Optiflux 4000
Optiflux 4000
Optiflux 4000
The documentation is only complete when used in combination with the relevant
documentation for the converter.
1 Product features 3
2 Technical data 8
3 Installation 23
4 Electrical connections 28
5 Notes 31
Highlights
• Standard device in the process industry
• Robust and reliable
• More than 300,000 units operating in the field
• Works reliably under demanding conditions: High temperatures (up to 180°C / 356°F) and low
conductivity (non-water from 1 µS/cm, water from 20 µS/cm)
• Quick and easy to install and operate
• Chemically resistant to alkaline solutions and acids
Industries
• Chemicals
• Pulp & Paper
• Water
• Wastewater
• Minerals & Mining
• Iron, Steel & Metals
• Pharmaceuticals
• Oil & gas
Applications
• For clean liquids
• For slurries and pastes with high solids content
• For abrasive and aggressive products
1.2 Options
The solution for any industry
The OPTIFLUX 4000 has been designed for
measuring any application in any industry and is
often referred to as "mother of all electromagnetic
flowmeters", suitable for all standard and special
applications, including "commercial" use according
to European Directive MI-001 and -005. The level of
precision complies with the industry demands,
furthermore does the modular concept allow tailor-
made solutions.
All meters are passing specific trials,
measurements and tests that go beyond the legal
specifications - and on which our customers can rely
100%. For example, we subject the converter
electronics to a series of extensive temperature
change tests, in which the converter is exposed to
cyclical fluctuations ( -20...+65°C / -4...+149°F)
Every meter that leaves our factory is first wet-
calibrated on our officially certified calibration rigs
(EN 17025).
Communication
Electromagnetic flowmeters can be installed in
random locations, demanding the instrument to be
suitable for any kind of environment. These locations
vary from indoor use (integrated in complex
processes) to outdoor use (straight forward
transport, but possibly buried or submerged). The
ability to read the measured results on-site may be
standard, but it does not always meet the current
and actual needs of the user or operator. That is why
this meter comes with optional, state-of-the-art bus
communication systems. The stored data is
transmitted by HART®, Profibus, Fieldbus or
Modbus, and then forwarded to a management
system.
Construction
The measuring tube of the sensor has a smooth,
cylindrical shape. This design, consisting of a
circular cross section (no internal or moving parts)
and a homogeneous magnetic field, forms the basis
for a flow-optimized pipe cross section, thereby
providing reliable measurements that are largely
independent of the flow profile. This design allows
the sensor to measure the flow bi-directional.
As an additional benefit, there is no possibility for
product to build up and the free cross section sized
to the customer pipeline creates no pressure drop.
In addition, the required straight inlet and outlet
runs are only 5D and 2D. The liner of the measuring
tube is made of high tech plastic or hard rubber and
is resistant to vacuum, corrosion, aging and
abrasion. The surface and shape of the measuring
tube also minimize mineral deposits, resulting in
exemplary measurement quality - even over the long
term.
in which:
v = mean flow velocity
k = factor correcting for geometry
B = magnetic field strength
D = inner diameter of flow meter
The signal voltage U is picked off by electrodes and is proportional to the mean flow velocity v
and thus the flow rate q. A signal converter is used to amplify the signal voltage, filter it and
convert it into signals for totalising, recording and output processing.
Measuring system
Measuring principle Faraday's law
Application range Electrically conductive fluids
Measured value
Primary measured value Flow velocity
Secondary measured value Volume flow
Design
Features Fully welded maintenance-free sensor.
Flange version with full bore flow tube
Standard as well as higher pressure ratings
Broad range of nominal sizes
Industry specific insertion lengths
Modular construction The measurement system consists of a flow sensor and a signal
converter. It is available as compact and as separate version.
Compact version With IFC 040 converter: OPTIFLUX 4040 C
With IFC 100 converter: OPTIFLUX 4100 C
With IFC 300 converter: OPTIFLUX 4300 C
Remote version In wall (W) mount version with IFC 100 converter : OPTIFLUX 4100 W
In field (F), wall (W) or rack (R) mount version with IFC 300 converter:
OPTIFLUX 4300 F, W or R
Nominal diameter With IFC 040 converter: DN10... 150 / 3/8... 6"
With IFC 100 converter: DN2.5...1200 / 1/10...48"
With IFC 300 converter: DN2.5...3000 / 1/10...120"
Measurement range -12...+12 m/s / -40...+40 ft/s
Measuring accuracy
Reference conditions Flow conditions similar to EN 29104
Medium: water
Electrical conductivity: ≥ 300 μS/cm
Temperature: 10...30°C / 50...86°F
Inlet section: ≥ 5 DN
Operating pressure: 1 bar / 14.5 psig
Wet calibrated on EN 17025 accredited calibration rig by direct
volume comparison
Accuracy curves For detailed information on the measuring accuracy, see chapter
"Measuring accuracy".
Option: verification to MID MI-001
(Only in combination with IFC 300)
Option: calibration according to OIML R49
(Only in combination with IFC 300)
Related to volume flow (MV = Measured Value)
These values are related to the pulse / frequency output
The additional typical measuring deviation for the current output is
±10 μA
With IFC 040 converter:
v < 1 m/s / 3.3 ft/s: ± 5 mm/s
v > 1 m/s / 3.3 ft/s: ± 0.5 % of MV
With IFC 100 converter:
DN2.5...6: ± 0.4% of MV + 1 mm/s
DN10...1200: ± 0.3% of MV + 1 mm/s
With IFC 300 converter:
DN2.5...6: ± 0.3% of MV + 2 mm/s
DN10...1600: ± 0.2% of MV + 1 mm/s
DN1800...3000: ± 0.3% of MV + 2 mm/s
Repeatability ±0.1% of MV, minimum 1 mm/s
Long term stability ±0.1% of MV
Special calibration Better accuracies optional
Operating conditions
Temperature
Process temperature PTFE: -40...+180°C / -40...+356°F for remote versions
PTFE: -40...+140°C /-40...+284°F for compact versions
PFA: -40...+180°C / -40...+356°F for remote versions
PFA: -40...+140°C / -40...+284°F for compact versions
ETFE: -40...+120°C / -40...+248°F
Hard rubber: -5...+80°C / 23...+176°F
PU: -5...+85°C / 23...+185°F
For Ex versions different temperatures are valid. Please check the
relevant Ex documentation for details.
Ambient temperature Standard (with aluminium converter housing):
-40…+65°C / -40…+149°F (Protect electronics against self-heating
with ambient temperatures above 55°C)
Option (with stainless steel converter housing):
-40...+55°C / -40…+130°F
For Ex versions different temperatures are valid. Please check the
relevant Ex documentation for details.
Storage temperature -50…+70°C / -58…+158°F
Pressure
EN 1092-1 DN2200...3000: PN 2.5
DN1200...2000: PN 6
DN200...1000: PN 10
DN65 and DN100...150: PN 16
DN2.5...50 and DN80: PN 40
Other pressures on request
ISO insertion length Optional for DN15...600
ASME B16.5 1/10...24": 150 lb RF
Other pressures on request
JIS DN50...1000: 10 K
DN2.5...40: 20 K
Other pressures on request
Vacuum load For information on pressure limits depending on liner material see
chapter "Vacuum load".
Pressure ranges for secondary For DN25...150:
containment
Pressure resistant up to 40 bar / 580 psi
Burst pressure up to approx. 160 bar / 2320 psi
Pressure drop Negligible
Chemical properties
Physical condition Electrically conductive liquids
Electrical conductivity Water: ≥ 20 μS/cm
Non water: ≥ 1 μS/cm
Permissible gas content (volume) ≤ 5%
Permissible solid content ≤ 70%
(volume)
Installation condtitions
Installation Take care that flow sensor is always fully filled
For detailed information see chapter "Installation"
Flow direction Forward and reverse.
Arrow on flow sensor indicates positive flow direction.
Inlet run ≥ 5 DN
Outlet run ≥ 2 DN
Dimensions and weights For detailed information see chapter "Dimensions and weights".
Materials
Sensor housing DN2.5...15: Stainless steel 1.4408
DN20: GTW-S 30
DN25...3000: Sheet steel, polyurethane coated
Other materials on request
Measuring tube Austenitic stainless steel
Flange Standard: Carbon steel, polyurethane coated
Other materials on request
Liner Standard
DN2.5...15: PFA
DN20: PTFE
DN25...150: PFA
DN200...3000: ETFE
Option
DN200...600: PTFE
DN200...1800: PU
DN200...3000: Hard rubber (Ex only)
Other materials on request
Connection box (only remote Standard: polyurethane coated die-cast aluminium
versions)
Option: Stainless steel
Measuring electrodes Standard: Hastelloy® C
Option: Platinum, stainless steel, titanium, tantalum, low noise
Other materials on request
Process connections
Flange
EN 1092-1 DN2.5...3000 in PN 2.5...40
ASME 1/10...120" in 150...2500 lb RF
JIS DN2.5...1000 in JIS 10...20 K
Design of gasket surface RF
Other sizes or pressure ratings on request
Electrical connections
Signal cable
Type A (DS) Standard cable, double shielded.
Max. length: 600 m / 1950 ft (dep. on electrical conductivity and
measuring sensor). See documentation of the converter for more
information.
Type B (BTS) Optional cable, triple shielded.
Max. length: 600 m / 1950 ft (dep. on electrical conductivity and
measuring sensor). See documentation of the converter for more
information.
Hazardous areas
ATEX Please check the relevant Ex documentation for details.
Compact version with IFC 040 C converter
II 2 GD
Compact version with IFC 100 C converter
II 2 GD
Compact version with IFC 300 C converter
II 2 GD or II 2(1) GD
Remote version
II 2 GD
FM In combination with IFC 300 converter
Class I, Div 2, groups A, B, C and D
Class II, Div 2, groups F and G
CSA In combination with IFC 300 converter
Class I, Div 2, groups A, B, C and D
IEC-Ex Compact version with IFC 100 converter
IIC T4
NEPSI GYJ05234 / GYJ05237
Ex me ia IIC T6...T3
Ex de ia IIC T6...T3
Ex qe ia IIC T6...T3
Ex e ia IIC T6...T3
Other approvals and standards
Custody transfer Standard: without verification
Only in combination with IFC 300 converter
For diameters: DN25...500 (other diameters on request)
Cold water
MI-001 type examination certificate
OIML R49 certificate of conformity
Conformity with ISO 4064 and EN 14154
Liquids other than water
MI-005 type examination certificate
OIML R117 certificate of conformity
Hygiene PFA liner is FDA approved.
Protection category acc. to Standard: IP 66/67 (NEMA 4/4X/6)
IEC 529 / EN 60529
Option: IP 68 (NEMA 6P)
IP 68 is only available for separate design and with a stainless steel
connection box
Vibration resistance IEC 68-2-6
Random vibration test IEC 68-2-34
Shock test IEC 68-2-27
[mm] [bar] 40°C 60°C 70°C 80°C 90°C 100°C 120°C 140°C 180°C
Liner in PTFE
DN10...20 50 0 0 0 0 0 0 500 750 1000
DN200...300 50 500 750 1000 1000 1000 1000 1000 1000 1000
DN350...600 50 800 1000 1000 1000 1000 1000 1000 1000 1000
Liner in PFA
DN2.5...150 50 0 0 0 0 0 0 0 0 0
Liner in ETFE
DN200...2000 150 100 100 100 100 100 100 100 - -
Liner in Hard rubber
DN200...300 150 250 400 400 400 - - - - -
DN350...3000 150 500 600 600 600 - - - - -
Liner in PU
DN200...1800 1500 500 600 - - - - - - -
[inches] [psi] 104°F 140°F 158°F 176°F 194°F 212°F 248°F 284°F 356°F
Liner in PTFE
3/8...3/4" 725 0 0 0 0 0 0 7.3 10.9 14.5
8...12" 725 7.3 10.9 14.5 14.5 14.5 14.5 14.5 14.5 14.5
14...24" 725 11.6 14.5 14.5 14.5 14.5 14.5 14.5 14.5 14.5
Liner in PFA
1/10...6" 725 0 0 0 0 0 0 0 0 0
Liner in ETFE
8...72" 2176 1.5 1.5 1.5 1.5 1.5 1.5 1.5 - -
Liner in Hard rubber
8...12" 2176 3.6 5.8 5.8 5.8 - - - - -
14...120" 2176 7.3 8.7 8.7 8.7 - - - - -
Liner in PU
8...72" 21756 7.3 8.7 - - - - - - -
2.3 MI-001
All new designs of water meters that are to be used for legal purposes in Europe require
certification under the Measuring Instruments Directive (MID) 2004/22/EC. Annex MI-001 of the
MID applies to: water meters intended for the measurement of volume of clean, cold or heated
water in residential, commercial and light industrial use. A type examination certificate to the
MID certificate is valid in all the countries of the European Union.
TheOPTIFLUX 4300 has a type examination certificate to the MI-001 for DN25...300.
The EC type examination certificate for the OPTIFLUX 4300 is valid for the compact and the
remote version and applies for forward and reverse flow.
According to MI-001 the maximum permissible error on volumes delivered between Q2
(transitional) flow rate and Q4 (overload) flow rate is ±2%. The maximum permissible error on
volumes delivered between Q1 (minimum) flow rate and Q2 (transitional) flow rate is ±5%.
Figure 2-1: ISO flow rates added to figure as comparison towards MID
X: Flow rate; Y: Maximum measuring error [%]
2.4 MI-005
MI-005
DN Span (R) Flow rate [m3/h]
Figure 2-2: ISO flow rates added to figure as comparison towards OIML
X: Flow rate; Y: Maximum measuring error [%]
1 ±3% for class 1, ±5% for class 2 devices
2 ±1% for class 1, ±2% for class 2 devices
Q1 = Q3 / R
Q2 = Q1 * 1.6
Q3 = Q1 * R
Q4 = Q3 * 1.25
• All data given in the following tables are based on standard versions of the sensor only.
• Especially for smaller nominal sizes of the sensor, the converter can be bigger than the
sensor.
• Note that for other pressure ratings than mentioned, the dimensions may be different.
• For full information on converter dimensions see relevant documentation.
EN 1092-1
Nominal size Dimensions [mm] Approx.
weight [kg]
DN PN [bar] L H W
DIN ISO
2.5 40 130 - 142 90 3
4 40 130 - 142 90 3
6 40 130 - 142 90 3
10 40 130 1 - 106 90 6
15 40 130 1 200 106 95 6
20 40 150 200 158 105 7
25 40 150 200 140 115 4
32 40 150 200 157 140 5
40 40 150 200 166 150 5
50 40 200 200 186 165 9
65 16 200 200 200 185 9
80 40 200 200 209 200 12
100 16 250 250 237 220 15
125 16 250 250 266 250 19
150 16 300 300 300 285 27
200 10 350 350 361 340 34
250 10 400 450 408 395 48
300 10 500 500 458 445 58
350 10 500 550 510 505 78
400 10 600 600 568 565 101
450 10 600 - 618 615 111
500 10 600 - 671 670 130
600 10 600 - 781 780 165
700 10 700 - 898 895 248
800 10 800 - 1012 1015 331
900 10 900 - 1114 1115 430
1000 10 1000 - 1225 1230 507
1200 6 1200 - 1417 1405 555
1400 6 1400 - 1619 1630 765
1600 6 1600 - 1819 1830 1035
1800 6 1800 - 2027 2045 1470
2000 6 2000 - 2259 2265 1860
1 150 mm for construction according to order code VN03.
150 lb flanges
Nominal size Dimensions [inches] Approx. weight
[lb]
ASME PN [psi] L H W
1/10" 284 5.12 5.59 3.50 6
1/8" 284 5.12 5.59 3.50 6
¼" 284 5.12 5.59 3.50 6
3/8" 284 5.12 1 5.08 3.50 12
½" 284 5.12 1 5.08 3.50 12
¾" 284 5.91 5.28 3.88 18
1" 284 5.91 5.39 4.25 7
1 ¼" 284 5.91 5.98 4.62 7
1 ½" 284 5.91 6.10 5.00 11
2" 284 7.87 7.05 5.98 18
3" 284 7.87 8.03 7.50 26
4" 284 9.84 9.49 9.00 40
5" 284 9.84 10.55 10.0 49
6" 284 11.81 11.69 11.0 64
8" 284 13.78 14.25 13.5 95
10" 284 15.75 16.3 16.0 143
12" 284 19.69 18.78 19.0 207
14" 284 27.56 20.67 21.0 284
16" 284 31.50 22.95 23.5 364
18" 284 31.50 24.72 25.0 410
20" 284 31.50 26.97 27.5 492
24" 284 31.50 31.38 32.0 675
1 5.91" for construction according to order code VN03
300 lb flanges
Nominal size Dimensions [inches] Approx. weight
[lb]
ASME PN [psi] L H W
1/10" 741 5.12 5.59 3.75 6
1/8" 741 5.12 5.59 3.75 6
¼" 741 5.12 5.59 3.75 6
3/8" 741 5.12 1 5.24 3.75 15
½" 741 5.12 1 5.24 3.75 15
¾" 741 5.91 5.67 4.62 20
1" 741 5.91 5.71 4.87 11
1 ½" 741 7.87 6.65 6.13 13
2" 741 9.84 7.32 6.50 22
3" 741 9.84 8.43 8.25 31
4" 741 11.81 10.00 10.0 44
6" 741 12.60 12.44 12.5 73
8" 741 15.75 15.04 15.0 157
10" 741 19.69 17.05 17.5 247
12" 741 23.62 20.00 20.5 375
14" 741 27.56 21.65 23.0 474
16" 741 31.50 23.98 25.5 639
20" 741 31.50 28.46 30.5 937
24" 741 31.50 33.39 36.0 1345
1 5.91" for construction according to order code VN03
3.2.4 T-section
3.2.5 Vibration
3.2.7 Bends
3.2.11 Pump
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
4.2 Grounding
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.
Possible if:
• ≥ DN10
• Electrical conductivity ≥ 200 µS/cm
• Electrode cable max. 50m., type DS
Mass flowmeters
• Vortex flowmeters
• Flow controllers
• Level meters
• Temperature meters
• Pressure meters
• Analysis products
• Measuring systems for the oil and gas industry
• Measuring systems for sea-going tankers