Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Fisher Fieldvue Dvc6200F Digital Valve Controller PST Calibration and Testing Using Valvelink Software

Download as pdf or txt
Download as pdf or txt
You are on page 1of 24

Instruction Manual Supplement DVC6200f Digital Valve Controller

D104217X012 February 2017

Fisher FIELDVUE DVC6200f Digital Valve


Controller PST Calibration and Testing using
ValveLink Software
The test procedure contained in this Instruction Manual Supplement is to be considered as a guideline only and should
be modified to address sitespecific requirements. Use this procedure in conjunction with the DVC6200 Series quick
start guide (D103556X012) and the DVC6200f instruction manual (D103412X012). In addition, exercise good
engineering practices and abide by specific plant safety guidelines for safe operation.
For additional information on Partial Stroke Testing and associated parameters refer to Partial Stroke Test Information
on page 19.

PST Calibration
This document covers the basic PST calibration, as well as details for making adjustments to the normal end, using
Advanced Settings (see figure 11).

Figure 1. Calibration > Partial Stroke

Confirm PST Tier

www.Fisher.com
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Figure 2. Set Transducer Block to Manual

2
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

Figure 3. Enter the Desire Outgoing Ramp Rate

Supports separate ramp


rates for outing and
incoming strokes

Figure 4. Enter the Desired Incoming Ramp Rate

Additional ramp
rates for the
incoming stroke

3
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Figure 5. Enter the Minimum Travel Movement

Figure 6. Partial Stroke Calibration progress

4
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

Figure 7. Partial Stroke Calibration progress

Will calibrate to 30% from the normal


end if the Minimum Travel Movement +
Set Point Over Drive is less than 30%

If it is greater than 30% the user-defined


point will be calibrated

Figure 8. Step or Ramp to Normal Position on Failed PST

5
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Figure 9. Partial Stroke Calibration Progress

Figure 10. Calibration Procedure Complete

List of parameter
values being
downloaded to
the device

If you need to make adjustments to the normal end default settings, select Advanced Settings, as shown in figure 11,
and make the necessary adjustments.

6
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

Figure 11. Select Advanced Settings to Make Adjustments to the Normal End

Select to define the


normal end

Figure 12. Set Travel High End

Defines the
end of travel

7
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Figure 13. Enter Minimum Travel Movement

If Minimum Travel Movement + Set Point Overdrive is more than


Maximum Allowable Travel then the below message is presented
and you are directed to the next screen as shown in figure 14

8
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

Figure 14. Review Inputs

Minimum Travel Movement +


PST Overdrive Amount
is greater than the
Maximum Allowable Travel

Figure 15. Enter New Maximum Allowable Travel

Enter a value greater


than Minimum Travel
Movement + PST
Overdrive Amount

9
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Figure 16. Partial Stroke Calibration Progress

Configuration
After successfully calibrating the valve for PST go to the Detailed Setup FST/ PST tab and verify that the values for the
parameters in the Valve Stroke Test group box are correct. Then go to the Partial Stroke group box and enable any of
the behaviors required for the PST.

Figure 17. Verify FST/PST Values

10
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

Once the PST has been setup and calibrated go to the VST Abnormal Criteria group box and select the criteria to be
used to evaluate PST after they are run. Then, select the criteria to be used to abort a PST immediately on initiation,
and the criteria to be used to prohibit a PST before initiation.

Figure 18. Select VST and PST Criteria

11
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Categorize the stroke alerts from the Valve Stroke Alerts tab into one of the Field Diagnostic Alert categories and
suppress them if publishing on the segment is not desired.

Figure 19. Set Valve Stroke Alerts

12
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

Initiating a PST Diagnostic


If an Auto PST is desired then go to the Auto PST tab and set up the PST schedule. The instrument will present a
message when the next PST becomes due. If a scheduled PST is not initiated then a PST overdue alert is generated.

Figure 20. Setting Auto PST

13
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

To run a manual PST select the Partial Stroke Test icon as shown in figure 21.

Figure 21. Initiating a Manual PST

Figure 22. Upload or Delete Saved Partial Stroke Diagnostics

Before the manual PST initiates, you are given the


opportunity to upload and save the instrument
diagnostic data to ValveLink software. Select the
datasets that are Not In Database and upload. Deletion of
the data from the instrument is not necessary as new
data will overwrite the oldest dataset. Datasets that are
not uploaded to ValveLink software will not be accessible
in the instrument once they are overwritten by new data

14
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

Figure 23. Partial Stroke Test Progress

Outgoing Stroke

Incoming Stroke

Figure 24. PST Analyzed Data

PST Result

15
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Figure 25. Resulting Press/Tvl Graph Results

Minimum PST Pause Time Enabled PST starts


incoming stroke on reaching the Minimum
Travel Movement (10% in this case)

Figure 26. Resulting Tvl/Time Graph Results

Select Mark Golden to set dataset as a


reference for future comparison. The
golden dataset will be stored in the
instrument and will not overwritten
by subsequent datasets

Minimum Pause Time


enabled with return lead

16
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

A PST Diagnostic can be run with Minimum PST Pause Time disabled, as shown in the figure below. When Minimum
PST Pause Time is disabled the pause time will be in effect when the valve reaches the desired test point, resulting in a
slower PST.

Figure 27. Disable Minimum PST Pause Time in FST/PST Partial Stroke

Minimum PST Pause


Time disabled

17
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Figure 28. Resulting Press/Tvl Graph Results with Minimum Pause Time Disabled

PST travels to Minimum PST Travel + Set Point Overdrive


(20% in this case)

Figure 29. Resulting Tvl/Time Graph Results with Minimum Pause Time Disabled

PST Pause Time

18
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

Figure 30. PST Analyzed Data with Minimum Pause Time Disabled

Select a PST style, either with or without Minimize PST Pause Time, and standardize on that style, as the data may be
difficult to compare between the two styles. If the desire is to minimize the amount of time the valve is away from the
normal end, then enabling Minimize PST Pause Time is recommended. If the amount of time away from the normal
end is not a concern, then disabling Minimize PST Pause Time will cause the set point to pause at the end of the
outgoing stroke for the travel to catchup to the set point. The results of the test with the Minimize PST Pause Time
disabled will be similar to the PST as offered in earlier versions of SIS instruments.

Partial Stroke Test Information


Valve Stroke Test

A valve stroke test is the process of taking the valve from the normal end to another target position at a preconfigured
ramp rate before returning to the normal end while gathering data. The data is analyzed to evaluate the condition of
the valve assembly against a set of user defined thresholds. A valve stroke test is only run if everything is normal in the
instrument. A safety demand signal will always take precedence over a valve stroke test.

DValve Stroke Test, select Partial Stroke Test, Full Stroke Test, or Disable to select the test to run when the test is
initiated using the VST_COMMAND parameter.

DPartial Stroke Start Point defines the normal end of the valve. The valve needs to be at this end for a PST to be
initiated. When a FST is initiated the valve will be moved by the test to this end before being ramped to the opposite
end and ramped back. Setting this value to Not Configured will disable partial stroke tests.

19
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

DTravel Open End defines, in percent (%) of calibrated travel, the point above which the valve is considered to have
reached the high end.

DTravel Closed End defines, in percent (%) of calibrated travel, the point below which the valve is considered to have
reached the low end.

DTest Pause Time is the time between the outgoing and incoming strokes of the test. The default value is 5 seconds.
Pause Time will not be used if Minimum PST Pause Time is enabled. The outgoing stroke is from the normal end to
the PST target and the incoming stroke is the return stroke to normal. See figure 31.

DVST High Friction Breakout Pressure indicates that the breakout required a higher force than configured by the user.
Refer to figure 31.

DVST Low Friction Breakout Pressure indicates that the breakout required a lower force than configured by the user.
Refer to figure 31.

DAction On a Failed Test defines if the valve should step or ramp back on a failed stroke test.

Figure 31. Valve Signature Representation


PST START POINT
j
k
INCOMING STROKE
l
m
OUTGOING STROKE

n
PRESSURE

RANDOMIZATION

TRIPPED o NORMAL

TRAVEL
j SUPPLY PRESSURE
k INCOMING PRESSURE THRESHOLD
l LOW FRICTION BREAKOUT PRESSURE THRESHOLD
m HIGH FRICTION BREAKOUT PRESSURE THRESHOLD
n OUTGOING PRESSURE THRESHOLD
o TARGET TRAVEL MOVEMENT  100%

VST Abnormal & Abort Criteria

DVST Abnormal Criteria

A partial stroke test is marked as abnormal if it fails one of the following criteria.

The device always evaluates a PST on the following criteria:


1. Target Travel achieved

20
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

2. Return to the normal end


In addition to the above, any of the following can be selected to evaluate a Partial Stroke Test.
1. Breakout Time
2. Outgoing Pressure Threshold
3. Incoming Pressure Threshold
4. High Friction Breakout Pressure
5. Low Friction Breakout Pressure
DVST Abort Criteria

The PST is terminated and the valve is returned to the normal end. The return to the normal end will be per the user
configuration for an aborted test. The abort criteria will only be active if it is added as a criteria to be evaluated during
PST by adding it to the PST Abnormal Criteria.

The device always aborts a PST if the Max Travel displacement is exceeded.

In addition to the above, any of the following can be selected to abort a Partial Stroke Test:
1. Breakout Time
2. Incoming Pressure Threshold
3. High Friction Breakout Pressure
Partial & Full Stroke

DPartial Stroke

PST Max Travel defines how much travel displacement is allowed before the PST aborts (see figure 32).

PST Minimum Travel is the percentage of total span that the valve moves away from its normal operating end of travel
towards its tripped end of travel during the test. The default value is 10%.

Set Point Overdrive defines the extent of the set point overdrive over the Minimum Travel Movement when the early
turn around is enabled. When the early turn around is not enabled it defines the travel target.

Freeze Analog / Discrete Feedback when enabled, freezes the corresponding feedback during a partial stroke test.

Minimum PST Pause Time, when enabled, the incoming stroke is initiated as soon as the travel reaches the minimum
travel movement. Refer to figure 32 for a time series representation of this parameter.

Randomized PST, when enabled the instrument randomizes the target travel, for each PST.

PST Randomization is defined in percent (%) of calibrated travel span, it defines the extent of randomization from the
minimum travel movement towards the normal end. If the user defined randomization is too large the instrument will
cap the max randomization to ensure that there will be at least 1% travel movement away from the defined normal
end. Refer to figure 31.

21
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Figure 32. Time Series Representation of Minimum PST Pause Time

TRAVEL
TRAVEL

NORMAL NORMAL
INCOMING
OUTGOING RAMP RATE
RAMP RATE
REDUCED
PST TIME

RETURN
LEAD
j
j RETURN
SETPOINT LEAD
SETPOINT OVERDRIVE
OVERDRIVE k

TIME BREAKOUT EARLY TIME


PAUSE TIME TIMEOUT TURNAROUND

MINIMUM PST PAUSE TIME MINIMUM PST PAUSE TIME


DISABLED ENABLED

j MINIMUM TRAVEL MOVEMENT


k MAX. ALLOWABLE TRAVEL

Outgoing Ramp Rate is the rate at which the valve will move during the Outgoing stroke of the Partial Stroke test. The
default value is 0.25%/second.

Incoming Ramp Rate is the rate at which the valve will move during the Incoming stroke of the Partial Stroke test. The
default value is 0.25%/second.

PST Return Lead defines the percent (%) change in setpoint to overcome the hysteresis in the valve assembly. The error
between setpoint and actual error is added to this percent change. For example, if the Return Lead is set at 0.5% and
there is a 1% error this will be set at 1.5%

PST Breakout Timeout is the user configured amount of time before which the valve must leave the normal end during
a PST.

VST Outgoing Pressure Threshold defines the actuator pressure at which a partial stroke test will abort during the
outgoing stroke (see figure 31). This prevents the DVC6200f from exhausting (or building) excessive pressure from/to
the actuator in an attempt to move a stuck valve. During PST Calibration, the Partial Stroke Outgoing Pressure
Threshold will be set automatically.

VST Incoming Pressure Threshold defines the actuator pressure at which a partial stroke test will abort during the
incoming stroke (see figure 31). This prevents the DVC6200f from exhausting (or building) excessive pressure from /
to the actuator in an attempt to move a stuck valve.

22
Instruction Manual Supplement DVC6200f Digital Valve Controller
D104217X012 February 2017

DFull Stroke

Full Stroke Ramp Rate is the rate at which the valve will move during the full stroke test.

FST Wait Time is the amount of time to wait for the valve to move to the normal end after initiation of the full stroke
test.

Full Stroke Breakout Timeout is the user configured amount of time before which the valve must leave the normal end
during a full stroke test.

PST Prohibited
A partial stroke test will not be initiated if any of the following user-configurable conditions are active:
1. Check Bit Alert
2. Drive Current
3. Drive Signal
4. Processor Impaired
5. Travel Sensor
6. Output Pressure sensor
7. Supply Pressure Sensor
8. Temperature Sensor
9. Supply Pressure
10. Temperature Limit
11. Travel Deviation
12. Pressure Fallback
13. PST Abnormal

23
DVC6200f Digital Valve Controller Instruction Manual Supplement
February 2017 D104217X012

Neither Emerson, Emerson Automation Solutions, nor any of their affiliated entities assumes responsibility for the selection, use or maintenance
of any product. Responsibility for proper selection, use, and maintenance of any product remains solely with the purchaser and end user.
Fisher, FIELDVUE, and ValveLink are marks owned by one of the companies in the Emerson Automation Solutions business unit of Emerson Electric Co.
Emerson Automation Solutions, Emerson, and the Emerson logo are trademarks and service marks of Emerson Electric Co. All other marks are the property
of their respective owners.
The contents of this publication are presented for informational purposes only, and while every effort has been made to ensure their accuracy, they are not
to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are
governed by our terms and conditions, which are available upon request. We reserve the right to modify or improve the designs or specifications of such
products at any time without notice.
Emerson Automation Solutions
Marshalltown, Iowa 50158 USA
Sorocaba, 18087 Brazil
Cernay, 68700 France
Dubai, United Arab Emirates
Singapore 128461 Singapore
www.Fisher.com

24
E 2017 Fisher Controls International LLC. All rights reserved.

You might also like