Peak Purity 2
Peak Purity 2
Peak Purity 2
Product Note
References ......................................................... 15
Glossary .............................................................. 16
1
Introduction Method validation criteria
An essential requisite of a separa- Neglecting peak purity confirma-
tion analysis is the ability to verify tion means, in quality control, an Selectivity (peak purity determination)
the purity of the separated species, impurity hidden under a peak
that is, to ensure that no coeluting could falsify the results or, in re- Linearity
or comigrating impurity contrib- search analysis, important infor-
utes to the peaks response. The mation might be lost or scientific Limits of detection and quantification
confirmation of peak purity should observations rendered void should
be performed before quantitative an impurity remain undiscovered. Quality of data (accuracy and precision)
information from a chromato Validated analytical methods usu-
graphic or electrophoretic peak is ally include the peak purity check Ruggedness
used for further calculations. as a major item in the list of their
method validation criteria. Table 1
Peak purity determination a major
criterion in method validation
Several techniques are currently that involve overlaying signals to In the elution dimension:
used for peak purity determina- compare peak profiles and calcu- signals
tion in high performance liquid late signal ratios. These detectors
chromatography (HPLC)1 and in have some disadvantages and A valuable tool in peak purity
capillary electrophoresis (CE).2 these will be discussed in the fol- analysis is the overlay of separa-
With a conventional single wave- lowing section. They can be elimi- tion signals at different wave-
length detector (or a monitor pro- nated easily by a third approach lengths to discover dissimilarities
viding just one single output sig- based on diode-array technology:1 of peak profiles. The availability
nal) such as a refractive index- or on-line acquisition of UV-Visible of spectral data in the three-di-
conductivity detector, peak purity spectra during the peaks elution mensional matrix generated by the
can only be judged from the peak several signals at different diode-array detector enables sig-
profile of this signal. Peak profile wavelengths in parallel nals at any desired wavelength to
however is influenced by a variety signal extracts from a 3-D data be selected and reconstructed for
of parameters and depends heavi- matrix, containing spectral data peak purity determination after
ly on chromatographic or electro- and separation signals, for the analysis. A set of signals can
phoretic resolution. Therefore peak purity analysis. be interpreted by the observer
peak purity determination based best when, before being displayed,
on the peak profile of a single Peak purity determination can be it is normalized to maximum ab-
signal is a very unreliable and performed in different levels, sorbance or to equal areas. A good
insensitive method. This is espe- tailored to the complexity of the overlap, where peak shape and
cially true in CE, where mis- separation and the users needs. retention or migration time match,
matched sample and buffer zones Detailed descriptions of the indicate a pure peak while a poor
always result in peak skewing.2 As various peak purity routines for overlap indicates an impure peak,
a consequence, a second approach signals and spectra, such as as demonstrated in figure 1.
involving multiple wavelength normalization and overlay, the
detectors acquiring more than one mathematical operations and the
signal in parallel has been adop- different display modes will be
ted. Such detectors enable impuri- given in the following sections.
ties to be uncovered by methods
2
pure impure
signal A
(offset) signal A
signal B signal B
Figure 1
7.5 8.1 7.5 8.1 Normalized signals
Time [min] Time [min] for pure and impure
peak
mAU
200
150
100
50
signal A signal B
1.6
1.5
1.4
1.3
1.2 impure pure
1.1 Figure 2
7.6 8.0 8.4 8.6 Signal ratiograms
7.8 8.2
Time [min] for impure and pure
peaks
3
Peak purity determination in mined if the spectra are distinct Peak apexThe apex spectrum
the third dimension: enough. If the spectra resemble is subtracted from all other peak
spectra each other very closely, and the spectra (figure 3). This mode is
column or capillary system does recommended if spectra are
Comparing peak spectra is prob- not resolve the impurity and main collected in the peak controlled
ably the most popular method to component well, then only 5% mode. In figure 6 the different
discover an impurity. If a peak is impurity may be feasible. spectra acquisition modes of
pure all UV-Visible spectra ac- Hewlett-Packards diode-array
quired during the peaks elution or Before proceeding, the difference detectors are illustrated. The
migration should be identical, between the terms spectral im- disadvantage of this method is
allowing for amplitude differences purity and peak impurity should that the resulting spectra show a
due to concentration. The results be clearly defined. Spectral impu- lower absorption (about half of
obtained by comparison of these rity indicates a distortion of the the expected response) and for
spectra against each other should analyte spectrum by the near-con- weakly absorbing compounds, the
be very close to a perfect 100% stant presence of background ab- relative noise level increases.
match. Significant deviations can sorbance from solvents, and/or
Baseline spectrathe spectra at
be considered as an indication of matrix compounds and/or an im-
the start and at the end of the
impurity. Unfortunately the in- purity. Peak impurity, in contrast,
integrated peak are linearly inter-
verse is not necessarily true. If the refers to a distortion of the analyte
polated and subtracted from each
spectra are not significantly differ- spectrum by an additional compo-
peak spectrum (figure 4). This
ent, the peak can still be impure nent which partially or completely
mode is recommended if spectra
for one or more of three possible co-elutes or comigrates with the
have been collected with the di-
reasons: (1) the impurity is major compound.3
ode-arrays All spectra or All in
present in much lower concentra-
peak spectra acquisition modes
tions than that of the main com- Background correction of the
(see also figures 6 and 7). A limi-
pound, (2) the spectrum of the peak spectra
tation here is that peaks must be
impurity and the spectrum of the Before the spectra are used for the
baseline separated, otherwise
main compound are identical or peak purity analysis, they should
impurities from neighboring peaks
very similar, (3) the impurity com- be corrected for background ab-
are introduced. Selected baseline
pletely coelutes or comigrates sorption caused by the mobile
spectra should not lie too far
with the main compound, with phase or matrix compounds, by
away from the actual peak apex.
both having exactly the same peak subtracting the appropriate refer-
profile. ence spectra. Whether such a Nearest baselineThe baseline
background correction needs to spectrum recorded closest in time
Any peak purity algorithm can be applied depends on the separa- to the right or left of the integra-
only confirm the presence of tion system employed. For iso- tion borders is used here as the
impurities and never prove cratic separations with a properly reference spectrum (figure 4).
absolutely that the peak is pure. balanced diode-array detector, the This mode is recommended if
The likelihood of discovering an solvent's constant spectral contri- spectra acquisition has been
impurity rises with increasing bution will be eliminated by the performed in peak controlled
distinction between spectra and automatic subtraction of the sol- mode, and sufficient spectra are
peak profile, higher resolution vent spectrum at the beginning of stored in the spectral data file for
between the main compound and the run. For gradient separations there to be a baseline spectrum
the impurity, and with increasing where the mobile phases contri- close enough for every peak
absolute and relative concentra- bution to absorbance may change spectrum.
tion of the impurity. with time, background corrections
should be made for each peak
There are no hard and fast rules individually. The Hewlett-Packard
as to which concentrations of HPLC3D and 3DCE ChemStation
impurities can be detected and softwares offer five different
which not. In general, less than modes for setting reference
0.51% impurity can be deter- spectra.
4
without reference spectrum
peak peak
start end using nearest baseline
peak 3
Baseline Nearest
Peak segment baseline
peak 5 spectrum
1 b1 to b2 b1
peak 4
peak 1 2+3 b3 to b4 b3
peak 2 4 b3 to b4 b4
peak 6
5 b3 to b4 b3
6 b5 to b6 b5
Figure 4
b1 b2 b3 b4 b 5 b6 Reference spectra
for All spectra or
All in peak spectra
Manual referenceOne or two No referenceSpectral opera- The wavelength range for the
reference spectra can be specified tions are performed without any spectra can, and should be,
by the user. This mode is used for reference (recommended for selected carefully so that only the
interactive spectral evalu- ations exceptional situations only, even significant spectral area is under
of non-baseline-separated peaks. isocratic separations should use a observation. This eliminates the
Only the spectra in between the reference spectrum). high absorbance of eluants in the
two selected reference points are lower UV range that normally
used for the purity evaluation. cause high spectral noise. Higher
wavelength ranges should be
omitted if the compounds show
no absorbance to avoid increased
noise and calculation time.
5
Normalization of the peak
spectra (a) maximumminimum (b) match normalization (c) area normalization
normalization
Before the spectra are compared
they should be normalized. The maximum
ChemStation offers three modes
of normalization (figure 5):
area
maximum absorbance of the minimum
spectrum (or a particular wave- maximum
best possible match of the entire Both spectra have same Area of difference Both spectra have same
spectrum. This type of normaliza- absorbance range spectrum as small area
as possible
tion is recommended to display
spectra for peak purity evaluation
because it tries to make the
difference of both spectra as
small as possible by shifting and
rescaling the spectra. Figure 5
Three normalization modes, (a) maximumminimum normalization, (b) match
normalization and (c) area normalization
All spectra
Figure 6
Acquisition of spectra at different peak sections
6
The number of spectra per peak 3 peak spectra
to be processed can be three or apex
more. If the value is set to three - 23 width 2
+ 3 width
then three spectra are taken at
roughly equidistant points during
the peaks elution or migration. If
5 peak spectra apex
spectra have been acquired in - 38 width + 38 width
peak-controlled mode during the
- 45 width + 45 width
run, you should select All spectra
for purity determination because
in most practical cases this corre- apex
7 peak spectra
sponds to the three to five spectra - 38 width + 38 width
that will have been recorded. If all - 45 width + 45 width
spectra were acquired during the - width + width
run, a setting of five to seven
spectra to be processed is wise 9 peak spectra apex
(see figure7). Too many spectra - 38 width + 38 width
will only increase calculation time - 5
8 width + 58 width
7
- 98 width- 8 width + 78 width + 9 width
and display time, without provid- 8
Comparing the peak spectra ties is the display of difference files result from spectral noise
After selection, correction for spectra, generated by subtracting which may be caused by the in-
background influences and norma- these normalized spectra from the strument (figure8, upper section),
lizing, the spectra can be overlaid other peak spectra selected. The whereas systematic trends will be
to check for possible spectral im- profiles of the difference enable a observed if real spectral differ-
purities. Figure 8 (lower section) further conclusion to be drawn: ences caused by a spectral impu-
shows the normalized and overlaid randomly distributed residual pro- rity occur (figure 9, upper sec-
spectra of a pure peak and figure tion).
9 an impure peak. Any significant
dissimilarity encountered in the
comparison of the spectra re-
corded across the peak indicates
the presence of an impurity. How-
ever no conclusions can be drawn
concerning the kind, number and
level of impurity. An additional aid
in interpreting spectral dissimilari-
Figure 8
Normalized spectra and randomly Figure 9
distributed residual spectra resulting Systematic trends of difference spectra
from spectral noise indicating spectral impurity
7
More advanced spectral techniques
i=n i=n
and their differences, it is also
i=1
(Ai Aav)
2
i=1
2
( Bi Bav )
Absorbance
260 nm
able to calculate a numerical spectrum 2 Similarity 999.968
value to characterize the degree Slope 1.06818
50
of dissimilarity of the peak spec- and where Ai and Bi are measured Intercept 0.04693
tra, a so-called similarity factor, absorbances in the first and sec- 40
based on the match of the peak ond spectrum respectively at the 30
spectra to one another. same wavelength; n is the number 20
of data points and Aav and Bav the 10
Several statistical techniques are average absorbance of the first and 0
0 10 20 30 40 50 Absorbance
available for comparison of second spectrum respectively (see spectrum 1
spectra. Since UV-Visible spectra also figure 10). At the extremes, a
contain only a small amount of similarity factor of 0 indicates no
fine structure, the least-square-fit match and 1000 indicates identical
coefficient of all the absorbances spectra. Generally, values very 18 mAU
at the same wavelength gives the close to the ideal similarity factor
best result. The similarity factor (greater than 995) indicate that 260 nm
used in the ChemStation is defined the spectra are very similar, values
Figure 10
as: lower than 990 but higher than 900 Similarity of absorbance at the indivi-
Similarity = 1000 x r2 indicate some similarity and un- dual wavelengths plotted for a pair of
derlying data should be observed spectra gives the similarity factor
more carefully. Figure 11 shows
(a) very similar spectra (b) different spectra (c) spectra with impurity (d) spectra with noise
Spectral difference 0.6% Spectral difference 55% Spectral difference 4.8% Spectral difference 0.5%
Figure 11
Graphical display of similarity factor for different pairs of normalized spectra
8
examples of similarity factors for
(a) peak without impurity
identical, similar, different and and noise
(b) peak without impurity
but with noise
noisy spectra. The slope of the
regression lines represents the
ratio of the concentration of the
two spectra.
9
A threshold curve shows the ef-
(a) peak without impurity but with noise (b) peak with impurity and noise
fect of noise on a given similarity
curve. The effect increases rap-
idly towards both ends of the
peak. In essence, a threshold 5% impurity
curve is a similarity curve with
background noise contribution. 980 980 similarity
threshold curve
Figure 14(a) shows both the simi- curve
larity curve and the threshold similarity threshold
curve
curve for a pure peak with noise, curve
1000
figure 14(b) for an impure peak. 1000
10
Using specific target spectra
11
Calculating a similarity factor
13
containing 5% of an impurity with
a very similar spectrum to that of
the main compound. The overlay
of the normalized spectra and the
extracted signals clearly indicate
the presence of an impurity. In
this case the overlay of spectra
and their obvious differences
would be sufficient to recognize
the impurity. Residual spectra
confirm the impurity: a systematic
difference not originating from
random noise. The ratiograms
possess the skewed profile of an
impure peak and the similarity
curve exceeds the threshold
curve, leading to the warning mes-
sage. From the extracted signals
we can conclude that the impurity
Figure 20
is at the end of the peak. Peak purity determination of a 1% impurity
Threshold curve
Figure 21
Discovering of 0.5% impurity by
similarity and threshold curve
Similarity curve
14
Automation of peak purity
determination
3
H.-J. P. Sievert and
A. C. J. H. Drouen, Spectral
matching and peak purity in
Diode-Array Detection in High-
Performance Liquid Chromatog-
raphy, 1993, 51125, Marcel
Dekker, New York.
15
Glossary
Absorbance threshold Puts a limit on the signal response considered to represent background or baseline response. Analytes
(signal) showing an absorbance stronger than the limit trigger the integrator's peak recognition function.
Absorbance threshold Puts a limit on the spectra used for peak purity analysis, so that only those showing an absorbance stronger
(spectrum) than a particular level are considered in the calculations, thereby reducing the contribution of noise.
Apex The point of maximum response of an analyte's peak at a particular wavelength.
Background absorption Absorption caused by the mobile phase, electrolyte or matrix compounds. Signals and spectra can be
corrected by subtracting the appropriate reference signal or reference spectrum recorded at a point in the
separation where no analyte absorbance was recorded.
Difference spectra The residual absorbance resulting from the subtraction of one spectrum from another.
Downslope The portion of the peak after the peak apex but before the peak end recognised by the
integrator.
Impurity Anything co-eluting or co-migrating with the analyte under study.
Method A method for a separation should, ideally, contain all the conditions needed to reproduce the analytical
results. For post-run data evaluation, such as peak purity analysis, the evaluation conditions can be stored
together with the separation conditions in the HPLC3D or CE3D ChemStation's method.
Peak-controlled acquisition Spectra can be recorded selectively during the separation, on the peaks only, with an absorbance threshold
for the signal. Any response above this level triggers the acquisition of spectra at regular intervals.
Peak purity analysis By observing an analyte's absorbance characteristics during elution, an analyst can study a peak's homo-
geneity, and make conclusions about the absence of co-eluants or co-migrants. This is what is meant by peak
purity. Non-homogeneous characteristics indicate peak impurity.
Ratiogram A plot of the ratio of the absorbance at two distinct wavelengths over time. A flat trace indicates a homo-
geneous peak, while any distortions indicate an impurity at the wavelengths chosen.
Reference signal Control signal recorded at a wavelength unaffected by analyte absorbance.
Reference spectrum Control spectrum recorded at a point on the baseline near to the analyte peak.
Signal Absorbance measured at a specific wavelength over time. When plotted, this trace gives a profile of the
separation, a chromatogram in HPLC, an electropherogram in CE.
Similarity curve The degree of similarity of one particular analyte spectrum, that at the apex of the peak for example, to all the
other spectra recorded during the separation. The similarity curve can also be plotted as a natural logarithm
to enhance the detail at the peak's apex, or as a ratio to the threshold curve.
Similarity factor Mathematical operations can reduce an analyte's absorbance characteristics to a numerical expression of
(also known as purity factor) peak purity, such that for example, an ideal pure peak would have a purity factor of 1000. A purity factor less
than 990 indicates a degree of uncertainty or probable impurity.
Spectrum Absorbance measured at several discrete wavelengths over a wavelength range. Diode-array technology
can record this data at 100 ms intervals, fast enough for on-line acquisition during the separation.
Spectral similarity The degree of agreement between the absorbance values of a set of spectra.
Target spectrum The spectrum where the most difference in peak homogeneity will be apparent, and therefore where the peak
purity analysis should be targeted at. If an impurity is suspected in the tail, then the target should be the tail.
Threshold curve Shows the effect of noise on a given similarity curve. The threshold curve can also be plotted as a natural
logarithm to enhance the detail at the peak's apex, or as a ratio to the similarity.
Upslope The peak front, the portion of the peak after the peak start recognised by the integrator but before the
peak apex.
Copyright Hewlett-Packard
Company, 1993. All Rights Reserved.
Reproduction, adaption or translation
without prior written permission is
prohibited, except as allowed under the
copyright laws.
Publication Number
12-5091-7502E
16