Edwards Vacuum Pump EDP Dry Pump Instruction Manual
Edwards Vacuum Pump EDP Dry Pump Instruction Manual
Edwards Vacuum Pump EDP Dry Pump Instruction Manual
Instruction Manual
Drystar EDP60, EDP120, EDP200 and EDP300
Chemical Dry Vacuum Pumps
Amen dment to
A705-20 -880
A705-21 -880
A705-51 -880
A705-52 -880
2 . C orr ect i on
Pump operating temperature 55C 540 l/h 600 l/h 600 l/h 720 l/h
Pump operating temperature 65C 120 l/h 180 l/h 120 l/h 240 l/h
Pump operating temperature 90C 12 l/h 51 l/h 42 l/h 120 l/h
CONTENTS
PAGE
Section Page i
1 INTRODUCTION 1
2 TECHNICAL DATA 7
3 INSTALLATION 15
3.1 Safety 15
3.2 Unpack and inspect 15
3.3 Locate the pump 16
3.4 Check the coolant level 16
3.5 Check the gearbox oil-level 18
3.6 Electrical connections 18
3.6.1 Introduction 18
3.6.2 Connect the electrical supply to the pump-motor 19
3.6.3 Connect the thermal snap-switch 19
3.7 Check the direction of pump rotation 22
3.8 Connect the cooling-water supply 22
3.9 Connect the shaft-seals purge nitrogen supply 24
3.10 Connect the pump-inlet and outlet 24
Ipsitech 8140-04
Feb 04 i Issue L
Drystar EDP Chemical Dry Vacuum Pumps
CONTENTS (CONTINUED)
PAGE
Section Page ii
4 OPERATION 31
5 MAINTENANCE 37
5.1 Safety 37
5.2 Maintenance plan 37
5.3 Check the gearbox oil-level and fill the gearbox
with oil (if necessary) 39
5.4 Check the shaft-seals purge supply 39
5.5 Remove the inlet filter 39
5.6 Inspect the pipelines and connections 41
5.7 Inspect the pressure relief valve and replace the
hinge bushes, flap and'O' ring (if necessary):
EDP200 and EDP300 only 41
5.7.1 Remove the valve from the pump 41
5.7.2 Clean and inspect the valve 41
5.7.3 Refit the valve to the pump 41
5.8 Change the pump oil and clean the oil-level sight-glasses 43
5.9 Relubricate/inspect the rotor bearings 44
5.9.1 Introduction 44
5.9.2 Remove the bearing cover and bearing caps 45
5.9.3 Clean, inspect and relubricate the rotor bearings 45
5.9.4 Refit the bearing caps and bearing cover 45
5.10 Check the coolant level and refill if necessary 46
5.10.1 Refill the pump with BOC Edwards coolant 46
5.10.2 Fill the pump with a different coolant 46
5.10.3 Check for coolant leaks 48
5.11 Flush the pump with cleaning solution 48
5.11.1 Introduction 48
5.11.2 Flush a pump which has been operating 48
5.11.3 Flush a seized pump 49
5.12 Clean the cooling system 50
5.12.1 Drain the coolant from the cooling system 50
5.12.2 Clean the cooling-water filter 52
Feb 04 ii Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
Section Page iii
6.1 Storage 71
6.2 Disposal 71
7.1 Introduction 73
7.2 Service 73
7.3 Spares and maintenance kits 73
7.4 Accessories 73
7.5 Ordering option 73
PAGE
iv
ILLUSTRATIONS
Figure Page
Feb 04 iv Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
v
TABLES
Table Page
If you change this manual, you may also have to change A705-51-880.
Feb 04 v Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
vi
Issue L vi Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
INTRODUCTION
maintenance instructions for the BOC Edwards
frame (6); this configuration ensures that liquids in
Drystar EDP60, EDP120, EDP200 and EDP300 the process gases easily drain through the pump.
Chemical Dry Vacuum Pumps. You must use the
Drystar pumps as specified in this manual. The pump is a three-stage, positive displacement
rotary pump in which pairs of intermeshing rotors
Read this manual before you install and operate your (mounted on common shafts) are held in correct
pump. Important safety information is highlighted as phase relation by a pair of timing-gears. The timing-
WARNING and CAUTION instructions; you must gears and the adjacent double-row angular contact
obey these instructions. The use of WARNINGS ball bearings are oil lubricated.
and CAUTIONS is defined below.
1.3 Shaft-seals purge
WARNING
Refer to Figure 1. The purge pipeline (17) delivers a
Warnings are given where failure to nitrogen purge to the shaft-seals. This nitrogen
observe the instruction could result purge: ensures that the shaft-seals are maintained at
in injury or death to people. a positive pressure during pump operation; prevents
the entry of corrosive or toxic process vapours into
CAUTION the pump gearbox; prevents contamination of the
process gases by pump oil; prevents damage to the
Cautions are given where failure to observe the shaft-seals by debris.
instruction could result in damage to the
equipment, associated equipment and process. Accessory kits are available to provide final stage gas
ballast and inlet purge facilities on the pump: refer to
Section 7.4.
The following IEC warning labels appear on the
pump:
1.4 Temperature control system
(Continued on page 2)
Feb 04 1 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
An adjustable thermal snap-switch (18) is fitted to The EDP200 and EDP300 pumps have a torque
the pump-body. The output of the thermal snap- limiter. The basic EDP60 and EDP120 pumps have a
switch will go open circuit when the temperature of drive coupling, however you can order these pumps
the pump-body is too high. Use the thermal snap- to be supplied with a torque limiter fitted (refer to
switch output to shut down the pump if the pump is Section 1.10). Refer to Section 1.6.2 for a
too hot. description of the drive coupling operation, and
refer to Section 1.6.3 for a description of the torque
1.5 Pressure relief valve (EDP200 and limiter operation.
EDP300 only)
1.6.2 Drive coupling operation
Refer to Figure 1. The EDP200 and EDP300 pumps
have a pressure relief valve (26) fitted in a pipe The basic EDP60 and EDP120 pumps have a flexible
between the pump-outlet and the last stage of the drive coupling which transmits the drive from the
pump. The valve is normally held closed by its own pump-motor to the pump rotors.
weight, but opens depending on the pump-inlet
Refer to Figure 20. A coupling hub (16) is fitted to
pressure, as follows:
the pump shaft (17) and a drive hub (5) is fitted to
At pump-inlet pressures of 300 torr and the motor shaft (7). Flexible drive coupling bushes
above on the EDP200, and at pump-inlet (18) fit in the coupling hub and engage in the drive
pressures of 188 torr and above on the dogs (4) on the drive hub.
EDP300, the interstage pressure forces the
valve open. This allows process gases to pass 1.6.3 Torque limiter operation
directly from the second stage into the pump-
CAUTION
outlet, without compression in the third stage
of the pump. If you continue to operate the pump-motor when
At pump-inlet pressures below those the torque limiter has tripped and the pump-motor
is decoupled from the pump, you may cause
specified above, the interstage pressure is low
excessive wear to the torque limiter.
and the valve is held closed. Process gases
pass through all stages of the pump; that is the
Refer to Figure 1. The pump-motor (12) drives the
process gases are compressed in the third
pump rotors through a torque limiter. When the
stage before they pass into the pump-outlet.
torque required to turn the pump rotors is
The pressure relief valve allows the pump to provide excessive (for example, if the pump is accidentally
a constant pumping speed from atmospheric flooded with process liquids), the torque limiter
pressure down to 7.5 torr and also prevents automatically trips. When the torque limiter trips,
excessive electrical power consumption by the it decouples the pump-motor from the pump, and
pump-motor when the pump starts. the pump-motor no longer drives the pump rotors.
This prevents damage to the pumping mechanism.
Issue L 2 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
When the torque limiter trips, the pump-motor will 1.8 Explosion-proof and standard
continue to operate, however there will be a versions of the pump
PAGE
significant fall in the electrical current consumed by 3
the pump-motor. We recommend that: European certification
Your control equipment should monitor the The pumps are available in both explosion-proof and
INTRODUCTION
current consumed by the pump-motor, in standard versions. The explosion-proof pumps have
order to determine when the torque limiter been certified for safe use in Zone 1 and Zone 2
has tripped. hazardous areas and (when fitted with the correct
pump-inlet and pump-outlet flame arrestors) are
When the torque limiter has tripped, your
permitted to pump gases and vapours allocated to
control equipment should switch off the
Group IIA and IIB as classified by BS 5345 Part 2:
pump-motor; operation in the decoupled
1990. Certification was carried out on the basis of
state may cause excessive wear to the torque
BS 5501 Part 5 (EN50018).
limiter.
If all of the following conditions apply, all surfaces of
The torque limiter is enclosed in the coupling
the pump will be below 275 oF at all times:
housing which has two removable coupling cover
guards (14). You can remove these covers to reset Ambient temperature 77 oF.
the torque limiter and recouple the pump-motor to
the pump (refer to Section 4.4). Pump outlet pressure 2.2 psig.
Cooling-water supply temperature 68 oF.
The coupling hub and flexible drive coupling bushes
used in the torque limiter (see Figure 19) are Cooling-water flow 158 US gallons h-1.
identical to those used in the EDP60 and EDP120
If any of the above conditions do not apply, under
drive coupling.
certain operating conditions, even though the main
1.7 Pump frame body of the pump will remain below 275 oF, parts of
the exhaust manifold may exceed this temperature.
Refer to Figure 1. The pump is vertically mounted in
a robust frame (6), which has lifting-bolts (5). The If the cooling-water supply does not meet the
frame on EDP200 and EDP300 pumps has lower requirements of Section 2.3, or if there is reduced
cross-members (11) which can be used to move the cooling-water flow through the pump (due, for
pump with a fork-lift truck. example, to a blockage), the maximum surface
temperature of the pump body will not exceed
The cooling-water and gas purge connections are on 392 oF.
a services panel (8) on the frame. The TCV (7) is
fitted to the rear of the services panel. (Continued on page 6)
Feb 04 3 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
4
INTRODUCTION
Issue L 4 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
5
INTRODUCTION
17. Shaft-seals purge pipeline 23. Torque limiter reset tool * and steel rod *
18. Coolant jacket thermal snap-switch 24. Frame lower cross members
19. Coolant overflow pipe 25. Pump-outlet
20. Bleed plug (EDP60/EDP120 only) 26. Pressure relief valve (EDP200/EDP300 only)
21. Heat exchanger 27. Pump case temperature measurement point
22. Pump-motor terminal-box
Feb 04 5 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
classification US equivalent
Zone 1 Division 1
Zone 2 Division 2
1.9 Accessories
Issue L 6 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
2 TECHNICAL DATA
PAGE
2.1 Operating and storage conditions 7
TECHNICAL DATA
Maximum ambient operating humidity 90% RH
Maximum operating altitude 3281 ft (1000 m)
Environment External, industrial location
Warm-up time * 45 min
Cool-down time 3h
* To a pump operating temperature of 131 oF.
To a safe temperature for maintenance.
2.2 Performance
Full load and no-load current ratings Refer to the motor rating label
Feb 04 7 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
8 Supply pressure 29 to 147 psig
Thermal snap-switch
Operating temperature range 120 to 212 oF adjustable
Contact rating
Maximum voltage 240 V
Maximum current (inductive load) 5 A (inductive), 12 A (resistive)
Issue L 8 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
The pump is supplied filled with BOC Edwards coolant which is monopropylene based, and is both an PAGE
9
antifreeze and a corrosion inhibitor. Spare BOC Edwards coolant is available: refer to Section 7.3. If you
will use another coolant type to replenish lost coolant:
TECHNICAL DATA
It must prevent the formation of visible oxide sludge and scale deposits.
It must provide frost protection down to a temperature of 7 oF, when mixed in accordance with
the manufacturers recommendations.
It must be compatible with the materials of construction of the cooling-jacket; that is: SG iron,
copper, brass and fluoroelastomer (Viton) seals.
The coolants shown in Table 8 may be suitable for use in the EDP pumps. All of these coolants are
amine free, automotive grade, ethylene glycol antifreezes and must be diluted to between 35% to 50%
by volume with water to provide the required cooling protection. However, we recommend that you
use BOC Edwards coolant; BOC Edwards cannot guarantee that other types of coolant will provide the
best corrosion protection for the pump.
Manufacturer Product
BP (Chemicals) NAPGEL C2230 (Universal)
Texaco Lubricants Texaco Engine Coolant ETX 6024
Castrol Castrol Antifreeze
Note: BOC Edwards Material Safety Data Sheets for some of the oils and greases specified below are
available on request
Gearbox
Oil capacity: minimum 0.89 US gal
Oil capacity: maximum 1.02 US gal
Recommended oil (supplied) Mobil SHC 629
Feb 04 9 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
10 EDP60 EDP120 EDP200 EDP300
2.11 Connections
Pump-inlet connection
EDP60 2 inch ANSI 150 lbf raised face flange
EDP120, EDP200, EDP300 3 inch ANSI 150 lbf raised face flange
Pump-outlet connection
EDP60, EDP120 1.5 inch ANSI 150 lbf raised face flange
EDP200, EDP300 2 inch ANSI 150 lbf raised face flange
Recommended pump-inlet and PTFE envelope gaskets: KLINGER milled type with a
pump-outlet seals 1.5 mm full-face insert
Cooling-water supply
Inlet 1/ inch NPT female
2
Outlet 1/ inch NPT female
2
2.12 Area classification in accordance with Article 500, National Electrical Code
Explosion-proof pumps
Standard pumps
Issue L 10 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
Component Materials 11
Stator
Rotor BS2789, GR420/12, 420 N mm-2, 12% ELONG
TECHNICAL DATA
Heat exchanger housing
Bearing housing
Shaft sleeves
Throwers BS 970 Pt. 1 1991
Valve body Grade 416S21
Valve seat
Valve pad
Feb 04 11 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
12
TECHNICAL DATA
1. Pump-inlet
2. Pump-outlet
3. Fixing hole: 0.7 inch (4 off)
4. Cooling-water outlet connection
5. Cooling-water inlet connection
6. Nitrogen supply inlet connection
Issue L 12 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
13
TECHNICAL DATA
Key EDP60 EDP120 EDP200 EDP300
A 621/2 633/4 661/4 673/4
B 553/4 571/4 591/2 61
C 447/8 447/8 451/4 451/4
Feb 04 13 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
14
Issue L 14 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
INSTALLATION
Obey the safety instructions listed 3.2 Unpack and inspect
below and take note of appropriate
precautions. If you do not, you can WARNING
cause injury to people and damage to
equipment. Use suitable lifting equipment to
move the pump. If you do not, you
can injure yourself or damage the
A suitably trained and supervised technician
pump. Refer to Section 2.3 for the
must install your EDP pump. mass of the pump.
Ensure that the installation technician is
familiar with the safety procedures which 1. Use a fork-lift truck or a pallet truck to place
relate to the products pumped. Wear the the pallet in a convenient position.
appropriate safety-clothing when you come
2. Remove the packing material from around the
into contact with contaminated components.
pump.
Dismantle and clean contaminated
components inside a fume-cupboard. 3. Remove small items packed with the pump.
Vent and purge the process system before 4. Inspect the equipment. If the pump or any of
you start installation work. the other items is damaged, notify your supplier
and the carrier in writing within three days;
Check that all the required components are
state the Item Number of the pump together
available and of the correct type before you
with your order number and your suppliers
start work.
invoice number. Retain all packing materials for
Disconnect the other components in the inspection. Do not use the pump if it is
process system from the electrical supply so damaged.
that they cannot be operated accidentally.
5. Check that you have received the items listed in
Do not reuse 'O' rings if they are damaged. Table 12. If any of these items is missing, notify
your supplier in writing within three days.
Leak test the system after installation and seal
any leaks found, to prevent leakage of 6. If the pump is not to be used immediately,
hazardous substances out of the system and replace the packing materials. Store the pump in
leakage of air into the system. suitable conditions as described in Section 6.1.
If you will pump substances that may attack or
react with the materials of construction of the
pump (see Section 2.13), there may be a risk of
a leak, or of contamination of the pump oil, or of
a loss of pump performance. If you are in any
doubt, consult BOC Edwards for advice.
Feb 04 15 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
!
* Pumps with a torque limiter only.
Table 12 - Checklist of items
3.3 Locate the pump Use suitable bolts through the four fixing holes
(Figure 1, item 9) to secure the pump in position.
WARNING
3.4 Check the coolant level
Use suitable lifting equipment to
move the pump. If you do not, you CAUTION
can injure yourself or damage the
pump. Refer to Section 2.3 for the Ensure that you correctly fill the EDP60 and
mass of the pump. EDP120 pumps with coolant, as described below.
If you do not, an air-lock may develop in the
WARNING cooling-jacket and the pump may overheat.
When you move or install the pump, Note: BOC Edwards coolant is available as a spare:
the baseframe of the pump must be refer to Section 7.3.
no more than 10o from horizontal. If
it is, it may topple.
1. Refer to Figure 3, detail A. Press down the
coolant filler-cap (1), turn it anticlockwise and
WARNING remove it from the pump.
Ensure that the cooling-air flow 2. Refer to Figure 3, detail A. Remove any dirt or
around the pump-motor is not water-scale from the seal of the filler-cap (1)
restricted. and from the filler-tube (4).
3. Refer to detail B. Check the level of coolant (6);
Note: If you will operate the pump in an environment
if it is more than approximately 1 inch below
with an ambient temperature of 32 oF or
the bottom of the filler-neck (5), continue at
lower, contact your supplier or BOC Edwards
Step 4, otherwise continue at Step 10.
for advice.
4. Remove the cap from a new container of
Use suitable lifting-equipment to move the pump; coolant. Place a clean funnel into the container
use one of the following methods: and fill the container with clean water.
Use the four lifting bolts on the frame 5. Replace the cap on the container, shake the
(Figure 1, item 5). container to fully mix the water and coolant,
then remove the cap.
On EDP200 and EDP300 pumps, use a fork-
lift or pallet truck through the frame lower 6. For EDP200 or EDP300 pumps continue at Step
cross members (Figure 1, items 11 and 24). 9, otherwise continue at Step 7.
Locate the pump on a firm, level surface. Ensure that (Continued on page 18)
the surface is clean and free from debris and
contamination (such as oil).
Issue L 16 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
17
INSTALLATION
Feb 04 17 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
7. Refer to Figure 1. Remove the bleed plug (20) On pumps without a torque limiter, we recommend
from the heat exchanger of the EDP60 or that you configure the high current setting on the
PAGE
18
EDP120 pump. current monitor to switch off the pump-motor at a
suitable overload current. This overload current
8. Refer to Figure 3. Pour the coolant mix into the
must not exceed the maximum current rating
pump through the filler-tube (4) until coolant
shown on the rating plate on the pump-motor.
INSTALLATION
9. Refer to Figure 3. Pour the coolant mix into the Configure the low current setting on your
pump through the filler-tube (4) until the current monitor to switch off the electrical
coolant level (6) is approximately 1 inch below supply at a current consumption consistent
the bottom of the filler-neck (5). If necessary with the no-load current of the pump-motor.
repeat Steps 4 to 9 to mix and add more Fit a rotation sensor to the coupling housing,
coolant. and connect the sensor to your control
10. Refit the coolant filler-cap (1); press it down and equipment, which must be configured to
turn it clockwise to secure it to the pump. switch off the electrical supply when the
rotation sensor indicates that the torque
3.5 Check the gearbox oil-level limiter has tripped.
The pump is supplied filled with oil. Before you On pumps with a torque limiter, if you do not
operate the pump, check that the gearbox oil-level configure your current monitor to automatically
is correct. Refer to Figure 1 for the locations of the switch off the pump-motor, note that you should
two oil-level sight-glasses on the pump. The oil-level not continue to operate the pump for more than 8
must be between the MIN and MAX marks on the hours if the torque limiter has tripped: refer to
bezel of either of the two oil-level sight-glasses (see Section 4.
Figure 11, detail A). If necessary, pour more oil into
the gearbox: refer to Section 5.3. Earth (ground) studs are available on the pump
(refer to Figure 5). If required, you can connect
3.6 Electrical connections suitable earthing (grounding) strips to these earth
(ground) studs to provide additional earthing
3.6.1 Introduction (grounding) protection.
Issue L 18 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
3.6.2 Connect the electrical supply to the 3.6.3 Connect the thermal snap-switch
pump-motor
PAGE
WARNING 19
WARNING
Obey the safety instructions given
On explosion-proof pumps, you below. If you do not, the installation
INSTALLATION
must use a suitable explosion-proof may not be safe, and there may be a
cable gland to connect the electrical risk of fire or explosion and injury to
supply to the pump-motor. If you do people.
not, the installation will not be
explosion-proof. You must connect the thermal snap-switch so
that the pump stops when the thermal snap-
Note: You must make the wiring connections to the switch opens. If you do not, there may be a
pump-motor in accordance with the National risk of fire or explosion.
Electrical Code and with approved local and
Incorporate a manual reset device in your
site practices.
control equipment. If you do not (and a fault
Connect your electrical supply to the pump-motor which causes the thermal snap-switch to open
through a suitable contactor. The contactor must is not corrected), the pump will automatically
incorporate a motor protection circuit-breaker switch on again when it cools down. If you
which meets the full load current shown on the have started maintenance or fault finding on
motor rating label. the pump, there will then be a risk of fire or
explosion and injury to people.
Use a suitably rated six-core cable (three phase On explosion-proof pumps, you must use a
wires, earth (ground) wire and two control wires) suitable 1/2 inch NPT cable gland to connect
to connect the electrical supply and your control to the thermal snap-switch. If you do not, the
equipment to the pump-motor, as described in the installation will not be explosion-proof.
motor instruction manual supplied with the pump.
Do not connect the thermal snap-switch into
an intrinsically safe circuit if the snap-switch
has been previously connected in any other
type of circuit. If you do, the contacts in the
snap-switch may have been damaged, and the
circuit may not be safe.
Feb 04 19 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
You must connect the thermal snap-switch to the If you connect to the thermal snap-switch as
electrical-overload control-loop of your contactor, described below, the output from the thermal snap-
PAGE
20
so that the contactor will automatically switch off switch will be normally closed and will open when
the pump if it is too hot: refer to Figure 5. the pump is too hot.
The thermal snap-switch will reset (that is, close 1. Remove the four bolts which secure the cover
INSTALLATION
again) when the pump cools down to a preset to the top of the thermal snap-switch (Figure 1,
temperature (see Section 2.6). We therefore item 18), then remove the cover.
recommend that your control equipment
2. Pass a suitably rated two-core cable through the
incorporates a manual reset device so that the pump
cable-gland.
does not automatically switch on again when it cools
down. 3. Refer to Figure 4. Connect one wire of the
cable to terminal 1 on the terminal block (2).
Use the following procedure to connect to the Connect the other wire of the cable to terminal
thermal snap-switch. 3.
Issue L 20 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
21
INSTALLATION
A Pump-motor connections
B Control circuit
Feb 04 21 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
3.7 Check the direction of pump rotation 4. If the torque limiter plates or coupling hub do
not rotate in the correct direction (shown by an
PAGE
22
WARNING arrow on the pump-motor, Figure 1, item 10),
the direction of rotation is incorrect. If the
You must ensure that the direction direction of rotation is incorrect:
of rotation of the pump is correct
INSTALLATION
before you operate the pump. If you Isolate the pump from the electrical supply.
do not, and the pump direction of
rotation is incorrect, the inlet Reverse any two of the electrical supply
pipeline will be pressurised and may phase-wires in the pump-motor terminal-box:
be damaged and there will be a risk of refer to the motor instruction manual.
injury to people or explosion or fire.
Repeat Steps 3 and 4 to ensure that the
direction of rotation is now correct.
WARNING
5. Refit the coupling cover guard (Figure 1, item
When you check the direction of 14) and secure with the four bolts. Tighten the
rotation with the coupling cover bolts to a torque between 2.2 and 3.6 lbf ft (3
guard removed, take suitable
and 5 N m).
precautions to prevent injury from
the exposed rotating mechanisms.
3.8 Connect the cooling-water supply
Issue L 22 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
23
INSTALLATION
Feb 04 23 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
3.9 Connect the shaft-seals purge Ensure that your system can provide adequate
nitrogen supply gas ballast and/or inlet purge to dilute toxic
PAGE
24
gases to safe limits. If you do not, there will be
WARNING a risk of emission of dangerous gases.
Your nitrogen supply pressure must If you will pump flammable and explosive
INSTALLATION
comply with the requirements of gases, fit flame arrestors and ensure that your
Section 2.5. If it does not, the shaft- system can provide adequate gas ballast and/
seals purge pipelines may become or inlet purge to dilute to the gases to safe
over-pressurised and may explode. levels. If you do not, there may be a danger of
fire or explosion.
Use a rigid metal (such as stainless steel) pipeline
Fit BOC Edwards flame arrestors if you will
with an outside diameter of 1/4 inch for your
nitrogen supply. If you will fit an acoustic enclosure, pump Class I Groups C and D gases. If you do
ensure that the pipeline wil not obstruct th not, there may be a risk of flame propagation
enclosure. along the inlet and exhaust pipelines.
We recommend that you install an automatically 3.10.1 Connect the pump-inlet to your
operated isolation-valve in your nitrogen supply process system
pipeline, configured so that:
WARNING
The shaft-seals purge nitrogen supply is on
whenever the pump is on. Do not exceed the load limits on the
pump-inlet flange as specified in
The shaft-seals purge nitrogen supply is off
Figure 7. If you do, there will be a risk
whenever the pump is off. of leakage of process gases from the
pump, or of damage to the pump.
Refer to Figure 6 and use the following procedure to
connect your shaft-seals purge nitrogen supply:
Note: Flame arrestors suitable for use with the pump
1. Remove the mating connector from the fittings are available as accessories: refer to
Section 7.4.
kit and fit it onto your nitrogen supply pipeline.
2. Remove the red blanking cap from the nitrogen When you connect the pump to the process system:
inlet (7) on the services panel and fit the
connector on your nitrogen supply pipeline to Support process pipelines to stop the
the nitrogen inlet connector (7). transmission of stress to pipeline joints.
Issue L 24 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
25
INSTALLATION
1. Pump-inlet
Maximum force Pump-inlet Pump-outlet
2. Pump-outlet
FX 530 N 120 lbf 357 N 80 lbf
F Force FY 446 N 100 lbf 290 N 65 lbf
M Moment FZ 669 N 150 lbf 446 N 100 lbf
R Resultant FR 959 N 215 lbf 647 N 145 lbf
X Load orientation Maximum moment Pump-inlet Pump-outlet
Y Load orientation
Z Load orientation MX 476 Nm 350 lbf ft 231 Nm 170 lbf ft
MY 360 Nm 265 lbf ft 177 Nm 130 lbf ft
MZ 238 Nm 175 lbf ft 115 Nm 85 lbf ft
MR 645 Nm 475 lbf ft 313 Nm 230 lbf ft
Feb 04 25 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
To get the best pumping speed, ensure that 3.10.2 Connect the pump-outlet to your
the pipeline which connects the process exhaust system
PAGE
26
system to the pump is as short as possible and
has an internal diameter not less than the WARNING
pump-inlet.
Pipe the exhaust to a suitable
INSTALLATION
Use a flexible connection in the pipeline from treatment plant to prevent the
the process system to the pump to reduce discharge of dangerous gases or
vibration and stress in the system pipelines. vapours to the surrounding
atmosphere.
Use the following procedure to connect the pump-
inlet to your process system. Do not allow debris WARNING
to get into the pump during installation. Refer to
Figure 1. Incorporate safety devices to prevent
operation of the pump when the
exhaust pipeline is restricted or
1. The pump-inlet has an inlet filter:
blocked. If you do not, the exhaust
If you install the pump in a new process pipeline may become over-
system, leave the filter (4) in the pump-inlet pressurised and may burst.
(3) to prevent the entry of weld particles or
other debris into the pump. You will remove WARNING
the filter in Section 5.5.
Do not exceed the load limits on the
If you install the pump in an existing process pump-outlet flange as specified in
system which you are sure does not contain Figure 7. If you do, there will be a risk
any debris which could damage the pump, of leakage of process gases from the
pump, or of damage to the pump.
remove the filter (4) from the pump-inlet (3).
2. Use four suitable bolts to connect the pump-
inlet to your process system. Use a suitable CAUTION
PTFE gasket to seal the connection. We
Install an outlet catchpot to prevent the drainage of
recommend that you use the gasket type condensate back into the pump. If you do not,
specified in Section 2.11. condensate which drains back into the pump may
damage it or cause it to seize.
Issue L 26 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
Incorporate flexible bellows in the exhaust pipeline If you want to change the operating temperature of
to reduce the transmission of vibration and to the pump (see Section 3.12.3) or if you want to
PAGE
prevent loading of coupling-joints. If you use flexible change the temperature at which the thermal snap- 27
bellows, you must ensure that you use bellows switch opens, use the following procedure.
which have a maximum pressure rating which is
greater than the highest pressure that can be 1. Refer to Figure 8. Remove the four bolts (4)
INSTALLATION
generated in your system. which secure the cover (3) on the thermal snap-
switch, then remove the cover.
3.11 Leak test the installation 2. Turn over the cover (3) and remove the metal
dowel (7) from its storage position in the back
WARNING
of the cover.
Leak-test the system after installation 3. Push one end of the dowel into one of the holes
and seal any leaks found to prevent on the adjuster (6), then turn the adjuster until
the leakage of dangerous substances
the required operating temperature (shown on
out of the system and leakage of air
into the system. the adjuster scale) is indicated by the arrow (5).
Feb 04 27 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
extraction system.
10. Turn off the pump and the cooling-water and
3. Check that there are no leaks in the water,
shaft-seals purge nitrogen supplies.
nitrogen system, and exhaust-extraction system
connections. Seal any leaks found.
3.12.3 Adjust the TCV (thermostatic
4. Refer to Figure 6. Turn the adjuster knob on control-valve)
the TCV (1) to the '0' position. Check that the
ball in the cooling-water flow indicator (11) Notes: As supplied, the TCV is adjusted for a pump
moves; this indicates that there is a flow of operating temperature of 149 oF (measured at
cooling-water through the cooling-water flow the point shown in Figure 1, item 27). If you
indicator. adjust the TCV for a different pump operating
temperature, you must also adjust the thermal
5. Adjust the TCV to the required pump operating
snap-switch (refer to Section 3.12.1).
temperature; refer to Table 13 for the TCV
settings for a pump operating temperature of The minimum pump operating temperatures
149 oF. For a different pump operating stated in Table 13 are with a cooling-water
temperature, refer to Section 3.12.3. flow of 32 US gal h-1 and with a cooling-water
6. Switch on the pump. supply temperature of 68 oF.
7. Check that the pressure shown on the shaft- The TCV regulates the flow of water through the
seals purge nitrogen pressure gauge (8) is water cooling system to maintain the pump at the
between 5 and 7 psig, above the exhaust back required operating temperature and is adjustable
pressure. If you need to adjust the pressure: between 0 (minimum) and 5 (maximum). These
Loosen the locknut (14) on the rear of the TCV settings correspond to the minimum and
nitrogen pressure regulator (12). maximum pump operating temperatures shown in
Table 6.
Turn the adjuster (13) until the required
pressure is shown on the pressure gauge. If you need to adjust the TCV to suit your operating
Tighten the locknut (14). conditions, turn the adjuster knob on the TCV
(Figure 6, item 1) clockwise or anticlockwise to the
8. Leave the pump to operate for approximately
position corresponding to your required pump
60 minutes to allow the pump operating
operating temperature. Note that it takes
temperature to stabilise.
approximately 60 minutes for the pump to stabilise
at its final operating temperature.
Issue L 28 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
29
INSTALLATION
1. Thermal snap-switch
2. Cable gland (not supplied): to fit 1/2 inch NPT electrical entry
3. Cover
4. Bolt
5. Arrow
6. Adjuster
7. Metal dowel
Feb 04 29 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
30
Issue L 30 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
CAUTION PAGE
4.1 Operational safety 31
OPERATION
condensable vapours. If you do not, the vapours
Obey the safety instructions given may condense in the pump and corrode or damage
below. If you do not, the installation the pump.
may not be safe, and there may be a
risk of injury or death to people.
Use the following procedure to start the pump:
During operation, parts of the pump can 1. Check the gearbox oil-level in the sight-glass on
become very hot. Under abnormal the side of the pump: refer to Section 3.5.
conditions, the temperature of the pump- 2. Turn on your cooling-water supply, shaft-seals
body can reach 300 oF. Ensure that you do purge nitrogen supply and exhaust-extraction
not touch the pump. system (if fitted).
Because of the high gas temperatures in the 3. Check that the pressure of the shaft-seals purge
pump, we recommend that you only pump nitrogen supply is correct and adjust if
gases in temperature classes T1 to T3 in necessary: refer to Section 3.12.2.
normal operating conditions. Do not pump
4. Switch on the pump.
gases in temperature class T4 unless you have
contacted your supplier or BOC Edwards for 4.3 Allow the pump to warm up
advice.
Do not operate the pump with a coupling WARNING
cover guard removed. If you do, there will be
During pump warm-up, coolant may
a danger of injury or death from the rotating drain out of the coolant overflow
mechanisms. pipe. Take all necessary precautions if
Do not operate the pump with the pump- this m,ay present a hazard.
inlet or -outlet open to atmosphere. If you do,
there will be a danger of injury or death from CAUTION
the rotating mechanisms, from the exposure
Allow the pump to warm up and use full gas inlet
to vacuum, or from hot exhaust gases. purge and gas ballast (if fitted) before you pump
If a pump with a torque limiter is located in a condensable vapours. If you do not, the vapours
may condense in the pump and corrode or damage
Division 1 hazardous area, do not let the
the pump.
pump-motor operate for more than 8 hours if
the torque limiter has tripped and decoupled
Leave the pump to operate (with the inlet isolation-
the pump-motor from the pump. If you do,
valve closed) to allow the pump to warm up to its
the torque limiter may get very hot and there operating temperature, that is, until there is a flow
may be a risk of fire or explosion. of cooling-water through the heat exchanger:
Note: The procedures in the following sections Without the inlet purge, warm-up takes
assume that you have a pump-inlet isolation- approximately 30 to 45 minutes, depending
valve fitted to your pump. on the ambient temperature.
With an inlet purge pressure of 0.7 to 3.6 psig,
warm-up time can be reduced to as little as 10
minutes.
(Continued on page 32)
Feb 04 31 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
Refer to Figure 6. When the pump has warmed up 4.5 Inspect and reset the torque limiter
to its operating temperature (indicated by a flow of (pumps with a torque limiter only)
PAGE
32
water in the cooling-water flow indicator, 11), you
may start process pumping. WARNING
In the first 30 seconds of pumpdown, open the inlet Switch off the pump and isolate the
OPERATION
isolation-valve slowly if there is liquid in the inlet pump from the electrical supply
before you reset the torque limiter.
pipeline, in order to minimize the ingress of liquid
Refit the coupling cover guard before
(entrained in the process gas stream) into the pump. you switch on the pump again. If you
do not, there will be a danger of
4.4 Shut down the pump injury or death from the rotating
coupling mechanism.
CAUTION
Note: If the torque limiter continually trips and
Purge the pump before you shut it down. If you do
decouples the pump-motor from the pump,
not, process vapours may condense in the pump
and corrode or damage it. refer to Section 5.19.
1. Isolate the pump-inlet from the process gases. 1. Switch off the pump and isolate the pump from
2. Purge the pump of contaminants. Use one of the electrical supply.
the following methods: 2. Undo and remove the four M8 bolts which
Operate the pump at or close to atmospheric secure one of the coupling cover guards
pressure for at least 15 minutes. (Figure 1, item 14), then remove the guard.
Operate the pump with full inlet purge (if 3. Refer to Figure 9. Check the relative positions
fitted) for at least 15 minutes. of the notches in the cage plate (2) to the
openings in the slide plate (3):
Operate the pump with full gas ballast (if
fitted) for at least 15 minutes. If the notches in the cage plate are offset to
the left of the openings in the slide plate, the
3. Switch off the pump.
torque limiter has tripped in the normal
4. When the pump has cooled down, turn off the direction: continue at Step 4.
cooling-water supply and the shaft-seals purge
If the notches in the cage plate are offset to
nitrogen supply.
the right of the openings in the slide plate, the
torque limiter has tripped in the reverse
direction: do not continue with this
procedure, but contact your supplier or BOC
Edwards for advice.
Issue L 32 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
4. Inspect the torque limiter. You must not 4.5.2 Reset the torque limiter
continue with this procedure, but must contact
PAGE
your supplier or BOC Edwards (to arrange for CAUTION 33
service/replacement of the torque limiter) if:
Only use the special tools supplied to reset the
The rim of the slide plate (3) is cracked or torque limiter. If you do not, you may damage the
OPERATION
broken. pump.
If the torque limiter does not have to be 1. Refer to Figure 10. Turn the drive plate (2) so
serviced/replaced, continue at Step 5. that a notch (1) in the plate aligns with a notch
on the cage plate (3).
5. Refer to detail B. If the pump-motor has been
operated for more than 8 hours with the 2. Place the steel rod (7) in one of the holes in the
torque limiter tripped, check the gap (5) thrust plate (6), then place the torque limiter
between the lower face of the drive plate (1) reset tool (5) in the notches (1) in the drive
and the rim of the slide plate (3): plate (2) and cage plate (3).
If the gap (5) is 0.08 inches (2 mm) or more: 3. Turn the torque limiter reset tool to the right
continue at Section 4.4.2 to reset the torque while you turn the steel rod to the left, until the
limiter. torque limiter recouples. You will hear a click
when the torque limiter resets and the pump-
If the gap (5) is less than 0.08 inches (2 mm):
motor recouples to the pump.
do not continue with this procedure, but
contact your supplier or BOC Edwards (to 4. Refer to Figure 9, detail A. Check that the
arrange for service/replacement of the torque torque limiter has reset correctly:
limiter). If the notches in the cage plate (2) are centred
in the openings in the slide plate (3), continue
at Step 5.
If the notches in the cage plate (2) are not
centred in the openings in the slide plate (3),
do not continue with this procedure, but
contact your supplier or BOC Edwards to
arrange for service/replacement of the torque
limiter.
Feb 04 33 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
34
OPERATION
Issue L 34 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
35
OPERATION
Feb 04 35 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
Issue L 36 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
MAINTENANCE
Obey the safety instructions given misused, if it malfunctioned, or if it was in a
below and take note of appropriate fire. BOC Edwards Material Safety Data
precautions. If you do not, you can Sheets for the fluorinated materials used in
cause injury to people and damage to
the pump are available on request: contact
equipment.
your supplier or BOC Edwards.
Ensure that the ambient atmosphere around Leak-test the system after maintenance work
the pump is not flammable before you start is complete and seal any leaks found to
maintenance. prevent leakage of dangerous substances out
of the system and leakage of air into the
A suitably trained and supervised technician
system: refer to Section 3.11.
must maintain the pump.
Ensure that the maintenance technician is 5.2 Maintenance plan
familiar with the safety procedures which
relate to the synthetic oils and greases used The plan in Table 14 details the maintenance
and the products pumped. Wear the operations required to maintain the pump in normal
appropriate safety-clothing when you come operation. Instructions for each operation are given
into contact with contaminated components, in the section shown. In practice, the frequency of
grease and pump oil. Dismantle and clean maintenance is dependent on your process. In clean
contaminated components inside a fume- processes, you may be able to decrease the
cupboard. frequency of maintenance operations; in harsh
processes you may have to increase the frequency of
Use suitable lifting equipment and wear safety
maintenance operations. Adjust the maintenance
shoes when you replace the pump-motor.
plan according to your experience.
Allow the pump to cool for at least three
hours before you start maintenance work. When you maintain the pump, use BOC Edwards
maintenance and service kits. These contain all of
Isolate the pump and other components in
the necessary seals, lubricating grease and other
the process system from the electrical supply components necessary to complete maintenance
so that they cannot be operated accidentally.
operations successfully. The Item Numbers of these
Recheck the pump rotation direction if the kits are given in Section 7.3.
electrical supply has been disconnected.
Do not reuse 'O' rings or gaskets if they are
damaged.
Feb 04 37 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
Refer to
Operation Frequency
PAGE Section
38
Check the gearbox oil level and fill the gearbox with oil (if Weekly 5.3
necessary)
MAINTENANCE
Inspect the pressure relief valve and replace the hinge Yearly or when necessary 5.7
bushes, flap and 'O' rings (if necessary) *
Change the pump oil and clean the oil-level sight-glasses # Yearly or when contaminated, 5.8
whichever occurs first
Replace the pump-motor and drive coupling bushes $ When necessary 5.16
Issue L 38 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
5.3 Check the gearbox oil-level and fill 5.4 Check the shaft-seals purge supply
the gearbox with oil (if necessary)
PAGE
WARNING 39
WARNING
Your nitrogen supply pressure must
The gearbox may be pressurised. To comply with the requirements of
MAINTENANCE
remove the oil filler-plug, undo it Section 2.5. If it does not, the shaft-
slowly until the vent hole is visible, seals purge pipelines may become
allow the gearbox to vent to over-pressurised and may explode.
atmosphere to reduce the pressure
in the gearbox, then fully undo and 1. Inspect the shaft-seals purge gas supply pipelines
remove the oil filler-plug. If you do
and connections; check that they are not
not, hot (up to 194 oF) oil may be
ejected under pressure from the corroded or damaged. Replace any pipelines
gearbox and cause injury. and connections that are corroded or damaged.
Feb 04 39 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
40
MAINTENANCE
Issue L 40 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
5.6 Inspect the pipelines and connections 3. Remove the valve flap (7) and the valve body (2)
assembly from the exhaust manifold (3).
1. Inspect all cooling-water pipelines and PAGE
41
connections; check that they are not corroded 5.7.2 Clean and inspect the valve
or damaged. Replace any of the pipelines and
connections that are corroded or damaged. 1. Refer to Figure 12. Inspect the valve:
MAINTENANCE
Check that all cooling-water connections are
If the retainer (9) or valve body (2) are
secure. Tighten any connections that are loose.
corroded or damaged, you must replace the
2. Inspect all nitrogen supply pipelines and pressure relief valve: refer to Section 5.14.
connections; check that they are not corroded
If any of the valve flap (7), the hinge bushes (1)
or damaged. Replace any pipelines and
or the valve 'O' ring (6) are damaged, replace
connections that are corroded or damaged.
the damaged component with a new
Check that all nitrogen supply connections are
component.
secure. Tighten any connections that are loose.
2. Clean the inside of the exhaust manifold (3) to
3. Inspect all electrical cables; check that they are
remove any deposits; if necessary, use a suitable
not damaged and have not overheated. Replace
cleaning solution. If you use a cleaning solution,
any cables that are damaged or have
ensure that all of the solution is removed before
overheated. Check that all electrical
you fit the new pressure relief valve
connections are secure. Tighten any
components.
connections that are loose.
3. Inspect the 'O' rings (8, 5, 4) and, replace them
4. Inspect all process and exhaust pipelines; check
if necessary.
that they are not corroded or damaged.
Replace any pipelines that are corroded or 5.7.3 Refit the valve to the pump
damaged. Check that all process and exhaust
connections are secure. Tighten any 1. Refer to Figure 12. If necessary, refit the two
connections that are loose. hinge bushes (1) to the valve flap (7), then fit the
valve flap to the valve body (2).
5.7 Inspect the pressure relief valve and
replace the hinge bushes, flap and 2. Refit the valve body (2) in the exhaust manifold
'O' ring (if necessary): EDP200 and (3).
EDP300 only 3. Fit the retainer (9) to the exhaust manifold (3)
and secure with the four bolts (11). Tighten the
If you think that the pressure relief valve does not bolts to a torque of 7.4 lbf ft (10 N m).
operate correctly, use the procedures in the
following sections to remove, inspect and refit the
valve.
Feb 04 41 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
42
MAINTENANCE
Issue L 42 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
5.8 Change the pump oil and clean the 1. Refer to Figure 11. Remove the oil filler-plug (3)
oil-level sight-glasses and bonded seal (2): see Note above.
PAGE
2. Place a suitable container under the oil drain- 43
WARNING
port (6); the container must have a capacity of
Obey the safety instructions given at least 1.2 US gal.
MAINTENANCE
below and take note of appropriate 3. Unscrew and remove the oil drain-plug (4) and
precautions. If you do not, you can
bonded seal (5) and allow the oil to drain from
cause injury to people and damage to
equipment. the gearbox. Dispose of the bonded seal (5).
Feb 04 43 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
14. Fill the gearbox through the funnel or nozzle, 5.9.1 Introduction
with the correct grade and quantity of oil.
PAGE
Allow the oil to drain into the gearbox and then The frequency of relubrication of the rotor bearings
44
check the level on the oil sight-glass (refer to depends on the operating temperature of the pump.
Section 5.3). Table 15 shows the recommended frequencies.
MAINTENANCE
15. Remove the funnel or nozzle from the oil filler- For maximum pump reliability, we recommend that
port (1) and refit the oil filler-plug (3) and you change the gearbox oil when you relubricate the
bonded seal (2) to the oil filler-port (1). rotor bearings: refer to Section 5.3.
16. Dispose of the used oil safely: refer to Section
6.2. Use the procedures in the following sections to
relubricate the rotor bearings.
5.9 Relubricate/inspect the rotor
bearings Pump
operating Relubrication frequency
Note: If you use the pump on a harsh application, we temperature
recommend that a BOC Edwards service
up to 158 oF Every 8800 hours (1 year)
engineer removes and cleans the bearing
assembly before it is relubricated: contact your 158 to 185 oF Every 4400 hours (6 months)
supplier or BOC Edwards to arrange this.
185 to 194 oF Every 2200 hours (3 months)
1. Gearbox
2. Compression ring
3. 'O' ring
4. Sight-glass
5. 'O' ring
6. Bezel
7. Screw (4 off)
Issue L 44 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
5.9.2 Remove the bearing cover and 2. Inspect the bearings for obvious signs of wear
bearing caps or the presence of debris. If the bearings are
PAGE
worn, contact your supplier or BOC Edwards 45
WARNING for advice.
Allow the pump to cool down and 3. If the bearings are in a staisfactory condition,
MAINTENANCE
release the pressure in the cooling use a suitable syringe to force approximately
jacket before you remove the bearing 1.8 oz of new Fomblin CR861 grease into the
cover. If you do not, hot coolant may bearings so that a smooth layer of grease covers
be ejected from the cooling jacket the case and bearings. Do not put too much
and may cause injury.
grease in the bearings or the pump will run hot.
1. Refer to Figure 3. Wear thermal protective 4. Use a 24 mm A/F spanner to turn the drive
gloves and push down the coolant filler cap (1) rotor shaft (9) in the correct direction (13)
and turn it anticlockwise by a 1/4 of a turn (as three or four complete revolutions.
shown in detail A) to release the pressure in the 5. Press any grease forced out of the bearings back
cooling jacket. into the bearings.
2. Refer to Figure 14. Remove the ten M10 bolts
5.9.4 Refit the bearing caps and bearing
and spring-washers (15) which secure the
cover
bearing cover (5) to the high vacuum head-
plate.
1. Refer to Figure 14. Refit the bearing cap (7)
3. Remove the M6 screws and nuts which secure with a new bearing cap 'O' ring (8) on the drive
the thermal snap-switch to the support bracket. rotor shaft. Apply a suitable thread sealant (for
example, Loctite Screwlok 242 or equivalent)
4. Remove the bearing cover (5), the bearing
to each of the three bearing cap retaining bolts
cover 'O' ring (14) and the cooling-jacket 'O'
(6) and refit the bolts to secure the bearing cap.
ring (11). Discard the 'O' rings.
2. Repeat Step 1 to refit the bearing cap (12) on
5. Remove the three M8 bearing cap retaining
the driven rotor shaft.
bolts (6) which secure the bearing cap (7) on
the drive rotor shaft; ensure that you do not 3. Apply a light wipe of vacuum grease to the new
accidentally remove one or more of the bearing bearing cover 'O' ring (14) and the water jacket
carrier bolts (16, see detail A), which are next 'O' ring (11), then place them in position on the
to the bearing cap retaining bolts. Remove the high-vacuum head-plate.
bearing cap (7) and the bearing cap 'O' ring (8).
4. Refit the bearing cover (5) to the high vacuum
6. Repeat Step 5 to remove the bearing cap (12) head-plate and secure with the ten bolts and
and bearing cap 'O' ring on the driven rotor spring washers (15).
shaft.
5. Secure the snap-switch to the support bracket
with the two M6 screws (4) and nuts (1).
5.9.3 Clean, inspect and relubricate the
rotor bearings 6. Refer to Figure 3. Turn the coolant filler cap (1)
clockwise to tighten it.
1. Refer to Figure 14. Use a clean lint-free cloth to
7. Dispose of the old grease safely: refer to
remove as much old grease as possible from the
Section 6.2.
inside face of the bearing caps (7, 12) and from
the top of the bearings. Do not use your fingers
for this operation.
Feb 04 45 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
5.10 Check the coolant level and refill if 5.10.2 Fill the pump with a different coolant
necessary
PAGE Note: You will require a Routine Maintenance Kit
46
WARNING (see Section 7.3) to drain and refill the pump
with a different coolant.
Allow the pump to cool down and
MAINTENANCE
release the pressure in the cooling If you will refill the pump with a different coolant,
jacket before you remove the refer to Figure 3 and use the following procedure.
coolant filler-cap. If you do not, hot
The coolant you use must comply with the
coolant may be ejected from the
cooling jacket and may cause injury. requirements of Section 2.4.
Issue L 46 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
47
MAINTENANCE
Feb 04 47 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
5.10.3 Check for coolant leaks 5.11.2 Flush a pump which has been
operating
PAGE Check the pump for obvious signs of a coolant leak
48
(for example, a small stream of coolant at one of the WARNING
joints on the pump-body).
Use a cleaning solution which is
MAINTENANCE
If there are signs of a coolant leak, contact your suitable for the process materials
supplier or BOC Edwards for advice. pumped. If you do not, there may be
a risk of explosion or pump damage.
5.11 Flush the pump with cleaning
solution WARNING
Issue L 48 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
5. Undo and remove the blanking plug from the 5.11.3 Flush a seized pump
middle stage purge port (5). Use a suitable tool
PAGE
to inject the cleaning solution into the port until WARNING 49
the cleaning solution comes out of the port,
then refit the plug to the port. Use a cleaning solution which is
suitable for the process materials
MAINTENANCE
6. Remove the plug from the inlet flush port (2), pumped. If you do not, there may be
use a suitable tool to inject the cleaning solution a risk of explosion or pump damage.
into the port until the cleaning solution comes
out of the port, then refit the plug. WARNING
7. Refer to Figure 10. On pumps with a torque Ensure that you correctly refit the
limiter, fit the steel rod (7) supplied into the blanking plugs to the purge ports on
hole (4) in the coupling adaptor, on other the pump. Ensure that the threads in
pumps, fit a suitable steel rod (not supplied) into the purge ports are clean and free
the hole in the drive hub (Figure 19, item 5), from damage, and that the plug is
undamaged. Replace damaged plugs.
then turn the pump clockwise through three or
four complete revolutions.
Note: If the pump has a torque limiter and it has
8. Repeat Steps 4 to 7 above as necessary, until decoupled the pump-motor from the pump, do
the pump rotates freely. not recouple the pump-motor to the pump
9. If you have a pump-inlet isolation-valve, open until you have finished this procedure. If you
the valve, otherwise, remove the blanking- leave the pump-motor decoupled from the
flange from the pump-inlet and reconnect the pump, it will be easier to turn the pump by
pump-inlet to your process system. hand.
10. Refer to Figure 1. If the pump has a torque 1. Refer to Figure 1. Remove the four bolts which
limiter and it has tripped and decoupled the secure one of the coupling cover guards (14)
pump-motor, reset the torque limiter: refer to and remove the guard.
Section 4.5.
2. If you have a pump-inlet isolation-valve in your
11. Use the four bolts to refit the coupling cover system, close the valve. If you do not have a
guard (14) to the pump. Tighten the bolts to a pump-inlet isolation-valve, disconnect the
torque between 2.2 and 3.6 lbf ft (3 and 5 Nm). pump-inlet (3) from your process system and fit
12. Refit the pump-outlet (25) to your exhaust a blanking-flange over the pump-inlet.
pipeline, then dispose of the deposits removed 3. Refer to Figure 15. Remove the plug from the
from the pump. inlet flush port (2).
4. Disconnect the pump-outlet (3) from your
exhaust pipeline and fit a blanking-cap to the
pump-outlet.
Feb 04 49 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
6. Repeat Step 5 until you can pour no more 5.12 Clean the cooling system
cleaning solution into the pump (see Table 3 for
PAGE
the capacity of the pump), then leave the pump 5.12.1 Drain the coolant from the cooling
50
for at least one hour for the cleaning solution to system
fully react with the deposits in the pump.
WARNING
MAINTENANCE
Remove one or more of the purge port 3. Turn the coolant filler-cap (1) anticlockwise and
blanking plugs (Figure 15, items 4 and 5) to remove it from the pump.
allow the cleaning solution to drain out of the 4. Refer to Figure 6. Place a suitable container
purge port(s), then refit the plug(s) to the under the cooling-water connections (3, 5) on
purge port(s). the services panel, then remove the connectors
10. If necessary, repeat Steps 5 to 9 to flush the (4, 6) on your cooling-water supply and return
pump again. pipelines from the connections on the services
panel and allow the cooling-water to drain from
11. Remove the blanking-flange from the pump- the pump.
inlet (Figure 1, item 3) and reconnect the pump-
inlet to your process system (refer to 5. Drain the coolant from the pump: refer to Steps
Section 3.10.1) or open the pump-inlet 3 and 4 of Section 5.10.2.
isolation-valve. 6. Refer to Figure 18. Place a suitable container
12. Reconnect the pump-outlet to your exhaust under the drain-plug (7), then remove the
pipeline: refer to Section 3.10.2. drain-plug (7) and allow the cooling-water to
drain from the heat exchanger.
13. If necessary, recouple the pump-motor to the
pump: refer to Section 4.5.
Issue L 50 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
51
MAINTENANCE
1. Pump inlet-flange
2. Inlet flush port
3. Pump outlet-flange
4. Gas ballast purge port
5. Middle stage purge port
Feb 04 51 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
Issue L 52 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
5.12.3 Clean the cooling-water flow 4. Clean the ball (6), the inside of the dome (4) and
indicator the inside of the body (7) to remove any
PAGE
deposits; if necessary, use a suitable cleaning 53
The location of the cooling-water flow indicator is solution.
shown in Figure 6, item 11.
5. Use the new 'O' ring and refit the ball (6), 'O'
MAINTENANCE
1. Refer to Figure 17. Remove the four bolts (1). ring (5), dome (4), gasket (3) and bezel (2) to
the body (7) of the cooling-water flow
2. Remove the bezel (2), gasket (3), dome (4), 'O' indicator.
ring (5) and ball (6) from the body of the
cooling-water flow indicator. 6. Use the four bolts (1) to secure the bezel to the
body (7) of the cooling-water flow indicator.
3. Dispose of the 'O' ring and the gasket (3).
Feb 04 53 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE Refer to Figure 18 and use the following procedure We recommend that the pump is given a major
54
to clean the heat exchanger. overhaul every three years.
1. Undo the fitting nuts (2, 9) of the cooling-water Such an overhaul is outside the scope of this manual
MAINTENANCE
connectors to disconnect the cooling-water and should be done by qualified BOC Edwards
pipes from the top and bottom end-caps of the service personnel: contact your supplier or BOC
heat exchanger. Edwards.
3. Remove the four bolts (8) which secure the You must replace the pressure relief valve if it is
bottom end-cap (6) to the heat exchanger and damaged. The Pressure Relief Valve is available as a
remove the end-cap and the 'O' ring (5). spare: refer to Section 7.3.
4. Use a pipe cleaner which will pass through a 16 1. Refer to Figure 12. Undo and remove the four
inch (400 mm) length of 0.2 inch (5 mm) M8 bolts (11) which secure the valve retainer
diameter tube to clean and remove any scale (9) to the exhaust manifold (3).
from each tube in the tube stack (3).
2. Place two of the bolts in the jacking holes (10)
5. Clean the 'O' ring sealing faces of the top and and tighten the bolts to remove the retainer (9)
bottom end-caps (11, 6) and the top and from the exhaust manifold.
bottom of the heat exchanger.
3. Remove the 'O' ring (8), valve flap (7), hinge
6. Fit the new 'O' ring (5) and then use the four bushes (1), valve body (2) and 'O' rings (5, 4)
bolts (8) to refit the bottom end-cap (6) to the from the exhaust maniifold (3).
heat exchanger.
4. Clean the inside of the exhaust manifold to
7. Fit the new 'O' ring (10) and then use the four remove any deposits; if necessary, use a suitable
bolts (1) to refit the top end-cap (11) to the cleaning solution. If you use a cleaning solution,
heat exchanger. ensure that all of the solution is removed before
8. Reconnect the fitting nuts (2, 9) of the cooling- you fit the new pressure relief valve.
water connectors to the top and bottom end- 5. Fit the new pressure relief valve to the exhaust
caps (11, 6). manifold and secure with the four M8 bolts (11).
Tighten the bolts to a torque of 7.4 lbf ft
5.12.5 Prepare the pump for operation
(10 Nm).
Issue L 54 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
55
MAINTENANCE
Feb 04 55 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
5.15 Inspect the torque limiter 4. Refer to Figure 1. Use the four bolts removed
in Step 1 to refit the coupling cover guard (14).
PAGE 1. Refer to Figure 1. Undo and remove the four Tighten the bolts to a torque between 2.2 and
56
M8 bolts which secure one of the coupling 3.6 lbf ft (3 and 5 N m).
cover guards (14), then remove the guard.
5.16 Replace the pump-motor and the
MAINTENANCE
The edges of the cage plate (2) are burred. Note: You must ensure that the pump-motor, motor
coupling and coupling housing assembly is
Any of the notches (Figure 10, item 1) in the
adequately supported in the following
drive plate are damaged.
procedures. The mass of the motor coupling
Any of the components show signs of and coupling housing is 68 lb on all pumps. The
discolouration (due to overheating). total mass of the pump-motor, motor coupling
and coupling housing assembly on your pump
If the torque limiter does not have to be is therefore (68 + M) lb, where M is the mass
serviced/replaced, continue at Step 3. of your pump-motor (in lb). Refer to the motor
manual or to the data plate on the pump-
3. Refer to detail B. Check the gap (4) between motor for the the pump-motor mass.
the lower face of the drive plate (1) and the rim
of the slide plate (3): 5.16.1 Lower the pump-motor and coupling
If the gap (4) is 0.004 to 0.016 inches (0.1 to housing (without the motor change frame)
0.4 mm), the torque limiter is correctly
WARNING
engaged: continue at Step 4.
If the gap (4) is > 0.016 inches (4 mm), the Ensure that the pump-motor, motor
coupling and coupling housing
torque limiter is not engaged correctly: re-
assembly does not fall over when you
engage the torque limiter (refer to move it. If it falls over, it can cause
Section 4.5), then recheck the gap. If the gap injury to people. Refer to the Note in
is still incorrect after you have re-engaged the Section 5.16 to calculate the mass of
torque limiter, do not continue with this the assembly.
procedure but contact your supplier or BOC
Edwards to arrange for service/replacement 1. Refer to Figure 19 or 20. Remove the fixing
of the torque limiter. bolts (12) which secure the coupling housing to
the gearbox and lower the pump-motor and
If the gap (4) is < 0.004 inches (0.1 mm): do coupling housing assembly until the pump-
not continue with this procedure, but contact
motor rests on the two cross-members at the
your supplier or BOC Edwards to arrange for
bottom of the pump frame (4). Ensure that the
service/replacement of the torque limiter. pump-motor, motor coupling and coupling
housing assembly is adequately supported
throughout and does not fall over.
2. Continue at Step 13 of Section 5.16.2.
Issue L 56 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
57
Figure 19 - Exploded view of the pump-motor, torque limiter and coupling housing
(pumps with a torque limiter)
Feb 04 57 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
58
MAINTENANCE
Figure 20 - Exploded view of the pump-motor, coupling drive and coupling housing
(pumps without a torque limiter)
Issue L 58 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
5.16.2 Lower the pump-motor and coupling 6. Position the motor change frame (6) so that the
housing (with the motor change two locating pins (12, detail B) are upwards.
PAGE
frame) 7. From the oil filler elbow side of the pump, slide 59
the motor change frame between the top and
WARNING
bottom flanges of the coupling housing, as
MAINTENANCE
Ensure that the pump-motor, motor shown in detail B.
coupling and coupling housing
8. Refer to detail B. Move the motor change frame
assembly does not fall over when you
move it. If it falls over, it can cause (6) up over the studs until the two locating pins
injury to people. Refer to the Note in (12) fit into the holes (11) from which the two
Section 5.16 to calculate the mass of fixing bolts were removed.
the assembly.
9. Hold the motor change frame (6) in place, then
fit a washer and nut (5) to the bottom of each
1. Refer to Figure 21. Undo and remove the three
stud (3) and spin the nuts upwards until the
nuts and washers from the tops of the studs (1)
motor change frame is held tightly in place
and remove the studs and the motor change
against the top flange of the coupling housing.
frame (7) from the pump frame (4).
10. Remove the remaining two coupling housing
2. Refer to detail A. Remove the two M16 fixing
M16 fixing bolts (10). The mass of the pump-
bolts (8) which secure the coupling housing to
motor and coupling housing assembly is now
the gearbox; one bolt is located next to the oil
supported by the motor change frame (6).
drain port and the other bolt is on the opposite
side of the pump. Do not remove the other two 11. Keep the top of the pump-motor level
bolts (10), as these secure the pump-motor and throughout and undo the three lower nuts (5)
coupling housing assembly to the pump. on the studs to lower the pump-motor and
coupling housing assembly until the pump-
3. Refer to Figure 19 or 20. Loosen the four M16
motor rests on the two cross-members at the
bolts (9) which secure the coupling housing (12)
to the pump-motor (8). Do not over-loosen or bottom of the pump frame (4). While you undo
the three nuts, continually check that the top
remove the bolts, as they secure the pump-
nuts are tight and that the studs are firmly
motor in place.
secured to the pump frame.
4. Refer to Figure 21. Pass the three studs (3)
through the holes in the pump frame (4); there 12. Ensure that the pump-motor rests securely on
are three holes in the pump frame, as follows: the pump frame and remove the motor change
frame (6), then remove the three studs (3) from
Two on the main cross-member, either side
the pump frame.
of the exhaust manifold.
13. Carefully slide the pump-motor and coupling
One on the main cross-member on the
housing assembly out of the pump frame, then
opposite side of the pump, near the oil filler
use suitable lifting equipment to lower the
(as shown in Figure 21).
pump-motor and coupling housing assembly so
5. Secure the tops of the three studs to the cross- that it rests on the floor in an upright
members of the pump frame (4). Use a nut and orientation (that is, with the top flange of the
washer above and below the cross members coupling housing at the top).
and ensure that the nuts are tight, so that the
studs cannot turn and become detached from
the cross-members. The studs should now hang
down from the pump frame.
Feb 04 59 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
5.16.3 Remove the coupling housing from 5.16.5 Remove the drive hub and fit to the
the pump-motor new pump-motor (pumps without a
PAGE
60
torque limiter)
1. Refer to Figure 19 or 20. Remove the four bolts
(9) which secure the pump-motor (8) to the 1. Refer to Figure 20. Undo and remove the grub-
coupling housing (13). screw (11) and the grub-screw (6) on the drive
MAINTENANCE
hub (5).
2. Take note of the orientation of the coupling
housing on the pump-motor (so that you can 2. Use a suitable puller tool to remove the drive
refit it later in the same orientation), then lift hub (5) from the motor shaft (7).
the coupling housing from the pump-motor.
3. Remove the key (10) from the shaft and dispose
of the key.
5.16.4 Replace the drive coupling bushes
4. Dispose of the key supplied with the new pump-
1. Refer to Figure 19 or 20. Undo and remove the motor.
two grub-screws (15) in the coupling hub (16).
5. Inspect the motor shaft (7) of the new pump-
2. Slide the coupling hub (16) off the pump shaft motor (8). The motor shaft must be free of
(17) and remove the key (14) from the pump burrs and dirt. If necessary, clean or refinish the
shaft. Dispose of the key. motor shaft.
3. Refer to detail A. Pull the drive coupling bushes 6. Inspect the motor shaft bore of the drive hub
(18) out of the coupling hub (16). Dispose of (5). The bore must be free of burrs and dirt. If
the bushes. necessary, clean or refinish the bore.
4. Fit the new drive coupling bushes (18) into the 7. Fit the new key (10, supplied in the Motor
coupling hub (16). Fitment Kit) into the motor shaft (7).
5. Apply a suitable thread sealant (such as Loctite 8. Fit the drive hub (5) onto the motor shaft (7)
242 Nutlock) to the new grub-screws (15) and and push it down until it will go no further.
screw the grub-screws into the coupling hub Check that the lower face of the drive hub is 1
(16). inch above the motor flange (refer to detail C);
6. Fit the new key (14) into the pump shaft (17). if the gap is not correct, contact your supplier
or BOC Edwards for advice.
7. Refer to detail B. Slide the coupling hub (16)
onto the pump shaft (17) and position it so that 9. Apply a suitable thread sealant (such as Loctite
the bottom face of the coupling hub (16) is 0.47 242 Nutlock) to the new grub-screws (11 and
inches (12 mm) above the end of the pump 6), then screw the grub-screws into the correct
shaft. screw holes in the drive hub (5). Fully tighten
the grub-screws to secure the drive hub (5) to
8. Fully tighten the two grub-screws (15) to the motor shaft (7).
secure the coupling hub (16) and key (14) in
place.
Issue L 60 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
61
MAINTENANCE
1. Stud (on pump frame) 7. Motor change frame (on pump frame)
2. Nut and washer 8. Bolt
3. Stud (removed from pump frame) 9. Pump-motor terminal-box
4. Baseframe 10. Bolt
5. Nut and washer 11. Bolt hole
6. Motor change frame (removed from pump frame) 12. Locating pin
Figure 21 - Use the motor change frame to lower the pump-motor and coupling housing assembly
Feb 04 61 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
5.16.6 Remove the torque limiter and fit to 5.16.7 Fit the coupling housing to the new
the new pump-motor (pumps with a pump-motor
PAGE
62
torque limiter)
CAUTION
1. Refer to Figure 19. Undo and remove the grub-
screw (11) and the grub-screw (6) on the Ensure that the pump-motor flange and the
MAINTENANCE
torque limiter (5). bottom flange of the coupling housing are clean and
free of burrs. If you do not, the pump-motor and
2. Use a suitable puller tool to remove the torque coupling housing may be misaligned and you may
limiter (5) from the motor shaft (7). damage the pump-motor, the pump or the torque
limiter or drive coupling when you operate the
3. Remove the key (10) from the shaft and dispose pump.
of the key.
4. Dispose of the key supplied with the new 1. Refer to Figure 19 or 20. Inspect the bottom
pump-motor. flange of the coupling housing (13) and the
flange of the pump-motor (8). The flanges must
5. Inspect the motor shaft (7) of the new pump-
be free of burrs and dirt. If necessary, clean or
motor (8). The motor shaft must be free of
refinish the flanges.
burrs and dirt. If necessary, clean or refinish the
motor shaft. 2. Fit the coupling housing (13) onto the flange of
the pump-motor (8). Ensure that the coupling
6. Inspect the motor shaft bore of the torque
housing is orientated correctly (as noted in
limiter (5). The bore must be free of burrs and
Section 5.15.3): one of the coupling cover
dirt. If necessary, clean or refinish the bore.
guards (2) must be at 90o to the terminal-box
7. Fit the new key (10, supplied in the Motor on the pump-motor.
Fitment Kit) into the motor shaft (7).
3. Fit the four bolts (9) and tighten the bolts to
8. On the EDP80 pump, fit the torque limiter (5) secure the coupling housing (13) to the pump-
onto the motor shaft (7) and push it fully down motor (8); note that you will fully tighten the
until the lower face of the torque limiter is 0.94 bolts in Section 5.16.8.
inches (24 mm) above the motor flange.
Issue L 62 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
5.16.8 Refit the pump-motor and coupling 5. Turn the torque limiter or coupling hub (16)
housing to the pump (without the until the drive dogs (4) align with the gaps
PAGE
motor change frame) between the drive coupling bushes (18) in the 63
coupling hub (16).
WARNING
6. Fully raise the pump-motor and coupling
MAINTENANCE
Ensure that the pump-motor, motor housing assembly so that the top flange of the
coupling and coupling housing coupling housing (13) locates against the
assembly does not fall over when you bottom flange of the pump gearbox, and so that
move it. If it falls over, it can cause the bolt holes in the coupling housing align with
injury to people. Refer to the Note in the bolt holes in the gearbox.
Section 5.16 to calculate the mass of
the assembly. 7. Use the bolts (12) to secure the coupling
housing and pump-motor assembly to the pump
CAUTION gearbox (1).
Ensure that the bottom flange of the pump gearbox
8. Refer to Figure 19 or 20, detail C. Check the
and the top flange of the coupling housing are clean
and free of burrs. If you do not, the pump gearbox gap between the bottom face of the coupling
and coupling housing may be misaligned and you hub (16) and the inner faces of the drive dogs
may damage the pump-motor, the pump or the (4). If the gap is correct, continue at Step 10.
torque limiter or drive coupling when you operate
the pump. 9. If the gap is not correct:
3. Use suitable lifting equipment to lift the pump- Check that the drive coupling or torque
motor and coupling housing onto the bottom of limiter (5) is correctly located on the motor
the pump frame. Carefully slide the pump- shaft (7): refer to Section 5.16.5 or 5.16.6.
motor and coupling housing assembly so that it
If the coupling housing and the drive coupling
is directly under the pump and so that the
or torque limiter are correctly fitted, loosen
pump-motor terminal-box is directly below the
the two grub-screws (15) on the coupling hub
gearbox oil drain-plug.
(16) and adjust the position of the coupling,
4. Look through the access hole in the coupling then tighten the two grub-screws (15) again.
housing (13) and use suitable lifting equipment Continue at Step 8 to check the gap is now set
to raise the pump-motor and coupling housing correctly.
assembly, until the drive dogs (4) are almost
10. Ensure that all of the fixing bolts (9, 12) are
engaged in the coupling hub (16). Ensure that
tightened to a torque between 94.4 and 97.4 lbf ft.
the top flange of the coupling housing is level at
all times. 11. Use the four bolts (3) to secure each coupling
cover guard (2) to the coupling housing.
Tighten the bolts to a torque between 2.2 and
3.6 lbf ft.
Feb 04 63 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
5.16.9 Refit the pump-motor and coupling 6. Hold the motor change frame (6) in place, fit a
housing to the pump (with the motor washer and nut (5) to the bottom of each stud
PAGE
64 change frame) (3) and tighten the nuts up the studs until the
motor change frame (6) is secured against the
WARNING top flange of the coupling housing.
MAINTENANCE
Ensure that the pump-motor, motor 7. Ensure that the top flange of the coupling
coupling and coupling housing housing is level at all times and alternately and
assembly does not fall over when you evenly tighten the three nuts (5) to start to raise
move it. If it falls over, it can cause the motor change frame (6) and the pump-
injury to people. Refer to the Note in motor and coupling housing assembly.
Section 5.16 to calculate the mass of
the assembly. 8. Refer to Figure 19 or 20. Look through the
access hole in the coupling housing (13) and
CAUTION continue to raise the motor change frame and
Ensure that the bottom flange of the pump gearbox the pump-motor and coupling housing assembly
and the top flange of the coupling housing are clean until the drive dogs (4) are almost engaged in
and free of burrs. If you do not, the pump gearbox the coupling hub (16).
and coupling housing may be misaligned and you
may damage the pump-motor, the pump or the 9. Turn the torque limiter or drive hub (5) until
torque limiter or drive coupling when you operate the drive dogs (4) align with the gaps between
the pump. the drive coupling bushes (18) in the coupling
hub (16).
1. Refer to Figure 19 or 20. Undo and remove the
10. Refer to Figure 21. Fully tighten the three lower
four bolts (3) which secure each of the two
nuts (5) on the studs (3) to raise the pump-
coupling cover guards (2) and remove the guard
motor and coupling housing assembly until the
from the coupling cover (13).
top flange of the coupling housing locates
2. Inspect the top flange of the coupling housing against the bottom flange of the pump gearbox.
(13) and the bottom flange of the pump gearbox
11. Use the two M16 bolts (10) to secure the
(1). The flanges must be free of burrs and dirt.
coupling housing and pump-motor assembly to
If necessary, clean or refinish the flanges.
the pump gearbox.
3. Refer to Figure 21. Use suitable lifting
12. Refer to Figure 19 or 19, detail C. Check the
equipment to lift the pump-motor and coupling
gap between the bottom face of the coupling
housing onto the bottom of the pump frame (4).
hub (16) and the inner faces of the drive dogs
Carefully slide the pump-motor and coupling
(4). If the gap is correct, continue at Step 14.
housing assembly so that it is directly under the
pump and so that the pump-motor terminal-
box (9) is directly below the gearbox oil drain-
plug.
Issue L 64 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
13. If the gap is not correct: 5.17 Relubricate the motor bearings
Check that the bottom flange of the coupling PAGE
WARNING 65
housing (13) is correctly located against the
top flange of the pump-motor: refer to Do not relubricate the motor
Section 5.16.7. bearings unless you have been
MAINTENANCE
suitably trained in the necessary
Check that the top flange of the coupling
procedures (and for explosion-proof
housing (13) is correctly located against the motors, unless you are fully
bottom flange of the pump gearbox (1): refer accredited/authorised to do so by
to Steps 1 to 11 and to the WARNING at the the motor manufacturer). If you are
start of this section. not suitably trained, you may damage
the motor and it may not operate
Check that the torque limiter or drive hub (5) correctly or safely.
is located on the motor shaft (7), so that the
lower face of item (5) is against the shoulder 5.17.1 Introduction
on the motor shaft: refer to Section 5.16.5 or
5.16.6. You must only relubricate the motor bearings of
If the coupling housing and the torque limiter standard motors if you have been suitably trained in
or drive hub are correctly fitted, loosen the all of the necessary procedures.
two grub-screws (15) on the coupling (16)
You must only relubricate the motor bearings of
and adjust the position of the coupling, then
explosion-proof motors if:
tighten the two grub-screws (15) again.
Continue at Step 12 to check the gap is now
You are fully accredited/authorised to do so
set correctly.
by the motor manufacturer.
14. Refer to Figure 20. Undo and remove the three
You have been suitably trained in all of the
nuts and washers (5) from the studs (3) and
necessary procedures.
remove the motor change frame (6) from the
studs. Undo and remove the three nuts and 5.17.2 General requirements
washers (2) and remove the three studs (3)
from the pump frame (4). Take note of the following when you relubricate the
15. Fit the remaining two M16 fixing bolts (8) to motor bearings:
fully secure the coupling housing and pump-
Refer to the guidelines given in Section 5.17.3
motor assembly to the pump.
for the relubrication intervals.
Feb 04 65 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
Only relubricate the bearings if they are in a To determine the required intervals between
satisfactory condition. The amount of grease relubrication:
PAGE
66
you should use is shown in Table 16. Do not
put too much grease into the bearings, or the Refer to Table 17 which provides general
motor may overheat. guidelines on the required interval between
relubrication.
MAINTENANCE
Issue L 66 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
WARNING 67
MAINTENANCE
in the necessary procedures (and for
explosion-proof motors, unless you
are fully accredited/authorised to do
so by the motor manufacturer). If
you are not suitably trained, you may
damage the motor and it may not
operate correctly or safely.
Feb 04 67 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
The pump-motor starts but the Has the torque limiter Recouple the pump-motor to the pump:
pump does not operate. decoupled the pump- refer to Section 4.5.
MAINTENANCE
The pump suddenly stops or the Is there a hydraulic lock in Switch off and drain the fluid from the
torque limiter continually the pump ? pump. If necessary, turn the pump by
decouples the pump-motor hand (refer to Step 7 of Section 5.11.3),
from the pump when reset. The then recouple the pump-motor to the
pump-motor may be tripping pump (refer to Section 4.5).
due to excessive electrical
current consumption. Has the pump seized due Switch off and then flush the pump (refer
to deposits ? to Section 5.10), then recouple the pump-
motor to the pump (refer to Section 4.5).
The pump stops but the torque Has the thermal snap- Check that the thermal snap-switch is
limiter does not decouple the switch operated to stop correctly set for the required operating
pump-motor from the pump. the pump because the temperature. If necessary, adjust the
pump is operating at too thermal snap-switch (refer to
high a temperature ? Section 3.12.1).
Is the TCV incorrectly set ? Check that the TCV is set to the required
operating temperature. If necessary,
adjust the TCV (refer to Section 3.12.3).
Is the coolant level too Check that the coolant level in the pump
low ? is correct. If necessary, add coolant (refer
to Section 5.10).
Issue L 68 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
MAINTENANCE
opens to release pressure in the
cooling-jacket. Has the TCV been set Adjust the TCV to a lower temperature
correctly ? setting (refer to Section 3.12.3).
The pump continues to operate Is the pressure relief valve Inspect the pressure relief valve and clean
at a high temperature which stuck in the open position ? it or replace it if necessary (refer to
may result in seizure. Sections 5.7 and 5.14).
The pump only achieves an Is the pressure relief valve See above.
ultimate pump-inlet pressure of stuck in the open position ?
22.5 to 37.5 torr.
The pump-motor trips out due Is the pressure relief valve See above.
to excessive electrical current stuck in the closed
consumption when the EDP200 position ?
pump is operating with pump-
inlet pressure in the range 337
to 750 torr, or when the
EDP300 pump is operating with
pump-inlet pressure in the range
187 to 750 torr.
The gearbox and oil are Has the shaft-seals purge Check that there is a flow of nitrogen
contaminated with the process nitrogen supply failed ? purge to the shaft-seals (look at the
substances pumped. nitrogen flow indicator); if necessary
adjust the pressure regulator (refer to
Section 3.12.2). Change the gearbox oil
before you restart the pump (refer to
Section 5.8).
Have the seals in the pump The seals must be replaced. Contact your
failed ? supplier or BOC Edwards for advice.
Feb 04 69 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
The pump does not operate. Is the pump-motor faulty ? Make all the other appropriate checks in
MAINTENANCE
Issue L 70 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
Feb 04 71 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
PAGE
72
Issue L 72 Feb 04
Drystar EDP Chemical Dry Vacuum Pumps
Feb 04 73 Issue L
Drystar EDP Chemical Dry Vacuum Pumps
You will need the correct Motor Bearing Kit to replace the motor
bearings as described in Section 5.18.
Issue L 74 Feb 04