Module 5. Draft Protocol For Gamma Scanning of Industrial Process Columns
Module 5. Draft Protocol For Gamma Scanning of Industrial Process Columns
Select a proper reference (zero) point (manhole, external pipes, temperature probes, etc.).
Include the mechanical design data into an industrial design software (Autocad, Autosketch,
Labview, Excell etc.); all positions of internal and external structures are referred to the point zero.
For example, some of the characteristics of a column are listed as follows
Internal diameter
Wall thickness
Trays (type)
Temperature profile
Refluxes
Note: Approximately 200 mm needs to be added to the column diameter to take into account the
distance of source and detector positions from the column.
The value of D, mentioned in the above equation corresponds to the transmitted radiation intensity
which will give 5000 - 7000 cps (counts per second) in vapour space of the column (vapour base line);
and will decrease to around 1000 - 2000 cps in the tray level (liquid base line). This will give a good
compromise between the contrast (between liquid & vapour phases) and the accuracy of counting.
5.1.4. Example of source activity calculation
A gamma source is needed to scan a distillation column, which has the following parameters:
Internal diameter = 2.9 m, Wall thickness = 15 mm.
Normally the sensitivity of 2x 2 NaI(Tl) detector for 60Co is 7500 cps/(mR/h). The NaI(Tl) detector
sensitivity is determined by placing a sealed source of 60Co or of 137Cs at a distance where the dose
rate is 1 mR/h. The Na(Tl) detector sensitivity for 60Co is lower than for the 137Cs, because the
detector Na(Tl) is more sensitive to low energies of 137Cs than to high energies of 60Co.Calculation
of the 60Co activity for scanning a column of 2.9 m internal diameter gives:
=
2 2
1 2.9 + 2 2 2
1.332
30 22
18.6
Table 5.1 gives the 60Co activity (estimated, rounded off) in function of the column diameter (wall
thickness of 15 mm and the dose rate of 1 mR/h):
TABLE 5.1. COLUMN DIAMETER AND ESTIMATED GAMMA SOURCE ACTIVITY
smaller and for gross scanning of a whole column larger slit opening shall be selected. For example
one shall select 5 mm slit opening for source collimator coupled with 10 mm slit opening of detector
collimator to scan a column up to 2 meter diameter.
Since higher activities sources are required for the examination of large diameter and thick-walled
columns, it is necessary to design source collimators which could be transported in the source housing
itself. Such source housing-cum-collimator devices shall be designed using heavy alloy or depleted
Uranium to reduce the weight of such housing and to minimize the radiation hazard to the scanning
team and the general public. For example, 100 mm diameter source collimator made of tungsten of
less than 10 kg weight is recommended.
Figure 5.1 shows a set of source (right) and detector (left) collimators in use during a normal column
scanning.
FIG. 5.1. Example of detector holder (left) and source holder (right)
5.1.6. Selection of traversing mechanism
Source and detector traversing mechanism shall be selected so that source and detector collimators
openings always remain aligned in the same horizontal plane. Great care shall be taken to minimise
the swing of source and detector particularly while examining the columns with small diameters (less
than 1.5 m).
Depending upon availability of the equipment with a particular scanning group the traversing
mechanism shall be selected. Preference should be given to the ease of handling and comfort of the
scanning team being familiar with the type of traversing mechanism. The complexity of external
hardware and the working conditions in the plant mostly have an impact on the selection of the
mechanism.
Manual Traversing Mechanism
Normally manual scanning (as shown in Fig. 5.2) shall be selected for scanning a smaller section of a
column as well as where there is complex external obstructions, thus limiting the movement of source
and detector collimators.
FIG. 5.4. Typical orientations of source and detector for tray-typed columns
However, scanline may cut through the downcomers when the information about downcomer flooding
is required.
Scanning of packed-bed columns:
To inspect packed beds, the grid scanning shall be applied. This technique can also be used for the
inspection of the correct installation of distributors as well as the correct distribution of incoming
liquid feed of the sieve-tray column. The schematic of grid scanning is shown in Fig. 5.5.
In general, four grids are recommended for the inspection of packed columns. However, the number
shall be increased when more information required.
FIG. 5.5. Typical orientations of source and detector for packed-bed columns.
FIG. 5.7. Traversing mechanism and data acquisition system for automatic scanning system
Scanning detection system shall be checked in laboratory with a check source. Details on this
procedure may be referred to Protocol on Maintenance and Calibration of Data Acquisition System
for Radiotracers and Sealed source Application, RAS/8/111-P4. However, a brief guideline is given
in the following paragraphs.
Verification of the operational condition of detection system.
Calibration for NaI(Tl) scintillation detector:
Scaler-Ratemeter check:
5.2. EXECUTION
Upon arrival at the work site, supervisor will ensure that a suitable permit to work is obtained.
He/she will inspect the work site and ensure that there is safe access. The scanning team will carry out
the scan in the agreed manner. Any deviation from the agreed scanning procedure must be approved
by the project manager after due consultation with the client.
Following steps need to be taken in carrying out gamma scanning:
The form acts as a record and is useful for future repetitions of the scanning of the same
column.
10
11
12
13
Damaged tray
14
Displaced packed-beds
15
5.4. REPORTING
The report shall contain a detailed description of the column as well as the description of the scanning
performances (scan number, source and detector direction scan lines in relation to the trays).
Scanning report should include as minimum the following information:
16
Executive summary
Column characteristics
References
Annexure