Service Manual of PM-600 Portable Patient Monitor: Shenzhen Mindray Bio-Medical Electronic Co., LTD
Service Manual of PM-600 Portable Patient Monitor: Shenzhen Mindray Bio-Medical Electronic Co., LTD
Service Manual of PM-600 Portable Patient Monitor: Shenzhen Mindray Bio-Medical Electronic Co., LTD
of
PM-600
Portable Patient Monitor
Version: 1.0
Issued date: 2002/06/11
P/N No.: 6003-20-11803
Statement
In the event of inadvertent or deliberate publication, Mindray intends to enforce its rights to this work
under copyright laws as a published work. Those having access to this work may not copy, use, or
disclose the information in this work unless expressly authorized by Mindray to do so.
All information contained in this publication is believed to be correct. Mindray shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the furnishing,
performance, or use of this material. This publication may refer to information and protected by
copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights
of others. Mindray does not assume any liability arising out of any infringements of patents or other
rights of third parties.
I
PROPERTY OF
SHENZHEN MINDRAY® BIO-MEDICAL ELECTRONICS CO., LTD. 2002
ALL RIGHTS RESERVED
Mindray is responsible for safety, reliability and performance of this equipment only in the condition
that:
all installation, expansion, change, modification and repair of this equipment are conducted by Mindray
qualified personnel; and,
applied electrical appliance is in compliance with relevant National Standards; and,
the monitor is operated under strict observance of this manual.
′ NOTE ′
This equipment is not intended for family usage.
Warning
This monitor is not a device for treatment purpose.
It is important for the hospital or organization that employs this equipment to carry out a reasonable
maintenance schedule. Neglect of this may result in machine breakdown or injury of human health.
Upon request, Mindray may provide, with compensation, necessary circuit diagrams, calibration
illustration list and other information to help qualified technician to maintain and repair some parts,
which Mindray may define as user serviceable.
II
Warranty
Mindray guarantees new equipment other than accessories to be free from defects in workmanship and
materials for a period of one year (six months for multi-site probes and SpO2 sensor) from date of
shipment under normal use and service. Mindray's obligation under this warranty is limited to repairing,
at Mindray’s option, any part which upon Mindray's examination proves defective.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the improper
use or application of the product or the substitution upon it of parts or accessories not approved by
Mindray or repaired by anyone other than a Mindray authorized representative.
This warranty shall not extend to any instrument which has been subjected to misuse, negligence or
accident; any instrument from which Mindray's original serial number tag or product identification
markings have been altered or removed, or any product of any other manufacturer.
Mindray is not responsible for the effects on safety, reliability and performance of the PM-9000
Portable Patient Monitor if:
assembly operations, extensions, re-adjusts, modifications or repairs are carried out by persons other
than those authorized by Mindray.
the PM-9000 Portable Patient Monitor is not used in accordance with the instructions for use, or the
electrical installation of the relevant room does not comply with NFPA 70: National Electric Code or
NFPA 99: Standard for Health Care Facilities (Outside the United States, the relevant room must
comply with all electrical installation regulations mandated by the local and regional bodies of
government).
I
Return Policy
Return Procedure
In the event that it becomes necessary to return a unit to Mindray, the following procedure should be
followed:
Obtain return authorization. Contact the Mindray Service Department and obtain a Customer Service
Authorization (Mindray) number. The Mindray number must appear on the outside of the shipping
container. Return shipments will not be accepted if the Mindray number is not clearly visible.
Please provide the model number, serial number, and a brief description of the reason for return.
Freight policy. The customer is responsible for freight charges when equipment is shipped to Mindray
for service (this includes customs charges).
Company Contact
I
Table of Contents
I. General ........................................................................................................................................1-1
II. Buttons .......................................................................................................................................1-3
III. Hardware Principle ...................................................................................................................1-7
IV. Software Principle ....................................................................................................................1-3
I. NIBP ............................................................................................................................................2-1
II. SpO2 ...........................................................................................................................................2-1
I. System Checks.............................................................................................................................3-1
II. Testing And Calibrating Each Parameter...................................................................................3-9
Classification ..................................................................................................................................... 1
Specifications..................................................................................................................................... 1
Introduction
Chapter 1 Introduction
I. General
PM-600 portable patient monitor is lightweight and compact, which weights 1.8kg including battery but
not cable and accessories. The AC power for the monitor is 100~250V. Also, the monitor can be
connected to the recorder via RS232 interface.
Electrical Description
PM-600 can be powered both by the AC or DC power supply. The external AC power supply
(100~250V, 50/60Hz) is connected to the monitor via a 3-core wire. When connected to AC power, the
AC indicator on the front panel lights on indicating that the instrument is charging the built-in battery.
During the charging process, the user may press POWER button to start up the monitor.
When no AC power is connected, the monitor may also work because it can be powered by the built-in
battery. The battery with full capacity can support the monitor to work for 2 hours (operating conditions:
temperature is 25℃, all monitoring mode, perform NIBP measurement once every 15 minutes.)
When powered by the built-in battery, the yellow BATT. indicator on the front panel lights on. At the
same time, the monitor will constantly check the remaining battery capacity. When the remaining
capacity can only support the monitor to work normally for 5 minutes, the yellow BATT. indicator
begins flashing together with audible alarm sound. In this situation, the operator should connect the AC
power to the monitor to avoid it sudden shutdown.
If the monitor stops working suddenly when powered by the external AC power, the built-in battery can
automatically begin to power the monitor to ensure its continuous operation.
Display
The screen of PM-600 series monitors consists of some displays made up of LED nixie tubes. These
displays are used to show the system status and measured results of those parameters in numerical form.
The appearance of the screen is shown in figure 1-1.
NOTE
When only measuring NIBP, the display of PULSE is the measured result from NIBP
measurement. The system updates the pulse value until the end of NIBP measurement. The value
will be also saved until the end of next NIBP measurement.
The measured results of SpO2 include pulse rate (PULSE) and %SpO2. These values are updated
at least once per second.
II. Buttons
PM-600 has 8 buttons. Of different models, the functions of these buttons may be different. See
following table.
Model
LED display and buttons
PM-600I PM-600II PM-600III
AC Available Available Available
BAT. Available Available Available
mmHg Available Available Not available
kPa Available Available Not available
ADULT Available Available Not available
NEO. Available Available Not available
LED Display
SYSTOLIC Available Available Not available
DIATOLIC Available Available Not available
MAP/CUFF Available Available Not available
TIME/CYCLE Available Available Not available
PULSE Available Available Available
%SpO2 Not available Available Available
Pulse Strength Not available Available Available
Operating POWER Available Available Available
buttons Available but
STAT Available Available
not for operation
SETUP Available Available Available
∧ Available Available Available
∨ Available Available Available
Available but
MEMORY Available Available
not for operation
SILENCE Available Available Available
The table below gives the meanings of LED symbols on the front panel. LED display has two kinds:
LED indicator and LED nixie tube.
Meanings of LED Symbols
This indicator lights on in green when the external AC power is connected implying
AC that in this condition the monitor can work normally and charge the internal battery
at the same time.
This indicator lights on when the monitor works powered by the internal battery
BATT. instead of the external AC power supply. This indicator flashes when the battery
has only low capacity so as to raise the operator’s attention.
mmHg Used to indicate NIBP measurement unit. Red color.
kPa Used to indicate NIBP measurement unit. Red color.
Used to indicate adult patient. Red color. Pediatric and adult belong to the same
ADULT
patient type.
NEO. Used to indicate neonate patient. Red color.
SYSTOLIC Used to display the measured rsullt of systolic pressure of NIBP measurement.
DIASTOLIC Used to display the measured result of systolic pressure of NIBP measurement.
Used to display the measured result of mean pressure of NIBP measurement.
MAP/CUFF
During measurement, it is used to display the cuff pressure.
Used to display 24-hour system clock and countdown the time of auto NIBP
TIME/CYCLE
measurement, i.e., display the serial No. when reviewing the measured results.
Used to display the pulse rate of SpO2 and NIBP measurement. For PM-600I,
PULSE display the pulse rate of NIBP measurement. For PM-600II, display the pulse
rate of SpO2 measurement.
SpO2 (%) Used to display the measured result of SpO2 measurement.
Pulse Strength Indicator of the pulse strength of SpO2 measurement.
Rear panel:
Connectors for connecting the external power supply and RS-232 serial port are on the rear panel
(figure 1-3). The label of the device is also on the rear panel. See figure below. Table below lists out the
description of each connector on the rear panel.
Power board
CPU main
control board
LED display board
4
P03
AC P03
6
P07
P07 Key 1
P05
4
P08 P01
COM accumulator
3 8
accumulator
P02
P01 Key 8
SPO2
SPO2 6 probe cable
module
Probe socket
2
P1 P06
33X2
NIBP
Probe socket
P04 L04
airway
NIBP
module P02
(J1)
Front side
speaker
Power board
NIBP module
Communication
interface SPO2 module
keyboard
Speaker
Accumulator 2
NIBP
SPO2
Probe socket
Probe socket
Figure 1-5 Internal structure
3. Power board:
Parameters:
Input voltage: 100~250Vac,50/60HZ
DC output: 12V/450mA,3.3V/550 mA
Lead-acid accumulator charging management
Power on/off control, over-voltage protection, over-current protection, short-circuit protection
Power on/off
logical control and
+5V
protect circuit
Battery
Charging
Managerent
3.3V BUCK
converter
Battery
Parameters and waveforms at some critical points: Use multimeter to measure the voltage value
of the critical points.
After connecting the AC power, the DC voltage across C12 is about 1.3 ~1.4 times as much as as the
incoming AC voltage. The voltage between 1PIN of PCB2 to earth is within 11~15V, which is the
working voltage of U1. The voltage across C11 is 17.6V. When battery is not connected, the voltage
across C20 is 13.8V. Voltage across C46 is 5V, which is used as the working voltage of U7, U8, U9
and U14. The voltages of 8PIN of U1, 1PIN of U10 and 11PIN of U11 to earth are 5V. Following
waveforms can be viewed using oscillograph. The waveform of the voltage of 4PIN of U1 is a standard
sawtooth, whose oscillating frequency is about 120KHz. After connecting a rating load as the output,
the waveform of the voltage of 1PIN of Q1 to earth shall be a square one, whose frequency is about
120KHz. The waveform of the voltage across D12 should be a square one, whose frequency is about
280KHz.
using the software to control the brightness of LEDs and nixie tubes. Nevertheless, to realize the
function of audio alarm, we have to use a special alarm circuit to drive a speaker. The signal is acquired
by the frequency-sharing circuit in FPGA. FPGA can output frequency signals respectively having four
different output frequencies, i.e., 2048Hz, 1024Hz, 512Hz and 256Hz.
AC power/battery monitoring and indication
After undergoing AC/DC conversion in the power board, a DC signal is introduced into the main
control board. This DC signal can be used to directly drive AC LED. After this signal has been
potential-divided by resistance, we can obtain a voltage of about 3V ---- ACS, which is used by MCU.
When the system is powered by battery, ACS is at the low level, based on which, MCU can judge
whether the power has been switched and accordingly make the power indicator light on.
Linear voltage regulating circuit:
The main control board has two power inputs, 3.3V and 12V. However, components on the main
control board require 5V. Accordingly, we select a linear voltage regulator---7805, which are used to
generate a 5V voltage from 12V voltage.
Display board:
LED display board consists of 7-segment LED nixie tubes of either big or small size (red, common
cathod, followings are the same), LED indicators (red, green, yellow), and scan driving circuit. By
pluging the socket P04 into the corresponding pin L04 on the CPU board, 5V working voltage and
display data can be obtained. LED display uses the way of scanning drive.
The LED nixie tubes and LED lamps of the same color have even and consistent display brightness.
The frame scanning frequency of display is ≥100. The display is flicker-free. When the system has
alarm, the LED nixie tubes will flash regularly on the LED numerical display position of the
corresponding parameter. At the same time, speaker gives sound to realize audio and visual alarm
function.
5. Battery
Use two 6V 1.0Ah closed rechargeable lead-acid accumulators, which are serial connected.
Measure again.
E04 SpO2 sensor off Check if the sensor is still connected to the patient. Check if the
sensor is loosely connected. Check if the sensor cable is incorrectly
connected or damaged.
E05 There is Check if the sensor on the patient is correctly connected. Change
interference when the measuring position. Check if the red tube inside the sensor no
meausring SpO2. longer flashes (flashing means working normally). Measure again.
E06 The recorder does Check if the recorder has paper. Check if the paper is installed
not work. correctly. Check if the recorder cable is correctly and safely
connected. Print again.
E77 NIBP unit failure If there is SpO2 unit, the system can continue to execute SpO2
measurement but cannot execute NIBP measurement. NIBP unit
needs repair.
E88 SpO2 unit failure If there is NIBP unit, the system can continue to execute NIBP
measurement but cannot execute SpO2 measurement. SpO2 unit
needs repair.
E99 Monitor system The monitor needs repair.
failure
I. NIBP
The monitor measures non-invasive blood pressure using the oscillometric method. Following are
detailed measurement procedures. Inflate the cuff encircled around the upper arm until the pressure in
the cuff blocking the blood flow in the artery of the upper arm. Then deflate the cuff according to the
requirement of a certain algorithm. With the decreasing of the pressure in the cuff, the artery blood will
palpitate with the pulse, which results in palsation in the cuff. Through the pressure sensor connected
with the inflating pipe of the cuff, a palsation signal palpitating with the pulse will be generated. After
being filtered by a high-pass filter (about 1Hz), this signal becomes pulsating signal and is amplified.
Then the amplified signal is converted into digital signal by A/D. After using the single chip to process
this digital signal, we may obtain systolic pressure, diastolic pressure and mean pressure. Be careful to
choose appropritate cuffs for neonatal, pediatric and adult patients so as to avoid the generation of
measurement error. NIBP module also has protection circuit to prevent the cuff from being inflated to a
very high pressure. The main operating modes of NIBP are;
1) adult/pediatric/neonate: select according to the patient shape, weight and age.
2) manual measurement, auto measurement, continuous measurement. Manual measurement is also
called single measurement. It means the monitor only performs one measurement for each time. Auto
measurement means to perform one measurement within selected cycle. Time interval can be set up
as 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240 and 480 minutes. Continuous measurement means
after being activated, the monitor will perform quick measurement continuously within 5 minutes.
Continuous measurement is effective in monitoring changes in blood pressure.
II. SpO2
By tracing the pulse waveform in the fingertip, using specified algorithm and consulting the clinical
data table, we can obtain the SpO2 value. The SpO2 sensor consists of two LEDs and a photodetecter.
The two LEDs are respectively red diode and infrared diode, which are lighted according to certain time
sequence. When the capillary vessel of the fingertip congests repeatedly, the light of the LEDs is
absorbed by blood vessels and organs and then projected onto the photodetecter. The photodetecter can
detect the light intensity varying with pulse changes and display the changing light intensity in the form
of changing electronic signals. The ratio between the DC and AC components of the two types of
signals is the % oxygen in the blood. Then we can calculate correct SpO2 value by using specified
algorithm and also calculate pulse rate according to the SpO2 waveform.
The SpO2 module mainly consists of following four parts: sensor, signal processing, control unit of
LED driving sequence, singlechip.
1) Sensor: the two LEDs alternatively emit red light and infrared light onto the part being monitored.
The photocell then converts the received lights into electro signals.
2) Signal processing: The electronic signal output from the photocell are amplified by the measurement
amplifier, high-pass filter and program controller and finally converted into digital signal by A/D.
The D/A converter is used to control the baseline of the signal, amplify the AC component and
finally output an appropriate pulse waveform. The digital signal output from A/D are to be processed
by the singlechip.
3) Control unit of LED driving sequence: The time sequence circuit and DAC together control the
alternative operation of the two LEDs and the light intensity. The singlechip control the driving
current based on the algorithm.
4) Singlechip: it is made up of CPU, RAM, ROM and interface circuit.
I. System checks
For the conventional testing contents of PM-600 portable patient monitor, please refer to its Operation
Manual. The information in this chapter is only a brief introduction. The following sections are used to
emphasize important tests and the information not clearly specified in the Operation Manual.
1. Device appearance and installation checks
1)The shell of the device is clean and has no scratches. The installation is stable. When shaking the
device, these are no inside leftovers.
2)Buttons are smooth and free for operation.
3)Labels are complete and sufficient and correct in delivering information.
4)Standard configuration is complete, the sockets are installed safely.
5)Perform vibration test on the machine before performing following operating tests.
2.Safety tests
2.1.Test equipment
1.Safety analyzer 501 PRO 1
2.Leakage current/grounding resistance measurement kit: 1
3. Connection kit of the application part: 1
4.Tinsel 20cm X 10cm 1
2.2.Test procedures
2.2.1 Leakage current to earth
2.2.1.1 Connection graph for testing is as shown in figure 1:
Connect one end of the 3-core power wire to AC220V network power, the other end to the leakage
current testing kit (A). Insert the 3-core power wire of safety analyzer (B) into power output socket of
(A). Connect the 3-core power wire of the device being tested ① into AC output of 501. Connect the
sensor of the application part based on the requirements of (C). Connect the red measurement clip RED
of 501 to the ground protection PE terminal. Connect the SUM terminal of (C) to the P terminal of (A).
Locate all switches to “OFF” position.
(A)-----grounding resistance/leakage current measurement kit
(B)-----501 safety analyzer
(C)-----application part processing kit
①------device being tested
⑤------application part
RED---501 red measurement clip
SUM--- kit post
2.2.1.2 Adjust input voltage: When the device being tested is in shutdown state, connect leakage current
measurement kit (A) with the input network voltage AC220V. Adjust the booster to raise the testing
voltage to 110% (that is 253V) of the nominal voltage 230V. This voltage is monitored by the voltage
meter of (A). Then turn on the device to let the device be in the operating state, micro-adjust the booster
to make the output voltage keep stablely at 253V. Press the [Ground] key of the 501 tester and
disconnect the grounding wire.
2.2.1.3 Leakage current between network source and earth in normal state: press the [Leakage] key of
the 501 tester and read the leakage current value on it. Connect SW12, in the condition that the
application part is connected to the earth, respectively press and release [Polarity] key to toggle
between the null line and the live wire. Then disconnect SW12 and cut off the connection between the
application part and the earth. Respectively press and release the [Polarity] key. The maximum value of
these four measurements should be less than 0.5mA.
2.2.1.4 Leakage current between network source and the earth in single fault condition:
Leakage current when connection between null line and live wire is being cut off: Press the [Leakage]
key of the 501 tester. Then press the [Neutral] key of the 501 tester, disconnect N line. Respectively
press and release the [Polarity] key to toggle between null line and live wire, and imitate the condition
that L line is disconnected. Read the leakage current value on the 501 tester. Connect SW12, press and
release the [Polarity] key. Disconnect SW12, press and release the [Polarity] key. The maximum value
of these four measurements should be less than 1.0mA.
of the application part based on the requirements of (C). Stick the tinsel A on any position of ① (never
let A touch live part, protection earth or the application part).Connect the red clip RED of the 501 tester
onto the tinsel A. Connect the SUM terminal of (C) to the P terminal of (A). Locate all switches to
“OFF” position.
Disconnect SW, press and release the [Polarity] key. The maximum value of these four measurements
should be less than 0.5mA.
the [ECG leak] key of 501, then press the arrow on the panel of 501 to select the RA-GND item. The
measured leakage current should be less than 0.01mA.
2.2.3.4 Patient leakage current of the application part when network voltage is added.
2.2.3.4.1 Press the [ECG leak] key of 501, then press the arrow on the panel of 501 to select the
RA-GND item. Press the [Isolation] key and add network voltage. Test the leakage current of the added
network voltage. The maximum current value should be less than 0.05mA.
2.2.4.2 Patient auxiliary current in the normal state (adjust the input voltage by referring to 3.1.2)
AC auxiliary current of the RA lead of ECG to other application parts: position the RA lead on the
connection kit of the application part to “ON” and other switches to “OFF”. Connect RA-P to the RA
binding post of 501. Connect other patient parts to LL through SUM. Press the [ECG leak] key of 501,
then press the arrow on the panel of 501 to select the RA-LL item. The tested current value should be
less than 0.01mA.
2.2.4.3.2 DC auxiliary current of the RA lead of ECG to other application parts (DC value of RA):
Press the [DC Only] key of 501, take turns to operate the [Ground] key (for disconnecting the ground
wire), the [Neutral] key (for disconnecting the null line) and the [Polarity] key (for toggling between
null line and live wire). Test the current in the above three fault conditions, the maximum current value
should be less than 0.05mA.
2.2.5 Testing grounding resistance
2.2.5.1 Connection graph for testing ground resistance is shown in figure 5:
Note: In the graph, P1 and P2 are two binding posts of grounding resistance testing kit. Keep the
measuring wires “Black” and “Red” as short as possible. The sectional area of the wire should be larger
than 10mm2. It is acceptable to use more than 3 pieces of parallel-connected 10WAG wires. GND is the
grounding terminal of either the power wire of the device being tested or the power plug. EP is the
3-6 Service Manual of PM-600 Portable Patient Monitor (V1.0)
Checks & Tests
grounding terminal to the device (for the current patient monitor, EP is equipotential binding post). C
are all the protecting metal covers (shells) that are connected to PE. M are all metal screws that are
connected onto EP. C and M are all on the device shell.
3.Testing SpO2
1)Testing tool
SpO2 simulator
2)Testing procedures
① Connect SpO2 simulator with the SpO2 probe of PM-600
② Set up the parameters of SpO2 simulator as following:
SpO2=98
PR=70
③ Check if the displayed SpO2 and PR values on PM-600 are consistent with those on the
simulator.
(Note: To observe the PR value, select PLETH as the HR source in the ECG menu.)
④ Change the setup values of SpO2 and PR on the simulator.
⑤ Check the displayed values on PM-600 are consistent with the setup values.
⑥ Make SpO2 sensor fall off, in this condition, PM-600 should immediately give alarm.
Chapter 4 Troubleshooting
Chapter 5 Installation
I. Unpack inspection
Open the package and take out the packing list. Check if the names, quantity and specifications of the
goods in the package are consistent with those on the packing list. Please note that:
1) If the user buys optional parts or other accessories, he should also verify if they are placed in the
package.
2) If the goods in the package are not consistent with those on the packing list, please contact the
supplier.
3) If the device or any part is damaged during transportation, please save all packing material and
goods for future inspection and immediately contact the supplier.
Chapter 6 Cleaning
The user should use specified method to clean the PM-600 portable multi-parameter monitor. To
prevent the monitor from being contaminated or damaged, please use recommended materials. To
guarantee patient safety, do not use cables, sensor or accessories that have already been deteriorated or
damaged.
I. Cleaning
1) Use soft cloth to clean the monitor.
2) Use soap, amino, or ethanol based material to clean the host.
3) Use soap, amino or ethanol based material to clean the reusable cuff.
4) Use soap, amino or ethanol based material to clean the reusable SpO2 sensor.
NOTE
1) Ammonia solution: diluted ammonia solution <3%;
Ethanol: ethonal 70%, isopropanol 70%
2) Don't use the grinding material, such as steel wool etc.
3) Do not let cleaner enter the device or immerse the part being cleaned.
III. Disinfection
Disinfect the PM-600 when necessary. Because disinfection process is complex, please execute the
procedures according to the hospital regulation. Pay special attention to safety. EtO gas is usually used
for disinfection.
Chapter 7 Maintenance
PM-600 portable patient monitor is a type of precision electronic medical device having complex
structure. Maintaining PM-600 carefully will not only let the device develop its performance to the best
but also ensure the long-term operating accuracy of the device and avoid various errors. To prevent
cross contamination, ensure that the device has undergone cleaning and disinfection before
maintenance.
1. Frequently check the device, cables, sensors and wires for damage.
2. Clean the device irregularly according to the actual requirement.
3. Perform safety test annually.
4. Perform NIBP parameter calibration test annually.
5. Test overall functions of the device annually.
6. Perform safety test once after each opening-chassis repair.
7. If finding problems during maintenance, contact your supplier in time.
Appendix: Specifications
Following is the specifications of PM-600II monitor, which has complete configuration of both NIBP
and SpO2. For PM-600III (measure SpO2 only) and PM-600I (measure NIBP only), without special
declariation, their specifications are included in the specifications of PM-600II.
1.General specifications:
Size: 60mm×180mm×220mm
Weight: 1.8 kg
Display Specification:
Display: LED
Display Information:
Power Supply Indicator: AC & BATT
NIBP Measurement Unit: mmHg & kPa
NIBP Patient Kind:: ADULT & NEO (neonate)
NIBP Measurement Display: SYSTOLIC; DIASTOLIC, MAP
Pulse: PULSE
SpO2: %SpO2
Clock: 24-hour
Cuff Pressure: CUFF
Pulse Intensity: LED Indicator
2.Safety Specifications
Standard compliable: IEC60601-1
Anti-electroshock type Class I equipment and internal powered equipment
EMC type Class A
Anti-electroshock degree SpO2 and NIBP are CF degree
Against water ingress degree Ordinary equipment (sealed equipment without liquid proof)
Operation mode Continuous running equipment
3.Power Supply
AC input:
100-250VAC, 50/60Hz, 40VA
FUSE: T 0.5A
Battery:
Two rechargeable 1.3 A/Hr 6V Lead-Acid battery
Operating time: ( Condition: 25℃, NIBP working period is 10 min. )
PM-600I Monitor: working time is 4 hours with full battery capacity.
PM-600II Monitor: working time is 2 hours with full battery capacity.
PM-600III Monitor: working time is 3 hours with full battery capacity.
Operating time after the first alarm of low battery is about 5 minutes
4.Environment
Temperature
Working 0 ~ 40 (°C)
Transport and Storage -20 ~ 60 (°C)
Humidity
Working 30% - 85 %
Transport and Storage 30% - 93 %(no condensation)
EMC According with the requirement of EN60601-1-2.
5.Monitoring Parameter
5.1 NIBP
Method Oscillometric
Mode Manual, Auto, STAT
Measuring Interval in AUTO Mode
1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480
(min)
Measuring Period in STAT Mode: 5 min
Pulse Rate Range 40 - 240 (bpm)
Cuff Pressure Range: 0 - 300 (mmHg)
NIBP Measurement:
Measuring and Alarm Range
Adult Mode
SYS 40 ~ 270 (mmHg)
DIA 10 ~ 215 (mmHg)
MAP 20 ~ 235 (mmHg)
Neonatal Mode
The SYSTOLIC and DIASTOLIC values measured using this device shall be equal to those measured
by trained medical personnel using cuff and stethoscope. The MEAN value measured using this device
is equal to that measured using endartery blood pressure measurement device.
5.2 SpO2
5.3 Trend
Storage method of trend data: combined storage in two ways specified below:
NIBP: store NIBP data each time after one measurement. Maximumly store NIBP data of 600 times.
SpO2: sotre once every 30 seconds (the average value of the measured results within these
6. Signal Interface
To connect Recorder: RS-232 serial Interface
7. Recorder (optional )
Type: thermal
Record Width 48 mm
Recording types:
Real-time recording
Trend data printing
Power Supply:
100-250VAC, 50/60Hz