Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Sl. Quantity Measured / Instrument Range/Frequency Calibration Measurement Capability ( ) Remarks

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 4

Laboratory RK Technologies, F-26, 2nd Floor, New Ashoka Vihar, XLO Point,

MIDC Ambad, Nashik, Maharashtra


Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2497 Page 1 of 4
Validity 26.12.2017 to 25.12.2019 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

ELECTRO-TECHNICAL CALIBRATION

I. SOURCE

1. DC Voltage # 1 m V to 10V 7.6 % to 0.16% Using 5½ Digit Zeal


10 V to 1000 V 0.16 % MFC By Direct Method

2. DC Current# 1mA to 1A 0.73 to 0.25% Using 5½ Digit Zeal


1A to 10 A MFC By Direct Method

10 to 100 A 1.34 to 1.11% Using Current Coil


100 to 900 A 1.11 to 0.67%

3. AC Voltage# 50 Hz
5mV to 1 V 1.90% to 0.79% Using 5½ Digit Zeal
1V to 1000 V 0.79% to 0.23% MFC By Direct Method

4. AC Current# 50 Hz Using 5½ Digit Zeal


1 mA to 1 A 0.8% to 1.06% MFC By Direct Method
1A to 10 A 1.06% to 0.32%

10 A to 100 A 1.25% Using Current Coil


100 A to 900 A 1.25%

5. DC Resistance# 1 Ω to 1000 Ω 1.01 % to 0.3% Using Decade


(2 Wire & 4 Wire) 1 kΩ to 1MΩ 0.3% to 0.58% Resistance Zeal By
1MΩ to 1000 MΩ 0.58% to 2.37% Direct Method

6. Frequency# 45 Hz to 1000 Hz 0.55 % to 0.06% Using 5½ Digit Zeal


MFC By Direct Method

Rajeshwar Kumar Avijit Das


Convenor Program Director
Laboratory RK Technologies, F-26, 2nd Floor, New Ashoka Vihar, XLO Point,
MIDC Ambad, Nashik, Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2497 Page 2 of 4
Validity 26.12.2017 to 25.12.2019 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

7. Temperature
Simulation#

RTD - PT-100 (-)200°C to 800°C 1.0°C Using Universal


J Type (-)100°C to 750°C 1.1°C Calibrator By Direct
K Type (-)100°C to 1300°C 1.27°C Method
T Type (-)100°C to 350°C 1.1°C
R Type 300°C to 1750°C 1.77°C
S Type 300°C to 1750°C 1.77°C

II. MEASURE

1. DC Current# 2 mA to 20 mA 0.7% to 0.12% Using Universal


Calibrator By Direct
Method

2. Time# 2 s to 3600 s 3.7 s Using Digital Stop


(Stop Watch) Watch By Comparison
Method

Rajeshwar Kumar Avijit Das


Convenor Program Director
Laboratory RK Technologies, F-26, 2nd Floor, New Ashoka Vihar, XLO Point,
MIDC Ambad, Nashik, Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2497 Page 3 of 4
Validity 26.12.2017 to 25.12.2019 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

MECHANICAL CALIBRATION

I. PRESSURE INDICATING DEVICES


1. Hydraulic Pressure- 0 to 700 bar 0.29 bar Using Digital Pressure
Analog /Digital Gauge ,Hydraulic pump
Pressure Based on DKD-R-6-1
Gauge#

2. Pneumatic Pressure- 0 to 30 bar 0.08 bar Using Digital Pressure


Analog /Digital Gauge ,Pneumatic pump
Pressure Based on DKD-R-6-1
Gauge#

3. Pneumatic Pressure- 0 to 1000m bar 1.17 mbar Using Digital Pressure


Analogue/ Digital Calibrator using low
Low Pressure pressure Polltech air
Gauge, Digital pump ,
Manometer# Based on DKD-R-6-1

Rajeshwar Kumar Avijit Das


Convenor Program Director
Laboratory RK Technologies, F-26, 2nd Floor, New Ashoka Vihar, XLO Point,
MIDC Ambad, Nashik, Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2497 Page 4 of 4
Validity 26.12.2017 to 25.12.2019 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

THERMAL CALIBRATION

I. TEMPERATURE

1. RTD/PT-100/ 50°C to 200 °C 1.20 °C Using 4 Wire RTD & Read


Thermocouples unit Universal Calibrator
With or Without By Comparison Method
Indicator,
Temperature
Gauge # 200°C to 1000°C 2.68°C Using R Type
Thermocouple & Read
unit Universal Calibrator
By Comparison Method

2. Indicator of Oven / 50°C to 200 °C 1.20 °C Using 4 Wire RTD & Read
Furnace/Baths at unit Universal Calibrator
Single Specified By Comparison Method
Position #
200°C to 1000°C 2.68°C Using R Type
Thermocouple & Read
unit Universal Calibrator
By Comparison Method
* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%
#
The laboratory is also capable for site calibration however; the uncertainty at site depends on the
prevailing actual environmental conditions and master equipment used.

Rajeshwar Kumar Avijit Das


Convenor Program Director

You might also like