Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Belz

Download as doc, pdf, or txt
Download as doc, pdf, or txt
You are on page 1of 24

Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd.

, 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 1 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

ELECTRO-TECHNICAL CALIBRATION

I. SOURCE

1. DC Voltage# 1mV to 300mV 0.55% to 0.01% Using Fluke 9100 MF


300mV to 300V 0.01% to 0.03% Calibrator
300V to 1000V 0.03% to 0.014%

2. DC Current# 1µA to 300mA 1.5% to 0.03% Using Fluke 9100 MF


300mA to 10A 0.03% to 0.12% Calibrator with current Coil
10A to 1000 A 0.12% to 0.70%

3. AC Voltage# 50 Hz /60 Hz Using Fluke 9100 MF


10mV to 300mV 1.15 % to 0.2% Calibrator
300mV to 100V 0.2% to 0.08%
100V to 1000V 0.08%

4. AC Current # 50 Hz /60 Hz Using Fluke 9100 MF


100µA to 300mA 0.45% to 0.13% Calibrator with current Coil
300mA to 10A 0.13% to 0.31%
10A to 1000 A 0.31% to 0.9%

5. Resistance# 50µΩ 0.3% to 0.13% Using Low Resistance box


(4 Wire) 100µΩ (Fixed Value)
0.001Ω
0.01Ω
0.1Ω
1Ω

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 2 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

6. Resistance# 1Ω to 4MΩ 1.2% to 0.07% Using Fluke 9100 MF


(2 Wire) 4MΩ to 40MΩ 0.07% to 0.18% Calibrator
40MΩ to 400MΩ 0.18% to 0.32%

High Resistance 500 MΩ, 1 GΩ, 2 2.5% Using High Resistance Box
(2 Wire) # GΩ, 10 GΩ, 20 GΩ,
100 GΩ, 200 GΩ
500 GΩ,1TΩ

7. Frequency# 10Hz to 10MHz 0.07% to 0.008% Using Fluke 9100 MF


Calibrator

8. Capacitance# 1kHz Using Fluke 9100 MF


1nF to 100µf 7.2% to 2.5% Calibrator

9. DC Power# 10W to 19 kW 0.31% Using Fluke 9100 MF


10V to 1000V Calibrator
1A to19A

10. AC Power# 10V to 640V 0.25% to 0.8% Using Fluke 9100 MF


(Active Power 1ɸ) 1A to 19A Calibrator
UPF
240 V
0.1 to 1A
0.5 Lead/Lag

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 3 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

11. Temperature Simulation#


(Indicator/controller/
Recorder)
RTD (-)200oC to 800oC 0.43oC Using Fluke 9100 MF
‘T’ Type Thermocouple (-)200oC to 400oC 0.7oC Calibrator
‘S’ Type Thermocouple (-)0oC to 1700oC 0.5oC
‘J’ Type Thermocouple (-)200oC to 1200oC 0.3oC
‘N’ Type Thermocouple (-)200oC to 1300oC 0.3oC
‘K’ Type Thermocouple (-)200oC to 1300oC 0.37oC
‘R’ Type Thermocouple (-)0oC to 1700oC 0.5oC
‘B’ Type Thermocouple 600oC to 1700oC 0.5oC
‘E’ Type Thermocouple (-)100oC to 600oC 0.5oC

12. Oscilloscope# 10 ns to 5 sec. 0.63s Using Fluke 9100 MF


Time Base Calibrator
(Marker)

Amplitude 1kHz 1.0%


(Deflection Factor) 1MΩ
5mV to 120V

Sin/Square 50kHz to 20MHz 1.0%


(DC) 50Ω
Bandwidth 20mV to 2.5V

1kHz to 250MHz 5.0%

13. Turn Ratio Meter# 1.11 0.65% Using Ratio Calibration


1.22 0.43% Standard
1.44 0.41%
1.110 0.31%
1.220 0.25%

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 4 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

II. MEASURE

1. DC Voltage# 1mV to 100mV 0.5% to 0.01% Using 6½ DMM


100mV t 10V 0.01% to 0.06%
10V to 1000V 0.06% to 0.02%

2. DC Current# 1µA to 10mA 3.0%to 0.08% Using 6½ DMM


10mA to 3A 0.08% to 0.16%
3A to 10 A 0.16% to 0.2%

3. AC Voltage# 50 Hz / 60Hz Using 6½ DMM


10mV to 10V 0.53% to 0.12%
10V to 1000V 0.12% to 0.18%

4. AC Current# 50 Hz / 60Hz Using 6½ DMM & Current


50µA to 100mA 0.9% to 0.5% Transformer
10mA to 10A 0.5% to 0.25%

5. Resistance# 100µΩ to 1Ω 0.6% to 0.06% Using Microohm meter &


(4 Wire) 1Ω to 10Ω 0.06% to 0.05% Low Resistance Standard

6. Resistance# 10Ω to 1MΩ 0.06% to 0.13% Using 6½ DMM & Micro


(2 Wire) 1MΩ to 100MΩ 0.13% to 0.9% ohm meter
100MΩ to 1GΩ 0.9%to 2.36%

7. AC Resistance# 1 kHz Using LCR Meter


1Ω to 100kΩ 0.23%

8. Frequency# 10Hz to 1MHz 0.06% Using 6½ DMM

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 5 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

9. Capacitance# 1kHz Using LCR Meter


1nF to 1µF 0.2%

10kHz
1nFto1µF 0.2%

10. Inductance# 1 kHz Using LCR meter


100µH to 1H 0.45% to 0.8%

11. Power/Energy# 50Hz 1.0% Using Power /Energy Meter


(1Ф. 3Ф.) UPF to 0.5 PF Accuchek & Power source
(Lag/Lead)
Voltage : 240V/63.5V
Current: 1A/5A

12. Timer / Stop Watch# 100 ms 0.003s Using Digital Timer


(Digital/Analog) 1s 0.003s
60s 0.03s
600 s 0.27s
1800 s 0.55s
3600 s (1 hour) 1.3s
7200 s (2 hour) 2.3s
14400 s (4 hour) 4.3s
86400 ( 24 hour) 20.5s

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 6 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

13. Temperature Simulation# Using


(Temperature source, Temperature Indicator
Indicator/Controller/
Recorder)
RTD 200oC to 800oC 0.42 oC
‘T’ Type Thermocouple (-)200oC to 400oC 0.60 oC
‘S’ Type Thermocouple 50oC to 1700oC 0.63 oC
‘J’ Type Thermocouple (-)200oC to 1200oC 0.61 oC
‘N’ Type Thermocouple (-)200oC to 1300oC 0.65 oC
‘K’ Type Thermocouple 50oC to 1300oC 0.66 oC
‘R’ Type Thermocouple 600oC to 1700oC 0.63 oC
‘B’ Type Thermocouple 600oC to 1700oC 0.62 oC
‘E’ Type Thermocouple 100oC to 600oC 0.61 oC

14. AC High Current# 50 Hz Using 6½ DMM & Current


10A to 1000A 1.3% to 0.65% Transformer

15. DC High Current# 10A to 750A 1.0% Using 6½ DMM &


900A/75mV Shunt

16. AC High Voltage# 50 kHz Using HV Probe with DMM


1kV to 28 kV 2.4%

17. DC High Voltage# 1kV to 37 kV 1.7% Using HV Probe with DMM

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 7 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

MECHANICAL CALIBRATION

I. DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)

1. Vernier/Dial/ Digimatic
Caliper$
L.C.: 0.01 mm Upto 300mm 8.0 µm Using Slip Gauge Grade ‘0’,
Upto 1000mm 14.70 µm Long Gauge Block &
L.C.: 0.02 mm Upto 2000mm 49.0 µm Caliper Checker

2. External Micrometer$
L.C.: 0.001 mm Upto 100mm 1.50 µm Using Slip Gauge Grade ‘0’
100 to 150mm 2.5 µm & Long Gauge Block
150 to 300mm 3.8 µm

3. Internal Micrometer$
L.C.: 0.001 mm Upto 300mm 4 µm Using Slip Gauge Grade ‘0’
& Accessories Set

4. Depth Micrometer$
L.C.: 0.01 mm Upto 300mm 7 µm Using Slip Gauge Grade ‘0’,
Long Gauge Block &
Caliper Checker

5. Depth Caliper / Depth


Gauge$
L.C.: 0.01 mm Upto 300mm 7.5 µm Using Slip Gauge Grade ‘0’,
Long Gauge Block &
Caliper Checker

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 8 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

6. Height Gauge$
L.C.: 0.001 mm Upto 600mm 6.0 µm Using Slip Gauge Grade ‘0’
L.C.: 0.01 mm Upto 1000mm 12.8 µm & Caliper Checker

7. Dial Gauge$
L.C.: 0.001 mm Upto 100mm 3.0 µm Using Single Axis M/c

8. Dial Test Indicator$


L.C.: 0.001 mm 0 to 1.0mm 3.0 µm Using Single Axis M/c

9. Dial Bore Gauge$


L.C.: 0.001 mm Upto 1mm Travel 3.0 µm Using Single Axis M/c

10. Plain Plug Gauge$


L.C.: 0.001 mm Upto 100mm 3.0 µm Using Single Axis M/c
100 mm to 185mm 3.5 µm

11. Plain Ring Gauge$ 3mm to 100mm 1.8 µm Using Single Axis M/c
100mm to 280mm 4.6 µm

12. Scale & Tape Up to 1000mm 60 µm Using Slip & Long Gauge
Calibration Unit$ Blocks

13. Radius Gauge$ Up to 100mm 96 µm Using Profile Projector

14. Snap Gauge$ 1mm to 150mm 2.2 µm Using Slip Gauge Grade 0

15. Measuring Pin Gauge$ 0.5mm to 20mm 1.5 µm Using Single Axis M/c

16. Three Wire Unit$ 0.17mm to 4 mm 1.5 µm Using Single Axis M/c

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 9 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

17. Feeler Gauge/ Up to 1mm 1.5 µm Using Single Axis M/c


1.6 µm
Foils$ 1.5 mm

18. Dial Thickness Gauge$


L.C.: 0.001 mm Up to 50mm 1 µm Using Slip Gauge Grade 0

19. Thread Plug Gauge$ Up to 100mm 5.6 µm Using Three Wire set,
Micrometer

20. Thread Pitch Gauge$ 0.40 to 7.0 mm 7.0 µm Using Profile Projector
3.0 min. of arc

21. Length Gauge / Up to 100mm 2.0 µm Using Single Axis M/c


Setting Master$ 100 mm to 280mm 2.6 µm

22. Steel Scale$


LC: 1mm Upto 1000mm 154√L µm Using Standard Scale with
L in meter DRO

23. Measuring Tape$


LC: 1mm Up to 50 mtr. 154√L µm, Where L is in Using Standard Scale with
meter DRO

24. Bevel Protractor/ 0° - 90° - 0o 5 min. of arc Using Slip Gauge Grade 0,
Combination$ Sine Bar & Master Cylinder

25. Electronic Probe$ Up to 25 mm 0.6 µm Using Slip Gauge Grade 0

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 10 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

26. Dial Calibration Tester$


LC: 0.001 mm Up to 25mm 1.0 µm Using Slip Gauge Grade 0
& Electronic Probe

27. Coating Thickness Up to 1500 µm 5 µm Using Standard Foils


Gauge$

28. Angle Plate Up to 300mm 10 µm Using Slip Gauge Grade ‘0’,


Squarness$ Master Cylinder & Dial Test
Indicator

29. Electronic Level$ +/-2mm/mtr 3.7 µm/mtr Using Electronic Level


Spirit Level (Sensitivity 0.001 Comparison Method
mm/mtr)
+/-200mm/mtr 7.0 µm/mtr
(Sensitivity 0.01
mm/mtr)
+/-100mm/mtr 12.0 µm/mtr
(Sensitivity 0.02
mm/mtr)

30. Test Sieves$ 32 µm to 4000 µm 6.2 µm Using Profile Projector/


(Aperture Size) 4mm to 50mm 20 µm Digital Caliper
50mm to 125mm 30 µm

31. Thread Ring Gauge$ 4mm to 100mm 3.4 µm Using Single Axis M/c.
(Effective Dia Only)

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 11 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

32. Sine Bar Flatness Up to 300 mm 3.80 µm Using Elect.Probe & Slip
/ Parallelism / 3.80 µm Gauge Grade 0
Central Distance$ 4.00 µm
Setting Angle$ 6.2 sec. Using Surface Plate, Angle
Gauge, Electronic Probe &
Slip Gauge Grade 0

33. Straight Edge Up to 1000mm 3.4 µm Using Surface Plate & Slip
Straightness$ Gauge Grade 0

34. Engineering Up to 500 mm Using Dial Test Indicator &


Square/Right Angle$ Surface Plate
Flatness 5.7 µm
Perpendicularity Using Slip Gauge Grade 0
5.0 µm & Master Cylinder

35. Test Mandrel$ Up to 300 mm 2.4 µm Using Sine Centre & Dial
Variation in diameter Test Indicator
Lack of trueness of
rotation 16 µm
(between centers)
Lack of trueness of
rotation
(between centers)

36. Magnetic V Block$ Up to 150mm 5.7 µm Using Slip Gauge Set, Dial
Flatness Test Indicator & Surface
Parallelism 8.3 µm Plate

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 12 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

Parallelism 5.7 µm Using Slip Gauge Set, Dial


Test Indicator, Test Mandrel
& Surface Plate

Symmetricity Error 8.3 µm Using Slip Gauge Set, Dial


Test Indicator, Test Mandrel
& Surface Plate

Perpendicularity 6.0 µm Using Slip Gauge Grade 0,


Master Cylinder & Surface
Plate

37. Standard Steel Ball$ 4 mm to 65 mm 2.6 µm Using Digital Micrometer

38. Profile Projector# 0-300mm (X,Y Axis) 3.9 µm Using Slip Gauge & Long
Gauge Block
Linear Dimension
Magnification 50 x 1.00% Using Slip Gauge Grade 0
& Digital Caliper/Glass
Scale

Angle Measurement 0-360° 75 sec of arc Using Angle Gauges

39. Universal Length Up to 100 mm 1.0 µm Using Slip Gauge Grade 0


Measuring Machine # Resolution: 0.0001

40. Surface Plate# 6000mm x 4000mm 1.3√(L+W)/125 Using Electronic Level


where L & W are in mm

41. Comparator Stand# 300 mm x 300 mm 1.2√(L+W)/150 Using Electronic Level


Where L & W are in mm

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 13 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

42. Bench Centre# Up to 300mm 10.0 µm Using Mandrel & Dial Test
Parallelism Indicator
Coaxiality 10.0 µm
43. Sine Centre# Up to 300mm Using Test Mandrel & Dial
Parallelism of working 4.0 µm Test Indicator, Electronic
surface Probe & Surface Plate,
Coaxiality of Dead 4.0 µm Electronic Probe, Slip
center Gauge Grade 0 &Angle
Central Distance 3.8 µm Gauge

II. WEIGHTS

1. Mass $ 1 mg 0.01mg Using E2 Class Weight &


(Weights) 2 mg 0.010 mg Mass Comparator
5 mg 0.010 mg (Readability 0.01 mg) As
10 mg 0.010 mg per OIML R-111 Calibration
20 mg 0.010 mg of Class F2 accuracy and
50 mg 0.010 mg coarser
100 mg 0.010 mg
200 mg 0.010 mg
500 mg 0.013mg
1g 0.015 mg
2g 0.02 mg
5g 0.02mg
10 g 0.023mg
20 g 0.042mg

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 14 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

50 g 0.10mg Using E2 Class Weight &


100 g 0.15mg Mass Comparator
200 g 0.24mg (Readability 0.01 mg) As
per OIML R-111 Calibration
of Class F2 accuracy and
coarser

500 g 2mg Using F1 Class Weight &


1 kg 10mg Mass Comparator
(Readability 0.001 g) As per
OIML R-111 Calibration of
Class F2 accuracy and
coarser

2 kg 13mg Using F1 Class Weight &


5 kg 13mg Mass Comparator
(Readability 0.01 g) As per
OIML R-111 Calibration of
Class F2 accuracy and
coarser

10 kg 89mg Using F1 Class Weight &


Mass Comparator
(Readability 0.1 g) As per
OIML R-111 Calibration of
Class M1 accuracy and
coarser

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 15 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

20 kg 1.07g Using M1 Class Weight &


50 kg 1.27g Mass Comparator
100 kg 1.27 (Readability 0.001 k g) As
per OIML R-111 Calibration
of Class M2 accuracy and
coarser

III. VOLUME

1. Micro Pipette$ 10 µl to 100 µl 0.06 µl Using E2 Class Weight &


100 µl to 1000 µl 0.10 µl Precision Weighing Balance
( Readability
0.01mg & 0.1 mg) and
Distilled Water as per ISO
8655

2. Glassware$ 1 ml to 10 ml 0.5 µl Using Standard weight &


(Measuring Cylinder / 10 ml to 100 ml 8.2 µl weighing balance
Jar, Volumetric Flask, 100 ml to 200 ml 0.04 ml (readability 0.1mg)
Beaker, Glass Pipette, 200 ml to 500 ml 0.14 ml Readability: 0.001g
Dispenser & Burette) 500 ml to 1 Ltr. 0.2 ml Readability: 0.01g
1 Ltr to 2 Ltr. 0.4 ml (Readability: 0.1g) with
2 Ltr. To 5 Ltr. 0.6 ml distilled water by
5 Ltr to 10 Ltr. 1.1 ml gravimetric method as per
ISO 4787

IV. WEIGHING SCALE AND BALANCE

1. Spring Balance$
L.C. 0.1 g Up to 100 kg 60g Using F1 & M1 Class

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 16 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

weight Comparison Method

2. Weighing Balance#
Readability 0.01 mg & Up to 30 g 0.06mg Using E 2 Class Weight As
Coarser Per OIML R76-1
Readability 0.1 mg & >30g to 220g 0.14mg
Coarser
Readability 1 mg & >220g to 1000g 0.006g Using E2 class weight &
Coarser F1 Class Weight As Per
OIML R76-1

Readability 0.01 g & >1kg to 6kg 0.01g Using F1 Class Weight As


Coarser Per OIML R76-1

Readability 0.1 g & >6kg to 12 kg 0.06g


Coarser
Readability 1 g & >12 kg to 100 kg 0.8g Using M1 Class Weight As
Coarser Per OIML R76-1

V. DENSITY AND VISCOSITY

1. Density / 0.6 g/ml to 1.8 g/ml 0.0015g/ml Using Hydrometer &


Hydrometers / Compatible Liquid
Lactometers$ Comparison Method
IS 3104-1 & 2

III. PRESSURE INDICATING DEVICES

1. Vacuum Gauge, (-)0.95 bar to 0bar 0.0036bar Using Digital Pressure


Transmitters Digital / Gauge
Analog#

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 17 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

2. Barometric Pressure 200 to 1050 mbar 1.32 mbar Using Pressure Calibrator
Gauge#

3. Pneumatic Pressure (-) 1300 to 0 pascal 2.0 Pa Using Pressure Calibrator &
Gauge (Digital / 0 to 2000 pascal 2.0 Pa Diff. Pressure Gauge by
Analog Manometer, 20mbar to 100 mbar 0.1 mbar comparison method
Diff. Gauge,
Transmitters,
Magnehalic Gauge#

4. Pressure Gauge , 0 to 3 bar 0.0036 bar Using Digital Pressure


Transmitters 0 to 30 bar 0.036 bar Gauge & Pressure
Digital/Analog# Comparator by comparison
Pneumatic method

5. Pressure Gauge 0 to 700 bar 1.15 bar Using Digital Pressure


(Analog/Digital)- Gauge & Pressure
Hydraulic# Comparator by comparison
method

6. Hydraulic Pressure 1 to 41.1 kg/cm2 0.052% rdg Using Dead Weight Tester
Digital / Analog 40 to 800 kg/cm2
Pressure Gauge$

IV. ACOUSTICS

1 Sound Level Meter# 94 dB 0.55dB Using Sound Calibrator


114 Db

V. DUROMETER

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 18 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

1. Rubber Hardness 0 to 100 Shore A 0.7 Shore A Using Weighing Balance,


Tester# 0 to 100 Shore D 0.7 Shore D Readability : 1g

VI. ACCELERATION AND SPEED

1. Speed 10 to 40 rpm 5% Using RPM Source by Direct


Contact Type RPM 40 to 100 rpm 0.35% Method
/Tachometer / 100 to 1000 rpm 0.25%
Stroboscope , 1000 to 9950 rpm 0.09%
Centrifuge#

2. Non-Contact Type 10 to 40 rpm 5.0 % Using Tachometer Calibrator


RPM 40 to 1000 rpm 1.7 % by Direct / Comparison
Stroboscope/Pulse 1000 to 10000 rpm 0.09 % Method
Engine Tachometer/ 10000 to 99000 rpm 0.06 %
Digital Tachometer,
Centrifuge#

VII. TORQUE GENERATING DEVICES

1. Torque Wrench# Type II 4 to 2000 Nm 1.3% Using Torque Transducer with


Type I: Class Type I 4 o 500 Nm 1.3% indicator
B,C,D,E
Type II: Class
A,B,D,E

VIII. MOBILE FORCE MEASURING SYSTEM

1. Push Pull Gauge$ 1N to 100N 0.1N Using Newtonian Weight &

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 19 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

100N to 900N 1.1N Frame Fixture & Hangers

THERMAL CALIBRATION

1. Temperature (-)196oC to -40oC 0.32oC Using RTD


Transmitter RTD’s (Pt-100) 4 wire Digital
Thermocouples with & Thermometer & Liquid
without Nitrogen Bath
(Controller/Indicator/D
ata Logger/Recorder),
Temperature Gauge,
Glass Thermometer &

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 20 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

Digital Thermometer#

2. Temperature (-)40oC to 50oC 0.20°C Using RTD (Pt-100) 4 wire


Transmitter RTD’s Digital Thermometer and Low
Thermocouples with & Temp Bath
without
(Controller/Indicator/D
ata Logger/Recorder)
Temperature Gauge,
Glass Thermometer &
Digital Thermometer#

3. Temperature >50oC to 200oC 0.20°C Using RTD (Pt-100) 4 wire


Transmitter RTD’s Digital Thermometer and Oil
Thermocouples with & Temperature Bath
without
(Controller/Indicator/D
ata Logger/Recorder),
Temperature Gauge,
Glass Thermometer &
Digital Thermometer#

4. Temperature 200oC to 300oC 0.80°C Using RTD (Pt-100) 4 wire


Transmitter RTD’s Digital Thermometer and
Thermocouples with & Dry Block Bath
without
(Controller/Indicator/
Data
Logger/Recorder/),
Temperature Gauge,
Glass Thermometer &
Digital Thermometer#

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 21 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

5. Thermocouples, >300oC to 1200oC 1.83°C Using Thermocouple wire


RTD’sTemperature Digital Thermometer and
Transmitter with & Dry Block Bath
without
(Controller/Indicator/ /
Recorder), Data
Logger & Digital
Thermometer#

6. Temperature (-)196oC to -40oC 0.32°C Using RTD (Pt-100) 4 wire,


Indicators of Freezers, Digital Thermometer
Oven, Environment
Chamber, Incubator,
BOD Incubator, Liquid
Bath / Dry Block
Furnaces#

7. Temperature Indicator (-)40oC to 250oC 0.2°C Using RTD (Pt-100) 4 wire


of Freezers, Oven, Digital Thermometer and
Environment Chamber, Dry Block Bath
Incubator, BOD
Incubator, Liquid Bath /
Dry Block Furnaces#

8. Temperature Indicators 250oC to 1200oC 1.8°C Using S Type


of – (Muffle Furnace, Thermocouple & Digital
Dry Block Furnace) # Thermometer

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 22 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

9. Radiation Pyrometer 50oC to 500oC 3.3°C Using Radiation


(Non-Contact Type) # Pyrometer, Digital
Thermometer & Black Bod
Source)

10. Freezers, Oven, Cold (-)80oC to 50oC 0.5°C Using Multi-point Data
Chamber, Environment 50oC to 250oC 1.3°C Logger with RTD (Pt-100)
Chamber# Sensors

11. Digital & Analog 10% - 95% RH 0.7%RH Using Std. RH Sensor with
Hygrometer, RH @ Indicator & RH Calibrator
Sensors / Transmitters approx. 25 oC
with
controller/Indicator/Re
corder/Data Logger#

II. SPECIFIC HEAT AND HUMIDITY

1. (Humidity 10% - 95% RH 1.95%RH Using STD. RH Sensor


Calibrator /Generator with Indicator
Chamber) #

2. Freezers, Oven, Cold -80oC to 50oC 0.5°C Using Multi-point Data


Chamber, Environment 50oC to 250oC 1.3°C Logger with RTD (Pt-100)
Chamber, Centrifuged Sensors
Chamber, Incubator &
BOD Incubator♣

3. Industrial Furnaces / 4.2°C Using Multi-point Data


Spatial Thermal 200oC to 1200oC Logger with N-Type

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 23 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

Mapping♣ Sensors

4. Environment Chamber 15% - 95% RH 1.95%RH Using Std RH Sensors /


/ Humidity Chamber♣ @ Transmitters with Data
approx. 25 oC Logger

FLUID FLOW CALIBRATION

1. Rotameter / Flow 1 LPM to 5 LPM 1.40% rdg Using Fluke Flow Meter
Meters (Flow >5 LPM to 50 LPM 0.58% rdg consist of Molbloc sensor
Rates) # by Comparision method

2. Volume Flow 0.8m3/hr to 80 m3/hr 1.98 % rdg Using Transducer based


Rate >80m3/hr to 200m3/hr 1.18% rdg portable Ultrasonic Flow
(Medium of >200m3/hr to 1450 m3/hr 0.41 % rdg Meter
Calibration
Liquid-Water) ♣
* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Belz Calibration Laboratory, Belz Instruments Pvt. Ltd., 5L-123, NIT,
Faridabad, Haryana

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2733 (In lieu of C-0100, C-0158,C-0407, Page 24 of 24
C-1424)

Validity 28.06.2018 to 27.06.2020 Last Amended on 25.07.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

$
Only in Permanent Laboratory

Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the
prevailing actual environmental conditions and master equipment used.

Dheeraj Chawla Avijit Das


Convenor Program Director

You might also like