1083ch8 2 PDF
1083ch8 2 PDF
1083ch8 2 PDF
Sampling
Costs: Compressed air filters cost about $60; stainless steel bypass filters, $300 to $600;
continuous-flow ultrasonic homogenizers, $3000; sludge centrifuge, $600; interval
sampling pump only, $750; automatic liquid sampler system, $1500 and up; complete
sampling system for single-process gas stream, $3000 to $6000; complete sampling
system for single-process liquid stream, $3500 to $7000. The per-stream cost in a
multistream sampling system drops as the number of streams increase.
1170
2003 by Bla Liptk
8.2 Analyzer Sampling: Process Samples 1171
TABLE 8.2a
Sample Analyzer Data Form
Stream name or identification _________________________________________________________________________
STREAM COMPOSITION DATA
CONCENTRATION RANGE OF COMPONENT
COMPONENT IN MOL%, WT%, PPM TO BE MEASURED
OPERATING PROCESS DATA
Temperature _____________________________ Pressure __________________________________________________
Phase: Liquid _____________________________ Vapor ___________________________________________________
Corrosive components/solids __________________________________________________________________________
Stability (polymerize, decompose, etc.) _________________________________________________________________
Sample bubble point___________________________ Dew point ____________________________________________
SAMPLE CONDITIONS
Maximum distance: Tap to analyzer _____________________ Analyzer to return ______________________________
Speed loop required: Yes _______________________________ No __________________________________________
Sample return pressure point _________________________________________________________________________
Sample probe requirements: Connection size ______________________ Orientation ____________________________
Materials of construction: Stainless steel ________ Teflon ________ Viton ________ Glass ________ Other _________
Electrical areas classification _________________________________________________________________________
Power supply ______________________________________________________________________________________
Output signal ______________________________________________________________________________________
Utilities available: Stream __________________ Air ___________________ Cooling water_______________________
Process Vessel
or Pipe Wall Gate Valve
Pipe-to-Tubing
Connection
To
Sampling
System
Tubing
FIG. 8.2b
Sample probe assembly with process shutoff value.
Sample Takeoff Point practice to install a sampling probe (Figure 8.2b) for most
applications as a precautionary measure to prevent particu-
Sample conditioning begins with the location of a suitable lates from entering the sample transport system. Sampling
sample takeoff point. To obtain a representative sample, the processes that are still reacting chemically or pyrolysis gases
takeoff is usually located at the side of a process line, espe- may require reaction quenching, or fractionation, at the sam-
cially in the case of liquid samples where there is the possi- ple takeoff. This is done by cooling or back-flushing with an
bility of vapor on the top of a horizontal line and dirt or inert gas or liquid to keep the sample takeoff clean and
solids on the bottom of the line. For sampling vapors, the reliably active (Figure 8.2c), while drawing off a reproduc-
connection may be located in the side or top of the process ible sample for analysis.
line, but in both cases with due consideration to accessibility With the advent of in situ analyzer detectors, the sample
for maintenance. takeoff becomes the point of analysis and, for locating
Ideally, the sample at the appropriately selected takeoff in situ analyzers, the above considerations must be carefully
point will require little or no conditioning; however, it is good evaluated.
Sample to
Analyzer Analyzer Vent
Conditioning or
Coalescer System
Drain
Single-Line Transport
Vortex
Cooling
Vent
Tube Analyzer
Temperature or
System
Controller Cooling Drain
Section
By-Pass Stream
Instrument Vent, Drain
Warm
Air Supply Sample Return
Coolant
Exhaust Point or Sample
Analyzer Recovery System
Reflux
Action Filter System
Reflux
Section By-Pass Return Fast Loop
Sample Condensibles FIG. 8.2d
Sample transport methods.
TABLE 8.2e
Dimensions and Volumes of Tubing and Pipe Used in Sample Systems
Nominal Internal Area
Volume per Inner
2
Diameter, in. Diameter, in. in. cc
1
316 stainless steel tubing /8 .0787 .0048 .9571
1
/4 0.1850 0.0268 5.3035
3
3 /8 /8 0.0253 0.0684 13.4417
1
/2 0.4055 0.1290 25.2984
1
Schedule 40 pipe /4 0.3642 0.1040 20.4521
3
/8 0.4921 0.1891 37.4904
1
/2 0.6220 0.3038 59.7408
3
/4 0.8268 0.5363 106.6800
Storage of the test sample may be a consideration, especially The successful design of sample systems requires careful
for unstable liquids or gases with low dew points. Treatment analysis of the physical and chemical conditions of the
of the containers is very important for trace analysis samples. stream, as well as serious consideration of the ambient and
In all cases, the calibration provisions must be incorporated transport conditions, to ensure integrity of the sample. There-
into the systems and a sample provided that is compatible with fore, care should be taken in evaluating the above consider-
the desired stream composition and suitable for analysis by the ations with respect to a given stream and in applying the
analyzer used for the system. It is desirable, but not essential, correct analyzer to provide the desired measurement.
that the calibration sample be introduced automatically from Sample system design is based on experience, and when-
a remote location so that the instrument can be periodically ever possible, previous experience should be given prime
checked; however, in most systems the introduction of the consideration in the selection and application of components.
sample is done by manual switching at the analyzer. A successful sample system normally results in a successful
analyzer system. Therefore, no effort must be spared in
proper sample systems design, which requires a careful selec-
COMPONENTS OF SAMPLING SYSTEMS tion of the system components. These are discussed below,
starting with filters.
A number of devices are discussed here that are components
of analyzer sampling systems. These devices include gas and Filter Designs
liquid filters, bypass filters, liquid homogenizers, liquid grab
sample collectors, chemical reactor sampling systems, and Most analyzer sampling systems will include a filter with at
solids samplers. Later, under Applications, the various sam- least one wire mesh strainer (100 mesh or finer) to remove
pling probe packages and single-stream and multistream pro- larger particles that might cause plugging. Available filter
cess sampling systems are described. materials include cellulose, which should only be considered
The more components there are in a sampling system, in applications where it does not absorb components of inter-
the less reliable it is likely to be. The mean time between est. Sintered metallic filters can remove particles as fine as
failure and maintenance needs of the overall system will 2 microns; cellulose filters can remove down to 3 microns;
improve as the number of pumps, ejectors, regulating valves, and ceramic or porous metallic elements can trap particles
coolers, heaters, filters, coalescers, dryers, knockout traps, of 13 microns or larger. When the solids content is high, two
manifolds, timers, and other components that comprise the filters can be installed in parallel, with isolation valves on
system are reduced. each. Motorized self-cleaning filters are also available for
such services.
Selecting the System Components
Sampling systems require certain components, which are Separating Liquids from Gases Glass microfiber filter tubes
commercially available. One source is the analyzers manu- efficiently separate suspended liquids from gases. The filter
facturer, who through the years has developed systems com- tubes capture the fine droplets suspended in the gas and cause
patible with its analyzer, such as filter coalescers, condensers, the droplets to run together to form large drops within the
and washing and treating systems. depth of the filter tube. The large droplets are then forced by
A second source is the analyzer systems vendor, who had the gas flow to the downstream surface of the filter tube, from
designed special components, such as kinetic separators, fil- which the liquid drains by gravity. This process is called
1
ter probes, and the like, for use with analyzers for applications coalescing.
in rather hard service. The coalesced liquid drains from the tubes at the same
A third source is the specialty vendor, who has developed rate as liquid droplets enter the tubes. Therefore, the tubes
unique sampling components, such as pyrolysis gas sample have an unlimited life when coalescing liquids from relatively
conditioners; permeation devices for water removal systems; clean gases. The filters operate at their initial retention effi-
high-efficiency, self-cleaning filters; and so on. It is desirable ciency even when wet with liquid. The flow direction is inside
to check whether specialty items are available before trying to outside to permit the liquid to drip from the outside of the
to design new components. Most of the specialty components filter to the housing drain (Figure 8.2g).
have taken years of field testing to develop and modify for The filter tube grade should be selected for maximum
successful application. liquid drainage rate, rather than maximum filtration efficiency
When a large number of analyzers are used, the compo- rating, because a liquid drainage rate decreases with increas-
nents selected must be of the same type and manufacture for ing filtration efficiency rating. If liquid is carried into the
interchangeability and stocking of spare parts. Documenting filter in slugs rather than dispersed as droplets in the gas, a
a sample system with complete flow schematics, part identi- filter that is properly sized for steady-state conditions can be
fication, and manufacturing of various components is an flooded and permit liquid carryover. If slugging of liquid is
essential part of being able to properly start up a system and expected, a filter with a relatively large bowl should be selected
maintain it successfully over a long period. to provide adequate liquid-holding capacity, and provisions
Gas Out
Gas Out
Out In
Filtering Media
Metal
Perforated Cylinder
Gas In Gas In
Packed Tower
Sparger
Scrubber
Liquid Drain
FIG. 8.2i
FIG. 8.2g Removal of corrosive gases or of condensable vapors.
2
Clean
High Pressure PI PCV
Chlorene Gas
Polymerization To Analyzer
Fiber Product in Sampling
Liquid Phase System
Dirty
Chlorene
Gas
Screens Jacket
Liquid
Seal Overflow Molten
Pot Polymer
Automatic Low
Gas Vent Filter Flowrate
High Flowrate
Inlet
Pressure
Reducing Analyzer
Liquid Out Process Valve
Stream
Bypass
Filtered Fluid
Filtering Element
(Can be Porous
Metal with 400 Quick
Mesh Size) Disconnect
Sample to
By-Pass
Analyzer
Filter
Restriction
Orifice Nitrogen FIG. 8.2r
or Slide or Steam Rotary disc filter.
Gate Valve Blow-Back
Backflush
Vibration
Filtrate Horn
Continuous-
Flow Cell
Signal
Wire Overflow
2"
(50 mm)
Orifice
Outlet
FIG. 8.2t
Ultrasonic homogenizer. (Courtesy of Cole-Parmer Instrument Co.)
Standpipe
Inserted
Into Process might otherwise plug the system, they also remove process
constituents and make the sample less representative.
The newer approach is to eliminate the potential for plug-
ging by reducing the size of solid particles (homogenization)
while maintaining the integrity of the sample. Thus, when a
pulverizer is used to replace the filter, the sample becomes
representative.
Homogenizers serve to disperse, disintegrate, and reduce
10 Porosity the particle size of solids and thereby reduce agglomerates
Sintered and liquefy the sample. Homogenizers can be mechanical,
Electrode
Metal Filter
Cup
using rotor-stator-type disintegrator heads. In this design, the
rotor acts as a centrifugal pump, which is recirculating the
slurry while the shear, impact collision, and cavitation at the
Filtrate disintegrator head provide homogenization.
in Cup
In ultrasonic homogenizers, high-frequency mechanical
vibration is introduced into the probe (horn), which creates
1" pressure waves as it vibrates in front of an orifice (Figure 8.2t).
(25 mm)
As the horn moves away, it creates large numbers of micro-
FIG. 8.2s scopic bubbles (cavities), and when it moves forward, these
Electrode is protected from material buildup by back-flushed porous bubbles implode, producing powerful shearing action and agi-
filter cup. (Courtesy of TBI-Bailey Controls.) tation due to cavitation.
Such homogenizers are available with continuous-flow
When the flow direction is reversed, the filtrate flows back cells for flow rates up to 4 g/h (16 l/h) and can homogenize
into the filter cup and out through the porous filter. During liquids to less than 0.1-micron particles sizes. The flow cell
back-flush, the buildup is quickly removed and another fil- is made of stainless steel and can operate at sample pressures
tering cycle is initiated automatically. of up to 100 psig (7 bars).
A frequent problem of sampling systems is plugging. There are The extracted sample begins conditioning at the takeoff point,
two ways to eliminate it. The older, more traditional approach continues through the transport, and finishes conditioning at the
is filtering. Unfortunately, as the filters remove materials that analyzer location prior to entering the analyzer (Figure 8.2u).
FIG. 8.2u
Typical sample conditioning system with remote preconditioning unit.
Analyzer
Cooling Water
Sample In Out Pressure Temperature
Gauge Flow
Out In Gauge
FI Indicator
PI T1 With
Control Flow Needle
Valve FI Indicator Valve
Self Calibration
Cleaning Sample
Filter
Sample Shut-Off
Cooler Coalescer Valves
Pressure
Shut-Off Regulator Lab Sample
Valves Pressure Take Off
Flow Relief Valve
Flow Indicator Pressure
FI FI Gauge
Indicator Check with
Valve Needle PI
Valve Check
Valve
Shut-Off
To Valve
Drain
FIG. 8.2v
Typical liquid product sample system for refinery applications.
All samples require some form of conditioning to make them of the components in the liquid used to wash the sample. The
suitable for the analyzer and to assure reliable on-stream oper- conditions of flow, temperatures, and pressures must be con-
ation. The conditioning is done at the appropriate location in trolled to maintain a relatively constant predetermined rela-
the sample system loop in order to maintain the integrity of the tionship of the composition. When washing, care must be
sample (Figure 8.2v). taken to keep the sample in the vapor phase by providing
Sample washing is usually limited to dirty, particle-laden heated transport lines or by making provisions for final mois-
streams whose composition will not be affected by the solubility ture removal as the analyzer may require.
1/4 NPT
Temperature Vapor Outlet
Adjusting Screw O Ring Seal
Temperature
Sensor
AC - IN
Heater
1/8 NPT
Liquid
Sample Inlet
FIG. 8.2w
Vaporizing regulator assembly (electrically heated).
The sample tap should be located on the side of the Condensate Reservoir
Stack Gas Temperature
(Optional)
pipeline to minimize liquid or dirt entrainment, and the sam- to 900 F
Valve Normally Open
ple should be taken from the center of the pipe. If it is
necessary to periodically remove the probe for cleaning or
to perform a sampling traverse, it is desirable that the probe FIG. 8.2aa
be inserted through a packing gland and block valve, as Stack gas sampling.
shown in Figure 8.2y.
Stack Gas Sampling
The volume of the sampling system should be kept to a
minimum, while the velocities through the sample lines should When sampling hot, wet stack gas, a filter capable of with-
be high to protect against the settling of entrained liquids or standing the gas temperature should be installed in the stack
particulates that can cause plugging. Sample line tubing can at the tip of the sample line to prevent solids from entering
be small as 1/8 in. (3 mm) in diameter, and sample flow the gas sample line. After the sample is cooled, a coalescing
velocities should be between 5 and 10 ft/sec (1.5 to 3 m/sec). filter is used to remove suspended liquids before the sample
When the sample is taken from a combustion zone or other goes to the analyzer (Figure 8.2aa).
dirty processes, a filter is usually provided at the tip of the The sample flow direction is from inside to outside. Filter
sample probe, and a high-pressure filter blow-back line is housings with Pyrex glass bowls are often used in this appli-
provided for periodic filter cleaning (Figure 8.2z). cation to permit visual check of the liquid level in the filter
housing. Since there is often a considerable amount of liquid
Small Bead Around End of Probe to Prevent present in the sample, steps must be taken to drain the housing
Pulling the Probe Completely Through the to ensure that liquid does not build up and carry downstream
Packing Gland to the analyzer.
The liquid coalescing filter should be located as close to
the analyzer as possible to minimize the chance of condensa-
Probe tion between the filter and the analyzer. Additional precautions
that can be taken to avoid downstream condensation are to
cool the sample below ambient temperature upstream of the
Packing Gland coalescing filter and to heat the line gently between the filter
Gate Valve and the analyzer.
Condensate
Pressure Stainless Steel Filter
Regulators Housing, 800 psig (56 barg)
Rating, With 25 Micron
Samples to Filter Tube
Conductivity Cells
Slipstream
FIG. 8.2dd
High-pressure stream sampling.
Sampler
FIG. 8.2ee
4
Continuous or intermittent sampling of chemical reactor.
The design illustrated in Figure 8.2ee requires only one installed below the process liquid (or sludge) level and traps
nozzle (3 in. or 75 mm) and utilizes a Teflon-coated sampler some air at the top of the chamber as it fills with the process
assembly that can be used up to 150 psig (10.5 bars) and fluid. When a sample is required, compressed air is intro-
350F (177C). For continuous sampling applications, such duced, which closes the Duckbill inlet and discharges the
as for closed-loop pH control, a Teflon diaphragm pump is sample from the bottom of the chamber. When a new sample
used to continuously return the analyzed sample. is to be drawn into the sample chamber, the compressed air
is vented and a new fill cycle is initiated.
Duckbill Samplers This sampler should be considered An automatic controller is provided on which the user
when liquid or sludge samples are to be collected at remote can adjust the frequency at which samples are to be taken
locations, from below the level in tanks, sewers, channels, into a composite sample collection bottle.
5
sumps, lakes, or rivers. As shown in Figure 8.2ff, this device
has no moving mechanical components, only a rubber (EPT,
Solids Sampling
Buna-N, or Viton) bucket-shaped Duckbill , which is inside
a housing made of polyvinyl chloride (PVC), aluminum, or When solids in bins or silos need to be sampled, one of the
stainless steel. This rubber insert closes around fibers or various solids feeders (described in Section 2.23) can be con-
particulate matter without jamming. It is operated by com- sidered. When solids are to be sampled while flowing by
pressed air. gravity or while pneumatically conveyed under the pressure,
The sample enters by gravity through the bottom con- the choice of sampling devices becomes more limited. The
nected Duckbill inside the sampling chamber, which is screw-type solids sampler can be used on these applications
Process
Stream
Motor
Sample
FIG. 8.2gg
Archimedian screw solid sampler.
Analyzer
Sample Return
Actuator #1
Actuator #2
FIG. 8.2hh
The tray-type solids sampler.
4. Tubing sizes are critical, especially where low flows are analyzer. However, because this system is more complex than
used, because they limit the amount of increase in the single-stream sample conditioning, the following consider-
adsorptiondesorption phenomenon. A rule of thumb ations should be reviewed to determine if multistreaming is
is to use the smallest possible tubing to achieve maxi- feasible:
mum flow to accommodate the sample loop design.
5. Packless shutoff valves with a diaphragm or bellows 1. The potential problem of cross-contamination among
seals should be used; however, due to their high cost, multiple streams
serious consideration must be given to this area, and 2. The importance of each analysis and frequency of
conventional valves are suitable for most applications. analysis
6. Filtration of the sample in ppm analysis can create 3. The loss of information from more than one analysis
significant problems unless the filter is totally inert. in case of analyzer failure
Therefore, stainless steel filters using a high flow rate 4. The cost of an additional analyzer vs. the cost of
or dense fluorocarbon inert materials are recommended multistreaming
for such applications. 5. Maintenance requirements
7. Conditioning of the lines and system to accommodate
the ambient temperature requirements for preventing After reviewing the above, one can decide if multistream-
condensation of components of interest in the sample ing is feasible and whether it should be manual or automatic.
must be considered; if necessary, heat tracing of the Whether manual or automatic, multistreaming requires good-
lines must be furnished. quality valves for stream switching.
8. When it is necessary to provide an aid for transporting A typical multistream sampling system is shown in
the sample from its sample point to the sample system Figure 8.2ii. A common and important requirement in all
through the analyzer, as is frequently done in ambient such systems is that a continuous bypass be provided for each
monitoring systems, a sample pump, ejector, or aspirator sampling point to avoid dead-end sample lines. The sample
is necessary. In such cases, the pump must be a dia- system should be laid out in such a manner that contamina-
phragm. If practical, an ejector to pressurize the sample tion between streams is avoided. This is best accomplished
or an aspirator to aspirate it through the measuring device by arranging the solenoid valve in a double-block double-
can be used. Both are more desirable than a sample pump. bleed configuration, which is rather expensive.
More often, a three-way solenoid valve is used for each
Multistream Switching Multistream switching is usually stream, with the venting port always at low pressures to create
used when it is practical to analyze several streams using one a relief in case of a leak. To prevent contamination, dead
Instrument
PI Air
Field Analyzer House
Supply
F
TI
Liquid PI Steam
PI
Sample
200 PSIG
F
70F Drain
(1378 kPa, 21C) TC
Steam FI
Drain To
Heated Enclosure Steam Atmosphere Return to Header FI A
Mounted at Traced Vent
Sample Tap
FI
Saturated PI ATM
Vapor Bleed
Sample F
4 PSIG Calibration
120F Sample
(27.6 kpa,
49C) Drain Steam Heated Enclosure
Bubbler for
Leak Detection
FIG. 8.2ii
Two-stream sampling system for trace analysis with a double-block, double-bleed configuration.
PI
Stream 1 FI Flow
Indicator
with
FI
Adjusting
PI Valve
Stream 2 FI FI Analyzer
Pressure Bleed
Relief
Gauge Orifice
Valve
PI Shut-Off
Valves Calibration Gas
Stream 3 FI
Pressure
Regulator Flow
3-Way
Indicator
Solenoid
with
Adjusting
Valve
FIG. 8.2jj
Typical multistream automatic sampling system.
volume of the sample system should be considered, as well Converse, J. G., Calibration & Maintenance Are Part of a Reliable Sample
as equalization of the pressures upstream of the three-way Preparation Systems Design, ISA Conference, Houston, October 1992.
Cornish, D. C. et al., Sampling Systems for Process Analyzers, London:
valves. The problem is more severe in ppm sampling systems Butterworths, 1981.
because of the adsorptiondesorption effects, and careful Dubois, R. et al., The New Sampling Initiative, 47th Annual ISA Analysis
consideration should be given to the design criteria described Division Symposium, April 2002.
above. Dubois, R., van Vuuren, P., and Tatera, J., New Sampling Sensor Initiative,
An Enabling Technology, 47th Annual ISA Analysis Division Sympo-
sium, Denver, April 1418, 2002.
References Erk, G. F., Engineering Analyzer Systems, InTech, August 1979.
Foundos, A. P., Reliable Sample Conditioning Improves Process Analysis,
InTech, January 1990.
1. Baker, G., Erk, G., Hudelson, J., Manka, D., Siebert, K., and Wachel,
Fussell, E., Molding the Future of Process Analytical Sampling, InTech,
L., Automated Stream Analysis for Process Control, Vol. 2, Orlando, FL:
August 2001, 32.
Academic Press, 1984, pp. 97101, 105106.
Houser, E. A., Principles of Sample Handling and Sampling Systems Design
2. American Petroleum Institute, API RP 550, Part II, Process Stream
for Process Analysis, Pittsburgh, PA: Instrument Society of America, 1977.
Analyzers.
Lodge, J. P., Methods of Air Sampling and Analysis, Chelsea, MI: Lewis
3. Nichols, J. H. and Brink, J. A., Jr., Use of Fiber Mist Eliminators in
Publishers, 1988.
Chlorine Plants, Electrochemical Technology, JulyAugust 1964.
McMahon, T. K., The New Sampling/Sensor Initiative, Control, August 2001.
4. System Samples Materials without Opening Manway, Chemical Pro-
Meyers, R. A., Encyclopedia of Analytical Chemistry: Instrumentation
cessing, mid-November 1986.
Applications, New York: John Wiley & Sons, 2000.
5. www.sludgecontrols.com.
Sherman, R. E., Process Analyzer Sample-Conditioning System Technology,
New York: John Wiley & Sons, 2002.
Strauss, R., Filtering Samples to On-Line Analyzers: Advanced Control and
Bibliography Instrumentation, Houston, TX: Gulf Publishing Co., 1987.
Sugar, J. W. and Brubaker, J. H., Sampling for Waste Water Analyzers,
American Petroleum Institute, Manual on Installation of Refinery Instru- InTech, June 1973.
ments and Control Systems: Part IIProcess Stream Analyzers, latest Van den Berg, F. W. J. et al., Selection of Optimal Process Analyzers for
edition. Plant-Wide Monitoring, Analytical Chemistry, 74(13), 2002.
Anderson, R., Sample Pretreatment and Separation, 1987. Webster, J. G., The Measurement and Instrumentation and Sensors Hand-
Clevett, K. J., Handbook of Process Stream Analyzers, Chichester, U.K.: book, Boca Raton, FL: CRC Press, 1999.
Ellis Horwood, Ltd., latest edition. Woodget, A. and Cooper, D., Sample and Standards, 1987.