Calog Manual
Calog Manual
C A L I B R AT O R S
/RDGFHOO
&DOLEUDWRU
/RDGFHOO'LVSOD\
7HVW/RDGFHOO
7HVW,QVXODWLRQ
www.zemic.nlDFRP
Contents
1
8.0 Weighing System Pre-calibration ...15
8.1 Sourcing mV (and Measure Excitation Voltage) ...15
(and Measure Milliamps)
8.2 Sourcing mV & Mass or Mass & mV ...16
(and Measure Excitation Voltage)
(and Measure Milliamps)
2
1.0 Introduction
Simply connect a loadcell (or strain gauges in a Wheatstone bridge) to the spring
terminals and press ‘test’... The back-lit graphic display will show four or six wire,
zero balance, input and output resistance and bridge balance.
Connect leads to the housing and screen and press ‘insulation test’ for a
50 volt dc test to display bridge to shield, bridge to housing and shield to housing
insulation resistance in megohms.
Now that we have the power of L-ion batteries it is possible to use the hand-held
calibrator as a loadcell indicator for a reasonable period of time. For longer
applications, such as with a cattle scale, auxiliary power can be supplied via an
external 12 volt battery, charger or car cigarette lighter socket. In order to display
mass, force, strain or torque from loadcells the unit allows you to set the
sensitivity in mV/V, range, decimal points and units. Zero trim can offset
deadweight and span trim is available to increase the reading accuracy with the
help of calibrated weights. Selections of software options such as tare, peak hold
and auto zero maintenance are included. The indicator can be used for one 350
Ohm or three 1000 Ohm loadcells with a fixed 5 Volt DC excitation.
For weighing system measurement connect the test leads to the loadcell output to
show millivolts. Alternatively include excitation measure or value and set up to
display mass in a variety of selectable units. The loadcell transmitter output
milliamps can also be selected for simultaneous display.
Finally, the calibrator can simulate a loadcell transmitter by injecting milliamps into
a loadcell display or scada system. The display can be in mA or set to display
mass and output mA.
Note : The CALOG - LC II is not suitable for AC or square wave pulse amplifiers.
Excitation volts and millivolts are measured and sourced in DC only.
3
2.0 Getting Started
2.1 Keypad
2.2 Display
The Main Menu (Icon Based)
Menu
Submenu
4
2.3 Safety
Your calibrator is not suitable for use in explosive atmospheres. Sparks at the
terminals could cause ignition.
Test leads with 4mm safety terminals are supplied with the unit
Electric shock injury may result from improper use while connected to the mains
power supply.
To ensure reliability of the unit please use anti-static equipment when working
with the circuit boards.
Press and hold down the enter key for approx. 2 seconds to switch the unit on.
Press and hold down the enter key for approx. 4 seconds to switch the unit off.
3. When the power bar stops scrolling, remove the charger from the unit and
the wall socket.
4. The charging voltage can be between 8 - 15 VDC with the inner pin as +ve.
This means that the CALOG can be charged from a DC source, such as a
car cigarette lighter socket using a suitable cable.
5
3.0 Flow Diagram
6
4.0 Wiring Schematics
This section shows examples of wiring configurations. These are also found
in the Help Menu.
4.1 Test Loadcell and LC Display
1 2 3 4 5 6
1. Sense + 4. Millivolt -
2. Excitation + 5. Excitation -
3. Millivolt + 6. Sense -
+ +
MEASURE MEASURE
mV mA
SOURCE + + SOURCE
1 2 3 4 5 6
1. Housing 4. Millivolt -
2. Excitation + 5. Excitation -
3. Millivolt + 6. Shield
7
4.3 Weighing System Measurement
Measure -5 - 35.00mVdc
Measure (0-20.000Vdc)
Measure (0-24.000mA)
- mA
+
-
+
+ +
MEASURE MEASURE
mV mA
SOURCE + + SOURCE
+
CALOG Ex+
mV- mV+
Ex-
8
4.4 Weighing System Pre-calibration
mV + mV -
Source -5 - 50.000mVdc
Measure (0-20.000Vdc) mA
Measure (0-24.000mA)
℀ Used to test or pre-
-
+
calibrate loadcell
transmitters or indicators
℀ Excitation can be MEASURE
mV
+ +
MEASURE
mA
measured or set on the SOURCE + + SOURCE
keypad
℀ Measure excitation on
LCT sense terminals for
6-wire system
-
+
Sense + Sense -
Use for electrically pre-calibrating a loadcell system. In the simplest form inject the
desired mV after doing all your own calculations. However, the CALOG - LC II can
do mV&mass (enter mV mass) or mass&mV (enter mass & display mV) by
measuring the excitation (on the sense terminals) or set a value for excitation with
mV/V and select units in the submenu.
9
4.5 Source Milliamps
0 to 24.000mA
℀ Used to send milliamp
-
signal to control room or
+
to test or calibrate mA
input devices
℀ In source mA mode the MEASURE
mV
+ +
MEASURE
mA
milliamp output changes SOURCE + + SOURCE
from measure mA to
source mA with 24V DC
driving voltage. Ensure
receiving end (SCADA,
PLC, indicators, recorder, -
etc.) is a passive input.
+
SCADA mA
Control Room
Use to send a mA signal to the control room or to test or calibrate milliamps input
devices such as indicators, recorder or PLCs. In the simple mA mode you set a
value on the display using the value set keys, output is activated immediately.
Loop error will flash if circuit open. Should you wish to set a mA value and display
the mass equivalent then you must select mA&mass in the sub-menu and enter
the mA range, 0 (4)mA value, 20mA value, decimal point and units. If you wish to
set the mass value and display the mA equivalent then select mass&mA and enter
the same sub-menu values as above.
10
5.0 Test Loadcell
Select TEST LOADCELL in the main menu.
11
7.0 Weighing System Measurement
Select MEASURE in the main menu then press the enter key.
Measure mV only
or
Measure mV
Measure excitation V
Measure mA
12
7.2 Measure mVdc & Mass (and Measure Excitation Voltage)
(and Measure Milliamp)
Used to display mV and mass from LC system for calibration. (For loadcell display
go to LC Display)
mV/V
Set loadcell mV/V from data or nameplate.
Zero Value
Set low scale value.
13
Span Value
Set high scale value.
14
8.0 Weighing System Pre-calibration
Select SOURCE mV in the main menu then press the enter key.
15
8.2 Sourcing mV & Mass or Mass & mV
(and Measure Excitation Voltage)
(and Measure Milliamps)
mV/V
Set Loadcell mV/V from data or nameplate
Zero Value
Set low system value for 0mV
16
Span value
Set high value.
(If 3 x 1 000kg Loadcell = 3 000kg).
17
9.0 Source Milliamps
Select Source mA in the main menu then press the enter key.
9.1 Source mA
18
Loop Power (left or right keys to change)
On (24V loop is active)
or
Off (Simulates mA)
O(4) mA value:
Set to low scale value for 0 (or 4) mA
19
20 mA value:
Set to full scale mass value for 20mA
Press enter
Set mA value and mass is also displayed.
20
Dec Point (left or right keys to change)
Select required position for decimal point
O(4) mA value:
Set to low scale value for 0 (or 4) mA
20 mA value:
Set to full scale mass value for 20mA
Press enter
21
10.0 Loadcell Display
Select LC Display in the main menu then press the enter key.
mV/V
Set to sensitivity value on loadcell
or datasheet
Display Zero
Set zero value
22
Display Span
Set span value
Filter Time
Set value for 1, 3, 5 or 10 seconds
Zero Trim
Use to to tare off deadweight
Press enter to trim zero
wait 10 seconds to trim
After Zero Trim completed goes to Span Trim
23
Span Trim
Press enter, load test weights and key in value,
press enter to trim span.
After approx. 10 seconds the unit reverts
back to the top of the LC Display Setup menu.
Press enter to go to the display screen.
11.0 Datalog
This menu allows the user to setup the CALOG - LC II so it may be used to datalog
it’s Loadcell display values. Downloading is done via a SD card and the onboard USB
port. Set Date & Time and format the SD Card in the Settings menu before
continuing.
24
Item: Use Left and Right keys to select block
of five characters and use the Set Value keys
to scroll
This identifies the type of instrument,
e.g. Cattle scale
is write protected
For instant datalogging whilst in the LC Display mode, insert the SD card
and press enter to start and enter to stop.
Note, that B y, Tag, Item and Interval will be the same as selected in
the Datalog menu
25
12.0 Datalogging Interface
After Datalogging is complete, the SD card can be removed from the
CALOG - LC II and inserted into a card reader to download the logged data to a PC.
Alternatively the .csv files can be downloaded directly from the CALOG via a USB
cable plugged into the USB port. The SD card MUST remain inserted in the
calibrator.
Once downloaded to a PC, the directory is the start date and time.
Click once to list the files.
Double click to open the .csv file (comma separated variables)
Use a suitable program to read the .csv files such as Excel or Open Office.
Display the file as a spreadsheet or graph.
13.0 Settings
This section will let you customise the CALOG - LC II to your personal work style.
26
Language (left or right keys to change)
English
or
German
Contrast
Press the enter key to gain access to the
contrast screen.
27
Battery Information
Press the enter key to view the battery
type and voltage level.
Unit information
Press enter to view the units serial number,
model type, firmware version, the date it was
last calibrated and its current temperature.
28
14.0 Maintenance
14.1 Opening your Calibrator
There are a few cases in which you may have to open your calibrator.
1. To change the battery pack.
2. To change the fuse.
In order to open it, place it face down on a flat, hard surface and unscrew the
four screws as shown. Then gently pull the two halves apart.
Caution: be very careful not to tear the keypad ribbon or damage the electronics.
If the battery pack fails you must order a new pack from your dealer. Use only
the factory supplied lithium Ion battery pack as it contains a temperature sensor.
Open the calibrator as shown above and then disconnect the old pack and replace
it with the new one. Be careful not to damage the electronics while the unit is
open, an anti-static mat and wrist band should be used.
Do not force the connector.
29
14.3 Resetting the Unit
The unit may fail if it passes through a strong electro-magnetic field or is
subjected to high voltages or current. In this event there is a reset switch within
the unit that should restart it. Press the button through the small hole in the
housing (near the power connector) using a paper clip.
The 250mA input protection fuse, marked F1, can be found at the top left of the
sub PCB
Take the blown fuse out and replace it with a fuse of the same rating.
See 17.0 Accessories for the part number.
1. The batteries can be recharged hundreds of times but will eventually fail.
2. Only use dealer approved batteries that contain a temperature sensor.
3. Do not store your calibrator in extreme heat, above 50ºC as the batteries may
become unstable and rupture .
14.6 Cleaning
Do not clean the unit while it is switched on. Do not use harsh abrasives or
solvents on the unit especially on the display window. The outside of the unit
may be safely wiped using a damp cloth. Do not attempt to clean the inside of
this unit.
30
15.0 Technical Specifications
15.1 General
Physical
Size: 147X33X75mm without rubber boot, 155X43X85 with rubber boot.
Weight: approx 480 g
Environmental Ranges
Working temperature: 0ºC to 35ºC
Storage temperature (battery included): -10ºC to 50ºC
Humidity: 0 to 85% (non-condensing)
Battery Pack
Type: Lithium Ion (LION) 2100mAH.
Charge time: 5 hours.
Battery run life: approximately 6 hours.
17.0 Accessories
The following items can all be ordered from your dealer or purchased separately.
CALOG - PRESSURE II: for measuring pressure in the ranges 0 - 3 bar, 0 - 30 bar, or
0 - 100bar. Displays actual value or pressure trend. Contains all CALOG - LOOP II
features.
CALOG - PRO: the process unit which measure, sources and simulates milliamps,
volts, counts and frequency.
RALLYTIME GPS: a GPS synchronized rally control and timing device for motor
sport. Now with radio and internet linked times. See www.rallytime.co.za
CALOG
C A L I B R AT O R S
www.zemic.nlDFRP