Kurz Model 504FTB Technical Specifications
Kurz Model 504FTB Technical Specifications
Kurz Model 504FTB Technical Specifications
SPECIFICATIONS
Information subject to change without notice. Contact or visit Kurz online for complete specifications and ordering information. | 367547E
Series 504FTB
Temperature influences the physical properties of gases, so temperature All options include USB interface with ASCII text and Modbus protocol
compensation is required for a thermal sensor to accurately measure gas through RS-485.
flow rates.
The 4-20mA analog outputs (AO) are used for flow rate and/or
• Standard Temperature Compensation (STC) is used for process temperature, or one AO for PID flow control. All AO are NAMUR NE-43
temperatures from 0°C to 125°C over a moderate velocity range. compliant.
• Velocity Temperature Mapping (VTM) is used when the process Relay digital outputs (DO) can be alarms, EPA zero-mid-span drift is
active, or pulsed totalizer function. PID uses one 4-20mA output for the
temperature and gas velocity vary widely. Multiple velocity
flow controller. The EPA zero-mid-span drift check requires a contact
calibrations are stored in the meter. VTM compensation is based on closure to start the drift check. All 4-20mA outputs are used during the
air; specific gas correlations are required to ensure accuracy at high Drift Check Calibration process.
temperatures.
EPA zero-mid-span drift check can be initiated using digital inputs (DI),
SPECIALTY GAS VELOCITY CALIBRATION elapsed runtime automatic drift check, Modbus, or HART.
The 4-20mA analog input (AI) supports feedback to the device.
There are two types of gas calibration:
• Laboratory gas calibrations are performed with gases of high
purity and are NIST traceable. Values above the calibrating facility
limit are correlated up to the specified range. Customers must
specify the calibration process pressure.
• Correlation gas calibrations are based on experimental
data correlated to an Air calibration at ambient pressure and
temperature. The flow element is calibrated in Air, and then an
additional calibration data sheet is generated using the correlation
factors. All correlation calibrations include velocity-temperature
mapping.
Add ±5% of reading to the accuracy specification when using a
correlation calibration.
For Oxygen gas, the customer is responsible for ensuring the mass flow
sensor is clean of hydrocarbons and safe for Oxygen use.
ELECTRONICS ELECTRONICS
ENCLOSURE H ENCLOSURE H
CAUTION LABEL CAUTION LABEL
EXTERNAL EXTERNAL
GROUND LUG GROUND LUG
#10-32 SCREW #10-32 SCREW FLOW DIRECTION
ARROW
MODELS 504FTB-6A, -6, -8, -12, & -16 MODELS 504FTB-24, & -32
SHOWN DIRECTLY ATTACHED WITH STANDARD SHOWN DIRECTLY ATTACHED WITH STANDARD
DISPLAY ORIENTATION (NOTE 5) DISPLAY ORIENTATION (NOTE 5)
FLOW
L1
L
(NOTE 1)
5.03 [127.76mm]
5.03 [127.76mm] CUSTOMER
HOOK-UP 3/4" FNPT 4.75 [120.65mm]
3/4" MNPT 4.75 [120.65mm] SIDE. 3x (Typical)
3x (Typical)
5.25
5.25 MASS FLOW TRANSMITTER [133.35mm]
[133.35mm] CLEAR GL0SS DISPLAY
4.61
4.61 [117.09mm]
[117.09mm]
4.37
4.37 [110.99mm]
[110.99mm]
1.13 W2 W1
[28.70mm] MOUNTING FEET (SEE TABLE 5) (SEE TABLE 5)
W3
1.06
(SEE TABLE 5)
SUBJECT TO DUAL CHAMBER
CHANGE
[26.90mm]
SENSOR WIRE TERMINAL ELECTRONICS ENCLOSURE
JUNCTION BOX (SHOWN W/ WINDOW LID)
(NOTE 5)
(NOTE 4)
504FTB-6 3/8 0.675 [17.15mm] 1/2 7.00 [177.80mm] 2.50 [63.50mm] 9.94 [252.48mm] 7.00 [3.18kg] 9.00 [4.08kg] 11.00 [4.99kg]
504FTB-8 1/2 0.840 [21.34mm] 1/2 8.00 [203.20mm] 3.00 [76.20mm] 10.02 [254.51mm] 7.50 [3.40kg] 9.50 [4.31kg] 11.50 [5.22kg]
504FTB-12 3/4 1.050 [26.67mm] 3/4 10.00 [254.00mm] 3.00 [76.20mm] 10.13 [257.30mm] 8.00 [3.63kg] 11.50 [5.22kg] 13.50 [6.12kg]
504FTB-16 1 1.315 [33.40mm] 1 12.00 [304.80mm] 3.50 [88.90mm] 10.25 [260.35mm] 8.50 [3.86kg] 12.50 [5.67kg] 14.50 [6.58kg]
504FTB-24 1-1/2 1.900 [48.26mm] 1-1/2 18.00 [457.20mm] 4.00 [101.60mm] 9.30 [236.22mm] 10.50 [4.67kg] 19.00 [8.62kg] 23.00 [10.43kg]
504FTB-32 2 2.375 [60.33mm] 2 24.00 [609.60mm] 5.00 [127.00mm] 9.54 [242.32mm] 14.00 [6.35kg] 24.00 [10.89kg] 30.00 [13.61kg]
504FTB-40 2-1/2 2.875 [73.03mm] 2-1/2 24.00 [609.60mm] 5.00 [127.00mm] 11.78 [299.21mm] N/A 32.50 [14.74kg] 42.50 [19.28kg]
504FTB-48 3 3.500 [88.90mm] 3 24.00 [609.60mm] 5.00 [127.00mm] 11.78 [299.21mm] N/A 40.00 [18.14kg] 54.00 [24.49kg]
504FTB-64 4 4.500 [114.30mm] 4 24.00 [609.60mm] 5.00 [127.00mm] 11.78 [299.21mm] N/A 62.50 [28.35kg] 82.50 [37.42kg]
Table A: Standard Full-Scale Flow Rate Ranges (Qmax) To compute the expected pressure drop for a flow rate, multiply the full
scale pressure drop by the square of the flow ratio.
Laboratory 1 Correlated 2
Model Flow Area Calibration Range Calibration Range
Number (ft2) Table B: Estimated Pressure Drop (DPs)
SCFM NCMH SCFM NCMH
504FTB-06A 0.00044 2 3.1 8 12.4
Gas Type Inches H2O mm H2O kPa
504FTB-06 0.00107 10 15.6 40 62.4 Air 30 762 7.46
504FTB-08 0.00179 20 31 80 124 Argon 40 1016 9.95
504FTB-12 0.00328 40 62 160 248 Butane 18 457 4.47
504FTB-16 0.00546 75 117 300 468 Carbon Dioxide 45 1143 11.19
504FTB-24 0.01337 150 234 600 936 Dry Ammonia 18 457 4.47
504FTB-32 0.02253 300 468 1200 1872 Dry Chlorine 51 1295 12.69
504FTB-40 0.03248 400 624 1600 2496 Ethane 18 457 4.47
504FTB-48 0.05057 600 936 2400 3744 Ethylene 20 508 4.97
504FTB-64 0.08763 1000 1560 4000 6240 Helium 2 51 0.49
Hydrogen 1 25 0.24
Note: 1. The baseline maximum flow rate for each transmitter model number. Methane 15 381 3.73
2. Compressed air only. Digester Gas: 50% CH4, 50% CO2 28 711 6.96
Digester Gas: 60% CH4, 40% CO2 25 635 6.22
SCFM Standard Cubic Feet Per Minute, Reference: 77°F, 14.69 PSIA Digester Gas: 70% CH4, 30% CO2 23 584 5.72
NCMH Normal Cubic Meters Per Hour, Reference 0°C, 760 mmHg Nitrogen 29 737 7.21
NCMH = 1.56 x SCFM (approximate) Oxygen 33 838 8.21
Propane 14 356 3.48
Parent Number Model Pipe Size & Length F2 Sensor Material / Flow Body and Flange Material
755960 504FTB-06A 0.375” x 7” (10 x 178 mm) Choose one option from each category.
755961 504FTB-06 0.375” x 7” (10 x 178 mm)
Option Sensor Material
755962 504FTB-08 0.5” x 8” (13 x 203 mm)
755963 504FTB-12 0.75” x 10” (19 x 254 mm) 3 C-276 alloy
755964 504FTB-16 1” x 12” (25 x 305 mm)
Option Flow Body and Flange Material
755965 504FTB-24 1.5” x 18” (38 x 457 mm)
2 316L stainless steel
755966 504FTB-32 2” x 24” (51 x 610 mm)
755967 504FTB-40 2.5” x 24” (64 x 610 mm)
755968 504FTB-48 3” x 24” (76 x 610 mm) F3 Option Flow Body Connection Type Pressure Rating
755969 504FTB-64 4” x 24” (102 x 610 mm) A Male NPT pipe ends (MNPT) 300 PSIG
B Class 150 ANSI B16.5 flanges (CL150) 150 PSIG
Note: Models 504FTB-06 to 504FTB-32 use the MD sensor type. C Class 300 ANSI B16.5 flanges (CL300) 300 PSIG
Model s 504FTB-40 to 504FTB-64 use the FD2 sensor type.
F4 Option Process Temperature Compensation
Electronics Enclosure Configuration and Standard temperature compensation over process range of
F1 Option
Input Power -40°C to 125°C for all gases.
1
Directly attached dual-chamber electronics enclosure, Accuracy: ± (1 + 2000 x A/F) %,
A where F = SCFM, A= flow body area (ft2)
AC power, display / keypad
Directly attached dual-chamber electronics enclosure,
B
AC power, without display / keypad Note: An accuracy specification of ±0.025%/°C = (A*)(0.25 SFPM/°C)
should be added for temperatures above or below standard.
Directly attached dual-chamber electronics enclosure
C
rotated 180° for viewing, AC power, display / keypad
F5 Gas Flow Rate Calibration Data Range
Remote dual-chamber electronics enclosure,
D Option Description Option Description
AC power, display / keypad
Remote dual-chamber electronics enclosure, A 100% Qmax I 60% Qmax
E
AC power, without display / keypad B 95% Qmax J 55% Qmax
Directly attached dual-chamber electronics enclosure, C 90% Qmax K 50% Qmax
F
DC power, display / keypad
D 85% Qmax L 45% Qmax
Directly attached dual-chamber electronics enclosure
G E 80% Qmax M 40% Qmax
rotated 180° for viewing, DC power, display / keypad
Directly attached single-chamber electronics enclosure, F 75% Qmax N 35% Qmax
H
DC power, without display / keypad G 70% Qmax P 30% Qmax
Remote dual-chamber electronics enclosure, H 65% Qmax Q 25% Qmax
I
DC power, display / keypad
Remote single-chamber electronics enclosure, Note: Refer to Tables A and B to determine the correct selection for calibration
J
DC power, without display / keypad range selection.
Remote polycarbonate electronics enclosure,
R
AC power, with display / keypad
Remote polycarbonate electronics enclosure,
S
AC power, without display / keypad
Stainless steel sensor and electronics enclosures,
V
AC power, with display / keypad
Stainless steel sensor and electronics enclosures,
W
AC power, without display / keypad
Stainless steel sensor and electronics enclosures,
X
DC power, with display / keypad
Stainless steel sensor and electronics enclosures,
Y
DC power, without display / keypad
Digester Gas 70% CH4 Remote — Transmitter and sensing element separate
37 87 Sensor enclosure: Stainless Steel Type 4X, IP66
30% CO2
Electronics enclosure: Stainless Steel Type 4X, IP66
– 8K User-Defined Binary Gas Composition Explosion-Proof / Flame-Proof sensor: CSA, ATEX, and IECEx
– 8M One Gas Curve H Ex d IIB + H2 Tx ; Ex d IIB + H2 Tx Gb; Class I Zone 1 AEx d IIB + H2 Tx Gb
Sensing element, Tp: -40°C to 45°C: T4 or to 110°C: T3
– 8N Two Gas Curves DC power electronics housing, Ta: -40°C to 65°C: T4
AC power electronics housing, Ta: -40°C to 50°C: T4 or to 65°C: T150°C (T3)
– 8O Three Gas Curves
CSA Approved for Class I, Division 1, Groups B, C, D
– 8P Four Gas Curves
– 8Q Five Gas Curves F8 Option Process Pressure
40 90 Nitrogen
---
Enter the Absolute Pressure (PSIA) rounded to 3 digits.
44 94 Oxygen For example, a process Absolute Pressure of 14.7 PSIA,
46 96 Propane round to 15.0 and enter 015; for 150 PSIA enter 150.
---
Enter the Absolute Temperature (°Rankin = °F + 460)
rounded to 3 digits. For example, a Process Temperature of
77°F is written as 537 (77 + 460).