5070 - 1.6-1.7 AST Sample Tubes
5070 - 1.6-1.7 AST Sample Tubes
5070 - 1.6-1.7 AST Sample Tubes
Application
These sample tubes are used in combination with the Ankersmid ASP probes
in order to sample the gas in the optimal section of the gas-stream. For a
typical installation of the probe-tip, a position in the middle third of the gas
stream is advised.
For dust concentrations higher than 2g/m, we advise fitting a pre-filter in
combination with an extension tube.
Description
Different solutions for different applications:
Maximum sample tube temperature is depending on the horizontal length inside this temperature zone.
Application
Tube type
ADT 080
(Demistor tube)
Max. T
[C]
Material
Connection
Outer
diameter
90C
PTFE
G o
40-70
150C
PTFE
G o
22
600C
SS316
G o
22
AST 431-432-433-434
900C
Hastelloy C
AST 435
HCl or high
corrosive gas
AST 421-422-423-424
G o
22
1400C
Kanthal
G o
20
450C
Titanium
G o
22
Application
The electrically heated Ankersmid sample probe tube AST 05x-10x15x-20x are used in extractive sampling systems to avoid cooling and
condensation of the sample.
Condensation, in combination with a high dust load, can result in blockage
of the probe. This is to be strictly avoided as sample gases may be
absorbed into the condensate after cooling and will be undetectable. An
extra stainless filter can be mounted on top of the tube in case of very high
dust levels (> 10 g/m).
Description
Technical data
AST05x
AST10x
AST15x
AST20x
AST 002
AST 003
AST 004
AST 005
Optional
500mm
1000mm
1500mm
2000mm
500C
500C
450C
400C
200C
200C
200C
200C
Pre-filter
Optional
G3/4"i
Dust loading
max. 2 g/m3
200ml/m
5 bar g
Ambient temperature
-20C to +70C
Storage temperature
-30C to +70C
Approx. 0,5h
Power supply
Electrical connection
Electrical standard
Degree of protection
Mounting flange
Material of gas wetted parts
230VAC, 500W
230VAC, 800W
230VAC, 1200W
230VAC, 1500W
115VAC, 500W
115VAC, 800W
115VAC, 1200W
115VAC, 1500W
2 x 2.5mm + 2.5mm
EN 61010, EN60519-1
IP54 EN 60529
DN65 PN6
Stainless steel 316