mode of ventilation is particularly a
problem in intermittent mandatory ventilation and pressure support, because VT and flows change dramatically
from breath to breath. Volume control breaths with a constant flow allow for fairly steady concentrations of INO.
However, pressure control breaths or volume control
breaths with a descending flow waveform present a much
more difficult task. The slow-response NO analyzers that
are typically used clinically make it impossible to detect
changes in NO concentration during the ventilatory cycle.
Because investigators have used different delivery systems
and analysis methods, it is difficult to determine the actual
dose administered to the patient in many instances. This
makes interpretation of dose-response studies arduous. To
avoid complications due to inaccurate dosing, NO delivery
systems should provide a precise concentration regardless
of ventilator mode, breath type, or alterations in inspiratory flow.
We believe that the following are important considerations when building a system for delivery of INO.11,56 59
Dependability and safety. INO is used for critically ill
patients. Complex systems will more likely permit errors
that could compromise ventilation, oxygenation, or delivery of the correct NO dose. The function of NO delivery
systems must be thoroughly evaluated in the laboratory
prior to patient use.
Precise and stable NO dose delivery. It is important to
deliver a precise and stable dose to avoid complications
associated with INO. The dose should not vary with changes
in ventilatory pattern or FIO2.
287
AND
MONITORING
Fig. 2. Ambient NO and NO2 levels for one hour with 100 ppm NO delivered into an ICU room at 8 L/min. Note that concentration readings
are in parts per billion. NO and NO2 measured by chemiluminescence 10 ft from NO flow. Note that ambient NO and NO2 levels remain low
without scavenging other than the usual room ventilation. (From Reference 11, with permission.)
288
AND
MONITORING
Fig. 3. Schematic representation of a ventilator system illustrating the various points at which NO can be introduced.78 The premixing
system (pre) introduces NO into the high pressure air inlet of the ventilator. NO can also be introduced into the inspiratory limb of the
ventilator circuit (I) or at the Y-Piece (y). This injection can occur either throughout the ventilatory cycle or electronically coupled to the
ventilator so that flow occurs only during the inspiratory phase. Wall suction is used to scavenge expired NO, which may be unnecessary
if low NO doses are used. (From Reference 78, with permission.)
289
AND
MONITORING
Risks:
Special Controls:
Fig. 4. NO delivery patterns with five systems.78 The peaks represent inspirations. The target NO concentration is 20 ppm in all
cases. Note that only the premixing system (pre) delivers a constant NO dose regardless of the ventilatory pattern. Also note the
very high spikes of NO with some delivery systems. pre 5 premixing; ii 5 inspiratory injection into the inspiratory circuit; iy 5
inspiratory injection at the Y-piece; ci 5 continuous injection into
the inspiratory circuit; cy 5 continuous injection into the Y-piece.
(From Reference 78, with permission.)
290
AND
MONITORING
Instead, the NO bleeds out the expiratory limb of the ventilator during expiration. Perhaps more importantly, continuous injection at the Y-piece prevents the accurate measurement of inspired NO concentration and the dose can
only be approximated by mathematical calculation. This
system suffers many of the same limitations of NO titration into the inspiratory circuit, specifically, augmentation
of VT, reduction in FIO2, decreased ability to trigger the
ventilator, and changes in delivered NO dose with changes
in inspiratory flow and minute ventilation (see Figs. 4 and
5). Using this INO delivery technique, errors of up to
322% in delivered INO have been identified at an inspiration-expiration ratio of 1:4.95
Inspiratory Phase Injection into the Y-Piece (iy)
Fig. 5. NO delivery with five systems during synchronized intermittent mandatory ventilation (SIMV) and pressure support ventilation (PSV).78 The peaks represent inspirations. The first and third
breaths during synchronized intermittent mandatory ventilation represent mandatory breaths. Note that only the premixing system
provides a constant NO dose with changes in ventilatory pattern.
pre 5 premixing; ii 5 inspiratory injection into the inspiratory circuit; iy 5 inspiratory injection at the Y-piece; ci 5 continuous
injection into the inspiratory circuit; cy 5 continuous injection into
the Y-piece. (From Reference 78, with permission.)
291
292
AND
MONITORING
AND
MONITORING
Nomogram to Determine External Blender Setting to Achieve Approximate Delivered INO Concentration
External Blender Setting (800 ppm Source Gas)
100
90
80
70
60
50
40
30
29
28
27
26
25
24
23
22
101
88
76
63
50
100
37
74
24
49
73
97
12
23
35
46
58
69
81
91
10
21
31
41
51
62
72
81
9
18
27
36
45
54
63
71
8
15
23
31
38
46
54
61
6
13
19
26
32
38
45
51
5
10
15
21
26
31
36
41
4
8
12
15
19
23
27
30
3
5
8
10
13
15
18
20
1
3
4
5
6
8
9
10
Notes: 800 parts per million (ppm) nitric oxide (NO) added to O2 inlet of external blender and N2 (or air) added to air inlet of external blender. Outlet of external blender leads to air inlet of
ventilator. The actual delivered dose must always be confirmed by analysis. The numbers in the body of the table represent approximate delivered NO concentration. This nomogram has been used
successfully at Massachusetts General Hospital since 1993 in conjunction with a Puritan Bennett 7200 Ventilator and Bird Blender. For NO source gas concentrations other than 800 ppm, appropriate
mathematical adjustments can be easily made to the target doses in the body of the Table. For example, if a 400 ppm source gas cylinder is used, the target doses in the body of this table should be
divided in half. FIO2 5 fraction of inspired oxygen. (From Reference 11, with permission.)
Table 6.
Flow
Rate
(L/min)
16.0
15.0
14.0
13.0
12.0
11.0
10.0
9.0
8.0
7.5
7.0
6.5
1.50
1.25
1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.30
0.20
0.18
0.15
0.13
0.10
0.08
0.05
0.03
79
69
73
77
58
62
66
70
75
82
47
50
53
57
62
67
73
80
43
45
48
52
56
61
66
73
81
38
41
43
46
50
54
59
65
73
77
82
34
36
38
41
44
48
52
58
64
68
73
78
29
31
33
35
38
41
45
50
56
59
63
68
24
26
28
30
32
35
38
42
47
50
53
57
20
21
22
24
26
28
31
34
38
41
43
46
15
16
17
18
20
21
23
26
29
31
33
35
10
11
11
12
13
14
16
17
20
21
22
24
9
9
10
11
11
13
14
15
17
18
20
21
7
8
8
9
10
11
12
13
15
16
17
18
6
7
7
8
8
9
10
11
12
13
14
15
5
5
6
6
7
7
8
9
10
11
11
12
4
4
4
5
5
5
6
7
7
8
8
9
2
3
3
3
3
4
4
4
5
5
6
6
1
1
1
2
2
2
2
2
2
3
3
3
Notes: 800 parts per million (ppm) nitric oxide (NO) added to O2 inlet of external blender and N2 (or air) added to air inlet of external blender. Outlet of external blender leads to air inlet of
ventilator. The actual delivered dose must always be confirmed by analysis. The numbers in the body of the table represent approximate delivered NO concentration. This nomogram has been used
successfully at Massachusetts General Hospital since 1993 in conjunction with a Puritan Bennett 7200 Ventilator and a Bird Blender. For NO source gas concentrations other than 800 ppm,
appropriate mathematical adjustments can be easily made to the target doses in the body of the Table. For example, if a 400 ppm source gas cylinder is used, the target doses in the body of this table
should be divided in half. (From Reference 11, with permission.)
293
Fig. 7. NO delivery system used for mechanically ventilated neonatal patients at Massachusetts General Hospital. NO (800 ppm) is
introduced into the inspiratory circuit of the ventilator with continuous flow. The NO flow and circuit flow determine the delivered
NO. NO, NO2, and O2 are sampled near the Y-piece. (From Reference 11, with permission.)
294
AND
MONITORING
ventilators,115 we have generally found this to be unnecessary with the low INO doses used in ARDS. We have
also found that many patients can be disconnected briefly
(ie, less than one hour) from INO during transport for
diagnostic tests or operations if FIO2 is increased to support
the patients oxygenation.
NO Delivery with Anesthesia Ventilators
The architecture of the anesthesia ventilator poses significant obstacles to the delivery of a stable dose of INO.
Anesthesia ventilators are configured to recirculate expired gas so as to conserve anesthetic and prevent release
of anesthetic into the room air. Systems that introduce a
continuous flow of NO into the inspiratory limb95 could
administer an unintended high NO dose because the recirculated expired gas may already contain a significant
NO concentration.
Anesthesia ventilators provide a continuous flow of oxygen, to which air and/or anesthetic gases can be added.
This continuous flow (termed fresh gas flow) is supplied
at a rate that can be varied (0.3 to . 12 L/min) depending
on the style of ventilation and ventilatory requirements. If
this were the only gas source available during inspiration,
NO delivery could be accomplished in the same fashion as
a continuous flow ventilator (ie, by introducing a continuous flow of NO into the fresh gas flow). The patients
expired gas passes into a bellows (or bag if employing
manual ventilation) for storage until needed for the next
inspiration. During the expiratory phase, the continuous
fresh gas flow is also diverted from the inspiratory limb
into the bellows (Fig. 9), mixing with the patients expired
gas. If the bellows capacity is exceeded, excess gas is
released through a pressure relief valve into the ventilators scavenging system. On inspiration, a combination of
AND
MONITORING
Several manufacturers are developing systems for delivery of INO, and these are at various stages of development. At the time of this writing, most have not been
approved for use by the FDA, and none can be used by
institutions that do not have a valid FDA Investigational
New Drug Number for the use of INO.
I-NOvent Delivery System
The I-NOvent Delivery System (INO Therapeutics) is a
universal NO delivery system designed for use with most
conventional critical care ventilators. It can be used with
either phasic flow ventilators (eg, adults) or continuous
flow ventilators (eg, neonates). In its typical configuration,
the delivery system is mounted on a transport cart that
holds two 800 ppm NO gas cylinders (Fig. 10). The system is factory-configured for the specific cylinder concentration required in the intended country of use. NO concentrations of 0 80 ppm can be delivered using the 800
ppm source gas cylinder, 0 40 ppm can be delivered using a 400 or 450 ppm cylinder, and 0 20 ppm can be
delivered using a 300 ppm cylinder. The system is delivered with 800 ppm tanks containing 1963 L and pressurized to 2000 psig. An integral battery allows 30 minutes of
operation in the absence of an external power source. At
the time of this writing, I-NOvent Delivery System is the
only commercial NO delivery system that has received
FDA approval for purchase by hospitals who have a valid
Investigational New Drug Number for administration
of INO.
The principle of operation of the I-NOvent Delivery
System is shown in Figure 11. An injection module is
inserted into the inspiratory circuit between the ventilator
output and the humidifier. The injection module includes
a hot film flow sensor and a gas injection tube. Flow in the
ventilator circuit is precisely measured and NO is injected
295
AND
MONITORING
Fig. 11. Schematic representation of the Ohmeda I-NOvent Delivery System in a mechanical ventilator circuit. Flow is measured at
the ventilator outlet and NO is injected proportional to that flow to
achieve the desired NO dose. Gas is sampled from the inspiratory
circuit, near the Y-piece, and analyzed for NO, NO2, and O2.
Fig. 10. The Ohmeda I-NOvent Delivery System. Top: User interface and monitoring panel. Middle: Flow sensor inspiratory limb of
ventilator circuit. Bottom: The I-NOvent Delivery System interfaced
with a mechanical ventilator system.
296
NO concentration is accurate over a wide range of ventilator flows and desired NO concentrations. By delivering
NO proportional to the ventilator flow, residence time is
reduced and NO2 production is minimized.
The I-NOvent Delivery System includes monitoring of
O2, NO, and NO2. Gas is sampled downstream from the
point of injection, near the Y-piece in the inspiratory circuit. Gas concentrations are measured using electrochemical cells that can be calibrated at regular intervals by the
user. Alarms can be set by the user for high NO, low NO,
high NO2, high O2, and low O2. Additional alarms include
loss of source gas pressure, weak or failed electrochemical
cells, calibration required, delivery system failures, and
monitoring failures.
The I-NOvent Delivery System uses a dual-channel design: one channel controls INO delivery and the other
controls monitoring. This design permits INO delivery independent of monitoring, which is an important safety
feature, and allows that the monitoring system can be calibrated without interruption of INO delivery.
A manual INO delivery system is also available with the
I-NOvent Delivery System. With an oxygen flow to the
manual ventilator set at 15 L/min, the I-NOvent injects gas
to provide an INO concentration of 20 ppm. As with any
manual ventilator system for INO, it is important to squeeze
the bag 35 times to clear residual NO2 before attaching it
to the patient.
Two independent evaluations of the I-NOvent Delivery
System have been published. Kirmse et al128 evaluated the
I-NOvent connected to the Puritan Bennett 7200 and Siemens Servo 900C at inspired NO concentrations of 2 ppm,
5 ppm, and 20 ppm, and a variety of ventilator modes.
Table 7.
AND
MONITORING
Cylinder Concentration
100 ppm
1000 ppm
50 ppm
200 ppm
Setting range
0.1 to 9.9 ppm
1 to 99 ppm
0.1 to 9.9 ppm
1 to 99 ppm
Delivered concentration
0.1 to 9.9 ppm
1 to 99 ppm
0.05 to 4.95 ppm
0.2 to 18.9 ppm
297
AND
MONITORING
298
INO can be delivered to spontaneously breathing patients. Although ARDS patients requiring INO are usually
mechanically ventilated, other patients with primary pulmonary hypertension might not be intubated. Several systems have been described to deliver INO to spontaneously
breathing patients. A high-flow system ($ 60 L/min) with
a tight-fitting face mask can be used (Fig. 13).11,70 The
expected [NO] can be calculated as described previously:
[NO] 5 (NO flow 3 source ppm)/(total flow)
Exhaled gas can be scavenged by the hospital vacuum
system. Inspiratory reservoir bags should be avoided with
these systems because of the likelihood of NO2 generation
within the bag.
INO can also be administered to spontaneously breathing patients using a transtracheal O2 catheter100 or a nasal
cannula.69,136 There are several limitations of these systems. It is not possible to analyze the delivered dose, and
the dose varies with the ventilatory pattern of the patient.57
NO/N2 can be delivered directly to the cannula from a
source cylinder, provided that this NO concentration is
AND
MONITORING
The principles of operation for each analyzer will be reviewed and recommendations for use will be provided.
Electrochemical Analyzers
Electrochemical analyzers have some similarity to other
fuel cell technologies used in respiratory care. When NO
diffuses into a reactive electrolyte solution, electrons are
released or absorbed.1 The current generated between electrodes, caused by this release or absorption, is proportional
to the concentration of NO or NO2 in the gas sample.
Electrochemical cells for NO and NO2 monitoring utilize
a semi-permeable diffusion membrane, acid or alkaline
electrolyte, and a series of 3 electrodes. The electrodes
include an anode (sensing electrode), cathode (counter electrode), and reference electrode (Fig. 14).
Gas contacts the diffusion barrier and a small amount of
NO or NO2 passes into the electrolyte. The sensing electrode is positioned nearest the membrane, and the counter
electrode and reference electrode are positioned deeper in
the electrolyte. The reference electrode provides a bias
voltage so that the sensing electrode is kept at a pre-determined operating voltage. The current flow through the
circuit is proportional to NO or NO2 concentration, and is
measured as voltage across a resistance.1
The electrons generated during the oxidation reaction at
the sensing electrode are consumed at the counter elec-
299
AND
MONITORING
Fig. 15. Diagram of a single reaction chamber chemiluminescence analyzer. In this device, sample flow is controlled by the capillary tube.
In the NO2 mode, the chemical or catalytic converter is used to convert NO2 to NO. Ozone is produced from ambient air by an ozonator.
In the sample chamber excited NO2 is created and the photomultiplier measures the emitted photons. (From Reference 1, with permission).
Chemiluminescence Analyzers
2NO2 3 2NO 1 O2
Electrochemical analyzers are small, portable, relatively
inexpensive, easy to calibrate, and use a small sample
volume. The response time of electrochemical sensors is
very slow, with a typical response time of 30 to 40 seconds. They can be sidestream or mainstream sensors. The
sidestream sensors can have an active withdraw system
such as a vacuum pump, or can utilize the pressure in the
ventilator circuit to deliver gas to the analyzer. The latter
system has considerably slower response time and is ineffective in systems at ambient pressures.
Electrochemical sensors can be adversely affected by
humidity, temperature, and pressure in the ventilator cir-
300
AND
MONITORING
Accuracy
Range of measurement
Sources of error
Clinically acceptable
NO 3100 ppm
NO2 0.510 ppm
Humidity, pressure
Response time
Configuration/sample flow
Size
Consumables
1030 seconds
Sidestream, Mainstream
Small, light weight
Fuel cells
Ozone production
Suitable for expired gas measurement
Ease of use/calibration
Oxygen percent correction required
No
No
Simple/quick calibration
No
Chemiluminescence
High (ppm, ppb, ppt)
NO 1023 2100 ppm
NO2 1023 2100 ppm
Quenching (oxygen)
Viscosity (humidity, oxygen)
0.1520 seconds
Sidestream
Large, heavy
Chemical converters, sample lines, capillaries, ozone
generator, scrubber contents
Yes
Yes
Difficult/prolonged calibration
Yes
and the requirement of a scrubber to eliminate ozone contamination from the work area. Table 9 compares the features
of electrochemical and chemiluminescence analyzers.
Several authors have evaluated commercially available
electrochemical and chemiluminescence analyzers.138 146
301
Summary
From a practical standpoint, technical issues related to
NO delivery are as important as therapeutic issues. The
therapeutic benefits will be consistently obtained only with
a reliable delivery system, and hazards and toxicity may be
more problematic with an unreliable delivery system. It is
incumbent upon clinicians to ensure that their INO delivery system is safe and reliable.
REFERENCES
1. Body SC, Hartigan PM, Shernan SK, Formanek V, Hurford WE.
Nitric oxide delivery: measurement and clinical application. J Cardiothor Vasc Anesth 1995;9(6):748763.
2. al-Ali MK, Howarth PH. Nitric oxide and the respiratory system in
health and disease (review). Respir Med 1998;92(5):701715.
3. NIOSH recommendations for occupational safety and health standards,
1988. MMWR Morb Mortal Wkly Rep1988;37(Suppl 7):129.
4. Tentative standard for the classification of toxic gas mixtures. Arlington, VA: Compressed Gas Association; 1991.
5. Austin AT. The chemistry of the higher oxides of nitrogen as related to the manufacture, storage, and administration of nitrous
oxide. Br J Anaesth 1967;39(5):345350.
6. Evans MJ, Stephens RJ, Cabral LJ, et al. Cell renewal in the lungs
of rats exposed to low levels of NO2. Arch Environ Health 1972;
24(3):180188.
7. Kelly FJ, Blomberg A, Frew A, Holgate ST, Sandstrom T. Antioxidant kinetics in lung lavage fluid following exposure of humans
to nitrogen dioxide. Am J Respir Crit Care Med 1996;154(6 Pt
1):17001706.
8. Kain ML. Higher oxides of nitrogen in anesthetic gas circuits. Br J
Anaesth 1967;39(5):382387.
9. Clutton-Brock J. Two cases of poisoning by contamination of nitrous oxide with the higher oxides of nitrogen during anaesthesia.
Br J Anaesth 1967;39(5):388392.
10. Shiel FO. Morbid anatomical changes in the lungs of dogs after
inhalation of higher oxides of nitrogen during anaesthesia. Br J
Anaesth 1967;39(5):413424.
11. Hess D, Ritz R, Branson RD. Delivery systems for inhaled nitric
oxide. Respir Care Clin N Am 1997;3(3):371410.
12. Young JD, Dyar OJ. Delivery and monitoring of inhaled nitric
oxide. Intensive Care Med 1996;22(1):7786.
13. Zapol WM, Hurford WE. Inhaled nitric oxide in the adult respiratory distress syndrome and other lung diseases. New Horiz 1993;
1(4):638650.
14. Krebs C, Escuyer M, Urak G, Zimpfer M, Germann P. Technological basis for NO application and environmental security. Acta
Anaesthesiol Scand Suppl 1996;109:8487.
302
AND
MONITORING
54.
55.
56.
57.
58.
59.
60.
61.
62.
63.
64.
65.
66.
67.
68.
69.
70.
71.
AND
MONITORING
303
72.
73.
74.
75.
76.
77.
78.
79.
80.
81.
82.
83.
84.
85.
86.
87.
88.
89.
90.
91.
304
AND
MONITORING
92. Hovenga S, Koenders ME, van der Werf TS, Moshage H, Zijlstra
JG. Methaemoglobinemia after inhalation of nitric oxide for treatment of hydrochlorothiazide-induced pulmonary oedema. Lancet
1996;348(9033):10351036.
93. Levy B, Bollaert PE, Bauer P, Nace L, Audibert G, Larcan A.
Therapeutic optimization including inhaled nitric oxide in adult
respiratory distress syndrome in a polyvalent intensive care unit.
J Trauma 1995;38(3):370374.
94. Puybasset L, Rouby JJ, Mourgeon E, Stewart TE, Cluzel P, Arthaud
M, et al. Inhaled nitric oxide in acute respiratory failure: doseresponse curves. Intensive Care Med 1994;20(5):319327.
95. Fernandez R, Artigas A, Blanch L. Ventilatory factors affecting
inhaled nitric oxide concentrations during continuous-flow administration. J Crit Care 1996;11(3):138143.
96. Rich GF, Murphy GD Jr, Roos CM, Johns RA. Inhaled nitric oxide.
Selective pulmonary vasodilation in cardiac surgical patients. Anesthesiology 1993;78(6):10281035.
97. Putensen C, Rasanen J, Thomson MS, Braman RS. Method of
delivering constant nitric oxide concentrations during full and partial ventilatory support. J Clin Monit 1995;11(1):2331.
98. Benzing A, Geiger K. Inhaled nitric oxide lowers pulmonary capillary pressure and changes longitudinal distribution of pulmonary
vascular resistance in patients with acute lung injury. Acta Anaesthesiol Scand 1994;38(7):640645.
99. Puybasset L, Rouby JJ, Mourgeon E, Cluzel P, Souhil Z, LawKoune JD, et al. Factors influencing cardiopulmonary effects of
inhaled nitric oxide in acute respiratory failure. Am J Respir Crit
Care Med 1995;152(1):318328.
100. Snell GI, Salamonsen RF, Bergin P, Esmore DS, Khan S, Williams
TJ. Inhaled nitric oxide used as a bridge to heart-lung transplantation in a patient with end-stage pulmonary hypertension. Am J
Respir Crit Care Med 1995;151(4):12631266.
101. Day RW, Lynch JM, Shaddy RE, Orsmund GS. Pulmonary vasodilatory effects of 12 and 60 parts per million inhaled nitric oxide
in children with ventricular septal defect. Am J Cardiol 1995;75(2):
196198.
102. Adatia I, Lillehei C, Arnold JH, Thompson JE, Palazzo R, Fackler
JC, Wessel DL. Inhaled nitric oxide in the treatment of postoperative graft dysfunction after lung transplantation. Ann Thorac Surg
1994;57(5):13111318.
103. Bigatello LM, Hurford WE, Kacmarek RM, Roberts JD Jr, Zapol
WM. Prolonged inhalation of low concentrations of nitric oxide in
patients with severe adult respiratory distress syndrome. Effects on
pulmonary hemodynamics and oxygenation. Anesthesiology 1994;
80(4):761770.
104. Blomqvist H, Wickerts CJ, Andreen M, Ullberg U, Ortquist
A, Frostell C, et al. Enhanced pneumonia resolution by inhalation of nitric oxide? Acta Anaesthesiol Scand 1993;37(1):
110114.
105. Putensen C, Rasanen J, Downs JB. Effect of endogenous and inhaled
nitric oxide on the ventilation-perfusion relationships in oleic-acid lung
injury. Am J Respir Crit Care Med 1994;150(2):330336.
106. Putensen C, Rasanen J, Lopez FA, Downs JB. Continuous positive
airway pressure modulates effect of inhaled nitric oxide on the
ventilation-perfusion distributions in canine lung injury. Chest 1994;
106(5):15631569.
107. Channick RN, Newhart JW, Johnson FW, Moser KM. Inhaled nitric oxide reverses hypoxic pulmonary vasoconstriction in dogs. A
practical nitric oxide delivery and monitoring system. Chest 1994;
105(6):18421847.
108. McIntyre RC, Jr, Moore FA, Moore EE, Piedalue F, Haenel JS,
Fullerton DA. Inhaled nitric oxide variably improves oxygenation
109.
110.
111.
112.
113.
114.
115.
116.
117.
118.
119.
120.
121.
122.
123.
124.
125.
and pulmonary hypertension in patients with acute respiratory distress syndrome. J Trauma 1995;39(3):418425.
Johannigman JA, Campbell RS, Davis K Jr, Hurst JM. Combined
differential lung ventilation and inhaled nitric oxide therapy in the
management of unilateral pulmonary contusion. J Trauma 1997;
42(1):108111.
Johannigman JA, Davis K Jr, Campbell RS, Luchette FA, Hurst
JM, Branson RD. Inhaled nitric oxide in acute respiratory distress
syndrome. J Trauma 1997;43(6):904910.
Dellinger RP, Zimmerman JL, Taylor RW, Straube RC, Hauser DL,
Criner GJ, et al. Effects of inhaled nitric oxide in patients with acute
respiratory distress syndrome: results of a randomized phase II trial.
Inhaled Nitric Oxide in ARDS Study Group. Crit Care Med 1998;
26(1):1523.
Graybeal J, Schuler GH, Larach DR. Instability of nitric oxide
concentration delivered from a two-blender system (letter). Crit
Care Med 1995;23(9):16111612.
Beghetti M, Habre W, Friedli B, Berner M. Continuous low dose
inhaled nitric oxide for treatment of severe pulmonary hypertension
after cardiac surgery in paediatric patients. Br Heart J 1995;73(1):
6568.
Betit P. Nitric oxide administration during pediatric mechanical
ventilation. Respir Care Clin N Am 1996;2(4):587605.
Betit P, Adatia I, Benjamin P, Thompson JE, Wessel DL. Inhaled
nitric oxide: evaluation of a continuous titration delivery technique
for infant mechanical and manual ventilation. Respir Care 1995;
40(7):706715.
Fierobe L, Brunet F, Dhainaut JF, Monchi M, Belghirth M, Mira
JP, et al. Effect of inhaled nitric oxide on right ventricular function
in adult respiratory distress syndrome. Am J Respir Crit Care Med
1995;151(5):14141419.
Kinsella JP, Ivy DD, Abman SH. Inhaled nitric oxide improves gas
exchange and lowers pulmonary vascular resistance in severe experimental hyaline membrane disease. Pediatr Res 1994;36(3):402
408.
Lonnqvist PA, Jonsson B, Winberg P, Frostell CG. Inhaled nitric
oxide in infants with developing or established chronic lung disease. Acta Paediatr 1995;84(10):11881192.
Lonnqvist PA, Winberg P, Lundell B, Sellden H, Olsson GL. Inhaled nitric oxide in neonates and children with pulmonary hypertension. Acta Paediatr 1994;83(11):11321136.
Leclerc F, Riou Y, Martinot A, Storme L, Hue V, Flurin V, et al.
Inhaled nitric oxide for a severe respiratory syncytial virus infection
in an infant with bronchopulmonary dysplasia. Intensive Care Med
1994;20(7):511512.
Miller OI, Celermajer DS, Deanfield JE, Macrae DJ. Guidelines for
the safe administration of inhaled nitric oxide. Arch Dis Child Fetal
Neonatal Ed 1994;70(1):F47F49.
Miller OI, Celermajer DS, Deanfield JE, Macrae DJ. Very-lowdose inhaled nitric oxide: a selective pulmonary vasodilator after
operations for congenital heart disease. J Thor Cardiovasc Surg
1994;108(3):487494.
Inhaled nitric oxide in full-term and nearly full-term infants with
hypoxic respiratory failure. The Neonatal Inhaled Nitric Oxide Study
Group. N Engl J Med 1997;336(9):597604. Published erratum
appears in N Engl J Med 1997;337(6):434.
Roberts JD Jr, Fineman JR, Morin FC 3rd, Shaul PW, Rimar S,
Schreiber MD, et al. Inhaled nitric oxide and persistent pulmonary
hypertension of the newborn. The Inhaled Nitric Oxide Study Group.
N Engl J Med 1997;336(9):605610.
Mortimer TW, Math MCM, Fajardo CA. Inhaled nitric oxide delivery with high-frequency jet ventilation: a bench study. Respir
Care 1996;41:895899.
AND
MONITORING
126. Ritz R, Hess D, Head CA, Kacmarek R. Nitric oxide and nitrogen
dioxide levels during manual ventilation (abstract). Respir Care
1995;40:1184.
127. Grover R, Murdoch I, Smithies M, Mitchell I, Bihari D. Nitric
oxide during hand ventilation in a patient with acute respiratory
failure. Lancet 1992;340(8826):10381039.
128. Kirmse M, Hess D, Fujino Y, Kacmarek RM, Hurford WE. Delivery of inhaled nitric oxide using the Ohmeda INOvent Delivery
System. Chest 1998;113(6):16501657.
129. Young JD, Roberts M, Gale LB. Laboratory evaluation of the INOvent nitric oxide delivery device. Brit J Anaesth 1997;79(3):
398401.
130. Sydow M, Bristow F, Zinserling J, Allen SJ. Flow-proportional
administration of nitric oxide with a new delivery system. Chest
1997;112(2):496504.
131. Gerlach H, Pappert D, Lewandowski K, Rossaint R, Falke KJ.
Long-term inhalation with evaluated low doses of nitric oxide for
selective improvement of oxygenation in patients with adult respiratory distress syndrome. Intensive Care Med 1993;19(8):443449.
132. Gerlach H, Rossaint R, Pappert D, Falke KJ. Time-course and
dose-response of nitric oxide inhalation for systemic oxygenation
and pulmonary hypertension in patients with adult respiratory distress syndrome. Eur J Clin Invest 1993;23(8):499502.
133. OHare B, Betit P, Thompson J, et al. In-vivo evaluation of a
composite pediatric nitric oxide delivery monitoring system (abstract). Crit Care Med 1997;25:A54.
134. Lindberg L, Larsson A, Steen S, Olsson SG, Nordstrom L. Nitric
oxide gives maximal response after coronary artery bypass surgery.
J Cardiothorac Vasc Anesth 1994;8(2):182187.
135. Lindberg L, Rydgren G, Larsson A, Olsson SG, Nordstrom L. A
delivery system of inhalation of nitric oxide evaluated with chemiluminescence, electrochemical fuel cells, and capnography. Crit
Care Med 1997;25(1):190196.
136. Channick RN, Newhart JW, Johnson FW, Williams PJ, Auger WR,
Fedullo PF. Pulsed delivery of inhaled nitric oxide to patients with
primary pulmonary hypertension. An ambulatory delivery system
and initial clinical tests. Chest 1996;109(6):15451549.
137. Matthews R, Sawyer R, Schefer R. Interferences in chemiluminescent measurement of NO and NO2 emissions from combustion
systems. Environ Sci Tech 1977;11:10921099.
138. Bathe D, Berssenbrugge A, Kohlman T, et al. Monitoring accuracy
for the measurement of nitric oxide in the 0 to 80 ppm range
(abstract). Crit Care Med 1996;24:A103.
139. Etches PC, Harris ML, McKinley R, Finer NN. Clinical monitoring
of inhaled nitric oxide: comparison of chemiluminescent and electrochemical sensors. Biomed Instrum Tech 1995;29(2):134140.
140. Miyamoto K, Aida A, Nishimura M, Nakano T, Kawakami Y,
Ohmori Y, et al. Effects of humidity and temperature on nitrogen
dioxide formation from nitric oxide. Lancet 1994;343(8905):1099
1110.
141. Moutafis M, Hatahet Z, Castelain MH, Renaudin MH, Monnot A,
Fischler M. Validation of a simple method assessing nitric oxide
and nitrogen dioxide concentrations. Intensive Care Med 1995;
21(6):537541.
142. Nishimura M, Imanaka H, Uchiyama A, Tashiro C, Hess D, Kacmarek RM. Nitric oxide (NO) measurement accuracy. J Clin Monit
1997;13(4):241248.
143. Nelin LD, Christman NT, Morrisey JF, Dawson CD. Electrochemical nitric oxide and nitrogen dioxide analyzer for use with inhaled
nitric oxide. J Appl Physiol 1996;81(3):14231429.
144. Petros AJ, Cox PB, Bohn D. Simple method for monitoring concentration of inhaled nitric oxide (letter). Lancet 1992;340(8828):
1167.
305
Discussion
Stewart: One of my concerns is that
were starting to use INO with high
frequency oscillation ventilation, and
we deliver INO in the inspiratory limb
and measure it just prior to the Ypiece. Im concerned with dynamic hyperinflation, or gas trapping, about
whats actually going on at the alveolar level, and how much NO2 is being produced. Do you have any comments on the use of INO in techniques
prone to gas trapping, such as inverse
ratio or high frequency oscillation or
high frequency jet ventilation?
Branson: I think those are 3 very
different things. In high frequency oscillation, the delivery of INO is actually fairly easy because its a constant
flow. Now, what effects happen in the
lung with air trapping and especially
delivery of INO to alveolar units that
eventually have airways collapse in
front of them, thats a good question.
What happens to that INO? With high
frequency jet ventilation it is virtually
impossible to deliver an accurate INO
concentration unless youre going to
deliver INO through the jet ventilator
and INO through the entrainment gas
at a constant rate. If you try to rely on
entraining INO and the patients compliance changes, the position of the
catheter changes, the endotracheal
tube changes, all bets are off, and you
dont know what INO youre delivering. The same thing goes if you just
deliver it through the jet catheter because you dont know what happens
to the gas that youre entraining. Inverse ratio ventilation, again, isnt
unique unto itself in terms of INO delivery, but in the systems that you have
air trapping, especially behind closed
airways, like in the patient with chronic
obstructive pulmonary disease, the fate
of that INO Im unsure about.
306
AND
MONITORING
146. Strauss JM, Krohn S, Sumpelmann R, Schroder D, Barnert R. Evaluation of two electrochemical monitors for measurement of inhaled
nitric oxide. Anaesthesia 1996;51(2):151154.
AND
MONITORING
307