Method Validation and Development
Method Validation and Development
whenever the conditions change for which the method has been validated (e.g., an
instrument with different characteristics or samples with a different matrix); and
whenever the method is changed and the change is outside the original scope of the
method.
Method validation has received considerable attention in the literature and from industrial
committees and regulatory agencies.
The U.S. FDA CGMP request in section 211.165 (e) methods to be validated: The
accuracy, sensitivity, specificity, and reproducibility of test methods employed by the
firm shall be established and documented. Such validation and documentation may be
accomplished in accordance with Sec. 211.194(a). These requirements include a
statement of each method used in testing the sample to meet proper standards of
accuracy and reliability, as applied to the tested product. The U.S. FDA has also
proposed industry guidance for Analytical Procedures and Methods Validation.
ISO/IEC 17025 includes a chapter on the validation of methods with a list of nine
validation parameters. The ICH has developed a consensus text on the validation of
analytical procedures. The document includes definitions for eight validation
characteristics. ICH also developed guidance with detailed methodology.
www.intechopen.com
The U.S. EPA prepared guidance for methods development and validation for the
Resource Conservation and Recovery Act (RCRA). The AOAC, the EPA and other
scientific organizations provide methods that are validated through multi-laboratory
studies.
www.intechopen.com
Are there interfering substances expected, and, if so, should they be detected and
quantified?
Which type of equipment should be used? Is the method for one specific instrument, or
should it be used by all instruments of the same type?
Will the method be used in one specific laboratory or should it be applicable in all
laboratories at one side or around the globe?
www.intechopen.com
5.
Acceptance criteria for equipment checks, system suitability tests and QC sample
analysis
6. Action plan in case criteria 2, 3 and/or 4 are not met.
In many cases, methods are developed and validated in service laboratories that are
specialized in this task. When the method is transferred to the routine analytical laboratory,
care should be taken that the method and its critical parameters are well understood by the
workers in the departments who apply the method. A detailed validation protocol, a
documented procedure for method implementation and good communication between the
development and operation departments are equally important. If the method is used by a
number of departments, it is recommended to verify method validation parameters and to
test the applicability and usability of the method in a couple of these departments before it is
distributed to other departments. In this way, problems can be identified and corrected
before the method is distributed to a larger audience. If the method is intended to be used
by just one or two departments, an analyst from the development department should assist
the users of the method during initial operation. Users of the method should be encouraged
to give constant feedback on the applicability and usability of the method to the
development department. The latter should correct problems if any arise.
2.2 Transferring validated routine methods
Validated routine methods are transferred between laboratories at the same or different sites
when contract laboratories offer services for routine analysis in different areas or when
products are manufactured in different areas. When validated routine methods are
transferred between laboratories and sites, their validated state should be maintained to
ensure the same reliable results in the receiving laboratory. This means the competence of
the receiving laboratory to use the method should be demonstrated through tests, for
example, repeat critical method validation experiments and run samples in parallel in the
transferring and receiving laboratories. The transfer should be controlled by a procedure,
The recommended steps are:
www.intechopen.com
yes
no
Perform part or full
revalidation
www.intechopen.com
www.intechopen.com
measure of precision normally encountered). It may be that some in-between measure is the
most useful in particular cases; for example precision measured between different analysts,
over extended timescales, within a single laboratory. This is sometimes known as
intermediate precision, but the exact conditions should be stated. Precision is usually
stated in terms of standard deviation or relative standard deviation. Both repeatability and
reproducibility are generally dependent on analyte concentration, and so should be
determined at a number of concentrations and if relevant, the relationship between
precision and analyte concentration should be established. The Purpose of carrying out a
determination is to obtain a valid estimate of a true value. When one considers the criteria
according to which an analytical procedure is selected, precision and accuracy are usually
the first time to come to mind. Precision and accuracy together determine the error of an
individual determination. They are among the most important criteria for judging analytical
procedures by their results. Precision refers to the reproducibility of measurement within a
set, that is, to the scatter of dispersion of a set about its central values. The term set is
defined as referring to a number (n) of independent replicate measurements of some
property. One of the most common statistical terms employed is the standard deviation of a
population of observation. Standard deviation is the square root of the sum of squares of a
deviations of individual results for the mean, divided by one less than the number of results
in the set.
The standard deviation S, is given by
S = =
1 N
(xi x )2 ,
N i =1
Standard deviation has the same units as the property being measured.
The square of standard deviation is called variance (s2). Relative standard deviation is the
standard deviation as a fraction of the mean, i.e. S/x. It is some times multiplied by 100 and
expressed as a percent relative standard deviation. It becomes a more reliable expression of
precision.
% Relative Standard Deviation (RSD)= S * 100/x
3.3 Accuracy and recovery
The accuracy of an analytical method is the extent to which test results generated by the
method and the true value agree. Accuracy can also be described as the closeness of
agreement between the value that is adopted, either as a conventional, true or accepted
reference value, and the value found. The true value for accuracy assessment can be
obtained in several ways. One alternative is to compare the results of the method with
results from an established reference method. This approach assumes that the uncertainty of
the reference method is known. Secondly, accuracy can be assessed by analyzing a sample
with known concentrations (e.g., a control sample or certified reference material) and
comparing the measured value with the true value as supplied with the material. If certified
reference materials or control samples are not available, a blank sample matrix of interest
can be spiked with a known concentration by weight or volume. After extraction of the
analyte from the matrix and injection into the analytical instrument, its recovery can be
determined by comparing the response of the extract with the response of the reference
material dissolved in a pure solvent. Because this accuracy assessment measures the
effectiveness of sample preparation, care should be taken to mimic the actual sample
www.intechopen.com
10
preparation as closely as possible. If validated correctly, the recovery factor determined for
different concentrations can be used to correct the final results.
3.3.1 Calibration
Calibration is the most important step in bioactive compound analysis. A good Precision
and accuracy can only be obtained when a good calibration procedure is adopted. In the
Spectrophotometric methods, the concentration of a sample cannot be measured directly,
but is determined using physical measuring quantity y (absorbance of a solution). An
unambiguous empirical or theoretical relationship can be shown between this quanity and
the concentration of an analyte. The calibration between y = g(x) the calibration function can
be obtained by fitting an adequate mathematical model through the experimental data. The
most convenient calibration function is linear, goes through the origin and is applicable over
a wide dynamic range. In practice however, many deviations from the ideal calibration line
may occur. For the majority of analytical techniques uses the calibration equation.
Y = a + bX
In calibration, univarate regression is applied, when means that all observations are
dependent upon a single variable X.
3.3.2 Standard deviation of slope (Sb)
The standard deviation of slope is proportional to standard error of estimate and inversely
proportional to the range and square root of the number of data points.
3.3.3 Standard deviation of intercept, (Sa)
Intercept values of least squares fits of data are often to evaluate additive errors between or
among different methods.
3.3.4 Correlation coefficient, (r)
The correlation coefficient r (x,y) is more useful to express the relationship of the chosen
scales. To obtain a correlation coefficient the covariance is divided by the product of the
standard deviation of x and y.
( xi x ) ( y i y )
n
r=
i =1
n
(xi x )2 ( yi y )2
i =1
i =1
The absolute recovery of analytical method as the response of a processed spiked matrix
expressed as a percentage of the response of pure standard. Which has not been subjected to
sample pre-treatment and indicates whether the method provides a response for the entire
amount of analyte that is present in the sample. It is best established by comparing the
responses of extracted samples at low, medium and high concentrations in replicates at least
6 with those non- extracted standards, which represent 100% recovery.
Absolute recovery =
www.intechopen.com
11
Slope = sensitivity
LOQ
LOD
Intercept
Amount
www.intechopen.com
12
accuracy and linearity using the method as written. The range is normally expressed in the
same units as the test results (e.g., percentage, parts per million) obtained by the analytical
method. For assay tests, the ICH (5) requires the minimum specified range to be 80 to 120
percent of the test concentration, and for the determination of an impurity, the range to
extend from the limit of quantitation, or from 50 percent of the specification of each
impurity, whichever is greater, to 120 percent of the specification.
3.6 Limit of detection
The limit of detection is the point at which a measured value is larger than the uncertainty
associated with it. It is the lowest concentration of analyte in a sample that can be detected
but not necessarily quantified. The limit of detection is frequently confused with the sensitivity
of the method. The sensitivity of an analytical method is the capability of the method to
discriminate small differences in concentration or mass of the test analyte. In practical terms,
sensitivity is the slope of the calibration curve that is obtained by plotting the response
against the analyte concentration or mass.
In chromatography, the detection limit is the injected amount that results in a peak with a
height at least two or three times as high as the baseline noise level. Besides this
signal/noise method, the ICH (4) describes three more methods:
1. Visual inspection: The detection limit is determined by the analysis of samples with
known concentrations of analyte and by establishing the minimum level at which the
analyte can be reliably detected.
2. Standard deviation of the response based on the standard deviation of the blank:
Measurement of the magnitude of analytical background response is performed by
analyzing an appropriate number of blank samples and calculating the standard
deviation of these responses.
3. Standard deviation of the response based on the slope of the calibration curve: A
specific calibration curve is studied using samples containing an analyte in the range of
the limit of detection. The residual standard deviation of a regression line, or the standard
deviation of y-intercepts of regression lines, may be used as the standard deviation.
3.7 Limit of quantitation
The limit of quantitation is the minimum injected amount that produces quantitative
measurements in the target matrix with acceptable precision in chromatography, typically
requiring peak heights 10 to 20 times higher than the baseline noise. If the required precision
of the method at the limit of quantitation has been specified, the EURACHEM (22) approach
can be used. A number of samples with decreasing amounts of the analyte are injected six
times. The calculated RSD percent of the precision is plotted against the analyte amount.
The amount that corresponds to the previously defined required precision is equal to the
limit of quantitation. It is important to use not only pure standards for this test but also
spiked matrices that closely represent the unknown samples. For the limit of detection, the
ICH (5) recommends, in addition to the procedures as described above, the visual inspection
and the standard deviation of the response and the slope of the calibration curve.
Any results of limits of detection and quantitation measurements must be verified by
experimental tests with samples containing the analytes at levels across the two regions. It is
equally important to assess other method validation parameters, such as precision,
reproducibility and accuracy, close to the limits of detection and quantitation. Figure 6
illustrates the limit of quantitation (along with the limit of detection, range and linearity).
Figure 7 illustrates both the limit of detection and the limit of quantitation.
www.intechopen.com
13
Limit of detection
Signal/Noise = 2-3
Limit of quantitation
Signal/Noise = 10-20
Signal
Noise
Limit of quantitation
www.intechopen.com
14
3.11 Sensitivity
This is effectively the gradient of the response curve, i.e. the change in instrument response,
which corresponds, to a change in analyte concentration. Where the response has been
established as linear with respect to concentration, i.e. within the linear range of the method,
and the intercept of the response curve has been determined, sensitivity is a useful
parameter to calculate and use in formulae for quantitation. Sensitivity is sometimes used to
refer to limit of detection but this use is not generally approved.
3.12 Ruggedness (or robustness)
Ruggedness is normally evaluated during method development, typically by the originating
laboratory, before collaborating with other laboratories and is a measure how well a method
stands up to less than perfect implementation. In any method there will be certain stages,
which, if not carried out sufficiently carefully, will have a severe effect on method
performance, and may even result in the method not working at all. These stages should be
identified, usually as part of method development, and if possible, their influence on
method performance evaluated using ruggedness tests, sometimes also called robustness
tests. This involves making deliberate variations to the method, and investigating the
subsequent effect on performance. It is then possible to identify the variables in the method,
which have the most significant effect and ensure that, when using the method, they are
closely controlled. Where there is a need to improve the method further, improvements can
probably be made by concentrating on those parts of the method known to be critical.
Ruggedness tests are normally applied to investigate the effect on either precision or accuracy.
Reagent blanks: Reagents used during the analytical process (including solvents used
for extraction or dissolution) are analysed in isolation in order to see whether they
www.intechopen.com
2.
3.
4.
5.
6.
7.
15
contribute to the measurement signal. The measurement signal arising from the analyte
can then be corrected accordingly.
Sample blanks: These are essentially matrices with no analyte. They are difficult to
obtain but such materials are necessary to give a realistic estimate of interference that
would be encountered in the analysis of test samples.
Samples / test materials: Test materials taken from real samples are useful because of
the information they yield on interferences etc. which could be realistically encountered
in day-to-day work. If the true analyte content of a test material is accurately know it
can be used as a way of assessing the accuracy of the method. However the true analyte
content is usually difficult to determine unless there is the possibility of using other
methods which are known to show negligible bias.
Spiked material: These are material or solutions, which have been fortified with the
analyte(s) of interest. These materials or solutions may already contain the analyte of
interest so care is needed lest fortification inadvertently leads to levels outside of the
range of applicability of the method. Fortification with a known amount of analyte
enables the increase in response to the analyte to be measured and calculated in terms
of the amount added (assuming 100% recovery), even though the absolute amounts of
analyte present before and after the fortification are not know. Note that most methods
of fortification add the analyte in such a way that it will not be as closely bound to the
sample matrix as it would be if it was present naturally. Therefore, recovery
determinations obtained by fortification can be expected to be over-optimistic. The
nature of the spike obviously needs to be identified.
(Measurement) Standards: These are traditionally thought of as solutions of single
substances but in practice can be anything in which a particular parameter or property
has been characterized to a sufficient extent it can be used for reference or calibration
purposes.
Reference materials: frequently confused with certified reference materials. Reference
materials can be virtually any material used as a basis for reference, and could include
laboratory reagents of known purity, industrial chemicals, or other artefacts. The
property or analyte of interest needs to be stable and homogenous but the materials
does not need to have the high degree of characterisation, traceability and certifi cation
more properly associated with certified reference materials.
Certified reference materials: The characterisation of the parameter of interest in a certified
reference material is generally more strictly controlled than for a reference material, and in
addition the characterised value is certified with a stated uncertainty by a recognised
institution. Characterisation is normally done using several different methods, so that as
far as possible, any bias in the characterisation is reduced or even eliminated.
5. References
[1] Eric Reid, Ian D. Wilson, Methodological Survey in Biochemistry and Analysis Volume
20: Analysis for Drug and Metabolites, Including Anti-infective Agents, 1990,1-57
[2] U.S. Department of Health and Human Services, Food and Drug Administration.,
Guidance for Industry, Bioanalytical Method Validation, May2001
[3] Nowatzke W, Woolf E, Best Practices during Bioanalytical Method Validation for the
Characterization of Assay Reagents and the Evaluation of Analyte Stability in Assay
Standards, Quality Controls, and Study Samples, AAPS Journal. 9(2), 2007, E117-E122.
[4] Braggio S., Barnaby R. J., Grosi P, Cugola M., A strategy for validation of bioanalytical
methods,Journal of Pharmaceutical and Biomedical Analysis 1996, 14, 375- 388
www.intechopen.com
16
[5] Breda C.A., Breda M., Frigerio E.,Bioanalytical method validation: a risk-based
approach?, Journal of Pharmaceutical and Biomedical Analysis,35, 2004, 88789
[6] Nakashima Kenichiro, High-Performance Liquid Chromatography of drug of abuse in
biological samples, Journal of Health Science, 51(3) 272-277 (2005)
[7] Boulanger B., Chiap P. Dewe W., Crommen J., Hubert Ph., An analysis of the SFSTP
guide on validation of chromatographic bioanalytical methods: progresses and
limitations Journal of Pharmaceutical and Biomedical Analysis,32, 2003,753-765
[8] Causon Roger, Validation of chromatographic methods in biomedical analysis viewpoint
and discussion, Journal of Chromatography B, 689 (1997) 175-180
[9] Hartmann C., Smeyers-Verbeke J., Massart D. L., McDowall R.D., Validation of
bioanalytical chromatographic methods, Journal of Pharmaceutical and Biomedical
Analysis17, 1998,193218
[10] G. C. Hokanson, A life cycle approach to the validation of analytical methods during
pharmaceutical product development, Part II: Changes and the need for additional
validation, Pharm.Tech., Oct. 1994, pp. 92100.
[11] J. M. Green, A practical guide to analytical method validation, Anal. Chem. News &
Features, 1 May 1996, pp. 305A309A.
[12] B. Renger, H. Jehle, M. Fischer and W. Funk, Validation of analytical procedures in
pharmaceutical analytical chemistry: HPTLC assay of theophylline in an
effervescent tablet, J. Planar Chrom. 8:269278 (July/August 1995).
[13] Wegscheider, Validation of analytical methods, in: Accreditation and quality assurance
in analytical chemistry, edited by H. Guenzler, Springer Verlag, Berlin (1996).
[14] S. Seno, S. Ohtake and H. Kohno, Analytical validation in practice at a quality control
laboratory in the Japanese pharmaceutical industry, Accred. Qual. Assur. 2:140145
(1997).
[15] AOAC Peer-Verified Methods Program, Manual on policies and procedures, Arlington,
Va., USA (1998). http://www.aoac.org/vmeth/PVM.pdf
[16] P. A. Winslow and R. F. Meyer, Defining a master plan for the validation of analytical
methods, J. Validation Technology, pp. 361367 (1997).
[17] Zhoua Shaolian, Songb Qi, Tangb Yong, Weng Naidonga, Critical Review of
Development, Validation, and Transfer for High Throughput Bioanalytical LCMS/MS Methods,Current Pharmaceutical Analysis, 2005, 1, 3-14 3
[18] Kelley M, DeSilva, B.,Key Elements of Bioanalytical Method Validation for
Macromolecules., AAP Journal. 2007, 9(2), E156-E163.
[19] Findlay J.W.A., Smith W.C., Lee J.W.,Nordblom G.D., Das I., DeSilva B.S., Khan M.N.,
Bowsher R.R., Validation of immunoassays for bioanalysis: a pharmaceutical industry
perspective Journal of Pharmaceutical and Biomedical Analysis, 2000, 21,12491273
[20] Shah Vinod P., Ph. D, History of Bioanalytical Validation and Regulation: Evolution of a
Guidance Document on Bioanalytical Method Validation, AAPS 3rd Bioanalytical
Workshop on Quantitative Bioanalytical Methods Validation and Implementation:
Best Practices for Chromatographic and Ligand Binding Assays Crystal City,
Arlington, VA, May 1-3, 2006
[21] Mohammad A.,Tabrizi-Fard, Ho-Leung, Fung, Reversed-phase high-performance liquid
chromatography method for the analysis of nitro-arginine in rat plasma and urine,
Journal of Chromatography B, 679, 1996, 7-12
[22] Bmscheck Torsten, Meyer Hartmut, Wellhrner Hans Herbert, a High-performance
liquid chromatographic assay for the measurement of azathioprine in human
serum samples, Journal of Chromatography B, 675, 1996,287-294
www.intechopen.com
ISBN 978-953-307-682-9
Hard cover, 282 pages
Publisher InTech
How to reference
In order to correctly reference this scholarly work, feel free to copy and paste the following:
Kapil Kalra (2011). Method Development and Validation of Analytical Procedures, Quality Control of Herbal
Medicines and Related Areas, Prof. Yukihiro Shoyama (Ed.), ISBN: 978-953-307-682-9, InTech, Available
from: http://www.intechopen.com/books/quality-control-of-herbal-medicines-and-related-areas/methoddevelopment-and-validation-of-analytical-procedures
InTech Europe
InTech China