APX HF3 Technical Manual
APX HF3 Technical Manual
APX HF3 Technical Manual
APX HF 3
TECHNICAL MANUAL
Ref : 81-40-001-F
2
Surgical C-arm APX HF 3
Contents
Safety........................................................................................ 5
Protection against Radiation .................................................. 5
Transport .................................................................................. 5
Installation ................................................................................ 6
Operating ................................................................................. 6
Cleaning and Disinfection ....................................................... 6
Storage ..................................................................................... 6
Life time .................................................................................... 7
Characteristics....................................................................... 11
1 General characteristics ...................................................................... 11
2 Radioscopy ....................................................................................... 11
3 Radiography ......................................................................................12
4 Protection devices .............................................................................12
5 Functioning indicators ........................................................................12
6 Other characteristics ..........................................................................12
Ref : 81-40-001-F
3
Technical Manual
Configuration ......................................................................... 28
1 Use of the configuration mode ............................................................28
2 Exit from configuration mode ..............................................................28
Maintenance ........................................................................... 40
1 How to replace the 63A fuse ...............................................................40
2 Description of test points ....................................................................40
3 Process for centring the laser module ................................................41
Diagrams ................................................................................ 48
Ref : 81-40-001-F
4
Surgical C-arm APX HF 3
Thank-you
Ref : 81-40-001-F
5
Technical Manual
Safety
The user is requested to observe national laws concerning the use and installation
of x-ray equipment.
This equipment is not intended to function in an atmosphere containing explosive gas.
A check of the reliability and efficiency of the apparatus every 12 months is recommended
for the safety of both patient and operator.
Worn parts may become dangerous; they must be checked and replaced by new ones.
Before using the unit, the operator is to ensure that all safety devices are in working order.
The condition of the cables and connections is to be regularly checked; they are to be
replaced if they show signs of wear.
In case of any doubt, call the Technical Department.
Before each utilisation, the operator must check the leaded collimator. To do it, make a
fluoroscopy, press on keys 24 and 23 (see page 10), and watch on the screen, the openning
and closing of collimator
Transport
The equipment does not be moved without taking precautionary measures.
For more information, please contact the representative of APELEM Company.
Ref : 81-40-001-F
6
Surgical C-arm APX HF 3
Installation
The installation of the equipment must be realised only by technicians formed or
agreed by APELEM, in a room without any humidity and dust. Please, refer to the chapter
"Safety" of this manual.
Operating
The conditions of use are the following :
- Ambient temperature : 10 to 40C
- Relative humidity : 30 to 75%
- Atmospheric pressure range : 500 to 1060 hPA
Storage
The conditions of storage are the following :
- Temperature : 0 to 40C
- Relative humidity : 10 to 80%
- Pressure : 700 to 1060 hPA
Ref : 81-40-001-F
7
Technical Manual
-Then carry out the following phases in fluoroscopy (manual scopie mode) :
50 kV and 0.5 mA for 5 minutes followed by a pause of 5 minutes.
70 kV and 2 mA for 5 minutes followed by a pause of 5 minutes.
90 kV and 2 mA for 5 minutes followed by a pause of 15 minutes.
- Slowly, (over the next 5 minutes) increase from 90 to 110 kV then respect a 2 hour
pause before using the machine.
Life time
The life time of the APX HF3 is 10 years. Beyond this time APELEM does not guarantee
technical characteristics of the product.
Ref : 81-40-001-F
8
Surgical C-arm APX HF 3
General description
Wheel steering
ang blocking
Fluoroscopy
foot switch
Console
Ref : 81-40-001-F
9
Technical Manual
Ref : 81-40-001-F
10
Surgical C-arm APX HF 3
Monitors Trolley
Monitors
Video printer
CCD Processor
APX HF 3
Connexion Wire
Locker Wheels
Ref : 81-40-001-F
11
Technical Manual
Characteristics
1 General characteristics
Aparatus class Class 1, Type B
Console / trolley
LC10... / 44 cm
2 monitors 17
LC 8000 et DSA 500 / 44 cm (or 21
2 monitors 17
/51 cm optionial)
2 Radioscopy
kV from 40 to 120 kV by 1 kV-steps
Ref : 81-40-001-F
12
Surgical C-arm APX HF 3
3 Radiography
mA maximum mAs Focal spot
4 Protection devices
Temperature of the X-rays tube assembly temperatures indicators
Operating error
Protection of the frequency inverter
5 Functioning indicators
X-ray emission
Stop
Temperature of the X-rays tube assembly temperatures indicators
6 Other characteristics
Flash radioscopy
Pulsed radioscopy
Anatomical memories in radiography
Ref : 81-40-001-F
13
Technical Manual
1 47
41 39 37 43 36
13 44
2 42 40 38 35 34 33
45
3
30
32
4 31
29 28
10 46
5
11 14 17 19 21 23 25 27
6
12 15
7 8 9
16 18 20 22 24 26
1 Thin patient 25 Lower c-arm
2 Normal patient 26 Radioscopy
3 Big patient 27 Raise c-arm
4 Profile 28 Timer reset
5 Child 29 mA or mAs setting
6 Anatomical memory search 30 Automatic scopy
7 Anatomical memory search 31 Contour enhancement
8 Anatomical memorisation 32 kV setting
9 Anatomical memorisation 33 Radiography mode
10 Storage B monitor 34 Flash radioscopy
11 Image intensifier field 35 Pulsed radioscopy
12 Rotation to the left 36 On-off switch
13 Operating mode 2 (future version) 37 Safety-stop
14 Scan inversions 38 High-energy radioscopy
15 Rotation to the right 39 Housing temperature
16 Less light 40 Low-energy radioscopy
17 Close iris 41 X-ray emission
18 More light 42 Radioscopy mode
19 Open iris 43 On-off indicator light
20 Close shutters 44 Operating mode 1 (future version)
21 Shutter rotation to the right 45 Reprography (optional)
22 Open shutters 46 Memorised image search
23 Shutter rotation to the left 47 Anatomical panel
24 Laser centering (optional)
Ref : 81-40-001-F
14
Surgical C-arm APX HF 3
1 Thin patient
This key is used to select, for a thin patient, the values recorded in the anatomical
memory of high-tension radiography and of the quantity of exposure mAs, for
15 different examination sections, head-on or profile.
2 Normal patient
This key is used to select, for a patient of normal corpulence, the values
recorded in the anatomical memory of high-tension radiography and of the
quantity of exposure mAs, for 15 different examination sections, head-on or
profile.
3 Big patient
This key is used to select, for a big patient, the values recorded in the
anatomical memory of high-tension radiography and of the quantity of exposure
mAs, for 15 different examination sections, head-on or profile.
4 Head-on/Profile
This key is used to select from the anatomical memory the type of examination
required, head-on or profile.
5 Child
This key is used to select, for a child, the values recorded in the anatomical
memory of high-tension radiography and of the quantity of exposure mAs, for
15 different examination sections, head-on or profile.
Ref : 81-40-001-F
15
Technical Manual
Both these keys are used to search for an anatomical memory amongst the
15 observation sections possible
10 Storage B monitor
This key is used to recopy the direct montitor image onto the recopy (B)
monitor.
This key with indicator lights controls the field selection of the
image intensifier when the c-arm is equipped with a bi-field or tri-
field image intensifier.
These keys are used to rotate the image in order to obtain the
best radioscopic observation.
Ref : 81-40-001-F
16
Surgical C-arm APX HF 3
13 Integration x 4
This key allows the camera integration rate selection (x4). (only available on
the compact version)
Ref : 81-40-001-F
17
Technical Manual
26 Fluoroscopy
28 Timer reset
This key is used to reset the timer at 5 minutes. The number of
minutes remaining is shown on the display left of this key.
When the timer reaches zero, fluoroscopy is stopped. The
operator is warned of this in advance by a beeper sounding off
during the last thirty seconds.
29 mA or mAs setting
Both these keys are used to select anodic current.
If the machine is in automatic radioscopy mode, press either
of these keys to invalidate.
The values of high-tension current that can be obtained manually
range from 0.1 mA to 3 mA in low energy mode and from 0.1
mA to 6 mA in high energy mode, by steps of 0.1 mA (keys 40
and 38).
These two keys are equipped with an automatic repetition, i.e. press briefly to obtain the
increase or decrease of the value by one unit, but if either key is pressed down and held
for a longer time, the value will increase or decrease rapidly.
Ref : 81-40-001-F
18
Surgical C-arm APX HF 3
31 Edge enhancement
This key is used to validate the system emphasising the contours of the
image.
32 Manual kV setting
Both these keys are used to set the kV in fluoroscopy or
radiography.
The values of kV that can be obtained manually range from 40
kV to 120 kV, by steps of 1 kV.
These two keys are equipped with an automatic repetition, i.e.
press briefly to obtain the increase or decrease of the value by
one unit, but if either key is pressed down and held for a longer time, the value will increase
or decrease rapidly.
33 Radiography mode
Press this key to switch the machine into radiography mode. There is an
indicator lamp above the key to show if the mode is active or not.
34 Flash fluoroscopy
This key is used to validate or invalidate flash fluoroscopy. There is an
indicator lamp above the key to show if the mode is active or not.
Flash fluoroscopy is used to obtain one x-ray pulse , of short duration but
high intensity.
The image memorised automatically at the end of the pulse is strongly
contrasted.
35 Pulsed fluoroscopy
This key is used to validate or invalidate pulsed fluoroscopy, a technique
used to reduce the amount of X-rays received by the patient. There is an
indicator lamp above the key to show if the mode is active or not.
Pulsed fluoroscopy consists in sending a number of x-ray pulsations at
fixed intervals.
Ref : 81-40-001-F
19
Technical Manual
36 On-off
This key is used to switch the machine on and off, on condition that the lamp
on the left of this key indicating it is plugged in is alight.
Caution : Switch on the machine only after plug on the trolley.
37 Safety-stop
If this indicator is alight, no new operation can be effected. The following is a
list of error codes. In case of a safety-stop, the corresponding error card will
be shown on the kV and mA/mAs displays:
Code -1- : Keyboard error; a key is activated when the machine is switched on.
Code -2- : Radiation error; the x-ray key is activated when the machine is switched on.
Code -4- : Housing temperature error; the temperature of the X-rays tube has reached
the limit of 65C. Stop the machine until the X-rays tube has cooled down.
Code -5- : Heating error; either the x-ray tube filament is cut or there is a faulty
connection.
Code -6- : Frequency error; there is something wrong with the frequency inverter or with
the X-rays tube (high voltage discharge).
Ref : 81-40-001-F
20
Surgical C-arm APX HF 3
38 High-energy fluoroscopy
Press this key to enter the high-energy fluoroscopy mode. There is an
indicator light above showing if this function is activated.
This operating mode is used if in low-energy fluoroscopy the image is
flecked or if the x-rays have difficulty in penetrating the subject.
This technique varies the anodic current from 0.1 mA to 6 mA.
Do not use high-energy fluoroscopy all the time, to avoid the high value of
the current overheating the X-rays tube and the patient receiving a too
high radiation dose.
40 Low-energy fluoroscopy
42 fluoroscopy mode
This key is used to change from the radiography to the fluoroscopy mode.
There is an indicator light above showing if this function is activated.
Ref : 81-40-001-F
21
Technical Manual
44 Integration x 8
This key allows the camera integration rate selection (x 8). (only available
on the compact version).
45 Video reprography
A video printer can deliver a paper copy of the image seen on the television
screen in a few seconds.
If your machine is equipped with this option, press this key to obtain a print.
47 Anatomical panel
Ref : 81-40-001-F
22
Surgical C-arm APX HF 3
Use in fluoroscopy
The image intensifier and the X-rays tube are to be covered with protection bags to
avoid any liquid getting inside.
When switched on, the machine will always start up in fluoroscopy mode.
If in the course of use you find yourself in a different mode, press the fluoroscopy mode
key.
1 Manual fluoroscopy
The manual fluoroscopy mode is obtained when the automatic fluoroscopy indicator
light is off.
If this indicator is alight, just press either the corresponding key or one of the four kV- or
mA-setting keys.
The operator adjusts the high tension and anodic current values using the appropriate keys
so as to obtain a good operating image on the video screen, and thus disactivates the
automatic mode if that was the mode the machine was in. If the mode is changed, the
operator is informed by a short beep sounding off.
If the automatic fluoroscopy mode is exited with the automatic radioscopy key there is no
beep.
2 Automatic fluoroscopy
In the automatic mode, the operator does not need to adjust the high-tension and
anodic current values; this is done by an electronic system which examines the video in a
circular zone in the screen centre and adjusts the fluoroscopy settings to obtain the best
image. There is two Automatic fluoroscopy mode :
low-energy : from 40 to 120 kV and from 0.1 to 3 mA
high-energy : from 40 to 120 kV and from 0.1 to 6 mA
3 Flash radioscopy
The flash fluoroscopy mode is used to obtain a highly contrasted image of the subject
thanks to x-ray emission of short duration but high intensity.
This operating mode can only be used if the subject under observation moves slowly.
Ref : 81-40-001-F
23
Technical Manual
The operator presses the fluoroscopy pedal and an x-ray emission takes place.
Once an image is obtained, it is stocked in the memory and the x-ray emission ceases,
even if the operator continues to press the fluoroscopy command.
After flash fluoroscopy, the machine returns automatically to continuous fluoroscopy mode.
The scan inversion mode required is selected by pressing the corresponding key.
Each time it is pressed one of four different inversion combinations is obtained. Two indicator
lights indicate the activated inversions.
The first memory is connected to the live TV monitor and enables the last image obtained
to be frozen. If this memory is of the continuous integration type, the quality of the image
can be improved by reducing the noise swarming. Memorisation is automatic and takes
place each time the radioscopy control is released ( L.I.H. ).
The second memory is connected to the second monitor and enables to compare
examinations with the first monitor by stocking, in the course of the observation, the different
images obtained.
Two keys on the console control the functioning of this memory: one key is used to transfer
the image from the direct monitor to the copy memory, the other key is used to search for
images.
Both these keys may be activated at any time, during radioscopy or not.
Ref : 81-40-001-F
24
Surgical C-arm APX HF 3
If your machine is equipped with a video printer, a print can be obtained at any moment by
pressing the corresponding key on the keyboard.
7 Fluoroscopy timer
This machine is equipped with a five-minute fluoroscopy timer.
After five minutes of x-ray emission in fluoroscopy mode, the timer must be reseted by
pressing the corresponding key on the keyboard.
If the fluoroscopic timer is not reseted, the fluoroscopy will be stopped and blocked, a
beeper will bring this to your attention if you attempt to start another fluoroscopy by pushing
the pedal.
During fluoroscopy, you are informed that one minute has been spent by a short beep.
When you have only 30 seconds left, a beep will sound every second, and then when the
time is up, a long beep will sound off and the stop light will come on.
Ref : 81-40-001-F
25
Technical Manual
Ref : 81-40-001-F
26
Surgical C-arm APX HF 3
Use in radiography
Enter the radiography mode by pressing the radiography mode key.
The mA/mAs displays show in this mode the quantity of mAs selected for the following
exposure.
In this operating mode the high-tension value and the mAs quantity are selected for the
following exposure.
To recall a memory, simply press the anatomical memory search keys which validate a
zone on the anatomical panel, the head-on or profile indicator and one of the patient sizes.
The high-tension and mAs quantity values previously memorised are then displayed.
The constants of an anatomical memory can be modified by setting the high tension value
or the mAs quantity. If the contents of the anatomical memory differs from the radiographic
constants displayed, the patient size indicator lamp will blink.
The values really in the memory can then be obtained by pressing the key corresponding
to the blinking lamp.
2 Memory search
There is a total of 120 anatomical memories. These memories are divided into 4
groups of patients: normal, thin, child and fat.
The anatomical panel is used to pinpoint the zone to be examined. There is a choice of 15
different zones.
Ref : 81-40-001-F
27
Technical Manual
3 Exposure setting
Take the hand switch and move back of the machine. Press on the first button, keep on
pressing and wait the start of the exposure.
The time between pressing the hand switch and the effective begining of the exposure is
3 secondes (parameter 15)
The second button of the hand switch is not use (no 2sd time fonction)
3 Memorization
To transfer the graphic constants to the anatomical memory, first select an
examination zone, head-on or profile, size of patient. Then display your own constants,
and press the memorisation keys.
Ref : 81-40-001-F
28
Surgical C-arm APX HF 3
Configuration
The use of a micro-controller has led to the suppression of all the settings which had to be
made with adjustable resistors before.
The configuration mode is used to personalise each machine to adapt it to the uses of its
operator. It is used to define the present options and the functioning modes at each start-
up.
When a parameter is modified, the scan inversion indicator lights will blink to remind
the operator that the scan inversion key must be pressed to enter the new value into the
memory.
Precautions
Access into the configuration mode without precaution may affect the proper functioning
of the machine.
Access into the configuration mode, only be done by a qualified Technical Department
personnel
Ref : 81-40-001-F
29
Technical Manual
In the configuration mode, the displays and keys usually used for the adjustment of the
high-voltage, serve to the selection of the parameter, while keys and the displays usually
used for the adjustment of the mAs or the mA, serve to the modification of the parameter
value.
parameter 8 : NU
Ref : 81-40-001-F
30
Surgical C-arm APX HF 3
N.B. The ways in which the different fluoroscopy modes develop are as follows:
Normal radioscopy:
time actions
Ref : 81-40-001-F
31
Technical Manual
N.B. The keys used in the configuration mode have a repetition speed independent
of the programming.
This parameter allows to define the APX type used ( compact or DSA version).
Parameter n 85 dsa_int :
DSA_INT = 0 (integration x 2)
DSA_INT = 1 (integration x 4)
DSA_INT = 2 (integration x 8)
Ref : 81-40-001-F
32
Surgical C-arm APX HF 3
Parameter n 86 ACQ_A :
This parameter allows to define the active level of the ACQ A signal
VERSION
2 0 0 3 inversions on start-up:
0 = no inversion
1 = horizontal inversion
2 = vertical inversion
3 = horizontal and vertical inversion
Ref : 81-40-001-F
33
Technical Manual
HE AUTO
0
1 NU NU
2 X
3 X
4 NU NU
5 X X
Ref : 81-40-001-F
34
Surgical C-arm APX HF 3
Ref : 81-40-001-F
35
Technical Manual
Ref : 81-40-001-F
36
Surgical C-arm APX HF 3
71 0 0 1 0 = normal; 1 = NU
83 0 1 0 = CCD, 1 = CRM86
Ref : 81-40-001-F
37
Technical Manual
Ref : 81-40-001-F
38
Surgical C-arm APX HF 3
Ref : 81-40-001-F
39
Technical Manual
Ref : 81-40-001-F
40
Surgical C-arm APX HF 3
Maintenance
1 How to replace the 63A fuse
Locate the frequency inverter block situated at the bottom left of the generator.
Press the red push-button on its side for at least 30 seconds to discharge the
reservoirs capacitors, in conformity with the instructions on the safety label close to the
fuse.
Unscrew the 2 screws on the fuse cover and change the fuse.
Ref : 81-40-001-F
41
Technical Manual
All necessary measures must be applied for avoiding damage to the pieces, particularly
painted and visible pieces of the finished unit .
These measures include protection and attention to the pieces during their handling, and
during the cleaning of the work station.
During all the mounting the unit must be stored in a place perfectly cleaned.
Reference documents :
Special tooling
Pieces to be prepared :
Preparing of the shift
WARNING :
The setting of the unit must be realised only by qualified technicians.
DANGER :
The laser beam may be dangerous to patient and operator if it is saw in a front view (in its
axis), and may create a burn of the retina.
2 - Place the at 25 cm from the collimator output, in the X-rays axis, the hole of a ruler
situated vertically.
3 - Center the laser centring cross at the middle of the input face of the intensifier.
6 - Engage a fluoroscopy and move the C so as to visualise the leaded cross exactly in
the centre of the ruler hole. This action allows to materialise the X-rays axis(fig.1and 2).
7 - Tighten the arms and check that the ruler is fixed in position. Check the X-rays axis.
8 - Adjust each arm support of the module so that the line generated by each of thus
crosses the ruler in its centre, and also be superimposed with the corresponding line of
the cross designed on the centring disk.
9 - If the centring has been yet realised in the factory, it is enough to set the two supports.
Ref : 81-40-001-F
42
Surgical C-arm APX HF 3
10 - If the centring is made for the first time, you have to displace the two arm supports
and to direct each module in order making the cross orthogonal, and also adjust the
position of each module with the one of its output window. This window has a hole which
allows to limit the dispersion of the laser beam out of the useful image intensifier field.
11- Check again the X-rays axis for compensating an inopportune decentring.
13 - Check again the setting, and applied a varnish for blocking each screw.
When you install the superior cover of the housing, take care of not knocking against the
module support.
X-rays axis
Tube Intensifier
Ruler
Leaded cross
on the disk
Ruler
Ref : 81-40-001-F
43
Technical Manual
Fuses List
Generator
Camera X EYE
V700 BOARD
Moniteur BG2644
M moire LC10XX
Ref : 81-40-001-F
44
Surgical C-arm APX HF 3
Preventive maintenance
The user is requested to observe national laws concerning the use and installation of x-
ray equipment.
Worn parts may become dangerous; they must be checked and replaced by new
ones.
Before using the unit, the operator is to ensure that all safety devices are in working
order. Particular attention is to be paid to indicator and display lamps by using the lamp
test .
The indicator lamp (x-ray emission) on the console is to remain alight only when the
x-ray emission controls are pressed. If it does not go out, disconnect the plug and call the
Technical Department.
The condition of the cables and connections is to be regularly checked; they are to
be replaced if they show signs of wear.
The vertical displacement mechanics of the structure, the wheel movements, the
mechanical brakes and the safety switches are to be checked once a year.
Ref : 81-40-001-F
45
Technical Manual
Monitor
Serial N Type of monitor
1 monitor
Type of power supply 2 monitor s Side by side
Serial Npower supply 2 superposed monitors
100 Hz option
Memory
Serial Nmonitor 1
DSA 500 2 images Serial Nmonitor 2
1 image 10 images
5 images Option text
Serial N
Options
Centrage Laser Serial N
Laser centering
Video rec SVO 9500 MP
Generator Video printer paper
Video printer film 1 view
Serial Nof inverter 2 views
3 views
Serial Nof monobloc
Cassette holder
Grid for cassette holder
Serial Ntube head Steril sheets
Ref : 81-40-001-F
46
Surgical C-arm APX HF 3
Grounding
point
This terminal, link at electric ground of the apparatus allows, with the yellow /green
wire gift, to realise a grounding connection with an other equipment (examples: tables,
monitoring, etc).
Ref : 81-40-001-F
47
Technical Manual
N of the body
notifier for APELEM
products
Protection Type
B Address
Manufacturer
Power supply
Ref : 81-40-001-F
48
Surgical C-arm APX HF 3
Diagrams
1 Generator unit
2 Keyboard 271
Ref : 81-40-001-F
49
Technical Manual
1 Generator unit
REFERENCE DESCRIPTION
Ref : 81-40-001-F
50
Surgical C-arm APX HF 3
Ref : 81-40-001-F
51
Technical Manual
2 Keyboard 271
REFRENCE
81 52 271
81 51 271
Ref : 81-40-001-F
52
Surgical C-arm APX HF 3
Ref : 81-40-001-F
53
Technical Manual
REFERENCE
81 52 270
81 51 270
Ref : 81-40-001-F
54
Surgical C-arm APX HF 3
Ref : 81-40-001-F
55
Technical Manual
REFERENCE
81 52 255
81 51 255
Ref : 81-40-001-F
56
Surgical C-arm APX HF 3
Ref : 81-40-001-F
57
Technical Manual
REFERENCE
81 52 S1C
81 51 S1C
81 52 S2C
81 51 S2C
81 52 S4C
81 51 S4C
Ref : 81-40-001-F
58
Surgical C-arm APX HF 3
Ref : 81-40-001-F
59
Technical Manual
6 Housing assemblies
REFERENCE
81 58 004
81 51 S3C
81 52 S3C
Ref : 81-40-001-F
60
Surgical C-arm APX HF 3
Ref : 81-40-001-F
61
Technical Manual
REFERENCE DESCRIPTION
81 58 001 Schematic
Ref : 81-40-001-F
62
Surgical C-arm APX HF 3
Ref : 81-40-001-F
63
Technical Manual
8 Power supply
REFERENCE
01 57 013
Ref : 81-40-001-F
64
Surgical C-arm APX HF 3
Ref : 81-40-001-F
65
Technical Manual
9 Console interface
REFERENCE
01 51 307
01 52 307
Ref : 81-40-001-F
66
Surgical C-arm APX HF 3
Ref : 81-40-001-F
67
Technical Manual
Ref : 81-40-001-F
68
Surgical C-arm APX HF 3
Ref : 81-40-001-F
69
Technical Manual
Description
The VARIAN x-ray rotating-anode tubes of the A 145 series are equipped with 0.3
and 0.6 mm focal spots.
The anode disc diameter is 71 mm.
Characteristics
Anode inclination 10
Ref : 81-40-001-F
70
Surgical C-arm APX HF 3
Ref : 81-40-001-F
71
Technical Manual
Ref : 81-40-001-F
72
Surgical C-arm APX HF 3
Ref : 81-40-001-F
73
Technical Manual
Ref : 81-40-001-F
74
Surgical C-arm APX HF 3
Ref : 81-40-001-F
75
Technical Manual
Ref : 81-40-001-F
76
Surgical C-arm APX HF 3
Ref : 81-40-001-F
77
Technical Manual
Ref : 81-40-001-F
78
Surgical C-arm APX HF 3
Ref : 81-40-001-F
79
Technical Manual
Ref : 81-40-001-F
80
Surgical C-arm APX HF 3
Ref : 81-40-001-F
81
Technical Manual
81 57 07 TROLLEY
81 57 08 CABLE N2
Ref : 81-40-001-F
82
Surgical C-arm APX HF 3
Ref : 81-40-001-F
83
Technical Manual
Personal notes
Ref : 81-40-001-F
84
Surgical C-arm APX HF 3
Ref : 81-40-001-F