Lab No. 5 Vector Measurements
Lab No. 5 Vector Measurements
Lab No. 5 Vector Measurements
5 Vector measurements
Vector measurements
Vector measurements allow finding out the components of a vector in one of its forms:
cartesian or polar coordinates in real or polar plane. The result is given as 2 components: X and
Y (Re and Im) or and . Vector quantities are:
- impedance, admittance
- transfer functions (gain, attenuation)
The vector instruments are AC bridges (manual or automated), Q-meters, vector
impedance meters, vector voltmeters. Automated AC bridges allow measuring the impedance or
admittance components R, L and G, C. Manual bridges allow determining the components by
calculation. Q-meters measure the vector by resonance. They measure the frequency, the
capacitance and the quality factor. Automated Q-meters use the damped oscillation produced by
loses in the resonant circuit. Vector impedance meters display the result in polar coordinates
and as functions of frequency. They maintain one quantity constant, voltage for admittance
(the current is proportional with the admittance) or current for impedance (the voltage is
proportional with the impedance). The phase shift between voltage and current and the ratio
between the amplitudes are the modulus and the phase. Vector voltmeters have two voltage
channels. One is the reference and the other is evaluated in respect with the reference.
1. AC bridges
Z x Z 3 = Z 2 Z 4 (a)
Z x Z3= Z2 Z4 (1)
x + 3 = 2 + 4 (b)
One variable element balance the modules and the
other the phases. From equations 1.1 and 1.b it results
Fig. 1
z=Re(Zx)+j·Im(Zx). The impedances depend on frequency, so
the accuracy depends also on frequency. The sensitivity of an AC bridge is a complex parameter:
S = U 2 /U1 = F (2)
where: - ΔU2 is the unbalancing voltage that appear when one of the variable impedances is
varied around the equilibrium
1
Lab No. 5 Vector measurements
In audio frequency the components are modeled by two simple elements of circuit using
a series or parallel configuration that approximates as good as possible their operation. There 2
types of bridges: for capacitors and for inductors, either for series equivalent configuration or
parallel configuration.
a) b)
Fig. 2
The equations for the series configuration bridge are:
R3 (a); R4 (b); tgδ
C x = C2 R x = R2 ωR x C x = ω R2 C 2 (c)
R4 R3
(3)
1 1
Z x = Rx - j = R4 R2 -j
ω Cx R3 ω C 2
2
Lab No. 5 Vector measurements
a) b)
Fig. 3
b) Audio frequency bridges for inductances
The most used bridges for measuring the inductances are the Maxwell-Wien for the
3
Lab No. 5 Vector measurements
a) b)
Fig. 4
R4 (b); L
L x = C 3 R2 R4 (a); R x = R2 Q x = ω x = ω R3 C 3 (c)
R3 Rx
(7)
R 4 R 4
Z x = R x +jω L x = R2 ( 1+jω C 3 R3 ) = R2 ( 1+jQ)
R3 R3
The Hay bridge, fig. 4 b), is used for high quality factor inductances (Q ≥ 10). The
equations under balance are:
R 4 (b); Rx = 1
L x = R 2 R 4 C 3 (a); R x = R 2 Qx = (c)
R3 ω Lx ω C 3 R3
(8)
1 1 1 R3 1
Yx = -j = 1-j
Rx ω Lx R 2 R 4 ω C 3 R3
The the series model can be transformed into the parallel model:
1
L p = L s 1+ 2 (a); R p = R s 1+ Q (b)
2
(9)
Q
For Q > 5 the two models have close values: Lp ≈ Ls.
Observations
1. In AC it is not possible to reach the zero value at the balance, but a minimum value.
This value reaches the lowest value when the bridge power supply has small distortions from a
sinusoidal wave and the disturbing signals are completely annealed. The null indicator is a
4
Lab No. 5 Vector measurements
The most used bridge in RF is the Schering bridge. According to the model of the
impedance there are two types: the Schering bridge for the series model and the Schering bridge
for the parallel model: fig. 5 a) and b). The balance is achieved C1, C3 in 2 steps: without Zx and
then with Zx connected in the bridge. The bridge for the series model is balanced in the first step
with the connections Zx short circuited, while the bridge for the parallel model is balanced with
the connections opened. When balanced, the series Schering bridge equations are:
a) b)
Fig. 5
- with the connections Zx short circuited:
1 R2
Z 31 = Y 11 Z 2 Z 4 = +jω C11 = R 2 C11 + R 2 (10)
R1 jω C 4 C 4 jω R1 C 4
- with the impedance Zx connected:
5
Lab No. 5 Vector measurements
C 12 R2 Δ C1
Z x + Z 32 = R 2 + = Z 31 + R 2 (a)
C 4 jω R 1 C 4 C4
1 1 1 1 1 1 (11)
Z 31 - Z 32 = - = j - (b)
jω C 31 C 32 ω C 32 C 31
The unknown impedance is:
Δ C1 1 1
Z x = R x +jX x = R 2 j Δ
ω C 3
(12)
C4
When balanced, the parallel Schering bridge equations are:
- with Zx connections opened:
1 C4
Y 31 = Z 11Y 2 Y 4 = +jω C 31 = +jω C 4 R1 (3)
R3 C11 R2 R2
- with Zx connected:
C4 C 1 1
Y x +Y 32 = Z 12 Y 2 Y 4 = +jω C 4 R1 = Y 31+ 4 - (4)
C12 R2 R2 R2 C12 C11
From the above equations (12) and (13) the admittance value is:
1 1
Yx = = G x + jB x = C 4 Δ +jω Δ(C 3) (5)
Zx R 2 C1
The Schering bridges work for resistors, capacitances or inductances for frequencies up
to 300 MHz.
4. Q-meter
The Q- meter uses the series resonance of a circuit. Figure 6 shows a schematic diagram
of a Q-meter. At resonance, the voltage over CV is Q times the voltage U1 – neglecting the
resistance r – the order of miliohms:
1
U 2 =Q U 1
2
ω= (16.a); (16.b)
L Cv
Fig. 6
6
Lab No. 5 Vector measurements
Knowing the frequency, the capacity and measuring Q, the inductance L can be found:
1 ωL ωL 1
L= 2
(a); Q = (b); r = = (c) (17)
2
4 π f Cv r Q 2πf C v Q
1 1 1 1
Z L = r+jω L = +j = +j (18)
2πf C v Q 2πf C v 2πf C v Q
Fig. 7
The oscillating circuit is a second order system with a self oscillating angular frequency
of:
a0 1/C 1
ω0 = = = (20)
a2 L L C
and damping ratio of:
a1 r r 1
ξ= = = = (21)
2 a0 a 2 2 L/C 2 L 0 2Q
7
Lab No. 5 Vector measurements
For Q>0,5 and ξ<1 – the circuit is underdamped, leading to sinusoidal oscillations as solution of
equation (15):
ω0 r
i = Ae0t sin(ω0 1- ξ 2t+ ) ; ξ ω0 = = (22)
2Q 2L
Choosing the initial moment t1, resulting:
r 1
i = I 1 exp - t cosω0 1- ξ 2 t (23)
2L 2
2L 2Q 2π
t 2 - t1 = ln K = ln K = N T = N (24)
r ω0 ω
If I2=I1 /K, the angular frequency is:
1
ω = ω0 1- ξ 2 because ξ 1, ω ω0 (25)
2
Q
N= ln K (26)
Choosing I2 such that lnK = π, equation (26) becomes: N=Q (27).
The schematic in figure 7b) contains, besides the oscillating circuit, there is a peak
detector that detects the levels I1 and I2 and drive a flip-flop (FF) that opens or closes the gate P.
The pulses triggered from the damped oscillation pass through the gate and are counted by a
counter, the final counted value N being displayed. Switching S on the position f the pulses are
counted for given time interval (1s) and the frequency is measured Tm = Nf·T (28) Nf = Tm/T
=Tm·f (29). The instrument can have 2 counters to perform both measurements in the same time
like in figure 8.
Fig. 8
Having the values C, f and Q, the inductance components can be computed:
1 2πfL 1 1 1
L= (a); r = = (b); Z L = +j (c) (30)
2 2
4π f C Q 2πfCQ 2πfC Q
8
Lab No. 5 Vector measurements
The accuracy of this kind of Q- meters is around 1- 2 % for Q and 0,1 - 0,2% for f. The
display error has to be added ± 1/N.
This is an instrument that gives the measurement result in polar coordinates (modulus
and phase). The measurement can be performed for a single frequency or for a frequency band.
The block diagram of a digital vector impedance meter in shown in fig. 9:
Fig. 9
1= signal generator; 2= phase detector; 3= pulse shaper; 4= limiter; 5= filter; 6= output stage; 10= analog
outputs; 11= level stabilizer; 12= programmable amplifier; 14= detector; 15= amplifier; 16= calibration;
Z= modulus display; φ = phase display.
The measuring mode is different for small impedances (under 1 kΩ) in respect with high
value impedances. For the first mode the current through the impedance is kept constant, the
voltage being the measure of the impedance/admittance, while in the second case the voltage is
kept constant, the current being the measure of the impedance/admittance.
- for impedances with small value:
Ux Ux U
Z x= = (argU x - arg I x ) (a); I x =I=ct. (b); Z x = x argU x (c) (31)
Ix Ix I
9
Lab No. 5 Vector measurements
1
L = 10-i ρ sin (38) or: C = 10-i (39)
ρ sin
The measurement errors are in the range of 5% for modulus and ± 6o for phase.
6. Vector voltmeter
The vector voltmeter measures 2 voltages, one being the reference and both having a
constant frequency. The measurement is performed through synchronous sampling. If the
sampling period is chosen as:
T
T E = m T + , m,n N-{0} (40)
n
after sampling, the signal’s frequency will be:
T FI = m n +1 (41).
In fig. 10 the principle of synchronous sampling is presented. Fig 11 presents the block
diagram of a vector voltmeter. Such instrument is used to measure voltages between 1mV and
1V and frequencies up to 1GHz. Because the signals are simultaneously sampled the phase shift
between them is kept with enough accuracy. The channel A is the reference signal that is used to
synchronize the vector voltmeter. The intermediary frequency is 20kHz. The amplitudes and the
10
Lab No. 5 Vector measurements
phase shift of the two signals are converted into digital, processed by the processor and
displayed.
Fig. 10
Fig. 11
1. Use of the vector voltmeter to measure the impedance
In order to measure the impedance, a coaxial test fixture like in fig. 12 must be used. The
instrument can be interfaced with a digital frequency to compute Ls, Rs respectively Cp, Rp. The
unknown impedance can be computed with:
U A -U B U
= B (42)
Zx Z0
1 1
Z x = R x + j X x = R0 cos-1 - j R0 sin (43)
U B / U A UB /UA
11
Lab No. 5 Vector measurements
Fig. 12
12
Lab No. 5 Vector measurements
Abstract: …………………………………………………………………………………………………………………………………………..
……………………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………………
………………………………………………………………………………………………………………
Keywords:…………………………………………………………………………….
Experimental work
1. Inductances and capacitances measurement using the vector impedance meter, figure 1.
k
X
Y
|Z|
Fig. 1
Table 1
Det. f φ |Z| CX RX LX Obs.
no. [Hz] [] [Ω] [nF] [Ω] [mH]
1 159
2 1.591
3 15.915
4 159.155
13
Lab No. 5 Vector measurements
Fig. 2 Fig. 3
Table 3
Det. f φ VA VB VB/VA ZL CX RX LX Obs.
nr. [MHz] [] [mV] [mV] [Ω] [pF] [Ω] [nH]
1 1,591 0 Z0=50
2 1,591 Z0+ZX
3 3,183 0 Z0=50
4 3,183 Z0+ZX
Conclusions:
………………………………………….…………………………………………………….
…………………………………………………………………………………………………
…………………………………………………………………………………………………
……………….………
Observation: Schematics must be completed with instrument types and components
values used in the experiments.
14