Beneheart R12 Operator's Manual V2.0 en
Beneheart R12 Operator's Manual V2.0 en
Beneheart R12 Operator's Manual V2.0 en
Electrocardiograph
Operators Manual
Copyright 2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
For this Operators Manual, the issue date is December 2013.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in
this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaption and translation of this manual in any manner
whatsoever without the written permission of Mindray is strictly forbidden.
by Mindray in China and other countries. All other trademarks that appear in this manual are used only for editorial
purposes without the intention of improperly using them. They are the property of their respective owners.
All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained
herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this
manual.
Mindray is responsible for the effects on safety, reliability and performance of this product, only if:
All installation operations, expansions, changes, modifications and repairs of this product are conducted by
Mindray authorized personnel;
The electrical installation of the relevant room complies with the applicable national and local requirements;
WARNING
This equipment must be operated by skilled/trained clinical professionals.
It is important for the hospital or organization that employs this equipment to carry out a reasonable
service/maintenance plan. Neglect of this may result in machine breakdown or personal injury.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability
for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or
the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized
personnel.
Malfunction or damage caused by improper operation or repair by unqualified or unauthorized service people.
Malfunction of the instrument or part whose serial number is not legible enough.
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address Mindray Building, Keji 12th Road South, Hi-tech industrial park,
Nanshan, Shenzhen 518057,P.R.China
Website www.mindray.com
E-mail Address: service@mindray.com.cn
This manual is based on the maximum configuration and therefore some contents may not apply to your product. If
you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the equipment so that it can be
conveniently referenced when needed.
Intended Audience
This manual is geared for clinical professionals who are expected to have corresponding working knowledge of
medical procedures, practices and terminology as required for the treatment of patients.
Illustrations
All illustrations in this manual serve as examples only. They may not necessarily reflect the setup or data displayed on
your equipment.
Conventions
Italic text is used to quote the referenced chapters or sections.
12 Accessories..................................................................................................................................................................... 12-1
12.1 ECG Accessories ..........................................................................................................................................................................................12-1
12.2 Others .............................................................................................................................................................................................................12-2
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or
product/property damage.
NOTE
Provides application tips or other useful information to ensure that you get the most from your product.
WARNINGS
This equipment is used for a single patient at a time.
Before putting the system into operation, the operator must verify that the equipment, connecting cables
and accessories are in correct working order and operating condition.
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective
earth. If the installation does not provide for a protective earth conductor, disconnect it from the power line
and operate it on battery power, if possible.
To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres,
flammable anesthetics, or other flammable agents, such as gasoline.
Do not open the equipment housings. All servicing and future upgrades must be carried out by the
personnel trained and authorized by our company only.
Do not touch the patient when connecting peripheral equipment via the I/O signal ports to prevent patient
leakage current exceeds the requirements of applicable standards.
This equipment is not intended for use with high frequency surgical units.
Do not contact the patient during defibrillation. Otherwise serious injury or death could result.
For paced patients, the equipment may mistake a pace pulse for a QRS complex if several adverse conditions
exist simultaneously. Always keep these patients under close surveillance.
The physiological data and waveforms displayed on the equipment are for reference only and cannot be
directly used for diagnostic interpretation.
To avoid electric shock or equipment malfunction liquids is not allowed to enter the equipment. If liquids
have entered the equipment, remove the equipment from use and have it checked by service personnel
before it is used again.
To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard. Wrap and
secure excess cabling to reduce the risk of entanglement or strangulation by patients or clinical personnel.
Properly dispose of the package material according to applicable waste control regulations and keeping it
out of childrens reach.
CAUTIONS
Use only parts and accessories specified in this manual.
This equipment contains no user serviceable parts. Refer servicing to qualified service personnel.
At the end of its service life, the equipment, as well as its accessories, must be disposed of in compliance
with the guidelines regulating the disposal of such products. If you have any questions concerning disposal
of the equipment, please contact us.
Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For
this reason make sure that all external devices operated in the vicinity of the equipment comply with the
relevant EMC requirements. Mobile phone, X-ray equipment or MRI devices are a possible source of
interference as they may emit higher levels of electromagnetic radiation.
Before connecting the equipment to the power line, check that the voltage and frequency ratings of the
power line are the same as those indicated on the equipments label or in this manual.
Always install or carry the equipment properly to avoid damage caused by drop, impact, strong vibration or
other mechanical force.
1.1.3 Notes
NOTES
Locate the equipment where you can easily see the screen, access the operating controls, and disconnect the
equipment from AC power.
Keep this manual in the vicinity of the equipment so that it can be conveniently referenced when needed.
The software was developed in compliance with IEC60601-1-4. The possibility of hazards arising from
software errors is minimized.
This manual describes all features and options basing on the maximum configuration. Your equipment may
not have all of them.
NOTE
The equipment is intended to be used by clinical professionals or under their guidance. It must only be used by persons
who have received adequate training in its use. Anyone unauthorized or untrained must not perform any operation on
it.
Provide ECG algorithm to automatically analyze the acquired ECG waveforms; output measurements and diagnosis.
Print ECG reports through either an internal thermal recorder or an external printer.
Present messages in case of lead off, interference, low battery, or other abnormity.
6
1
5. Soft keys: for the equipment configured with a touchscreen. Press the soft keys to select the options.
Soft key labels: for the equipment not configured with a touchscreen.
6. Soft keys: only for equipment not configured with a touchscreen. The soft keys illuminate when the equipment is
powered on. Press the soft keys to select the options that appear on the right side of the screen. For the equipment
configured with a touchscreen, there are no keys in this area.
7. USB connector: connects USB devices, such as a USB drive, external printer, or barcode reader.
8. Patient cable connector: connects the patient cable for ECG acquisition.
Key Function
Power switch Turns on the equipment when the equipment is powered off.
Turns off the equipment by pressing and holding this key for 0.5 second when the equipment
is power on.
Forcefully shuts down the equipment by pressing and holding this key for 10 seconds when it
cannot be shut down normally.
Setup key Accesses the main menu.
Exits a menu and returns to the normal screen when the menu is open.
Indicator Description
Power indicator On: when the equipment is powered on.
Off: when the equipment is powered off.
Battery indicator Green: when the equipment operates on battery power or the battery is being charged.
Yellow: when the equipment operates on battery power and the battery is low.
Yellow and blink: when the equipment operates on battery power and the battery is
depleted.
Off: when no battery is installed or the battery is fully charged.
AC indicator On: when the AC mains is connected.
Off: when the AC mains is not connected.
2 13
4 12
5 5
6
11
7 8 9 10 9
Key Function
Starts printing.
Stops printing.
1 2 3 4 5 6 7
1. Battery compartment
2. USB connector: connects USB devices, such as a USB drive, external printer, or barcode reader.
4. Network connector: a standard RJ45 connector for software upgrade and sending ECG data.
6. Equipotential Grounding Terminal: when using the monitor together with other devices, connect their
equipotential grounding terminals together to eliminate the potential differences between them.
7. AC power input: connects the power cord to run the equipment on AC power supply.
2 5
3 4
This area shows patient information including patient ID, gender, age, and heart rate; system information including
network and USB device connecting status, system time, battery status; etc.
5. Soft key area: shows the soft keys. For the equipment not configured with a touchscreen, this area shows the labels
of the soft keys located rightward.
3 4
1. Heading: shows the menu heading and system information including network and USB device connecting status,
system time, battery status; etc.
Soft key labels: for the equipment not configured with a touchscreen.
In the Normal mode, you can acquire the patients electrocardiographic information, record ECG waveforms,
measurements, and diagnoses. You can also configure the equipment and export data.
2. Select [Maintenance], and then select [Demo Mode 1] or [Demo Mode 2].
To exit the Demo mode, turn off the equipment and restart it.
WARNING
The Demo mode is for demonstration purpose only. To avoid the potential risk of the simulated data being
mistaken for the patient data, do not enter the Demo mode during ECG acquisition.
WARNING
The equipment shall be installed by personnel authorized by us.
Do not open the equipment housings. All servicing and future upgrades must be carried out by the
personnel trained and authorized by our company only.
The software copyright of the equipment is solely owned by us. No organization or individual shall resort to
altering, copying, or exchanging it or to any other infringement on it in any form or by any means without
due permission.
Devices connected to the equipment must meet the requirements of the applicable IEC standards (e.g. IEC
60950 safety standards for information technology equipment and IEC 60601-1 safety standards for medical
electrical equipment). The system configuration must meet the requirements of the IEC 60601-1-1 medical
electrical systems standard. Any personnel who connect devices to the equipments signal input/output
port is responsible for providing evidence that the safety certification of the devices has been performed in
accordance to the IEC 60601-1-1. If you have any question, please contact us.
If it is not evident from the equipment specifications whether a particular combination with other devices is
hazardous, for example, due to summation of leakage currents, please consult the manufacturers or else an
expert in the field, to ensure the necessary safety of patients and all devices concerned will not be impaired
by the proposed combination.
If the packing case is intact, open the package and remove the equipment and accessories carefully. Check all materials
against the packing list and check for any mechanical damage. Contact us in case of any problem.
WARNING
When disposing of the packaging material, be sure to observe the applicable waste control regulations and
keep it out of childrens reach.
The equipment might be contaminated during storage and transport. Before use, please verify whether the
packages are intact, especially the packages of single use accessories. In case of any damage, do not apply it
to patients.
The equipment operating environment should be reasonably free from noises, vibration, dust, corrosive, flammable and
explosive substances. If the equipment is installed in a cabinet, sufficient space in front and behind should be left for
convenient operation, maintenance and repair. Moreover, to maintain good ventilation, the equipment should be at
least 5 cm away from around the cabinet.
When the equipment is moved from one place to another, condensation may occur as a result of temperature or
humidity difference. In this case, never start the system before the condensation disappears.
WARNING
Make sure that the operating environment of the equipment meets the specific requirements. Otherwise
unexpected consequences, e.g. damage to the equipment, could result.
To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres,
flammable anesthetics, or other flammable agents, such as gasoline.
Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For
this reason make sure that all external devices operated in the vicinity of the equipment comply with the
relevant EMC requirements. Mobile phone, X-ray equipment or MRI devices are a possible source of
interference as they may emit higher levels of electromagnetic radiation.
The mains plug is used to isolate the equipment circuits electrically from the SUPPLY MAINS. Do not position
the equipment so that it is difficult to operate the plug.
Before connecting the equipment to the power line, check that the voltage and frequency ratings of the
power line are the same as those indicated on the equipments label or in this manual.
NOTES
Put the equipment in a location where you can easily see the screen and access the operating controls.
Keep this manual in the vicinity of the equipment so that it can be conveniently referenced when needed.
Before connecting the equipment to the AC mains, check that the voltage and
frequency ratings of the power line are the same as those indicated besides the
AC power input.
1. Connect the female end of the power cord with the AC power input on the equipments back.
2. Connect the male end of the power cord with a wall AC outlet.
WARNING
Use only the supplied power cord.
Where the integrity of the external protective conductor in the installation or its arrangement is in doubt,
the equipment shall be operated from the battery. Otherwise the patient or operator might be shocked.
The battery must only be installed by service personnel trained and authorized by our company. No battery is installed
when the equipment leaves the factory. Contact your service personnel to install the battery before putting the
equipment into use.
To prevent data loss in case of sudden power failure, we recommend you always install a fully charged battery in the
equipment.
The battery is charged whenever the equipment is connected to an AC power source regardless of whether or not the
equipment is currently turned on.
When the battery is being charged, the battery indicator is illuminated in green. The on-screen battery symbol
dynamically shows the charging status if the equipment is powered on.
NOTE
Charge the battery before it is first put into use.
Black mark
The equipment can print either on A4
(295 mm 210 mm) or US Letter (8.5 11) paper. The paper tray is configured to meet the appropriate paper size for
the destination location when the equipment leaves the factory.
To change the paper size, move the white plastic spacer bar in the paper tray to limit the paper tray.
For A4 sized paper, insert the spacer in the slot at the top of the paper tray.
For US Letter sized paper, insert the spacer in the slot at the bottom of the paper tray.
NOTE
Use only thermal recording paper we supply.
2. Tight the screws to securely attach the patient cable to the equipment.
NOTE
Restore the barcode reader to factory default configuration before using it.
Operating environment
Check and make sure that there is no electromagnetic interference source around the equipment, especially large
medical electrical equipment such as radiological equipment and magnetic resonance imaging equipment etc.
Switch off these devices when necessary.
Keep the examination room warm (no less than 18 C) to avoid muscle action voltages in ECG signal caused by cold.
Power supply
Check that power supply specification is met and the power cord is securely connected if the mains power is used.
Use only power socket that is properly grounded.
Check that a battery is installed and fully charged if you want to run the equipment on battery power.
Patient cable
Recording paper
WARNING
This equipment is not intended for use with high frequency surgical units.
1. Connect the equipment with AC mains. If you run the equipment on battery power, ensure that the battery is
sufficiently charged.
WARNING
Do not use the equipment on a patient if you suspect that it is not working properly, or if it is mechanically
damaged. Contact your service personnel or us.
You can also set other items as needed. Refer to 4 System Setup for details.
Then press and hold the Power switch for approximately 0.5 second to turn off the equipment.
CAUTION
Although not recommended, you can press and hold the Power switch for 10 seconds to forcibly shut down
the equipment when it could not be shut down normally or under some special situations. This may cause
loss of data.
Press the [Prev] or the [Next] soft key to move to the previous or the next menu item.
Press the [Back] soft key or the [Esc] key on the keyboard to return to the previous menu.
Press the [Select] or the [Cancel] soft key to select or deselect a menu item.
Press the [Enter] soft key or the Enter key on the keyboard to confirm the selection.
The settings in the main menu are saved as user defaults and remain effective even after the equipment is turned off
and restarted.
Muscle Artifact 20 Hz , 35 Hz, Off 35 Hz Sets the default frequency of muscle artifact filter.
Filter Muscle artifact filter attenuates noise in the
waveform by restricting the frequencies that are
included.
The muscle artifact filter is a low-pass filter. That is to
say signals that exceed the set frequency are filtered
out.
[35 Hz]: only signals at 35 Hz or less display. Signals
exceeds 35 Hz are attenuated.
[20 Hz]: only signals at 20 Hz or less display. Signals
exceeds 20 Hz are attenuated.
[Off]: signals at 150 Hz or less display.
Baseline Drift Selected, not selected Selected Select whether the baseline drift removal (BDR)
Removal process or 0.05-Hz filter is used.
If selected, BDR is enabled. This process suppresses
most baseline drift interference and also is able to
preserve the fidelity of the ST-segment level.
If not selected, BDR is disabled and the 0.05-Hz filter
is used.
NOTE: BDR or 0.05-Hz selection applies to the
displayed ECG, printed report, and analyzed and
stored data.
BDR introduces around 1-second delay. We
recommend use of BDR except when the delay is
unacceptable.
Both BDR and 0.05-Hz selections meet requirements
of the 1990 American Heart Association
Recommendations for Standardization and
Specifications in Automated Electrocardiography:
Bandwidth and Signal Processing pertaining to
lower-frequency response in electrocardiography.
AC Filter Selected, not selected Selected Selects whether electrical interference is filtered
from AC line voltage.
If selected, the AC filter is enabled to filter electrical
interference from AC line voltage.
Note: The AC filter should be on. Turn off only if
necessary.
Screen Waveform 31, 61, 34+1, 34+3, 34+1 Selects the default format of ECG waveforms
Format 62, 62+1, 121 displayed on the screen.
[31]: displays 12-lead ECG waveforms in four
pages, with 3 waveforms in one column in each
page.
[61]: displays 12-lead ECG waveforms in two
pages, with 6 waveforms in one column in each
page.
[121]: displays 12-lead ECG waveforms in one
page in one column.
[62]: displays 12-lead ECG waveforms in one page
in two columns, with 6 lines in each column.
[34+1]: displays 12-lead ECG waveforms in one
page in 4 columns, with 3 lines in each column, and
one rhythm lead waveform at the bottom.
So it is with [34+3] and [62+1].
Speed 5 mm/s, 12.5 mm/s, 25 mm/s, 25 mm/s Selects the default printing speed.
50 mm/s
Gain 2.5 mm/mV, 5 mm/mV, 10 10 mm/mV Select the default amplitude of 1mV ECG signal.
mm/mV, 20 mm/mV, Auto, L The larger the setting is, the larger the waveform
10 C5, L=20 C=10 size. However, only the appearance of the waveform
changes. The signal strength is not affected.
[L=10 C=5]: displays the limb lead waveforms at an
amplitude of 10 mm/mV; displays chest lead
waveforms at an amplitude of 5 mm/mV.
[L=20 C=10]: displays the limb lead waveforms at
an amplitude of 20 mm/mV, displays chest lead
waveforms at an amplitude of 10 mm/mV.
[Auto]: automatically selects the gain as per the
amplitude of ECG waveforms.
Pacemaker Label Selected, not selected Selected Selects whether a mark is placed on each ECG
waveform when a pace pulse is detected.
If selected, a pace pulse mark is placed on each
ECG waveform when a pace pulse is detected.
If not selected, no mark is placed when a pace pulse
is detected.
Lead Sequence Standard, Cabrera Standard Select ECG lead sequence for displaying and
printing.
[Standard]: the sequence is I, II, III, aVR, aVL, aVF, V1,
V2, V3, V4, V5, V6.
[Cabrera]: the sequence is aVL, I, -aVR, II, aVF, III, V1,
V2, V3, V4, V5, V6.
1
Fridericia: QTc QT HeartRate 3
60
The detailed information helps you to know more about the patient.
Menu item Option Default Description
Secondary ID Selected, not selected Not selected Selects whether the patients secondary ID is
included on the ECG report as patient information.
DOB Selected, not selected Not selected Selects whether the patients date of birth is
included on the ECG report as patient information.
Race Selected, not selected Not selected Selects whether the patients race is included on the
ECG report as patient information.
Medication 1 Selected, not selected Not selected Selects whether the medication taken by the patient
is included on the ECG report as patient information.
Medication 2 Selected, not selected Not selected Selects whether the medication taken by the patient
is included on the ECG report as patient information.
Class 1 Selected, not selected Not selected Selects whether the patients class is included on the
ECG report as patient information.
Class 2 Selected, not selected Not selected Selects whether the patients class is included on the
ECG report as patient information.
V3 Placement Selected, not selected Not selected Selects whether the setting of V3 placement is
included on the ECG report as patient information.
Physician Selected, not selected Not selected Selects whether the physician who supervises the
ECG is included on the ECG report as patient
information.
Technician Selected, not selected Not selected Selects whether the technician who conducts the
ECG measurement is included on the ECG report as
patient information.
Department Selected, not selected Not selected Selects whether the patients department is
included on the ECG report as patient information.
Room Selected, not selected Not selected Selects whether the patients room number is
4.6 Maintenance
Menu item Option Default Description
Network Type LAN, WLAN LAN Selects the type of network through which the
equipment is connected.
Network Name / / When connects WLAN, enters the SSID.
(SSID)
Password / / Enters the password to connect the WLAN.
IP Address 0 - 255 192.168.0.100 Enters the IP address of the equipment.
Subnet Mask 0 - 255 255.255.255.0 Enters the subnet mask of the equipment.
Default Gateway 0 - 255 192.168.0.254 Enters the IP address of the default gateway.
Server IP Address 0 - 255 192.168.0.101 Enters the IP address of the FTP server.
FTP Port 0 - 65535 21 Enters FTP port.
FTP Username / / Enters FTP username.
FTP Password / / Enters FTP password.
Demo Mode 1 / / Enters the password to access Demo Mode 1. To
exit the Demo mode, turn off the equipment and
restart it.
Demo Mode 2 / / Enters the password to assess Demo Mode 2. To
exit the Demo mode, turn off the equipment and
restart it.
Restore Default / / Restores the factory default configuration.
Configuration This does not change the current language setting.
Load / / Imports the configuration file on the USB drive to
Configuration the internal memory.
Export / / Exports the configuration file on the internal
Configuration memory to the USB drive.
Print / / Prints the current configuration.
Configuration
Language ENGLISH, SIM. CHINESE, ENGLISH Selects UI language.
FRENCH, GERMAN, ITALIAN,
POLISH, SPANISH,
PORTUGUESE, RUSSIAN,
CZECH, TURKISH,
HUNGARIAN, ROMANIAN
AC Filter 50 Hz, 60 Hz 50 Hz Selects the frequency of the AC power line filter.
Patient information is classified as required information and detailed information. The required information must be
entered. In the [Patient Info] menu, an asterisk (*) is placed before the required information. The detailed information
helps you to know more about the patient.
2. Select [Basic Setup] [Patient Info Setup] to enter the [Patient Info Setup] menu.
3. Select the required patient information and detailed patient information as necessary.
You can:
NOTE
You can save patient information only when all the required patient information is entered.
We recommend using pediatric lead placement V4R, V1, V2, V4 - V6 if the patient is under 16 years of age.
Please record V4R using the V3 electrode. Also set [V3 Electrode Placement] to [V4R]. This is a normal
practice for a patient of this age.
2. Press down the button on the reader handle, and target the reader to the barcode.
Then the [Patient Info] menu pops up with the patient ID entered.
After the patients information is saved, the patient is added to the Patient List. The Patient List can include up to 500
patients.
2. Select [Edit Patient ID] to enter the [Edit Patient ID] sheet.
Editing patient information updates the information of corresponding patient in the Patient List.
Assure the patient that there is no danger or discomfort involved. Explain that full cooperation will produce a
valuable diagnostic record.
Lay the patient on a bed with arms rest at the side and legs lying flat and not touching. Ensure the patient is
comfortable and relaxed.
Once the electrodes and lead wires are applied, instruct the patient to:
Breathe normally.
The more relaxed the patient is, the less the ECG will be affected by noise.
3. Degrease each electrode site with alcohol and abrade slightly with dry gauze to remove dead skin cells.
CAUTION
Ensure that all leads are connected and all electrodes are applied to correct sites. Ensure the conductive
parts of the patient cable and electrodes, including the neutral electrode, do not contact other conductive
parts, including earth.
Polarizing electrodes may cause the electrodes to retain a residual charge after defibrillation. Residual
charge will block the acquisition of ECG signal.
Never mix patient electrode types or brands. Dissimilar metals or other incompatibilities may cause
considerable baseline drift and may increase trace recovery time after defibrillation.
Do not reuse disposable electrodes. Reuse may cause a risk of contamination and affect the measurement
accuracy.
Reusable electrodes shall be cleaned and disinfected before applying to the patient.
The patient cable consists of a connector, a trunk cable, 4 limb lead wires and 6 chest lead wires. The lead wires are
color-coded. Refer to 6.4.3 Lead Wire Color Code.
1
2
2. Trunk cable
The chest electrode consists of a bulb and a metal electrode. On the metal electrode, there are two lead wire
connectors: one for lead wire with 3.0 mm connector; the other for lead wire with 4.0 mm connector.
1. Bulb
1
2. Lead wire connector (3.0)
Limb Electrode
The limb electrode consists of a plastic clamp and a metal electrode. On the metal electrode, there are two lead wire
connectors: one for lead wire with 3.0 mm connector; the other for lead wire with 4.0 mm connector.
1
1. Lead wire connectors
2. Metal electrode
3. Clamp
2 3
Note
The limb electrodes are color coded. Make use limb lead wire and limb electrode of the same color are
connected.
RL(N) LL(F)
V4R
Limb electrodes should be placed on fleshy areas above the inside wrists and ankles, not on the bone.
2. Connect the four limb electrodes with corresponding lead wires as indicated by the color. Route the lead wires to
avoid twisting.
7. Place the electrodes on the limb sites above the inside ankles and wrists.
8. Make sure the patient cable is tightly connected to the equipment and electrodes are correctly connected with
the lead wires.
2. Connect the six chest electrodes with the chest lead wires. Route the lead wires to avoid twisting.
5. Apply a thin layer of conductive gel on each electrode site. Ensure the gel from one site does not touch another
site.
7. Apply the electrodes by squeezing the rubber bulb and allowing suction to hold the electrodes in place.
8. Make sure the patient cable is tightly connected to the equipment and electrodes are correctly connected with
the lead wires.
The reusable electrodes contain nickel, a material that can cause skin irritation. Monitor the electrode sites
and, if irritation occurs, use an alternate electrode.
NOTE
To obtain high-quality ECG signal, make sure that the metal electrodes firmly contact the skin.
When placing the chest electrodes, ensure that the metal electrodes do not touch each other and the
conductive gel from one application site does not touch another site.
The metal plate of the limb electrode may be loose due to frequently plugging and unplugging the lead
wire. Make sure the lead wire is firmly connected with the electrode.
Limb electrodes should be placed on fleshy areas above the inside wrists and ankles, not on the bone.
4. Route the lead wires to avoid twisting. Connect the lead wires with the electrodes.
5. Make sure the patient cable is tightly connected to the equipment and electrodes are correctly connected with
the lead wires.
When any of the electrodes on the patients left arm, left leg, or right arm is detached, or any of LA/L, LL/F, RA/R
lead is off, the system respectively prompts LA Lead Off (L Lead Off), LL Lead Off (F Lead Off), or RA Lead
Off (R Lead Off).
When any of the chest electrodes or leads is detached, the system respectively prompts V (X) Lead Off (C (X)
Lead Off), in which X represents 1 - 6.
When RL/N electrode or lead is off, or two or more limb leads are detached, or the patient cable detaches the
equipment, the system prompts Limb Lead Off.
In this case, check that the electrodes are firmly attached to the skin, the lead wires are properly connected with the
electrodes, and the patient cable is tightly connected to the equipment.
Do not contact the patient during defibrillation. Otherwise serious injury or death could result.
For paced patients, the equipment may mistake a pace pulse for a QRS complex if several adverse conditions
exist simultaneously. Always keep these patients under close surveillance.
Ensure that all leads are connected and all electrodes are applied to correct sites. Ensure the conductive
parts of the patient cable and electrodes, including the neutral electrode, do not contact other conductive
parts, including earth.
To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard. Wrap and
secure excess part to reduce the risk of entanglement or strangulation by patients or clinical personnel.
The bulb of the chest electrode contains latex, a material that can cause skin irritation. Monitor the
electrode site and, if irritation occurs, use an alternate electrode.
The auto measurements and diagnoses are for reference only and cannot be directly used for patient
treatment.
Press the first soft key to adjust the current waveform speed.
Press the second soft key to adjust the current waveform size.
Press the third soft key to adjust the current frequency of the muscle artifact filter.
You can also configure the ECG waveforms by accessing the [Wave Setup] menu. Refer to 4.2 Waveform Setup for
detail.
3. Adjust waveform speed, waveform size, and the frequency of muscle artifact filter.
4. Check other waveform and report settings by selecting Setup [Waveform Setup] and [Report Setup].
If the preview option is disabled, the equipment automatically prints the ECG report after ECG data is acquired and
analyzed.
If the preview option is enabled, the preview of the ECG report displays. You can:
Select the [Home] soft key or the [Esc] hard key to discard the report and return to the normal screen.
Select the [Send] soft key to send the report to the external device.
Select the [Next Page] soft key to display the next page of the report, if there is any.
If the auto save function is disabled, select the [Save] soft key to manually save the report to the internal storage.
The equipment automatically stops recording when the ECG report has been printed. You can also press the [Stop] soft
key to interrupt printing.
4. Adjust waveform speed, waveform size, and the frequency of muscle artifact filter.
5. Check other waveforms and report settings by selecting Setup [Waveform Setup] and [Report Setup].
Select the [1 mV] soft key to place a 1 mV square wave on each waveform.
The rhythm measurement provides only printed report. There are no measurement results and diagnoses. You cannot
save the report or send it to the external device.
3. Set [Rhythm Format], [Rhythm lead 1], [Rhythm Lead 2], and [Rhythm Lead 3] by selecting Setup [Report
Setup].
If you set [Rhythm Format] to [One Lead], the waveform of the selected rhythm lead displays in 6 cascade
lines, with each line including 10 seconds of waveforms on the report.
If you set [Rhythm Format] to [Three Leads], the waveforms of the selected rhythm leads display in 3 cascade
lines, with each line including 20 seconds of waveforms on the report.
4. Check other waveforms and report settings by selecting Setup [Waveform Setup] and [Report Setup].
Then the equipment starts acquiring ECG data and a countdown displays. When 60 seconds are reached, printing starts.
The rhythm measurement automatically stops when the report is finished. You can also select the [Stop] soft key to
manually interrupt it.
NOTE
Do not touch the metal electrodes or connectors when acquiring and recording an ECG. Otherwise
inaccurate measurements may results.
To use an external printer, set [Printing Device] to [External printer] by selecting Setup [Report Setup].
Before printing a report, check that the paper is properly loaded. Refer to 3.2.3 Loading the Paper for loading the paper
for the thermal recorder. To load the paper for the external printer, refer to the printers accompanying instructions for
use.
NOTE
For LaserJet M401n, on the printer select [System Setup] [Paper Setup] [Tray 1]/[Tray 2], set [Paper
Size] to [Any Type]..
To print another copy of the latest auto or rhythm ECG report, press on the keyboard.
You can copy the report using the current configuration, or change the settings before printing another copy.
If auto save is disabled, you can manually save a report when a preview of the report is generated.
Measurements, including:
PR Interval (ms)
P/QRS/T Axes ()
Acute MI/Ischemia
Extreme Tachycardia
Extreme Bradycardia
Significant Arrhythmia
Diagnoses
Median Complex
Measurement Matrix
Pon (ms), Pdur (ms), QRSon (ms), QRSdur (ms), Qdur (ms), Rdur (ms), Sdur (ms), R'dur (ms), S'dur (ms), P+dur (ms),
QRSdef (ms), P+amp (V), P-amp (V), QRSp2p (V), Qamp (V), Ramp (V), Samp (V), R'amp (V), S'amp (V),
STamp (V), 2/8STT (V), 3/8STT (V), T+amp (V), T-amp (V), QRSarea (V*ms), Rnotch, DWconf (%), STslope (deg),
Ton (ms), Tdur (ms), T+dur (ms), QTint (ms).
The diagnoses of 12-lead ECG analysis is included on the ECG report by default, see Report Analysis Setup in 4.3 Record
Setup.
Resting 12-lead ECG analysis is not intended for the manual measurement and rhythm measurement.
Refer to12-Lead ECG Interpretive Program Physicians Guide (PN: 046-004817-00) for details.
1 2 3 4
5 6 7 8 9 10 11
Select [Export Configuration] to export the current configuration to the USB drive.
You can send the patients reports in either of the following ways:
Automatically
During auto measurement, the equipment automatically sends out the current report in XML format through the
network after the measurement is finished.
Manually
You can manually send the selected files to the FTP server through the network, or send them to the USB drive
connected to the equipment. The format of the files sent to the USB drive can be XML, Mindray, or PDF. Refer to [File
Format] as described in 4.4 File Management.
If you have problems to send out the patients reports, contact your service personnel.
The equipment does not power up. 1. The equipment is not turned on. 1. Verify the equipment is turned on.
2. The equipment is not connected to 2. Verify the equipment is properly
the AC mains or the power cord is connected to the AC mains.
poorly connected. 3. Verify the equipment receives power
3. External power supply problems, from the AC mains. Replace the power
such as damaged power cord or AC cord or AC power outlet if necessary.
power outlet. 4. Verify the battery is installed and has
4. Battery is not installed or has no sufficient charge. Otherwise, connect
charge when the AC mains is not the equipment to the AC mains to run
connected. the equipment and charge the battery.
The display is completely blank. 1. The equipment is power off. 1. Check that the equipment is turned
2. The equipment is in the Standby on.
mode. 2. Press any key to exit Standby.
The display is frozen. Software failure. 1. Press and hold the Power switch for
10 seconds to forcibly shut down the
equipment.
2. Restart the equipment.
Wrong characters are entered. Wrong entering method. Verify the entering method is correct.
No response to keystroke. 1. One or more keys on the keyboard 1. Verify no other key is pressed and
are being pressed and held. held.
2. Software failure. 2. Press and hold the Power switch for
10 seconds to forcibly shut down the
equipment.
3. Restart the equipment.
The barcode reader cannot read the The barcode reader is not properly Properly connect the barcode reader to
patient ID. connected to the equipment. the equipments USB port.
The recorder does not work. 1. Paperless recording is enabled. 1. Select Setup [Record Setup] and
2. Recording paper is not loaded. disable [Paperless Recording].
3. The paper tray is not snap in place. 2. Verify the recording paper is properly
5. The thermal recorder is disabled due 3. Verify the paper tray is snap in place.
Paper jammed or misaligned. 1. Unspecified paper is used. 1. Verify specified paper is used.
2. Recording paper is not properly 2. Take out the paper and tear off the
loaded. jammed part. Reload the paper as
3. The paper tray spacer is not properly described in 3.2.3 Loading the Paper.
placed. 3. Verify the paper tray spacer is placed
appropriately for the paper size. Refer
to 3.2.3 Loading the Paper for detail.
Some or all leads have no waveforms. 1. Defective or broken ECG cable. 1. Replace the ECG cable with a new
2. ECG cable is not connected. one.
Baseline drift for one or more leads. 1. Unspecified electrodes are used or 1. Use specified accessories. Do not mix
mix electrode types and brands. electrode types or brands.
2. Poor skin preparation. 2. Prepare the patient skin before ECG
3. Electrode problems. acquisition as described in 6.2
Preparing the Skin.
3. Verify electrodes are correctly
applied as described in 6.4.1Electrode
Placement. Check for defective or
expired electrodes. Replace with
disposable electrodes if necessary.
ECG data displays unacceptable noise. 1. Patient movement. 1. Ensure the patient remains
2. AC interference from external motionless during ECG acquisition.
devices or improper AC filter setting. 2. Turn of the adjacent devices or move
3. Muscle artifact or improper muscle the electrocardiograph away from the
artifact filter setting. interference if possible. Properly set the
4. Poor skin preparation. AC filter.
The equipment automatically shuts 1. Auto shutdown is enabled and the 1. Enable or disable [Auto Shut Down]
down. equipment is inactive for the by selecting Setup [Basic Setup] as
predefined time. desired.
2. The battery is depleted when the 2. Connect the equipment to AC mains
equipment runs on battery power. to run the equipment and charge the
battery.
9.2 Messages
The equipment prompts messages to indicate the current system status.
Some messages, see 9.2.1 Message List 1, are important and urgent, and need you to acknowledge or take actions in
time. The system pops up a dialog box when these messages happen. In this case, you cannot operate the equipment
unless you press any key to clear the pop-up message or wait till the triggers disappear.
Some pop-up messages also display in the message area and disappear till the triggers disappear.
Some messages, see 9.2.2 Message List 2, are less urgent. These messages are shown in the message area. They
disappear automatically when the triggers disappear.
The equipment can give a notification tone when a message is presented. The notification tone is switched off by default.
You can enable it by accessing the [Basic Setup] menu. Refer to Notification Tone in 4.5 Basic Setup.
However, the equipment always gives a notification tone when some messages occur regardless of the setting of
[Notification tone]. Refer to the messages followed by an asterisk in the messages lists below.
Indicates that the battery works properly. The solid green portion represents the current battery charge level. Each
block represents a charge of approximately 20% capacity.
Indicates that the battery has low charge level and needs to be charged. In this case, the LED turns yellow and the
message Low Battery shows at the bottom of the screen.
Indicates that the battery is almost depleted and needs to be charged immediately.
When the battery is depleted, the system pops up the message Battery Depleted, the battery indicator flashes in
yellow, and the recorder is disabled. At this moment, connect the equipment to the AC mains to run the equipment and
charge the battery. Otherwise the equipment will shut down.
When the battery is being charged, the battery indicator is illuminated in green. The on-screen battery symbol
dynamically shows the charging status if the equipment is powered on.
To get the most out of the battery, observe the following guidelines:
Perform the battery performance test once a year, before equipment repairs, or whenever the battery is suspected
as being the source of the problems.
Condition the battery once when it is used or stored for 3 months, or when its operating time becomes noticeably
shorter.
Take out the battery before the equipment is transported or will not be used for more than 3 months.
Store the battery with battery power about 50% of the full charge. Every 6 months, fully charge the battery, and
then run the equipment on this battery till its power becomes 50% of the full charge. Remove the battery from the
equipment and store it.
When storing batteries, make sure that the battery terminals do not come into contact with metallic objects. Store
the batteries in a cool place, ideally at a temperature of 15 C. Storing batteries in a cool place slows the aging
process, while storing batteries at high temperature for an extended period of time will significantly shorten
battery life. Do not store the battery at a temperature beyond -20 C 60 C.
WARNING
Keep the battery out of childrens reach.
2. Connect the equipment to the AC mains. Allow the battery to be charged uninterrupted till the battery is full and
the battery indicator is off.
3. Disconnect the AC mains and allow the equipment to run from the battery until it shuts off.
4. Again connect the equipment to the AC mains. Allow the battery to be charged uninterrupted till the battery is full
and the battery indicator is off.
NOTE
The actual battery capacity decreases over time. For an old battery, the full capacity battery symbol does
not indicate the capacity and operating time of this battery can still fulfill battery specifications in the
operators manual. Please replace the battery if its operating time is significantly lower than the specified
time.
2. Connect the equipment to the AC mains. Allow the battery to be charged uninterrupted till the battery is full and
the battery indicator is off.
3. Disconnect the AC mains and allow the equipment to run from the battery until it shuts off.
Battery operating time directly reflects its performance. If the operating time of a battery is noticeably shorter than that
stated in the specifications, contact your service personnel.
NOTE
Battery operating time depends on the device configuration and operation. The battery might be damaged
or malfunctioned if its operating time is too short after being fully charged.
When a battery has visual signs of damage, or no longer holds a charge, it should be replaced.
WARNING
Do not disassemble, puncture or incinerate batteries. Do not short the battery terminals. They may ignite,
explode, or leak, causing personal injury.
WARNING
Failure for the responsible individual, hospital or institution employing this equipment to implement a
satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
This equipment contains no user serviceable parts. Refer servicing to qualified service personnel.
The safety checks or maintenance involving any disassembly of the equipment should be performed by
professional servicing personnel. Otherwise, undue equipment failure and possible health hazards could
result.
If you discover a problem with any of the equipment, contact your service personnel or us.
Always dilute the cleaning and disinfecting agent according to the manufacturers instructions or use the lowest
possible concentration.
Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or
acetone-based cleaners).
WARNING
Be sure to shut down the system, disconnect power cord and other cables before cleaning the equipment.
Use only the substances approved by us and methods listed in this chapter to clean or disinfect your
equipment. Warranty does not cover damage caused by unapproved substances or methods.
We make no claims regarding the efficacy of the listed chemicals or methods as a means for controlling
infection. For the method to control infection, consult your hospitals Infection Control Officer or
Epidemiologist.
CAUTION
Remove the equipment from use if liquid is spilled on the equipment or accessories. Contact your service
personnel.
Water
Mild soap
Do not use any of the following materials to clean the equipment because equipment damage may result.
Ammonia-based solvents
Acid or alkaline cleaning agents such a sodium hypochlorite and peroxide solvents
For the recommended cleaning agents for the reusable accessories, refer to the instructions for use delivered with the
accessories.
Your equipment should be cleaned regularly. If there is heavy pollution or lots of dust and sand in your place, the
equipment should be cleaned more frequently. Before cleaning the equipment, consult your hospitals regulations for
cleaning the equipment.
1. Shut down the equipment and disconnect the power cord, accessories, and other devices that are connected
with the equipment.
3. Soak a clean and soft cloth in the solution and wring out excess solution.
4. Thoroughly wipe the surface of the equipment with the damp cloth, avoiding the connectors.
Do not drip the solution or any liquid on the keyboard and the opening of the thermal recorder.
Remove the cables and lead wires from the equipment before cleaning.
1. Gently wipe the cables and lead wires with a soft cloth dampened with the cleaning agent, avoiding the metal
connectors.
3. Dry the cables and lead wires in a ventilated and cool place.
1. Gently wipe the electrodes surface with a soft cloth dampened with 75% ethanol, avoiding the metal connectors.
Dirty print head deteriorates printing quality. Clean the print head at least once per month or as needed. Check the
printout to ensure the printing is legible and dark. Light printing may indicate a dirty print head.
2. Pull out the paper tray. Take out the recording paper.
3. Gently wipe the print head with cotton swabs dampened with water or ethanol to remove the dust and foreign
particles.
5. Reload the recording paper and push back the paper tray after the print head is completely air dry.
CAUTION
The print head gets hot when recording. Do not clean the print head immediately after recording.
11.1.2 Disinfecting
Disinfection may cause damage to the equipment and is therefore not recommended for this equipment unless
otherwise indicated in your hospitals servicing schedule. Cleaning equipment before disinfecting is recommended.
The recommended disinfectant for the equipment is 75% ethanol. For the recommended disinfectant agents for the
reusable accessories, refer to the instructions for use delivered with the accessories.
11.1.3 Sterilization
Unless otherwise specified in the instructions for using an accessory, do not sterilize the equipment or the accessories.
The housing and display screen are free from cracks or other damages.
Connectors are not loose, cracked, or bent and cables have no cuts, nicks, or fraying.
Power cord and patient cable are securely connected with the equipment.
Chest electrode bulbs are free from cracks and limb electrodes can properly clamp.
After your equipment has been used for 6 to 12 months, or whenever your equipment is repaired or upgraded, a
thorough inspection should be performed by qualified service personnel to ensure the reliability.
Make sure that the environment and power supply meet the requirements.
Inspect power cord, patient cable and lead wires for damage, and make sure that their insulation is in good
condition.
Make sure that the recorder functions correctly and the recorder paper meets the requirements.
In case of any damage or abnormity, remove the equipment from use. Contact the hospitals biomedical engineers or
your service personnel immediately.
Then the symbol appears at the top left corner of the screen.
2. Tap the center of the symbol to align the touchscreen. Then the symbol moves to the next position.
The equipment automatically exits touchscreen calibration and displays the message Touchscreen Calibration
Completed after the calibration is completed. You can press the Setup key to interrupt touchscreen calibration.
Store in a cool, dark, and dry place, avoiding high temperature, moisture and direct sunlight.
Avoid contact with cleaning fluids and solvents, such as alcohols, ketones, esters, ether, and so on.
Do not store thermal paper with polyvinyl chloride or other chemicals which cause yellowing and fading.
Store each report separately in a paper bag. Avoid long-term overlapping or pressing by weight.
NOTE
Use only specified thermal paper. Using other paper may result in print head wearing out prematurely or
recording of poor quality.
Hang cables and lead wires vertically or around a big wheel, avoiding twisting or sharp-angle bending.
Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination and affect
the measurement accuracy.
Check the accessories and their packages for any sign of damage. Do not use them if any damage is
detected.
The accessory material that contacts the patients has undertaken the biocompatibility tests and is verified
to be in compliance with ISO 10993-1.
Patient Cable
A.5.3 Indicators
Power indicator 1 (green)
AC indictor 1 (green)
Battery indictor 1 (two colors: yellow and green)
0.4 dB
Frequency response 0.05 Hz-150 Hz ( 3.0 dB )
Overall system error is tested using the method described in AAMI EC11 3.2.7.1.
Accuracy of input signal Overall system error is5%.
reproduction Frequency response is tested using the method described in AAMI EC11 3.2.7.2
methods A and D.
Common mode rejection ratio 110 dB
AC filter 50/60 Hz
1 kHz (A/D)
ECG sampling rate
Accuracy: 1V/LSB
Pacer detection sampling rate 16 kHz/channel, two channels
Input signal range 10 mV (peak-to-peak value)
Input impedance 50 M @10 Hz, any two electrodes
Note
Using accessories, transducers and cables other than those specified may result in increased
electromagnetic emission or decreased electromagnetic immunity of the device.
The device or its components should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the device or its components should be observed to verify normal operation in the
configuration in which it will be used.
The device needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided below.
Other devices may interfere with this device even though they meet the requirements of CISPR.
When the inputted signal is below the minimum amplitude provided in technical specifications, erroneous
measurements could result.
Portable and mobile communication equipment may affect the performance of this device.
Other devices that have RF transmitter or source may affect this device (e.g. cell phones, PDAs, and PCs with
wireless function).
WARNING
This equipment/system is intended for use by healthcare professionals only. This equipment/ system may
cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or relocating the [ME EQUIPMENT or ME SYSTEM] or shielding the
location.
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Note 3: The device that intentionally receives RF electromagnetic energy at the exclusion band (2395.825MHz-2487.645MHz)
is exempt from the essential performance requirements, but remains safe.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the [ME EQUIPMENT or ME SYSTEM] is used exceeds the applicable RF
compliance level above, the [ME EQUIPMENT or ME SYSTEM] should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the [ME EQUIPMENT or
ME SYSTEM].
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than3V/m.
WARNING
The device is configured with a wireless network connector to receive wireless signal. Other devices may
interfere with this device even though they meet the requirements of CISPR.
Description
Item
IEEE 802.11b IEEE 802.11g IEEE 802.11n
Operating Frequency Band
2412 - 2472 2412 - 2472 2412 - 2472
(MHz)
Modulation DSSS and CCK OFDM OFDM
Transmitter Output Power
<20 <20 <20
(dBm)
The radio device used in this product is in compliance with the essential requirements and other relevant provisions of
Directive 1999/5/EC (Radio Equipment and Telecommunications Terminal Equipment Directive).
WARNING
Keep a distance of at least 20cm away from the device when Wi-Fi function is in use.
V microvolt
s Microsecond
A ampere
Ah ampere hour
C centigrade
cm centimeter
dB decibel
F fahrenheit
g gram
GHz gigahertz
h hour
Hz hertz
in inch
k kilo
kg kilogram
kPa kilopascal
L litre
m meter
Mb mega byte
mg milligram
min minute
ml milliliter
mm millimeter
mV millivolt
mW milliwatt
M megaohm
s second
V volt
VA volt ampere
ohm
W watt
C.2 Symbols
Minus sign, negative, or hyphen
% percent
/ per; divide; or
+ plus or positive
= equal to
plus or minus
multiply
copyright
AC alternating current
CE Conformit Europenne
DC direct current
ECG electrocardiograph
HR heart rate
ID identification
IP internet protocol
LA left arm
LL left leg
MR magnetic resonance
R right
RA right arm
RL right leg
All tests can be performed using commercially available safety analyzer test equipment. These procedures assume the
use of a 601PROXL International Safety Analyzer or equivalent safety analyzer. Other popular testers complying with IEC
60601-1 used in Europe such as Fluke, Metron, or Gerb may require modifications to the procedure. Follow the
instructions of the analyzer manufacturer.
The electrical safety inspection should be periodically performed every two years. The safety analyzer also proves to be
an excellent troubleshooting tool to detect abnormalities of line voltage and grounding, as well as total current loads.
LIMITS
ALL COUNTRIES R = 0.2 Maximum
The following outlet conditions apply when performing the Earth Leakage test.
LIMITS
For UL60601-1,
For IEC60601-1,
The following outlet conditions apply when performing the Patient Leakage Current test.
LIMITS
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity;
Reversed Polarity
LIMITS
The following outlet conditions apply when performing the Patient Auxiliary Current test.
LIMITS
NOTE
Make sure the safety analyzer is authorized comply with requirement of IEC61010-1.