Content: Standard Operating Procedure (SOP) Use and Maintenance of A PH Meter
Content: Standard Operating Procedure (SOP) Use and Maintenance of A PH Meter
Content: Standard Operating Procedure (SOP) Use and Maintenance of A PH Meter
Content
1. Scope
2. Definitions and abbreviations
3. Personnel qualifications
3.1 Medical fitness
3.2 Education and training
4. Procedure
4.1 Principle
4.2 Samples
4.3 Equipment and materials
4.4 Reagents and solutions
4.5 Detailed instructions for use
4.6 Reading and recording
4.7 Quality control and maintenance
4.8 Waste management
5. Related documents
Annex. Maintenance logbook
1. Scope
Institution Code:
Laboratory name Standard Operating Procedure (SOP) Version: no.
Location Date: of release
Head/Responsible person Use and maintenance of a pH meter Page: 2 of 6
This SOP describes how to use and maintain a pH meter. The pH meter is intended
for accurate measurement of the pH of buffer solutions and/or culture media
solution used in the general diagnostic TB laboratory.
pH
Negative log10 of the hydrogen ion (H+) concentration.
NA: not applicable
3. Personnel qualifications
NA
hygiene requirements;
wearing and use of protective equipment and clothing;
prevention of incidents and steps to be taken by workers in the case of
incidents (biohazard incidents, chemical, electrical and fire hazards);
good laboratory practice;
organization of work flow;
use of equipment (operation, identification of malfunctions, maintenance).
The training shall be:
given before a staff member takes up his/her post;
strictly supervised;
adapted to take account of new or changed conditions; and
repeated periodically, preferably every year.
4. Procedure
4.1 Principle
4.2 Samples
NA
4.5.1 Calibration
Calibration must be carried out, in accordance with the manufacturer’s instructions,
once daily or, in case of infrequent use, at least on the day of pH testing. It must be
performed before the first measurement of the day. Record in the logbook (Annex).
Temperature variation affects pH measurement. Calibration and testing of the
sample must therefore be done at the same temperature ± 2 °C.
Select for calibration two buffer solutions that are within 3 pH units of the solution to
be tested. Discard contaminated or cloudy standard buffers.
Calibration results are acceptable if the pH of the buffer solution is within 0.1 pH
units of the expected value.
4.5.2 pH measurement
Before use, rinse the electrode with deionized water and blot dry with a soft, clean
paper towel.
Transfer the electrode to the test solution.
Compensate for the temperature if necessary.
Record the pH when the reading is stable (5–20 seconds after insertion of the
electrode into the solution)
Rinse the electrode with deionized water and store according to the manufacturer’s
instructions.
4.5.3 Cleaning and maintenance
Clean the pH meter with a soft, clean, damp paper towel after use. No solvents
should be used.
Replace the electrode filling solution on a regular basis, according to the
manufacturer’s instructions. Record in the logbook (see Annex).
Repair and service of the pH meter should be done by a qualified service
technician.
Calibration results are acceptable if the pH values of the reference buffer solutions
are within 0.1 pH units of the expected values.
NA
5. Related documents
FAILURE EVENTS
Corrective
Date Event Operator
action taken