Mailing Address: P.O. BOX 21145 Waco, TX 76702 Shipping Address: West Hwy. 6 at Spur 412 Waco, TX 76712 Phone: (254) 848-4435 Fax: (254) 848-4209
Mailing Address: P.O. BOX 21145 Waco, TX 76702 Shipping Address: West Hwy. 6 at Spur 412 Waco, TX 76712 Phone: (254) 848-4435 Fax: (254) 848-4209
Mailing Address: P.O. BOX 21145 Waco, TX 76702 Shipping Address: West Hwy. 6 at Spur 412 Waco, TX 76712 Phone: (254) 848-4435 Fax: (254) 848-4209
WP-609
WP-612
WP-621
WP-629
WP-639
WP-641
The duplexer is factory-tuned to the exact frequencies appearing on the decal with use of a
spectrum analyzer/tracking generator and no further tuning or "touching up" is required.
(2)
(3)
The duplexer includes interconnecting cables between cavities which are critical in length.
Do not change.
(4)
With some transmitters, the length of the cable between the transmitter chassis and the
duplexer may have to be optimized to obtain a proper impedance match (see instructions).
INSTALLATION
The three input connectors are marked "Low Freq. Input", "High Freq. Input" and "Antenna".
Connect your equipment to the appropriate connector. The frequency is the only concern. All of
the four cavities are built to handle the power output of a transmitter.
(2)
Loosen the hex nut which locks the threaded tuning rod of the four
cavities and then rotate the threaded tuning rod of cavities 1 & 2 for
minimum loss into the analyzer at the pass frequency.
(3)
Loosen the hose clamps on each of the four stubs so the rexolite rod
can be adjusted (in and out). Leave the clamp slightly tight to
prevent the rexolite rod from sliding out of the stub. Use the rexolite
rods to adjust the notches on cavities 1 & 2 so as to obtain a
symmetrical response about the center frequency as shown in Plot 1.
(4)
Connect the equipment to the duplexer as shown in Figure 2. Set the analyzer to the
higher of the two frequencies.
(5)
Rotate the threaded rods on cavities 3 & 4 for minimum loss into the analyzer at the
pass frequency.
(6)
With analyzer still tuned to the higher frequency, adjust the rexolite
rod of stub on cavities 3 and 4 to obtain a sym-metrical response
about the center frequency as in Plot 1.
The duplexer is now in the proper preliminary state to "fine tune" to the distinct transmit and
receive frequencies.
NOTE: For all "fine tune" steps use horizontal and vertical scales on the analyzer which will yield
highly accurate and reliable frequency and amplitude measurements. Typical response curves are
given in Plots 1-5. The duplex curve (Plot 5) shows the full response of one side of the duplexer.
(8)
Rotate the threaded tuning rod of Cavities 3 & 4 for minimum loss at
center frequency into the analyzer. (See Plot 2) Tune one cavity then
the other since there is some interaction between the two cavities.
(10)
Set the analyzer center frequency to the lower of the two frequencies.
(11)
Rotate the threaded tuning rod of Cavities 1 & 2 for minimum loss at
center frequency into the analyzer. (See Plot 3)
(12)
Tighten the hex nuts which lock the tuning rods of Cavities 1 & 2.
(14)
Tighten the hex nuts which lock the tuning rods of Cavities 3 & 4.
(16)
(17)
(18)
(20)
Set the center frequency of the analyzer to the lower of the two duplex
frequencies.
(21)
(22)
(23)
(2)
The transmitter output power might become erratic, either too high or too low as
measured on a wattmeter.
(3)
The insertion loss of the duplexer might measure normal with sweep equipment but
measure too high or too low on a wattmeter when connected to the transmitter.
(4)
The reflected power might change when the length of cable between transmitter and
first cavity is changed.
(5)
Use of a ferrite isolator at output of transmitter solves the forward power and
reflected power problem, and eliminates change in reflected power when length of
cable between the isolator and the first cavity is changed. (In this instance, the
ferrite isolator can be used to reveal the symptoms or as a solution to the problem.)
The above problems might be reduced or minimized by optimizing the length of cable between the
transmitter and first cavity of the duplexer. The optimum length of cable can be found by the
following procedure:
(1)
Tune the transmitter into a 50 ohm dummy load according to the manufacturers
instruction.
(2)
Connect the duplexer to the transmitter. The transmitter output signal should feed
through the wattmeter, then through the duplexer then into a dummy load. If there
is an impedance mismatch the duplexer will detune the transmitter and the cable
length should be optimized.
(3)
Using short lengths of cable (no longer than 1" @ 900 MHz, 2" @ 460 Mhz, 6" @
160 MHz, 24" @ 40 MHz) or right angle elbow connectors, gradually increase the
length of the cable between the transmitter and duplexer until the optimum length
(no detuning effect) is found. Up to four of these short cables should be tried.
(4)
When the proper cable length is determined, i.e., when the system is operating as it
should, replace all of the short and long lengths of cables, right angle elbow
connectors and wattmeter with one continuous length of cable of equivalent
electrical length. Do not overlook the fact that the wattmeter was a part of the cable
length between the transmitter and duplexer during this process. If the wattmeter is
removed from the circuit, the cable length without the wattmeter must be
lengthened by some amount in order to be the same equivalent electrical length as
the cable with the wattmeter in the circuit.