Coleman Vacu-print™
T E C H N I C A L N O T E S
Background
The Coleman Vacu-Print™ equipment allows you to glue-fume evidence in an air-evacuated, sealed chamber. Vacuum
technology for glue fuming is relatively new. This method will develop latent fingerprints without having excessive
residue coat the surface of the evidence, and it will be easier to handle the evidence. Because there is no residue
buildup on the evidence, dye-staining for fluorescent examination is more effective. Dyes such as Rhodamine 6G and
Ardrox adhere to the glue residue on the item. When there is excessive buildup of the glue residue, the dye stains all of
it, causing the entire surface to fluoresce, perhaps obscuring ridge detail.
Safety
CAUTION: Do not place pressurized items such as sealed soda cans, sealed glass bottles or aerosol cans in the
chamber. They can expand rapidly or explode as air pressure is evacuated from the chamber, potentially
causing injury.
Equipment
The components of the Coleman Vacu-print set consist of a vacuum pump and a fuming chamber. The fuming chamber
has a gauge, a release valve and a hose mount on it. The gauge indicates the amount of vacuum pressure measured in
inches of mercury. The release valve is used to allow air back into the chamber. The hose mount is used to connect the
hose from the pump to the chamber. The vacuum pump (if purchased with this equipment) is 1⁄3 horsepower rated 3
CFM and includes oil. Read the instruction books that come with the vacuum pump to learn about care and mainte-
nance. However, ignore the part about changing the oil in the pump after every use. It might not be necessary to change
the oil but once per year, depending on your usage. Note the names of the various parts of the vacuum pump in the
diagram shown in this instruction sheet. Follow the directions for tightening the Oil Filter Plug and Drain Valve and fill
the pump with the oil included in the shipping box.
Method
Place the items of evidence inside the Coleman Vacu-Print™ chamber. It is not necessary to unfold garbage bags or to
leave large amounts of space between items of evidence. If using liquid glue, put a small volume (2 or 3 ml) in a small
dish or folded-up piece of aluminum foil. Or, if using the Hard Evidence™ pouches, cut off a half-inch strip, open it up
and attach it to the inside wall of the chamber with tape. For best results, we recommend using Hard Evidence™
pouches. Put the lid on the chamber and close the release valve.
Turn on the vacuum pump. Open the Gas Ballast Valve one half turn and open the Isolation Valve (to up
position). To help the lid seal properly, press on the lid until the chamber begins to evacuate. This
may or may not be necessary, depending upon the ambient tempera-
Isolation Valve
ture. About 5 to 10 seconds, after the vacuum pump has begun to
(Up is open and
evacuate the chamber, close the Gas Ballast Valve. Evacuate the
down is closed)
chamber to approximately 25 inches of Mercury as shown on the
gauge. This should take no more than 60 to 75 seconds. When the
pressure reaches 25 inches of mercury, close the Isolation Valve.
Open the Gas Ballast Valve, wait 2 to 3 seconds and turn off the
pump. Then, close the Gas Ballast Valve. Leave the items in the
chamber under vacuum for about twenty minutes. If items are left
Gas Ballast Valve (slightly hidden longer (even overnight), there is no danger of over-fuming.
by the handle)
When ready, allow air back into the chamber by slowly opening the release valve on the chamber. Remove the lid,
taking care to stand back from the opening of the chamber. There may be a strong odor of glue fumes, and it may be an
irritant to the eyes, nose and throat. Take out the evidence. It is recommended to allow the items to sit in the air for
about 10 to 15 minutes before dusting or processing with liquid staining chemicals. This helps to harden the developed
latent prints.
Dual Connector
A Dual Connector (in a “Y’ shape) is available if you wish to operate the two chambers with the same pump. The Dual
Connector attaches to the pump and the hoses from the chambers are attached to it. Place evidence in both chambers.
Close the hose on the Dual Connector to one of the chambers and start the pump. It will evacuate the other chamber.
When the first chamber is under vacuum, close the hose on the Dual Connector to it. Open the hose on the Dual
Connector to the second chamber and run the pump to evacuate it. It is not recommended to operate the pump and
evacuate both chambers simultaneously. It would take too much time to do so. Set up your evidence and glue in one
chamber and evaluate it. Then, you can set up and evaluate the second chamber.
Examination
Most latent prints developed with glue fuming in the aquarium style of processing will have white ridges and will be
quite noticeable, but the disadvantage is that there will also be residue on the background surface. With this vacuum
method the latent prints will not be as visible, but they will show up as sharp, clear prints when viewed with a Forensic
Light Source or UV light after treatment with the appropriate dye or powder. Sometimes, however, the latent prints will
be very difficult to see; therefore, each piece of evidence must be carefully inspected. Oblique lighting from a flashlight
will be helpful. It is suggested that these prints be photographed before other processes are used. The latent prints can
be visually enhanced by dusting with a regular or magnetic fingerprint powder contrasting with the background. Once
dusted with powder, some glue-fumed latent prints can be lifted with regular lifting tape. The latent prints are quite
durable and usually can be lifted more than once. Sometimes the second lift is clearer than the first.
Photography
Visible, contrasting prints can be photographed in the usual way. However, photography of white, glue-developed
prints on light-colored or transparent objects may need some different lighting. If the object is transparent, place a piece
of paper of contrasting color behind it and photograph as normal. Or, move the light source behind the object. The light
will shine through the transparent areas, but the latent prints will appear dark, as no light will shine through them.
If the object is light-colored and not transparent, it may be necessary to dust with black powder or use fluorescent dyes
and powders on the latent print to provide the needed contrast. If using fluorescent dyes or powders, it will be necessary
to photograph the latent prints using the same forensic light source wavelength as used to visualize the latent print with
the eye.
Helpful Hints
• Be careful not to drop or chip the plastic lid. If it is chipped too badly, it will not seal properly and a vacuum will not
be achieved.
• Clean the chamber with a household cleaner between each use to avoid heavy build up of residue.
• Let the evidence sit for 10-15 minutes in the air before dusting or dye staining.