TomoView 2.10R21
Release Notes
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 1/22
Table of Contents
1. Introduction...............................................................................................................................4
2. What is TomoView?....................................................................................................................4
5. Requirements.............................................................................................................................7
5.1 Computer.....................................................................................................................................7
5.2 Supported Windows Operating System.......................................................................................7
5.3 Dongle (Hardware Security Key)..................................................................................................7
6. Languages...................................................................................................................................8
7. User’s Manual............................................................................................................................8
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 2/22
10. Features no longer supported.....................................................................................................9
11. Changes......................................................................................................................................9
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 3/22
1. Introduction
This document contains information related to the newly released edition of TomoView 2.10R21.
For more information on the related NDT Remote Control Library and NDT Data Access Library
versions, refer to the corresponding release notes.
2. What is TomoView?
TomoView is a PC-based software for data acquisition with the following units:
o TomoScan FOCUS LT (standard or rackmount)
o OmniScan MX with PA1 and UT1 modules, OmniScan MX2 with PA1, PA2 (32:128 and
32:128PR) and UT1 modules
o TomoScan FOCUS, Tomoscan III PA, µTomoscan™ and MultiScan MS5800 (UT only)
► Important Note: TomoView does not support the OmniScan iX for data acquisition.
TomoView is a PC-based software for UT data visualization and reporting for data acquired
with the following units:
o TomoScan FOCUS LT (standard or rackmount)
o OmniScan MX with PA1 and UT1 modules (version 2.0R2 or higher), OmniScan MX2
with PA1, PA2, UT1 and UT2 modules and OmniScan iX
o TomoScan FOCUS, Tomoscan III PA, µTomoscan™ and MultiScan MS5800 (UT only)
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 4/22
3.4 OmniScan MX2 – PA2 32:128 and 32:128PR modules support
Starting from TomoView 2.10R12, the OmniScan MX2 equipped with a PA2 32:128 and 32:128PR
modules can be used in inspection mode. Note that the OSTV PA2 1.0R1 software must be installed on
the instrument storage card for this mode to be supported.
3.6 OmniScan MX2 – PA2 & UT2 modules data files support
Starting from TomoView 2.10R4, data files generated using an OmniScan MX2 with PA2 and UT2
modules will be supported by TomoView. These data files can be associated with MXU 3.2 or higher
OmniScan MX2 software versions.
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 5/22
3.12 TCG Curves import/export from .csv files
The TCG curves can now be easily imported from or exported to .csv files for more flexibility in custom
TCG curves configurations.
5. Requirements
5.1 Computer
The followings are the minimum computer requirements to install TomoView 2.10:
Computer and processor 500 megahertz (MHz) processor or higher
Memory 2 gigabyte (GB) RAM or higher (4 GB RAM recommended for large data files and high-
speed acquisition)
100 gigabyte hard disk (GB) for optimal performance in analysis and acquisition
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 6/22
Interface USB port for HASP software protection device
A 1280 x 1024 display or higher resolution monitor
Network 100 megabits Ethernet adaptor for acquisition mode
Keyboard and mouse
Additional actual requirements and product functionality may vary based on system configuration and
operating system.
6. Languages
7. User’s Manual
A completed TomoView 2.10 revised user’s manual is now available in English. German, French and
Chinese versions will follow.
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 7/22
8. Compatibility with Previous Versions
8.1 Installation
TomoView 2.10 is compatible with the previous versions of TomoView (from TomoView 2.2 to
TomoView 2.10). Both revisions can be installed on the same computer.
However, a specific Bootp configuration needs to be saved and reloaded for each software version.
Before installing TomoView 2.10, save the Bootp configuration into a file (for instance bootp file
TomoView2.7.cfg). Install TomoView 2.10 and save new Bootp configuration in another file (for
instance bootp file TomoView2.10.cfg). Before turning on the acquisition units, load the proper file
into the Bootp, and then click the restart button. You can now open the related version of TomoView.
Note: the Bootp shortcut is found in the configuration panel.
► Important Note : You can be kept informed of the new TomoView 2.10 software version releases at
the following link: https://www.olympus-ims.com/en/service-and-support/downloads/subscribe/
► Important Note: To fully benefit from the new display features of TomoView, it is recommended to
convert OmniScan data files with the new converter even though they were converted into an .rdt
format using a previous version of the converter. The multibeam C-scan image can be displayed for
any sector scan (multiple refracted beam angles or constant beam angles) only if the file was
converted with the TomoView 2.10.
► Important Note: In order to install the NDT Data Access Library 1.12 version (which will replace the
NDT Data Access Library 1.11 version from TomoView 2.10R16), the NDT Data Access Library 1.11
should be fully uninstalled prior to the installation of the new version. The NDT Data Access Library
enables the access to the raw data from OmniScan iX, MX, MX2 & SX (*.oud and *.opd) and TomoView
(*.rdt) data files.
8.2 Files
Setup files and data files saved in a previous version of TomoView (higher than 2.2) can be loaded in
TomoView 2.10 (using the same acquisition unit). In some cases, it might be necessary to rebuild the
layout using the predefined display layout of TomoView 2.10.
► Set-up files or data files created in TomoView 2.10 should not be loaded into a previous version
which could create software instability and unexpected behavior.
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 8/22
9. Installation Software and Bootp Server
11. Changes
A few names have been changed in TomoView 2.10 from previous versions:
Channel was renamed Group
Focal Law was renamed Beam
Sequence was renamed Scan
Pane was renamed View
Contour was renamed Zone tool
Signal rectification possibilities were renamed
o Not Rectified -> RF
o Unipolar Positive -> HW+ (Positive Half Wave)
o Unipolar Negative -> HW- (Negative Half Wave)
o Bipolar -> FW (Full Wave)
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 9/22
12. Bug Fixes
The following bugs were corrected for the different TomoView versions.
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 10/22
P29153: [LawFile] Importing Law files doesn't correctly set the Velocity
P29199: [DDF] TomoView crashes when using DDF
P29245: [Skew][Analysis] Cannot change skew angle offline for UT groups
P29566: Error when using Windows XP SP3
P29728: [Converter] OmniScan converted Opd to Rdt that are loaded as setup show an
innacurate error message
P29770: [Advanced Calculator] Error message when opening Advanced Calculator
P29771: [Launcher] TomoView 2.10R14 splash screen display issue
P29835: [PostInspection] Impossible to perform Volumetric merge in Analysis after Inspection
mode
P29919: [Reading][12Bit][SoftGain] Some A% readings on A-scan and C-scan view are wrong
above 100% after ajusting Soft Gain
P30044: [IndexOffset] Changing Probe Index offset offline affect the BeamIndex Offset value
P30098: [DataAccess] A-scan TOFD amplitude .opd to .rdt correction
12.5 TomoView 2.10R14
P29770 [Advanced Calculator] Error message when opening Advanced Calculator
Set encoders in TomoView through RCLIB unlocked in Inspection mode
12.6 TomoView 2.10R13
P29008 Crash when loading specific setup files
12.7 TomoView 2.10R12
P27509: [TCG] TCG points disappear in certain conditions
P27858: [DATA] Missing warning that the Scan stop (and length) has been changed
P28485: [Preferences] Remove option for old indication table option
P27860: [DATA] Incorrect TOFD velocity
P28211: [OVERLAY] Weld overlay doesn't work on Side (B) views
P28216: [DATA] Incorrect OmniScan iX data when imported in TomoView
P27443: [Presynchronization] Missing some validation when programming Gate in
presynchronization causing error in from driver
P28109: [DATA] "Minimum Amplitude" C-Scan merge doesn't work
P27125: [OVERLAY] Weld overlay is lost when selecting the LITE WELD layout
P28204: [CRASH] Crash when using the Binarizer on a data file in Inspection mode
P28364: [REPORT] Incorrect and missing Gates information
P27782: Reading only of LAW file version 5.2
P27859: Alarms should be available for consulting during the Inspection
P28445: Importing TCG is ~100x slower than calibrating (using 10laws)
P28889: Corrected bug when sending more than 32 laws to program in PA2 mode with
OmniScan MX2
P28942: add save law file in 5.0 or 5.2 versions from the Advanced Calculator.
P27443: Missing some validation in TomoView when programming Gate in presynchronization
causing error in FPGA or Driver
Corrected TCG bug when importing phased array TCG configuration file
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 11/22
12.8 TomoView 2.10R9
P28404: New version of NDT Data Access Library that allows TomoView to convert OmniScan
data files
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 12/22
12.12 TomoView 2.10R5
P26950: Incorrect TOFD groups recognition by TomoView
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 14/22
P24332: Clicking on an Available group doesn’t refresh the "VariableBackwall" section (List,
ActivateCheckbox,B or D) to reflect current state of the group.
P24331: Activate checkbox isn’t remembered.
P24327: Can't add point on End(D) view without VolumeCorrected.
P24323: Inaccurate error message description for Update button while not on A-scan view.
P24323: Fix invalid messages.
P24313: Fix invalid ultrasound settings start and range.
P23936: Fix TCG points list refresh problem.
P24072: TCG error messages in TomoView don't appear with Windows 7
P23310: Remove second field for "gate saved"
P23310: Use XML Report 2.0R8T2 for Conditional A-Scan support.
23936: TCG control not properly refreshed
23920: Soft C-Scan created with multiple lengths does not correspond to soft C-Scan created
with regular soft C-Scan tool.
23912: Imported TCG file does not support changing range
23933: Multiple length file can't be imported if there are no ";" at the end of every lines.
23849: Invalid TCG messages.
23919: Limits not correctly validated for 0 degrees setup.
23914: Able to have negative TCG gain.
23883: 3D readings are not giving refracted angles and skews.
23983: TCG curve isn't drawn correctly when the last TCG point is outside the upper range
limit.
23850: Invalid Multiple length points messages.
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 15/22
P21617: Separation changes to 1000 after setting the value
P21752: Soft C-scan now works correctly in RF mode.
P21755: Remove checkbox and reload of option for now
P22193: Fixing another problem with the "insufficient space disk" message.
P22206: Print out quality in report is bad (rulers are blurry)
P22222: Resolve async socket reception in internal clock of A-Scan vs C-scan
P22253: Convert thickness correctly when using inches
P22262: Corrected Display and AFiSiMo dialog initialisation of probe offset
P22271: Export data in RF does not work
P22338: BUGS LINKED TO INCHES VS MM CONVERSION
P22476: File in analysis now validate correctly the DDF point limit Use same #define for
message ID for TomoView and SetupManager
P22526: Fix synchro on gate A
P22530: Remove dialog and add code to check for exception from merge operation
P22540: Support OmniScanMx2 module
P22580: Increase static box size only (title problem still present)
P22590: PassServer.tlb is now registered in the Chinese version of TomoView.
P22667: wrong behavior of gates if modified from S-Scan
P22674: Alarms aren't working with OSTV
P22711: Amplitude axis in sample goes from -128 to +128, it should stop at +127
P22769: TV takes 100% CPU in release mode with OSTV 3.0D0T29 because of socket error
P22832: Configuration selection dialog bypass preference not working
P23020: Support for UT-8 module
P23103: Wrong TOFD color palette in TomoView
P23199: Change from pulse/echo sync cause crash
P23216: Add DDF support for 32/128
P23629: Fix for corrupted file
P23641: Type of focus with DDF
P23645: Wrong offset in drawing of Beam display info.
P23913: Data is saved even if data corresponding are unchecked
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 16/22
22123: Binarizer error
22333: SNR crash
20951: Thickness on echo cannot be imported
22193: Insufficient disk space
22195: Index/Scan not refreshed correctly
22412: Layout with Dependent Setup
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 17/22
18753: TomoView Manager tooltip is "Unexpected error occured.":
18770: Wrong column header title in TCG:
18894: ReportError function ws not returning valid error message:
18900: Edit User Field menu stay disable in setup mode:
18914: On uTomo, change of range cause applied TCG from DGS to be invalid:
18921: TCG from DGS is not valid if range is using start other then zero:
18925: On Reset button in DGS dialog, the TCG display is not refreshed:
18926: TCG points displayed when done from DGS are not equidistant:
18959: DGS - Refresh of ERS in A-scan view causing screen flicker:
18960: DGS - TCG points displayed in A-scan view should not be editable:
18961: DGS - ERS is changing when gain is modified with TCG created from DGS:
18965: DGS - ERS is not taking into account the booster activation correctly:
18966: Cannot change the wedge speed on different group without weird behavior:
18967: DGS - ERS should be valid even if the TCG applied from DGS is zero:
18968: DGS - Only change/lose DGS calibration on specific changes:
18969: DGS - Add dialog option for not losing DGS calibration:
19036: Hiding TomoView main window in Remote Control:
19057: DGS - Booster gain is not used for limit in calibration:
19084: ERS info. view with TCG from DGS are not valid with AScan start different then zero:
19119: Scrolling View lost data display on mode change from inspection:
19120: Add Din support for mode change between setup & inspection:
19121: Link gate level to scrolling view first color limit:
19122: Add supervisor mode to TomoView:
19124: Add maximum in channel option for scrolling view instead of only one law:
19125: Analog input are enable on OmniScan unit:
19191: UT probe : unmodifiable preset angle:
19202: Support DGS reading in online C-scan view as A-scan:
19209: Crash when loading a palette:
19212: Need position of amplitude to be valid even if gate not crossed in ScrollingView:
19244: Avoid losing threshold level color when deactivated for strip chart:
19245: Problem with validation of level in threshold dialog for strip chart:
19260: Add minimum value support for PA in Scrolling View:
19313: Support Soft-Gate in ScrollingView:
19315: Scrolling view type addition & validation modifications:
19316: Linear merge for PA group are not working in Scrolling View:
19318: Decimation & compression mode support:
19321: Missing folder for elementary A-scan creation:
19322: Viewer must not contain in the toolbar the zone control:
19405: Missing refresh for A-Scan when moving focal law cursor:
19432: Empty pane button is not working:
19441: Scrolling view color selection limitation are not correct:
19451: DGS dual probe curve is not valid at the beginning:
19488: Crash with viewer on specific OmniScan (.oud) file:
19511: Icon of Adv. Cal. in debug mode is wrong:
19512: Problem with importing data file from OmniScan iXU to TomoViewer:
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 18/22
19637: Change of beam with shorcut F8/F7 is slower:
19690: Add special color support for scrolling view:
19723: Problem when loading a iXU 2.1R12T5 data file:
19732: "Auto ending" encoder function doesn't work:
19756: C-Scan immediately stops when started:
19762: TV2.9R4: Error with C-Scan merge for linear file imported with OmniScan Converter:
19769: TV2.9R4: Move of gate cause software failure ("Crash"):
19778: No trace in scrolling view after gate move in A-scan:
19780: Change in order of data in scrolling view corrupt color display:
19783: Zone color are set using wrong gate for AMP-TOF view in Scrolling View displays:
19821: Moving gate A after loading a specific setup cause application "Crash":
19875: Wrong color palette error message (French message):
19876: Wrong NO DETECTION color in Scrolling View:
19880: NO DETECT color should be replaced for some Scrolling View types:
19885: Group Creation Wizard buttons are in the wrong language:
19888: When adding a new group, TV passes to acquisition mode:
19889: When adding a new group and then a new A-Scan, TV changes all the A-Scans:
19893: Alarm cannot be set using gate 4 (D):
19896: Compression checkbox is unavailable (not displayed) in some configuration:
19963: Set range not working in inches.:
19968: TomoView crashes after a certain time on screensaver:
19975: TomoView crashes when opening table of measurements after moving gate A:
19988: Missing values when loading a .cal file:
19992: A-Scan and S-Scan signals are flicking:
19993: Wrong index resolution for the merged group:
19994: Scan position incoherence between merged and not merged C-Scans:
19996: Display problem (black lines):
19998: Problem when merging data files with different resolutions:
19999: Problem when merging data files with different angles:
20000: Patch at the beginning of merged C-Scans:
20004: Log all rpc and socket errors in TomoView:
20016: Problem with specific "Data + Layout" files:
20018: "No detection" zones become "No data" zones in the merged C-Scan:
20031: Problem when merging after having moved gate A:
20039: Display doesn't work with some file:
20045: Wrong background display for C-Scan merge:
20046: Wrong background color for C-Scan merge:
20090: When no HASP key present, title of application stay TomoView even if in viewer
mode:
20094: Signal tearing in the A-Scan while in acquisition mode:
20095: Using set range feature on sector Scan does not work if blue cursor is set before red
cursor in Time Base:
20106: User numbers in the setup vs user numbers from active pods (Multipod):
20111: Color dialog need to have controls hidden:
20112: C-Scan filter cut gate C & D from list:
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 19/22
20119: Velocity type in material is not kept between operation:
20144: Viewer layout:
20150: Simulation size is too big for display (cannot zoom out enough):
20153: Display always unzoom when changing options:
20292: Cascading, file save, tile window toolbar need to be removed from viewer mode:
20293: Layout of converter OmniScan file should not be loaded into viewer mode:
20298: Data (RDT) file pops up error messages when corresponding report (R01) file
doesn't exist:
20300: Enable in TVManager the button to load layout, activate only the viewer as choice:
20317: Default path created are invalid in preference for the palette path:
20331: TomoView pops up a message dialog in Standalone mode which prevents
automated test
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 20/22
17443 Changing layout with maximized view has no effect
17453 Excel info. view are not working
17461 iTest development (Remote Control automatic random test unit)
17552 Version of Windows Vista (64 bits) are not supported
17559 Default.acq error message on first start
17680 Report: Sound speed displayed in inches/sec.
17682 Duplicate readings in the readings selection dialog
17691 Help not working with Adv. Cal. only installation
17693 Uninstall of Adv.Cal. crash when unregistering file PassServer.tlb
17748 Crash in reporting
17754 Stacked Cscan VC (zero degree only) support needed
17765 PASSServer is not working with the new HASP drivers
17772 Bring back the binarizer
17783 Binarizer : Block component with PA data file
17798 TV Manager gate value is incorrect with echo synchronization
17819 Incredibuilt support needed
17823 Reading HR Filter
17826 Unable to see pane content when loading a PWZ40 in TV29
17840 Add AFiSiMo parameters in GUI
17892 Delay of element in Pulse-Echo mode is wrong
17897 Loading file with view maximized does not load the layout
17946 Crash when deleting values in the calculator
17948 Optimize when simulation not in all laws of XCal file
17967 Preference dialog do not has its caption translated
17975 AFiSiMo not working with latest PassServer version
17981 Indication Table: Additional info crashes with empty view
18005 TomoView’s main window title only contains the application version but should also
contain the name.
18051 Converted file cannot be opened immediately after conversion
18054 TomoView hangs when trying to close it (must kill process)
18055 XmlReportToolbox error message on start
18056 Errors messages when opening a data file in Analysis
18057 Invalid default number of element used with default probe
18058 Calculator hangs when trying to close it (must kill process)
18068 Cannot close TV29 anymore, process stay stuck in memory
18069 Cannot close AdvCal29 anymore, process stay stuck in memory
18070 Permit DGS reading to work with TCG active
18071 Angle change in Analysis, angle displayed is 0 in UT
18072 Support true depth programming for gate
18080 Crash after closing TomoView
18103 Crash using scrolling views
18128 Remove scrolling view HASP key validation
18132 The CustomServices.dll cannot be loaded through the RemoteControl library when
TomoView 2.8 is installed
18143 Gate Programmation in True Depth cause crash in UT
This document is the exclusive property of Olympus NDT inc. No part of this document may be
used or reproduced entirely or in part without the prior written consent of Olympus NDT inc.
385977586.doc Page 21/22
18154 Asserts are thrown when trying to load a data file that contains ZoneDiscrimination
information.
18159 Weird error message without English text
18168 AFiSiMo crashes with the standalone calculator
18176 Add possibility in analysis to change offset of a channel
18177 Echo dynamics not drawn correctly with 1:1 ratio
18178 Help is not appearing when the installation folder is not
"C:\OlympusNDT\TomoView29"
18213 Add DGS reflector and parameters readings
18246 Load of file cause "invalid pointer" dialog to appear
18275 Crash with specific client DDF setting
18285 TomoView crashes when opening given file.
18299 Cannot close and re-open TV when using a MS5800
18341 Gate are not programmed correctly from cursor move in VC Scan
18342 Crash on display of Soft C-scan
18361 Gate threshold is reset when using "Set Gate" buttons
18367 Crash reading old acq file
18383 VC S-scan & uncorrected S-scan are not always refreshing
18389 Minor interface English corrections
18409 PA programming could fail if more than one group
18410 Reprogramming channel by adding DDF corrupts S-Scan display
18414 Crash upon file opening (multipod)
18417 Channel control is not displayed correctly
18465 Setting any gate on linear merge data group is causing crashes
18495 Evaluate TCG conversion from TV27 to TV29 and list differences
18502 HASP key detection is not working for standalone version
18512 TomoView crashes when using ERS display with C-scan gate B position C-scan
18520 Custom layout cannot be saved in Lite mode
18541 TCG resolution is different from previous version